Fukuda FL-296 Series, FL-286 Installation Instructions Manual

10527-T-800E-00
http://www.fukuda-jp.com
COMPACT AIR LEAK TESTER
DESIGN DOCUMENT
Contents
MODEL ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 6 SPECIFICATIONS ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 7
DO YOU HAVE EVERYTHING? ・・・・・・・・・・・・・・・・・・・・・・・ 8 OPTION ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 9 INSTALLATION INSTRUCTIONS ・・・・・・・・・・・・・・・・・・・・ 14-15 AIR PIPE CONFIGURATIONS ・・・・・・・・・・・・・・・・・・・・・・ 16-17
ELECTRIC WIRING CONFIGURATIONS ・・・・・・・・・・・・・・・・・ 18-19
LINE CONNECTIONS ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 20-35
RS-232C CONNECTION ・・・・・・・・・・・・・・・・・・・・・・・・・・ 60-63 PRINTER CONNECTIONS ・・・・・・・・・・・・・・・・・・・・・・・・ 65
EXTERNAL DIMENSIONS ・・・・・・・・・・・・・・・・・・・・・・・・ 72-74
Declaration
・This design documents is outline specification for system design which is extracted from operation manual (10527-T-001E-05). The above contents page numbers is mentioned the page numbers of operation manual. For about details of operation and testing procedure, please refer to operation manual.
・The contents of this manual may be modified without notice due to function and performance improvements.
・This manual has been written with care to provide the user with helpful and useful information. Please
contact your local FUKUDA distributor if there are any inaccuracies or questions of concern.
・Do not use this tester for any purpose other than testing air leaks. ・For best operating results and optimum efficiency, read this instruction manual thoroughly to understand the tester’s many functions prior to operation. ・Reprinting or duplicating this manual partially or completely is prohibited unless permitted by FUKUDA CO., LTD.
Trademark
Company names and product names being used in this manual are the registered trademarks of their respective holders.
6
OUT LINE TESTER MODEL
OUT LINE TESTER MODEL
MODEL
FL-
2
MODEL
Symbol Permissive Leak rangeApplication Application
8
±0.01~9.99 kPa
Middle Leak Detection
9
±1~999 Pa
Small Leak Detection
② CALIBRATOR (BUILT IN)
Symbol With/Without Calibrator Remarks
6 Without Calibrator 6A 1.0ml Calibrator 6B 0.1ml Calibrator
③ SET PRESSURE RANGE
Symbol
TEST PRESSURE RANGE SET PRESSURE RANGE DISPLAY PRESSURE RANGE
UL
-10.0~10.0kPa -10.0~10.0kPa -10.0~10.0kPa
L
-100~200kPa -100~200kPa -100~200kPa
H
-100~900kPa -100~900kPa -100~900kPa
POWER SUPPLY
Symbol POWER SUPPLY
0
AC100V ±1050/60Hz
2
AC90~110V AC200~240V 50/60Hz
INTERNAL OPTIONS
Symbol OPTION REMARKS
R RS-232C Output
Select either RS-232C output (R) or printer output (P).
P Printer Output B1 Special Buzzer Sounding
Select either a “Beep” (B1) or a pulsed tone (B2) for notification.
B2 Special Buzzer Sounding
E
Piping modification to correspond with the external exhaust unit
Only for use with a vacuum
V
Manual valve position indication switch
Only at work side
EXTERNAL OPTIONS see page 9.
