FLUKE COMPASS FOR MOLBOX User Manual

COMPASS® for molbox™
Flow Calibration Software
© 1998-2003 DH Instruments, Inc.
COMPASS for molbox™ User’s Manual
© 1998 - 2003 DH Instruments, Inc. All rights reserved. Information in this document is subject to change without notice. No part of this document may be reproduced or transmitted in any
form or by any means, electronic or mechanical, for any purpose, without the express written permission of DH Instruments, Inc. 4765 East Beautiful Lane, Phoenix AZ 85044-5318 USA.
DH Instruments makes sincere efforts to the ensure accuracy and quality of its’ published materials; however, no warranty, expressed or implied, is provided. DH Instruments disclaims any responsibility or liability for any direct or indirect damages resulting from the use of the information in this manual or products described in it. Mention of any product does not constitute an endorsement by DH Instruments of that product. This manual was originall y composed in English and subsequently translated into other languages. The fidelity of the translations cannot be guaranteed. In case of conflict between the English version and other language versions, the English version takes precedence.
DH Instruments, DH, DHI, molbox, molbloc and COMPASS are trademarks, registered and otherwise, of DH Instruments, Inc.
Windows is a registered trademark of Microsoft Corporation.
030918 550087h Printed in the USA.
© 1998-2003 DH Instruments, Inc.
COMPASS for molbox™ User’s Manual
T
AABBLLEE OOFF
T
C
OONNTTEENNTTS
C
S
TABLE OF CONTENTS................................................................................I
TABLES................................................................................................... VI
FIGURES ............................................................................................... VIII
USER REGISTRATION ..............................................................................XI
ABOUT THIS MANUAL ............................................................................. XII
1. INTRODUCTION.................................................................................................................1
1.1 PRODUCT OVERVIEW.........................................................................................................................1
2. GETTING STARTED...........................................................................................................3
2.1 OVERVIEW............................................................................................................................................3
2.2 SYSTEM REQUIREMENTS...................................................................................................................3
2.3 INSTALLING COMPASS FOR molbox................................................................................................3
2.3.2 UPGRADING FROM VERSIONS LOWER THAN 4.00.............................................................................3
2.3.3 [TOOLS], [FILE UPGRADE WIZARD]......................................................................................................5
2.4 RUNNING THE PROGRAM ..................................................................................................................5
2.5 UNINSTALLING....................................................................................................................................5
2.3.1.1.1.1 OR..........................................................................................................................................3
3. OPERATING PRINCIPLES.................................................................................................7
3.1 OVERVIEW............................................................................................................................................7
3.1.1 BASICS SUMMARY..................................................................................................................................7
4. QUICK TIPS........................................................................................................................9
4.1 HOW DO I?............................................................................................................................................9
5. A P P L I CA T I O N E X A M PL E S ..........................................................................................13
5.1 OVERVIEW..........................................................................................................................................13
Page i © 1998-2003 DH Instruments, Inc.
5.1.1 EXAMPLE #1...........................................................................................................................................13
5.1.1.1 SET UP AN RS-232 DUT ........................................................................................................14
5.1.1.2 SET UP A molbox AS THE REFERENCE...............................................................................15
5.1.1.3 RUN THE TEST.......................................................................................................................16
5.1.2 EXAMPLE #2...........................................................................................................................................18
5.1.2.1 SETUP <VOLTAGE/CURRENT> OUTPUT TYPE..................................................................18
5.1.2.2 REFERENCE SETUP..............................................................................................................18
5.1.2.3 RUN THE TEST.......................................................................................................................18
5.1.3 EXAMPLE #3...........................................................................................................................................19
5.1.3.1 DUT DEFINITION SETUP.......................................................................................................19
5.1.3.2 DMM CONFIGURATION SETUP............................................................................................19
5.1.3.3 RUN THE TEST.......................................................................................................................20
5.1.4 EXAMPLE #4...........................................................................................................................................20
5.1.4.1 DUT DEFINITION SETUP.......................................................................................................20
5.1.4.2 TEST DEFINITION SETUP .....................................................................................................21
5.1.4.3 TEST HARDWARE SETUP.....................................................................................................21
5.1.4.4 RUN THE TEST.......................................................................................................................22
5.1.5 EXAMPLE #5...........................................................................................................................................22
5.1.5.1 DUT DEFINITION SETUP.......................................................................................................22
5.1.5.2 TEST DEFINITION SETUP .....................................................................................................23
5.1.5.3 TEST HARDWARE SETUP.....................................................................................................23
5.1.5.4 RUN THE TEST.......................................................................................................................24
5.1.6 EXAMPLE #6...........................................................................................................................................24
5.1.6.1 DUT DEFINITION SETUP.......................................................................................................24
5.1.6.2 TEST DEFINITION SETUP .....................................................................................................24
5.1.6.3 RUN THE TEST.......................................................................................................................25
5.1.6.4 PLOT THE DATA.....................................................................................................................25
5.1.7 EXAMPLE #7...........................................................................................................................................25
5.1.7.1 DUT DEFINITION SETUP.......................................................................................................25
COMPASS for molbox™ User’s Manual
5.1.7.2 TEST DEFINITION SETUP .....................................................................................................25
5.1.8 EXAMPLE #8...........................................................................................................................................26
5.1.7.3 RUN THE TEST.......................................................................................................................26
5.1.8.1 DUT DEFINITION SETUP.......................................................................................................26
5.1.8.2 TEST DEFINITION SETUP .....................................................................................................27
5.1.8.3 TEST HARDWARE SETUP.....................................................................................................27
5.1.8.4 SET UP DATA GRID ...............................................................................................................27
5.1.8.5 RUN THE TEST.......................................................................................................................28
6. MAIN SCREEN .................................................................................................................29
6.1 OVERVIEW..........................................................................................................................................29
6.2 MAIN MENU BAR................................................................................................................................29
6.3 STATUS BAR......................................................................................................................................29
6.4 RUN MODES.......................................................................................................................................30
6.5 RUN TOOLBAR...................................................................................................................................31
6.6 CONTROL TOOLBAR.........................................................................................................................33
6.7 RUN SCREENS...................................................................................................................................42
6.5.1 RUN TOOLS............................................................................................................................................31
6.5.2 DISPLAY TOOLS ....................................................................................................................................32
6.6.1 CHANGE/RELOAD molbloc...................................................................................................................34
6.6.2 TARE molbox..........................................................................................................................................34
6.6.3 PURGE FUNCTION.................................................................................................................................35
6.6.4 LEAK TEST .............................................................................................................................................35
6.6.5 MODIFY molbox VALVE DRIVERS .......................................................................................................36
6.6.6 FLOW CONDITIONS...............................................................................................................................36
6.6.7 UNITS SELECTION.................................................................................................................................37
6.6.8 FLOW CONTROL....................................................................................................................................38
6.6.9 ENTER FLOW DATA ..............................................................................................................................39
6.6.10 DATA ACQUISITION TOOLS.................................................................................................................39
6.7.1 DUT/REFERENCE COMPARISON RUN SCREEN................................................................................42
