High pressure liquids an d gases a re pot ential ly haz ardous. En erg y stored in th ese
liquids and gases can be released unexpectedly and with extreme force. High
pressure systems sh ould b e assemb led and op erated only by p ersonnel w ho hav e
been instructed in proper safety practices.
form or by any means, electronic or m echanical, for any purpose, without the express written permission of Fluke Calibration, 4765
East Beautiful Lane, Phoenix, Arizona 85044-5318 USA.
Fluke Calibration makes sincere efforts to ensure the accuracy and quality of its published materials; however, no warranty,
expressed or implied, is provided. Fluke Calibration disclaims any responsibility or liability for any direct or indirect damages
resulting from the use of the information in this manual or products described in it. Mention of any product or brand does not
constitute an endorsement by Fluke Calibration of that product or brand. This manual was originally composed in E ngl ish and was
subsequently translated into other languages. The fidelity of the translation cannot be guaranteed. In case of conflict between the
English version and other language versions, the English version predominates.
Products described in this manual are manufactured under international patents and one or m ore of the following U.S. patents:
6,701,791, 5,142,483, 5,257,640, 5,331,838, 5,445,035. Other U.S. and international patents pending.
Fluke Calibration, FCAL, DH, DHI, PG7000, PG7102, PG7202, PG7302, PG7601, COMPASS, CalTool are trademarks, registered
and otherwise, of Fluke Corporation.
LabVIEW is registered trademark of National Instruments Corporation.
Swagelok is a registered trademark of the Swagelok Company.
1.2.1 General Specifications ............................................................................................................................. 2
1.2.1.1 Embedded Features ............................................................................................................................... 3
1.2.1.2 Ambient and Instrument Condition Measurements ................................................................................. 4
1.2.3 Mass Sets ................................................................................................................................................. 7
2.1.1.2 Mass Set .............................................................................................................................................. 13
2.1.2.2 Mass Set .............................................................................................................................................. 19
2.2 Site Requirements ................................................................................................................................ 22
2.3.1 Preparing for Operation ......................................................................................................................... 23
2.3.1.1 Setting Up the Platform ........................................................................................................................ 23
2.3.1.2 System Pressure Interconnections ....................................................................................................... 24
2.3.1.3 Setting Up a Mass Set .......................................................................................................................... 24
2.3.2 installing a Piston-Cylinder Module into the Platform ......................................................................... 25
2.3.3 Switching a PG7202 Between Gas Operation and Oil Operation ........................................................ 26
2.4 Power Up and Verification ................................................................................................................... 28
2.4.1 Powe r UP ................................................................................................................................................ 28
2.4.2 Check that On-Board Piston-Cylinder Module and Mass Set Information are Correct ...................... 28
2.4.3 Set Local Gravity Value ......................................................................................................................... 28
2.4.11 Additional Precautions to Take Before Making Pressure Measurements .......................................... 31
2.5 Short Term Storage .............................................................................................................................. 32
3. General Operation ........................................................... 33
3.1 Fundamental Operating Principles ..................................................................................................... 33
3.5 Pis to n Po sition ..................................................................................................................................... 38
3.6 Ma ss Loading Protocol ........................................................................................................................ 39
3.7 Main Run Screen .................................................................................................................................. 42
3.8 General Function/Menu Flow Chart .................................................................................................... 43
3.9 Direct Function Keys............................................................................................................................ 44
3.9.1 Direct Function Keys Summary ............................................................................................................ 44
3.9.5.1 First System Run Screen...................................................................................................................... 68
3.9.5.2 Second System Run Screen ................................................................................................................ 69
3.9.11 [ENTER /SET P] from Run Scre e n .......................................................................................................... 80
3.9.11.1 [ENTER/SET P] in Pressure to Mass M ode .......................................................................................... 81
3.9.11.2 [ENTER/SET P] in Mass to Pressure Mode .......................................................................................... 83
3.9.11.3 Commands for Zero Pressure, Ending a Test ....................................................................................... 83
3.9.12 [P OR M] .................................................................................................................................................. 84
3.9.13 [ ] and [], [←] ................................................................................................................................. 84
3.10 [SETUP] Menu ....................................................................................................................................... 86
3.11 [SPECIAL] Menu ................................................................................................................................... 91
3.11.1.5 Select the active piston-cylinder module ............................................................................................... 96
3.11.1.6 Add Mass Set ....................................................................................................................................... 97
3.11.1.7 Edit mass set ......................................................................................................................................101
3.11.1.8 View Mass Set ....................................................................................................................................101
3.11.1.9 Delete Mass Set ..................................................................................................................................101
3.11.1.10 Select Mass Set ..................................................................................................................................102
3.11.1.11 Add Mass Loading Bell ........................................................................................................................102
3.11.1.12 Edit Mass Loading Bell ........................................................................................................................104
3.11.1.13 View mass loading bell ........................................................................................................................104
3.11.1.14 delete mass loading bell ......................................................................................................................104
3.11.1.15 Select Mass Loading Bell ....................................................................................................................105
3.11.5.3 RS232 Self Test ..................................................................................................................................114
4.2.1.2 COM2 AND COM3 ..............................................................................................................................126
4.4 Statu s System ..................................................................................................................................... 160
4.4.1 Status Reporting System ......................................................................................................................160
4.4.1.1 Status Byte Register............................................................................................................................160
4.4.1.2 Standard Event Register .....................................................................................................................161
5.2.2 Piston Position Detection Adjustment .................................................................................................170
5.2.3 Emptying Oil Run-Off Tray (PG7202 and PG7302 Only) .....................................................................171
5.2.4 Purge Mounting Post Liquid Run Off (PG7202 Only) ..........................................................................172
5.2.5 Drive Belt Replacement ........................................................................................................................172
5.4 Mass Sets ............................................................................................................................................ 189
5.4.2.1 Updating Mass Set Files .....................................................................................................................189
5.5 Reloading Embedded Software into PG7000 Flash Memory .......................................................... 189
5.6 Disassembly and Reassembly of PG7000 ........................................................................................ 190
7.2.3.1 Fluid Head Components ......................................................................................................................199
7.2.3.2 Overall Fluid Head Correction .............................................................................................................200
Table 1. PG7102 Parts List ........................................................................................................................ 15
Table 2. PG7202 Parts List
Table 3. PG7302 Parts List
Table 4. PG7601 Parts List
Table 5. Manual Mass Set Parts List (excluding 80 and 100 kg)
Table 6. Manual Mass Set Parts List (80 and 100 kg)
Table 7. AMH-38 Mass Set Parts List
Table 8. AMH-100 Mass Set Parts List
Table 9. Mass Set Compositions
Table 10. Mass Set Compatibility
Table 11. PC-7100/7 600 P iston-Cylinder Modules Parts List
Table 12. PC-7200 Piston-Cylinder Modules Parts List
Table 13. PC-7300 Pisto n -Cylinder Modules Parts List
Table 14. Summary of PG7000 Direct Function Key Operations
Table 15. Pressure Units of Measure Available
Table 16. Valve Settings for Setting Differential Mode Static Pressure
Table 17. Valve Settings to Apply PG7000 Pressure to the RPM for Differential Mode Offsetting
Table 18. Valve Settings for Operating in Differential Mode
Table 19. SETUP File Choices, Factory Preferred Choice and Normal Value
Table 20. Security Levels - Functions NOT Executed Per Function/Level
Table 21. COM1, COM2 and COM3 Available Settings
Table 22. COM1 DB-9F Pin Designation
Table 23. COM2 and COM3 DB-9M Pin Designation
Table 24. Command Summary
Table 25. Error Messages
Table 26. Status Byte Register
Table 27. Standard Event Register
Table 28. Mounting Post Wire Colors, Description and Location
Table 29. PG7000 Troubleshooting Checklist
Table 30. Pressure Unit of Measure Conversions
Table 31. PG7000 Defined Pressure Calculation Variables
Table 32. DHI Authorized Service Providers
This manual provides the u ser wit h the inf orm ation nec essar y to operate v arious PG7000 P iston Ga uges.
It also includes a great deal of additional infor mation pr ovided to help you optimi ze PG7000 use and tak e
full advantage of its many features and functions.
This manual covers four PG7000 m odels: PG7102, PG7202, PG7302 and PG7601. The four m odels
have many features and characteristics in com mon as well as individual dif ferences. When discussing
features th at are comm on to all four m odels, they ar e referred to c ollectively as PG7000. W hen providing
information pertaining to a specific model, that model is referred to by its specific model number.
Before using the manual, take a moment to familiarize yourself with the Table of Contents structure.
All first time PG7000 users should read Sections 1 and 2. Section 3 provides a comprehensive
description of general PG7000 operating principles. Section 4 covers remote communication with an
external computer. Section 5 provides maintenance and calibration information. Section 6 is a quick
troubleshooting guide. Use the information in Section 6 to troubleshoot unexpected PG7000 behavior
based on the symptoms of that behavior.
Certain words and expressions have specific meaning as they pertain to PG7000s. The Glossary
(see Section 7) is useful as a quick referenc e for the definition of specific words and ex pressions as the y
are used in this manual.
Note
For those of you who “don’t read manuals”, go directly to section Error! Reference source not found., initial setup, to set up your PG7000. Then go to section 2.4,
power up and verification. This will get you running quickly with minimal risk of
causing damage to yourself or your PG7000. THEN… when you have questions or
start to wonder about all the great features you might be missing, get into the
manual!
Manual Conventions
Caution
“Caution” is used in throughout the manual to identify conditions or actions that
could cause harm to the PG7000 or to the devices that are connected to it.
Warning
“Warning” is used in throughout the manual to identify actions that could pose a
hazard to the user of the PG7000.
Note
“Note” is used throughout the manual to identify operating and applications advice
and additional explanations.
[ ] Indicates direct function keys (e.g., [RANGE]).
< > Indicates molbox1+ screen displays (e.g., <1yes>).
PG7000 Piston Gauges are reference level pressure standards that operate on the piston
gauge principle. Pressure is defined by balancing it against the force exerted by a known mass
accelerated by gravity on the effective area of a piston-cylinder.
A PG7000 piston gauge co ns is ts of the PG7000 Platform , one or s everal pis to n-cylinder modules, a mass
set. An automated m ass handling system is available. A PG7000 system normally also includes the
means to generate and adjust pr essures and to interconnec t the system components and a device bein g
calibrated or tested. The p ressure gen eration com ponent c an be m anual or a utom ated. COMPASS
Pressure software m ay also be included to assist i n executing test sequences, acquir ing test data and
producing test reports.
There are four PG7000 Platforms: PG7102, PG7202, PG7302 and PG7601. These have a common
PG7000 presentation and features . They are distinguis hed by their nor mal operating medium ( oil and/or
gas) and the capability to define pressures relative to a vacuum reference.
•PG7102 - Gas operated with gas lubricated piston-cylinder modules (PC-7100/7600 modules)
- Maximum pressure is 11 MPa (1 600 psi)
- Does not support definition of pressure against a vacuum reference
®
for
•PG7202 - Gas operated, liquid lubricated piston-cylinder modules (PC-7200 modules)
- Maximum pressure is 110 MPa (16 000 psi) when operate d w it h a P C-7200 module
- Maximum pressure is 200 MPa (30 000 psi) when operated with a PC-7300 module
- Does not support definition of pressure against a vacuum reference
•PG7302 - Oil operated (PC-7300 modules)
- Maximum pressure is 500 MPa (72 500 psi)
•PG7601 - Gas operated, gas lubricated piston-cylinder modules (PC-7100/7600 modules)
- Maximum pressure is 7 MPa (1 000 psi)
- Supports definition of pressure against a vacuum reference
PG7000 platforms, piston-cylinder modules, mass sets and mass handling systems are designed to
maximize metrologic a l per f or mance and ease of oper atio n. They include m an y featur es that enhance the
fundamental precision and stability of pressure measur ements as well as simplifying use and r educing
operator influence on the m easurements. Extensive m onitoring and controlling capabilit y and advanced
local and remote user interfaces are integrated into PG7000 Platforms.
Operator interaction with P G7000 and its extensive capabil ities and peripherals is ac complished through
a single display and keypad on the PG Terminal or from a computer via a single standard RS232 or
IEEE-488 interface.
100 kg, while not exceeding 11 MPa (1 600 psi)
100 kg, while not exceeding 110 MPa (16 000 psi) when operated
with PC-7200 piston-cylinder modules or 200 MPa (30 000 psi) when
operated with PC-7300 piston-cylinder modules
100 kg
38 kg
T es t port : DH200
T es t port : DH500
Drain port: DH500
T es t port : DH500
T es t port : DH200
Bell Jar Vent Port: DH200
Vacuum Reference
Pump Down Port: KF25 (KF40 available on optional AMH
automated mass handler)
External Vacuum Port: Optional KF25 on bell jar (KF40 available
on optional AMH automated mass handler)
Note
DH200 and DH500 are gland and collar t ype fittings for
1/4 in. (6.35 mm) coned and left hand threaded tubes.
DH200 is equivalent to AE SF250C, HIP LF4, etc.
DH500 is equivalent to AE F250C, HIP HF4, etc.
Available, must be specified.
1.2.1.1 Embedded Features
•Local contr ol with 2 x 20 vacuum fluorescent dis play and 4 x 4 function
driven keypad.
•Real tim e (1 second update rate) display and m easurement of ambient
(pressure, temperature, humidity) and instrument (piston-cylinder
temperature, piston posit ion, piston drop rate, pis ton rotation rate, piston
rotation decay rate, reference vacuum) conditions.
•Real time (1 second update rate) mass-to-pressure and pressure-to-
mass calculations taking into consideration all environmental and
operational variables.
•Full gas and liquid fluid head cor rections including D UT head correction
and piston position head correction.
• Adjustable mass loading resolution (0.01 g to 0.1 kg).
• Audible prompts of instrument status (piston movement, Ready/Not Ready
indication) with override capability.
• Integra ted aut omated mass handling option (AMH-38 or AMH-100).
• Interfacing and automatic exploitation of external barometer via RS232.
• Interf acing and automatic exploi tation of any external vac uum gauge via
RS232 (PG7601 only).
•Automated differential mode to define low differential pressures at
various static pressures between vacuum and two atmospheres.
•Automated high line differential mode to define differential pressure at
high line pressure.
•Storage and one step activation of metrological data on up to 18 piston-
cylinder modules, (3) mass sets and (3) mass loading bells.
Gas operated, liquid lubricated piston-cylinder module characteristics. Used in
PG7202 platform.
Though not recommended for day-to-day operation, PC-7200
modules can also be filled completely with oil and operated with
oil as the test medium (see Section 2.3.3).
Note
PC-7200-100
Piston and Cylinder Material
PC-7200-200
Piston and Cylinder Material
PC-7200-500
Piston and Cylinder Material
PC-7200-1
Piston and Cylinder Material
PC-7200-2
Piston and Cylinder Material
Operation
Lubricating Liquid
Nominal Diameter
Nominal Area
Mounting System
Operation
Lubricating Liquid
Nominal Diameter
Nominal Area
Mounting System
Operation
Lubricating Liquid
Nominal Diameter
Nominal Area
Mounting System
Operation
Lubricating Liquid
Nominal Diameter
Nominal Area
Mounting System
Operation
Lubricating Liquid
Nominal Diameter
Nominal Area
Mounting System
Gas operated, liquid lubricated
Synturion 6 (Krytox
®
optional)
Tungsten carbide
11.2 mm
98.1 mm
2
Negative free deformation
Gas operated, liquid lubricated
Synturion 6 (Krytox
®
optional)
Tungsten carbide
7.9 mm
49.0 mm
2
Negative free deformation
Gas operated, liquid lubricated
Di-2-ethylhexyl Sebacate (Krytox
Tungsten carbide
5.0 mm
19.6 mm
2
Negative free deformation
Gas operated, liquid lubricated
Di-2-ethylhexyl Sebacate (Krytox
Tungsten carbide
3.5 mm
2
9.8 mm
Negative free deformation
Gas operated, liquid lubricated
Di-2-ethylhexyl Sebacate (Krytox
Tungsten carbide
All masses are delivered in molded, reusable, transit cases with custom inserts.
Masses > 50g
Material
Finish
Adjustment Tolerance
Uncertainty of Measured Values
Masses < 50g
304L non-magnetic stainless steel
Electropolished
± 20 ppm of nominal value (manual mass sets, AMH
automated mass handler mass sets do not have fixed
adjustment tolerances)
± 5 ppm or 1 mg, whichever is greater
± 1 mg
PG7000™ OPERATION AND MAINTENANCE MANUAL
Masses desig nat ed “ t ar e” are d el ive red w ith out rep ort ed m easu r ed va lu es
and are intended only for use on the “tare” PG7000 in high line differential
pressure measurement mode.
1.2.4 Pressure Measurements
1.2.4.1 PC-7100/7600
For uncertainty in piston-cylinder effective area and typical
measurement uncertainty in pressure defined by the piston
gauge, see the piston-cylinder calibration report and current
revision of Technical Note 7920TN01.
Note
Note
PC-7100/7600-10
PC-7100/7600-10-L
Sensitivity
Reproducibility
Typical Drop Rate (35 kg)
PC-7100/7600-20
Sensitivity
Reproducibility
Typical Drop Rate (35 kg)
PC-7100/7600-50
Sensitivity
Reproducibility
Typical Drop Rate (35 kg)
PC-7100/7600-100
Sensitivity
Reproducibility
Typical Drop Rate (35 kg)
PC-7100/7600-200
Sensitivity
Reproducibility
Typical Drop Rate (35 kg)
1 Sensitivity: The smallest variation in input detectable in output.
1
0.02 Pa + 0.5 ppm
2
± 2 ppm
0.2 mm/min
1
0.04 Pa + 0.5 ppm
2
± 2 ppm
0.3 mm/min
1
0.1 Pa + 0.5 ppm
2
± 2 ppm
0.5 mm/min
1
0.2 Pa + 0.5 ppm
2
± 3 ppm
0.7 mm/min
1
0.4 Pa + 0.5 ppm
2
± 3 ppm
1.0 mm/min
Piston-cylinder modules designated “tare” are delivered without
reported values and are intended only for use on the “tare”
PG7000 in high line differential pressure measurement mode.
For uncertainty in piston-cylinder effective area and typical
measurement uncertainty in pressure defined by the piston
gauge, see the piston-cylinder calibration report and current
revision of Technical Note 7920TN01.
PC-7200-100
Sensitivity
Reproducibility
Typical Drop Rate (50 kg)
PC-7200-200
Sensitivity
Reproducibility
Typical Drop Rate (50 kg)
PC-7200-500
Sensitivity
Reproducibility
Typical Drop Rate (50 kg)
PC-7200-1
Sensitivity
Reproducibility
Typical Drop Rate (50 kg)
PC-7200-2
Sensitivity
Reproducibility
Typical Drop Rate (50 kg)
1 Sensitivity: The smallest variation in input detectable in output.
2 Reproducibility: Combined long term stability of piston-cylinder effective area and masses.
1
2 Pa + 1 ppm
2
± 5 ppm
0.10 mm/min
1
4 Pa + 1 ppm
2
± 5 ppm
0.15 mm/min
1
10 Pa + 1 ppm
2
± 5 ppm
0.20 mm/min
1
20 Pa + 1 ppm
2
± 5 ppm
0.25 mm/min
1
40 Pa + 1 ppm
2
± 5 ppm
0.50 mm/min
1.2.4.3 PC-7300
Note
For uncertainty in piston-cylinder effective area and typical
measurement uncertainty in pressure defined by the piston
gauge, see the piston-cylinder calibration report and current
revision of Technical Note 7920TN01.
1 Sensitivity: The smallest variation in input detectable in output.
2 Reproducibility: Combined long term stability of piston-cylinder effective area and masses.
1
20 Pa + 1 ppm
2
± 5 ppm
0.20 mm/min
1
40 Pa + 1 ppm
2
± 5 ppm
0.40 mm/min
1
100 Pa + 1 ppm
2
± 5 ppm
1.00 mm/min
1.3 Terminal and Platform Front and Rear Panels
1.3.1 Terminal Front and Rear Panels
1.3.1.1 PG Terminal Front Panel
The front panel ass embly provides a 2 x 20 vacuum fluorescent display and a
4 x 4 membrane keypad for local user interface. The terminal front panel assembly
is the same for all PG7000 models.
The rear panel assembly provides the communications connection to the
PG7000 Platform and the power connection module. The terminal rear panel
assembly is the same for all PG7000 models.
2. Fuse
Figure 2. PG Terminal Rear Panel
5. Cooling fan
1.3.2 Platform Rear Panels
The PG7000 Platform rear panels provide the connection to the PG Terminal, remote
communication connections and pressure connection ports. The rear panels of all PG7000
models are identical except for the pressure connections (see Figure 3, # 7).
COM2 (RS232) - External
Barometer, External Vacuum
Gauge (PG7601 only) and Pass
Through Communications
COM3 (RS232) - Automated
Pressure Generation/Control
Component
Temperature - Humidity Probe
COM1 (RS232) - Remote Hos t
Communications
IEEE-488 - Remo te Host
Communications
AMH Connection
Pressure Ports:
PG7102 - TEST port: DH200
PG7202 - TEST and DRAIN
ports: DH500
PG7302 - TEST port: DH500
PG7601 - TEST and VACUUM vent
A t ypical PG7000 s ystem incl udes the PG7000 P latform ( see Sect ion 2.1.1 .1), a mas s set, (see
Section 2.1.1.2), one or more piston-cylinder modules (see Section 2.1.1.3) and other accessories
such as an AMH a utomated mass han dler and/or pres sure generation a nd control com ponents
(see the accessory Operation and Maintenance Manual or Instruction Sheet).
2.1.1.1 Platform
Caution
The mass loading bell is a metrological element that is part of
the mass set. Like all of the masses, it is preferable not to
handle it with bare hands. Protective gloves are provided in the
accessory kit of each PG7000 Platform.
The PG7000 Platform is shipped in a reusable, molded shipping and storage case.
Open the PG7000 shipping and storage case (it is th e large, 66 cm x 53 cm
x 47 cm case).
Remove the PG Terminal and accessories from upper packing insert.
Inspect and inventory the acc ess or ies (see Secti on 2.1 .2).
Remove the upper packing insert.
Carefully lift the PG7000 Platform from its position in the lower packing
insert. Note the orientatio n so that the same orientation will be used when
PG7000 is repacked.
Reinstall the upper packing insert into the shipping and storage case and
store in a safe place.
2.1.1.2 Mass Set
Caution
The stability over time of PG7000 pressure measurements is a
function of the stability of the masses loaded on the piston.
Precautions should be taken in handling the masses to minimiz e
influences that may change their mass. This includes always
wearing protective gloves when handling the masses to avoid
contaminating them with body oils and perspiration. Protective
gloves are pr ovid ed i n th e a cce ss ory k it s of PG700 0 P la tfor ms.
The mass set accessories are shipped in a separate corrugated container.
Open the corrugated container and inspect and inventory the accessories.
The PG7000 masses are shipped in reusable, molded shipping and
storage cases. The PG7000 masses should be removed from their shipping
cases and inventoried when actually setting up the PG7000 system.
The piston-cylinder modules are shipped in Acetal bullet cases that ar e p ac k ed in
corrugated containers with custom foam inserts.
Open the corrugated containers and remove the piston-cylinder modules
and accessories.
The bullet cases screw open by turning the lid counterclockwise.
When reinstalling an oil (PC-7300) or liquid lubricated (PC-7200)
piston-cylinder module in its bullet case, be sure to empty out
any liquid that may have collected in the hole in the bottom of
the case. Excess liquid wi ll not compress, making it difficult to
fully close the case and could result in damaging it.
2.1.1.4 Automated Mass Handler
See the AMH-38/AMH-100 Operation and Maintenance Manual.
2.1.2 Inspecting Contents
Check that all items ar e present and have NO visible signs of damage. A parts list of items
supplied is provided in Section 2.1.2.1 for PG7000, Section 2.1.2.2 for mass sets, and
Section 2.1.2.3 for piston-cylinder modules.
Caution
2.1.2.1 Platform
Each PG7000 Platform is delivered complete with acces sories as listed by part
number in Tables 1 through 4.
Platform 3117734 3117752
Manual Mass Bell 3071537
Terminal 3069735
PG Terminal to Platform Cable Non-CE (DB25M - DB25F,
≈ 1.8 meters)
CE (DB25M - DB25F,
≈ 1.5 meters)
Power Cable 3133781 (Black) 3153005 (Gray)
TH Probe Assembly 2106009
Accessory Kit 3117741
NIP, SS, DH200, 2.75 in. 3068377
ADPT, SS, DH200 F x 1/8 in. NPT F 3068547
O-ring, Buna 2-242 (2 ea.) 3135041
Storage Cover, 7600 Type 3135594
Allen Wrench, 2.5 mm 3136044
Allen Wrench, 3 mm 3135703
Allen Wrench, 5 mm 3136098
Spanner Wrench (Metrological) 3068940
Krytox GPL205/6 0.5 oz.
