These service instructions aim to provide assistance for troubleshooting
activities with the A
LIA ventilator.
It contains drawings, circuit diagrams and parts lists for a better
understanding of how the ventilator works.
Service and maintenance work can be carried out to the peripheral
components of the A
the ventilator can only be carried out by F. S
service partners. Otherwise the entire electronic systems in the A
LIA ventilator. Repairs to the electronic control of
TEPHAN GMBH or authorized
LIA
ventilator require no maintenance.
Every step of a specific procedure should be read through carefully
before starting to service the unit. Always only use the correct tools and
the stated measuring equipment. Failure to follow the instructions and/or
recommendations in these service instructions can result in the device not
working properly or cause damage to the device.
Only use original spare parts by F. S
spare parts list (see chapter
10 p. 75).
TEPHAN GMBH as contained in the
These service instructions do not replace the operating instructions.
Handling the device in any way presumes an exact knowledge and
compliance with the operating instructions. The device is only intended
for the described purpose.
%h Percentage hours Service life of the oxygen sensor in
hours depending on the oxygen
concentration
AIR Medical compressed air
aqua dest. Aqua destilata (lat.) Distilled, demineralized water
bar Unit of measurement for compressed
air
Batt Rechargeable battery Device for storing electrical energy in
the form of chemical energy
cmH2O Unit of measurement for compressed
air = mbar
FiO2 Inspired oxygen concentration
IGR Incremental transducer Push button and knob
PDU Power distribution unit Power Distribution Unit
PEEP Positive end expiratory
Refers to instructions drawing attention to important facts.
The following safety instructions appear at relevant points in the service
instructions and must always be heeded.
Refers to warnings which, if not heeded, can result in malfunctions,
damage or defects in the device, which can possibly also put the patient at
danger.
Warning
1 General information
Caution
Refers to precautions which, if not heeded, can result in damage to the
device and its accessories.
LIA ventilator consists of a basic housing and a front housing. The
The A
front housing also acts as the device cover.
Fig. 1: Basic and front housing: general
2 Mechanical structure
Front housing
Warning
Remove the front housing to open the A
LIA ventilator.
Risk of an electric shock
Disconnect from the power source before opening the housing!
The front housing is screwed to the frame with 15 countersunk screws
M4x10. The 9 screws used on the side of the device are also fitted with
rosettes to prevent damage to the plastic front housing when tightening
the screws.
Fig. 1: Opening the housing
The front housing can now be hinged open to the front.
Before hinging open the front housing, ensure that the connectors for the
tube heating, the temperature sensors and the pneumotachograph have
been disconnected.
2
The plastic front housing is coated with an EMC varnish on the inside. It
contains the input periphery of the A
LIA ventilator.
This input periphery consists of touch buttons, the IGR (push/turn knob),
the LCD display, the controller board, the double flowmeter ON/OFF
switch on the front pane.
The entire pneumatic system is accommodated in the basic housing. Only
the double flowmeter is in the front housing.
The pneumatic system consists of the compressor, the fresh gas reservoir
and its control valves. The pressure switches for detecting the intake
pressure are firmly integrated in the pneumatic system
2
1
3
4
Fig. 7: pneumatic components
1 compressor 2 fresh gas reservoir
3 gas inlets 4 service flap
The patient part is connected to the basic device from the front.
The patient part is held in position by the guide rail.
LIA ventilator uses an 8-bit micro controller. This controls the
The A
sensors and actuators in the system.
In addition to these components, the unit also has electronic components
which provide the electrical power supply.
These consist of a power supply unit and a device for bridging any power
failure. In the event of a power failure, the device switches over to battery
operation and regulates the battery charging process.
The sensor board generates voltages of ±15V and -5V, and the controller
board generates + 5V.
3.1.1 Sensor board
The sensor board acts as interface for triggering the actuators (control of
the fresh gas reservoir and the expiration valve) and for registering the
measured data (pressure and differential pressure sensors, AD converter).
An SPI bus system provides the link to the micro controller. This
simplifies linking the various components in terms of both hardware and
software.
The following illustration shows the components on the sensor board.
The main task of the sensor board is to convert the analog signals.
Rechargeable battery voltage
"DC_good"
signal
On/off signal
Input voltage
12 V (for heating and valves)
Inspired pressure
Expired pressure
Flow sensor differential pressure
Oxygen sensor cell voltage
Heating
Valves
"Batt_enable"
Pressure sensors
The power supply for the electronic components is generated by a
DC/DC converter on the sensor board, generating +15V, -5V and -15V.
The input signals for On/Off, DC_good, input voltage and "Batt_enable"
are sent to the SPI bus by a slide register. The input signals for pressure,
flow, O
and battery voltage are sent to the SPI bus following
2
amplification and conversion via an 8-channel AD/12 bit converter.
The AD converter is equipped with protective diodes (max. 5V) on the
input side.
A serial chargeable slide register with power amplifier is used for
triggering the heating and the valves, via the SPI bus.
The respiratory pressure is measured at the inspiration fitting of the
patient part. The pressure present here is conveyed via a tube to the
pressure sensors (respiratory pressure).
For safety reasons, two pressure sensors are provided for measuring the
respiratory pressure.
The differential pressure developing over the resistance body of the PNT
head is used for measurement of the volume flow. This differential
pressure is conveyed via two tubes to the differential pressure sensor.
3 pneumotachograph connector 5 pressure fittings
6 water bath connectors
The calibration block allows for offset calibration of the differential
pressure sensor and the oxygen sensor.
The interposed solenoids V4 and V5 switch the measuring lines of the
differential pressure sensor against the atmosphere. The valves are
switched at exactly the same time. The value measured in this way
corresponds to a flow of 0 l/min.
FiO
FiO2 measurement
measurement is carried out using an electrochemical cell (fuel cell).
2
This oxygen sensor generates a voltage depending on the prevailing
oxygen concentration. The voltage is then converted electronically into a
corresponding signal.
To calibrate the oxygen sensor, compressed air ( 21% O
) is blown
2
against it via solenoid V6. During this procedure, the display shows
"CAL" in the window of the FiO
value. After completing calibration, the
2
valve switches the oxygen sensor back to the output of the double
flowmeter.
Once the ON switch on the front of the unit has been pressed, the system
starts with a self-test which shows the user that the control elements are
functioning. The self-test consists of the following sequence:
1. green LED of the mains power supply lights up
2. green LED of the 12V vehicle power supply lights up
3. green LED of the battery lights up
4. red LED of the battery lights up
5. piezo sounds
6. everything goes off again
The battery test is only carried out when A
LIA starts up.
The supply voltages are switched off internally for this purpose.
The test is displayed by the red and green LEDs in the battery symbol
which flash alternately.
The battery test is carried out with load on the battery by switching the
compressor on (approx. 2 amps for 6 s). This starts immediately after the
self-test. A brief alarm is given if the system does not pass the battery
test.
Total discharge
protection for the
battery
Total discharge protection comes into force when the measured voltage
of the battery is <10.8V ( Batt_enable = 0). An acoustic and visual alarm
is given.
The user should stop using the ventilator and recharge the battery, or
provide another voltage source.
Compressor ON/OFF
The compressor is switched on/off without any contact.