![](/html/f3/f3df/f3df72ac26885a73bf50b5e70638be3a69da51c901bf47caee32e44b241a092b/bg2.png)
Smart, Real-Time Mass
Flow Measurement
The Fox Model FT3 measures two
important process variables with a
single instrument, providing isolated
4-20mA and pulse outputs for ow
rate, and a 4-20mA pulse output for
process gas temperature or a second
ow rate output.
Direct mass ow measurement,
exceptional low-ow sensitivity, fast
response, and low maintenance
requirements distinguish the Fox
Model FT3. Virtually immune to
changes in temperature and pressure,
the ow meter delivers repeatable,
accurate mass ow measurement
under varying loads. It’s rugged, no-
moving-parts design is also ideal for
high-vibration industrial environments,
and enhanced EMI immunity makes
it suitable for environments where
electric motors, ignitors or dirty power
may affect instrument performance.
Creative Solutions - Calibration
Validation with Fox Model FT3
Fox Thermal knows that customers
want creative and effective solutions
to some of their common process
problems. Downtime due to
equipment maintenance or calibration
costs money and wastes valuable time.
Calibration Validation of ow meters
in the eld provides assurance of the
functionality of the meter and avoids
the downtime associated with annual
calibrations.
Used in succession, Fox’s CAL-V™
and Zero CAL-CHECK® Tests can
give you the reassurance that your
meter is performing accurately in the
eld without the need to interrupt
ow or send the meter back to the
factory. The tests are quick and easy
to perform at any time and help with
CALIBRATION VALIDATION:
Typical Requirements of Competitive Models
Stop the ow*
Remove meter from pipe*
Disconnect wires from ow meter
Look up data on ow meter’s calibration certicate
Measure electrical characteristics with volt ohm meter
Perform calculations to evaluate ow meter performance
Set process pressure to manufacturer’s calibration pressure
Connect auxiliary test equipment and/or test gases to ow
meter
*When using a packing gland assembly for Zero CAL-CHECK®
NORMAL MEASUREMENT MODE
Sensor Control Signal
SENSOR
SIGNAL
PROCESSING
ELECTRONICS
PowerPro™ Sensor
In normal measurement mode the signal processing electronics control the sensor.
CAL-V™ MODE
Sensor Control Signal
SENSOR
SIGNAL
PROCESSING
ELECTRONICS
PowerPro™ Sensor
In CAL-V™ mode, the microprocessor controls the sensor and determines the resulting electrical
characteristics.
the challenging requirements for
measuring the ow of air and gases.
look-up at any time.
CAL-V™ is an operator-initiated test
MICROPROCESSOR
MASS FLOW
MEASUREMENT
MICROPROCESSOR
CALIBRATION
VALIDATION
Calibration
Validation
PASS/FA IL
that can be performed at any ow rate,
CAL-V™
The CAL-V™ feature is an in-situ
calibration routine that validates the
ow meter’s calibration accuracy by
testing the functionality of the sensor
and its associated signal processing
circuitry. This innovative approach lets
you validate instrument calibration
in the pipe, at process conditions,
with just a push of a button. At the
conclusion of the test, the meter will
display a pass/fail message and the
CAL-V™ data is saved in the meter for
Other Thermal
Flow Meters
FT3 with CALV™ & Zero CALCHECK®
including zero, and is completed in just
three to four minutes. During the test,
the meter’s microprocessor adjusts
the signal to the sensor elements and
determines the resulting electrical
characteristics. These site-determined
characteristics are compared with the
data that was collected and stored in
the instrument electronics during the
original factory calibration. Matching
data within established tolerances
conrms the meter is measuring
accurately.
Zero CAL-CHECK® Tests
The Zero CAL-CHECK® tests are
used to ensure that the ow meter
still retains its original NIST-traceable
calibration at zero ow. The test can
be performed in one of two ways: In-
situ or Out-of-Pipe.
Required Not Required
In-situ Zero CAL-CHECK®
If zero ow can be established, the
sensor does not need to be removed
and the procedure can be done in
the pipe. If zero ow cannot be
established, a Fox Packing Gland
Assembly is used to remove the sensor
Outputs