Gift Kit with Gloves 3123777
ADPT, DH200 M x 1/8 in. swage 3069062
P/N 3069572
NON-CE
3068724
3072235
2493420
PG7102
P/N 3072317
CE
Calibration Report (PG Platform)
Calibration Report (Mass Bell)
Technical Data
PG7000 Operation &
Maintenan ce Ma nu al
≈ 1.5 meters)
Power Cable 3133781 (Black) 3153005 (Gray)
TH Probe Assembly
Accessory Kit 3120011
DH500 M x 1/8 in. NPT F 3142684
O-ring, Buna 2-242 (2 ea.) 3135041
Storage Cover, 7600 Type 3135594
Allen Wrench, 2.5 mm 3136044
Allen Wrench, 3 mm 3135703
Allen Wrench, 5 mm 3136098
Wrench, 5/8 in. 3139417
Collar, SS, DH500 3068607
Krytox GPL205/6 0.5 oz.
Gift Kit with Gloves 3123777
Documentation
Calibration Report (PG)
Calibration Report (Mass Bell)
Technical Data
PG7000 Operation &
Maintenan ce Ma nu al
Documentation CD
Platform 3118073 3118086
Manual Mass Bell 3071537
Terminal 3069735
PG Terminal to Platform Cable Non-CE (DB25M - DB25F,
≈ 1.8 meters)
CE (DB25M - DB25F,
≈ 1.5 meters)
Power Cable 3133781 (Black) 3153005 (Gray)
TH Probe Assembly
Accessory Kit 3120011
DH500 M x 1/8 in. NPT F 3142684
O-ring, Buna 2-242 (2 ea.) 3135041
Storage Cover, 7600 Type 3135594
Allen Wrench, 2.5 mm 3136044
Allen Wrench, 3 mm 3135703
Allen Wrench, 5 mm 3136098
Wrench, 5/8 in. 3139417
Collar, SS, DH500 3068607
Krytox GPL205/6 0.5 oz.
Gift Kit with Gloves 3123777
Documentation
Calibration Report (PG)
Calibration Report (Mass Bell)
Technical Data
PG7000 Operation &
Maintenan ce Ma nu al
Documentation CD
Platform 3117525 3117540
Manual Mass Bell 3071603
Bell Jar and Seal 3068933 and 3068634
Terminal 3069735
PG Terminal to Platform Cable Non-CE (DB25M - DB25F,
≈ 1.8 meters)
CE (DB25M - DB25F,
≈ 1.5 meters)
Power Cable 3133781 (Black) 3153005 (Gray)
TH Probe Assembly
Accessory Kit 3117533
NIP, SS, DH200, 2.75 in. 3068377
ADPT, SS, DH200 F x 1/8 in. NPT F 3068547
O-ring, Buna 2-242 (2 ea.) 3135041
Storage Cover, 7600 Type 3135594
Allen Wrench, 2.5 mm 3136044
Allen Wrench, 3 mm 3135703
Allen Wrench, 5 mm 3136098
Spanner Wrench (Metrological) 3068940
Krytox GPL205/6 .5 oz.
Gift Kit with Gloves 3123777
ADPT, DH200 M x 1/8 in. swage 3069062
Valve, Vacuum Relief 3124573
Documentation
Calibration Report (PG)
Calibration Report (Mass Bell)
Technical Data
PG7000 Operation &
Maintenan ce Ma nu al
Documentation CD
PG7000 mass sets are compos ed of different c ombinations of ind ividual mas ses
and accessories depending on the specific mass set ordered (see Tables 5 - 9).
Table 5. Manual Mass Set Parts List (excluding 80 and 100 kg)
DESCRIPTION PART NO.
Mass Set Refer to Table 9
Reusable Molded Transit Case with Foam Inserts
35 kg set
40 kg set
45 kg set
55 kg set
Mass Set Tray and Spindle 3147461 and 3148764
Dust Covers 3138017 and 3138130
Calibration Report 3152121
3068969
1 ea.
1 ea.
1 ea.
1 ea.
Table 6. Manual Mass Set Parts List (80 and 100 kg)
DESCRIPTION PART NO.
Mass Set Refer to Table 9
Reusable Molded Transit Case with Foam Inserts
80 kg set
100 kg set
Mass Set Tray and Spindle 3147461 and 3148764
Dust Covers 3138017 and 3138127
Calibration Report 3152121
3068969
1 ea.
1 ea.
3068991
1 ea.
1 ea.
1 ea.
2 ea.
3068984
2 ea.
3 ea.
Table 7. AMH-38 Mass Set Parts List
DESCRIPTION PART NO.
Mass Set Refer to Table 9
Reusable Molded Transit Case with Foam Inserts
13 kg set (MS-AMH-13)
25 kg set (MS-AMH-25)
39 kg set (MS-AMH-38)
Calibration Report 3152121
3123990
1 ea.
1 ea.
1 ea.
3069004
1 ea.
1 ea.
1 ea.
Table 8. AMH-100 Mass Set Parts List
DESCRIPTION PART NO.
Mass Set Refer to Table 9
Reusable Molded Transit Case with Foam Inserts
40 kg set (MS-AMH-40)
60 kg set (MS-AMH-60)
80 kg set (MS-AMH-80)
100 kg set (MS-AMH-100)
The mass loading bell and piston make up part of the total mass
load. The mass loading bell for loading manual mass sets is
delivered with the PG7000 platform. The mass loading bell for
AMH mass sets is d elivered with the m ass set. Piston -cylinder
modules are purchased and delivered separately.
2.1.2.3 Piston-Cylinder Module(s)
Table 11. PC-7100/7600 Piston-Cylinder Modules Parts List
Table 13. PC-7300 Piston-Cylinder Modules Parts List
100 kPa
PC-7300-100
Piston-Cylinder Kit
Piston-Cylinder
Module
Hermetic Acetal
Bullet Case
Accessory Kit 3119102 3119102 3119102 3119366 3119366 3119366
O-rings
Calibration
Reports
3070039 3070042 3070056 3070063 3070074 3070088
3118918 3118976 3119032 3119116 3119178 3119229
3071865 3071865 3071865 3071865 3071865 3071865
2527053
3134022
3152121 3152121 3152121 3152121 3152121 3152121
200 kPa
PC-7300-200
2527053
3134022
500 kPa
PC-7300-500
2527053
3134022
1 MPa
PC-7300-1
1785497
(5) 927863
2 MPa
PC-7300-2
1785497
(5) 927863
5 MPa
PC-7300-5
1785497
(5) 927863
2.2 Site Requirements
The exact PG7000 s ystem installation is af fected by the elements ot her than the PG7000 Platf orm that
make up the PG7000 system.
When selecting and preparing a site to set up the PG7000 system, the following should be considered:
•Ambient conditions: To a chieve optimum metrologic al performance, ambient c onditions should be
controlled and maintained within the following:
♦ Temperature: 19 to 23 °C, minimize rate of change of temperature.
♦ Relative Humidity: 10 to 60 %RH (non-condensing).
♦ Ambient Pressure: Minimize external influences that will cause barometric instability.
♦ Air Currents: Do not i nstall the PG70 00 Pla tform un der a sour ce of v ertica l air curr ents s uch as an
overhead air conditioning duct. These can blow on the mass load and add unquantified forces.
♦Vibration:Minimize local v ibration. Excessive vibration will red uce the stabilit y of the pressures
defined by PG7000 (v ibration aff ects the floating piston). Exces sive high fr equency vibration , for
example from a vacuum pump on the same table as the PG7000, may affect piston sensitivity.
•Bench st ability: Up to 100 kg ma y be loaded and unloaded onto the PG7000 Platf orm. The bench
on which the PG7000 sits should not def lect significa ntly under the m ass load changes. T his can be
verified by setting the PG7 000 Platform on the bench, leveling it, l oading and u nloading the c omplete
mass set while observing whether the level setting changes.
•Location of other components: Plan the space requir ed and a convenient layout f or the complete
PG7000 system inclu ding t he PG T erm inal, m ass s et, pressure gen erat ion/co ntrol com ponent( s), tes t
instrument connection and computer (if present). If using a PPC, MPC, GPC or OPG to
generate/control pres sure, see its Operat ion and Maintenanc e Manual for infor mation on installin g it.
If a Fluke Calibration interconnections kit is being used to interconnect the components, see its
instruction sheet.
•Elect rical and pressu re suppli es: Plan the supply of elect rical power to th e PG Term inal and to the
pressure generation/control component(s), if needed. If using a PPC, MPC, GPC or OPG to
generate/control pressure, see its Operation and Maintenance Manual for information on the
pressures source(s) it needs and how to connect them. Gas supplied to a PC-7100/7600
piston-cylinder module m ust be clean and dry (instrument grade minim um, high purity preferred) to
avoid contaminating the piston-cylinder gap.
•Referen ce vacuum supply (PG7601 only): Plan for the vacuum connection to the pl atform or the
optional AMH automated mas s handler and the location of the reference vacuum pump.
•Bell jar placement (PG7601 only): Plan a location for the bell jar when it is removed from the
platform to load and unload masses. A small shelf is often used for this purpose.
•AMH automated mass handler (optional) placement: If an AMH is being used, plan electrical and
pneumatic connectio ns to it as well as a locat ion to p lace it when it is rem oved fr om the pl atform (s ee
the AMH-38/AMH-100 Operation and Maintenance Manual).
Before setting up the PG7000 system, see Section 2.2 for information on
site requirements.
To prepare PG7000 for check out and operation:
Set up the PG7000 Platform (see Section 2.3.1.1).
If an optional AMH-100 m ass set is being used, set it up with its m ass set (see the AMH -
38/AMH-100 Operation and Maintenance Manual).
Make the system pressure interconnections (see Section 2.3.1.2).
If a manual mass set is being used, set up the manual mass set (see Section 2.3.1.3).
2.3.1.1 Setting Up the Platform
To set up the PG7000 Platform proceed as follows:
Place the PG7000 Platform on the site table in the desired orientation.
Though the rear panel is usually in the back, any orientation can be used.
Place the PG7000 Terminal at the desired location.
Connect the PG7000 Terminal to the PG7000 Platform using the 25-pin
cable supplied.
Connect the PG7000 Temperature - Humidity Probe per Figure 3.
Connect electrical power ( 85 to 264 VAC, 50/60 H z) to the PG700 0 Term inal
using the power cable sup plied. Any grounde d power cable with a sta ndard
IEC320-313 connection may be used.
(PG7601 Only) - Install the vacuum vent valve kit on the vacuum vent port
on the rear of the PG7000 Platform. Refer to the inst ruction sheet provide d
with the vent valve assembly.
Connect the reference vacuum source and shutoff valve to the reference
vacuum port. Tak e measures to ass ure that vacu um oil cannot ret urn to the
PG7601. If an opti onal AMH-38 automated mas s handler is b eing used, th e
reference vacuum may be connect ed to the larger KF40 v acuum connection
on the AMH-38 vacuum chamber.
If an external barom eter and/or vacuum gauge is/are being used, establish
communications between the barometer/vacuum gauge and the PG7000
Platform by connecting the external device RS232 port to the PG7000
Platform Com2 port and setting up PG7000 to read and use an external
barometer and/or vacuum gauge (see Section 3.11.5.4, 3.11.5.5). Set the
external barometer head height (see Section 3.11.3.3).
If an automated pressure generation/control component is being used,
establish communications between the automated pressure
generation/control component and the PG7000 Platform by connecting the
generation/control com ponent RS232 port to the PG700 0 Platf orm COM3 p ort
and setting up PG7000 to use an automated pressure generation/control
component (see Section 3.9.9).
If an AMH automated m ass handling s ystem is being used, set u p the AMH
mass set and the AMH mass handler f ollowing the instructions in the AMH 38/AMH-100 Operation and Maintenance Manual.
Level the platform using the PG7000 Platform’s two leveling feet and the
level mounted on the front of the platform. (Or on the optional AMH mass
handling system.)
Interconnect the PG7000 P latform, press ure generation/control com ponents and
a test connection.
The pressure connection on the PG7000 TEST port is:
•PG7102/PG7601: DH200 (DH200 is equivalent to AE SF250C,HIP LF4,
etc.). Adapt ers to c onver t the DH 200 c onnec tion t o 1/8 in. NPT fem ale and
1/8 in. swage are provided with PG710 2 and P G7601 P lat fo r m a cce sso ries .
•PG7202: DH500 (DH500 is equivalent to AE F250C, HIP HF4, etc.).
Connect the gas test medium press ure control/generation com ponent to
the TEST port. If the PG7202 is to be operated in oil, connect the oil
pressure control/gener ation com ponent to the DR AIN port (the gas or oil
component is disconnected when not in use, see Section 2.3.3).
•PG7302: DH500 (DH500 is equivalent to AE F250C, HIP HF4, etc.).
If using a standard interconnections kit such as PK-7600PPC/MPC P/N 3069508, PK-7600-PPC/MPC-DIF P/N 3070127 or
PK-7100-MPCD-DIF P/N 3070226, see the instruction sheet
provided with the kit for installation instructions. If an OPG1 or
GPC1 generator/controller is being used, pressure
interconnecting hardware is included with it. See its Operation
and Maintenance Manual for connection instructions.
Note
2.3.1.3 Setting Up a Mass Set
Note
If installing an MS-AMH-xxx mass set for use with an AMH
automated mass handling system, see the AMH-38/AMH-100
Operation and Maintenance Manual.
To install a PG7000 manual mass s et, plac e the mass loading tra y (provi ded wit h
the mass set accessories) at the desired location then install the individual
masses on the mass loading tray. See the AMH-38/AMH-100 Operation and
Maintenance Manual for instructions on installation of an AMH mass set for
automated mass handling.
Caution
It is VERY IMPORTANT that the individual masses be installed
on the mass loading tray in proper sequence. This will ensure
that PG7000 mass loading instructions are executed properly
(carefully follow the instructions provided in Section 2.3.1.3,
Installing M asses on t he Mass L oading Tray).
Installing Masses on the Mass Loading Tray (Manual Mass Set)
PG7000 masses are shipped in reusable, molded shipping and storage cases.
One of the cases conta ins a 4.0, 4.5 or 5 kg m ass and the masses of 2 k g and
under, the ot he r c ase(s) con ta i n(s ) the main m ass es of 1 0 or 5 kg each. Eac h mass
is packed in a sealed plastic bag and then placed in a protective shipping insert.
Caution
The stability over time of PG7000 pressure measurements is a
function of the stability of the masses loaded on the piston.
Precautions should be taken in handling the masses to minimiz e
influences that may change their mass. This includes always
wearing protective gloves when handling the masses to avoid
contaminating them with body oils and perspiration. Protective
gloves are pr ovid ed i n th e a cce ss ory k it s of PG700 0 P la tfor ms.
To install the masses on the mass loading tray, proceed as follows:
Open the shipping cases.
Install the main masses: The main masses (a series of 10 k g mass es if the
mass set is > 55 kg, a series of 5 kg masses if the mass set is < 60 kg)
are installed horizontally aligned on the mass loading spindle.
The main masses are sequentially numbered starting with he number 1.
The main mass with the highest sequential num ber is installed first at the
bottom of the stack (i.e., the first m ass loaded on the tra y). The rest of the
main masses should be stacked upwards in descending order endin g with
main mass sequential number 1. Be careful NOT to confuse the makeup
mass (refer to next item) with main mass 1.
Install the make up mass: The make up mass is a single mass sequentially
numbered 1. It has the same diameter as the main masses. It is a 9 kg
mass if the main masses are 10 kg. It is 4 or 4.5 kg if the main masses are 5 kg.
The make up mass is placed on top of the main mass stack . It is always the
top of the stack.
Install the fractionary masses: The fractionary masses are all the mas ses
of lower value than the main mass es and makeup mass. These are m asses
of 5 kg and under for m ass sets with 10 k g main masses. T hey are mass es
of 2 kg and under f or mass sets with 5 k g main masses. Fractionary masses
of 1 to 5 kg are discs with a c entral ho le. Fractionary masses of 100 to 500 g
are solid, small diam eter pucks. Fractionary masses of 50 g and under are
grams masses packed and stored in their own separate storage case.
Fractionary mass discs and pucks are installed vertically in the corresponding slots
in the mass loading tray. Use a consistent s etup for the sequence number
when there are two masses (e.g., always load sequential number 1 in the front).
2.3.2 installing a Piston-Cylinder Module into the Platform
To operate the PG7000 Pl atform, a piston-cylinder m odule must be installed in i ts mounting
post. To install a piston-cylinder module in the PG7000 Platform, proceed as follows:
Remove the PG7000 Platform mounting post plug. Unscrew the ORANGE plastic
mounting post plug that is installed in the PG7000 Platform mounting post.
Rotate counterclock wise to r emove.
Remove the piston-cylinder module from its bullet case. Select a piston-cylinder
module. Open the piston-c ylinder m odule bullet cas e by rotatin g its lid counterc lock wise.
Remove the piston -cylinder module from the bullet case base by unthreading it from the case.
Hold the piston-cylinder module body by the knurled area and rotate it counterclockwise.
Note
PC-7200 gas operated, liquid lubricated piston-cylinder modules are
delivered with their lubricating liquid reservoir drained. The reservoir must
be filled prior to using the pi ston-cylinder module. If installing a PC-7200
gas operated, liquid lubricated piston-cylinder module, see Section 5.3.3
for instructions on how to fill the module’s liquid lubr ication re servoir prio r
to install atio n, the n conti nue t he proc edure from thi s poi nt.
Caution
When reinstalling an oil or liquid lubricated gas piston-cylinder module in
its bullet case, be sure to empty out any liquid that may have collected in
the hole in the bottom of the case. Th e liquid will not comp ress, m aking i t
difficul t to ful ly clos e the c ase an d co ul d res ul t in da magi ng it.
will be encountered in the
second half of travel as the
Place the piston-cylinder module in the PG7000 Platform mounting post. Place
the piston-cylinder module (thread down) into the PG7000 Platform mounting post
(see Figure 4 below).
If installing the piston-cylinder module into a PG7302 or a PG7202 to be
operated with oil, first use the oil generat ion control component to fill the
counter bore in the mounting post with oil until it just starts to overflow.
Purge the air from under the piston-cylinder module (PC-7300 modules only in
PG7302 or PG7202). Rotate the piston-cylinder m odule clockwise until all threads are
engaged and there is no gap bet ween the pis ton-c ylinder m odule and the m ounting post.
Back the piston-cylinder module off slightly by rotating it counterclockwise 3/4 turn.
Slowly supply additional oil to the mounting post from the pressure generation/control
component. Watch the oil run of f tube on the bottom r ight of the platf orm. As soon as oil
appears or can be seen flowing, stop the oil supply.
Screw the piston-cylinder
Rotate the piston-cylinder
threads are engaged and there
is NO gap between the pistoncylinder module an d the P G7000
mounting post. Slight res istance
Note
2.3.3 Switching a PG7202 Between Gas Operation and Oil
piston-cylinder module O-rings
seat in the mounting post.
Figure 4. Piston-Cylinder Module Installation
Caution
•Low torque manual rotation is all that should be required to fully seat the
piston-cylinder module into the PG7000 mount ing post. Never force the
piston-cylinder module into the mounting post.
•Alwa ys maintain PG7202 and PG7302 p iston- cylinder modules v ertical
with the piston cap up. Do not invert the assembly, as this might allow
liquid to run up into the piston head and into the adjustment mass and
cap. Liquid contamination of the piston head and cap changes the
mass of the piston assembly and could lead to out of tolerance
pressure definitions at low mass loads. If liquid contaminates the
adjustment mass and cap, disassemble the module and clean it (see
Sections 5.3.1, 5.3.4, 5.3.2.2).
Operation
PG7202 can be operated w ith gas as the pressuri zed medium using PC-7200 gas operated,
liquid lubricated m odules or with oil as the pressurized m edium using oil operated PC-7300
modules.
PC-7200 gas operated, liquid lubricated piston-cylinder modules can be
operated in gas oil filled with oil and opera ted in oil. Oil operation of PC7202 modules is not recommended for routine operation. It can be useful
in very specific crossfloating circumstances, particularly when
establishing a calibration link between independent gas operated and oil
operated piston gauges. When switching PC-7200 module between oil and
gas and oil operation, see Section 5.3.3 for information on emptying
excess oil from the modu le.
Switching a PG7202 from gas to oil operation
Note
Only the hig h pressure PC -7300 oil modules m ay be used in the PG7202
platfor m (P C-7300-1, -2, -5)
Disconnect the gas generation/control system from the TEST port. Disconnec t the
tube at the DH500 T EST port connection on the back of the PG7202. Loosely i nstall a
DH500 plug in the TEST port.
Connect the oil generation/control system to the DRAIN port. Connect a tube from
the oil generation/contro l system to the PG 7202 DRAI N port and t ighten i t ( torque DH 500
glands to 15 Nm (12 ft.lb)).
Fill the PG7202 mounting post with oil. Hand tighten the DH500 plug in the TEST
port. Use the oil ge neration/contro l component to fill the PG720 2 mounting p ost with oil.
Fill to the edge of the second step in the m ounting post. Place a paper to wel under the
TEST port plug and l oosen the D H500 plug allowing o il to run out until it is at the level of
the first step in t he mounting post and t her e is no air in the run off oil. If there is still air in
the run off, repeat the o peration. After f illing the mounting post, ti ghten the DH500 plug
(torque DH500 glands to 15 Nm (12 ft.lb)).
Install a PC-7300 oil operated piston-cylinder module in the PG7202 platform.
Install the module and purge the air from underneath it (see Section2.3.2).
Operate with oil as the test medium. The PG7202 can no w be operated with oil as th e
test medium.
Caution
The maximum working p ressure of the PG7202 platform when used in
oil with a PC-7300 module is 200 MPa (30 000 psi). The maximum
pressure when using a PC-7200 module is 110 MPa (16 000 psi). Do not
exceed this limit.
Switching a PG7202 from oil to gas operation
Remove the piston-cylinder module. Disinstal l t he P C-73 00 oil oper ate d pist on-cylinder
module for the PG7202 platform.
Disconnect the oil generation/control system from the DRAIN port. Place a paper
towel under the DRAIN por t and disconnec t the tube t o the oil genera tion/con trol s ystem.
Let all the oil run out of the PG7202 platform.
Remove the DH500 plug from the TEST port. Place a paper towel under the TEST
port and remove the DH500 plug. Let all the oil run out of the PG7202 platform..
Install a PC-7200 gas operated, liquid lubricated piston-cylinder module in the
PG7202 platform. Install the piston-cylinder module (see Section 2.3.2).
Connect the gas generation/control system to the TEST port. Connect a tube from
the oil generation/cont rol system to the PG7202 TES T port and tighte n it (torqu e DH500
glands to 15 Nm (12 ft.lb)).
Prepurge the oil from the PG7202 mounting post. Hold a paper towel lightly over the
DRAIN port opening. Leaving the drain port open, use the gas generation/control
component to flow enough gas through the mounting post to just lift the piston.
Purge the PG7202 mounting post. Purge the PG7202 m outing pos t usin g t he s t andard
purging procedure (see Section 5.2.4) and finish by tightening the plug in the DRAIN port.
Operate with gas as the test medium. The PG72 02 can now be operated with gas as
the test medium.
Caution
The maximum working pressure of the PG7202 platform when used with a
PC-7200 module is 110 MPa (16 000 ps i). Do no t exceed this limit.
2.4 Power Up and Verification
2.4.1 Power UP
Turn the PG7000 power ON by pressing th e power ON/OFF switc h on the rear panel of th e
PG Terminal. Observe the PG terminal display as the terminal connects with the PG7000
Platform, tests, initializes and goes to the main run screen (see Section 3.7).
If <….Searching…..> displays for more than 5 seconds, the communications between the
PG7000 and the PG T erminal are failing. Check that the PG7000 to PG Terminal cable is
properly installed.
If PG7000 fails to reach the m ain run sc reen, ser vice m ay be required. Recor d the s equence
of operations and displays observed and contact a Fluke Calibration Authorized Service
Provider
(see Table 32).
2.4.2 Check that On-Board Piston-Cylinder Module and
Mass Set Information are Correct
PG7000 uses stored pist on-cylinder and mass set metrologica l information to calculate the
reference pressures it defines (see Section 3 .1). For the pressur e values to be correct, the
stored metrological information on the piston-c ylinder, mass set and m ass loading bell must
be correct. Before using PG 7000 for accurate pressure definitio n, the validity of the stored
information should be v erified. This cons ists of com paring the piston-c ylinder, mass set and
mass loading be ll inform ation stor ed in PG7 000 to the i nform ation in t he current p iston-cylinder
and mass set calibration reports.
