©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 1
Applying Measurement Uncertainty To
Digital Multimeter Calibration
An introductory study of measurement
uncertainty and its application to digital
multimeter calibration
Teleconference:
US & Canada Toll Free Dial-In Number: 1-(866) 230-5936
International Dial-In Number:+1-281-913-1100
Conference Code: 1010759559

©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 2
Welcome
Greetings from –
Fluke Corporation
Everett, Washington, USA
We are very pleased to bring you this
presentation on measurement
uncertainty for DMM Calibration.

©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 3
Welcome
This presentation is based on Fluke’s
extensive experience with:
− Use and design of calibration
Instruments
− Our experience and understanding of the
problems faced when applying
measurement uncertainty for both
regular and accredited metrology
Thanks for your time, we hope you find it
both valuable and useful.
Welcome and Thanks!

©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 4
Presented by
Fluke’s Calibration Business Unit
and Jack Somppi
Electrical Calibration Instruments
Product Line Manager
jack.somppi@fluke.com

©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 5
Web seminar etiquette
• Choice of Audio – VOIP or Teleconference
− VOIP receives audio only while teleconference is two way
sound
• Don’t mute your phone if you have background
music enabled
• Use Q&A or chat to send me questions or request
clarification
• There will be an opportunity throughout the
discussion to pause and ask questions.
• You can view the material using either full screen
or multi window methods

©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 6
Applying Measurement Uncertainty To
Digital Multimeter Calibration
An introductory study of measurement
uncertainty and its application to digital
multimeter calibration

©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 7
Objectives
In this session you will -
• Be introduced to the concept of measurement
uncertainty and why it is important
• Observe the basic elements that influence
measurement uncertainty for DMM calibration
applications
• Study a simple but detailed example of calculating
measurement uncertainty
• Consider some benefits of automating measurement
uncertainty calculations
• Receive a variety of references for further research on
this topic

©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 8
Benefits
• Introduce measurement uncertainty to
those in calibration/metrology who are
not familiar with it
• Understand why measurement
uncertainty is important for quality
metrology
• Understand measurement uncertainty
with respect to DMM calibration
• Appreciate to the benefits of automation
• Have technical references for more
detailed information
• Obtain copies of this presentation via
email

©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 9
Measurement Uncertainty
& Why It Is Important

©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 10
Facts regarding measurement -
• Can you ever measure the true value of
something?
− No, there will always be errors
• How important is this fact?
− Very important, as measurement is never complete
unless you know how good it is!
• How is this taken into account in today’s
calibration & metrology?
− By applying & documenting the measurement uncertainty
process to the tests being done

©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 11
Measurement uncertainty in
metrology today…
Measurement errors were not rigorously evaluated in all
cases. Often in industrial labs, accuracy ratio analysis
(referred to as TUR’s or TAR’s or TSR’s) had been frequently
used to evaluate the significance of the calibrator’s errors on
the measurements. Other errors were sometimes ignored.
Individually analyzed, calculated, & documented measurement
uncertainties are more thorough and are required to be
considered - as stated in
− ANSI/ISO/IEC 17025:2005 General Requirements for the
Competence of Testing and Calibration Laboratories

©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 12
ISO 17025 –
about measurement uncertainty…
5.4.6 Estimation of uncertainty of
measurement
− 5.4.6.1 A calibration laboratory, or a testing
laboratory performing its own calibrations, shall
have and shall apply a procedure to estimate the
uncertainty of measurement for all calibrations
and types of calibrations.

©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 13
… about the sources of
uncertainty…
ISO 17025, Section 5.4.6.3:
− NOTE 1: Sources contributing to the uncertainty
include, but are not necessarily limited to,
• The reference standards and reference
materials used
• Methods and equipment used
• Environmental conditions
• Properties and condition of the item being
tested or calibrated
• Operator
There are many contributors to uncertainty

©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 14
ISO 17025, Section 5.10.4
Calibration Certificates shall include …
for the interpretation of calibration results
a. The conditions of the test
b. The uncertainty of measurement &
compliance statements to metrological standards
c. Evidence of traceability
When statements of compliance are made, the
uncertainty of measurement shall be taken into account
…about calibration certificates…

©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 15
An example of an accredited
calibration certificate –
“Measurement uncertainties at the
time of test are given in the following
pages, where applicable. They are
calculated in accordance with the
method described in NIST TN1297,
for a confidence level of 95% using a
coverage factor of approximately 2
(K=2).”

©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 16
To summarize the importance of
measurement uncertainty….
From the NPL UK - “A Beginner's Guide to
Uncertainty of Measurement”
• Uncertainty of a measurement tells us something about
its quality
• Uncertainty of measurement is the doubt that exists
about the results of any measurement
• For every measurement – even the most careful – there
is always a margin of doubt
• You need to know the uncertainty before you can
decide whether the tolerance is met

©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 17
“How is this Measurement Uncertainty
obtained?”

©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 18
Properly Calculating Measurement
Uncertainty – a topic often discussed &
debated among metrologists
Initially, there were no standardized
process to quantify measurement
uncertainty….
But a standard technique was agreed
upon & published in October 1993:
ISO Guide 98 - Guide to the
Expression of Uncertainty in
Measurement (a.k.a. GUM)

©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 19
In the USA, refer to one of the Guides
relating to expressing of Uncertainty in
Measurement
ANSI/NCSL Z540.2-1997 (R2002) U.S.
Guide to Expression of Uncertainty in
Measurement
http://www.ncsli.org and find it in the store
under NCSLI publications
NIST Technical Note 1297
http://www.physics.nist.gov/Pubs/guidelines/
contents.html
Recommendation:
Refer to the GUMs -
Internationally, many metrology
organizations publish similar GUMs

©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 20
Questions?
- about measurement uncertainty
or why it is important

©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 21
Measurement Uncertainty &
Calibrating DMMs
A study of applying the GUM to DMM
calibration

©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 22
First – lets look at the concept
Our initial look –
• Consider verifying a
precision digital multimeter
• With a hypothetical study
of verifying the DMM’s
measurement performance
at 100 millivolts DC
• Let’s briefly look at what
measurement uncertainty
could be in this case

©Fluke Calibration 2011 Basics Of Measurement Uncertainty for DMM Calibration 23
Some sources of measurement
“doubt” when verifying a DMM
• The most obvious & significant sources of doubt:
− Inaccuracy of the calibrator’s output value
• 100.0000 mV might actually be 100.0000 mV .0030 mV
− Repeatability or randomness in measurement values from the DMM
• 100.0003 mV, 99.9995 mV, 100.0010 mV, etc.
− Resolution or sensitivity limits on the DMM
• It’s value is ½ the least significant digit,
• in this example it represents 0.05 V
• Many other factors that could also contribute to uncertainty:
− ambient temperature effects, thermal emfs, noise, loading, power line
conditions, etc.
• Consider all factors and include if they significantly contribute to
measurement uncertainty