Fluke Y5020 Service manual

Instruction Manual
Model Y5020
Current Shunt
®
P/N 516591 September 1979
Y5020
Y5020
The Model Y5020 is a precision, four-terminal, current shunt designed for use in calibrating ac/dc current sources in the range of 0 to 20 amps, do to 5 kHz. The shunt is housed in a forced-air cooled, PTI (Portable Test Instrument) case. This case is compatible with all instruments in the Fluke PTI product line and can be stacked with other PTI products to form a portable test system.
Forced-air cooling is accomplished by a fan mounted on the rear panel. Air is pulled in through the fan, passed over the shunt, and exhausted through the rear panel. Power for the fan is derived from the power line. Two voltage
configurations are available, 115 or 230V ac ±10%, 50 to 60 Hz.
Four front panel terminals provide electrical access to the shunt; two are current input terminals and two are voltage output (sense) terminals. When the shunt is connected into a circuit and operating at its maximum current limit (20 amps), the input connections and the shunt present a burden voltage of less than 0.25V dc.

SPECIFICATIONS

Specifications for the Y5020 Current Shunt are given in Table 1.
1. Locate and pull the black latches on both sides of the unit to their extended position.
2. Place the unit on top of the PTI stack with the front panel toward the front of the stack.
3. When the unit is properly seated push both latches to the in position. This will lock the Y5020 to the stack.

CONTROLS AND CONNECTORS

The front- and rear-panel controls and connectors on the Y5020 are shown in Figure 1 and described in Table 2.
OPERATING NOTES Input Power
Line power requirements for the Y5020 are specified on a decal located on the bottom side of the unit. These include line voltage requirements (115V or 230V ac ±10%, 50 to 60 Hz) and replacement fuse types. Input power is used solely to power an internal cooling fan. A front panel POWER switch allows the fan to be turned on and off.

Fuse Replacement

The Y5020 includes a fuse to protect the ac power line from an accidental overload. If replacement is required, use a 1/8 amp, 250 volt (AGC) fuse. Use the following procedure to replace the fuse:

INSTALLAT ION

The Y5020 PTI case is designed to stack with other Fluke instruments in the PTI product line. Use the following procedure to stack the Y5020 with the other PTI instruments.
Table 1. Y5020 Specifications
NOMINAL RESISTANCE ACCURACY OF CALIBRATION VALUE
DC DC to 1 kHz
1 kHz to 5 kHz STABILITY OF CALIBRATION VALUE MAXIMUM CURRENT BURDEN VOLTAGE AT 20 AMPS TEMPERATURE COEFFICIENT POWER COEFFICIENT POWER
1. Set the POWER switch to OFF and disconnect the
unit from line power.
2. Locate the fuse holder on the rear panel.
0.01±1% ±.01% ±.01% ±.025% ±.035% (30 ppm per 6 months) 20A dc or rms ac Less than 0.25V
0.001% per °C0.005% at 20 amps
115V ac ±10% or 230V ac ±10%, 11 watts for cooling fan.
PROTECTION CLASS 3
Relates solely to insulating or grounding properties as defined in IEC 348.
1
Y5020
12
3
54
6
Figure 1. Controls and Connectors
2
Loading...
+ 8 hidden pages