Fluke VT650, VT900 User Manual

Page 1
VT650/VT900
Gas Flow Analyzer
Users Manual
FBC - 0099 October 2017, Rev. 1
©2017 Fluke Corporation. All rights reserved. All product names are trademarks of their respective companies.
Page 2
Fluke Biomedical warrants this instrument against defects in materials and workmanship for one year from the date of original purchase OR two years if at the end of your first year you send the instrument to a Fluke Biomedical service center for calibration. You will be charged our customary fee for such calibration. During the warranty period, we will repair or at our option replace, at no charge, a product that proves to be defective, provided you return the product, shipping prepaid, to Fluke Biomedical. This warranty covers the original purchaser only and is not transferable. The warranty does not apply if the product has been damaged by accident or misuse or has been serviced or modified by anyone other than an authorized Fluke Biomedical service facility. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY.
This warranty covers only serialized products and their accessory items that bear a distinct serial number tag. Recalibration of instruments is not covered under the warranty.
This warranty gives you specific legal rights and you may also have other rights that vary in different jurisdictions. Since some jurisdictions do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you. If any provision of this warranty is held invalid or unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
7/07
Page 3
Notices
All Rights Reserved
Copyright 2017, Fluke Biomedical. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any language without the written permission of Fluke Biomedical.
Copyright Release
Fluke Biomedical agrees to a limited copyright release that allows you to reproduce manuals and other printed materials for use in service training programs and other technical publications. If you would like other reproductions or distributions, submit a written request to Fluke Biomedical.
Unpacking and Inspection
Follow standard receiving practices upon receipt of the instrument. Check the shipping carton for damage. If damage is found, stop unpacking the instrument. Notify the carrier and ask for an agent to be present while the instrument is unpacked. There are no special unpacking instructions, but be careful not to damage the instrument when unpacking it. Inspect the instrument for physical damage such as bent or broken parts, dents, or scratches.
Technical Support
For application support or answers to technical questions, either email techservices@flukebiomedical.com or call 1-800- 850-4608 or 1-440-248-9300 (Europe +31-40-2965314).
Claims
Our routine method of shipment is via common carrier, FOB origin. Upon delivery, if physical damage is found, retain all packing materials in their original condition and contact the carrier immediately to file a claim. If the instrument is delivered in good physical condition but does not operate within specifications, or if there are any other problems not caused by shipping damage, please contact Fluke Biomedical or your local sales representative.
Returns and Repairs
Return Procedure
All items being returned (including all warranty-claim shipments) must be sent freight-prepaid to our factory location. When you return an instrument to Fluke Biomedical, we recommend using United Parcel Service, Federal Express, or Air Parcel Post. We also recommend that you insure your shipment for its actual replacement cost. Fluke Biomedical will not be responsible for lost shipments or instruments that are received in damaged condition due to improper packaging or handling.
Use the original carton and packaging material for shipment. If they are not available, we recommend the following guide for repackaging:
• Use a double–walled carton of sufficient strength for the weight being shipped.
• Use heavy paper or cardboard to protect all instrument surfaces. Use nonabrasive material around all projecting parts.
• Use at least four inches of tightly packed, industry-approved, shock-absorbent material around the instrument.
Page 4
Returns for partial refund/credit: Every product returned for refund/credit must be accompanied by a Return Material Authorization (RMA) number, obtained from our
Order Entry Group at 1-440-498-2560.
Repair and calibration:
To find the nearest service center, go to www.flukebiomedical.com/service
or
To ensure the accuracy of the Product is maintained at a high level, Fluke Biomedical recommends the product be calibrated at least once every 12 months. Calibration must be done by qualified personnel. Contact your local Fluke Biomedical representative for calibration.
Certification
This instrument was thoroughly tested and inspected. It was found to meet Fluke Biomedical’s manufacturing specifications when it was shipped from the factory. Calibration measurements are traceable to the National Institute of Standards and Technology (NIST). Devices for which there are no NIST calibration standards are measured against in-house performance standards using accepted test procedures.
WARNING
Unauthorized user modifications or application beyond the published specifications may result in electrical shock hazards or improper operation. Fluke Biomedical will not be responsible for any injuries sustained due to unauthorized equipment modifications.
Restrictions and Liabilities
Information in this document is subject to change and does not represent a commitment by Fluke Biomedical. Changes made to the information in this document will be incorporated in new editions of the publication. No responsibility is assumed by Fluke Biomedical for the use or reliability of software or equipment that is not supplied by Fluke Biomedical, or by its affiliated dealers.
Manufacturing Location
The VT650/VT900 Gas Flow Analyzer is manufactured at Fluke Biomedical, 6920 Seaway Blvd., Everett, WA, U.S.A.
