Fluke RUSKA 2456-LEM User Manual

Page 1
RUSKA 2456-LEM
Laboratory Environment Monitor
PN 3952272 November 2010 © 2010 Fluke Corporation. All rights reserved. Printed in USA. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
Page 2
LIMITED WARRANTY AND LIMITATION OF LIABILITY
Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days. This warranty extends only to the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries, or to any product which, in Fluke's opinion, has been misused, altered, neglected, contaminated, or damaged by accident or abnormal conditions of operation or handling. Fluke warrants that software will operate substantially in accordance with its functional specifications for 90 days and that it has been properly recorded on non-defective media. Fluke does not warrant that software will be error free or operate without interruption.
Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price. Fluke reserves the right to invoice Buyer for importation costs of repair/replacement parts when product purchased in one country is submitted for repair in another country.
Fluke's warranty obligation is limited, at Fluke's option, to refund of the purchase price, free of charge repair, or replacement of a defective product which is returned to a Fluke authorized service center within the warranty period.
To obtain warranty service, contact your nearest Fluke authorized service center to obtain return authorization information, then send the product to that service center, with a description of the difficulty, postage and insurance prepaid (FOB Destination). Fluke assumes no risk for damage in transit. Following warranty repair, the product will be returned to Buyer, transportation prepaid (FOB Destination). If Fluke determines that failure was caused by neglect, misuse, contamination, alteration, accident, or abnormal condition of operation or handling, including overvoltage failures caused by use outside the product’s specified rating, or normal wear and tear of mechanical components, Fluke will provide an estimate of repair costs and obtain authorization before commencing the work. Following repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges (FOB Shipping Point).
THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
Fluke Corporation P.O. Box 9090 Everett, WA 98206-9090 U.S.A.
Fluke Europe B.V. P.O. Box 1186 5602 BD Eindhoven The Netherlands
11/99
To register your product online, visit register.fluke.com
Page 3
Table of Contents
Chapter Title Page
1 Function ............................................................................................... 1-1
General Function................................................................................................ 1-1
How to Contact Fluke ........................................................................................ 1-1
Safety Information ............................................................................................. 1-2
Symbols Used in this Manual ............................................................................ 1-2
Components ....................................................................................................... 1-2
PC Requirements ............................................................................................... 1-2
LEM Basics........................................................................................................ 1-3
Calibration Data Storage ............................................................................... 1-3
Main Comm Port ........................................................................................... 1-3
Power............................................................................................................. 1-4
2 Getting Started .................................................................................... 2-1
Introduction........................................................................................................ 2-1
Interconnect ....................................................................................................... 2-1
Software Installation .......................................................................................... 2-1
Software Setup................................................................................................... 2-1
Help Menus ................................................................................................... 2-1
Serial Port Setup............................................................................................ 2-1
Communications Protocol ............................................................................. 2-2
LEM Addressing ........................................................................................... 2-2
LEM .......................................................................................................... 2-2
LEMCAL Software................................................................................... 2-3
Selecting a Pressure Source........................................................................... 2-3
Selecting Units............................................................................................... 2-3
Limits............................................................................................................. 2-3
Setup.......................................................................................................... 2-4
Error Limit Logging.................................................................................. 2-4
Password Protection ...................................................................................... 2-4
Calibration Password ................................................................................ 2-4
Program Exit Password ............................................................................. 2-5
Viewing Real Time Data ................................................................................... 2-6
Logging Data ..................................................................................................... 2-6
Setup.............................................................................................................. 2-6
i
Page 4
RUSKA 2456-LEM
Users Manual
3 RUSKA 2456 Operation with the LEM................................................ 3-1
4 Remote Pressure Capture .................................................................. 4-1
Intervals ......................................................................................................... 2-7
File Names..................................................................................................... 2-7
Default Location............................................................................................ 2-7
File Verification............................................................................................. 2-7
Multiple LEMs................................................................................................... 2-8
Cabling .......................................................................................................... 2-8
Addressing..................................................................................................... 2-8
Communications ................................................................................................ 3-1
Winprompt Interface.......................................................................................... 3-1
RUSKA 2456 and 2465 Upgrade for LEM Compatibility ................................ 3-1
Remote Pressure Capture................................................................................... 4-1
Use with a RUSKA Portable Pressure Gauge 7220 or 6220 ............................. 4-2
Initialization String ............................................................................................ 4-2
RUSKA 6220................................................................................................. 4-2
RUSKA 7220 Using SCPI Protocol .............................................................. 4-2
RUSKA 7220 Using 6220 Emulation Mode ................................................. 4-2
Pressure String ................................................................................................... 4-2
RUSKA 6220 and 7220 in Emulation Mode................................................. 4-2
RUSKA 7220 in SCPI Mode......................................................................... 4-2
Cabling............................................................................................................... 4-3
Connecting the RUSKA 6220 to the LEM Pressure Serial Port ................... 4-3
RUSKA 7220 to the LEM Pressure Serial Port............................................. 4-3
Use with Other RS232 Output Transducers....................................................... 4-3
Protocol ......................................................................................................... 4-3