To verify the PG7000 pist on-cylinder, mass set and mass loading bell information, use th e
piston-cylinder and mass set viewing capabilities accessed by pressing [SPECIAL],
<1PC/MS> (see Section 3.11.1). Compare all the information contained in the PG7000
piston-cylinder, mass set and mass loading bell files to the information on the current
piston-cylinder, mass set and mass loading bell calibration reports.
2.4.3 Set Local Gravity Value
PG7000 uses the value of local acceleration due to gravity (gl) in its calculation of the
reference pressure it d ef ine s (s ee Sect ion 3.1). The cor rec t va lue of local ac c eler atio n du e to
gravity at the site of PG7000 use must be entered. This is accomplished by pressing
[SPECIAL], <6gl> (see Section 3.11.6) and editing the value of local gravity.
2.4.4 Setup Pressure Equation Va riable Input Sources
PG7000 uses man y variables in calculating defined pres sures. The sources of the variables
are determined by the current SETUP file. SETUP files are viewed, created, edited and
selected using the SETUP function accessed by pressing [SETUP] (see Section 3.10).
A factory SETUP file is available and already selected on a ne w PG7000, but the operator
may desire to customize it. The factory setup file assumes that PG7000’s internal
measurement values will be used whenever possible.
2.4.5 Check Proper Operation of Ambient Condition
Measurements
PG7000 automatically measures ambient conditions and uses these conditions in its
pressure calculations.
To verify that the ambient condition measurements are operating properly proceed as follows:
•Displ ay current ambient condition readings: Press [AMBIENT]. The ambient conditions
run screen is displayed (see Section 3.9.6).
•Verify proper ambient condition readings: Compare the ambient condition values
displayed to the actual values of ambient conditions. Refer to the ambient condition
measurement specifications when evaluating the ambient readings (see Section 1.2.1.2).
Note
•The unit of measure in which ambient pressure is displayed is the same as
the unit selected by pressing [UNIT] ( see Section 3.9.3). Units of measure
in which other ambient condition values are expressed cannot be
changed.
•PG7000 allows the source of ambient condition value s used in reference
pressure calculations to be specified. The source may be PG7000’s onboard measurements, default values or operator entered values. See
Section 3.10 for information on specifying the source of ambient
condition values used by PG7000 in reference pressure calculations.
2.4.6 apply pressure to THE piston-cylinder module
Note
This section assumes that the PG7000 system has already been set up,
including pressure interconnection (see Section 2.3).
Caution
Before appl ying press ure to the PG 7000 syst em, be sure th at all pres sure
vessels and connections are rated for the pressure levels that will be
applied and that all connections have been properly tightened.
Continuing with the PG7000 set up and check out requires applying pressure to the pistoncylinder module and floating the piston.
Proceed as follows:
Turn OFF automated piston rotation (if present) and automated pressure
generation (if present). This will prevent the automated rotation and pressure
generation features (if present) from interfering durin g verification of these features (see
Sections 3.9.8 and 3.9.9 for information on automated piston rotation and
pressure generation).
Load mass on the piston. Install the m ass loading bell on the piston. Then load th e
make up m ass (9, 4.5 or 4 k g depending o n the mas s set) (see Sectio n 2.3.1.3). If the
PG7000 platform is equipped with AMH autom ated mass handling, press [ENTER] and
enter a low value of pressure or mass to cause mass to be loaded onto the piston.
Float the piston. Use the pressure generation/control component of the PG7000 system
to apply pressure un der the piston thr ough the PG7000 Platform TEST port. The piston
will float at a pressure approximately equal to the piston mass to pressure conversion
factor multiplied by the m ass load in kg. The piston-cylinder conversion f actor is marked
on the top of the piston cap and is in either kPa or MPa per kilogram [kPa/kg or MPa/kg].
If the piston cannot be floated becaus e it immediately sinks down from a
float position, there is a leak in the pressure system. Identify and eliminate
leaks until the piston, once floated, falls at a rate less than the nominal f all
rate given in the specifications of the piston-cylinder module being used
(see Section 1. 2. 2).
2.4.7 Check Proper Behavior of Mot orized Piston Rotation
The motorized rotation feature can be set to engage and disengage automatically as needed to
maintain the floating piston rotating within a set rotation rate range when floati ng, and to stop rotati on
before mass manipulati on to set a differen t pressure occ urs (see Sectio n 3.9.8). Motor ized piston
rotation c an a lso b e acti vat ed m anua ll y by press in g [
by pressing and holding [
Turn automated rotat ion OF F by press ing [ROTATE] and selecting <1off>. Loa d at least the
mass bell on the piston, float the piston and press [
motorized rotation system should engage and cause the mass bell and piston to begin
rotating. Within 5 to 30 seconds , depending o n the m as s load, th e rot ation r ate sh ould r each
80 rpm for PG7601 or abo ut 50 rpm for PG7102 or PG7302. Current rotat ion rate can be
observed by pressing [SYSTEM] (see Section 3.9.5).
] and then pressing [←] (see Section 3.9.13).
Note
]. Stopping piston rotation can be activated
]. When the piston is floating, the
2.4.8 Check Proper Operation of Pis t on Behavior
Measurements
Float the piston and rotate it (see Sections 2.4.6 and 2.4.7).
Press [SYSTEM] once to r each the f irst SYST EM run s creen. Ver ify that the piston p osition,
piston fall/rise rate, pisto n rotation rate and piston r otation decay rate ar e indicating correc tly
(see Section 3. 9.5). The piston positi on r ead in g s yst e m may be calibrated us ing an on-board
procedure if necessary (see Section 5.2.2). Calibration of piston position is recommended
when installing a new PG7000 and regularly after installation.
Press [SYSTEM] again to r each t he sec ond SYST EM run s cre en. Ver ify th at the pisto n-cylinder
temperature and temperature rate of change are indicating correctly (see Section 3.9.5).
If checking a PG7601 Platform, verify that the vacu um reference measurem ent is operating
correctly (see Section 2.4.8. 1).
2.4.8.1 Verify Vacuum Reference (PG7601 Only)
PG7601 includes provisions for establishing and measuring a vacuum reference.
To verify the vacuum reference capability:
Install a piston-cylinder module (see Section 2.3.2).
Install the bell jar on the PG7601 Platfor m (the bell jar aligns on th e PG7601
vacuum plate and seals its elf). If using AMH-38 autom ated mass handling,
install the AMH-38 mass handler (see the AMH-38/AMH-100 Operati on and
Maintenance Manual).
Apply a vacuum through the r eference vacuum port (KF25 port on front left
side of platform or KF40 if using the optional AMH-38 automated mass
handler).
Press [SYSTEM] twice to observe the value of vacuum read by the
PG7601’s built-in vacuum gauge.
If a vacuum pump of adequate capacity has been correctly connected to the
reference vacuum port, the vacuum read by the PG7601 built-in vacuum gauge
should go to 4 Pascal [Pa] or lower in less than five minutes on the f irst pump
down and two to three minutes on immediately subsequent pump downs.
Before appl ying press ure to the PG 7000 syst em, be sure th at all pres sure
vessels and connections are rated for the pressure levels that will be
applied and that all connections have been properly tightened.
To check automated pressure generation/control:
Verify that the automat ed pressure generation/control c omponent is properly connecte d
to the system (see Section 2.3.1.2).
When oper ation has returned to the main run screen, turn ON automated pressure generation
control, if available, by pressing [GEN] and selecting <1on> (see Section 3 .9.9) .
Press [P
[ENTER] and enter a pressure value to be generated and follow the mass loading
instruction (see Section 3.6) .
Verify the pressure genera tion/co ntrol com ponent prop erly gener ates press ure and f loats
the PG7000 piston.
OR M], <1pressure> to select press ure entr y mode (see Section 3.9 .12). Press
2.4.10 Check/Set Security Level
PG7000 has a security s ystem based on User Leve ls. By default, the secur ity system is set
to low and NO password is required to change the securit y level. See Section 3.11.4.5 for
information on the s ecurity s ystem. As part of the PG 7000 startup, set your desired securit y
level and a password.
Caution
PG7000 is delivered with the security level set to low to avoid inadvertent
altering of critical internal settings but with access to changing security
levels unrestricted. It is recommended that the low security level be
maintained at all times and passw ord prote ction b e imple mented if cont rol
over setting of security levels is desired.
2.4.11 Additional Pr e c a uti ons t o Tak e Bef ore Ma king
Pressure Measurements
Before using PG7000 to make accurate pressure measurements, consider the following:
•Select/activate the correct piston-cylinder module, mass set and mass loading bell
(see Sections 3.9.2, 3.11.1.10, 3.11.1.15).
•If us ing an automated pressure generation/ control component with automated pressure
generation, set the pressure controller’s upper limit (UL) (see Section 3.9.9.3).
• Enter the correct value of local gravity at the site of use (see Section 3.11.6).
• Consider head corrections (see Sections 3.9.7 and 3.11.3).
• Level the PG7000 Platform properly (see Section 2.3.1.1).
•Select the correct press ure unit of m easure and measurem ent mode (see Sect ions 3.9.3
and 3.9.5).
•Verify that the settings for the sources of variables to be used by PG7000 in its
calculations of reference pressures are those desired (see Section 3.10).
•Verify that the piston-cylinder module is correctly cleaned and operating properly
(see Section 5.3).
•PG7202 only: Ensure that the piston-cylinder module liquid reservoir is filled (see
Section 5.3.3).
•Verify that there are NO leaks in the pressure system.
2.5 Short Term Storage
The following is recommended for short term storage of PG7000.
Remove all masses from the PG7000 Platform (manual mass set only).
Vent all circuits to atmosphere.
Turn OFF power using the power switch on the rear of the PG7000 Terminal.
C over the PG7000 Platform and m as s set with th e du st c overs inc luded in th e pl atf orm and mass set
accessories (not used with AMH mass handling system).
PG7000s operate on the pr inciple of the pis ton gaug e in whic h pressur e is def ined b y balanc ing it agains t
a known force on a known ar ea (see Fig ure 5). T he known are a is def ined b y a vertica lly m ounted pis ton
rotating in a c ylinder an d t h e known force is app li ed t o the pis to n by loading it wit h known mass subj ec ted
to acceleration due t o gravi ty. W hen the f orce app lied b y the pr essur e and the forc e applie d b y the m ass
accelerat e d by gravit y are in eq u i li br i um, the pisto n f l o ats and the pr es sure und er t h e p iston remains constant.
The pressure can be calcul ated following th e equation i n Figure 5 (see also, Section 7.2). Generall y, the
pressurized f luid under the piston als o lubricates the gap bet ween the piston and the c ylinder. PG7202
uses a unique gas operated, liquid lubricated piston-cylinder (see Section 3.1.1).
Figure 5. Piston Gauge Operating Principle
The PG7000 Platform is designed to mount a variety of piston-cylinder sizes, al low pres s ure to b e ap pl ied
under the piston and allo w masses to be loaded on to p of the pis ton. Ther e are diff erent PG700 0 m odels
depending upon whether the pressure medium is oil or gas and whether a vacuum reference is needed.
The measurement uncer tainty in the press ure defined b y the piston gauge d epends on the uncerta inty in
the effective area of the piston-cylinder and the force applied by the mass accelerated by gravity.
PG7000 stores the calibrated values of the piston-c ylinders and masses it uses in on-board files (see
Section 3.11.1). To determine the effective area of the piston-cylinder and the force applied by the
masses under actual oper ating conditions, a number of influences on these values must be quantified
and taken into consideration. For this reason, PG7000 includes extensive features to monitor the
behavior and conditio ns of the pis ton-c ylinder as well as am bient c ondit ions that af fect pr essur e definition
(see Sections 3.9.5 and 3.9.6) . PG700 0 uses the pist on-cylinder, mass and am bient conditio n information
to calculate the pressure defined by a given mass load or the mass load needed to define a given
pressure (see Secti on 7 .2) . T he s our ce of each value used b y PG7000 in its c alculat io ns c an be se lec ted
by the user between PG7000’s internal measurements, default values or user entered values. These
sources are defined in SETUP files (see Section 3.10).
Once PG7000 has been set up, it is used in day-to-day operation either to define pressures applied to a device
or system under test or to measure a stable pressure. To interface with the PG Terminal , the oper ator:
Selects the appropriate piston-cylinder to cover the pressure range (see Section 3.11.1.5).
Selects the desired pressure unit of measure (see Section 3.9.3).
Selects the desired pressure measurement mode (gauge, absolute, differential) (see Section 3.9.4).
Sets the head difference between the PG7000 and the device under test (see Section 3.9.7).
Selects mass to pressure or pressure to mass operating mode (see Section 3.9.12).
Enters a pressure to define or a current mass load to calculate (see Section 3.9.11.1 and 3.9.11.2).
Loads masses by hand or with optional AMH automated mass handler, floats piston and defines
pressure.
3.1.1 Gas Operated, Liquid Lubric a t e d Piston-Cylinder
Operating Princi ple (PG7202)
PG7202 provides v ery high pressure gas pressur e operation using a unique gas operated,
liquid lubricated piston-cylinder system.
The principle of operation of the gas operated, liquid lubric ated piston-cylinder is simple but
very effective (see F ig ure 6 ) . T he measured gas pres s ure, Pg , is a pp lied to the bottom of the
piston and to the top of a liquid reservoir located around the cylinder. The reservoir is
connected to the gap between the piston and the cylinder through lateral holes near the
bottom of the cylinder, al low ing li qu id f r om the res er voir to ent er the gap. The pressur e of the
liquid in the gap, Pl, is equal to t he gas press ure Pg, plus the liquid head, h. Therefore, th e
liquid pressure in t he gap is al ways higher than the g as pressure b y the amount of the liquid
head regardless of the gas pressure value. Since h is small and the space between the
piston and cylinder is typicall y < 1 micron, the bleed of liquid f rom the bott om of the c ylinder
towards the gas press ure is extremely sm all. The mounting post of the 7202 pis ton gauge is
configured so this minute amount of liquid drops directly into a sump that is deadened (see
Section 5.2.4, 3. 1.1) and no t in the flow path of test gas into and o ut of the system . Though
molecules of the liquid m ay migrate through the gas, no sign ificant contaminatio n of the test
system occurs. Because the liquid reservoir is contained in the piston-cylinder module,
piston-cylinders can be removed and instal led in the PG7202 piston ga uge platform with no
loss of liquid fr om the reser voir. All PC-7 200 pist on-cylinder modules can be deliver ed using
Krytox, a fluorinate d synthetic fluid, to lu bricate the piston-cylinder in applicat ions where the
system must r emain perfectly free of hydrocar bons (e.g. when ca librating instrument ation for
oxygen service).
PC-7200 gas operated, liquid lubricated piston-cylinder modules are
delivered with their lubricating liquid reservoir drained. The reservoir must
be filled prior to using the piston-cylind er modul e an d then regula rly whe n
used (see Section 5.3.3)
PG7000 has a 4 x 4 keypad for local operator access to direct functions, function menus and for data entry.
The Function/Data keys
functions to be accessed directly from t he main run screen by
a single keystroke. The name of the function is on the bottom
half of the key (see Section 3.9.1). These keys enter
numerical values when editing.
The Editing and Execution keys
suspending execution, backing up in menus and editing
entries.
The Menu/Data keys provide access t o function menus from
the main run screen. The menu name is on the bottom half of
the key. The SETUP menu is for more frequently used
functions. The SPECIAL menu is for functions that are NOT
generally used as a part of day to day operation. These keys
enter numerical values when editing.
Figure 7. PG7000 Keypad Layout
Key press co nfirm ation is pro vide d by bo th tac tile an d audi ble fe edbac k. A sing le beep c onf irm s a valid e ntr y.
A descendin g two note tone signals an invalid entry. The audible val id entr y feedbac k can be s uppressed or
modified by pressing [SPECIAL] and selecting <5prefs>, <2sound> (see Section 3.11.4. 2).
Pressing the [ENTER/SET P] key generally causes execution or forward movement in the menu tree.
[ENTER/SET P] is also used to enter a command to set a pressure.
Pressing the [ESCAPE] key generall y allows movement back in th e menu tree and/or causes execution
to cease or suspend without changes being implemented. Pressing [ESCAPE] repeatedly eventually
returns to the main run screen. From the main run screen, pressing [ESCAPE] allows momentary
viewing of the PG7000 identification screen.
Pressing the [+/-] key changes a numerical sign when editing. It also togg les through multiple scr eens
when available.
Pressing the [←] and [→] keys when editing a llows reverse and f orward cursor movem ent when editing
data entry. These keys are also used to scroll through choices.
Menu selections can be made by pressing the n umber of the selection direc tly or by pressing [←] and
→] to place the cursor on the number of the desired selection and pressing [ENTER].
[
Note
Some screens go beyond the two lines provided by the display. This is indicated
by a flashing arrow in the second line of the display. Press [
←] and [→] to move
the cursor to access the lines that are NOT visible or directly enter the number of
the hidden menu choice if you know it.
3.3 Sounds
The PG Terminal is equipped with a variable frequency tone device to provide audible feedback
and alarms. Som e sounds can be modified and all sounds can be suppressed (see Section 3.11.4.2) .
Sounds are used for the following indications:
Valid key press
Invalid key press
Piston left end of stroke high
or low
Brief high frequency beep.
Three rapid, low frequency beeps.
Three rapid valid key press beeps. Piston was at low stop or high stop and
just entered spring zone (see Section 3.5).
3.4 Pressure Ready/Not Ready Indication
The three characters on th e top l ine, f ar lef t, of the m ain r un sc reen pr ovide a pr essur e Ready /Not R eady
indication. This indic ation is intended to g ive the user a clear and objective indication of when PG7000
conditions are such that the value of pressure it is defining is valid and in tolerance. T here are three
Ready/Not Ready ind ication c haracters t o indicat e the s tatus of the t hree main R eady/No t Ready criteria.
The Ready/Not Ready indication character s are shown o n the first line, top left hand side, o f the main run screen.
1. Pisto n pos ition and vertical movement.
2. Piston rotation.
3. Reference vacuum (PG7601 in absolute by vacuum measurement mode only).
For each Ready/ Not Rea dy in dicat ion c harac ter , <*> indicates a Ready condit ion. T heref ore, <***> or <** >
indicates that all cond itions necessar y for an in tolerance pr essure definition are present. An y indication
other than <*> indicates Not Ready.
See Sections 3.4.1, 3.4.2 and 3.4.3 for details on each of the three Ready/Not Ready indicating characters.
Note
The criteria used to distinguish between Ready and Not Ready conditions can be
customized by the user (see Sections 3.4.1, 3.4.2, 3.4.3).
The piston position Rea dy/Not Ready c haracter indicates Ready or Not R eady based on the
position of the piston in its vertical stroke (see Section 3.5) and an automated measure
generation status (see Section 3.9.9). This ensures that pressure definitions will be made with
the piston not mor e than a certain dist ance from m id-float position and that meas urements will
not be made while the action of an automated pressure controller may influence the pressure.
The piston position Ready/Not Ready character is t he first character f rom the left on the top
line of the main run screen.
3. GENERAL OPERATION
The piston position Ready/Not Ready criter ion is determined by the current S ETUP file and
can be customized by the user (see Section 3.10).
Piston position Ready /N ot Ready character indications include:
<*>Piston position Ready ( within the posit ion limits specified in the cur rent SETUP f ile)
(see Section 3.5).
<↓>Piston position Not Ready, low (below t he position limits specified in the current
SETUP file, see Sec ti on 3. 10) . T he <
position to alert the user that this indicator is Not Ready.
<↑>Piston position Not Ready, high (above the positio n limits specified in the current
SETUP file) (see Sec tion 3.10). The <
position to alert the user that this indicator is Not Ready.
<?>Piston position not known (current specified mass load is less than the load of the piston
+ bell). The bell must be installed for PG7000 piston position measurement to operate
correctly so piston posi tion values shown are not valid when the bel l is not in stalled.
<T> Tare PG7000 is Not Ready (PG7102 and PG7202 only). Indicates that piston
position or rotation rate is Not Ready on the tare PG7000 in high line differential
mode (see Section 3.9.4.2).
↓> flashes if the pis ton is not at t he b ottom stop
↑> flashes if the piston is not at the to p stop
Note
•Piston position always indicates Not Ready if automated pressure
generation is adjusting pressure, regardless of actual piston
position.
•In PG7102 and PG7202 high line differential pressure mode (see
Section 3.9.4.2), the piston position Ready/Not Ready indicator
applies to the pistons of both the tare and the reference PG7000.
For Ready to be indicated, both pistons must be within the ready
limits for piston position.
3.4.2 Piston Rotation Ready/ Not Re a dy
The piston rotation Rea dy/Not Ready character in dicates Ready or Not R eady based on the
rotation rate of the piston.
The piston rotation Ready/ Not Ready character is the sec ond character from the left on the
top line of the main run screen.
The piston rotation rate Ready/Not Ready criterion is specific to the currently active
piston-cylinder module and can be edited by the user (see Sections 3.11.1.1).
Piston rotation rate Ready/Not Ready character indications include:
<*> Rotation rate Ready: Rotation rate is within the r otation rate limits specified in the
current piston-cylinder module file) (see Section 3.11.1.1).
<> Rotation rate Not Ready, low: Rot ation rate is less than the lo wer rotation rate limit
<
specified in the current piston-cylinder module file (see Section 3.11.1.1)
or motorized rotati on system is currently engaged. Note that the low rotation limit is
automatically reduce d when the mass loaded on the piston is less than 3 kg. The
<> flashes if the piston is floating to alert the user that this indicator is Not Rea dy .
<
<?> Rotation rate not known: Current specified mass load is less than the load of the
(piston + bell). The bell must be installed for PG7000 piston rotation rate measurement
to operate correctly so piston rotation rate cannot be measured when the bell is not
installed.
At mass loads less then 3 kg the low rotation limit is automatically reduced
to an operational minimum to maximize free ro tati o n time .
In PG7102 and PG7202 high line differential pressure mode (see
Section 3.9.4.2), the piston rotation rate Ready/Not Ready indicator
applies to the pistons of both the tare and the reference PG7000. For
Ready to be indicated, both pistons must be within the ready limits for
piston rotation rate.
The vacuum reference Ready/Not Ready character is onl y used wit h PG760 1, as it is the only
PG7000 model with vac uu m ref erenc e capabilit y. It is onl y active whe n oper atin g in absolute by vacuum mode (see Sect ion 3.9. 4).
The vacuum reference Ready/Not Ready charac ter indicates Ready or Not Ready based o n
the value of refer ence vacuum when making measurem ents in absolute by vacuum mode.
This ensures that definition s of abs olute pres sure with a vacu um ref erence will b e made with
the vacuum under the PG7 601 bel l jar lo wer tha n a sp ecified value. When the vacuum is not
low enough, vacuum measurement errors may be excessive.
The vacuum reference Ready/Not Ready char acter is the third ch aracter from the left on the
top line of the main run screen.
The vacuum reference R eady/Not Ready criterion is determined b y the current SETUP f ile
and can be custom ized by the user (see Section 3.1 0). The vacuum reference criterio n is
a fixed value that c an b e c ustom ized b y the user whe n the vacuum ref erenc e selec tion i n the
SETUP f ile is internal. If th e selectio n in the SET UP file is NOT internal, the vacuum refere nce Ready/Not Ready character always indicates Ready and the value cannot be customized.
Vacuum reference Ready/Not Ready character indications include:
< > (Blank) Vacuum referen ce Ready/Not Ready is not in use. This is not a PG7601;
or if PG7601, current measurement mode is not absolute by vacuum.
<*> Vacuum reference Ready. Vacuum value is below limit specified in the current
SETUP file if s ource is internal OR sourc e is default or user and the vacuum value is
not a measured value (see Section 3.10).
< > >Vacuum refer ence Not Ready. Current SETUP file source for vacuum is internal
and vacuum value is above the limit spec ified (see Se ction 3.10). The <
the piston is floating to alert the user that this indicator is Not Ready.
Note
>> flash es if
3.5 Piston Position
When operating a PG 7000 piston gauge, referenc e pressure values are defined by loading k nown mass
values on the piston and adjusting the pressure to float the piston. Piston position is measured and
displayed real time on the MAIN run s creen (see Section 3.7) and in the first SYST EM run screen (see
Section 3.9.5). Piston position is used as a criterion for the Ready/Not Ready indication as valid
measurements can only be made when the piston is in the correct position (see Section 3.4.1).
The full piston stroke is ± 4.5 mm from the midstroke position. The stroke is divided into different
positions and zones as illustrated in Figure 8. These zones are:
•High and low stops: The piston is at the m inimum or m aximum end of str ok e. The press ur e app lied
to the piston is higher (at high stop) or lower (at low stop) than th at corr espond ing to the m ass loaded
on the piston. The high and low stop positions are fixed.