In the U.S.A. and Asia: Cleveland Calibration Lab Tel: 1-800-850-4608 x2564 Email: globalcal@flukebiomedical.com
In Europe, Middle East, and Africa: Eindhoven Calibration Lab Tel: +31-40-2675300 Email: ServiceDesk@fluke.com
Page 5
i
Table of Contents
Title Page
Introduction .................................................................................................... 1
Key Features .................................................................................................. 1
Safety Information .......................................................................................... 2
Unpacking and Inspection .............................................................................. 5
Accessories .................................................................................................... 6
The Analyzer .................................................................................................. 8
Power On the Analyzer ......................................................................... 10
Analyzer Connections ........................................................................... 12
Airway Flow (Inlet and Exhaust) .................................................... 12
Ultra-Low Flow + and - (VT900) .................................................... 13
High Pressure ............................................................................... 13
Low Pressure (+ and -) .................................................................. 14
Ultra-Low Pressure (VT900 only) .................................................. 14
Test Setup .................................................................................................... 15
Bi-directional Flow Mode ....................................................................... 15
Page 6
VT650/VT900
Users Manual
ii
Unidirectional Flow Mode ......................................................................16
Inspiratory Flow Connections ........................................................ 16
Expiratory Test Connections ......................................................... 17
Operations .................................................................................................... 18
Measured Signals ................................................................................. 18
Airway Flow ................................................................................... 19
Airway Pressure ............................................................................ 19
Airway Temperature and Humidity ................................................19
High Pressure ................................................................................ 19
Low Pressure ................................................................................ 20
Ultra-Low Pressure (VT900) .......................................................... 20
Ultra-Low Flow (VT900) ................................................................ 20
Barometric Pressure ...................................................................... 20
Oxygen Concentration ................................................................... 21
Calculated Breath Parameters ...................................................... 21
Excel Add-in .......................................................................................... 21
Measurements ...................................................................................... 22
Make a Measurement .................................................................... 22
Save a Measurement .................................................................... 22
Main Menu Functions ............................................................................ 24
Profiles Menu ................................................................................ 24
Setup Menu ................................................................................... 25
Units .............................................................................................. 31
Special Tests Menu ....................................................................... 32
Calibrate Oxygen ........................................................................... 33
Customize Breath Views ............................................................... 34
Page 7
Contents (continued)
iii
Memory Menu ............................................................................... 34
Test ID ........................................................................................... 34
Maintenance, Service and Calibration ......................................................... 35
Cleaning ................................................................................................ 35
Oxygen Sensor Replacement ............................................................... 36
Battery Status ....................................................................................... 36
Battery Replacement ............................................................................ 37
Replaceable Parts ................................................................................. 38
Service and Calibration ......................................................................... 38
Specifications ............................................................................................... 39
Pressure ................................................................................................ 40
Flow ...................................................................................................... 41
Page 8
VT650/VT900
Users Manual
iv
Page 9
1
Introduction
The VT650/VT900 Gas Flow Analyzer (the Analyzer or Product), is a general purpose gas flow analyzer with special features for testing mechanical patient ventilators. The Analyzer measures bi-directional air flow, high and differential low pressure, barometric pressure, Oxygen concentration and airway pressure, airway temperature and airway humidity. The VT900 also measures ultra-low flow (±750 ml/ min) and ultra-low pressure (0 mbar to 10 mbar). The Analyzer can be controlled externally using USB commands or automated with available software. The Analyzer operates on a rechargeable Li-Ion battery or external power supply for stationary or portable use. All figures show the VT900 unless otherwise noted.
Key Features
Full range, bi-directional air flow and volume channel
Ultra-low flow and pressure ranges (VT900)
High pressure, vacuum and differential low pressure
Airway pressure, Oxygen concentration, temperature, and humidity
Barometric pressure
External trigger input (VT900)
Rechargeable Li-ion battery with up to 8 hours battery life
•USB port
Customizable profiles that can be saved
Numerical and graphical screens with real time data
Automation software available
On-board memory
Page 10
VT650/VT900
Users Manual
2
Safety Information
A Warning identifies conditions and actions that pose hazards to the user; a Caution identifies conditions and actions that may damage the Product or the equipment under test.
XW Warning
To prevent possible electrical shock, fire, or personal injury:
Read all safety information before you use the Product.
Use the Product only as specified, or the protection supplied by the Product can be compromised.
Carefully read all instructions.
Do not use the Product around explosive gas, vapor, or in damp or wet environments.
Use this Product indoors only.
Do not use the Product if it operates incorrectly.
Disable the Product if it is damaged.
Do not use the Product if it is damaged.
The battery door must be closed and locked before you operate the Product.
Recharge the battery when the low battery indicator shows to prevent incorrect measurements. Low battery will also corrupt the memory card.
Remove all probes, test leads, and accessories before the battery door is opened.
Remove all probes, test leads, and accessories that are not necessary for the measurement.
Use only specified replacement parts.
Have an approved technician repair the Product.
Page 11
Gas Flow Analyzer
Safety Information
3
Batteries contain hazardous chemicals that can cause burns or explode. If exposure to chemicals occurs, clean with water and get medical aid.
Do not disassemble the battery.
Repair the Product before use if the battery leaks.
Use only Fluke approved power adapters to charge the battery.
Do not short the battery terminals together.
Do not disassemble or crush battery cells and battery packs.
Do not keep cells or batteries in a container where the terminals can be shorted.
Do not put battery cells and battery packs near heat or fire. Do not put in sunlight.
Remove the battery if the Product is not used for an extended period of time, or if stored in temperatures >50 °C. If the battery is not removed, battery leakage can damage the Product.
W Caution
To prevent possible damage, remove the O2 sensor if the Product is stored in temperatures >50 °C.
Page 12
VT650/VT900
Users Manual
4
Symbols used on the Analyzer and in this manual are explained in Table 1.
Table 1. Symbols
Symbol Meaning Symbol Meaning
W
WARNING. RISK OF DANGER.
X
WARNING. HAZARDOUS VOLTAGE. Risk of electric shock.
Consult user documentation.
P
Conforms to European Union directives.
+
Power button
)
Conforms to relevant North American Safety Standards.
Li-ion Battery
Conforms to relevant Australian Safety and EMC standards.
"
Power input 15 V dc 2.0 A
Conforms to relevant South Korean EMC Standards.
Conforms to the Appliance Efficiency Regulation (California Code of Regulations, Title 20, Sections 1601 through 1608), for small battery charging systems.
~
This product complies with the WEEE Directive marking requirements. The affixed label indicates that you must not discard this electrical/electronic product in domestic household waste. Product Category: With reference to the equipment types in the WEEE Directive Annex I, this product is classed as category 9 "Monitoring and Control Instrumentation" product. Do not dispose of this product as unsorted municipal waste.
Page 13
Gas Flow Analyzer
Unpacking and Inspection
5
Unpacking and Inspection
Make sure you do not damage the Analyzer as you unpack.
Inspect the shipping carton for damage.
If there is no damage, remove the
Analyzer from the shipping case. Save the box and packing materials.
If the shipping carton is damaged,
carefully continue to unpack the Analyzer. Note any dents and scratches on the Analyzer. Save the damaged shipping carton and packing material for the carrier’s inspection.