6.7.2 molbox OUTPUT RUN SCREEN............................................................................................................43
6.7.3 DUT OUTPUT RUN SCREEN .................................................................................................................45
6.7.4 AVERAGE DATA RUN SCREEN............................................................................................................46
6.7.5 DATA GRID RUN SCREEN ....................................................................................................................47
6.7.6 DATA PLOT RUN SCREEN....................................................................................................................48
6.7.7 SPY WINDOWS.......................................................................................................................................48
7. [RUN] MODES..................................................................................................................51
7.1 OVERVIEW..........................................................................................................................................51
7.2 [RUN], [RUN TEST].............................................................................................................................51
7.2.1 TEST INITIALIZATION............................................................................................................................51
7.2.2 RUN TEST ...............................................................................................................................................57
7.2.3 TEST CONCLUSION...............................................................................................................................59
7.2.1.1 INITIALIZE molbloc/molbox.....................................................................................................52
7.2.1.2 SELECT STARTING molbloc CHANNEL................................................................................52
7.2.1.3 SELECT DUT ..........................................................................................................................53
7.2.1.4 SELECT TEST.........................................................................................................................54
7.2.1.5 SETUP DUT ID/INTERFACE ..................................................................................................55
7.2.1.6 USER ID ..................................................................................................................................57
7.2.2.1 MANUAL ENTRY DATA ACQUISITION..................................................................................59
7.3 [RUN molbox].....................................................................................................................................60
7.3.1.1.1.1 FIGURE 36. <RUN MODE DATA SELECTOR> FORM.....................................................61
7.4 [RUN DUT] ..........................................................................................................................................61
8. [SETUP], [DUT] ................................................................................................................63
8.1 OVERVIEW..........................................................................................................................................63
8.2 CREATING DUT DEFINITIONS..........................................................................................................64
8.3 EDITING DUT DEFINITIONS...............................................................................................................64
8.4 DUT DEFINITION SELECTOR............................................................................................................64
8.5 DUT DEFINITION EDITOR..................................................................................................................65
8.5.1 TOOLBAR................................................................................................................................................66
8.5.2 [HEADER] TAB .......................................................................................................................................67
8.5.3 [CORRECTION] TAB ..............................................................................................................................69
8.5.4 [RANGE] TAB..........................................................................................................................................71
8.5.5 [TOLERANCE] TAB................................................................................................................................73
8.5.6 [TEST] TAB .............................................................................................................................................74
© 1998-2003 DH Instruments, Inc. Page ii
COMPASS for molbox™ User’s Manual
8.5.7 [CONDITIONS] TAB................................................................................................................................75
8.5.8 [READ] TAB ............................................................................................................................................76
8.5.9 [SET] TAB................................................................................................................................................78
9. [SETUP], [TEST]...............................................................................................................81
9.1 OVERVIEW..........................................................................................................................................81
9.2 CREATING TEST DEFINITIONS.........................................................................................................81
9.3 EDITING TEST DEFINITIONS.............................................................................................................82
9.4 TEST DEFINITION SELECTOR..........................................................................................................82
9.5 TEST DEFINITION EDITOR................................................................................................................82
9.5.1 TOOLBAR................................................................................................................................................83
9.5.2 <POINTS> TABLE...................................................................................................................................84
9.5.3 [POINTS] TAB.........................................................................................................................................86
9.5.4 [SEQUENCE] TAB ..................................................................................................................................88
9.5.5 [TARE] TAB.............................................................................................................................................89
9.5.6 [CONTROL] TAB.....................................................................................................................................91
9.5.7 [LEAK TEST] TAB ..................................................................................................................................93
9.5.8 [PURGE] TAB..........................................................................................................................................93
9.5.9 [DATA] TAB.............................................................................................................................................94
9.5.10 [AUXILIARY] TAB...................................................................................................................................96
9.5.10.1 AUXILIARY DATA DEVICE SETUP......................................................................................96
10. [SETUP], [SYSTEM]/ [SETUP], [HARDWARE]...............................................................99
10.1 OVERVIEW..........................................................................................................................................99
10.2 [SETUP], [SYSTEM]..........................................................................................................................100
10.3 [SETUP], [HARDWARE] ...................................................................................................................102
10.3.1 TOOLBAR..............................................................................................................................................103
10.3.2 HARDWARE DEFINITION EDITOR......................................................................................................104
10.3.3 [REFERENCE] TAB ..............................................................................................................................104
10.3.4 [DM M ] , [ F R EQ U E N C Y ] , [ P R E S S U R E ] , [ T E M PE R A T U R E ] T AB S...........................................107
10.3.5 [POWER SUPPLY] TAB .......................................................................................................................111
10.3.7 [AUXILIARY]..........................................................................................................................................118
10.3.8 [INTERFACE] TAB................................................................................................................................119
11. [TOOLS], [OPTIONS].....................................................................................................121
11.1 OVERVIEW........................................................................................................................................121
11.2 [molbox] TAB....................................................................................................................................121
11.3 [GRID] TAB .......................................................................................................................................124
11.4 [MAINTAIN LISTS] TAB....................................................................................................................126
11.5 [INITIALIZE TEST] TAB....................................................................................................................127
11.6 [RUN TEST] TAB ..............................................................................................................................128
11.7 [END TEST] TAB...............................................................................................................................130
11.8 [DATA FILE] TAB..............................................................................................................................132
11.9 [DISPLAY] TAB AND [DISPLAY] MENU..........................................................................................134
12. [DATA]............................................................................................................................137
12.1 OVERVIEW........................................................................................................................................137
12.2 [Data], [View Data File] ....................................................................................................................137
12.3 [Data], [Print Test Data] ...................................................................................................................138
12.4 [Data], [Plot Data File]......................................................................................................................138
12.4.2 PLOT PROPERTIES .............................................................................................................................141
12.4.2.1 GENERAL............................................................................................................................141
12.4.2.2 SUBSETS............................................................................................................................142
12.4.2.3 BEST FIT.............................................................................................................................142
12.4.2.4 CUSTOM PLOT...................................................................................................................143
13. REMOTE INTERFACING................................................................................................145
13.1 OVERVIEW........................................................................................................................................145
13.2 REMOTE COMMAND EDITOR .........................................................................................................146
Page iii © 1998-2003 DH Instruments, Inc.