Scaling ........................................................................................................... 4-4
Communication Port Configuration .............................................................. 4-4
5 Calibration............................................................................................ 5-1
Span ................................................................................................................... 5-1
Pressure - On Board Sensor........................................................................... 5-1
Temperature................................................................................................... 5-1
Zeroing .......................................................................................................... 5-1
Humidity........................................................................................................ 5-1
Calibration Data................................................................................................. 5-2
6 Connector Definition........................................................................... 6-1
Power Plug......................................................................................................... 6-1
DB9 Connectors................................................................................................. 6-1
Pressure Port ...................................................................................................... 6-1
7 Air Density Calculation ....................................................................... 7-1
LEM Internal Calculation vs. RUSKA PC Software Calculation ..................... 7-1
Air Density Available Directly from LEM........................................................ 7-1
8 Specifications ...................................................................................... 8-1
Sensors............................................................................................................... 8-1
9 Remote Commands............................................................................. 9-1
ii
Page 5
Contents (continued)
Remote Commands............................................................................................ 9-1
DruckBus Binary Standard............................................................................ 9-1
DruckBus Compatibility Standard................................................................. 9-1
Terminology: ................................................................................................. 9-1
Timing ........................................................................................................... 9-2
Global Command Responses......................................................................... 9-2
DruckBus Binary Mode Commands.............................................................. 9-2
@ - Set Address............................................................................................. 9-2
A – Read ADC Counts .................................................................................. 9-3
C – Configure Instrument.............................................................................. 9-3
D – Read Air Density .................................................................................... 9-4
E – Read EEPROM ....................................................................................... 9-4
K - Calibrate .................................................................................................. 9-4
P – Set Initialization and Remote Pressure Strings ....................................... 9-4
R - Read ADC Reference Values .................................................................. 9-5
S - Read Remote Pressure with Displayed Precision .................................... 9-5
V - Get Version Info...................................................................................... 9-5
W - Write EEPROM...................................................................................... 9-5
Z - Zero.......................................................................................................... 9-5
RS485 Multi-Drop Configuration...................................................................... 9-6
iii
Page 6
RUSKA 2456-LEM
Users Manual
iv
Page 7
List of Tables
Table Title Page
1-1. Symbols.................................................................................................................. 1-2
4-1. RUSKA 6220 25 Pin RS-232 Connector ............................................................... 4-3
4-2. RUSKA 6220 DB25 and LEM DB9 Connections ................................................. 4-3
6-1. DB9 Pinout............................................................................................................. 6-1
v
Page 8
RUSKA 2456-LEM
Users Manual
vi
Page 9
List of Figures
Figure Title Page
1-1. LEM General View ................................................................................................ 1-3
1-2. LEM Communication Switch................................................................................. 1-4
2-1. Serial Port Setup..................................................................................................... 2-2
2-2. View Coefficients................................................................................................... 2-2
2-3. Options ................................................................................................................... 2-3
2-4. Units ....................................................................................................................... 2-3
2-5. Limits ..................................................................................................................... 2-4
2-6. Select Password...................................................................................................... 2-5
2-7. Change Exit Password............................................................................................ 2-5
2-8. Main Screen ........................................................................................................... 2-6
2-9. Data Logging.......................................................................................................... 2-7
4-1. Remote String Selection......................................................................................... 4-1
4-2. Remote String Dialog............................................................................................. 4-2
5-1. View Coefficients................................................................................................... 5-2
9-1. RS485 Multi-Drop Configuration .......................................................................... 9-6
vii
Page 10
RUSKA 2456-LEM
Users Manual
viii
Page 11
General Function
This manual contains operation instructions for the RUSKA 2456 Laboratory Environment Monitors. The RUSKA 2456-LEM monitors the fundamental environmental parameters; Temperature, Pressure, and Relative Humidity. The sensor data is collected, processed, and made available on the selectable, RS232 or RS485, communication port. The data logging functions allow for easy collection, verification of data integrity and use of historical data.
The LEM can be used as a stand alone device or in conjunction with other monitors. The monitor can also be used as the Air Density Monitor for the RUSKA 2456 and 2465 models. If higher accuracy pressure data is needed, then a secondary RS232 communications port allows for use with RUSKA 7220 / 6220 Portable Pressure Gauges and other pressure standards.
Chapter 1
Function
How to Contact Fluke
To order accessories, receive operating assistance, or get the location of the nearest Fluke distributor or Service Center, call:
Technical Support USA: 1-800-99-FLUKE (1-800-993-5853)
Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31-402-675-200
China: +86-400-810-3435
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
Or, visit Fluke's website at www.fluke.com
To register your product, visit http://register.fluke.com
To view, print, or download the latest manual supplement, visit
http://us.fluke.com/usen/support/manuals
.
.
.
1-1
Page 12
RUSKA 2456-LEM
Users Manual
Safety Information
Symbols Used in this Manual
W Warning
Pressurized vessels and associated equipment are potentially dangerous. The apparatus described in this manual should be operated only by personnel trained in procedures that will assure safety to themselves, to others, and to the equipment.
W Caution
Do not use hydrocarbon lubricants. Use only supplied lubricant, unless otherwise specified in this manual. Always use replacement parts specified by Fluke.
W Warning
When any maintenance is performed, turn off power and remove power cord.
In this manual, a Warning identifies conditions and actions that pose a hazard to the user. A Caution identifies conditions and actions that may damage the Laboratory Environment Monitors.