Do a visual inspection. Make sure the Analyzer is intact. If there is any physical damage, such as bent or broken parts, dents, or scratches, call a Fluke Biomedical Service Center immediately. To return the Analyzer to Fluke Biomedical for service see Returns and Repairs.
Check the standard accessories. If any accessories are missing, contact a Fluke Biomedical Service Center.
Page 14
VT650/VT900
Users Manual
6
Accessories
Table 2 is a list of the standard accessories provided with the Analyzer.
Table 2. Standard Accessories
Item Part Number
USB Serial cable 4015274
AC power adapter 4760480
Accessory kit with: 4922115
Bacterial filter for external connection to the flow ports (1) 2133712
1.2 m (4 ft) silicon tubing (2) 2237172
22 mm ID x 22 mm ID tubing adapters (2) 2133305
22 mm OD x 22 mm OD tubing adapters (2) 2133291
Tapered 15 mm OD x 22 mm OD tubing adapters (2) 2133269
DISS Handtight Nut/Nipple to 6.4 mm (1/4 in) ID hose barb adapter (1) 2216329
Certificate of Calibration with test data --
Page 15
Gas Flow Analyzer
Accessories
7
Table 3 is a list of optional accessories.
Table 3. Optional Accessories
Item Part Number
Soft-sided carrying case for ACCU-LUNG 2397628
ACCU LUNG II test lung 4281291
ACCU-LUNG Lung Simulator with Soft-sided carrying case (2397628)
2387318
VESA system mount 4969657
Page 16
VT650/VT900
Users Manual
8
The Analyzer
Figure 1 shows the top of the Analyzer.
Figure 1. Top of the Analyzer
Touchsceen LCDFlow exhaustFlow inlet
D Low pressure + and - E High pressure and vacuum F Ultra-low pressure (VT900) G Ultra-low flow + and - (VT900) H Power button
FLOW INLET
±
300 lpm
FLOW
EXHAUST
ULTRA
LOW FLOW
LOW
PRESSURE
ULTRA
LOW PRESSURE
0 TO 10 mbar
HIGH
PRESSURE
750 ml/min 160 mb ar0.8 TO 10 bar
GAS FLOW ANALYZER
VT900
4
5
7
6
8
3
2
1
Page 17
Gas Flow Analyzer
The Analyzer
9
Figure 2 shows the back of the Analyzer.
Figure 2. Back of the Analyzer
Handle clipUSB portOxygen sensor door
D External trigger input (VT900) E Battery status indicator F External DC power input
O2 SENSOR ACCESS DOOR
1
2
1
3456
Page 18
VT650/VT900
Users Manual
10
Figure 3 shows the bottom of the Analyzer.
Figure 3. Bottom of the Analyzer
VESA mount points (FDMI MIS-C, fits WxH
of 75 mm x 35 mm)
Oxygen sensor door screwBail
D Battery door screws E Battery door
Power On the Analyzer
To power on the Analyzer, push +.
The Analyzer defaults to the Airway screen. Figure 4 shows the main screen of the Analyzer.
2 3
4
5
1
Page 19
Gas Flow Analyzer
The Analyzer
11
To make a selection, tap the screen.
Figure 4. Example of Airway Measurement
Measurement and Other MenusSelected ProfileMeasurement settings
D Battery symbol E Display area F Screen options
2
3
4
5
6
1
Page 20
VT650/VT900
Users Manual
12
Analyzer Connections
You can connect the Analyzer to a ventilator and test lung in either a bi-directional or unidirectional flow configuration.
Airway Flow (Inlet and Exhaust)
The outer diameter of the airway port is a standard 22 mm fitting for use with patient hoses. The inner diameter of the airway port accepts a 15 mm male respiratory fitting for use with endotracheal tubes, gas sampling adapters, and similar equipment.
WCaution
To avoid damage to the Analyzer and avoid adverse affects on the Analyzer performance:
Do not put metal objects into connectors.
Always use the external flow filter on the main airflow channel inlet. This helps reduce turbulence and keeps out small particles that could damage the flow sensor.
To avoid damage to the sensor, make sure the pressure inside the airway port does not exceed 5 psi.
Measure only dry gases with the airway port. Do not use this port to measure humidified gases.
Page 21
Gas Flow Analyzer
The Analyzer
13
Ultra-Low Flow + and - (VT900)
The + and - ultra-low-flow ports have a barbed hose fitting connection.
WCaution
To avoid damage to the Analyzer and avoid adverse affects on the Analyzer performance:
To avoid damage to the sensor, make sure the pressure inside the ultra-low-flow port does not exceed 25 psi.
Measure only dry gases with the ultra-low-flow port. Do not use this port to measure humidified gases.
High Pressure
The high-pressure port is primarily to test wall and tank pressurized gas sources. The connector works with standard oxygen DISS fittings, as used on oxygen supply hoses.
WCaution
To avoid damage to the Analyzer and avoid adverse affects on the Analyzer performance:
To avoid damage to the sensor, make sure applied pressure does not exceed 188 psi (13 bar).
Measure only dry gases with the high-pressure port. Do not use this port to measure fluid pressure.
Page 22
VT650/VT900
Users Manual
14
Low Pressure (+ and -)
Measure differential low pressure between the + and - ports or gauge pressure on either port. Connectors are barbed hose fittings.
WCaution
To avoid damage to the Analyzer and avoid adverse affects on the Analyzer performance:
To avoid damage to the sensor, make sure applied pressure does not exceed 5 psi.
Measure only dry gases with the low-pressure port. Do not use this port to measure fluid pressure.
Ultra-Low Pressure (VT900 only)
The ultra-low pressure port has a barbed hose fitting connection.
WCaution
To avoid damage to the Analyzer and avoid adverse affects on the Analyzer performance:
To avoid damage to the sensor, make sure applied pressure does not exceed 5 psi.
Measure only dry gases with the ultra-low-pressure port. Do not use this port to measure fluid pressure.