•High and low spring zones: The combination of pressure and the force of the high or low stop
springs have caused t he piston to leave the stop. The pres sure applied to the piston is withi n the
equivalent of 2 kg of the press ure cor respond ing to th e m ass loaded on t he pis ton. T he high an d low
spring zones indicate the piston is about to float. The spring zone positions are fixed.
•High and low measurement zones: The piston is within the zon e in which a Ready measurement
can be made (see Secti on 3.4.1). The pressur e applied to the pisto n is the pressure corr esponding
to the mass loaded on the piston. The default value of the high and low measurement zones is
± 2.5 mm around midstroke position. This value can be adjusted in the SETUP file (see Section 3.10).
•Midstroke: The piston is at the middle of its stroke. The bottom of the piston (or its equivalent
for hollow pistons) is at the reference level marked on the mounting post (see Section 3.9.7).
There is equal stroke available to the high and low stops.
Note
The piston position monitoring system is driven by the internal ring in the base of
the mass loading bell. When the bell is not installed, piston position cannot be
measured. If the current mass load is less than the mass of the piston + bell,
PG7000 will display < ---- > where piston position is normally displayed. If the
current mass load is greater than the mass of the piston + bell but the bell is not
actually installed, the piston position indication is not valid.
Figure 8. Piston Stroke and Zones
3.6 Mass Loading Protocol
PURPOSE
To provide the operator with mass loading instructions and allow mass entry in convenient nominal values
which PG7000 can accurately convert to measured mass values to determine the actual mass load.
PRINCIPLE
All PG7000 manual mas s set masses of 0.1 kg and above, including those in t he mass set, the mass
loading bell and the piston assembly, are adjusted so that their true mass is within ± 20 ppm of their
nominal value. The nominal value is m arked on each m ass. Each m ass’s actual individual value is
measured more accurately than ± 20 ppm and reported in the mass set, mass loading bell or pistoncylinder module calibration report. These measured actual values are used by PG7000 in all of its
defined pressure calcu lations. The nominal mas s values, and the sequent ial numbers of the nominal masses when there are severa l of the same value, ar e used by PG7000 to des cribe the m ass to load or
the mass that is loaded. Following a few simple mass loading rules allows PG7000 to accurately
transform actual mass values into nominal mass values and vice-versa so that mass loading
instructions to the operator and the operator mass entries can be made in simple nominal mass instructions while representing actual individual mass values.
Since the nominal mas s values written o n the masses and the actual values of the individual mass es
are different, in press ure to mass m ode (see Sec tion 3. 9.12) the a ctual m ass v alue loaded on t he pisto n
will be different from the mass loading instruction. In the same manner, in mass to pressure mode
the nominal value of the mass load that the operator enters is different fr om the actual mass loaded
on the piston. The mass value shown in the MAIN run screen (see Section 3.7) is always the
actual mass value. The mass loading instruction given i n pressure to mas s mode and th e mass value
entered by the operator in mass to pres sure mode is al ways the nominal m ass value. If PG7000 mass
loading protocol is f oll o wed , the c on vers i on f r om nominal mass to actual m a ss occurs corr ect ly al lo wing
very simple nom inal mass loading and accoun ting for the operator wh ile using very accurate m easured
mass values for metrological calculations.
OPERATION
Note
To avoid wear to masses and to the piston-cylinder module end of stroke stops,
the piston should not be rotating when masses are loaded or unloaded. Before
loading masses, stop piston rotation by hand or using the motorized rotation
system (see Section 3.9.13).
PG7000 instructions to the operator of m ass to load, and operator entr ies of the m ass that is loaded, are
always expressed i n terms of nominal mas s within 0.1 k g and in term s of grams for values under 0 .1 kg.
This protocol is also followed when using the AMH automated mass handler.
In pres sure to mass mode, the instruction of the mass to load to
achieve the requested pressur e is formatted:
In mass to pressure m ode, the entry of the mas s currently loaded
on the piston is formatted:
kkk.k
kkk.k
kkk.k and gg.gg must be expressed and loaded following the mass loading rules below:
Mass Loading Rules
See Sections 2.3.1. 3,
Installi ng Masses On The Mass Load ing Tra y
and 3.6. for PG 7000 m ass set pr inciples
and protocol information. Refer to display examples immediately above to identify kkk.k and gg.gg.
See EXAMPLES: Mass Loading below for mass loading examples.
PG7000 mass loading entries and instructions are always formatted:
kkk.k kg and nn.nn g
•T he mass set being us ed must be select ed as the ac tive mass s et (see Section 3.11.1.10) and must
be properly set up prior to use (see Section 3.11.1.6).
•kk k.k is made up of the piston, the mass loading bell and mas s set masses of 0.1 kg and greater.
When using an AMH mass set, the binary mass carrier and mass lifting shaft are also included.
• kkk.k always includes the mass loading bell and then the makeup mass, if possible.
• kkk.k is made up of the largest masses possible rather than a combination of smaller masses.
EXAMPLE: To load 5 kg on a PG7601:
USE the piston (0.2 kg) + the mass loading bell (0.3 kg) + the 4.5 kg makeup mass.
DO NOT USE the piston + the mass loading bell + 2 kg #1 + 2 kg #2 + 0.5 kg #1.
When several masses of the same nominal value are included in kkk.k, they are loaded in num erical
sequence, low to high.
EXAMPLE: When loading three 5 kg masses:
USE 5 kg mass #1, #2 and #3.
DO NOT USE 5 kg mass #4, #3, #1.
This rule will be followed automatically if masses are installed and used as described in Section 3.6.
•gg.gg is made up of mass from the gram trim mass set (masses of 50 g t o 0.01 g). Thes e masses
When PG7000 provides mass loading instruct ions and calculate s the true mass of
the mass currentl y loaded, it assumes that the m ass set in use has been set up
correctly (see Section 2.3.1.3). For PG7000 mass loading protocol to operate
properly, the mass set in use must be EXACTLY the mass set that has been
defined by the add and/or edit mass function (see Section 3.11.1.6).
Note
In PG7102 and PG7202 High Line Differ ential mode (see Secti on 3.9.4.2), there are
two types of mass loading instructions: line pressure setting and differential
pressure setting. Both follow conventional PG7000 mass loading protocol. The line
pressure mass instructions are always given with 0.1 kg resolution. This is to
preserve the smaller masses for setting the dif ferential pressure. Th e differential
pressure mass loading instructions follow conventional mass loading protocol
using the masses that remain after the line pressure has been set. It may be
necessary to adjust th e line pressu re value to a ssure th at ad equ ate ma ss es rem ain
to set the desired differential pressure. For example, if the line pressure requires
loading 10.6 kg, the 0.5 and 0.1 kg masses w ill b e used to set the line pressure and
will not be available if they are needed to set the differential pressure. If this
situation occurs, consider adjusting the line pressure so that it is set using a
whole number of kilograms, in this example 10.0 or 11.0 kg.
EXAMPLES: Manual Mass Loading
•Mass load ing i ns truc tion using a PG7601 an d a
35 kg mass set:
12.3 kg and 32.33 g
Load:
piston (0.2 kg)
bell (0.3 kg)
4.5 kg #1 (makeup mass)
5 kg #1
2 kg #1
0.2 kg #1
0.1 kg #1
32.33 g from trim mass set
• Mass loading instruction using PG7302 and an
80 kg mass set:
77.6 kg and 10.45 g
Load:
piston (0.2 kg)
bell (0.8 kg)
9 kg #1 (makeup mass)
10 kg #1 through #6
5 kg #1
2 kg #1
0.5 kg #1
0.1 kg #1
10.45 g from trim mass set
The PG7000 m ain run scr een is its hom e display that i s reached upo n powering u p and from which other
functions and menus are accessed. It is the top level of all menu structures.
The main run screen is where PG7000 is left in n ormal operation. It displays the pressure defined by
PG7000, Ready/Not Ready condition and a variety of other information.
Note
The screen described below is called the MAIN run screen. The SYSTEM and
AMBIENT screens (s ee Sections 3.9.5 and 3.9.6) are a lso run screens in the sense
that all the other functions may be accessed directly from them and pressure or
mass entries may be made from them. See Section 3.8 for a description of
PG7000’s main menu/function structure.
3. <unit>: Current pressure unit of measure (see Section
3.9.3).
4. Current measurement mode. <a> for absolute, <g>
for gauge, <d> for differential or high line differential
(see Section 3.9.4).
5. Automated pressure g eneration st atus. <G> if aut omated
pressure generation is ON. The <G> flashes if
automated pressure generati on is ON and acti ve, blank
if automated pressure generation is not ON (see
1. <prv>: Three Ready/Not Ready indication characters
(from left to right: piston position, piston rotation,
vacuum reference) (see Section 3.4). Eac h character
indicates as follows:
<p>: Piston Position Ready/Not Ready. <*> when
Ready, <
measuring zone, <
below measuring zone, <?> if piston position not
available or out of range. Flashes if condition is Not
Ready and piston is floating (see Section 3.4.1).
Indicates Not Ready when automated pressure
generation is adjusting pressure.
<r>: Piston Rotation Ready/Not Ready. <*> when
Ready, <<> when Not Ready due to pist on rotation rate
too low or motorized rotation engaged, <?> if piston
position not ava ilable or o ut of ran ge. Flas hes if con dition
is Not Ready and piston is floating. (see Section 3.4.2).
<v>: Vacuum Reference Ready/Not Ready
(PG7601 only). <*> when Ready, <v> if Not Ready due
to vacuum not below ready limit. Flashes if condition is
Not Ready and piston is floating (see Section 3.4.3).
2. <PRESSURE1>: Numerical value and sign of the
pressure defined by PG7000 with current mass loaded
on current piston when all Ready/Not Ready indication
characters indicate Ready.
↑> when Not Ready due to piston above
↓> when Not Ready due to piston
Section 3.9.9).
6. DUT head correction status. <h> if a head correction is
currently being applied, blank if head height is zero (see
Section 3.9.7).
7. <NN.NNNNN kg>: Current mass l oad in actual mass
(assuming mass loading instructions have been
followed and/or mass loading entries are c orrect (see
Section 3.6).
8. <± N.N mm>: Current piston position in millimeters
from midstroke position. Positive values are above
midstroke, negative values below. Indicates < ---- >
if piston position not available or out of range (see
Section 3.5).
Note
• When a number is too large to display in
the allocated screen space, PG7000
displays <********> or <OVERFL OW>.
• PG7000 has a screen saver function that
causes the display to dim if NO key is
pressed for 10 minutes. Pressing a key
restores full power to the display. The
screen saver activation time can be
changed or screen saving can be
completely disabled (see Section 3.4.5.1).
Local operator inter face with PG7000 is provided by PG T erminal keypad and dis play. Normal PG7 000
operation is organized around run screens from which functions and menus are accessed. The run
screens include:
• MAIN run screen: Displays Ready/Not Ready, defined pressure, piston position and mass load.
• (2) SYSTEM run screens: Display piston position, rotation, temperature and reference vacuum
(PG7601 only).
•AMBIENT run screen: Displays ambient pressure temperature, relative humidity and local gravity.
The flow chart below outlines the operating protocol of the run screens and associated key presses.
Local operator interface with PG7000 is provided by PG Terminal keypad and display.
To minimize the use of multi-layered menu structures, the keypad’s numerical keys also
provide direct access to the most commonly used functions. The function accessed is
labeled on the bottom half of the eac h k e y. Direct f unction keys are active whenever PG7000
is in a run screen (MAIN, S YSTEM or AMBI ENT). T able 14 summ arizes the ope ration of the
direct function keys. Section 3.8 provides a flow chart of PG Terminal general operation.
See corresponding manual Sections to 3.9.1 to 3.11.9.
It may be usefu l to keep a copy of Tab le 14 near the PG70 00, especially
when first becoming acquainted with its operation.
Table 14. Summary of PG7000 Direct Function Key Operations
Note
DIRECT FUNCTION KEYS ARE ACTIVE FROM ANY RUN SCREEN
See corresponding manual sections for full detail.
Menu of less frequently used internal
functions and settings including resets,
user preferences, internal calibration
and remote interface setup.
Edit and select files that determine the
sources of the values for the variables
PG7000 uses in calculations of reference
pressures.
Select pressure to mass or mass to
pressure operation.
Activate motorized control manually.
Activates when pressed, deactivates
when released.
From a run screen: Access mass or
pressure setting commands.
From other screens: Select menu
choice, enter value, confirm selection.
Turn automatic activation of motorized
rotation ON/OFF.
(main, ambient, system).
Set the resolution of PG7000 mass
loading commands (0.01 g to 0.1 kg).
Access SYSTEM run screens (2) which
display current measurements of piston
behavior, piston-cylinder temperature
and reference vacuum if present.
Toggles between the two screens.
Access AMBIENT run screen which
displays current values of barometric
pressure, ambient temperature, ambient
humidity and local gravity as specified in
the active SETUP file.
Adjust height of DUT fluid head
correction. Set to zero to defeat head
correction.
Select active piston-cylinder module
(range).
Set unit of measure in which pressures
are defined. Choice of units available
from this key can be customized.
Turn automated pressure
generation/control ON/OFF, adjust
control parameters, set upper limit for
automated pressure control. Has no
effect if an automated pressure
generation/control component is not part
of the PG7000 system.
Set pressure measurement mode
(gauge, absolute, differential).
3. GENERAL OPERATION
Active 10 kPa/kg 247
0.7 to 50 psi
Select 200kPa/kg 382
14 to 1000 psi
3.9.2 [P-C]
PURPOSE
To view and/or change the active piston-cylinder module. In most cases, changing the
piston-cylinder module is equivalent to changing the range.
Note
•[P-C] is for selecting the active piston-cylinder module only. Piston-
cylinder mod ules can b e added and d eleted and th eir charact eristics can
be edited by pressing [SPECIAL] and selecting <1PC/MS> (see Section
3.11.1).
•The active mass set and mass loading bell are selected by pressing
[SPECIAL] and selecting <1PC/MS> (see Section 3.11.1).
PRINCIPLE
To make valid pressure and mass load calculations, PG7000 must know the exact
characteristics of the piston-cylinder module that is currently in use. See Section 7.2 for
detailed information on PG7000 pressure and mass calculations.
Most PG7000s are used with m ore than one p iston-c ylinder m odule. Detai led characteristics
on up to 18 piston-cylinder m odules are stored in fi les. These files can be added, deleted,
viewed and edited by pressing [SPECIAL], 1pc/ms (see Section 3.11.1).
The P-C function pro vides rapid access, f rom any run screen, to view ing the piston-cylinder
modules available and selecting one to be active.
OPERATION
Pressing [P-C] activates the pisto n-cylin der module viewing and s electing f unction. Pr essing
the [P-C] key again or [+/-] while in the P-C function steps through displays of available
piston-cylinder modules.
When [P-C] is first pressed, a summary of the characteristics of the active piston-cylinder
module is displayed, for example:
1. Nominal pressure to m ass conversion coefficie nt of the pistoncylinder module t hat is currently s elected (acti ve). This value is
calculated f rom the effective area of the piston-cylinder and is
in kPa (if < 1 MPa) or MPa. Upgraded Type 5000 pistoncylinders may be displayed in psi.
2. Serial number of the active piston-cylinder module.
3. Current pressure unit of measure.
4. Nominal pressure in current pressure units resulting from
loading all the mass of the active mass set.
5. Nominal pressure in current pressure units resulting from
loading the piston and mass bell only.
Pressing [P-C] again or [+/-] causes the screen to step through the other available piston-cylinder
modules in the sequence that they were added.
1. Nominal pressure to m ass conversion coefficie nt of the pistoncylinder module. This value is c alculated f rom the eff ective ar ea
of the piston-cylinder and is always in kPa (if < MPa) or MPa.
2. Serial number of the piston-cylinder module.
3. Current pressure unit of measure.
4. Nominal maximum pressure using the piston-cylinder
module. Pressure, in current pressure units, resulting from
loading all the mass of the active mass set.
5. Nominal minimum pressure using the piston-cylinder module.
Pressure, in current pressure units, resulting from loading the
piston and mass bell only.
Pressing [ENTER] while in the P-C func tion caus es PG7000 t o select t he currentl y displa yed
piston-cylinder module as the active piston-cylinder module.
PG7000™ OPERATION AND MAINTENANCE MANUAL
function key from any run screen. The
1kPa 2Mpa 3Pa
4bar 5psi 6kcm2
Select inWa ref temp
4°C 20°C 20°F
Pressing [ESCAPE] while in the P-C function returns to the main run screen without
changing piston-cylinder modules.
•The pressure unit of measure in which the range offered by the piston-
cylinder mo dule using the active mass set i s d i splayed can b e changed by
pressing [UNIT] (see Section 3.9.3).
•When the current pressure unit of measure is an altitude unit, piston-
cylinder mo dule range i s e xp re s sed in kPa if t h e a ltitude un it is meters (m)
or psi if the altitude unit is feet (ft).
3.9.3 [UNIT]
PURPOSE
To specify the unit of measure in which defined pressures are expressed.
PRINCIPLE
PG7000 supports a wide variety of pressure units of measure. Internally, PG7000 always
operates in Pascal [Pa] (the SI uni t of pr essur e). Values of pressure ar e represent ed in other
units by the application of c onversion factors to convert from Pasc al. See Section 7.1.1 for
Unit of Measure Conversions .
The pressure unit of measure selection (e.g., psi, kPa, etc.) is separate from the pressure
measurement mode selection (gauge, absolute or differential). See Section 3.9.4 for
information on changing the measurement mode.
OPERATION
Note
To change the pressure unit of measure, press the
[UNIT]
display is (default units depend on model):
The cursor is on the n um ber corr es p o nd i ng t o t h e active uni t. T o change t he pr es sure unit , s el ect
the desired unit. The display returns to the previou s run scre en with the sel ected uni t active .
If the pressure unit selected is inWa, the reference
temperatur e f or wa ter densit y m us t be s pecified. When
inWa is selected as the unit , the next di splay is:
Select the desired r eferenc e temper ature for water de nsity usi ng the [←] or [→] key to move
the cursor. Pressing [ENTER] returns to the main run screen with inWa based on water
density at the selecte d reference tem perature as the activ e pressure unit. T he current inW a
reference temperature can be viewed by observing the position of the cursor in the reference
temperature screen.
Note
•See Section 7.1 for tables of the pressure unit of measure conversion
factors used by PG7000.
•The UNIT function provides rapid access to a choice of up to six units.
The choice of units can be customized from a wider s elect ion by th e user
(see Section 3.9.3.1). The default units of the UNIT function dep end on th e
PG7000 model. The default units are:
• PG7102 and PG7601: 1kPa 2MPa 3Pa 4bar 5psi 6kcm2
• PG7202 and PG7302: 1MPa 2k Pa 3bar 4 psi 5 kcm2
• To restore the default UNIT function settings, use the Unit Reset (see
pressure) ar e always represented in P ascal [Pa], regardle ss of the active
pressure unit of measure.
•When the current pressure unit of measure is an altitude unit, atmospheric
pressure in the AMBIENT run screen is expressed in kPa if the altitude unit
is meters (m) or psi if the altitude unit is feet (ft).
3.9.3.1 Customizing Pressure Units Available Under the
UNIT Function
PURPOSE
To customize the selection of pressure units that are ava ilable for selection fr om
the [UNIT] function key.
PRINCIPLE
The UNIT function provides a choice of different pressure units of measure
depending on the PG7000 model. The units that are available by default are
those indicated in th e Note in Section 3.9.3. Howe ver, PG7000 supports m any
other pressure units of measure. Ot her units can be m ade a vai la ble for selection
and units can be deleted b y custom izing the UNIT function. T his allows PG 7000
to offer a very wide selection of units while simplifying day to day operation.
The typical user w ill customize the [UNIT] function key to support the s ix most
commonly used units.
OPERATION
To customize the [UNIT] f unction key, from the m ain run screen press [SETUP]
and select <2PresU>. The display is:
1. The UNIT number in the [UNIT] menu that is
to be changed.
Enter the number of the unit position that
you would like to change. The display
becomes:
Select the desired pressure unit category. SI units include units based on SI
such as mmHg. Select the desired unit from the unit list (see Table 15).
The UNIT function display is not required to display six units. To delete the
current unit from the U NIT screen and show no un it, select <2other>, <6none>
for that unit number.
If <4user> was selected, the user unit must be defined. The display is:
1. Entry field.
Enter the num ber of user units per Pas ca l [Pa] in the e ntry fi eld. Pr essi ng [ENTER]
defines the user unit and returns to the <Set up unit #> screen.
•See Section 7.1.1 for the pressure unit of measure
conversion factors used by PG7000.
•The user defined unit can be assigned a user defined label
using the UDU remote command (see Section 4.3.4.2).
3.9.4 [MODE]
Note
PURPOSE
To specify the meas urement mode (gauge, absolute, diff erential) in which PG7000 defines
reference pressures.
Note
Pressure measurement mode and unit of measure a re not the same thing.
See Section 3.9.3 for i nf orm ati on on t he pr es sur e un it of me asu re .
PRINCIPLE
Pressure defined relative to absolute zero or vacuum is generally referred to as absolute
pressure.
Pressure defined relat ive to atm ospheric press ure is g enerall y ref erred to as gau ge press ure,
with pressure below atmosphere called negative gauge pressure.
Pressure defined relati ve to another pressur e that m a y be, but is not n ecess aril y atm ospheric
pressure, is generally referred to as differ e nti al pres sur e.
PG7000s (depending on the model) can define absolute, gauge, negative gauge and
differential pressures. These are referred to as measurement modes.
The specific PG7000 measurement modes are as follows:
•gauge: Defines gauge pre ssure by leaving the PG7000 mass load open to atm osphere.
This measurement m ode is supported b y all PG7000 m odels. It does not allo w negative
gauge pressures. The minim um gauge pressure is the pressure resulting f rom loading
the combined mass of the piston and the bell on piston-cylinder effective area.
•absolute by adding atmospheric pressure (absolute by ATM): Defines absolute
pressure by defining a pre ssure in the same manner as in gaug e mode and adding the
value of atmospher ic pressure measured by a barometer. T his measurement mode is
supported by all PG 700 0 m ode ls . T he b ar ometer can be PG700 0’s on-board sens or or a
remote RS232 barometer. The uncertainty on the value of atmospheric pressure
measured by the barometer must be considered but this value can become relatively
small as pressure goes u p . Fo r example, if using a ± 0 .0 1 % b arometer , t h e a dded uncer t a in ty
at 1 MPa (150 psi) is 10 ppm.
This mode is more convenient an d less c ostly than ab solute by v acuum si nce no vacuum
reference needs to be es tablished. Howev er, it does not allo w absolute pr essures under
atmosphere and the minimu m absolute press ure is atmospher ic pressure plus the pressure
resulting from loading the c ombined mass of the piston and the b ell on the piston-cylinder
effective area.
•absolute by vacuum (absolute by vac) (PG7601 on ly): Defines absolute pr essure by
measuring relative to an evac uated bel l jar. T his mode is tim e consum ing as the vacuum
under the bell jar mus t be made and brok en to adjust the m ass load for each pres sure to
be set. This mode is required for settin g absolu te pres sur es under a tm ospheric press ure
and for lowest uncertainty under about 1 MPa (150 psi).
•differential (dif) (PG7601 only): Defines differential pressures at an absolute static
pressure between vacuum and two atmospheres by defining an absolute pressure
relative to an evacuated bell jar and subtracting static pressure monitored by a digital
pressure indicator. An offsetting technique ensures that only the digital indicator’s
resolution and very short term repeatability influence the measurement results.
Allows positive and negative differential pressure (i ncluding gauge pressures) with one
common hardware setup. Cov ers pr ess ur es very near and at zero without lim itati ons due
to piston-cylinder size and mass loads (see Section 3.9.4.1).
•high line differential (HLdif) (PG7102 and PG7202 only): Defines differential
pressures and at g aug e s ta tic pres s ure bet w ee n t he lo wes t g aug e pr ess ur e s upp orted by
the piston-cylinder and t he maximum gauge pressure supported b y the piston-cylinder.
Requires the use of a PG7202 or PG7102 as the “refer ence” and a secon d gas oper ated
PG7000 as the “tare”. Differential pressures are defined by setting a common line
pressure on both PG7000s and then adding the differential pressure to the reference
PG7000 (see Section 3.9.4.2) .
OPERATION
To change th e press ure m easur em ent m ode, pr ess [MODE] from an y run scr een. The res ulti ng
display depends on the PG7000 Mode l:
The cursor is on the number corresponding to the current measurement mode.
For PG7102, PG7202 and PG7302, selecting <2abs> accesses absolute by ATM mode
(see PRINCIPLE above).