13.2.1 INITIALIZATION COMMANDS..............................................................................................................146
13.2.2 READ COMMANDS...............................................................................................................................146
13.2.3 SET COMMANDS..................................................................................................................................147
13.2.4 EDITOR FEATURES .............................................................................................................................148
13.2.4.1 <LEADING CHARACTERS TO STRIP> .............................................................................148
COMPASS for molbox™ User’s Manual
13.2.4.2 <INSERT SPECIAL CHARACTER>....................................................................................149
13.2.4.3 <COMMAND TERMINATOR>.............................................................................................149
13.3 RS-232 INTERFACE .........................................................................................................................149
13.3.1 RS-232 SETUP......................................................................................................................................149
13.4 IEEE-488 INTERFACE ......................................................................................................................150
13.5 DEVICENET INTERFACE.................................................................................................................150
13.6 [TOOLS], [REMOTE COMMUNICATIONS]......................................................................................150
14. [TOOLS], [FLOW UNIT EDITOR]...................................................................................153
14.1 OVERVIEW........................................................................................................................................153
14.2 FLOW UNIT EDITOR.........................................................................................................................154
14.3 UNIT CONVERTER ...........................................................................................................................156
15. [TOOLS], [PROCESS GAS EDITOR] ...........................................................................159
15.1 OVERVIEW........................................................................................................................................159
15.2 FEATURES........................................................................................................................................160
15.3 GAS BLENDS....................................................................................................................................162
16. [DATA], [REPORT EDITOR]..........................................................................................163
16.1 OVERVIEW........................................................................................................................................163
16.2 REPORT PRINCIPLES......................................................................................................................163
16.3 MAIN MENU BAR..............................................................................................................................163
16.4 REPORT EDITOR AUTOMATION ....................................................................................................176
16.3.1 [FILE].....................................................................................................................................................164
16.3.1.1 [FILE ], [ED IT RE P ORT ]................................................................................................165
16.3.1.2 [FILE ], [GE NER A TE RE P ORT ] .................................................................................165
16.3.1.3 [FILE], [EDIT TEMPLATE]...................................................................................................165
16.3.1.4 [FILE], [CREATE TEMPLATE].............................................................................................166
16.3.1.5 [FILE], [SAVE]......................................................................................................................167
16.3.1.6 [FILE], [SAVE AS]................................................................................................................167
16.3.1.7 [FILE], [SAVE ALL] ..............................................................................................................167
16.3.1.8 [FILE], [CLOSE]...................................................................................................................167
16.3.1.9 [FILE], [CLOSE ALL]............................................................................................................167
16.3.1.10 [FILE], [PRINT] ..................................................................................................................168
16.3.1.11 [FILE], [PRINT ALL]...........................................................................................................168
16.3.1.12 [FILE], [EXIT] .....................................................................................................................168
16.3.2 [EDIT].....................................................................................................................................................168
16.3.2.1 [EDIT], [UNDO]....................................................................................................................168
16.3.2.2 [EDIT], [CUT] .......................................................................................................................168
16.3.2.3 [EDIT], [COPY] ....................................................................................................................169
16.3.2.4 [EDIT], [PASTE]...................................................................................................................169
16.3.2.5 [EDIT], [FIND] ......................................................................................................................169
16.3.2.6 [EDIT], [FIND NEXT]............................................................................................................170
16.3.2.7 [EDIT], [SELECT ALL] .........................................................................................................170
16.3.3 [FORMAT]..............................................................................................................................................170
16.3.3.1 [FORMAT], [INSERT DATE/TIME] ......................................................................................170
16.3.3.2 [FORMAT], [INSERT FIELD KEY].......................................................................................170
16.3.3.3 [FORMAT], [SELECT TEMPLATE]......................................................................................171
16.3.3.4 [FORMAT], [MULTIPLE CYCLES].......................................................................................172
16.3.3.5 [FORMAT], [PAGE SETUP].................................................................................................173
16.3.4 [OPTIONS].............................................................................................................................................173
16.3.4.1 [OPTIONS], [TOOLBAR].....................................................................................................173
16.3.4.2 [OPTIONS], [FONT FORMAT BAR]....................................................................................173
16.3.4.3 [OPTIONS], [RULER]..........................................................................................................175
16.3.4.4 [OPTIONS], [PATH].............................................................................................................175
16.3.5 [WINDOW].............................................................................................................................................176
16.3.6 [HELP] ...................................................................................................................................................176
16.4.1 DRAG AND DROP ................................................................................................................................176
16.4.2 COMMAND LINE ARGUMENTS...........................................................................................................177
16.4.3 DDE (DYNAMIC DATA EXCHANGE)...................................................................................................177
17. [HELP].............................................................................................................................179
17.1 OVERVIEW........................................................................................................................................179
© 1998-2003 DH Instruments, Inc. Page iv
COMPASS for molbox™ User’s Manual
18. DATA FILES ...................................................................................................................181
18.1 OVERVIEW........................................................................................................................................181
18.2 DATA FILE CREATION.....................................................................................................................181
18.3 NAMING AND STORING DATA FILES ............................................................................................181
18.4 DATA FILE STRUCTURE .................................................................................................................182
18.4.1 DATA FILE FIELDS...............................................................................................................................183
19. CALCULATIONS............................................................................................................187
19.1 OVERVIEW........................................................................................................................................187
19.2 DUT FLOW ........................................................................................................................................187
19.3 DUT SET POINTS .............................................................................................................................188
19.4 DUT ERRORS ...................................................................................................................................188
19.5 DENSITY CORRECTIONS ................................................................................................................189
19.6 TOLERANCE.....................................................................................................................................191
19.7 BEST FIT........................................................................................................................................... 192
19.8 LINEARITY........................................................................................................................................192
19.9 HYSTERESIS ....................................................................................................................................193
19.10 REPEATABILITY...............................................................................................................................193
19.11 MEASUREMENT UNCERTAINTY ....................................................................................................193
19.12 FLOW UNIT CONVERSIONS............................................................................................................194
19.13 PRESSURE UNIT CONVERSIONS...................................................................................................200
19.14 WATER RATIO..................................................................................................................................200
20. TROUBLESHOOTING....................................................................................................203
20.1 OVERVIEW........................................................................................................................................203
21. GLOSSARY ....................................................................................................................207
22. END USER LICENSE AGREEMENT .............................................................................211
Page v © 1998-2003 DH Instruments, Inc.