Symbols used on the Laboratory Environment Monitors and in this manual are explained in Table 1-1.
Symbol Description
B AC (Alternating Current)
J Earth Ground
W Important Information: refer to manual
~
Components
The LEM product comes with the following accessories:
Power Plug (input: 110-240 Vac, 50/60Hz) Part #62-315
Communication Cable (F/F 9DSUB) Part # 8-823
Software Packet Part # 2456-LEMCAL
Manual Part # LEM-1D01
PC Requirements
Windows ™ 95, 98, 2000, or XP 130 MHz Pentium or higher
Table 1-1. Symbols
Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information.
1-2
Page 13
Function
LEM Basics 1
LEM Basics
Temperature Sensor
Local Pressure Port
Serial Pressure Input
Humidity Sensor
Main Comm. Port
Power Indicator
Calibration Data Storage
The calibration data is stored in the LEM. Refer to Chapter 5 for information on the Calibration process and how to access the coefficients.
Main Comm Port
The main communication port is user selectable for RS232 or RS485, using a switch located on the bottom of the LEM (see Figure 1-2). Cycle the power to activate a change in the communication port.
The provided communication cable allows for the monitor to be directly connected to a standard PC - DB9 - RS232 port.
Power Input
Figure 1-1. LEM General View
gky01.eps
1-3
Page 14
RUSKA 2456-LEM
Users Manual
Power
RS232 - C1
Figure 1-2. LEM Communication Switch
RS458 - C2
gky02.eps
Optional USB cables are available: Part # 8-825 - USB to RS232 (requires Windows 98 or higher).
The Power LED will illuminate when power is applied through either the power plug or the main communication port.
Refer to Chapter 6, Connector Specification, for connector and pin out details.
1-4
Page 15
Introduction
The following explains how to setup PC operation with RUSKA LEMCal software and a single LEM.
Interconnect
The standard interface cable for the LEM is a DB9 F/F cable. The green indicator light should come on when the power plug is connected.
Chapter 2
Getting Started
Software Installation
1. Insert CD into drive.
The AutoRun function should begin the installation.
2. If the AutoRun function does not begin then Select Start and Select Run.
Type D:\Setup (where D is your CDROM directory)
3. Double Click on the new LemCal icon in the RUSKA Instrument Program group.
At this point, the System should display data. If not, or if customization is desired then continue with the section below.
Software Setup
Help Menus
The LEM software contains extensive help menus that provide detailed information about each function.
Serial Port Setup
From the menu select Setup | Communication.
2-1
Page 16
RUSKA 2456-LEM
Users Manual
The default settings are shown. Change the Comm Port to match the host computer. The Baudrate, Databits, Parity and Stopbit defaults should be set as shown in Figure 2-1.
Communications Protocol
See Appendix, Remote Commands, for a description of the Communication Protocol.
LEM Addressing
From the Menu select Setup | Options.
LEM
The factory default address is 33.
This address can be changed to any address between 0 and 99.
The 0 address is the global address. When a command is issued with the 0 address all instruments on the communication bus will respond.
To view the address and Calibration Coefficients go to Calibrate | View Coefficients (Figure 2-2).
To change the address, double click on the Address Window and modify, answer “Yes” to editing coefficients, and enter a new address. Click on the “Write coefficients to Sensor” icon to write the new address to the LEM. This will take several seconds.
This function can be password protected. Refer to the section Password Protection.
Figure 2-1. Serial Port Setup
gky04.bmp
2-2
Figure 2-2. View Coefficients
gky05.bmp
Page 17
Getting Started
Software Setup 2
LEMCAL Software
The factory default address is 33. The LEM and LEMCAL Software default to this Address. The LEMCAL Software needs the address set in the SETUP/OPTIONS screen of the LEM with which it will communicate. If this address is set to 0 then all LEMs on the communication bus will respond.
Selecting a Pressure Source
From the Menu select Setup | Options.
Select the Local or Remote pressure sources for each enabled LEM by clicking on the appropriate column. The Local sensor location is shown in Chapter 1, Figure 1-1, and specified in Chapter 8. Refer to Chapter 4 for a description of the Remote Pressure function.
Selecting Units
From the Menu select Setup | Units.
Figure 2-3. Options
Figure 2-4. Units
gky06.bmp
gky07.bmp
Select desired units from the available choices for each environmental parameter. Changing units will also change the data written to the logfile. Refer to Chapter 2, Logging Data.
Limits
From the Menu select Setup | Limits.
2-3
Page 18
RUSKA 2456-LEM
Users Manual
Setup
Figure 2-5. Limits
gky08.bmp
Select values for the upper and lower limits for Pressure, Temperature, and Relative Humidity.
The Serial Number for the desired LEM is selected in the box in the top, left corner. Click the arrow, and the available LEMs will be displayed. When a specific instrument is selected, the current limits are displayed, and may be edited.
The Audio Alarm is enabled by checking the Audio Alarm checkbox. This alarm generates a standard beep using the computer speaker.
The Visual Alarm is enabled by checking the Visual Alarm checkbox. This alarm is displayed on the main screen by changing the background color of the Serial Number to yellow whenever a limit is exceeded. The background will remain yellow until it is either disabled or reset. Double-clicking the Serial Number will reset the visual alarm, but does not disable it.