① ② ③-④ ⑤ ⑥
7
TESTER SPECIFICATIONS
PERFORMANCE SPECIFICATIONS
Testing Method: Differential Pressure Comparison
Leak Pressure RangeMeasurement Accuracy
MODEL
Leak Pressure Measurement
Range
Measurement Sensor Accurac
y
(Including hysteresis)
FL-286
±0.01~9.99kPa ±5%F.S.10.00kPa
FL-296
±1~999Pa ±5%F.S.(1000Pa
Direct Pressure Sensor Accuracy
MODEL Pressure Measurement Range
Measurement Sensor Accurac
y
(Including hysteresis)
UL
-10.0~10.0kPa ±5%F.S.(-10.0kPa、10.0kPa
L
-100~200kPa ±5%F.S.(-100kPa、200kPa
H
-100~900kPa ±5%F.S.(-100kPa、900kPa
FUNCTION SPECIFICATIONS
Timer Setting Range
DLY
CHG
BAL DET
EXH
(Delay) (Charge) (Balance)
(Detection) (Exhaust)
0.0 ~ 999.9sec
0.1 ~ 999.9sec
0.1 ~ 999.9sec
0.1 ~ 999.9sec
0.2 ~ 999.9sec
Group Setting Function
4ch(DLY time, CHG time, BAL time, DET time, EXH time, -NG set, +NG set, -NG set(leak amount), +NG set(leak amout), P.NG value, Volume value, -NG unit (leak amount), +NG unit(leak amount), Volume unit, Leak amout indication unit)
Supplemental Functions
BUBL(Bubble),CAL(Calibration), Cylinder Control
Test circuit internal volume
8 ml(when rear panel stop valve is closed
Display Digital indication in four digi
t
External Input/Output
1. Contact input/output for control 38
P
2. Option input/output
Output for printer(Centronics-compliant output system) RS-232C input/output with PC
GENERAL SPECIFICATIONS Operation Temperature Range
0~40
Operation Humidity Range
35~85%RH(without condensation
Power Source
A
C100V±1050/60Hz
AC90~110V AC200~240V 50/60Hz Power Consumption 150VA Voltage enduranc
e
A
C1000V 1 minut
e
Insulating Resistance More than 50MΩ at DC500V Leak Current Less than 1mA at AC100V Pressure enduranc
e
UL Range:-50~+50kPa、L Range:-100~+600kPa、H Range:-0.1~+1.5MPa
Dimensions Function
W172.9 mm×H272.8 mm×D386 mm(Without external option
Weight
Power Source100V:About 9.5 kg
Power Source 200V:About 10.5 kg
Air Specifications
Please supply clean , stable , and dry air
Recommended Condition: Based on JISB8392-1:2000
Compressed Air Quality Class1,3,1
Class
A
rticle Referenc
e
1 Maximum Particle Diameter
0.1μm
3
Minimum Pressure Dew-point
-20
1 Maximum Oil Conten
t
0.01mg/
m
3
8
OUTLINE DO YOU HAVE EVERYTHING ?
DO YOU HAVE EVERYTHING?
Confirm the following points prior to operating the tester. If there are any problems such as incorrect specifications, missing accessory parts or damages with the delivered items, please contact your FUKUDA agent. Please compare the model numbers described below with the number indicated on the tester’s nameplate. Please specify the SER (serial number) when contacting your FUKUDA agent.
TESTER STRUCTURE
【FL-2□□□-0 FRONT】 【FL-2□□□-2 FRONT
【FL-2□□□-0 REAR】 【FL-2□□□-2 REAR
STANDARD ACCESSORIES
A
CCESSORIES
NUMBER
REMARKS
AC POWER CODE 1.5m 1
A
s the attached AC power cord is of Japanese domestic specification (rating 125V), if it does not fit to the specifications of the using area or place, kindly please prepare the suitable one by yourself.
A
DJUSTMENT SCREW DRIVER
1
MOUNTING BRACKET
1
FIXING SCREW M4×8L 4 OPERATION MANUAL
1
TEST DATA SHEET
1
Unit Stickers
1
9
OUTLINE OPTIONS
OPTION
EXTERNAL OPTIONS
MODEL COMMON OPTION REMARKS
C1
Manual Calibrator0.1mlF.S. Externa
Mounted in the rear of the work port manual valve. One of C1 or C3 is chosen.