Page 23
Gas Flow Analyzer
Test Setup
15
Test Setup
Use the airway port for ventilator testing. Use a test lung to set up a test that measures ventilator parameters in bi-directional or uni-directional modes. Fluke Biomedical recommends the bi­directional mode.
Bi-directional Flow Mode
For bi-directional flow connections, see Figure 5.
1. Use a Y-piece adapter to connect the ventilator to the flow inlet on the Analyzer.
2. Use a standard breathing hose to connect to the exhaust port on the Analyzer.
The Analyzer shows the flow of gas delivered by the ventilator.
Figure 5. Bi-Direction Flow Mode Connections
FLOW
INLET
±
300 lpm
FLOW
EXHAUST
ULTRA
LOW FLOW
LOW
PRESSURE
ULTRA
LOW PRESSURE
0 TO 10 mbar
HIGH
PRESSURE
750 ml/min 160 mbar0.8 TO 10 bar
GAS FLOW ANALYZER
VT900
Inspiratory Hose
Expiratory Hose
Ventilator
Test Lung
Inlet
Exhaust
Page 24
VT650/VT900
Users Manual
16
Unidirectional Flow Mode
Use the unidirectional flow to measure either inspiratory or expiratory gas flow.
Inspiratory Flow Connections
For inspiratory connections, see Figure 6.
1. Connect the inspiratory hose to the flow inlet on the Analyzer.
2. Use a standard breathing hose to connect the test lung to the exhaust port on the Analyzer.
Figure 6. Inspiratory Breathing Circuit Connections
FLOW
INLET
±
300 lpm
FLOW
EXHAUST
ULTRA
LOW FLOW
LOW
PRESSURE
ULTRA
LOW PRESSURE
0 TO 10 mbar
HIGH
PRESSURE
750 ml/min 160 mbar0.8 TO 10 bar
GAS FLOW ANALYZER
VT900
Inspiratory Hose
Ventilator
Inlet
Exhaust
Expiratory Hose
Test Lung
Page 25
Gas Flow Analyzer
Test Setup
17
Expiratory Test Connections
For expiratory connections, see Figure 7.
1. Connect the expiratory hose from the test lung to the flow inlet on the Analyzer.
2. Use a standard breathing hose to connect the ventilator to the exhaust port on the Analyzer.
The Analyzer shows the flow of gas delivered by the ventilator.
Figure 7. Expiratory Breathing Circuit Connections
FLOW
INLET
±
300 lpm
FLOW
EXHAUST
ULTRA
LOW FLOW
LOW
PRESSURE
ULTRA
LOW PRESSURE
0 TO 10 mbar
HIGH
PRESSURE
750 ml/min 160 mbar0.8 TO 10 bar
GAS FLOW ANALYZER
VT900
Inspiratory Hose
Ventilator
Inlet
Exhaust
Expiratory Hose
Test Lung
Page 26
VT650/VT900
Users Manual
18
Operations
Use the Analyzer to measure flow and pressure. For each test: select the test and do the setup.
Displayed options depend on the selected test:
•Zero—Corrects sensor offsets for the selected measurements.
Graph—Select the parameters to graph.
Back—Return to the previous screen.
Clear—Clear the graph and statistics
(minimum, maximum, and average).
Autoscale—Toggles between Autoscale and manual scale.
Profiles—Select a different profile.
Save—Saves the final result.
Measured Signals
The Analyzer measures these signals:
Airway flow
Airway pressure
Airway temperature and humidity
High pressure
•Low pressure
Ultra-low pressure (VT900)
Ultra-low flow (VT900)
Barometric pressure
Oxygen concentration
Calculated breath parameters
Page 27
Gas Flow Analyzer
Operations
19
Airway Flow
The Analyzer has a full-range flow (±300 lpm), with bi-directional flow measurement. Flow measurements are either static flows—no breath variations, or ventilator waveforms—both an expiratory phase and an inspiratory phase. The Analyzer uses a heat transfer sensor to measure the flow.
Use this measurement for pediatric or adult ventilators or to determine the performance of many types of flow meters.
Airway Pressure
The Analyzer measures airway pressure from a proximal tap of the channel (near the exhaust port).
Airway Temperature and Humidity
There is a temperature / humidity sensor in the airway channel on the exhaust side of the flow sensor. The Analyzer uses the temperature reading to automatically adjust for the gas corrections (see Table 6).
Tap More Data on the Airway screen to see temperature and humidity.
High Pressure
The high-pressure port measures gauge pressure in the range of -0.8 bar to 10 bar. The Analyzer can use this pressure measurement for any gauge pressure in the given range.
Page 28
VT650/VT900
Users Manual
20
Low Pressure
The low-pressure port is a dual-port connection consisting of a (+) positive and a (-) negative pressure port.
The differential pressure range is ±160 mbar. Use this pressure measurement for any pressure differential or gauge pressure in the given range.
Ultra-Low Pressure (VT900)
The ultra-low pressure sensor measures gauge pressure in the range of 0 mbar to 10 mbar. Use this pressure range for accurate, very low pressure measurements.
Ultra-Low Flow (VT900)
The Analyzer (VT900 only) has ultra-low flow (±750 ml/min) bi-directional flow measurement capability. The flow measurement is for accurate, high resolution, static low flow measurements, not for ventilator waveforms. The flow sensor uses the heat transfer method to measure air flow. Use the ultra-low flow to measure many types of low flow meters.
Barometric Pressure
The Analyzer makes barometric pressure measurements. The barometer reads absolute pressures from 8 psia to 18 psia. The Analyzer also uses the barometer in the automatic gas corrections for airway flow measurements.
Page 29
Gas Flow Analyzer
Operations
21
Oxygen Concentration
The Analyzer measures oxygen concentration through the airflow channel. An integrated sensor measures the oxygen percent of the gas in the airflow channel of the Analyzer. The range for this sensor is 0 % to 100 %. The oxygen cell is mounted inside the enclosure on the rear bulkhead for the high-flow circuit. The oxygen cell must be replaced approximately once a year (VT650) or every 2 years (VT900).