The Audio and Visual alarms can be set by double-clicking the message panel at the lower-left corner of the main screen, until the desired setting is selected.
The limit values for one instrument may be copied to another instrument's limit values, by using the Copy and Past buttons. Closing the Limit form will erase values from the copy buffer.
Pressing OK will check the new values for out-of-range and inconsistent values. If a problem is detected, an error message is displayed.
Pressing CANCEL will leave the Limits unchanged.
Pressure DEFAULT will return all limits, for the selected instrument, back to default values.
Error Limit Logging
Crossing a limit value, will trigger a record being entered in the log file, if enabled. Refer to Chapter 2, Logging Data.
Password Protection
Two types of password protection are supported—Calibration and Program Exit.
Calibration Password
From the Menu select Setup | Password | LEM
2-4
Page 19
Getting Started
Software Setup 2
The password protection will prevent the changing of the calibration data and address. The default password is “0” (zero). When the password is “0” the password protection is turned off.
When setting the password for the first time, “0” must be entered as the Old Password.
The password can be reset to “0” as long as the current password is known.
The password can be up to eight alpha/numeric characters long and is case sensitive.
Program Exit Password
From the Menu select Setup | Password | Program Exit
Figure 2-6. Select Password
Figure 2-7. Change Exit Password
gky09.bmp
gky10.bmp
This password protection will prevent closing LEMCal software. This is most useful when the Logfile function is being used, and a record without gaps is desired. The default password is “0” (zero). When the password is “0” the password protection is turned off.
When setting the password to the first time, “0” must be entered as the Old Password.
The password can be reset to “0” as long as the current password is known.
The password can be up to eight alpha/numeric characters long and is case sensitive.
2-5
Page 20
RUSKA 2456-LEM
Users Manual
Viewing Real Time Data
The Memo line allows a user to enter a descriptive to identify each LEM. This description is associated with the Serial Number and is stored in the *.ini file on the host computer.
The green indicator at the end of each line identifies if the LEM data is being logged. See Chapter 2, Data Logging
*** If the Data Line turns yellow then the data being displayed is out of the following
ranges: 17 0% < relative humidity < 100% 10 psi < pressure < 16.7 psi
*** If the Data Line turns red then communication has been lost with the LEM.
Logging Data
The Data Logging function captures Temperature, Pressure, Relative Humidity and Air Density (Chapter 7.1 - NBS Pub. 700-1)
Setup
From the Menu select Log (Figure 2-9).
Figure 2-8. Main Screen
o
C < temperature <29 oC
gky11.bmp
2-6
Page 21
Getting Started
Logging Data 2
The Module number, in the left hand column, is referenced to each LEM as defined in Figures 2-3 and 2-62-6. To begin logging data for a particular LEM, select the Enable button.
Intervals
LEM data can be logged into a Day, Week, or Monthly file.
Timing and data information are based upon the clock of the host computer.
Day: Readings are taken at 1 minute intervals. The active file is saved and a new file is crated at midnight of each day.
Weekly: Readings are taken at 5 minute intervals. The active file is saved and a new file is created at midnight of each
Monthly: Readings are taken at 15 minute intervals.
File Names
The file names are generated automatically using the following formats:
Day: SerialNumber_Year_Day of Year.LOG Example: SN054323_Y2002_D137.LOG
Figure 2-9. Data Logging
gky12.bmp
Saturday.
The active file is saved and a new file is created at midnight of the last day of each month.
Week: SerialNumber_Year_Week of Year.LOG Example: SN054323_Y2002_W47.LOG
Month: SerialNumber_Year_Month of Year.LOG Example: SN054323_Y2002_M05.LOG
Default Location
By default, the log files are placed in
C:\Program Files \ RUSKA Instruments \ LEM Cal \ LEMDataLog \ ...
To change this location, select <Browse Dir> button. Figure 2-9.
File Verification
From the Data Logging screen (Figure 2-9) select Verify Log.
2-7
Page 22
RUSKA 2456-LEM
Users Manual
Multiple LEMs
Cabling
Addressing
Select the *.LOG file that needs to be verified. Upon completion of the verification a notice of verification will be displayed or a file *.ERR will be created and saved in the LOG directory. This file will identify which lines of the *.LOG file have been corrupted or altered. If any errors are detected then the corrupted lines are stored in a *.ERR file.
The LEMCal software supports up to 8 LEM modules at the same time when using RS485. Refer to Chapter 1, Main Comm Port, for selecting RS485 mode.
For Multi-Drop Systems refer to Appendix A for available components. For custom cabling, refer to Chapter 6, DB9 Connections, for the pin-out of the Main Communication Port on the LEM.
Since the default address for each module is “33”, each module will need to have its address set to a unique number between 0 and 100. This must be done with only one non-unique module on the communication bus at a time. Refer to Chapter 2, LEM Addressing. Each LEM must have a unique address and LEMCAL Software must be told which addresses are active on the communication bus.
2-8
Page 23
RUSKA 2456 Operation with the LEM
Communications
Set the communications interface to RS485 for use with the 2456 and 2465. Refer to Chapter 1, Main Comm Port, for LEM communication setup. Power for the LEM is supplied from the 2456 or 2465, over the communications cable. Use a standard DB9M / DB9F cable.