C3
Manual Calibrator1.0mlF.S. External
FR53 Filter/Regulator Unit
Positive Pressure Standard Spec. Pressure Setting Range:0.02 ~ 0.2MPa
FR54 Filter/Regulator Unit
Positive Pressure Precision Spec Pressure Setting Range:0.005 ~ 0.2MPa
FR55 Filter/Regulator Unit
Positive Pressure Precision Spec. High Performance Filter Pressure Setting Range:0.005 ~ 0.2MPa
FR56 Filter/Regulator Unit
Positive Pressure Standard Spec. Pressure Setting Range:0.05 ~ 0.85MPa
FR57 Filter/Regulator Unit
Positive Pressure Precision Spec. Pressure Setting Range:0.005 ~ 0.8MPa
FR58 Filter/Regulator Unit
Positive Pressure Precision Spec. High Performance Filter Pressure Setting Range:0.005 ~ 0.8MPa
FR59 Filter/Regulator Unit
Positive Pressure Standard Spec. Pressure Setting Range:0.1 ~ 1.6MPa
FR61 Filter/Regulator Unit
Negative Pressure Standard Spec. Pressure Setting Range:-100 ~ -1.3kPa
FR62 Filter/Regulator Unit Negative Precision Spec. P-200
FE-20 External Exhaust Unit For Positive Pressure
FE-20C External Exhaust Unit For Positive Pressure with Drain
FE-20V External Exhaust Unit For Negative Pressure
PS-01 Pressure Switch
Small pressure switch for negative pressure, Accuracy: ±3% F.S. Pressure Setting Range:-101 ~ 0kPa (5~40)
PS-02 Pressure Switch
Small (low) pressure switch for positive pressure, Accuracy: ±3% F.S. Pressure Setting Range:0 ~ 100kPa (5~40℃)
PS-03 Pressure Switch
Small pressure switch for positive pressure, Accuracy: ±3% F.S. Pressure Setting Range:0 ~ 970kPa (5~40℃)
FP-100 Thermal Printer Print out measured data
CB-14 Manual Operation Box
One Circuit each for start and reset switch 4-group switching function.
P-200 Pilot Regulator
R5 Regulator for small pressure
Pressure Setting Range:0.5 ~ 10kPa
The high-performance filter consumes air at a rate of 0.3 m3/h.
INSTALLATION INSTRUCTIONS
INSTALLATION INSTRUCTIONS
Air leak testers detect leaks by reading out precise pressure (air condition) alterations.
Thus, it requires different handling instructions from other standard precision equipment.
WARNING
Tester must be in a stable position. Be sure to install the tester on a level surface.
Vibrations or earthquakes may cause the tester to fall and cause injury to operators. Do notplace anything on top of the tester. Do not place anything on top of the tester. Falling objects may
result in injury.
CAUTION
Do not install the tester in the following locations:
High moisture or dusty areas. Areas in which oil and water can cause contamination. Areas directly exposed to sunlight. Outdoors
May result in fire or electric shock.
Areas with obstructions or places in which the testers fanning
system may be blocked. The tester may overheat and may cause for fires. Moving the Tester: Be sure to shut off the power, unplug cords, cut off the
pneumatic source and remove piping. Be sure to carry the tester in a stable manner. (Do not carry the tester by the manual valve handle.) May cause damage to the tester or to its casing and may result in injury
.
OPERATION ENVIRONMENT
FL-286,296
14
FL-286,296
To keep the stable measurement condition, install the air leak tester where it will not receive any influences from temperature or humidity changes, such as at the center of the facility. Do not in stall the te ster n ear wi ndows or an area that is directly exposed to the sunlight. The operation temperature and humidity is as follows.
Ambient Temperature: 0~40
Ambient Humidity: 35~85%RH with no condensation.
In order to attain best results and accuracy, it is re commended that the operation
temperature and environment humidity condition is 23 ±5, 55±10%RH.
Rapid temperature change adds the noise on the measurement and specified
performance cannot be obtained.
Transferring the tester rapidly from high to low temperature locations and/or
humidity points may cause condensation. When such cases cannot be avoided, leave the tester in the environment for at least an hour in order for it to adjust to the surrounding temperature.