COMPASS for molbox™ User’s Manual
T
AABBLLEES
T
Table 1. COMPASS Version Compatibility..................................................................................................4
Table 2. Tools Available On The Run Toolbar .......................................................................................... 31
Table 3. Tools Available On The Display Toolbar.....................................................................................32
Table 4. <Tare molbox> Panel Fields......................................................................................................35
Table 5. <Flow Conditions> Panel Fields................................................................................................37
Table 6. <Units> Panel Fields...................................................................................................................37
Table 7. <Flow Control> Panel Fields......................................................................................................39
Table 8. <Control Toolbar> Data Acquisition Options.............................................................................40
Table 9. <DUT/Reference Comparison> Run Screen Fields.................................................................. 43
Table 10. <molbox Output> Run Screen Fields......................................................................................44
Table 11. <DUT Output> Run Screen Fields............................................................................................45
Table 12. <Average Flow Data> Run Screen Fields ...............................................................................47
Table 13. <Spy Window> Information......................................................................................................49
Table 14. <Test Complete> Panel Options..............................................................................................60
Table 15. DUT Definition Editor, <Toolbar> Features..............................................................................66
Table 16. DUT Definition Editor, [Header] Tab Fields..............................................................................68
Table 17. DUT Definition Editor, [Correction] Tab Fields ........................................................................ 70
Table 18. DUT Definition Editor, [Range] Tab Fields ............................................................................... 72
Table 19. DUT Definition Editor, [Tolerance] Tab Fields ......................................................................... 74
Table 20. DUT Definition Editor, [Test] Tab Fields...................................................................................75
Table 21. DUT Definition Editor, [Conditions] Tab Fields........................................................................76
Table 22. DUT Definition Editor, [Read] Tab Fields..................................................................................77
Table 23. DUT Definition Editor, [Set] Tab Fields.....................................................................................79
Table 24. Test Definition Editor, <Toolbar> Features ..............................................................................83
Table 25. Test Definition Editor, <Points> Table......................................................................................84
Table 26. Test Definition Editor, [Points] Tab Fields................................................................................87
Table 27. Test Definition Editor, [Sequence] Tab Fields.......................................................................... 88
Table 28. Test Definition Editor, [Tare] Tab Fields...................................................................................90
Table 29. Test Definition Editor, [Control] Tab Fields..............................................................................92
Table 30. Test Definition Editor, [Leak Test] Tab Fields..........................................................................93
Table 31. Test Definition Editor, [Purge] Tab Fields.................................................................................94
Table 32. Test Definition Editor, [Data] Tab Fields...................................................................................95
Table 33. Test Definition Editor, [Auxiliary] Tab Fields............................................................................96
Table 34. <System Configuration> Screen Fields................................................................................ 101
Table 35. Hardware Configuration, <Toolbar> Options..........................................................................104
Table 36. Hardware Definition, <Edit Reference> Screen Fields..........................................................106
Table 37. <Edit DMM>, <Frequency>, <Pressure>, And <Temperature> Screen Fields...................109
Table 38. Hardware Definition, <Edit Power Supply> Screen Fields....................................................112
Table 39. Hardware Definition, <Edit Controller> Screen Fields...........................................................116
Table 40. Hardware Definition, <Edit Controller> Screen Fields...........................................................117
Table 41. Auxiliary Definition, <Edit Controller> Screen Fields.............................................................119
Table 42. [Options], [molbox] Tab Selections.......................................................................................122
Table 43. [Options], [Grid] Tab Selections............................................................................................125
S
© 1998-2003 DH Instruments, Inc. Page vi .
COMPASS for molbox™ User’s Manual
Table 44. [O
ptions], [Maintain Lists] Tab Fields ..................................................................................126
Table 45. [Options], [Initialize Test] Tab Fields....................................................................................127
Table 46. [Options], [Run Test] Tab Fields...........................................................................................128
Table 47. [Options], [Run Test] Tab, [Manual DUT P and T Entry Method] Fields...........................130
Table 48. [Options], [End Test] Tab Fields...........................................................................................131
Table 49. [Options], [Data File] Tab Fields...........................................................................................132
Table 50. [Options], [Display] Tab Fields..............................................................................................135
Table 51. Data Plot Types.......................................................................................................................140
Table 52. [Plot Properties], [General] Tab Fields.................................................................................141
Table 53. [Plot Properties], [Best Fit] Tab Fields.................................................................................143
Table 54. [Plot Properties], [Custom Plot] Tab Fields.........................................................................144
Table 55. <Flow Unit Editor> Panel Fields............................................................................................ 155
Table 56. <Flow Unit Converter> Panel Fields .....................................................................................157
Table 57. <Process Gas Editor> Tool Fields.........................................................................................160
Table 58. <Find> Panel Fields................................................................................................................170
Table 59. <Multi Cycle Data File Options> Screen Choices................................................................172
Table 60. COMPASS Report Editor, Font Format Toolbar ................................................................... 174
Table 61. COMPASS Report Editor, Ruler Features ............................................................................ 175
Table 62. Data File/Grid Run Screen Main Sections...............................................................................182
Table 63. Data File/Grid Run Screen Fields/Columns ............................................................................ 183
Table 64. Flow Unit Conversion Formulas ..............................................................................................196
Table 65. Definition Of Terms In Flow Unit Conversion Formulas..........................................................197
Table 66. Flow Unit Conversion Scaling Factors ....................................................................................197
Table 67. Supported Gas Molecular Weights..........................................................................................198
Table 68. Conversions From kg/s To mole/s For Various Gases............................................................199
Table 69. Conversion From mole/s To pccm...........................................................................................199
Table 70. Pressure Unit Conversion Chart..............................................................................................200
Table 71. Troubleshooting.......................................................................................................................203
Page vii © 1998-2003 DH Instruments, Inc.