Winprompt Interface
Refer to the WinPrompt manual for detailed instructions. WinPrompt Version 2.12 or later is required. WinPrompt software calculates mass to pressure and pressure to mass values, and automatically displays real time piston gauge parameters measured by the Model 2456 Deadweight Gauge Monitor or the Model 2465 AutoFloat Controller.
Chapter 3
RUSKA 2456 and 2465 Upgrade for LEM Compatibility
To upgrade a 2456/2465 from the original Air Density Model to the LEM, use part number 2465-200-KIT1 for all 2465's and the Single Channel 2456. Use part number 2465-200-KIT2 for the Dual Channel 2456.
3-1
Page 24
RUSKA 2456-LEM
Users Manual
3-2
Page 25
Remote Pressure Capture
Remote Pressure Capture
From the Menu select Setup | Options (refer to Figure 4-1). Set the Pressure Source as Remote. Click the extension arrows to the right of the selected LEM. The Remote String Dialog box will appear.
Chapter 4
Figure 4-1. Remote String Selection
gky13.bmp
4-1
Page 26
RUSKA 2456-LEM
Users Manual
Use with a RUSKA Portable Pressure Gauge 7220 or 6220
Figure 4-2. Remote String Dialog
Initialization String Transmitted from LEM to remote pressure sensor to initialize
the remote sensor.
Pressure String Transmitted from LEM, on a continuous basis, to retrieve
pressure readings.
Remote Pressure Units Specifies the units that are being returned from the remote
pressure sensor. The selection of the User units requires the entry of a conversion factor. The User units, when multiplied by the conversion factor, result in kPa. (Refer to Chapter 9 for conversion factors.)
Initialization String
RUSKA 6220
‘UN,3\r’ - This sets the units to kPa.
RUSKA 7220 Using SCPI Protocol
‘UNIT:PRES KPA\r’
This sets the 7220 to kPa units.
RUSKA 7220 Using 6220 Emulation Mode
‘:SYST:LANG "6220"\rUN,3\r’
This sets the 7220 to 6220 emulation mode and to kPa units.
gky14.bmp
Refer to the 6220 and 7220 manuals for other options.
Pressure String
RUSKA 6220 and 7220 in Emulation Mode
The pressure command string is ‘PA\r’.
RUSKA 7220 in SCPI Mode
The pressure command string is ‘:MEAS?’.
4-2
Page 27
Remote Pressure Capture
Cabling 4
Cabling
Connecting the RUSKA 6220 to the LEM Pressure Serial Port
Table 4-1. RUSKA 6220 25 Pin RS-232 Connector
PIN DESCRIPTION
2 Transmitted data (to PPG)
3 Receive Data (from PPG)
4* RTS (must be set for PPG to transmit)
5* CTS (always set by PPG)
7 GND
*Connect Pins 4 & 5 together.
Table 4-2. RUSKA 6220 DB25 and LEM DB9 Connections
6220
DB25
2 3
3 2
7 5
LEM
DB9
Refer to 6220 Manual and Chapter 6, DB9 Connections, of this manual for additional information.
RUSKA 7220 to the LEM Pressure Serial Port
Connecting the RUSKA Model 7220 to the LEM pressure serial port requires a Null modem adapter or Null modem cable DB9 M / DB 9 F. Only Pins 2 and 3 must be swapped in the Null Modem cable.
Use with Other RS232 Output Transducers
Protocol
The LEM can retrieve pressure data from other "smart" pressure sensors, as long as the return string can be parsed. The string returned to the LEM must be in ASCII characters.
The LEM breaks down the returned string as follows:
1. Delete Prefix—all non-numeric characters are deleted up to the first numeric character.
2. The Exponent character (E or e) is located.
3. A one or two digit numeric exponent is read following the Exponent character.
Examples of valid returned strings:
101.23
101.23kPa
1.0123E02 PA, 1.0123e02 kPa PA, 101.23
4-3
Page 28
RUSKA 2456-LEM
Users Manual
Scaling
Communication Port Configuration
Once the raw value is retrieved from the string then units or a scaling factor must be given. Refer to Chapter 4, Remote Pressure Units.
This configuration applies only to the Remote pressure Input and is set as follows:
BAUD RATE - 2400
DATABITS - 8
PARITY - None
STOPBITS - 1
4-4
Page 29
Span
Pressure - On Board Sensor
Calibration Point 1: 10.5 +/-0.25 psia
Calibration Point 2: 15.0 +/-0.25 psia
Temperature
Calibration Point 1: 17 +/- 2 degC
Calibration Point 2: 29 +/- 2 degC
Chapter 5
Calibration
Zeroing
Temperature measuring devices are impacted by self heating. The rise in temperature, measured by the sensor, is a function of the amount of heat generated by the probe and the ability of the probe to dissipate this heat. Air flow around the probe improves the heat dissipation and thus lowers the rise in temperature. When a temperature probe is calibrated in a high flow chamber and used in a low flow environment, then the error due to self heating is maximized. This error can be minimized by zeroing the sensor in the operating environment with respect to an appropriate standard.