For PG7102, PG7202 and PG7302 selec ting <3HLdif> accesses high line differential mode
(see PRINCIPLE above and Section 3.9.4.2).
For PG7601, selecting <2avac> accesses absolute by vacuum mode. Selecting <3aatm>
accesses absolute by ATM mode. Selecting <4dif> accesses differential mode
(see PRINCIPLE above and Section 3.9.4.1).
Making a measurement mode select ion returns to the previous run scr een with the selected
mode active.
Note
When using an AMH automated mass handler, be sure to apply drive
vacuum to the AMH when switching operation to absolute by vacuum
mode (see the AMH-38/AMH-100 Operation and Maintenance Manual).
To define gauge press ures near and under atm ospheric press ure that cannot be
covered in conventional gauge mode; to define pressures at a static pressure
near atmospheric pressure but other than atmospheric pressure.
PG7000™ OPERATION AND MAINTENANCE MANUAL
(nominally equal to RPM
)
to RPM test port
Pg
=
Differential pressure defined by PG7601
Pg
=
Absolute pressure defined by PG7601
RPM
=
Current RPM indication of static pressure
the operating static pressure
Differential mode operation is only supported by PG7601.
Differential mode operation requires that the PG7601 SETUP
AtmP selection be for an external barometer connected to
COM2 (see Section 3.1 0, 3.11.5.4). Differential mode op eration
also requires the PG7000 Differential Mode Interconnections Kit
P/N 3070127, which includes the Differential Mode Controller.
PRINCIPLE
Differential mode is des igned to allow PG7601 to define pressures wh ich piston
gauges have typicall y not covered or for which they have be en difficult to use.
This includes positive and negative gauge pressures right around zero, for
example ± 2.5 kPa (10 in. H2O); as well as low differential pressures at low
absolute static pressur es, for exam ple 0 to 2.5 kPa (20 T orr) at 15 kPa absolute
(2.2 psia) static pressure.
The principal of differential mode is to define a differential pressure as the
difference between an absolute pressure relative to vacuum and a second
absolute pressure (the static pressure) measured by a precision Reference
Pressure Monitor (RPM). An offsetting procedure “tares” the RPM so that the
uncertainty contribute d to the differentia l pressure b y the RPM’s measurem ent is
a function of the RPM resolution and short term repeatability rather than its
absolute accuracy and long term stability.
For differential mode opera tion, the static press ure is set to the desired valu e as
measured by the RPM (atmospheric pressure for gauge pressure
measurements). Then RPM offset at the static pressure is determined by
applying the static pressure as defined by the PG7601 following:
Note
offset = PGRPMstart - RPMindicate
RPM
Where:
PGRPMstart
RPMindicate
Pressure applied by PG7601 to RPM
=
start
RPM reading when PGRPMstart is applied
=
Once the RPM offset has been det ermined, differentia l pressures are def ined by
subtracting the static pressure m easured b y the RPM f rom an absolute pr essure
defined by the PG7601 following:
dif = PGabs - RPMcurrent - RPMoffset
PG
Where:
dif
abs
current
RPMoffset
Disagreement between PG7601 and RPM at
=
For differential mode o peration, the PG7000 Differential Mode In terconnections
Kit should be used to pro vi de th e DU T high an d lo w te s t c onnec tions an d s upp or t
the RPM offsetting procedure (see Figure 9 and Section 3.9.4.1). A precision
Reference Pressure Monit or, s uch as a Dhi RPM, m ust be s electe d f or the Atm P
source in SETUP (see Section 3.10, 3.11.5.4).
PG7601 differential measurement mode manages the data acquisition and
handling to support differential mode operation without operator effort.
Differential mode operation requires setting the static pressure at which
differential meas urements are to be m ade and regu lar determ ination of the R PM
offset. These functions are supported under [MODE], <4dif>.
Technical Note 9940TN02 provides more detailed information on
differential mode principles and includes a complete
uncertainty analysis. Consult Fluke Calibration or visit
www.dhinstruments.com.
OPERATION
To select differential mode operation and access differential mode functions
press [MODE] and select <4dif>.
Differential mode operation includes:
•Select ing differential mode, setti ng static pressure and find ing the RPM
offset (see Section 3.9.4.1,
Pressure, Finding RPM Offset (<1run>)
•Operating in differential mode (see Section 3.9.4.1, . Operating In
Differential Mode
).
•Viewing differential mode static pressure and RPM offset (see
Section 3.9.4.1,
To operate in differential mode and/or execute the RPM offset
determination procedure, the AtmP selection under SETUP must
be external (COM2) (see Section 3.10). If thi s is not the case, an
error messa ge will be dis played (<Dif mode require s COM2 AtmP
SETUP>) when attempting to activate differential mode.
To select diff erential mode oper ation and/or execut e the RPM offset determ ination
procedure, press [MODE] and select <4dif>, <1run>. The display is:
1. Current value of RPM offset. Always in Pascal
[Pa].
Select <1no> to us e the existing RPM off set and return the previous run s creen
in differential mode (see Section 3.9.4.1,
The RPM offset should be redetermined any time the static
pressure value is change d, as well as at the beginning of each
calibration or test sequence. If the difference in the value of the
RPM offset in subsequent determinations is < 0.1 Pa, then less
frequent offset determinations may be considered.
Select <2yes> to proceed with a new determination of the RPM offset
(see Section 3.9.4.1, PRINCIPAL. The display is:
Note
Operating in Differential Mode
Note
).
1. Current pressure reading of the RPM in the
current units of.
Table 16. Valve Settings for Setting Differential Mode Static Pressure
VALVE VALVE SETTING
Control Selection Static P
PG Shutoff Closed
Test Bypass Open
Static P Selection Off
If static pressure is atmospheric
pressure VENT or OFF
With the Differential Mode Controller valves (see Figure 9) set as indicated
in Table 16 use the system’s pressure control component to set the desired s tatic
pressure value as read by the RPM.
If the desired static pr es sure is atmospheric pressure (for ex am ple, f or low ga uge
and negative gauge diff erential pressur es), set the STAT IC P SELECTION valve
to VENT.
If a REFERENCE VOLUM E is connected to the STAT IC P port and the desired
static pressure is less than atm ospheric pressure, c onsider settin g the STATIC P
SELECTION VAL VE to SUPPLY. This will connect the pressure s ettin g vacuum
source directly to the reference vol ume and reduce pull dow n time.
Be sure the TEST BYPASS valve is in the OPEN position before
adjusting the static pressure. Failure to do so will cause
differential pressure to be generated across the HIGH and LOW
TEST ports which may overpressure the DUT.
When the static pressure, as indicated
by the RPM, has been set to the desir ed
value, press [ENTER]. PG7000 reads
the RPM and calculat es th e mass to load
on PG7000 to set the pressure value
indicated by the RPM. The display is:
Note
When using an AMH autom ated mass handler, the mass value
to set the value indicated by the RPM is always loaded with
resolution of 0.1 kg. This is to avoid having to break the
reference vacuum to load trim masses. The differential
pressure values are loaded with the resolution specified in
[RES] (see Section 3.9.10).
This is the st andar d nom ina l mass load ing in i nstruc tio n displ ay of pr ess ure to m ass
mode (see Section 3.9.11.1). The mass loading instruction is always given with 0.01
g resoluti on regardless of t he current mass lo ading resolution s etting (see Sect ion
3.9.10). Lo ad t he m ass val ue f oll owi ng PG70 00 m as s lo adin g pr otoc o l (se e Sec tion
3.6). If an AMH-38 automated mass handler is being used, the mass is loaded
automatically and de fault mas s loading reasolu tion of 0. 1 kg is used.
Set the PG7000 Diff erenti al Mod e Contr oller (see F igur e 9) to a pp ly the pr essur e
defined by the PG7000 to the RPM (see Table 17).
Table 17. Valve Settings to Apply PG7000 Pressure
to the RPM for Differential Mode Offsetting
VALVE VALVE SETTING
Control Selection PG
PG Shutoff Open
Test Bypass Open
Static P Selection Off
If static pressure is atmospheric
pressure VENT or OFF
Once the valves have been properl y set and the mass has been loaded, install
the bell jar and establish vacuum under the bell jar. Then press [ENTER].
The display is:
1. Current pressure reading of the RPM in the
current units of measure.
2. Standard main run screen piston position
indication (see Section 3.7).
3.
Current RPM offset (pressure applied by
PG7000 - RPM indication). Indicates
******* if the value is > 9999.99 Pa.
Pressure unit of measure is always
Pascal [Pa].
Float and rotate the PG7000 piston. When the PG7000 is ready
(see Section 3.4), press [ENTER]. Wait about 6 seconds while the PG7000
takes and averages multiple readings. The display is:
1. RPM offset currently in use.
2. New RPM offset just measured.
To record the new RPM off set and go to the previous run screen in differential
mode with the new RPM offset active, press [ENTER].
To return to the <New offset?> screen, pres s [ESCAPE].
See Section 3.9.4.1,
Operating in Differential Mode
mode operation.
• The value of the RPM offset typically should be inside of ±
20 Pa. An RPM offset value outside of this range may
indicate a poorly calibrated RPM or a problem with the
offset determination procedure.
• The difference between the RPM offset at the beginning
and at the end of a test will be reflected as zero drift of the
differential pressures defined during the test. To quantify
this value, perform the RPM offset procedure at the
beginning and at the end of a test and evaluate the
difference between the starting and ending RPM offset
values.
Operating in Differential Mode
, for information on diff erential
Note
Caution
Differential mode is often used to calibrate DUTs with very low
differential pressure ranges. These can easily be over
pressured and damaged by inadvertently applying excessive
pressure to one port. OPEN the MPC1-D BYPASS valve to zero
the differential across the DUT when there is a risk of
overpressure.
From a practical standp oint, operating in differential mode is near ly identical to
operating i n absolute by v acuum mode (see Section 3.9.1 1). Unless and AM H-38
automated mass handler is used, vacuum under the bell jar must be broken to make
mass load changes and ree stabli shed to set a pressure .
Operation is in either pressure to mass or mass to pressure mode
(see Section 3.9.12).
To operate in differential mode, set the PG7000 Differential Mode Controller
(see Figure 9) to apply pres sure defined by PG7000 to the hi gh side of the DUT
(see Table 18).
Table 18. Valve Settings for Operating in Differential Mode
VALVE VALVE SETTING
Control Selection To adjust pressure under PG7000 piston: PG
To adjust static pressure: STATIC P
PG Shutoff Piston is floating or near floating: OPEN
Any other time: CLOSED
Test Bypass To make differential pressure measurements: CLOSED
To set or check DUT zero: OPEN
Static P Selection OFF
If static pressure is atmospheric pressure: VENT or OFF
Note
For the highest quality differential mode measurements, use
consistent conditions for static pressure and PG7000 vacuum
reference pressure at each point. Set the PG7000 Differential
Mode Interconnections Kit CONTROL valve to STATIC P to
readjust the static press ure to the same value at each point if
necessary. This valu e sh ou ld b e th e same as t he v alu e at which
the RPM offset was determined.
Typical Differential Mode Operation al Sequence
See Figure 9 for a schematic of the Differential Mode Controller and its valves.
Set RPM offset at desired static pressure (see Section 3.9.4.1, Selecting
Select pressure to mass or mass to pressure mode (see Section 3.9.12).
Put the Diffe re nti al M ode C ont roll e r P G SH U TOFF v alv e in CL OSED po siti on.
If desired, read DUT output at zero differential pressure with
TEST BYPASS valve open. Put TEST BYPASS valve in CLOSED position. If
desired, put CONTROL valve in STATIC P position and use system control
component to adjust static pressure to starting value.
Press [ENTER] and enter a pressur e or mass value . Load mass as instructed
on PG7000.
Install bell j ar on PG7000, shu t the PG7000 vac uum vent valve , open vacuum
reference shutoff valve. Wait for vacuum under bell jar to reach Ready
condition. If using AMH-38 automated mass handler and mass loading
resolution of 0.1kg, mass can be changed without breaking vacuum.
Put CONTROL va lve in PG posi tion and use s ystem contr ol compo nent to float
PG7000 piston. Slowly put PG SHUTOFF valve into OPEN position. Readjust
pressure to float the PG7000 piston if necessary.
If desired, p ut CONTROL valve in STATIC P posi tion and use system control
component to adjust static pressure to starting value.
When PG7000 indicates Ready on all Ready/Not Ready indicat ors, take DUT
reading at di ffe rent ial p re ssure in di cated on t he to p li ne o f th e P G7000 di s play .
Slowly put P G SHU TOF F v alve in C L OSED po si ti on.
Shut vacuum reference shutoff valve, open vacuum vent valve. Wait for
pressure un der bell jar to r eturn to ambie nt. Remove be ll jar. This ste p is not
necessary is using AMH-38 automated mass handler and mass loading
resolution of 0.1 kg.
Repeat Steps through for each desired differential pressure point.
See Figure 9 for a schematic of the Differential Mode Controller and its valves.
Zero differential pressure can be set by PG7000 or simply by setting the
Differential Mode Controller TEST BYPASS valve to the OPEN position which
connects the HIGH and LOW TEST ports together. The TEST BYPASS valve
OPEN position is a “true” zero. The zero differential pressure value set by
PG7000 has the same uncer tainty as any other differential pressure point (see
Section 1.2.4.1). Com paring the DUT reading at zero set b y TEST BYPASS in
the OPEN position and at zero set by PG7000 in differenti al mode like any other
differential pressure, can help quantify the quality of the current RPM
offset value. If the difference between the two zero readings is excessive, the
RPM offset value should be redetermined (see Section 3.9.4.1,
Differential Mode, Setting Static Pressure, Finding RPM Offset (<1run>)).
Also, consider correcting all the DUT readings by the difference in the zero
reading found in the two different conditions to correct for the defect in the
RPM offset.
Viewing Static Pressure and RPM Offset (<2view>)
To view the current RP M offset and the static pr essure value at which the offset
was recorded, press [MODE] and select <4dif>, <2view>. This display is:
Selecting
1. RPM offset value that is currently active.
2. PG7000 pressure at which RPM offset was
recorded.
Press [ESCAPE] to return to the <Differential mode:> menu.
Press [ENTER] to return to the previous run screen.
3.9.4.2 HIGH LINE Differential measurement mode (PG7102,
PG7302 and PG7202 ONLY )
PURPOSE
To define dif ferentia l pressu res at ele vated line pr essures up to 11 MPa ( 1 600 psi)
with PG7102 or 110 MPa (16 000 psi) with PG7202 in gas and up to 500 MPa
(72 500 psi) in oil.
Note
High line differential (HL dif) mode operation is only supported
by PG7102, PG7202 and PG7302. HL dif mode operation
requires the use of a two ident ical PG7102, PG720 2 or PG7302s
in tandem. Both PG7000s must have nominally identical pistoncylinders and one of the PG7000s must have a “tare” pistoncylinder. The HL dif mode setup for pressure up to 110 MPa (1
600 psi) in gas normally also includes an MPC1-D-1000 or
MPC1-D-3000 manual pressure controller and the PK-7100MPCD-DIF Interconnections Kit P/N 3070226. The HL dif mode
setup for pressure greater than 110 MPa uses GPC1 (gas),
OPG1 or MPG1 (oil) and custom interconnecting hardware.
Technical Note 0080TN03 p rov id es more d etailed i nformation on
high line differential mode principles including a complete
uncertainty analysis. Consult Fluke Calibration or visit our
website,
High line dif ferential mode is designed t o allow two PG7102, PG7202 or PG7302
piston gauges to be used together to define differential pressures relative to line
pressures significantly above atmospheric pressure. This capability is most often
used to test or c a l i br at e differential pressure devices designed to operate at elevated
line pressures under their normal operating line pressure conditions.
The main challenge of def ining differential pressures at elevated line pressures
comes from the very high r atio of li ne pressure to differ ential pr ess ur e. Re lat iv el y
small errors and instabilities in the line pressure are very large relative to the
differential pressure.
The principal of high line differential mode is to “cross float” two PG7000 piston
gauges so that they define a common line pressure. After the crossfloat is
completed, a BYPASS valve is closed isolating on e PG7000 from the other. T he
“tare” PG7000 maintains the line pressure. Mass is added to the “reference”
PG7000 to define differential pressures “on top of” the line pressur e. The ver y high
precision of the PG7000 gas operated piston gauges allows them to set and
maintain a common line pressure with uncertainty much lower than the overall
measurement uncertainty on either piston gauge. The two PG7000s are
crossfloated at the line pressure prior to making differential measurement to
minimize the contribution of line pressure errors to differential pressure. The
crossfloating procedure consists of making mass adjustments on the tare
PG7000 so that both pisto ns, when connected toget her at the line pressure, fal l
at their “natural” drop rate . PG7000’s embedded pist on fall rate measurem ents
and calculations capabilities are used to assist the operator in performing the
crossfloat.
The PG7000 that is us ed o nly to maintain the l in e pressure on the lo w s id e of t he
device under test (DUT ) is designated t he tare PG70 00. The tare PG 7000 is a
standard PG7000 but it is normally used with a tare piston-cylinder and a tare mass
set (a standard pist on-cylinder and m ass set may also be used) . A tare pistoncylinder effective area is entered as the nominal effective area and its
characteristics include k(P), which must be entered in the piston-cylinder
definition file (see S ec tio n 3.11. 1.1) . A tare mass s et has th e s ame configuration
as a standard m ass set but the exact values of the masses are not measured
and the mass set is s et up using nominal values for each m ass measured value
(see Section 3.11.1 .6). T he tare piston m ass is adj usted to b e slightl y under t he
nominal piston mas s. This assures that the tare side will always be th e light sid e
when crossfloated with the reference side. If a tare piston-cylinder and mass set
is not used on the tare PG7000, the tare side may not be the light side in a
crossfloat. In this case, the tare mass load on the reference side should be
increased slightly (ge nerally < 1 g) to ensure that i t is heavier, tak ing care not to
confuse the added tare mass with the subsequent differential mass load.
The PG7000 that is used to apply the added pr essure to create the differential
pressures on the high side of the DUT is designated the reference PG7000. The
reference PG7000 is a standard PG7102, PG7202 or PG7302 with a standard
piston-cylinder and mass set. The reference must be a PG7102, PG7202 or
PG7302. The tare may be a PG7102, PG7202, PG7302 or a PG7601.
The tare and reference PG7000s mus t have nom inally identic al piston-cylinders.
The two PG7000s operate together with the reference acting as “master” and
the tare as “slave”. For communications between the two to occur, C OM2 of the
reference m ust be connec ted to COM1 of the tare wit h a valid RS232 c able and
proper settings must be made on both interfaces (see Section 3.11.5.1).
For high line differential mode operation up to 11 MPa (1 600 psi) in gas, an
MPC1-D and the PK-7100-MPCD-DIF Interconnections Kit P/N 401645 should
be used to provide the DUT high and low test connections and support line
pressure and differential pressure setting procedures (see Figure 10). For
operation at pressure grea ter than 11 MPa, GPC1 for gas, OPG1 or MPG1 for
oil, and custom interconnecting hardware are used in a functionally similar setup.
PG7000 high line d ifferential measurement mode m anages the data acquisition
and handling necessar y to suppor t dif fer ential mode o perati on. Dif fer ential mode
operation requires setting the line pressure including crossfloating the two
PG7000 piston gauges and then setting differential pressures at the line
pressure. These two independent operations are supported under
[MODE], <3HLdif>, <1run>.
• High line differential mode relies upon the very high
sensitivity of PG7000 gas operated piston-cylinders to set
and stabilize low differential pressures relative to very high
line pressures. To meet the full performance potential of
PG7000 high line differential mode operation, external
influences on the piston gauges must be minimized. Air
currents and vibrations are the most significant
possible influences. Do not operate near an active air
conditioning or heating duct, avoid opening and closing
doors or any movement of personnel around the system.
Consider putting the PG7000s in an electrostatic free
enclosure if the environment cannot be adequately
controlled.
• See Section 7.2.1 for information on the exact calculations
used by PG7000 to obtain the differential pressures defined
in high line differential mode.
Differential Device Under Test (DUT
Reference PG7000 (PG7102, PG7202
or PG7302)
MPC1-D or other Hi Variable Volume (Hi VV)
MPC1-D or other VENT Valve
MPC1-D or other HI/LO BYPASS Valve
MPC1-D or other Lo Variable Volume (Lo VV)
MPC1-D or other INLET Valve
Figure 11. High Line Differential Mode Schematic
Note
Technical Note 0080TN03 p rov id es more detailed information on
high line differential mode principles using PG7102 and PG7601
piston gauges and includes a complete uncertainty analysis.
Consult Fluke Calibration or visit our website,
www.dhinstruments.com to obtain a copy.
OPERATION
Note
High line differential mode does not support operation with
AMH automated mass handler(s).
To select high line differential mode operation and access differential mode
functions press [MODE], <3HLdif> on the reference PG7000.
Note
When operating in high line differential pressure mode, the
reference PG7000 is “master” and the tare PG7000is “slave”.
All operator interaction is with the PG Terminal of the reference
PG7000. The tare’s display is for information only. The tare
PG7000 keypad is inactive.
•Entering differential mode (see Section 3.9.4.2,
Differential Mode
([MODE], <3HLdif>, <1run>)).
•Setting a line pressure (see Section 3.9.4.2,
([MODE], <3H Ldif>, <1run > , <2yes>)).
•Setting d if f er ential pr es sur e s at a h ig h line pressure (see Sec ti on 3.9 .4.2 ,
Setting Differential Pressures at a High Line Pressure
•Viewi ng current line pressure and na tural fall rates for the line pres sure
(see Section 3.9.4.2,
Temperatures and Natural Piston Fall Rates ([MODE], <3HLdif>,
<2view>).
Entering High Line Differential Mode ([MODE], <3HLdif>, <1run>)
To enter high line differential mode, press [MODE] ,<3HLdif>, <1run> on the
reference PG7000’s PG Terminal.
The reference PG7000 initializes high line differential mode. To successfully enter
high line differential m ode the reference PG7 000 must be able to communic ate
with the tare PG7000, the piston-cylinders in both PG7000s must be nom inally
identical and the piston-cylinder in the tare PG7000 must have a non-zero value
for k(P) in its piston-cylinder module file. W hen initializing high line diff erential
mode, the reference PG7000 tests for these conditions and provides error
messages as follows:
•<Cannot find PG7000 on COM2>: COM2 of the reference PG7000 must
be connected to COM1 of the tare PG7000 using a valid RS2 32 ca ble and
both interfaces must be properly set so that communications between
the two PG7000s can occur (see Sections 3.11.5.1, 3.11.5.4). If the
reference is unable to com municate with the tare, t his error message is
displayed. If this error m essage is obs erv ed, c orr ec t the communications
problem and retry (see Section 3.11.5.1).
•<T & R PCs not a pair, cannot run HLdif>: The piston-cylinder
modules in the tw o PG 700 0s must be nominally i den ti c al (ha ve th e s ame
nominal mass to pressure conversion coefficient). If the reference
PG7000 finds the two pis ton-cylinder modules to be different, this er ror
message is displayed. If this error message is observed, correct the
situation by changing piston-cylinder selection or adjusting one pistoncylinder’s definition (se e Sections 3.9.2, 3.11. 1.2).
•<No T PC in T PG7000, cannot run HLdif>: The piston-cylinder
module selected in th e tare PG 7000 m us t have a non-zero value for k (P)
in the active piston-cylinder module file. If the k(P) value is zero, this
error message is displayed. If this error message is observed, correct
the situation by providing a non-zero value f or k(P) in the piston-cylinder
module file of the tare piston-cylinder (see Section 3.11.1.2).
After high line differential mode is initialized, the displayof the tare PG7000 is:
1. Standard Ready/Not Ready indicators for the
tare PG7000.
2. Last line pressure set in the pressure unit of
measure that was active when the line
pressure was set. The unit of measure
remains the same, even if the reference
PG7000’s unit of measure is changed, until a
new line pressure is set.
3. Label indicating that the mass load includes whatever trim masses were loaded on the tare
piston in the crossfloating proc ess to reach equilibri um.
4. Tare mass load for the last line pressure set.
5. Current piston position of the tare PG7000.
While in high line differential pressure mode, the tare PG Terminal keypad
is inactive.
After high line differen tial mode is initializ ed, the displayof the re ference P G7000 is:
1. Current (last set) value of line pressure in
current pressure unit of measure.
Select <1no> to use the ex isting li ne press ure sett ings and return t o the pr evious
run screen in high line differential measurement mode (see Section 3.9.4.2,
Setting Differential Pressures at a High Line Pressure
).
Select <2yes> to set a new line pressure value or repeat the line pressure setting
procedure at the current lin e pressure va lue (see Sec tion 3.9.4.2,
Pressure ([MODE], <3HLdif>, <1run>, <2yes>)
).