COMPASS for molbox™ User’s Manual
F
IIGGUURREES
F
Figure 1. Example 1, DUT Range Setup...................................................................................................14
Figure 2. Example 1, DUT Read Command Setup...................................................................................15
Figure 3. Example 1, molbox1 Reference Setup ......................................................................................16
Figure 4. Example 1, System Configuration..............................................................................................16
Figure 5. Example 1, DUT Interface Setup...............................................................................................17
Figure 6. Example 1, RS-232 Setup with Ref Com2.................................................................................17
Figure 7. Example 2, DMM Setup.............................................................................................................19
Figure 8. Example 4, DUT Setup..............................................................................................................20
Figure 9. Example 4, System Configuration..............................................................................................21
Figure 10. Example 5, DUT Definition.......................................................................................................23
Figure 11. Example 5, System Configuration............................................................................................24
Figure 12. Example 7, Test Setup.............................................................................................................25
Figure 13. Status Bar ................................................................................................................................30
Figure 14. <Tare molbox> Panel.............................................................................................................34
Figure 15. <Toggle molbox drivers> Form............................................................................................. 36
Figure 16. <Flow Conditions> Panel.......................................................................................................36
Figure 17. <Units> Panel..........................................................................................................................37
Figure 18. <Flow Control> Panel............................................................................................................. 38
Figure 19. <Single Point Average Setup>.............................................................................................. 40
Figure 20. <Auto Point Average Setup> ................................................................................................ 41
Figure 21. <DUT/Reference Comparison> Run Screen.........................................................................42
Figure 22. <molbox Output> Run Screen...............................................................................................44
Figure 23. <DUT Output> Run Screen.....................................................................................................45
Figure 24. <Average Flow Data> Run Screen.........................................................................................46
Figure 25. <Data Grid> Run Screen.........................................................................................................48
Figure 26. <Data Plot> Run Screen.........................................................................................................48
Figure 27. <Spy Window> Screen...........................................................................................................49
Figure 28. <Select molbloc> Panel.........................................................................................................53
Figure 29. <DUT Selection> Tool ............................................................................................................53
Figure 30. <DUT Selector> In DUT Definition Editor...............................................................................54
Figure 31. <Select Test> Tool..................................................................................................................54
Figure 32. <Test Selector> In Test Definition Editor................................................................................55
Figure 33. <Setup DUT ID/Interface> Panel............................................................................................56
Figure 34. <Manual DUT Data Entry> Panel...........................................................................................59
Figure 35. <Test Complete> Panel..........................................................................................................59
Figure 36. <Run Mode Data Selector> Form..........................................................................................61
Figure 37. <DUT Selection> Tool ............................................................................................................65
Figure 38. DUT Definition Editor, <Toolbar>............................................................................................ 66
Figure 39. DUT Definition Editor, [Header] Tab.......................................................................................67
Figure 40. DUT Definition Editor, [Correction] Tab .................................................................................70
Figure 41. DUT Definition Editor, [Range] Tab.........................................................................................71
Figure 42. DUT Definition Editor, [Tolerance] Tab ..................................................................................73
Figure 43. DUT Definition Editor, [Test] Tab............................................................................................74
S
© 1998-2003 DH Instruments, Inc. Page viii
COMPASS for molbox™ User’s Manual
Figure 44. DUT Definition Editor, [Conditions] Tab.................................................................................75
Figure 45. DUT Definition Editor, [Read] Tab...........................................................................................76
Figure 46. DUT Definition Editor, [Set] Tab..............................................................................................78
Figure 47. Test Definition Editor, <Toolbar>............................................................................................83
Figure 48. Test Definition Editor, <Test Points> Tab...............................................................................84
Figure 49. Test Definition Editor, [Points] Tab.........................................................................................86
Figure 50. Test Definition Editor, [Sequence] Tab...................................................................................88
Figure 51. Test Definition Editor, [Tare] Tab ............................................................................................ 90
Figure 52. Test Definition Editor, [Control] Tab.......................................................................................91
Figure 53. Test Definition Editor, [Leak Test] Tab...................................................................................93
Figure 54. Test Definition Editor, [Purge] Tab.......................................................................................... 94
Figure 55. Test Definition Editor, [Data] Tab............................................................................................95
Figure 56. Test Definition Editor, [Data] Tab............................................................................................96
Figure 57. Auxiliary Data Device Selector.................................................................................................97
Figure 58. <System Configuration> Screen.........................................................................................100
Figure 59. Hardware Definition, [Reference] Tab..................................................................................105
Figure 60. Hardware Definition, <Edit Reference> Panel......................................................................105
Figure 61. Hardware Definition, [DMM] Tab...........................................................................................108
Figure 62. Hardware Definition, <Edit DMM> Panel .............................................................................. 108
Figure 63. Hardware Definition, [Power Supply] Tab............................................................................111
Figure 64. Hardware Definition, <Edit Power Supply> Panel...............................................................112
Figure 65. Hardware Definition, [Controller] Tab ..................................................................................114
Figure 66. Hardware Definition, <Edit Controller> Panel......................................................................115
Figure 67. Hardware Definition, <Auxiliary Data Device> ....................................................................118
Figure 68. Hardware Definition, [Interface] Tab.....................................................................................120
Figure 69. [Options], [molbox] Tab.......................................................................................................122
Figure 70. [Options], [Grid] Tab............................................................................................................124
Figure 71. [Options], [Maintain Lists] Tab ...........................................................................................126
Figure 72. [Options], [Initialize Test] Tab............................................................................................. 127
Figure 73. [Options], [Run Test] Tab!!!.................................................................................................128
Figure 74. [Options], [End Test] Tab!!!.................................................................................................131
Figure 75. [Options], [Data File] Tab ....................................................................................................132
Figure 76. [Options], [Display] Tab.......................................................................................................134
Figure 77. Data File Viewer.....................................................................................................................138
Figure 78. [Data], [Plot Data File] Screen.............................................................................................139
Figure 79. [Plot Properties], [General] Tab..........................................................................................141
Figure 80. [Plot Properties][Subsets] Tab...........................................................................................142
Figure 81. [Plot Properties], [Best Fit] Tab..........................................................................................142
Figure 82. [Plot Properties], [Custom Plot] Tab..................................................................................143
Figure 83. <Initialization Commands Editor> Panel............................................................................146
Figure 84. <Read Commands Editor> Panel........................................................................................147
Figure 85. <Set Commands Editor> Panel...........................................................................................148
Figure 86. <Insert Special Character> Panel.......................................................................................149
Figure 87. <RS232 Settings> Panel ...................................................................................................... 150
Figure 88. Remote Communications Tool...............................................................................................151
Figure 89. <Flow Unit Editor> Panel.....................................................................................................154
Figure 90. Unit Converter........................................................................................................................156
Figure 91. <Process Gas Editor> Tool..................................................................................................160
Figure 92. <Create Gas Blend> Panel...................................................................................................162
Page ix © 1998-2003 DH Instruments, Inc.
