Fox Thermal Instruments FC10 Operating Manual

FC10 Installation Instructions for Insertion Meters
Definition: The FC10 is a flow conditioner installed between two flanges upstream of the flow
meter that corrects irregular flow profiles ensuring flow meter accuracy.
Scope: The FC10 is used with flanges in schedule 40 pipe with insertion style meters. Purpose: This procedure describes proper installation of the FC10 to ensure flow meter
accuracy.
Directions:
1. Ensure proper installation of the branch outlet for the flow meter. See Fox Document 107590 for more information.
2. Determine the correct distance for dimension “X” for the appropriate schedule 40 pipe in the table below.
Figure 1.1: FC10 Minimum Straight Run Requirements
FC10 Flow Conditioner
installed between two anges
Irregular Flow Prole
FLOW
Schedule
40 Pipe
Size
1.5 6 (152) 8 (203) 2 6 (152) 10 (254)
2.5 9 (229) 15 (381) 3 9 (229) 15 (381) 4 9 (229) 20 (508)
“X”
Inch (mm)
5D Min. Up/Downstream
Inch (mm)
Proper Flow Prole
Schedule
Branch Outlet
F1
F2 F3 F4
“X”
“X”
40 Pipe
Size
5 10 (254) 25 (635) 6 12 (305) 30 (762) 8 16 (406) 40 (1016)
10 20 (508) 50 (1270)
Inch (mm)
(installed by customer)
5D5D
5D Min. Up/Downstream
Inch (mm)
Fox Thermal Instruments, Inc. 399 Reservation Road, Marina, CA 93933 www.foxthermalinstruments.com 831-384-4300
DOCUMENT NO.: 107679, REV.: B
Page 1 of 2
3. Position the FC10 Flow Conditioner correctly between the flanges.
(customer supplied)
• The FC10 must be installed between flanges and two gaskets for optimum seal.
• Center the flow conditioner between the flanges.
• The FC10 is equipped with an orientation marker. This marker must be aligned in parallel with the probe of the flow meter.
• The Orientation marker is also etched with “UPSTREAM” to indicate which side, or face, of the FC10 must be pointed upstream. When installed correctly, the louvers will point downstream.
Figure 1.2: Orientation of FC10 Flow Conditioner
Orientation marker must be parallel to
ow meter probe FC10 Orientation marker with “UPSTREAM” etching
FLOW
Gaskets
(supplied by customer)
(see table previous page)
4. Tighten bolts on the flanges to complete installation of the FC10.
“X”
F1
Flow meter
F2 F3 F4
Probe
Branch Outlet
Fox Thermal Instruments, Inc. 399 Reservation Road, Marina, CA 93933 www.foxthermalinstruments.com 831-384-4300
DOCUMENT NO.: 107679, REV.: B
Page 2 of 2
Loading...