Setting a Line
Note
The line pressure setting procedure must be executed each
time the line pressure is changed and should be executed at the
beginning of each differential pressure calibration sequence
even if the line pressure is the same as the line pressure
previously used. Select <2yes> to go to the line setting
procedure.
Setting a Line Pressure ([MODE], <3HLdif>, <1run>, <2yes>)
Caution
To meet the full performance potential of PG7000 high line
differential mode operation, external influences on the piston
gauges must be reduced or eliminated. Air currents and
vibrations are the most significant possible influences. Do not
operate near an active air conditioning or heating duct, avoid
opening and closing doors or any movement of personnel
around the system.
Leaks at any point in the test system are highly detrimental to
measurement results in high line differential mode. Thoroughly
leak check the system before operation and correct any leaks
detected.
PG7000™ OPERATION AND MAINTENANCE MANUAL
Target line P:
1000 psi
Load nominal mass on
T & R: 33.3 kg
Open bypass, float T
Close bypass,[ENTER]
Setting a line pressure has two steps:
Setting the line pressur e an d de ter mining the natural f all rates of the tare an d
reference pistons.
Performing a crossfloat of the pistons to reach equilibrium and minimize
differential pressure zero of f set .
To set a line pressure press [MODE], <3HLdif>, <1run>, <2yes> on the
reference PG7000’s PG Terminal. The display is:
1. Edit field for entry of desired line pressure
value in current pressure unit of measure.
Defaults to last value used.
Values of line pressure are always entered and displayed in
gauge mode (i.e., relative to atmospheric pressure). Line
pressure mass loading i nstructions are always giv en with 0.1 kg
resolution. The minimum line pressure is the pressure resulting
when the mass of the piston + mass bell is loaded.
Enter the desired line pressure value.
The next display indicates the mass to
load to set the line pressure.
Note
This display corresponds to the standard mass loading instruction display of
pressure to mass m ode (see Section 3.9.11.1). The mass loading instruction is
always given with 0.1 kg resolution regardless of the current mass loading
resolution setting (see Section 3.9.10). L oad the in dicated mass on both t he tare
and reference PG7000s following PG7000 mass loading protocol using main
and fractional mass es only (do not use trim masses) (see Section 3.6). Press
[ENTER] when ready.
Note
Line pressure mass instructions are always given with 0.1 kg
resolution. This is to preserve the smaller m asses for setting
the differential pressu re. The diff erential pressure m ass loading
instructions follow conventional PG7000 mass loading protocol
using the masses that remain after the line pressure has been
set. It may be necessary to select the line pressure value to
assure that adequate masses remain to set the desired
differential pressure. For example, if the line pres sure requires
loading 33.3 kg, a 0.2 and 0.1 kg mass w ill be used to set the
line pressure and will not be available if needed to set the
desired differential pressure. If this situation occurs, consider
changing the line pressure so that it is set using a whole
number of kilograms, in this example 33 or 34 kg.
rate difference” measurement. Select
difference” value and continue to the
Open bypass and
[ENTER] to xfloat
With the system’s B YPASS valve OPEN ( see Figure 10), use the INLET and/or
VENT valve and the h igh side v ariable volume to adjust the pressure to float the
piston of the tare PG7 000. This should cause the reference PG7000 piston to
be at its bottom stop. Once the tare piston is floating, close the system’s
BYPASS valve and press [ENTER]. The next display is:
1. Tare piston position.
2. Reference piston position.
Use the high and/or lo w s id e v aria ble v olume to set both p istons t o a pos i tio n j ust
above + 1.0 mm. Ensure that both pistons are rotating. As both pistons fall
through the + 1.0 position, press [ENTER]. PG7000 pauses for 5 seconds a nd
then measures the natural fall rates of both pistons for 30 seconds.
1. Indication of tare piston position.
2. Indication of reference piston position.
3. 30 second count down.
Wait for the 30 second t imer to count down while the measurements are m ade.
Be sure not to interfere with the free movement of the piston-cylinders during
the countdown. After the countdown completes, the results ar e disp layed:
1. Average fall rate of the taring piston.
2. Average fall rate of the reference piston.
3. Difference between the reference and taring
piston fall rates (R - T). This is the “natural fall
rate difference”.
Note
The “natural fall rate difference” is the difference between the
fall rates of the two pistons when they are floating naturally at
the line pressure. W hen performing the crossfloat between the
pistons in the next step of line pressure setting, the objective
will be to adjust the mass of the tare piston until the “crossfloat
fall rate difference” measured with th e system’s BYPASS valv e
open is equal to the “natural fall rate difference” measured with
the BYPASS closed.
Select <2no> to repeat the “natural fall
<2yes> to accept the “natural fall rate
second step of line press ure s etting whic h
is crossfloating the two piston-cylinders.
The display is:
Open the BYPASS val ve a nd pres s [ENTER] to pr oc e ed with t he c ros s f loat. T he
display is:
1. Tare piston position.
2. Reference piston position.
3. Difference between the “natural fall rate
difference” and the “crossfloat fall rate
difference” in mm/min. This is the “equilibrium
fall rate”.
The “equilibrium fall rate” is the difference between the “natural
fall rate difference” and the “crossfloat fall rate difference”:
Equilibrium fall rate = (R
When the “equilibrium fall rate” is zero, the two piston-cylinders
are at equilibrium at the line pressure.
Crossfloat the two pistons to find equi librium by mak ing mass adjustm ents on the
tare piston. Find ing equ ilibr ium is an iterat ive proces s wh ich is assis ted b y PG700 0.
Pressing [ENTER] starts an automated fall rate measurement sequence that
measures the fall rates m ore precisely than the rea l time indication. Follo wing a
30 second countdown or when on e of the pistons has moved more than 1 mm ,
PG7000 displays the measured “equilibrium fall rate” and suggests the mass
adjustment needed on th e tare piston to ac hi ev e a n “e qui li brium fall rate” of zer o.
This process can be repeated as many times as desired.
nat
Note
- T
nat
) - (R
xfloat
-T
xfloat
)
Note
If a standard piston-cylinder and mass set (rather than a tare
piston-cylinder and m ass set) is being used in the tare PG7000,
the tare piston may be the heavy one at the start of the
crossfloat (see Section 3.9.4.2 , PRINCIPLE). In t his case, rath er
than reduce mass on the tare PG7000, add a small amount of
mass to the referenc e PG7000 before starting th e crossfloat so
that the tare piston will be the lighter one (generally < 1 g is
needed). This will simplify the crossfloating procedure and
subsequent differential mass loading. Be sure not to confuse
the additional tare mass on the reference piston with the
subsequent differential mass loads.
Using trim masses ( 50 g an d be lo w) fr om the trim mass set, adj us t the mass load
on the tare PG7000 to establish equilibrium (“equilibrium fall rate” near zero).
When the mass adjustments are complete, close the BYPASS valve and put both
pistons at a position jus t above 0.0 mm. Ensure that both pisto ns are rotating.
As the pistons fall through the 0.0 mm point, OPEN the BYPASS valve, then
press [ENTER]. After a 5 second d elay for stabi lization, PG7000 begins the fall
rate measurement. The display is:
1. Indication of tare piston position.
2. Indic ation of ref erenc e piston position.
3. 30 second count down.
4. Average “equilibri um fall rat e” since start of
the countdown.
PG7000 is measuring the drop rate of the two pistons and calculating the
“equilibrium fall rate”. Be sure to not interfere with the free movement of the
piston-cylinders or to introduce any outside interference while the countdown
occurs. The fall rate measurement c onti nues unt il either piston has moved 1 mm
or 30 seconds have elapsed. The measurement can also be concluded by
pressing [ENTER]. When the fall rate m easurement has concluded, the resu lts
are displayed:
1. Average “equilibrium fall rate” measured i n
last crossfloat.
2. Suggested mass change (in grams) on the
tare (T) piston to achieve equilibrium based on
equilibrium fall rate just measured.
Evaluate the qu alit y of the equili brium. Check t he va lue of the “e quil ibrium fal l rate” .
Consider t hat the cl oser the ra te is to zer o, the b etter the eq uilibrium ; consider that
the suggested mass change on the tare piston times the mass to pressure
conversion c oeffic ient of th e pisto n approx imates t he press ure val ue of the d efect in
the equilibr ium. Check the difference between the DUT output with the BYPASS
open and BYPASS clo sed t o ev al uat e the z ero e r ror due to t he equ ili b riu m.
Select <1yes> to return to the crossfloat <B/P open, [ENTER]> screen.
If you are satisfied with the quality of the equilibrium , select <2no> to go to the
run screen in high l ine dif f er entia l pres s ure mode and set differ entia l pres s ures a t
this line pressure (see Sec tion 3.9.4. 2,
Line Pressure
).
Setting Differential Pressures at a High
Note
•T he value of the av erage “equi librium fal l rate” measu red by
the crossfloat should typically be inside of
± 5 mm/min
and/or the suggested mass adjustment should be inside of
50 mg.
•The device under test (DUT) can often be used to evaluate
the quality of the equilibrium between the two pistons and
to correct for the zero error caused by the defect in
equilibrium. Observe the change in the DUT output when
the system BYPASS valve is opened and closed with the
pistons floating. With the system BYPASS valve open, by
definition, the differential pressure applied to the DUT is
zero. The change in the DUT output observed when the
BYPASS valve is closed with the pistons floating is the zero
error due to the defect in the crossfloat equilibrium.
Consider correcting all of the DUT readings by the value of
this offset to correct for the crossfloat zero error. When
using the offset, keep in mind that, due to differential
evolution of the two PG7000 piston-cylinder temperatures,
the differential pressure at zero differen tial mass load may
not be zero.
Setting Differential Pressures at a High Line Pressure
Caution
•To meet the full performance potential of PG7000 high line
differential mode operation, external influences on the piston
gauges must be reduced or eliminated. Air currents and
vibrations are the most significant possible influences. Do
not operate near an active air conditioning or heating duct,
avoid opening and closing doors or any movement of
personnel around the system.
•It is not possible to operate in differential mode with leaks at
any point in the test system. Thoroughly leak check the
system before operation and correct any leaks detected.
•High line differential mode is often used to calibrate DUTs
whose differential pressure range is very low relative to the
line pressure. Th ese DUT s can easily be ov er pressured by
inadvertently applying excessive pressure to one port.
When operating in high line differential pressure mode,
OPEN the system BYPASS valve to zero the differential
across the DUT when there is any risk of overpressure.
Before operating in diff erential mode the line pressure must be s et (see Section
3.9.4.2,
Setting a Line Pressure ([MODE], <3HLdif>, <1run>, <2yes>)).
Once the line pressur e is set, from a practical s tandpoint, operating in high line
differential mode is very similar to operating in gauge mode (see Section 3.9.11).
All operat or interaction is with the PG Ter minal of the reference PG7000. The PG
Terminal of the tare PG7000 is for display only. The tare PG7000 PG Terminal
keypad is loc ked o ut du ring high line diffe ren tia l mode ope rati on.
Operation is in either pressure to mass or mass to pressure mode
(see Section 3.9.12).
To operate in differential m ode the system BYPASS valve must be closed (see
Figure 10).
All mass loading and unloading to define differential pressures is made on the
reference PG70 00. Mas s load ing and unloa ding i nstruc tions fol low co nvent ional
PG7000 mass loading pr ot oc ol (s ee S ectio n 3.6) b ut t he dif f er enti al mass loading
instructions are “on top of” the line pressure mass load. The mass load
corresponding to the line pressure on the tare and reference PG7000s must
remain exactly the same throughout the differential measurements.
While in high line differential mode, the piston position and piston rotation
Ready/Not Ready ind icators on the reference PG7000 ref lect the status of bot h
the tare a nd the reference PG 7000s. For the indicator to indicate Ready, bot h
the tare a nd the reference PG7000s m ust meet their current Ready/Not Ready
criteria (see Sectio n 3.4). A <T> in a Ready/Not Ready status indicator indicates
that the tare PG7000 is Not Ready.
•When operating in high line differential mode, the
measurement mode indicating character directly to the right
of the pressure unit of measure in the main run screen is
<d> (see Section 37).
•Zero differential pressure (BYPASS valve closed) can be set
by the system. In some cases, the dif ferential mass to lo ad
to set zero is not zero. This is normal and can occur due to
differential changes in the temperature of the tare and
reference piston-cylinders after the original line pressure
crossfloat. A differential piston-cylinder module
temperature change while running in high line differential
mode affects the line pressure causing an offset to the
differential pressure. The offset to the differential pressure
causes a non-zero differential mass load to be needed to set
zero differential pressure (see Section 7.2.1). Conversely,
and for the same reason, a differential mass load of zero
may not result in a differential pr essure of zero.
used in calculating corrections to the differential pressure to take into account differential
Typical Sequence to Set Differential Pressures at a High Line Pressure
Set the desired line pressure (see Section 3.9.4.2, Setting a Line
Pressure ([MODE], <3HLdif>, <1run>, <2yes>)).
Select pressure to mass or mass to pressure mode (see Section 3.9.12).
If desired, read DUT output at zero differentia l pressure with BYPA SS valve
OPEN. Then, put the BYPASS valve in CLOSED position. Float both
PG7000 pistons before oper ating the BYPASS valve. Read DUT output at
zero as defined by the PG7000s if desired (this value can be used as an
offset correction on subsequent DUT differential pressure outputs).
Press [ENTER] and enter a pressure or mass value. All mass loading and
unloading to define differential pressures is performed on the
reference PG7000. Follow normal PG7000 mass loading protocol (see
Section 3.6). Note, however, that the mass loading instructions are “on top
of” the mass already loaded to define the line pressure.
The mass corresponding t o the line pressure must remain ex actly the same
on both the tare and reference PG7000s throughout the differential
pressure measurements.
Float the tare and reference pistons. For best results put both pistons
slightly above the 0.0 point, for example +0.3 mm.
When the reference PG7000 indicates Ready on both Ready/Not Ready
indicators (see Section 3.4 ), take the DUT reading at the differential
pressure indicated on the to p line of the reference dis pla y. For best
results log DUT readings and average them over time while the PG7000
pistons fall through th e 0. 0 m m point. Averag ing f or 10 to 30 sec onds a llo ws
the random pressure noise from piston rotation to be integrated and
eliminated.
Repeat Steps through for each desired differential pressure point.
If the last differential press ure is zero, consider reading the DU T output with
the BYPASS valve CLOS ED. Then OPEN the BYPASS valve to set “true”
zero differential pressure.
Viewing Line Pressure, Starting Piston-Cylinder Temperatu res and Natural
Piston Fall Rates ([MODE], <3HLdif>, <2view>)
To view the last line pressure setting, the starting piston-cylinder module
temperatures, the natural piston fall rates and the “natural fall rate difference”
recorded at that line pressure, press [MODE] and select <3HLdif>, <2view>.
The display is:
1. Last line pressure value set (line pressure is
always in gauge mode).
2. Temperature of the reference piston-cylinder
logged when the line pressure crossfloat was
completed.
3. Temperature of the tare piston-cylinder
module logged when the line pressure
crossfloat was completed. The starting
temperatures are the temperatures logged at
the time the line pressure setting crossfloat
was completed. These temperatures are
temperature changes in the piston after completion of the crossfloat (see Section 7.2.1 for
differential pressure calculation information).
Press [ENTER] to continue to the next view screen or [ESCAPE] to go to the
previous screen. The next view screen is:
1. Average fall rate of the tare piston recorded at the
last line pressure.
2. Average fall rate of the reference piston recorded at
the last line pressure.
3. Difference between the reference and tare piston
average fall rates (R - T). This is the “natural fall rate
difference”.
The “natural fall rate difference” is the difference between the
fall rates of the two pistons when they are floating naturally at
the line pressure. The “natural fall rate” can change slightly as
operating conditions change so the “natural fall rate”
determination and piston crossfloat procedure should be
performed at the beginning of each sequence of differential
pressure definitions.
To return to the <HLDif mode:> menu, press [ESCAPE].
To return to the previous run screen, press [ENTER].
Note
3.9.5 [SYSTEM]
PURPOSE
To access the two SYSTEM run screens which display current values of PG7000 piston
behavior, piston-cylinder temperature and vacuum reference (PG7601 only) measurements.
OPERATION
To access the SYSTEM ru n sc reens, pr ess [SYSTEM] from any other run scr een. T here are
two SYSTEM run screens. Pressing [SYSTEM] or [
between the first and second SYSTEM screens.
See Sections 3.9.5.1 and 3.9.5.2 for detailed information on the contents of the first and
second SYSTEM run screens.
Note
The SYSTEM screens are run s creens. This means that other functions
can be accessed from the SYSTEM screens and the active SYSTEM screen
will be returned to when leaving functions. The MAIN screen and
AMBIENT screen are also run screens (see Sec ti on 3. 8 ).
3.9.5.1 First System Run Screen
OPERATION
The first SYSTEM r un screen provides real tim e display of piston rotation r ate,
decay in piston rotation rate, piston position and piston fall rate.
To access the first SYSTEM run screen, press [SYSTEM] from any other
run screen.
changed. Flashes when Not Ready and
piston is floating (see Section 3.4.2).
when information is
nnn rpm ±nn/min
±n.nn mm ±n.nn/min
when the piston is at the low stop
The first SYSTEM run screen displays:
1. <nnn rpm>:
piston rate of rotation. The unit of measure is
rotations per minute [rpm] and cannot be
Indicates < ---- >
unavailable or out of range.
2. <±nn/min>: Numerical value of current decay
in piston rotation rate (deceleration). The unit
of measure is rotations per minute [rpm/min]
and cannot be changed. < ---- > when
information is unavailable or out of range.
3. <±n.nn/min>: Sign and numerical value of current piston vertical rate of displacement. A
negative value i ndi cates piston f alling. A positive value indicates pis t on ri sing. The unit of measure
is millimeters per minute [mm/min] and cannot be changed. Indicates < ---- > when
information is unavailable or out of range.
4. <±n.nn mm>: Sign and numerical value of current position of the piston within the piston
stroke (see Section 3.5). The unit of measure is millimeters [mm] away from mid-stroke
position and cannot be changed. I ndicates <HSTOP> when the piston is at the high stop (all
the way up) and <LSTOP>
(all the way down). F lashes when Not Read y and pist on is floati ng (se e Section 3. 4.1). Indicates
< ---- > when information is unavailable or out of range.
Pressing [ESCAPE] in the f irst SYST EM r un sc re en re tur ns operat ion to th e M AIN
run screen. Press ing [SYSTEM] or [
±] toggles bet ween the first and secon d run
screen. All function keys are active from the first SYSTEM run screen and
operation returns to that screen when leaving functions the were accessed from it.
Note
•The measurement systems for piston behavior indications
rely on movement of the mass loading bell. Piston behavior
indications (piston position, piston rotation rate) are not
valid when the mass bell is not loaded on the piston.
•See Section 3.5 for information on the piston stroke and
measurement zone.
3.9.5.2 Second System Run Screen
OPERATION
The second SYSTEM run screen displays the values of piston-cylinder
temperature and temperature rate of change measured by PG7000.
PG7601 also displa ys reference vacuum and vacuum rate of change . If internal
or external measurement is specified for the vacuum values in SETUP (see
Section 3.10), the second S YST EM r un s cr een pro vi d es a real time display of the
values measured.
To access the second SYS T EM run scr een, pres s [SYSTEM] or [±] from the first
SYSTEM run screen. To acces s the first SYSTEM run sc reen, press [SYSTEM]
from any run screen.
negative value indicates temperature
temperat ure inc re asing. The unit of
n.nn°C ±n.nn/min
nnn.n Pa ±nnn.n/min
measure is degrees Centi grade per minut e [°C/min] and cannot b e changed. Indic ates < ---- >
when information is unavailable or out of range.
is out of range or greater than 20 Pascal.
The second SYSTEM run screen displays:
1. <n.nn°C>: Current piston-cylinder
temperature. Source of value can be internal
measurement, default or user depending on
current SETUP selection (see Section 3.10
The unit of measure is degrees Centigrade
[
°C] and c annot be changed. Indicates < ----
> when information is unavailable or out of
range.
2. <+n.nn/min>: Sign and numerical value of
current rate of change of temperature. A
decreasing. A positive value indicates
when information is unavailable or out of range. Indicates < ---- > when “user” or “default” is
the current SETUP selection for piston-cylinder temperature source (see Section 3.10).
3. <+nnn.n/min>: (Used for PG7601 only.) Sign and numerical value of current rate of change
of the vacuum reference pressure. A negat i ve value i ndi cates pressure decreasing. A positive
value indicates pressure increasing. The unit of measure is Pascal per minute [Pa/min] and
cannot be changed. Indicates < ---- > <Blank> if the PG7000 model is not PG7601 or if user or default is the current SETUP
selection for reference vacuum (see Section 3.10).
4. <nnn.n Pa>: (Used for PG7601 only.) Current vacuum reference value. Can be internal
measurement, external measurement, default or user depending on c urrent SETUP selection.
Flashes when Not Ready and piston is f loating (see Section 3.4.3). The unit of measure is
Pascal [Pa] and cannot be changed. Indicates < >20 Pa > if current SETUP selection is
internal or external and current measurement
<Blank> if PG7000 model is not PG7601.
).
Pressing [ESCAPE] in the s econd SYSTEM run sc reen returns operat ion to the
MAIN run screen. Pressing [SYSTEM] or [
second SYSTEM run screen. All function keys are active from the second
SYSTEM run screen and o peration ret urns to that screen w hen leaving f unctions
that were accessed from it.
Note
The current selection in SETUP determines the source of the
values used by PG7000 for piston-cylinder temperature and
vacuum reference values. If the SETUP setting is user or
default, the SYSTEM screen displays the user or default value,
not PG7000’s on-board measurement(s).
3.9.6 [AMBIENT]
PURPOSE
Access the AMBIENT r un screen which displa ys the current ambient conditio n values being
used by PG7000 for calculations of reference pressures.
PRINCIPLE
PG7000 uses ambient condition values to calculate the reference pressures that it defines
(see Section 7.2). The source of the ambient condition values is specified in the current
SETUP file (see Section 3.10). The AMBIENT run screen displays the current ambient
condition values. If the SE TUP selection for the am bient condition is internal m easurement,
then the AMBIENT r un screen pr ovides a rea l time dis play of the m easurem ent of PG7000’s
on-board sensor for that variable.
±] toggles between the first and
OPERATION
To access the AMBIENT run screen, press [AMBIENT] from any other run screen
(MAIN or SYSTEM).
1. <nnnnnnn uuuu>: Current numerical value and pressure
when information is unavailable or out of range.
nnnnnnn uuuu nn.n°C
nn%RH n.nnnnnn/s2
2. <n.nn°C>: Current ambient temperature. Source of value can be internal measurement, default or user
Note
The AMBIENT screen is a run screen. This means that other functions can
be accessed from the AMBIENT screen and the active AMBIENT screen
will be returned to when leaving functions. The MAIN screen and SYSTEM
screens are also run screens (see Sections 3.7 and 3.9.5).
The AMBIENT run screen displays:
unit of atmospheric pressure. Source of val ue can be internal
measurement, remote barometer, default or user depending
on current SETUP selection (see Section 3.10). The unit of
measure is determined by the UNIT function setting (see
Section 3.9.3). ATM head is applied to internal or external
barometer reading to correct atmospheric pressure to the
PG7000 reference level (see S ection 3. 11. 3.3). Indicates < --
-- >
Indicates <TIMEOUT> if the current SETUP selection is a
remote barometer and PG7000 communication with the
barometer times out.
depending on current SETUP selection (see Secti on 3.10). The unit of measure is degrees Centigrade [ºC]
and cannot be changed. Indicates < ---- > when information is unavailable or out of range.
3. <n.nnnnnnm/s2>: Value of local acceleration due to gravity. Can b e default or user depending on current SETUP
selection (see Section 3.10). The unit of measure is meters per second squared (m/s2) and cannot be changed.
4. <nn%RH>: Current ambient relative humidity. Can be internal measurement, default or user depending on
current SETUP selection (see Section 3.10). The unit of measure is percent relative humidity (%RH) and
cannot be changed. Indicates < -- > if current SETUP selection is internal and current measurement is
unavailable or out of range.
Pressing [ESCAPE] in the AMBIENT run screen retur ns operation to the MAIN run screen.
All function keys are active from the AMBIENT run screen and operation returns to that
screen when leaving functions that were accessed from it.
Note
•The curr ent selection in SETU P determines the source of the values used
by PG7000 for atmospheric pressure, ambient temperature and relative
humidity. If the SETUP setting for these values is user or default, the
AMBIENT screen displays the user or default value, not PG7000’s onboard measurement(s).
•To change the ambient pressure units of measure, see Section 3.9.3.