The LEM software allows the technician to perform a zero offset. The zero offset is prompted following the span calibration. After the span calibration is finished, the high air flow chamber can be turned off and the LEM can be zeroed in a low air flow environment that is typical of its operating environment.
The reference probe should be placed at a distance of 3 or more inches in front of the LEM temperature sensor.
Both the span calibration and zeroing operation are password protected.
Humidity
The humidity calibration must be done after the Temperature calibration. The two calibration points must be different by at least 25%RH.
A suggested method:
Point 1: applied vacuum -> 0%RH
Point 2: ambient humidity (25%RH or higher)
5-1
Page 30
RUSKA 2456-LEM
Users Manual
Calibration Data
In order to view the calibration coefficients, select Calibrate | View Coefficients.
Figure 5-1. View Coefficients
gky15.bmp
The menu bar selections allow the coefficients to be read and written to disk and read and written to the LEM, assuming that the proper password is known.
5-2
Page 31
Power Plug
The power plug is a barrel type receptacle:
ID = 2.5 mm, OD = 5.5 mm, Depth = 10.0 mm, Inside - Positive
DB9 Connectors
Chapter 6
Connector Definition
Table 6-1. DB9 Pinout
Pin #s DB9
1* +PWR IN +PWR IN NC
2** A TX RX
3** A RX TX
4 GND-PWR GND-PWR GND
5 GND-SIG GND-SIG GND
6 NC NC NC
7 NC NC NC
8** B RX NC
9 GND-PWR GND-PWR GND
* Supply pin (PIN 1 DB9 female) is fused with re-settable fuses. Maximum operating current is
50 mA.
** Pin 3 of the primary connector (DB9 male) is switched between Pins 2 and 8. This allows
compliance with RUSKA ER-3097 : "DB9-232/485-Pin Out Standard".
Pressure Port
The pressure port is a friction fit - press on tube connection.
Motorola
®
package definition CASE 482A-01
DB9 male
RS 485
DB9 male
RS 232
DB9 female
(RS 232)
Use RUSKA tubing, part # 86-1011 - 3mm ID tubing or equivalent.
6-1
Page 32
RUSKA 2456-LEM
Users Manual
6-2
Page 33
Air Density Calculation
LEM Internal Calculation vs. RUSKA PC Software Calculation
The Air Density calculation available directly from the LEM is a best fit curve approximation as described below in Chapter 7, Air Density Available Directly from LEM.
The Air Density output available from RUSKA PC Software (WinPrompt, LEM PC, etc.) uses the definition available per National Bureau of Standards, NBS Special Publication 700-1, Industrial Measurement Series (November 1984).
Chapter 7
Air Density Available Directly from LEM
Air density is available directly from LEM (without LEMCAL or WinPrompt Software). T = Temperature (16-bit integer, scaled °C*100)
H = True RH (16-bit integer, scaled %*100)
P = Pressure (16-bit integer, scaled kPa*100)
D = Density (16-bit integer, g/m
PD
4916*
 
Formula is valid for temperatures 18-28
3
)
T
 
T
Note
2096*
2
+
o
C.
=
 
15.273
H
**20
 
46460
1616
2
7-1
Page 34
RUISKA 2456-LEM
Users Manual
7-2
Page 35
Sensors
Operating Range:
Temperature: 18-28
Pressure: 690-1070 mbar (20.4 - 31.6 inHg)
Humidity: 5-95% RH
Performance:
Precision
Temperature: +/-0.1
Chapter 8
Specifications
o
C
o
C
Pressure: +/-1.4 mbar (+/-0.04 inHg)
Humidity: +/-3% RH
Total Uncertainty:
Temperature: +/-0.5
Pressure: +/-2.7 mbar (+/-0.08 inHg)/year
Humidity: +/-10% RH/year
Notes:
(1) Precision is defined as the combined effects of linearity, repeatability and
hysteresis.
(2) Expression of accuracy (uncertainty) conforms with the recommendation of the
ISO Guide to the Expression of Uncertainty in Measurement and includes RSS of precision, stability, temperature effects, and the calibration standard to 2 sigma (95%). The expanded uncertainty in the standards utilized by RUSKA to calibrate the LEM is typically 0.1
Storage Environment:
-20 - 70 ºC; 0 - 95% relative humidity, non-condensing
Electrical Power:
10 - 30 Volt DC input; 250mA max
o
C, 2.4% RH, 1.4e-3 kPa.
o
C/year
8-1
Page 36
RUSKA 2456-LEM
Users Manual
Data Update Rate:
New Data available at 1 second intervals. Includes Temperature, Humidity, Pressure, Air Density.
Available Units:
from Module : ºC x100, %RH x100, kPa x100, g/m
from *.dll: ºC, %RH, kPa, g/m
3
from PC Software : ºC, ºF;
%RH;
mmHg, cmHg, inHg, kPa, hPa, mbar, psi, kg/cm
3
g/cm
, kg/m3, lb/in3
Warm Up:
Typically 30 minutes.
Placement:
The proper orientations for the LEM are
3
2
;
1. Set flat on the table, sitting upon the rubber feet
2. Hanging on the wall with the temperature sensor pointing down.
Calibration Period:
1 year
8-2
Page 37
Remote Commands
There are two general standards supported.
Chapter 9
Remote Commands
The Binary protocol
requires fewer bytes, therefore requiring less time to transmit each
message.