Environment Temperature change is the enemy of leak test. Avoid the direct sun shine or the wind from outside on work, master and measurement piping. Also be careful against the wind from air conditioner (heating/cooling) duct or the entrance.
Warning
FL-286,29
6
1
5
INSTALLATION INSTRUCTIONS
INSTALLATION LAYOUT
z Use a pair of brackets and the supplied screws.
※Dimensions in front & in the rear differ in case Manual Calibrator is mounted, or not.
INSTALLATION MEASURMENT
Follow the remarks listed below for installation. Leave a sufficient amount of space in the rear for the work connection cables and pressurized air piping. See page 72 and page 74 for external dimensions.
Fixing the Brackets
Example of Front/Rear Fixing Fix the brackets using the screws into the tapped holes on the bottom of the unit.
Front/Rear Mounting Side Mounting
Without calibrator
With calibrator
Both with and without calibrator
Main body
Supplied Screw
Mounting bracket
1
6
AIR PIPE CONFIGURATIONS
Air leak testers are designed to detect a leak by sensing a small change in pressure. For that reason, you must ensure that the air leak tester’s piping differs from normal piping. Please keep this in mind to help obtain accurate and reliable leak testing.
TEST CIRCUIT PIPES
For the test circuit piping, it is necessary to use rigid, leakless pipes that will not change in size and
configuration under applied pressure.
For connection, use a high-pressure type coupler.
Arrange them to minimize thermal effect. Keep them away from heat generating elements such as solenoid valves. Keep the piping arrangement as short as possible. Wrap them with thermal insulation material.
(e.g. Spiral Tubes
Others
Stabilize the piping to avoid unnecessary movements during measurement. Allow “MASTER” and “WORK” piping to provide the same conditions. Do not connect inappropriate items to the testing piping.
Make sure you always use the optional pressure switch which checks the internal pressure of the work, if the equipment is used with a pressure of 10 kPa or lower.
PRIMARY PRESSURE REGULATOR
Adjust to: (approx.) Test pressur
e
100kPa
Pressure Supply Port (AIR SUPPLY) Rc 1/4
Work Port
(WORK)
Rc 1/4
Connect to a work (WORK)
This is a measurement pipe.
Master Port(MASTER) Rc 1/4
Connect to a master.
This is a measurement pipe.
Exhaust Port
(EXH)
Rc 1/4
AIR PIPING CONFIGURATION
17
IN CASE OF WATER/OIL INTRUSION IN WORK
Oil, water or chippings may intrude in works because of the previous process.
The tester may draw in unwanted objects during the exhausting process, and may cause possible break down. Install the external exhaust unit (optional) within the testing air circuit. While the FE-20 optional external exhaust unit is in use, close the exhaust port of the air leak tester with the attached plug.
FE-20 FE-20C FE-20
空圧源 供給エアーにはクリーンなものを使用下さい
エアリークテスターサーにとって、油分・水分は大敵です。 エアィルターはもちろん、必要に応じエアドラ空気清浄器を設備 下さい
ブリケータを使用するは、エアリークテスターへの供給配管と主配管
クランプ装置 シリンダー駆動圧力を調整する減圧弁は、個別に設け下さい シリンダーが動作する圧力変動が、エアリークテスター力供給を変動 させたり、軸のシリンダーのクランプ力を変動させ測定を不安定にする あります
漏れ測定部(ワーク、マスター)の内容積は、 測定感度を上げるを短くしたり、中子
を利用して出来るだけ小さくして下さ
Position close to the Work parts.