When the current pressure unit of measure is an altitude unit, atmospheric
pressure in the AMBIENT run screen is expressed in kPa if the altitude unit
is meters (m) or psi if the altitude unit is feet (ft).
3.9.7 [HEAD]
PURPOSE
To cause a pressur e value, representin g the fluid head res ulting from a dif ference in height,
to be added to the pressure defined by PG7000 at its reference level. To set the height of the
DUT head.
PRINCIPLE
The pressur e def ined b y the PG700 0’s flo atin g pist on is th e pres sure at the b ottom of the pist on.
This is referred to as the PG7000 r efer ence level. T he height of t he bottom of the pist on with
the piston in mid-stroke position is mark ed reference level on the PG7000 piston-cylinder
module mounting post. Frequently, when performing a calibration or test, the device or
system under test is at a different hei ght than t he PG7000 r eference level. T his differ ence in
height (referred to as DUT head) can cause a significant difference between the pressure
defined by the PG7000 at its reference level and t he pressure actuall y applied to the device
under test located at a different hei ght. In this case, it is useful to m ake a head correction to
the pressure defined b y the PG7000 at its reference level in order to ac curately predict the
pressure actually applied at a different height. The HEAD function allows this head correction
PG7000™ OPERATION AND MAINTENANCE MANUAL
Edit DUT head height
95 cm N2
REFERENCE
LEVEL
Reference
Line
(+)
(-)
to be applied a utom atical ly for a v ariet y of flui ds base d on oper ator e ntr y of the hei ght dif ferenc e.
The fluid used and the H EAD function units of m easure are set by pressin g [SPECIAL] and
selecting <3head> (see Section 3.11.3).
PG7000 can accurately determine head pressures for gases (nitrogen, helium and air) and liquids
(Di-2 ethylhexyl Sebacate oil, water and a user defined liquid) as the pressurized medium.
•When gas is the test fluid, use of the H EAD function is most important at
low absolute pressures. In this case, specifying the head height within
± 0.2 in. (5 mm) is adequate to ensure that, even in the worst case, the
uncertainty on the head correction will be insignificant relative to the
tolerance on the PG7000 measurement. Use of the HEAD function to
ensure tol eranc e mea sur ement s is p arti cula rly c ritic al wh en a liquid is th e
test fluid, due to the high density of liquids. To determine when and how
precisely a head correction for liquids must be made, 0.03 psi/inch (90
Pa/cm) may be used as an estimation of the liquid head value.
•Regardless of the head function’s setting, corrections are automatically
applied to the calculated reference pressure to compensate for the
deviation between the current piston position and the mid-stroke position
(see Section 3. 11 .3. 4 ).
•The pistons of certain gas operated piston-cylinder modules are hollow.
Due to their irregular shape, for these pistons, the natural reference level is
not at the bottom of the piston. So that, in pra ctice, the actual r eference
level is the s am e fo r all p ist on-cyl ind e r mod ules , a r efe ren ce lev el off set is
applied when a hollow piston is used. The reference level offset is
included in the piston-cylinde r file (see S ection 3.11.1 .1) and corr ects the
reference level back to the reference point marked on the mounting post.
OPERATION
To access the HEAD function, press [HEAD]. The display is:
Note
1. Test fluid currently specified for the head correction.
2. Entry field for head height (1 to 999 cm or in.).
Entering a value of zero turns the HEAD function OFF. Entering a value other than zero
turns the HEAD function ON using the height entered. Pressing [ESCAPE] returns to the
main run screen with NO change to the current head setting.
•The reference height of PG7000 pressure definition is the bottom of the
piston in m id-stroke float position. This position is mar ked on the piston cylinder module mounting post and on the optional AMH, mass automated
handler. The DUT head h eigh t sh ou ld b e ent e red a s a p osit iv e v alu e if t he
device or syst em und er te st i s h igh er t han th e PG 70 00 ref e renc e l ev el and
negative if it is lower.
•To change uni ts of DUT head height betw een inches and centimeters and
to change the test fluid, press [SPECIAL] and select <3Head> (see Se ction
3.11.3).
•When the HEAD function is ON (DUT head value different from 0), the
application of a head correction is indicated by <h> in the right side of the
top line of the MAIN run screen (see Section 3.7). When the HEAD function
is OFF, the <h> is NOT shown. PG7000’s also have a separate head
correction to compensate for the deviation between the current piston
position and mid-stroke (see Section 3.11.3.4). This P ISTON head can be
turned ON and OFF (see Section 3.11.3.4).
3.9.8 [ROTATE]
PURPOSE
To turn automatic control of motorized piston rotation (acceleration and brake) ON and OFF.
Note
See Section 3.9.13 for information on manual cont rol of motorized piston
rotation.
PRINCIPLE
The motorized piston rotat ion system is used to start or incr eas e pisto n r otat ion ra t e when t he
piston is floating. It is also used to stop pis ton rotation when necessar y, for example befor e
manipulating mass to set a new pressure. T he system operates b y engaging a m otor driven
belt around the bottom of the m ass loadin g bell to ac celerate or brak e the rotation r ate of the
mass bell and piston it is loaded on. The motorized rotation system can engage with the
piston at any position in its strok e and at any rotation speed with minimal impact on piston
position and the defined pressure. When the motorized rotation system disengages, the
piston is completely free.
With automatic motorized rotation ON, the motorized rotation system engages and
disengages automatically as needed when the piston is floating to maintain the piston rotation
rate above the minim um rate Ready limit (see Section 3.4.2). T he rotation rate i s measured
by PG7000 on board sens ors and the rot ation r ate lim it s are set in the f ile of the ac tive pisto n
cylinder (see Section 3.11.1.1). Whenever the piston is floating, the motorized rotation
system will attempt to m aintain the piston rotation rate within the Ready limits (except under
the cutoff mass load of 3 kg, at which the low limit is reduced to minim um to maximize free
rotation time. T he piston rotat ion Rea dy/Not Rea dy indicat ion charac ter indicat es N ot Ready
to alert the operator when the motorized rotat ion system is about to engage. The rotation
system will not engage when the current mass load is less than the mass of the piston +
mass loading bell.
The motorized rotatio n system is also used to brake and s top rotation when starting a ne w
pressure point. If the piston is f loating and rotating when [ENTER/SET P] is pressed, the
motorized rotation s ystem engages at a speed near t he rotation rate of the piston an d then
brakes it to a stop. Pist on rotation is stopped to a void loading and unloading m asses on the
rotating piston and to avoid stopping rotation by friction between the piston and the piston
end of stroke stops.
With automatic motorized rotation OFF, the motorized rotation s ystem engages only when
actuated by the operator . Pressing [
rotation (see Section 3.9.13).
] accelerates rotation or [] followed by [←] stops
Automatic motorized rotat ion is generally left ON for normal operation. It is t urned OFF in
situations where rotation system actuation independent of operator initiative is undesired
(e.g. when performing a crossfloat interc omparison with another pisto n gaug e).
OPERATION
[ROTATE]. The display is:
<AutoRotate ON> or <AutoRotate OFF> indicates the current state of AutoRotate. Use
<1on> or <1off> to change the state.
When automated rotati on is OFF, the PG7000 m otorized rotation s ystem will only engage if
the operator presses [
] or [] followed by [←] (see Section 3.9.13).
When automated rotation is ON, the automated rotation system engages automatically.
When the piston is floating, it enga ges as needed to maintain the piston rot ation rate above
the Ready limit (see Section 3.4.2). When [ENTER/SET P] is pressed, it engages to stop
rotation before masses are loaded and/or the pressure is adjusted (this function can be
turned ON and OFF, see Secti on 3.9.8.1). Manual control using [
←] is also still active.
[
] and [] followed by
Note
•Automatic motorized piston rotation is dependent on PG7000’s
measurements of piston position and rotation rate. These measurements
are only ava ilable when the mass lo ading bell is installed on the piston.
Automatic piston rotation is suspended when the current PG7000 mass
load does not include the mass loading bell.
•Proper operation of the rotation system is dependent on PG7000’s
measurements of piston position. When using AutoGen, be sure that the
piston position indication system is properly adjusted (see Section 5.2.2).
•When PG7 601 i s ope rating in abso lute b y v acuum m ode using th e intern al
vacuum sensor to measure reference vacuum, automatic motorized piston
rotation will not engage until the reference vacuum value is within the
Ready limit (see Section 3.4.3).
3.9.8.1 <2PRE-DECEL>
PURPOSE
To turn ON and OFF a function that causes the automated rotation system to
begin the piston rotation deceleration when [ENTER/SET P] is pressed rather
than at the time mass is to be loaded.
On a PG7000 Platform equipp ed with m otor ized ro tati on, when AutoR otate is on,
the piston rotation deceler ation function is used to stop rotation before mass es
are loaded or pressure is adjusted. As the decelerat ion function can take up to
one minute to execute, it can be initiated the moment that [ENTER/SET P] is
pressed to enter a new press ure or mass target. The decel eration function then
runs while the new target v alue is being en tered. Ho wever, when the nex t targ et
does not require chang ing main mass es, it may not be necessar y to stop piston
rotation. In these c ases, the running the decelera tion function is probably not
desired and it is not beneficial for piston deceleration to initiate it when
[ENTER/SET P] is pressed. For this reason, the function to start piston
deceleration when [ENTER/SET P] is pressed can be turned ON and OFF.
If pre-deceleration is ON and AutoRotate is ON, piston rotation deceleration
always initiates immediately when [ENTER/SET P] is pressed.
If pre-deceleration is OFF and AutoRotate is ON, piston rotation deceleration
occurs after entr y of the pres sure or m ass tar get value and on ly if the new target
requires changing m ain masses with a manual mass set or operation AMH if an
automated mass handling system is active.
To turn ON and OFF the f unction that causes stopp ing piston rotation to in itiate
when [ENTER /SET P] is p ressed, press [ROTATE], <2pre-decel>. The curs or
is on the choice corresponding to the current state. Select <2on> for piston
rotation deceleration to initiate when [ENT ER/SET P] is pres sed. Select <1off>
for deceleration to initiat e only after entry of a new pressure or m ass target and
only if main masses need to be m oved or AMH automated m ass handling needs
to be operated. The default is <2on>.
3.9.9 [GEN] (OPTIONAL)
PURPOSE
To turn ON and OFF automated pressure generation/control and view and edit autom ated
pressure control settings. Requires that an automated pressure generation/control
component be included in the PG7000 system and properly configured (see Section 2.4.9).
PRINCIPLE
PG7000s support automated pressure generation/control components. These components, when
properly configured and i nterfaced with the PG7000 platfor m, are controlled by the PG7000
platform to automatically set and adjust pressure to float the piston. Automated pressure
generation and control c omponents are interfaced vi a the PG7000 platform’s CO M3 RS232
port (see Section 3.11.5.1).
Once an automated pressure generation/control component has been properly configured
and interfaced with the PG7000 platform, the functions under [GEN] are used to turn
automated pressure generation/control ON and OFF and to set operating parameters
associated with automated pressure control.
With automated pressure generation/control ON, PG7000 uses the automated control
component to float and refloat its piston when a pressure or mass value is entered.
[ENTER/SET P] is pressed to initiate a new command. T he automated control function is
suspended when any funct ion key is pressed, when entering rem ote mode or if automated
pressure control is turned OFF.
After a target pressure or m ass has been enter ed (lo call y or rem otely) and t he r equired m ass
has been loaded, the GEN f unction controls the pressure control com ponent as needed to
float the PG7000 piston at the piston float target (see Section 3.9.9.1) and refloat it if
necessary. If the piston moves be yond the high or low piston posit ion Ready lim it the GEN
function refloats it to the piston float target. T he piston float targ et and piston po sition ready
limits are user adjustable (see Section 3.10 <6READY). See Section 3.5, Figure 8 for a
description piston position stroke zones and limits.
With automated pressure generation/control OFF, PG7000 leaves the automated
pressure generation/control component idle and does not attempt to use it.
If an automated pressur e control component is included in the PG7000 s ystem, automated
pressure generation/control is generally left ON for normal operation. It is turned OFF in
situations where pressure c ontrol independent of operator initiative is undesired (e.g. when
performing a crossfloat intercomparison with another piston gauge).
The [GEN] menu includes:
1. Turning the GEN function ON and OFF.
2. The adjustable piston f loat target that defines th e position to which the p iston is set when
floated (see Section 3.9.9.1).
3. A choice to hav e the pist on raised to the top of the strok e befor e manipulatin g m ass using
an AMH automated mass handler (see Section 3.9.9.2).
4. Viewing and setting t he UPP ER LI MI T of the automated pressure gener ati on component to
avoid accidental overpressure (see Section 3.9.9.2).
5. Viewing and setting the as sum ed press ure controll er tolera nce used to det er m ine press ure
setting limits when floating the PG7000 piston (see Section 3.9.9.4).
6. A c hoice to not readjust piston position if the pis ton is already floating within the Ready
position limit after a new target is executed (see Section 3.9.9.5).
7. Viewing and setting the volum e of the system to which the PG7000 system is c onnected
(PG7302 and PG7307 only) (see Section 3.9.9.6).
OPERATION
To access the GEN functions, press [GEN]. The
display is:
2target 3raise 4UL
<AutoGen ON> or <AutoGen OFF> indicates the c urrent state of AutoG en. Use <1on> or
<1off> to change the state.
If automated pressur e control is OFF, PG7000 attempts to turn automated pr essure control
ON when <1on> is pressed. To do so, communication must be established with a valid
automated pressure contr ol component o ver its COM3 RS232 por t (see Section 3.11.5 .1). If
PG7000 is not able to est ablish com munic ation with a valid press ure contr ol com ponent, <P control timeout, autogen off> is displayed momentarily. Correct th e communications error
with the pressure control component and try again. If PG7000 is able to establish
communications with a valid automated pressure control component, <Turning ON automated generation> is displayed momentarily and automated control is turned ON.
The PG7000 automated control function will attempt to float the piston within the piston
position ready limits (see Section 3.4.1) after any mass to pressure or pressure to mass
command once mass loading has been confirmed. It will continue to attempt to float its
piston until [ENTER] is pressed to initiate a new command, a function key is pushed to
interrupt AugoGen, or automated pressure control is turned OFF using [GEN].
If automated pressure control is ON and <1off> is pressed, <Turning OFF automated generation> is displayed m omentarily and automated control is turned OFF. The PG7000
automated control function is inactive.
Note
•Automated pressure generation/control ON is indicated by a <G> in the
right of the top line of the PG Terminal ma in run sc reen. The <G> flashes
when automated generation is active (see Section 3.7).
•Proper operation of the automated pressure generation/control function is
dependent on PG7000’s measurements of piston position. When using
AutoGen, be sure that the piston position indication system is properly
adjusted (see Section 5.2.2).
•Proper operation of the automated pressure generation/control function is
dependent on PG7000’s measurements of piston position and rotation
rate. These m easurem ents are onl y availa ble when the mass loading bell
is installed on the piston. Automated pressure generation/control is
automatica lly turned of f when the PG 7000 entere d pressure or m ass load
does not include the mass loading bell.
•Automated pressure generation/control is not available when operating in
PG7000 high line dif ferentia l press ure mode (s ee Sectio n 3.9.4.2 ).
To help protect against accidental overpressure, when using automated
pressure generation/control, set the upper limit (UL) of the pressure
control system using [GEN], <4UL> (see Section 3.9.9.3).
3.9.9.1 <2target>
PURPOSE
To adjust the distanc e from midstroke to which the pis ton position must be set
before the automated generation system considers the piston floating process
complete (see Sections 3.9.9, 3.5).
3. GENERAL OPERATION
Note
With PG7202, AutoGen pressure control using the PPCH-G
pressure control does not stop when the target piston position
is reached. The PPCH-G thermal pressure control unit (TPCU)
is used continuously to attempt to maintain the piston at the
target position. The default target piston position for a PG7202
is mid-float position (0.0 mm).
OPERATION
To adjust the AutoG en pist on float t arget pr ess [GEN], <2target>. Edit the val ue of
the target as desired. T he default is + 1 m m (0 mm for PG7202) . The entr y should
not exceed + 2.5 mm.
3.9.9.2 <3raise>
PURPOSE
To turn ON and OFF a function tha t causes the au to mated gene ra tion component (see
Section 3.9.9.) to raise the PG7000 piston to the top of its stroke before automated
mass manipulation by an AMH automated mass handler. This can avoid a large
pressure c han ge whi ch m ay occ ur in ce rta in circ um stanc es w hen AMH l ifts t he mas s
load off of the piston.
OPERATION
To turn ON and OFF the raise piston before mass loading fuction, press [GEN],
<3raise>. The cursor is on the choice corresponding to the current state. Select <1no>
for the piston not to be raised o r <2yes> to raise the piston. The defaul t is <1no>.
3.9.9.3 <4UL>
PURPOSE
To read and/or set the UPPER LIMIT (UL) of the automated pressure generation
component used by AutoGen. This function is used to protect against accidential
overpressure when using the PG7000 automated pressure generation function (see
Section 3.9.9 and the generation component’s Operation and Maintenan ce Manual, UL
section)
OPERATION
To view or set the UPPER LIMIT of the automated pressure control com ponent
used by AutoGen, press [GEN], <4UL>.
If AutoGen is not current ly ON, the control component UPPER LIMIT cannot be
accessed and an error message is displayed.
If AutoGen is ON, the curr ent UP PER LIM IT of the aut om ated control c om ponent
is displayed and can be edited.
The automated pressure control component will abort pressure generation and
beep repeatedly if its UPPER LIMIT is exceeded.
3.9.9.4 <5tol>
PURPOSE
To read and/or set the pressure measuring tolerance of the pressure controller used by
AutoGen to automate pressure control.
Note
This function is used only when the pressure controller is a
PPC3.
The control toler a nce funct i o n can be us e d t o reduce t h e t ime requir ed to set pr essure
and float the PG7000 piston when the pressure controller used is significantly more
accurate than the default tolerance value. The tolerance can also be described as the
degree of agreement between the pressure control’s pressure measurement and the
pressure value set by the PG7000 when its piston is floating. In general, making the
tolerance smaller reduces the t ime requi red to float the pi ston and making the tolerance
wider increases the time. However, if the tolerance is set too small so that it does not
correctly reflect the degree of agreement between the pressure controller pressure
measurement and the PG7000’s floating piston, overshoot of the piston float point or
inability to float the pi ston wi ll result .
The pressure controller tol erance dete rmines, when appl icable:
a) the distan ce from the actua l pressure targe t that pressure is s et before the
AutoGen function begins to seek move ment of the piston.
b) the amount of pressure overshoot that must occur before “forced rotation” is
used to overcome possible friction between the piston and the cy linder.
This function is used only when the pressure controller is a
PPC3.
Note
OPERATION
To view or set the pressure control tolerance, press [GEN], <5tol>.
The current value of the pr essure control t olerance is displa yed. The value is in
% of full scale of the pressure controller. If the pressure controller is
AutoRanged, the AutoRange maximum pressure is used as full scale (see the
pressure controller’s Operation and Maintenance Manual).
The default tolerance value is 0.05% of full scale.
Edit the toleranc e value if desired. Do no t make the pressure toler ance less than
the worse case agr eement between t he pressure controller pressure measurem ent
and the pressure indicated by the PG7000 when its piston is floating.
3.9.9.5 <6refloat>
PURPOSE
To turn ON and OFF a function tha t causes the au to mated gene ra tion component (see
Section 3.9.9.) to refloat the piston to the target piston position after a new pressure or
mass target is entered, even if the piston is already floating within the piston position
Ready limits. When Refloat is ON, the piston is always refloated to the target piston
position after a new pressure or mass target command. This gives the full stroke of the
piston to drop befor e refl oat is necess ar y. W hen Refl oat is O FF, t he time requ ired t o
activate t he pressure c ontrol com ponent an d refloat the p iston is elim inated if it is n ot
needed. This can result in very rapid pressure setting when the piston is still floating
after a mass load change.
OPERATION
To turn ON and OFF the Refloat function, press [GEN], <6refloat>. The cursor is on
the choic e corre spondi ng to the c urre nt state. Select <1no> for the piston NOT to be
controlled to the target position after a pressure or mass command if the piston is
already in the Ready piston position limits. Select <2yes> for the piston to alway s to be
controlled to the rai se the piston. The de fault is <2yes>.
3.9.9.6 <7Vol>
PURPOSE
To read and/or set the volume of the system to which the PG7000 system is
connected.
This function is used available only with PG7302 and PG7307.
PRINCIPLE
The piston floating routines of PG7302 or PG7307 oil operated piston gauge us ing a
PPCH automated pressure controll er are highly dependent upon the volume of the te st
system into which pressure is being controlled. The VOL function is used to specify the
test volume so that the PG7302 or PG7307 may scale the PPCH rates properly for the
volume.
Operating with an improperly specified volume will cause the piston floating
routine to be either very slow (specified volume to small) or to overshoot
(specified volume too large).
Note
•If the test system volume is not known, the PPCH volume
determination function may be used to measure it. This
function is run directly from the PPCH front panel (see the
PPCH Operation and Maintenance Manual, {INTERNAL],
<1CONFIG>).,
•The test volume can be read or set remotely (see Section
4.3.4, ).
OPERATION
To view or set the test volu me, press [GEN], <7vol>.
The current volume value is displayed. The value is in cubic centimeres (cc).
The default volume is 30 cc. The maximum volume is 300 cc.
Edit the volume value if desired.
3.9.10 [RES]
PURPOSE
To set the resolution with which PG7000 loads mass in response to pressure or mass
commands (see Section 3.9.12).
PRINCIPLE
PG7000 piston-cylinder s are sized such that there is a who le number, nominal relationship
between mass loaded on the piston in k ilograms [kg] and the press ure at w hich t h e pis to n will
float in kilo Pasc al [kPa] or Mega Pascal [MPa] . This relationship is cal led the pressureto
mass conversion coeffic ient and is expressed as k Pa/kg or MPa/kg. The pressure to mass
conversion coefficient is marked on the cap of each piston.
PG7000 mass sets are made up of masses in multiples and submultiples of the kilogram
making it simple to load mass values rounded to 0.01 g, 1 g, 10 g or 0.1 kg.
When using PG7000 to def ine pressure, the desired press ure value is entered (see Sectio n
3.9.11) and PG7000 prom pts the user with the mass value t o be loaded. Due to the many
variables that influence the exact pressure to mass relationship for a piston-cylinder
(even though there is nominally a whole number mass to pressure relationship) the mass
value to load to reach exactly the pressure requested is always an odd value. Therefore,
defining the exact pressure value requested always requires loading mass with 0.01 g
resolution.
When it is acceptable for the pressure values defined to not be exactly the nominal pressure value
of the point, operation can be simplified and mass loading errors can be reduced by loading mass
with a lower level of resolution and us ing the pressur e that the lower le vel resolution m ass load
generates. F or exam ple , on a pis ton-c ylinder with a nom inal pr essur e to m ass rel ati onship of 10
kPa/kg, defining a pressure of exactly 100 kPa, nominally requires loading 10 kg of mass.
However, o nce all the influences on the measurement are taken into cons ideratio n, the actual
mass to load to def i ne ex act ly 100 kPa wil l not be 10 . 000 00 kg, it wil l be a v alu e nea r 10 k g such
as 9.99731 kg. This value is diffic ult to load, as it r equires re latively com plex mass accounting
and the mani pulation of very sm all sub-gram masses. To avo id handling a diffic ult, odd mass
value, one might instead decide to load 10 kg and use whatever pressure results as the
reference pressure. In this example, loading 10 kg rather than 9.99731 kg would result in defining
100.0269 k Pa rat her tha n 100. 000 0 kP a. The pr es sur e d efi ned is onl y ver y s light l y dif fere nt f rom
the nominal value and there is no additional uncertainty if that value is us ed. The savi ngs in t ime
and reduction of possible mass loading errors are significant.
The PG70 00 RES function i s used to cause PG 7000 to automatic ally calculat e mass loads to a
whole number value starting at 0.01 g and increasing in powers of 10 to 0.1 kg. This function is very
useful to make operation more convenient and less error prone when it is not imperative that the
pressure defined by PG7000 be exactly the nominal p ressure of the te st or cali bration sequ ence.
The mass loading resolution of AMH automated mass handling systems is
0.1 kg. The default mass loading resolution when AMH is initialized is 0.1
kg. If resolution finer than 0.1 kg is set when AMH is active, the AMH loads
the required mass value wit 0.1 kg resolution and the operator is prompted
with an instruction to load the trim mass under 0.1 kg. For fully automated
operation, the mass loading resolution should always be set to 0.1 kg.
OPERATION
To access the resolutio n function, press [RES]. The
display is:
Press the [←] and [→] keys to select the desired level of resolution. [←] decreases
resolution and [
→] incr eases resolution. Press [ENTER] to set the selected resolution and
return to the main run screen. The resolution range is from 0.01 g to 0.1 kg in powers of 10.