COMPASS for molbox™ User’s Manual
Figure 93. COMPASS Report Editor.......................................................................................................164
Figure 94. <Find> Panel.........................................................................................................................169
Figure 95. <Field Insertion Tool> Panel ............................................................................................... 171
Figure 96. <Multi Cycle Data File Options> Panel...............................................................................172
Figure 97. Page Setup Screen................................................................................................................173
Figure 98. Report Editor, Main Toolbar.................................................................................................173
Figure 99. COMPASS Report Editor, Font Format Bar ........................................................................174
Figure 100. COMPASS Report Editor, Ruler........................................................................................175
© 1998-2003 DH Instruments, Inc. Page x
COMPASS for molbox™ User’s Manual
U
U
SSEERR
R
EEGGIISSTTRRAATTIIOON
R
N
COMPASS® for molbox™
Flow Calibration Software
Please fill out this registration sheet and return it to DH Instruments. Registering as a user will allow us to contact you with important information about COMPASS possible recalls and product announcements.
USER CONTACT
Name:
Company/Organization
Address 1: Address 2:
City: State/Province:
Postal Code: Country:
Tel: Fax:
®
including product upgrades,
Email:
1)
I am using COMPASS
I will run COMPASS
2)
Please return this form by mail or fax to:
NOTE: COMPASS molbox use. Running COMPASS simultaneously on two computers is a violoation of the license agreement.
®
for molbox is a licensed software product intended for single computer or
®
for molbox Ver. ______________.
®
on a computer whose operating system is:
(circle one)
Windows 95, 98, ME, NT, 2000, XP Other__________
ATTN: COMPASS Registration
DH Instruments, Inc.
4765 East Beautiful Lane Phoenix AZ 85044-5318 USA
or via Fax: 602.431.9559 email: dhi@dhinstruments.com
Page xi © 1998-2003 DH Instruments, Inc.
COMPASS for molbox™ User’s Manual
A
BBOOUUTT
A
This manual provides the user with the basic information necessary to set up and run COMPASS for molbox. It also includes a great deal of additional information provided to help you optimize COMPASS use and take full advantage of its many features and functions.
Before using the manual, take a moment to familiarize yourself with:
Table of Contents - structure;
Section 1 – to help you get started
Section 5 - gives examples of setting up COMPASS for typical applications;
Section 4 - answers frequently asked questions;
Section 20 - for troubleshooting;
Other sections - describe each of the six (6) main menu selections in detail.
Certain words and expressions have specific meaning as they pertain to COMPASS for molbox. The Glossary (see Section 21) is useful as a quick reference for specific terms and expressions as they are used in this manual and the program.
T
T
HHIISS
M
AANNUUAAL
M
L
Manual Conventions
For those of you who don’t read manuals, go directly to Section 2.3 to install COMPASS and then
Section 3 for a summary of operating principles. Later … when you have questions or start to wonder about all the great features you might be missing, get into the manual.
Cross references are used extensively to direct you towards additional information on a topic. Cross references are generally in parentheses and give the reference’s section number. For example: (see Section 11).
[ ] indicates COMPASS menu or tab selections (for example [D always described hierarchically from highest to lowest level. For example: [To
[Maintain Lists]. < > indicates COMPASS text displays such as screen names, field names, prompts, warnings and
instructions. For example: <Enter user ID>. File names are designated in quotation marks when they do NOT include a file extension. For example:
“dat” file or *.dat.
(CAUTION) is used in the manual to identify user warnings and cautions.
(NOTE) is used in the manual to identify operating and applications advice and additional
explanations.
© 1998-2003 DH Instruments, Inc Page xii .
ata]). Menu or tab selection paths are
ols], [Options],
COMPASS for molbox™ User’s Manual
11..
I
NNTTRROODDUUCCTTIIOON
I
N
1.1 PRODUCT OVERVIEW
Welcome to COMPASS® for molbox™, the complete flow calibration software package for users of DHI molbloc/molbox flow transfer standards. COMPASS for molbox is designed to supply the missing link needed to get from individual automated hardware components to an automated calibration system and from owning a great flow standard to performing a wide variety of special calibrations and tests. Using COMPASS for molbox, you can create any number of calibration test scenarios; define characteristics of various devices under test (DUT); run tests; analyze test data and generate reports without ever leaving the program or you can export test data for analysis using other applications. COMPASS’s flexible approach makes it easily adaptable to a wide variety of hardware and allows you to adjust the level of automation for many different tasks and Hardware Definitions from manual entry flow meters to fully automated mass flow controller calibrations.
Please fill in and return the User Registration form located in this manual immediately following the
Table of Contents.
Page 1 © 1998-2003 DH Instruments, Inc.
COMPASS for molbox™ User’s Manual
N
N
OOTTEES
S
© 1998-2003 DH Instruments, Inc. Page 2
COMPASS for molbox™ User’s Manual
22..
G
EETTTTIINNGG
G
S
TTAARRTTEED
S
D
2.1 OVERVIEW
This section explains how to install COMPASS for molbox on your computer.
2.2 SYSTEM REQUIREMENTS
COMPASS for molbox is an application designed for Windows 32 bit operating systems. This includes Windows 95, 98 and NT. Although any computer running one of these operating systems has the basic requirements to run COMPASS, it is recommended that the following minimum configuration be used:
Windows 95, 98, or NT
200 MHz, Pentium processor
32 MB RAM
10 MB free hard disk space

2.3 INSTALLING COMPASS FOR molbox

Insert the COMPASS for molbox CD into the CD drive. If the auto run feature is enabled on the
PC, the setup application runs automatically.
If the setup program does not run after 10 seconds, press the Windows [Start] button and select
[Run].
In the Run dialog box, type d:\setup or f:\setup, depending on the drive letter of the CD drive of the
installation disk. Select COMPASS for molbox from the list of applications and follow the installation prompts to install the program.
2.3.1.1.1.1 OR
Insert the COMPASS for molbox CD into the CD drive. Use the Add/Remove Programs feature in the Windows Control Panel. Press the Windows
[Start] button and select [Settings] followed by [Control Panel].