Calculated Breath Parameters
From the primary flow and pressure measurements, the Analyzer calculates breath parameters. A breath detection algorithm determines the various phases of a ventilator breath and calculates the parameters listed in Table 10.
Excel Add-in
Install custom Excel add-in, available for download at www.flukebiomedical.com.
Use the
Excel Add-in on a PC to view results data. The Excel Add-in has these worksheets:
Data – shows saved data
Recording – shows recorded data
Graph – shows graphs and data
Page 30
VT650/VT900
Users Manual
22
Measurements
Figure 8 is an example of the measurement screen.
Make a Measurement
To make a measurement:
1. Select the type of measurement.
2. To set the offset to zero, tap Zero. Make sure to remove all the connections.
Note
You may need to block the airflow channel if room air currents are present.
3. To switch between Auto and manual scaling, tap Autoscale.
Save a Measurement
The Analyzer can save readings or start a recording to save.
To save a measurement:
1. On the measurement screen, tap Save.
2. Select the type of reading you want to save.
Data—The current data points.
Graph—The current data points in a
graphical format.
Record—Set up the parameters and
start a new recording to save.
Page 31
Gas Flow Analyzer
Operations
23
Figure 8. Example of Measurement
MeasurementsChange scalingGraph area
D Toggle manual or automatic scaling
1
2
2
3
4
Page 32
VT650/VT900
Users Manual
24
Main Menu Functions
Use the Main menu to access the Analyzer functions, including:
•Profiles
•Setup
•Units
Special Tests
•Memory
Customize Breath Views
Calibrate Oxygen
Test ID
Profiles Menu
You can configure the settings on the Analyzer to create test profiles. The analyzer can save up to 20 profiles.
The Analyzer stores the profiles using a numeric code and a profile name. 00 is the default numeric code. You cannot change the 00 default profile. An asterisk (*) indicates the profile that will be loaded at start up.
To select a profile, use the arrow keys.
To manage profiles, select Menus > Profile or tap Profile on the Airway screen.
Options are:
•View Current —View the settings for the current profile loaded on the Analyzer. Tap
More for another page of settings.
View Selection—View the settings of the
highlighted profile in the profile list. Tap More for another page of settings.
Page 33
Gas Flow Analyzer
Operations
25
*Use on Power-Up—Make the selected profile the default profile.
Back—Go to the main Profile menu.
•Edit Name—Change the name of the selected profile.
Load—Select a profile to load. If there are no saved profiles, the option is Default.
Save—Saves the current profile.
Setup Menu
Use the Setup menu to make and view the Analyzer settings.
To setup the Analyzer, select Menus > Setup.
For Setup selections, see Table 4.
Table 4. Setup Menu
Setup Description
Gas Select the type of gas. See Table 5.
Correction Mode
Select the type of Gas correction (temperature and pressure) to different gas standards. See Table 6.
Page 34
VT650/VT900
Users Manual
26
Breath Detect
Sets the options for Breath Detection. Options are:
Mode - Sets the breath detection mode:
Bidirectional Uni-directional Inspiratory Uni-directional Expiratory Off - disables breath triggering
Trigger Source - Sets how the breath detection is triggered:
Flow - Set the Threshold Pressure - Requires back pressure, such as a test lung External - enables external TTL triggering input on the Analyzer
(VT900)
Patient - Sets the type of patient:
Adult Pediatric
Threshold - Enter the numeric value for the detection limit.
+ (Inspiratory threshold) - (Expiratory threshold)
Table 4. Setup Menu (cont.)
Setup Description
Page 35
Gas Flow Analyzer
Operations
27
Instrument Information
Displays basic information about the Analyzer including:
Model number
Serial number
Firmware version
Date of calibration
Battery charge level
Memory remaining
Auto Dim
Options are:
On - The LCD dims after 5 minutes of no use. Tap to restore brightness.
Off - The LCD does not auto dim.
Beeper
Options are:
•Off
•Soft
•Medium
Loud
Table 4. Setup Menu (cont.)
Setup Description
Page 36
VT650/VT900
Users Manual
28
Time and Date
Options are:
Set Time Format - Select 12 hour with am/pm or 24 hour.
Set Time - Tap the arrows to set the time. If in 12-hour format, select am or pm.
Set Date Format - Select MM/DD/YYYY, DD/MM/YYYY, or YYYY/MM/DD.
Set Date - Tap the arrows to set the date.
Brightness
Options are:
•Low
Medium
•High
Language Select the language.
Table 4. Setup Menu (cont.)
Setup Description
Page 37
Gas Flow Analyzer
Operations
29
Table 5. Gas Types
Gas Type Description
Air Standard room air
N2 100 % Nitrogen
N2O 100 % Nitrous Oxide
CO2 100 % Carbon Dioxide
O2 100 % Oxygen
Ar 100 % Argon
Heliox 21 % Oxygen, and 79 % Helium
O2 bal N2O mix
[1]
Measured Oxygen, balance Nitrous Oxide
O2 bal He mix
[1]
Measured Oxygen, balance Helium
O2 bal N2 mix
[1]
Measured Oxygen, balance Nitrogen
[1]
For these gases, the oxygen concentration is measured using the oxygen sensor in the airflow channel. The balance is the other gas. These
gases can only be measured in the airflow channel, not available in the ultra-low-flow channel of the VT900.
Page 38
VT650/VT900
Users Manual
30
Table 6. Gas Correction Mode
Gas Correction Description
ATP Ambient temperature and pressure (actual humidity)
ATPD Ambient temperature and pressure, dry (0 % humidity)
ATPS Ambient temperature and pressure, saturated (100 % humidity)
STP20 Standard temperature 20 °C, standard pressure 760 mmHg (actual humidity)
STP21 Standard temperature 21 °C, standard pressure 760 mmHg (actual humidity)
STPD0 Standard temperature 0 °C, standard pressure 760 mmHg, dry (0 % humidity)
STPD20 Standard temperature 20 °C, standard pressure 760 mmHg, dry (0 % humidity)
STPD21 Standard temperature 21 °C, standard pressure 760 mmHg, dry (0 % humidity)
BTPS Body temperature 37 °C, ambient pressure 760 mmHg, saturated (100 % humidity)
BTPD Body temperature 37 °C, ambient pressure 760 mmHg, dry (0 % humidity)
Note: Gas correction applies to the airflow channel. The ultra-low-flow channel of the VT900 is set to STPD21.