Note
Note
•The RES setting has no aff ect in mass to pressure m ode. T he RES setting
only affects the resolution of the mass commands that result from
pressure entries in pressure to mass mode (see Section 3. 9.12 ).
•In PG700 0 high line differential pre ssure mode (see Sect ion 3.9.4.2), line
pressures setting is not affected by the RES setting; line pressures are
always set with 0.1 kg resolution. Differential pressure mass loading
resolu ti on is de ter mi ned by t he RES setting.
3.9.11 [ENTER/SET P] from Run Screen
PURPOSE
To enter and execute pressure to mass or mass to pressure commands (see Section 3.9.12).
PRINCIPLE
PG7000 can calculate and display the mass to be loaded to achieve an entered pressure
value (pressure to mass m ode), or the pressure resulting f rom an entered mass load (mass to pressure mode) . The P
mode (see Section 3.9.12).
Pressing [ENTER/S ET P] from any run screen (MAIN, SYSTEM or AMBIENT ) accesses the
pressure or mass entry screen w hich allo ws the comm and value to be entere d and proc eeds
through the sequence to set or read the pressure defined by PG7000.
OPERATION
To access th e press ur e or m ass entry screen, press [ENTER/SET P] from any run scree n. T he
sequence after [ENTER/SET P] has been pressed varies between mass to pressure and
pressure to mass mode. The mode is selected by pressing [P
See Section 3.9.11.1 for details on [ENTER/SET P] in pressure to mass mode and Section
3.9.11.2 for mass to pressure mode. See immediately below for typical operational seque nc es in
gauge and absolute m odes. See Section 3.9.4. 1,
differential mode operational sequence.
OR M function is used to s e t pr es sur e to m as s or mass to pressur e
Typical Gauge and Absolute by ATM Mode Operational Sequence
Press [MODE] and select gauge or absolute by ATM mode as desired (see Section 3.9.4).
Press [P OR M] and sele ct pressure to mass or mass to pressure mode (see Section 3.9.12 ).
Press [ENTER/SET P] and enter a pressure or mass value. If the piston is floating and
AutoRoate is ON, the braking function engages to stop piston rotation (see Section 3.9.8).
Load mass as instructed ( see Sectio n 3.6). If an AMH automated m ass handling system
is active, the mass is loaded automatically with resolution of 0.1 kg.
Use the system pressure control com ponent to float the PG7000 pis ton. If the AutoGen
function is ON, the autom ated pr es sure cont rol c om ponent floats the piston aut o m atically
(see Section 3.9.9).
When PG7000 indicates Ready on a ll Ready/Not Ready indicators (see Section 3.4), t ake a
DUT reading at the pressure indicated on the top line of PG7000 display.
Repeat Steps through for each desired pressure value.
Typical Absolute by Vacuum Mode Operational Sequence (PG7601 Only)
Press [MODE] and absolute by vacumm (avac) mode (see Section 3.9.4).
Press [P OR M] and sele ct pressure to mass or mass to pressure mode (see Section 3.9.12).
Press [ENTER/SET P] and enter a pressure or m ass value. If AutoRoate is ON, the
braking function engages to stop piston rotation (see Section 3.9.8).
Load mass as instructed (see Section 3.6). If an AMH automated mass handling
system is active, the mass is loaded automatically to resolution of 0.1 kg.
Install bell jar on PG7000, shut PG7000 vacuum vent valve, open vacuum reference
shutoff valve. Wait for vacuum under bell jar to reach Ready condition (see Section 3.4.3).
Use system pressure control component to float the PG7000 piston. If the AutoGen
function is ON, the autom ated pr es sure cont rol c om ponent floats t he piston auto m aticall y
(see Section 3.9.9).
When PG7000 indicates Ready on a ll Ready/Not Ready indicators (see Section 3.4), t ake a
DUT reading at absolute pressure indicated on the top line of the PG7000 display.
Shut vacuum reference s hutoff valve, open vacuum vent valve. W ait for press ure under
bell jar to return to ambient. Remove bell jar. If an AMH automated mass handling
sys te m is active and mass loading r esolution is 0.1 k g, the vacuum does not need to be
broken and reestablished at each increment since the masses are moved automatically.
Repeat Steps through for each desired differential pressure point.
3.9.11.1 [ENTER/SET P] in Pressure to Mass Mode
PURPOSE
To enter and execute a press ure to mass command in pressure to mass mode
(see Section 3.9.12).
OPERATION
Put the PG7000 in pressure to mas s operation mode (see Section 3.9. 12), then
press [ENTER/SET P] in any run screen. If automated rotation is on, the
<DECELERATING> screen shows until piston deceleration is complete.
Note
Pressing [ENTER] in the run screen causes automated pressure
generation to be suspended if ON (see Section 3.9.9) and
AutoRotate to stop piston rotation if ON (see Section 3.9.8).
When [ENTER] is pressed to confirm mass entry, automated
pressure generation and/or motorized rotation resume.
3. Entry field for the target value of pressure to
be set.
Use the numerical keys and editing keys to enter the target pressure value desired.
Press [ENTER/SET P] to process the target pressure value. If the pressure
value entered cannot be executed, an explanatory error message is displayed
momentarily and op eration returns to the target pres sure entry screen with the
previous target pressure va lue displayed. If the target pressure value entered is
valid, the value is processed and operation proceeds to the mass loading
instruction screen. The display is:
1. Mass to load using trim masses.
2. Nominal mass to load using main and
fractional masses.
Note
If an AMH automated mass handlin g system is activ e and mass
loading resolution is set to 0.1 kg, the AMH loads the mass
automatically. As it does, the steps of its operation are
displayed. If resolution is set to higher than 0.1 kg, the operator
is prompted to load the trim mass value.
With PG7601 in absolute by vacuum measurement mode (see Section 3.9.4), the instruction
is <Load mass & vac:> indicating that the nominal mass value should be loaded and
then the bell jar should be installed and vacuum established under the bell jar.
Load the nominal m ass value foll owing the pr otocol d escribed in Section 3 .6 and
press [ENTER/SET P]. When [ENTER/SET P] is pressed confirming that the
nominal mass value has been loaded, operation returns to the previous run
screen with the new pressure target and mass value active.
Caution
To make “in tolerance” measurements, it is imperative that all
mass loading instructions be executed following the protocol
describe in Section 3.6. This ensures that the actual mass value
resulting from a nominal mass loading command will be
correct. Failure to load masses following the PG7000 mass
loading protocol is likely to result in out of tolerance mass load
determinations and pressure definitions.
Note
•The resolution with which the pressure to mass mode m ass
loading instruction is given depends on the resolution set in
the RES function. The RES function makes it possible to
avoid loading high resolution mass values when it is not
imperative that the pressure defined be exactly the nominal
pressure requested (see Section 3.9. 10) .
•The pressure to mass loading instruction is given in
nominal mass while the main run screen displays the true
mass loaded. For this reason, the nominal mass loading
instruction and the true mass displayed in the main run screen
are slightly different values. This is normal operation (see
Section 3.6).
To enter and execute a mass t o pressure command in mass to pressure mode
(see Section 3.9.12).
OPERATION
Put the PG7000 in mass to press ure operation mode (see Section 3.9. 12), then
press [ENTER/SET P] in any run screen. If automated rotation is on, the
<DECELERATING> screen shows until piston deceleration is complete.
Note
Pressing [ENTER] in the run screen causes automated pressure
generation to be suspended if ON (see Section 3.9.9) and
AutoRotate to stop piston rotation if ON (see Section 3.9.8).
When [ENTER] is pressed to confirm mass entry, automated
pressure generation and/or motorized rotation resume.
The display is:
1. Edit field for total trim mass currently loaded.
2. Edit field for nominal mass of main and
fractional masses currently loaded.
Use the numerical and editing keys to enter the nominal mas s to be loaded on
the piston following the mass loading protocol described in Section 3.6.
Press [ENTER/SET P] to process the mass value. If the mass value entered
cannot be executed, an ex planatory error mess age is dis played m om entar ily and
operation returns to the m ass entry scr een with the pr evious nom inal m ass value
displayed. If the mass value entered is valid, the value is processed and
operation proceeds to the previous run screen with the new mass value active.
Caution
To make “in tolerance” measurements, it is impe rative that the
value of mass loaded on the piston be the NOMINAL mass
following the protocol describe in Section 3.6. This ensures
that PG7000 will correctly determine the true mass value
loaded. Failure to enter nominal mass values following the
PG7000 mass loading protocol is likely to result in out of
tolerance mass load determination and pressure definitions.
Note
•The setting of the RES function has no effect on the
resolution of mass load entries in mass to pressure mode
(see Section 3.9.10).
•The mass to pressure mass entry is exp ressed in nominal
mass while the MAIN run screen displays the true mass
loaded. For this reason, the nominal mass loading
instruction and the true mass displayed in the main run
screen are slightly different values. This is normal operation
(see Section 3.6).
3.9.11.3 Commands for Zero Pressure, Ending a Test
Entering a value of zero as the target presure is a convenient w ay to end a test
and vent the automated pressure control component when AutoGen is ON.
PG7000™ OPERATION AND MAINTENANCE MANUAL
Unknown pressure except in absolute by
current value of atmospheric pressure is
atmospheric pressure when the system is
???----- psi a
--- mm --------kg
Select entry mode:
1pressure 2mass
Entering a value of zero in pressure to mass mode causes the following
sequence to occur:
Stop piston rotation if AutoRotate is ON.
Vent pressure cont rol co mponent if AutoG en is ON or prompt op erato r to vent.
A special run screen disp lays reflecting that the PG7000 is at rest an d the
mass loading bell may not be installed. The display is:
1. <?> in ready/not ready indicator positions as
status of piston is unknown since mass loading
bell may not be installed.
2.
atmosphere measurement mode in which the
displayed. This is the value of absolute by
vented.
3. Unknown/meaningless mass load.
4. Unknown/meaningless piston position.
3.9.12 [P OR M]
PURPOSE
To select between PG7000 pressure to mass or mass to pressure operation mode.
PRINCIPLE
Piston gauges are generally used either to define desired pressure set points (e.g. when
applying reference pressures to a device to be calibr ated) or to measure a static pressur e
(e.g. when performing a crossfloat intercomparison with another piston gauge).
PG7000s support thes e two typical situations with two operating modes: pr essure to mass
and mass to pressure.
In pressure to mass operating mode, the operator enters target pressure values and the
PG7000 provides instructions of the mass to load to achieve the desired target press ur e.
In mass to press ure operating m ode, the operator enters the mass currently l oaded and the
PG7000 determines the pressure resulting from the current mass load. Mass to pressure mode is also useful to determine the true mass resulting from a nominal mass load (see
Section 3.6).
OR M function is used to s et the PG 7000 op erating m ode to eit her pres sure to mas s or
The P
mass to pressure.
OPERATION
Note
For details on pressure to mass and mass to pressure operation, see
Section 3.9.11.
To access the P OR M function, press [P OR M],
the display is:
Selecting <1pressure> activates pressure to mass mode and returns to the previous run screen.
Selecting <2mass> activates mass to pressure mode and returns to the previous run screen.
PRINCIPLE
Motorized rotati on engages and disenga ges to rot ate or stop the rota tion of the PG7000 piston.
3. GENERAL OPERATION
12 rpm ACCELERATING
+ 2.05 mm 0.1/min
48 rpm DECELERATING
+ 2.05 mm 0.1/min
Motorized rotation can be set to operate automatically to maintain the piston within Ready
condition rotation rate lim its when the piston is f loating and stop rot ation before chang ing the
pressure or mass load (see Section 3 .9.8). Motorized rotation can a lso be eng aged m anuall y
to accelerate or brake piston rotation at any time under direct operator control.
] is used for m omentary acceleration of piston rotat ion. [] followed by [←] is used to
[
start a function that stops piston rotation.
OPERATION
To momentarily engage th e m ot ori zed pist on r ota tio n s ystem and accelerate the pi s ton, pres s
] from any run screen. If PG7000 is equipped with motorized rotation, the motorized
[
rotation system engages and remains engaged until maximum rotation rate has been
achieved or the key is released.
st
The display is a modified version of the 1
system run sc reen to indicate the p is to n rot a tio n is
being accelerated while showing rotation rate and position:
1. Current piston rotation rate.
2. Current rate of piston vertical displacement.
3. Current piston position.
To start the piston braking function press and hold [
], then press [←]. If PG7000 is
equipped with m otorized rotat ion, the motori zed rotatio n system engages and sta ys engaged
until the piston rotati on is s toppe d. O nce the braking func tion st ar ts the keys may be released
and the function will complete unless [ESCAPE] is pressed.
The display is a modified version of the 1st system run sc reen to indicate the p is to n rot a tio n is
being decelerated while showing rotation rate and position:
1. Current piston rotation rate.
2. Current rate of piston vertical displacement.
3. Current piston position.
To interrupt the piston rotation braking routine, press [ESCAPE].
•Most PG7000 platforms produced before 2006 use a solenoid based
rotation actuation system. The system is protected against overloading. If
the duty cycle becomes excessive, motorized rotation cuts OFF and
remains off for a 25 second delay. During the cutoff time, [
] has no
effect.
•The motorized rotation sy stem can be set to operate automatically based
on current piston position and rotation rate. See Section 3.9.8 for
information.
•In PG7000 high line differentia l mode (see Section 3.9.4.2), p ressing [
]
on the reference PG7102 engages motorized rotation on both the reference
and tare PG7000s.
PG7000™ OPERATION AND MAINTENANCE MANUAL
3.10 [SETUP] Menu
PURPOSE
To select, view and edit th e SETUP files that det ermine the so urce of the values that PG7000 uses in its
calculation of defined pressure and as criteria for Ready/Not Ready condition determination.
PRINCIPLE
PG7000 piston gauges perform continuous, real time calculations of the pressure defined by the floating
piston under current conditions (see Section 7.2 for the details on the calculations applied). The
calculations are used to determine the defined pressure displayed in the MAIN run screen, and to arrive
at the mass to load value when a target pressure is entered in pressure to mass mode.
The equations that PG700 0 uses to calculate pressur e include many instrument and am bient variables.
PG7000 allows the source of the value used f or each variable to be defined. For example, the value
of barometric pressure used to calculate current air density and to add to gauge pressure to define
absolute pressure in absolute by atmosphere m ode could come either from PG7000’s internal sensor,
an external barometer connected to PG7000’s COM2, a user entered value or be set to standard
atmospheric pressur e. The PG70 00 SETUP f unction allows the s ource (an d in som e cases t he value) of
the variables used in the pres sure equations to be specified. In order to allow var ious combinations of
sources and/or values to be s etup and r ecalled, SETUP f iles can be creat ed, stor ed, edited and reca lled.
The SETUP files also include certain variables used to determine PG7000 Ready/Not Ready status.
Table 19 identifies the vari ables included in the SETUP file, the f actory source setting for each variable
and the default value for each variable.
Variable source and value s electi ons are r ec orded in SET U P f iles. These files can be stored and rec al le d
so that specific com binations of variable sour ces and/or values ca n be convenientl y recalled. There are
20 SETUP files. File #1 is a factory SETUP file that cannot be edited. It includes the factory default
variable choices and values.
Note
SETUP files, USER values and PG7000’s calculation capabilities can be used to
calculate defined pressure in specific conditions independent of actual PG7000
operation.
Table 19. SETUP File Choices, Factory Preferred Choice and Normal Value
VARIABLE
Atmospheric pressure 1. Internal
Ambient temperature 1. Internal
Ambient relative humidity 1. Internal
Piston-cylinder
temperature
Gravity 1. Local
Vacuum
(PG7601 only)
Ready/Not Ready
Piston position
Ready/Not Ready
Maximum vac uum
reference pressure in
absolute by vacuum mode
(PG7601 only)
SOURCE
CHOICES
2. Normal
3. User
4. COM2
2. Normal
3. User
2. Normal
3. User
1. Internal
2. Normal
3. User
2. Normal
3. User
1. Internal
2. Normal
3. User
4. COM2
1. User 1. User
1. User 1. User 5 Pa (default)
FACTORY
DEFAULT CHOICE
1. Internal 101.325 kPa
1. Internal 20 °C
1. Internal 45 %RH
1. Internal 20 °C
1. Local 9.806650 m/s
1. Internal 0 Pa
NORMAL VALUE
(14.6959 psi)
± 2.5 mm from
midfloat position
(default)
2
Note
The factory default SETUP choice is the SETUP choice setting for all variables in a new
SETUP file. SETUP file #1 is f ixed to fa ctory SE TUP choices and cannot be altered.
OPERATION
To access the SETUP menu, press [SETUP] from the main run
screen. The display is:
#nn in the upper right hand corner displays the number of the SETUP file that is currently active.
Select <1select> to select a diff erent ac ti ve SET UP f il e. Se lect <2view> to vie w the v aria bl e c hoic es and
values of the active SETUP file. Select <3edit> to create or edit a SETUP file.
See Sections 3.10.1 to 3.10.3 for detailed information on each SETUP function operation.
Changing the current SETUP file will change the source and in some cases
the values of the variables used by PG7000 in calculating defined
pressure. SETUP files should be selected and changed only b y qualified
personnel. Incorrect SETUP file selection may cause out of tolerance
calculations and measurements. See Section 3.10, PRINCIPLE, for
information on SETUP files and their use.
OPERATION
To select a SETUP file num ber to make it the active SETUP file, press [SETUP] and select
<1select>. In the fol lowing screen enter the number of the desired SETUP file and press
[ENTER]. If a va lid SETUP file n umber has been entered, operatio n returns to t he previous
run screen with the selected SETUP file number displayed and active.
•Entering an invalid SETUP file number causes an error message to be
displayed briefly and returns to the select SETUP file screen.
•If the SETUP file selected specifies COM2 as the source of barometric
pressure, PG7000 will attempt to read a barometer on COM2 when
initializing the new SETUP file. If PG7000 is unable to read a barometer on
COM2, a communications time out message is displayed briefly and
operation r etu rns t o th e s ele ct S ETU P fi le s cre en. See S ect ion 3.1 1.5 .4 f or
information on configuring COM2 to read an external barometer.
Caution
Note
3.10.2 <2view>
PURPOSE
To view the variable source choices and values of any SETUP file number.
OPERATION
Note
•Selecting <2view> SETUP allows the variable source choices and values of
the active SETUP fi le for viewing o n l y. <2v i ew> does not m a ke the viewed
SETUP file active. To select the active SETUP file, press [SETUP] and
select <1select>.
•See Table 19 for a listing of the source choices for each variable.
To view an existing SETUP file, press [SETUP] and select <2view>. The display is:
1. Entry field for number of SETUP file to be viewed. Defaults
to SETUP file currently selected. Must be a number from 2 to
10.
Enter the desired SETUP file number and press [ENTER].
If a valid SETUP file number is entered, the display is:
Select <1atmP> to view the atmospher ic pressure variable source choic e and current value.
The value is displayed in the current pressure units (see Section 3.9.3).
Select <2ambT> to view the ambient temperature variable source choice and the
current value. The value is displayed in degrees Centigrade [
Select <3%RH> to view the am bient relati ve humidit y variable sourc e choice an d the current
value. The value is displayed in %RH.
Entry field for number of SETUP file to be edited or created.
Edit SETUP file:
#02
Edit #02 latmP 2ambT
7vac
Select <4PCT> to view the piston-cylinder temperature variable source choice and the
current value. The value is displayed in degrees Centigrade [
Select <5g> to view the gravity variabl e source choice and the c urrent value. The value is
displayed in meters per second squared [m/s
2
].
°C].
Select <6READY> to vie w the values of Ready/Not Ready status criteria (see Section 3.4).
These include:
<1Pposition>:
<2vac> (PG7601 only):
Piston position Ready/Not Ready limits (see Section 3.4.1). The value is
displayed as a band in millimeters about mid-float position [
3.5).
Maximum reference vacuum pressure limit when operating in absolute by
vacuum mode. The value is displayed in Pascal [Pa].
mm] (see Section
Select <7vac>( PG7601 only): to view the reference vacuum variable sour ce choice and the
current value. The value is displayed in Pascal [Pa].
3.10.3 <3edit>
PURPOSE
To edit an existing SETUP file and/or to create a new SETUP file.
Note
See Section 3.10, PRINCIPLE for information on SETUP files and their
purpose.
OPERATION
• SETUP file #1 is t he fac tory pr efe rr ed f il e an d it c an no t be edi ted .
• See Table 19 for a listing of the source choices for each variable.
To edit an existing SETUP file or create a new SETUP file, press [SETUP] and
select <3edit>. The display is:
1.
Defaults to SETUP file currently selected. Must be a number
from 2 to 10.
Enter the desired SETUP file number and press [ENTER].
If a valid SETUP file number is entered, the display is:
Selecting an item leads to t he varia ble sourc e choices menu for that item . The c ursor is on the
source choice that is currently selected in that SETUP file number. Select the desired
variable source. The selec tion causes the variable ch oice selection to be made and retur ns
to the edit menu. From the variable source choic e menus, pressing [ENTER] or [ESCAPE]
returns to the edit selection menu. This allows e asy, discreet movement between variable
source choices when editing a SETUP file.
Select <1atmP> to specify the atmospheric pressure var iable source f or the pres sure values
that are used to calculate am bient air density and to add t o g aug e pr es sur e measurements to
calculate absolute pr es sur e in abs o lute by addition of atmosphere mode. Selecti n g <1atmP>
offers the atmospheric pressure variable source choices:
Use real time readings from PG7000’s on-board barometer.
Use a fixed value of 101.325 kPa (14.6959) psi.
Use a fixed user entered value. If <3user> is selected, the user value must be
entered, in the current pressure unit of measure (see Section 3.9.3).
PG7000™ OPERATION AND MAINTENANCE MANUAL
°
°
<4COM2>
Use real time readings from the device connected by RS232 to PG7000’s COM2
port. The external barometer must be properly set up to communicate with PG7000
(see Section 3.11.5.4 for information on setting up to read an external barometer).
Note
For differential mode operation (see Section 3.9.4) the AtmP selection
must be <4COM2>.
Select <2ambT> to specify the ambient temperature variable source for the temperature
values that are used to calculate ambient air density. Selecting <2ambT> offers the
temperature variable source choices:
<1internal>:
<2normal>:
<3user>:
Use real time readings from PG7000’s on-board ambient temperature platinum
resistance thermometer (in the Temperature - Humidity (TH) Prob e on the platform rear panel).
Use a fixed value of 20 °C.
Use a fixed user entered value. If <3user> is selected, the user value must be
entered, in degrees Centigrade [
C].
Select <3%RH> to specify the relative humidity variable source for the relative humidity
values that are used to c alculate ambient air dens ity. Selecting <3%RH> of fers the relative
humidity variable source choices:
<1internal>:
<2normal>:
<3user>:
Use real time readings from PG7000’s on-board relative humidity sensor (in the
Temperature - Humidity (TH) Probe on the platform rear panel).
Use a fixed value of 45 %RH.
Use a fixed user entered value. If <3user> is selected, the user value must be
entered, in percent relative humidity [%RH].
Select <4PCT> to specif y the pist on-c ylinder tem per ature var iable sourc e for the tem per ature
values that are used to co mpensate piston-cylinder effective are a for tem perature. Selecting
<4PCT> offers the temperature variable source choices:
<1internal>:
<2normal>:
<3user>:
Use real time readings from PG7000’s on-board piston-cylinder temperature
platinum resistance thermometer (embedded in the piston-cylinder module
mounting post).
Use a fixed value of 20 °C.
Use a fixed user entered value. If <3user> is selected, the user value must be
entered, in degrees Centigrade [
C].
Select <5g> to spec ify the value of ac celeration due to gravity that is used in calc ulating the
force applied to the piston. Pressing <5g> offers the gravity variable source choices:
<1locall>:
<2normal>:
<3user>:
Use the fixed value of local gravity stored under [SPECIAL], <6gl> (see Section
3.11.6).
Use a fixed value of 9.806650 m/s
Use a fixed user entered value different from the current local gravity value
stored in the [SPECIAL], <6gl> menu option. If <3user> is selected, the user
value must be entered, in meters per second squared [m/s
2
.
2
].
Select <6READY> to edit the values of Ready/Not Ready status criteria (s ee Section 3.4).
These include:
<1position>:
<2vacl>
(PG7601 only):
Piston position limits. Edit the fixed value, in millimeters [mm], of the band
around mid-float position within which the piston position is Ready (see
Section 3.4.1). This also determines the limit at which the piston is refloated by
the GEN function when GEN is ON (see Section 3.9.9).
Maximum reference vacuum pressure when operating in absolute by vacuum
mode. Edit the value, in Pascal [Pa], under which the reference vacuum must
be for a vacuum Ready condition to occur (see Section 3.4.3).