This service information sheet provides the following information for the T5-600 and T5-1000 Electrical
Testers (hereafter referred to as "the tester").
• Safety information
• Parts and service information
• Specifications
• Cleaning procedure
• Required equipment
• Performance tests
• Parts and accessories list
• Battery replacement procedure
®
For operating instructions, refer to the T5-600/T5-1000 Electrical Tester Instruction Sheet.
Safety Information
Warning
To avoid possible electric shock or personal injury, follow these guidelines:
• Do not use the tester if it is damaged or operati ng abnormal l y. Protection
may be impaired.
• Before each use:
• Make sure the battery door is closed and latched.
• Inspect the tester and test leads. Look for cracks, mi ssing plastic, exposed
metal, or damaged insulation. Replace damaged test leads before using the
tester.
• Verify the tester’s operation by measuring a known voltage.
• Replace the batteries as soon as the low battery i ndi cat or (M) appears.
• Do not use the tester around explosive gas, vapor or dust.
• Do not apply more than the rated voltage, as marked on the tester, between
terminals or between any terminal and earth ground.
• Refer servicing to qualified personnel.
• Use caution when working above 30 V ac rms, 42 V ac peak, or 60 V dc.
• When using the probes, keep your fingers behind the finger guards on the
probes.
• Connect the common test lead before you connect the live t est lead.
Disconnect the live test lead fi rst .
• The display will not display hazardous voltages in display hold. The voltage
indicator LED (
) continues to operate.
PN 686953
March 1998 Rev.1, 9/01
1998-2001 Fluke Corporation. All rights reserved. Printed in U.S.A. Product names are trademarks of their respective companies.
T5-600/T5-1000
Electrical Tester
Parts and Service
The tester is warranted to be free from defects in material and workmanship for two years, while under normal
use. Parts and repairs are warranted for 90 days. For the complete warranty statement, refer to the
T5-600/T5-1000 Electrical Tester Instruction Sheet.
To order parts, or for warranty service, contact Fluke as follows:
USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
Or, visit Fluke’s Web site at www.fluke.com.
Specifications
Accuracy is specified for one year after calibration, at 18 °C to 28 °C (64 °F to 82 °F) with relative humidity
to 90 %. AC conversions are ac-coupled, average responding, and calibrated to the rms value of a sine wave
input. Accuracy specifications are given as follows:
±([ % of reading] + [number of least significant digits])
Temperature coefficient of 0.1 x (specified accuracy)/ °C for <18 °C or >28 °C (<64.4 °F or >82.4 °F)
Serial Numbers 79560000 and Above
General Specifications
CalibrationOne-year calibration cycle.
Maximum Voltage
Between any Terminal
and Earth Ground
Maximum Voltage
Between Current Fork
and Earth Ground
TemperatureOperating: -10 °C to +50 °C (14 °F to 122 °F); Storage: -30 °C to +60 °C (-22 °F to +140 °F)
AltitudeOperating: 2000 m (6562 ft); Storage: 10,000 m (32808 ft)
Relative Humidity0 % to 95 %, 5 °C to 30 °C (41 °F to 86°F); 0 % to 75 %, 30 °C to 40 °C (86 °F to 104 °F);
Battery Type and LifeAA (2); 360 hours continuous with alkaline; 125 hours continuous with zinc chloride
Shock, Vibration1 m drop at -10 °C to + 50 °C (14 °F to 122 °F) per ANSI/ISA-S82.01-1994 and EN 61010-1
T5-600: 600 V rms, Overvoltage Category III, Pollution Degree 2
T5-1000: 1000 V rms, Overvoltage Category III, 600 V RMS, OverVoltage Category IV,
Pollution Degree 2
T5-600 and T5-1000: 1000 V rms, Overvoltage Catergory III, also 600 V rms, Overvoltage
Category IV, Pollution Degree 2, (This dual rating applies to the current fork only.)
0 % to 45 %, 40 °C to 50 °C (104 °F to 122 °F)
1995. Random vibration per MIL-PRF-28800F for a Class 2 instrument
(5 Hz to 55 Hz, 3 g maximum)
Surge ProtectionT5-600: 6 kV per IEC 61010, T5-1000: 8 kV per IEC 61010
Enclosure RatingIP 52 per IEC 60529, no vacuum applied
2
Service Information
Serial Numbers 79559999 and Below
General Specifications (continued)
RF Field Specification0.5 % full scale + (specified accuracy) at 3 V/m
SafetyComplies with ANSI/ISA-S82.01-94, UL Classified to IEC 61010, CSA/CAN C22.2 No.1010.1-
600 V rms1000 V rms1 MΩ, <100 pF ac-coupled
600 V rms1000 V rms1 MΩ, <100 pF
600 V rms1000 V rms
Open Circuit Test VoltageShort Circuit Current
Input impedance (nomial
CAT III protects against transients in a fixed equipment installation such as a distribution panel, and lighting systems in
large buildings.