Page 39
Gas Flow Analyzer
Operations
31
Units
You can change the units of measure for all readings.
To set the units:
1. Select Menus > Units.
2. Select the type of reading.
3. Select the unit of measure. See Table 7.
4. Select OK.
Table 7. Available Units of Measure
Unit Description
Flow
lpm Liters per minute
lps Liters per second
mlpm Milliliters per minute
mlps Milliliters per second
cfm Cubic feet per minute
Volume
lLiters
ml Milliliters
cf Cubic feet
Temperature
°C Celsius
°F Fahrenheit
Table 7. Available Units of Measure (cont.)
Unit Description
Page 40
VT650/VT900
Users Manual
32
Special Tests Menu
Use the Special Tests menu for more tests.
Leak Tests calculate the lost volume over the specified time. To do a leak test:
1. Select Menus > Special Tests.
2. Select the type of leak test:
Airway Pressure
High Pressure
Low Pressure
Ultra-Low Pressure
3. Use the keypad to enter the length of the test.
4. Tap Compliance and use the keypad to enter the compliance limit.
Use compliance to calculate volume lost. For example, enter the ml of volume lost per cmH
2
O. If blank, volume lost is not
calculated.
5. Tap Start.
Pressure
mbar millibar = 0.001 bar
bar 14.7 psi
mmHg Millimeters of mercury
inHg Inches of mercury
cmH
2
0
Centimeters of water
inH
2
0
Inches of water
psi Pounds per square inch
atm Atmosphere
kPa Kilopascals
Table 7. Available Units of Measure (cont.)
Unit Description
Page 41
Gas Flow Analyzer
Operations
33
Trend Tests show how a particular breath parameter performs over time. To do a trend test:
1. Select Menus > Special Tests > Trend Test.
2. Use the arrows to select the parameter to measure.
3. Use the keypad to enter the % limit for maximum allowed deviation.
4. Tap Start.
The Stacked Volume Test measures the volume of multiple breaths. Breath volume determines the accuracy of the ventilator in accumulating volume.
1. Select Menus > Special Tests > Stacked Volume Test.
2. Use the key pad to enter the number of breaths.
3. Select Start.
The test stops when the number of breaths is reached or you select End.
The High Frequency Test measures high­frequency flow and airway pressure.
1. Select Menus > Special Tests > High
Frequency Test
2. Apply the high-frequency flow or pressure.
Calibrate Oxygen
Calibrate the oxygen sensor at the beginning of each day that oxygen will be measured, and after replacing the sensor.
To calibrate the sensor:
1. Select Menus > Calibrate Oxygen.
2. Follow the on-screen instructions. Make sure to complete both steps.
If the calibration fails, replace the Oxygen sensor. If the problem persists, contact Fluke Biomedical for repair.
Page 42
VT650/VT900
Users Manual
34
Customize Breath Views
Use Customize Breath Views to configure custom breath screens with up to eight readings each. Select the Airway or the Breath parameters to show.
Memory Menu
Use Memory to manage stored readings.
1. Select Menus > Memory.
2. Select List.
Options are:
Back—Go back to the previous screen.
View—Views the selected reading.
Delete—Deletes the selected reading.
Delete All—Deletes all readings.
Test ID
Use the Test ID to identify saved test data. Use the keyboard to enter a Test ID. When viewing Memory, select a Test ID to see the saved data.
Page 43
Gas Flow Analyzer
Maintenance, Service and Calibration
35
Maintenance, Service and Calibration
The Analyzer requires little maintenance or special care; however it is a calibrated measuring instrument and should be treated with care.
W Caution
To prevent damage to the Analyzer or adverse affects on the Analyzer performance, do not drop the Analyzer and avoid mechanical abuse that could cause a shift in the calibrated settings.
Fluke Biomedical recommends storing the Analyzer in the carry case. Do not store the Analyzer where there is vibration.
During use, always use the external flow filter on the main airflow channel inlet. The filter helps reduce turbulence and keeps out small particles that could damage the flow sensor.
Cleaning
Clean the exterior of the Analyzer occasionally with a cloth dampened with a mild detergent solution. To remove stains and clean the Analyzer, use a solution of 70 % isopropyl alcohol. Fluke Biomedical does not recommend any other solvents.
W Caution
To prevent damage to the Analyzer or adverse affects on the Analyzer performance, do not spray liquid directly on the Analyzer. Do not immerse the Analyzer.
Page 44
VT650/VT900
Users Manual
36
Oxygen Sensor Replacement
The Analyzer uses a galvanic oxygen sensor that is under warranty for 12 months. You must replace the oxygen sensor cell when the oxygen calibration cannot be successfully completed. The oxygen sensor can operate for longer than 12 months depending on use. With normal use the VT650 oxygen sensor will last 12 months and the VT900 oxygen sensor will last 24 months.
To replace the oxygen sensor:
1. Remove the screws in the oxygen sensor door on the back of the Analyzer.
The oxygen sensor is a round plastic cylinder about one inch in diameter with wires protruding from the back of the sensor.
2. Disconnect the electrical connector on the sensor wire.
3. Unscrew the old sensor and remove it.
4. Screw in the new sensor. (See Replaceable Parts).
5. Re-connect the sensor wire.
6. Replace the oxygen sensor door.
7. Re-calibrate the oxygen sensor. (See
Calibrate Oxygen.)
Battery Status
The Battery status indicator on the back of the Analyzer has these normal states:
Solid red—Battery is charging
Solid green—Battery is 100 % charged
The battery has these error states:
Blinking red—Battery charger error
Blinking red/green—Battery charger error (Battery is near full charge.)