The Compatibility protocol
requires more bytes per message, but allows the module to
co-exist on the same RS485 bus with other DruckBus compatible components.
DruckBus Binary Standard
Command:
&<Addr><Size><Cmd>[<Parm1>[<Parm2>...]]<Chk>
Reply:
%<Addr><Size><Cmd><Parm1>[<Parm2>...]<Chk>
Example: (hexadecimal bytes)
26 01 01 56 70 25 01 05 76 02 03 04 CB 99
DruckBus Compatibility Standard
Command:
$<Addr><Size><Cmd>[<Parm1>[<Parm2>...]]<Chk>CR
Reply:
!<Addr><Size><Cmd><Parm1>[<Parm2>...]<Chk>CR
Example:
$01015670cr !010576020304CB99cr
Terminology:
<Addr> Ascii 2 characters specifying the address (00 to 99, 00=Global).
Binary 1 byte specifying the destination address of the command
<Size> Binary Number of bytes after the size byte not including the
Replies always specify the address of the unit, never 00.
or the source address of the reply (00 to FF, 00=Global). Replies always specify the address of the unit, never 00.
<Chk>
9-1
Page 38
RUSKA 2456-LEM
Users Manual
Timing
<Cmd> Ascii 2 Ascii characters specifying command.
Binary 1 Ascii byte specifying command (Uppercase on
command, lowercase on reply).
<Parm x> Ascii Variable-length Ascii formatted parameters.
Binary Fixed-length binary parameter.
<Chk> Binary 1 binary byte LRC (exclusive-or of all bytes including start)
CR Ascii Carriage Return required
Compatibility Carriage Return required
<start> Start Binary mode
Command = '&' ; Reply = '%'
Compatibility mode
Command = '$' ; Reply = '!'
Slaves must wait at least 1 character time before replying (for RS-485 line turnaround) and must reply within 50 ms.
Global Command Responses
Units respond to Zero, the global address, the same as their assigned address. This means that global addressing can be used when only 1 unit is on the bus.
DruckBus Binary Mode Commands
(Commands may be translated to Compatibility Mode for transmission)
@ - Set Address
Command: <start><addr><Size>@<NewAddr><Model><Value><Chk> Reply: None.
Command <Size>: 8
<NewAddr> is an 8-bit unsigned number. <Model> is a 2-byte integer containing the model number, least significant byte first. <Value> is a 4-byte integer specifying a pressure in kPa * 100 or a serial number, least significant byte first. Normally, this command is sent using the global address. This command sets the address of the unit to the specified address with the following conditions:
1. If the model number and value are both zero, all units receiving this command will
set their address.
2. If the model number in the command is not zero and the model number of the unit
matches the specified model, the address is set.
3. If the model number matches and if the pressure specified in the command is not zero
and the full scale pressure of the unit is greater than, or equal to, the specified pressure and the current address of the unit is 254 or 255, the address is set.
4. If the model number is zero, and a non-zero value is specified, and the value matches
the serial number of the unit, the address is set.
These conditions allow for the addresses of each different model to be set individually and the units of the same model to be set by pressure range. If two units have the same model and pressure range the address can still be set by serial number.
9-2
Page 39
Remote Commands
Remote Commands 9
A – Read ADC Counts
Command: <start><addr><Size>A<chk> Reply: <start><addr><Size>a<status><temp><rh><pressure><chk> Command <Size>: 1 Reply <Size>: 10 <status> is three unsigned 8-bit values.
Status1:
Bit 0: ADC channel 0 Extended Input Range Indicator (EXR0) 0: 0 < V 1: V
> V
in
<= V
in
or Vin < 0
ref
ref
Bit 1: ADC channel 0 Sign Indicator (SIG0) 0: V 1: V
< 0
in
> 0
in
Bit 2: ADC channel 1 Extended Input Range Indicator (EXR1)
Bit 3: ADC channel 1 Sign Indicator (SIG1)
Bit 4: ADC channel 2 Extended Input Range Indicator (EXR2)
Bit 5: ADC channel 2 Sign Indicator (SIG2)
Bit 6: ADC channel 3 Extended Input Range Indicator (EXR3)
Bit 7: ADC channel 3 Sign Indicator (SIG3)
Status2:
Bit 0: 1 = Calibration in Progress
Bit 1: 1 = Zero in Progress
Bit 2: 1 = Remote Port Selected (v1.4)
Bit 3-7: Reserved
Status3: Bit 0: 1 = Command Error Bit 1: 1 = EEPROM Write Error Bit 2: 1 = EEPROM Read Error Bit 3: 1 = Power-on Reset Bit 4: 1 = RS232 Receive Error Bit 5: 1 = Calibration Error Bit 6: Reserved Bit 7: 1 = Long (multiple) EEPROM accesses
<temp> is a 16-bit unsigned value <rh> is a 16-bit unsigned value <pressure> is a 16-bit unsigned value
C – Configure Instrument
Command: <start><addr><Size>C<cfgByte><chk> Reply: <start><addr><Size>c<cfgByte><chk>
Command <Size>: 2 Reply <Size>: 2 Defines which pressure source to use. <cfgByte> is an 8-bit unsigned value. <cfgByte> = 0 No pressure source selected = 1 Local pressure = 2 Auxiliary pressure = 3 Remote pressure
= 254 Load LEM Memory from EEPROM = 255 Poll for currently selected pressure source
9-3
Page 40
RUSKA 2456-LEM
Users Manual
D – Read Air Density
Command: <start><addr><Size>D<chk> Reply: <start><addr><Size>d<status><density><chk> Command <Size>: 1 Reply <Size>: 6 <status> is three unsigned 8-bit values. (See Command A) Returned data is a 16-bit unsigned value, least significant byte first.