Double-click the [Add/Remove Programs] icon and click the [Install] button. Follow the prompts
to install COMPASS. Select COMPASS for molbox from the list of applications and follow the installation prompts to install the program.

2.3.2 UPGRADING FROM VERSIONS LOWER THAN 4.00

During the installation process, choose the directory of any existing version of COMPASS to automatically upgrade the program. When the new version runs for the first time, the COMPASS file Upgrade Wizard (see Section 2.3.3) will run to facilitate the conversion of
Page 3 © 1998-2003 DH Instruments, Inc.
COMPASS for molbox™ User’s Manual
older DUT and Test Files to the new DUT and Test Definitions format. If the new version of COMPASS is not installed in the same location as the previous version, the automatic update option will not occur. Run COMPASS and then select [T
ools], [File Upgrade Wizard] to
convert DUT and Test Files to the new format. Most aspects of user setup information and Data Files created with COMPASS versions
lower than 4.00 are compatible with release 4.00 and higher. To the extent possible, COMPASS automatically incorporates older setup information into the required format. Table 1 summarizes COMPASS version compatibility.
Table 1. COMPASS Version Compatibility
FEATURE COMPATIBILITY COMMENT
Data Files
DUT Files
Test Files
Reports
Automatic (see Section 12)
Use the Upgrade Wizard (see Section 2.3.3)
Use the Upgrade Wizard (see Section 2.3.3)
Only *.rtf format (see Section 16)
Data Files created with COMPASS Ver. 4.00 and higher contain substantially more data than in previous versions. For this reason, Data File information such as the molbloc Tare will display as “N/A” in the <Data Viewer> and completed reports.
DUT Files are no longer maintained as a directory of individual files. Instead COMPASS uses a single database to store all DUTs. The file Upgrade Wizard is required to incorporate the DUT File information into the new database format. The Upgrade Wizard assumes that all DUT’s both control and measure flow. Since this is not true in the case of MFM (Mass Flow Meters), it may be necessary to edit the DUT’s after the upgrade process.
Just like DUT Files, Test Files are no longer maintained as a directory of individual files. Instead COMPASS uses a single database to store all test records. The file Upgrade Wizard is required to incorporate the Test File information into the new database format. COMPASS for molbox versions 4.00 and higher do not contain DUT data acquisition information. This information is now contained in the DUT Definition. The upgraded tests will contain al l other compatible information. There is no distinction between validation tests (*.tsv) and adjustment tests (*.tsa). These files are treated the same.
Only reports specifically saved as *.rtf files are compatible with the new report Editor. If a specific report is desired, regenerate it using the original Data File and report template with the desired format (see Section 16).
Templates
Process Gas
Information
DUT Models
© 1998-2003 DH Instruments, Inc. Page 4
No compatibility
Automatic
Automatic
Report templates from older COMPASS versions are not supported. The new Data File format contains substantially more data and requires several more report template features. The new Report Editor can be used to create a template with the exact same features as the older version (see Section 16).
All releases of COMPASS use the same process gas file. During the installation process, the setup program displays an overwrite prompt if a progas.txt file is already present.
Any models set up in an existing version of COMPASS are automatically supported by the new installation.
COMPASS for molbox™ User’s Manual
2.3.3 [TOOLS], [FILE UPGRADE WIZARD]
When the COMPASS installation is complete, run the program and select [Tools], [File Upgrade Wizard] to convert old DUT and Test Files to the new database format. DUT and
Test Files can be selected individually or by directory. All selected files are automatically built into the appropriate COMPASS database. The file name of the converted file is set to the <Record Label> of the database equivalent record. Press the <Next> button to proceed with the next step in the process or press <Back> to repeat a previous step. There are four (4) sequential steps to the upgrade process as described below.
Select the type of file to upgrade: DUT(*.dut) or Test Files (*.tsv, *.tsa). Determine whether files should be selected individually or globa lly from a selected directory.
When the individual option is selected, a file dialog box displays allowing the selection of a single or multiple files. The directory option displays a directory browser that should be used to locate the target directory to copy from. All files with the proper file extension are included in the <Files to upgrade> list.
The files selected for upgrade display in the <Files to upgrade> list for final confirmation. COMPASS converts all listed files into the proper database. The number of files accessed
and the number of files copied is provided for troubleshooting purposes in the event a DUT or Test File is not converted and an error message is displayed.
The upgrade process is complete. <Back> can be used to select new files or to change from DUT Files to Test Files.
Selecting DUT and Test Files more than one time will create duplicate entries in the
respective DUT and Test databases.
2.4 RUNNING THE PROGRAM
When the software installation is complete, a new Windows group is created to hold the
COMPASS program. To run the program, select it by pressing the Windows [Start] button, selecting the [Programs] sub-menu followed by the [COMPA SS fo r mo lbox] grou p. Click the < COMPA SS fo r mo lbox> icon
to start the program.
2.5 UNINSTALLING
To uninstall COMPASS, use the <Remove COMPASS for molbox> icon in the program group created during installation. As an alternative, use the Add/Remove Programs feature in the Windows Control Panel and select COMPASS for molbox. In both cases, a series of prompts must be followed in the application removal program to uninstall COMPASS. All installation files and registry updates are removed. Data directories created by running COMPASS are not removed by uninstalling. It is up to the user to manually remove these data directories. The application removal program always displays a message indicating COMPASS was not completely removed when executed after new Data Files have been created. Any file management tool such as Windows Explorer can be used to remove these data directories.
Page 5 © 1998-2003 DH Instruments, Inc.
COMPASS for molbox™ User’s Manual
N
N
OOTTEES
S
© 1998-2003 DH Instruments, Inc. Page 6
COMPASS for molbox™ User’s Manual
33..
O
PPEERRAATTIINNGG
O
P
RRIINNCCIIPPLLEES
P
S
3.1 OVERVIEW
COMPASS for molbox is an application program intended to automate the test and calibration process to the extent supported by the available calibration hardware and appropriate for the DUT being tested. COMPASS supports automated operation of DHI molbloc/molbox transfer standards but can also be use d w it h a ny t yp e of flow measurement system from any manufacturer. This feature allows COMPASS to be used s ystematically to provide consistent data formats and report generation for all of your flow calibrations.
COMPASS sets up and maintains setup requirements on devices under test (DUTs) (see Section 8) and testing procedures (see Section 9) to associate with DUTs. This information is recalled to run tests. Tests are run using the hardware that has been configured (see Section 10) in COMPASS. While running a test, data is acquired and stored in a delimited Data File (see Section 12). Within COMPASS, Data Files may be viewed, used to generate plots and to generate reports which can be customized with user editable templates (see Section 16). COMPASS operations follow conventional Windows protocol for file management, menu and message formatting, graphics and editing.