For these errors, remove the ac adapter and reinstall. If the problem reoccurs, replace the battery pack. If the problem persists, contact Technical Support.
Page 45
Gas Flow Analyzer
Maintenance, Service and Calibration
37
Battery Replacement
To check the battery charge level, remove the battery door and push the TEST button on the battery. The battery displays the percentage of charge.
Use only the replacement battery listed in Table 8.
To replace the battery (See Figure 9):
1. Loosen the screws on the battery door. (The screws do not come out.)
2. Remove the battery door.
3. Use the strap on the battery to lift out the battery.
4. Insert the new battery and replace the battery door.
Figure 9. Battery Replacement
1
2
3
Page 46
VT650/VT900
Users Manual
38
Replaceable Parts
Table 8 is a list of the replaceable parts for the Analyzer.
Service and Calibration
To maintain accuracy, calibrate the Analyzer annually. If the Analyzer does not operate successfully or if it needs calibration, return it to the Fluke Biomedical Service Center, as indicated under Warranty and Product Support. As part of this service, hardware and software updates are automatically installed.
W Caution
To prevent damage to the Analyzer or adverse affects on the Analyzer performance:
Other than replacement of minor components, such as the oxygen sensor, all Analyzer service should be done by qualified service personnel.
Table 8. Replaceable Parts
Item Fluke Part Number
Oxygen sensor (VT650) 2138514
Oxygen sensor (VT900) 4917220
O2 sensor door 4916215
Battery Door 4916194
Battery Pack 4948749
Protective cap (High­pressure)
2133735
Protective caps (Flow) 2133758
Protective caps (Pressure) 4918975
Page 47
Gas Flow Analyzer
Specifications
39
To keep foreign debris from entering the Analyzer, use flow and pressure protective caps when transporting the Analyzer.
To keep the Analyzer under warranty, make sure the Analyzer is calibrated by qualified service personnel only.
Specifications
Specifications are based on a one year calibration cycle and apply to ambient temperature 18 °C to 28 °C unless stated otherwise.
Display ....................................7 in, 800 x 480 touchscreen LCD
Communications .....................USB, Micro-B Device Port
Environmental Conditions
Operating Temperature ......10 °C to 40 °C
Storage Temperature..........-20 °C to +60 °C
Note
For storage temperatures below -15 °C or above +50 °C, remove the oxygen sensor.
Operating Humidity .............10 % to 90 % non-condensing
Storage Humidity ................5 % to 95 % non-condensing
Altitude................................3000 m
Power
AC adapter
Input Voltage Range ....... 100 V ac to 240 V ac
Input Frequency
Range .............................50 Hz/60 Hz
DC Output.......................15 V, 2 A
Polarity............................Center positive (+)
Page 48
VT650/VT900
Users Manual
40
Battery
Rechargeable Li-Ion
Battery ........................... 10.8 V, 2.5 Ah, 27 Wh, 3ICR19/66
Discharge temperature... 0 °C to 50 °C
Charge temperature ....... 0 °C to 40 °C
Battery life ...................... 8 hours
Battery charge time ........ 5 hours, typical
Note
Battery life is dependent on backlight brightness, auto­dim, and other battery intensive settings.
Pressure
Ultra Low Pressure (VT900)
Maximum applied
pressure ............................. 400 mbar
Operating Pressure ............ 0 mbar to 10 mbar
Span Accuracy .................. ±1 % of rdg or ±0.01 mbar
(whichever is greater)
Resolution .......................... 0.001 mbar
Low Pressure
Maximum applied
pressure ............................. 400 mbar
Operating Pressure............ (Differential) ±160 mbar
Span Accuracy................... ±0.5 % of rdg or ±0.1 mbar
(whichever is greater)
Resolution .......................... 0.01 mbar
High Pressure
Maximum applied
pressure ............................. 13 bar
Operating Pressure............ -0.8 bar to 10 bar
Span Accuracy................... ±1 % of rdg or ±7 mbar (whichever
is greater)
Resolution .......................... 1 mbar
Airway Pressure
Maximum applied
pressure ............................. 400 mbar
Operating Pressure............ ±160 mbar
Span Accuracy................... ±0.5 % of rdg or ±0.1 mbar
(whichever is greater)
Frequency Response......... 10 % to 90 % rise time <10 ms
Resolution .......................... 0.01 mbar
Sample rate
 200 Hz
Pressure Temperature
Coefficient .............................. Add 0.01 % of range per °C
between 10 °C and 18 °C Add 0.01 % of range per °C between 28 °C and 40 °C
Page 49
Gas Flow Analyzer
Specifications
41
Flow
Airway Flow
Resolution...........................0.01<100 slpm and
0.1 >100 slpm
Accuracy .............................see Table 9
Frequency Response..........10 % to 90 % rise time <10 ms
Sample Rate
 200 Hz
Dynamic Resistance ..........<2.00 cmH2O @ 60 slpm
Ultra-Low-Flow (VT900)
Range .................................±750 ml/min
Accuracy .............................±1.7 % of rdg or 0.01 slpm
Resolution...........................0.001 lpm
Flow Temperature
Coefficient ...............................Add 0.2 % of reading per °C
between 10 °C and 18 °C Add 0.2 % of reading per °C between 28 °C and 40 °C
Table 9. Airway Flow Range and Accuracy
Gas Range Specification
Main Airway Flow
Air, Nitrogen (N2), Oxygen (O2)
±300 slpm
1.7 % of rdg or
0.04 slpm
Argon, Heliox, O2 bal He, O2 bal N2
±300 slpm
3.0 % of rdg or
0.08 slpm, typical
Carbon Dioxide (CO2), Nitrous Oxide (N2O), O2 bal N20,
±150 slpm
3.0 % of rdg or
0.08 slpm, typical
Ultra-low Flow (VT900)
Air, Nitrogen (N2), Oxygen (O2)
±0.750 slpm
1.7 % of rdg or
0.01 slpm
Argon ±0.750 slpm
3.0 % of rdg or
0.02 slpm, typical
Nitrous Oxide (N2O) ±0.400 slpm
3.0 % of rdg or
0.02 slpm, typical
Note: Ultra-low flow that is over range can be detected as unstable. If this occurs, reduce flow.