E – Read EEPROM
Command: <start><addr><Size>E<addr1><chk> Reply: <start><addr><Size>e<data><chk>
Command <Size>: 3 Reply <Size>: 9 <addr1> is 16-bit address, least significant byte first.<data> is 8 bytes read from EEPROM. To read the eeprom, set the high byte of <addr> to 00H To read low memory, set the high byte of <addr> to F0H To read extended memory, set the high byte of <addr> to F1H. Address should be an 8-byte boundary.
K - Calibrate
Command: <start><addr><Size>K<channel><point><value><chk> Reply: <start><addr><Size>k<status><counts><chk>
Command <Size>: 5 Reply <Size>: 6 <channel> and <point> are 8-bit values. <channel> selects channel to be calibrated. P for pressure, H for humidity, T for temperature. <point> is 1 for first point, 2 for second point. <value> is a 16-bit signed integer using scaled (*100) units matching select channel. <status> is three unsigned 8-bit values. (See Command A) <counts> is an unsigned 16-bit value.
P – Set Initialization and Remote Pressure Strings
Command: <start><addr><Size>P<seq><strlen><strseq><chk> Reply: <start><addr><Size>p<status><chk>
Command <Size>: 11 Reply <Size>: 4 <seq> identifies which 8 byte substring is being transmitted. Acceptable Values for the Remote String are 0, 1, 2. Acceptable Values for the Initialization String: 3, 4, 5.
<strlen> is the actual length of the command string, in bytes, that will be transmitted from the LEM to the remote pressure sensor over the software serial port. Acceptable values: 1-24.
<strseq> is an 8 byte section of the string to be transmitted. The string to be transmitted must be 24 characters long, using a pad character to fill out the string. The first eight bytes are transmitted with <seq> set to 0 for the Remote Pressure string and 3 for the Initialization string, the middle eight bytes are transmitted with <seq> set to 1 for the Remote Pressure string and 4 for the Initialization string, and the last eight bytes are transmitted with <seq> set to 2 for the Remote Pressure string and 5 for the Initialization string. All three <seq> substrings for each string must be transmitted in sequential order. Either string may be updated without updating the other.
<status> is three unsigned 8-bit values. (See Command A)
9-4
Page 41
Remote Commands
Remote Commands 9
R - Read ADC Reference Values
Command: <start><addr><Size>R<chk> Reply: <start><addr><Size>r<temp><rh><press><chk>
Command <Size>: 1 Reply <Size>: 7 All values are 16-bit signed integers using scaled (*100) units. <temp> is degrees C <rh> is % <press> is kPa.
S - Read Remote Pressure with Displayed Precision
Command: <start><addr><Size>S<chk> Reply: <start><addr><Size>s<status><press><decimals><chk>
Command <Size>: 1 Reply <Size>: 9 <status> is three unsigned 8-bit values. (See Command A) <press> is 32-bit unsigned integer using scaled (*10^decimals) units. Units is kPa. <decimals> is 8-bit signed integer
V - Get Version Info
Command: <start><addr><Size>V<chk> Reply: <start><addr><Size>v<versionmajor><versionminor><HWVer> <SubModel><model><chk> Command <Size>: 1
Reply <Size>: 7 <versionmajor> is an unsigned 8-bit unsigned value
<versionminor> is an unsigned 8-bit unsigned value <HWVer> is an unsigned 8-bit unsigned value <SubModel> is an unsigned 8-bit unsigned value <model> is a 2-byte flag used to differentiate between software with different command sets.
W - Write EEPROM
Command: <start><addr><Size>W<addr1><data><chk> Reply: <start><addr><Size>w<data><chk>
Command <Size>: 11 Reply <Size>: 9 <addr1> is a 16-bit address, least significant byte first. <data> is 8 bytes read from EEPROM.
To write the eeprom, set the high byte of <addr> to 00H. To write low memory, set the high byte of <addr> to F0H. To write extended memory, set the high byte of <addr> to F1H. Address should be an 8-byte boundary.
Z - Zero
Command: <start><addr><Size>Z<channel><value><chk> Reply: <start><addr><Size>z<status><counts><chk>
Command <Size>: 4 Reply <Size>: 6 <channel> is an 8-bit value. <channel> selects channel to be calibrated. P for pressure, H for humidity, T for temperature. <value> is a 16-bit signed integer using scaled (*100) units matching select channel. <status> is three unsigned 8-bit values. (See Command A) <counts> is an unsigned 16-bit value.
9-5
Page 42
RUSKA 2456-LEM
Users Manual
RS485 Multi-Drop Configuration
RS485 HARDWARE CONFIGURATION
Figure 9-1. RS485 Multi-Drop Configuration
gky03.eps
9-6
Loading...