主配管
マスター御用意下さい 漏れ測定を行うは、マスターを取り付けて下さい。マスターの容積・形状・接続方法に
り、漏れ測定の精度が変化します。マスターは、ワークと同一形状・同材質・同一の配管・ 同一取付が望ましいのですが、実際の設備では全てを満たすはなかなか出来ません。 マスター設置のポイントを以下に示します。
漏れの無いワー ワーク側とじ条件にする)
ワーク側より小さめの剛体タン
(早く安定になるようにする)
取付位置
ワークの近くで同じ高さにする ワーク側高さを同じにして周囲を囲う
(同じ温度で変動が無いようにする
配管
ワーク配管と一緒にして平面に這わ せる
ワーク側と長さを同じにする
手待ちワークの温度が変
しないように、リークテ ト装置の近くで同じ高さ 場所に保管し下さい
クランプシリンダー
Master
COMPRESSED AIR SOURCE: Water and oil intrusion often cause the pressure sensor in the air leak tester to malfunction. Be sure to provide an air filter, and if necessary, equip an air purifier such as an air dryer. Furthermore, it is important to contrive a bypass air piping system when using lubricator so that lubricating oil will not intrude the tester
CLAMPING EQUIPMENT Regulators to adjust the cylinder driving pressure must be provided separately for each cylinder. This minimizes the possibility of cylinder movement, which destabilizes the pressure supply to the air leak tester. Supplied air pressure fluctuation may interfere with the clamping powers of the other cylinders
Main air source
制御用電磁弁
逆止め
Check Valve
Regulator Lubricator
Sol. Valve for control
Clamp Cylinder
KEEPING THE INTERNAL VOLUME AS SMALL AS POSSIBLE In order to increase test sensitivity, keep the internal volume of the work and master parts as small as possible. Shortening pipes or using a core is ways to decrease internal capacity.
AVOID TEMPERATURE CHANGE In order to avoid temperature interference to work parts waiting to be tested, keep them close to leak tester and at about the same level.
PREPARE THE MASTER PARTS It is recommended to use master (criterion) parts for air leak testing. Leak testing accuracy will change according to the status of the applied master parts, so it would be ideal to use master parts equivalent to the work parts in shape, material, piping and mounting. Maintaining these conditions can be difficult in actual settings. Please refer to the following chart when choosing a master part.
Best Method
A
lternative
Master
Test Parts with no leak (same conditions as the work)
A rigid tank that has a slightly smaller volume than that of the work (to stabilize quickly)
Mounting
Position
Same height from the ground as the work
Near the work
Same height as the work and surround the peripheral Environment with the same temperature interference
conditions as the work parts
Piping
Arrangement
Same conditions and arrangement as the work piping and arrange piping in plane .
Same piping length as the work port (shorter when there is temperature interference)
ELECTRIC WIRING CONFIGURATION
ELECTRIC WIRING PRECAUTIONS
EXAMPLE
(AC 200V/100V step-down transformer is used for an air leak tester with AC 100V spec
Air Leak Tester
MODEL
AC100V AC110V AC200V
FL-2□□□-0 FL-2□□□-2
POWER LINEVOLTAGE
AC100V±10
50/60 Hz
AC90110V
50/60 Hz
AC200240V
50/60 Hz
Power consumption increases if valves and other parts are driven by an external control.
18
AC100V
BEFORE WIRING AIR LEAK TESTER, BE SURE TO TURN OFF THE MAIN POWER SWITCH. (Unplu
g
the tester from the AC power source)
Insulated Transformer
AIR LEAK TESTER with AC100V
Supply line voltage suitable to the power line voltage of the tester.
Select appropriate wires and material according to use, considering voltage and current
requirements for each of the terminal outlets located on the rear panel of the tester.
To avoid noise interference, run the input signal lines separately (pins #1 to #12 of the Terminal Board)
from the output signal lines.
To avoid noise interference, run the input signal lines separately from other power supply lines.
To avoid noise interference, it is recommended to use a shield wired for input signal lines.
Be sure to ground the frame grounding (pins #37 and #38 of Terminal Board).
Use an insulated transformer when the power for the tester is supplied through a step-down
transformer.
AC200V
AC100V L
AC100V N
F.G.
ELECTRIC WIRING CONFIGURATION
Control testing and bubble testing procedures
START Start of air leak test.