CAT IV protects against transients from a primary supply such as an electricity meter or an overhead or underground utility
service.
Serial Numbers 79559999 and Below
General Specifications
CalibrationOne-year calibration cycle.
Maximum Voltage
Between any Terminal
and Earth Ground
TemperatureOperating: -10 °C to +50 °C (14 °F to 122 °F); Storage: -30 °C to +60 °C (-22 °F to +140 °F)
AltitudeOperating: 2000 m (6562 ft); Storage: 10,000 m (32808 ft)
Relative Humidity0 % to 95 %, 5 °C to 30 °C (41 °F to 86°F); 0 % to 75 %, 30 °C to 40 °C (86 °F to 104 °F);
Battery Type and LifeAA (2); 400 hours continuous with alkaline; 200 hours continuous with zinc chloride
T5-600: 600 V rms, Overvoltage Category III
T5-1000: 1000 V rms, Overvoltage Category III
0 % to 45 %, 40 °C to 50 °C (104 °F to 122 °F)
3
T5-600/T5-1000
Electrical Tester
General Specifications (continued)
Shock, Vibration1 m drop at 15 °C to 35 °C (59 °F to 95 °F) per ANSI/ISA-S82.01-1994 and EN 61010-1 1993.
Sinusoidal vibration per MIL-PRF-28800F for a Class 2 instrument
(5 Hz to 55 Hz, 3 g maximum)
Surge ProtectionT5-600: 6 kV per IEC 1010-1, 1990-09: T5-1000: 8 kV per IEC 1010-1, 1990-09
Enclosure RatingIP 52 per IEC 60529, no vacuum applied
RF Field Specification0.5 % full scale + (specified accuracy) at 3 V/m
SafetyComplies with ANSI/ISA-S82.01-94 for use in overvoltage category III (CAT III) locations,
UL3111, CSA/CAN C22.2 No.1010.1-92, and EN61010-1 1993.
600 V rms1000 V rms1 MΩ, <100 pF ac-coupled
600 V rms1000 V rms1 MΩ, <100 pF
600 V rms1000 V rms
Open Circuit Test VoltageShort Circuit Current
Input impedance (nomial
CAT III protects against transients in a fixed equipment installation such as a distribution panel, and lighting systems in
large buildings.
CAT IV protects against transients from a primary supply such as an electricity meter or an overhead or underground utility
service.
Cleaning the Tester
WWarning
To avoid electric shock or damage to the tester, never allow w at er i nsi de t he
case. To avoid damaging the tester’s case, never use solvents on the tester.
If the tester requires cleaning, wipe it down with a cloth that is lightly dampened with water or a mild
detergent. Do not use aromatic hydrocarbons, chlorinated solvents, or methanol-based fluids when wiping
down the tester.
4
Service Information
Equipment Required for Performance Tests
Equipment Required for Performance Tests
The following equipment is required for performance tests:
• Fluke 5500A Multi-Product Calibrator, or equivalent
• 0 V to 5 V adjustable dc power supply
• Approximately 8 m (26 ft) of #14 single-conductor magnet wire wound into a butterfly-shaped coil.
Figure 1 shows how to make the coil.
Performance Tests
Use the following procedures to verify the tester’s performance. If the tester fails any of the tests, return it to
Fluke for calibration or repair.
Testing the AC Current Function
The tests in this section require the butterfly coil shown in Figure 1. Note that the 10 conductors in the middle
of the coil cause the tester to read 10 times more current than the calibrator supplies.
1. Put one edge of the butterfly coil in a vice to hold the coil during testing.
2. Set the calibrator to the current and frequency given in Step 1 of Table 1.
3. Connect the butterfly coil to the calibrator.
4. Set the tester to the amperage function.
5. Place the tester’s current fork around the middle of the coil so that the alignment marks are centered and
perpendicular to the wire bundle, as shown in Figure 1.
6. Verify that the tester reads within the display limits shown for Step 1 in Table 1.
7. Apply the currents and verify the tester’s readings for the remaining steps in Table 1.