Table 9. Airway Flow Range and Accuracy (cont.)
Gas Range Specification
Page 50
VT650/VT900
Users Manual
42
Volume (Airway Flow channel)
Range................................. 100 l
Accuracy ............................ ±1.75 % rdg or 0.02 l
(whichever is greater)
Resolution .......................... 0.001<1 and 0.01>1 l
Temperature and Humidity (Airway Flow channel)
Temperature range ............ 0 °C to 50 °C
Temperature accuracy*...... ±0.5 °C
Temperature resolution ...... 0.1 °C
Humidity range ................... 0 % to 100 % RH
Humidity accuracy* ............ 3 % RH from 20 % to 80 % RH
5 % RH < 20 %, and >80 % RH
Humidity resolution............. 0.1 % RH
* with steady state air flow
Oxygen concentration (Airway Flow channel)
Range................................. 0 % to 100 %
Accuracy ............................ ±1 % (VT900), ±2 % (VT650)
Resolution .......................... 0.1 %
Gas correction
See Table 6.
Gas types
Air, Nitrogen (N
2
), Nitrous Oxide (N2O), Carbon Dioxide (CO2),
Oxygen (O
2
), Argon, Heliox (21 % O2, 79 %He), Oxygen/
Nitrogen, Oxygen/Nitrous Oxide, Oxygen/Helium
Units of measure
Flow ................................... lpm (liter/minute), cfm (ft
3
/minute),
lps (liter/sec), ml/min, ml/sec
Pressure............................. psi, kPa, bar, mbar, atm, inH2O (at
4 °C, inHg (at 0 °C), cmH2O (at 4 °C), mmHg (at 0 °C)
Volume............................... l (liter), cf (ft
3
), ml
Temperature ...................... C, F
Humidity ............................. % RH
Safety
General .............................. IEC 61010-1: Pollution Degree 2
Lithium Battery ................... IEC 62133
Page 51
Gas Flow Analyzer
Specifications
43
Electromagnetic Compatibility (EMC)
International ........................IEC 61326-1: Controlled
Electromagnetic Environment
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency energy that is necessary for the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected to a low­voltage power supply network that supplies buildings used for domestic purposes. There may be potential difficulties in ensuring electromagnetic compatibility in other
environments due to conducted and radiated disturbances.
Caution: This equipment is not intended for use in residential environments and may not provide adequate protection to radio reception in such environments.
Emissions that exceed the levels required by CISPR 11 can occur when the equipment is connected to a test object.
Korea (KCC) ...................Class A Equipment (Industrial
Broadcasting & Communication Equipment)
Class A: Equipment meets requirements for industrial electromagnetic wave equipment and the seller or user should take notice of it. This equipment is intended for use in business environments and not to be used in homes.
USA (FCC)......................47 CFR 15 subpart B. This product
is considered an exempt device per clause 15.103.
Page 52
VT650/VT900
Users Manual
44
Table 10. Breath Parameter Accuracy
Parameter Abbreviation Resolution Range Accuracy
Inspiratory Tidal Volume Vti 0.1 ml 0 l to 60 l ±1.75 % of rdg or 0.02 l
Expiratory Tidal Volume Vte 0.1 ml 0 l to 60 l ±1.75 % of rdg or 0.02 l
Minute Volume (The Expiratory Tidal Volume times the
respiratory rate in breaths per minute. Reported as an eight-breath average.)
MV 0.001 lpm 0 l to 100 l ±1.75 % of rdg or 0.02 l
Breath Rate BPM 0.1 bpm 1 bpm to 1500 bpm 1 %
Inspiratory to Expiatory Time Ratio (I:E Ratio)
(The ratio of the Inspiratory Time to the Expiratory Time)
I:E 0.01 1:300 to 300:1 ±2 % or 0.1
Peak Inspiratory Pressure (The maximum pressure during the
Inspiratory Time.)
PIP 0.1 mbar ±160 mbar ±0.75 % of rdg or 0.1 mbar
Inspiratory Pause Pressure IPP 0.1 mbar ±160 mbar ±0.75 % of rdg or 0.1 mbar
Mean Airway Pressure (The mean pressure over the entire
breath.)
MAP 0.1 mbar ±160 mbar ±0.75 % of rdg or 0.1 mbar
Positive End-Expiratory Pressure (The average pressure during the End Expiratory State.)
PEEP 0.1 mbar ±160 mbar ±0.75 % of rdg or 0.1 mbar
Lung Compliance
[1]
CMPL 0.1 ml/mbar 0 ml/mbar to 1000 ml/mbar ±3 % or 0.1 ml/mbar
Page 53
Gas Flow Analyzer
Specifications
45
Inspiratory Time (The time of the inspiratory period including the inspiratory hold time.)
Ti 0.01 sec 0 sec to 60 sec 0.02 sec
Inspiratory Hold Time (The time of the End Inspiratory State.)
TiH 0.01 sec 0 sec to 60 sec 1 % or 0.1 sec
Expiratory Time (The time of the expiratory period including the expiratory hold time.)
Te 0.01 sec 0 sec to 90 sec 0.5 % or 0.01 sec
Expiratory Hold Time (The time of the End Expiratory State.)
TeH 0.01 sec 0 sec to 90 sec 0.02 sec
Peak Inspiratory Flow (The maximum of the absolute value of
the Flow Rate measured during the Inspiratory Time.)
PIF 0.01 lpm ±300 lpm ±1.7 % or 0.04 lpm
Peak Expiratory Flow (The maximum of the absolute value of the Flow Rate measured during the Expiratory Time.)
PEF 0.01 lpm ±300 lpm ±1.7 % or 0.04 lpm
[1] Inspiratory pause time >0.5 sec
Table 10. Breath Parameter Accuracy (cont.)
Parameter Abbreviation Resolution Range Accuracy
Page 54
VT650/VT900
Users Manual
46