Applies a pulse signal of 0.5 sec or longer during DLY sequence. BUBL Start of bubble test Performs a bubble test while a signal is being applied. RESET Stop the tester in case of an emergency.
Apply a break signal to open the air circuit. The RDY sequence
is established after EXH sequence. Test result output
・OK -NG set range < leak reange < +NG set range ・+NG leak reange ≧ +NG set range ・-NG -NG set range ≧ leak reange ・P.NG Test pressure is lower than the set value at the end of
PRES sequence.
・P.NG Test pressure is lower than the set value set up with
P.SW at time of BAL sequence.
Current status output
RDY Standby (Ready Status) ERR Error has transpired.
When the power is OFF, ERR can be used to confirm
ON/OFF status.
・END Testing has ended.
1 The air leak tester’s internal output status: power off. 2 The names, indicated with line(***)imply that the signal is effective when it is open. 19
Controlling clamping cylinders
LMT 1 Input of unclamp signals LMT 2 Input of clamp signals SOL 1 Cylinder Driving valve control signals
In the Internal Operation Mode, the tester cannot operate.
Specifies the Group # (03)
CH1 CH2
Group # selection
0 1 2 3
CH1 CH2
○:Open ●:Short
Controls the exhaust bypass unit
SUPPLIES POWER TO THE AIR LEAK TESTER
Apply power voltage specified by the manufacture
F.G Be sure to ground this terminal.
EXHAUST BYPASS UNIT CONTROL SIGNALS
POWER SUPPLY
GROUP SWITCHING SIGNALS
CYLINDER CONTROL SIGNALS
Pressure Error (P.NG) Verification Signal Input
P.SW Pressure switch signal input. If pressure verification function is not used,
keep the P.SW pin connected to the COM- pin.
PRESSURE VERIFICATION SIGNALS
AIR LEAK TESTER CONTROL SIGNAL
z REMOTE TERMINAL BOAR
FL-
2
□□□-0 FL-2□□□-
2No
Cod
e
Meanin
g
Connection *1 Specification Connection *1 Specification
1
2
3
4
5
6
7
8
9・10
11・12
RESET START BUBL P.SW LMT1 LMT2 CH1 CH2 COM­COM+
Reset signal
Measurement start signal
Bubble test
Pressure verification
Unclamp verification
Clamp verification
Group # switch
Group # switch
Input signal common
Power sup. for pres. switch
Input signal (Photo coupler) Voltage when open DC 12V type Voltage when shorted 5 mA type
COM+Power Supply Output voltage DC 12V type Output current 50 mA min
Input signal (Photo coupler) Voltage when open DC 12V type Voltage when shorted 5 mA type
COM+Power Supply Output voltage DC 12V type Output current 50 mA min
13
14
15
16
17・18
19
20
21
22
23・24
OK
+NG
-NG
P.NG
COM O
1
RDY
ERR
END
COM O
2
OK results output
+NG result output
-NG result output
Pressure malfunction. Output
Output signal common 1
Stand by (Ready) output
Error output
End output
Unconnected
Output signal common2
Output signal (Relay contact) Rated load AC 250V 0.4A DC 30V 0.4A
Output signal (Relay contact) Rated load AC 250V 0.4A DC 30V 0.4A
25・26
27
28
29
30
31
32
SOL1
SOL1
SOL2
SOL2
SOL3
SOL3
Unconnected
Cylinder drive valve output
Pressure valve output
Optional valve output
Output signal (SSR) Output Voltage Power source supply Actual ON current
1.0 Arms Leak current 3 mA max
Output si
g
nal (Relay contact) Rated load AC 250V 1A DC 30V 1A
To accord with external supply power source specs.
33・34
35・36
37・38
AC
AC
F.G.
AC power input
AC power input
Frame Ground
Power source for tester AC100V±10% 50/60Hz (Used also as external valve power source)
Power source for external valve AC100V DC24V (Power source for tester is inlet.)
Loading...
+ 24 hidden pages