To run a test, first a DUT Definition must be created (see Section 8) to define the DUT that will be tested and a Test Definition must be created (see Section 9) to provide COMPASS with the exact testing procedure for the DUT. The Hardware Definition, must be set up with the necessary instruments to support the DUT, including the setup of the molbox remote interface. If the DUT requires voltage or current measurement, a DMM must be set up using the Hardware Definition Tool (see Section 10). The devices set up in the Hardware Definition must be selected as the current System Configuration (see Section 10.2). Once a DUT and test have been set up, and the Hardware Definition has been defined and selected, tests can be run (see Section 7.2).
The results of the test are stored in a Data File (see Section 18) that is designed to be used by COMPASS to generate reports or easily exported to other applications by the user for further analysis. Over 40 individual data items are logged for each test point. COMPASS includes integrated analysis and reporting functions. Fully customizable reports and plots can be created using Data Files (see Section 16).
Other than running tests, COMPASS supports direct molbox and DUT run options to allow remote control of the molbox and/or a DUT with all COMPASS features active (see Sections 0, 7.4). This feature provides a remote interface to all measurement and control instruments required by the DUT and setup in the current System Configuration (see Section 10.2). The output information displays on various program run screens. Use this feature to perform tests or operations that do not require creation of a Data File or execution of a specific test script.
Many operational preferences relating to how tests are initialized and run, and how data is gathered and stored are configurable with user options (see Section 11). The use of these configurable preferences results in a customized test environment to meet a very wide variety of testing requirements. By selecting the proper options, a user can start an automated test and return to a custom formatted report.
Detailed information and assistance on COMPASS application and functions are available in this manual (see Sections 4, 5 and 20) and with on-line help (accessed by pressing [F1] or selecting [H
elp]).
3.1.1 BASICS SUMMARY
In summary, the following steps and sections should be referred to when setting up and
Page 7 © 1998-2003 DH Instruments, Inc.
running a test for the first time. Subsequent tests will only require a selection of the desired DUT and Test Definitions. The [Run Test] section of this manual has a more detailed explanation of the requirements to run a test in COMPASS (see Section 7.2). Several COMPASS examples and frequently asked questions are in the Quick Tips section of this manual (see Sections 4, 5). These sections are also very useful in getting started.
COMPASS for molbox™ User’s Manual
When first running COMPASS, make the minimum number of changes and modifications until
you are familiar with the program. The default values in all setup screens are designed for the most common COMPASS uses. This will help to avoid confusion and setup conflicts as you develop experience in using COMPASS.
Select [Setup], [DUT] to create a DUT Definition (see Section 8.2). A DUT that outputs
an electrical signal such as V, mV or mA will most likely be read using a digital multimeter (DMM) or the molbox analog option. The DMM must be set up in the Hardware Definition before it will be available in the System Configuration (see Section 10.3.4). The molbox analog option does not require any additional setup. Whenever the flow reference is a molbox1, the analog option is automatically available in the System Configuration.
Select [S
etup], [Test] to create a test script that describes flow points and other aspects
of the testing procedure (see Section 9).
Select [S
etup], [Hardware] to update the Hardware Definition of specific components
available for use by COMPASS (see Section 10.3). This includes setting up the molbox remote interface as well as any other peripheral devices needed to run the DUT. This option should be used to set up a DMM, power supply, pressure device, temperature device and any flow controllers that may be used to control the flow of non-flow controlling DUT’s. Once these entries are made, there is no need to access the Hardware Definition unless a change in one of these devices occurs.
Select [S
etup], [System] to set up the System Configuration of specific instruments that
will be used by COMPASS to measure reference flow, pressure, temperature, analog output and provide analog power (Section 10.2). Not all selections are mandatory. Only hardware specifically needed to run the test must be selected at any one time.
Select [R
un], [Run Test].
© 1998-2003 DH Instruments, Inc. Page 8
COMPASS for molbox™ User’s Manual
44..
Q
UUIICCKK
Q
T
T

4.1 HOW DO I?

Set up and run a test?
Create a Test Definition to define the test procedure (see Section 9).
Create a DUT Definition to define a device under test (see Section 8).
Configure the test hardware to be used to run the test (see Section 10).
Run test (see Section 7.2).
Display the flow unit I want on the molbox1 or molbox RFM front panel display
during COMPASS operation?
The various COMPASS run screens, rather than the molbox display, should be used to
obtain molbox output information. COMPASS sets the molbox to fundamental mass flow units and then performs conversions to the other units and other operating conditions when required. Do not refer to the molbox display when you are running COMPASS.
A list of standard mass flow units available for display on the molbox can be selected using the
ools], [Options, [molbox], <Display mass flow units> units feature (see Section 11.2).
[T
As an alternative, press [AVERAGE], [DISPLAY] on the molbox keypad. Select the unit of
measure that you would like displayed on the second (bottom) line of the front panel display.
(Refer to the molbox Operation and Maintenance Manual, DISPLAY Section, for details.)
Keep in mind that COMPASS does not transmit pressure and temperature values for
volume or actual flow units to the molbox even if COMPASS is making real time measurements.
Therefore, the volume flow units displayed by the molbox and by COMPASS will not agree.
S
IIPPS
Run a test in which the flow at each point is adjusted to a nominal flow value,
for example to the nominal point when calibrating a rotameter?
Specify “jog before dwell” in the Test Definition (see Section 9.5.6). This causes test
execution to pause at each point to allow flow adjustment (jog) prior to taking data at the
point.
Use COMPASS to run tests with a reference other than a DHI molbloc/molbox?
Though COMPASS only supports automated operation with DHI molbloc/molbox mass
flow references, COMPASS can be used to run manual entry tests with any flow reference.
Using the [S the <Manufacturer> (see Section 10.3.3). A test can then be run in which reference flow values are entered manually at each point. DUT readings can be automated or not as in any other test. This feature allows COMPASS to be used to standardize flow calibration procedures and data reporting even with references with which COMPASS cannot interface directly. Entry of reference flow values must be in the flow units that are selected for display on the molbox. To change these units, go to the [T tab and use <Display mass flow unit>.
Page 9 © 1998-2003 DH Instruments, Inc.
etup], [Hardware], [Reference] tab, set up a reference selecting <Other> as
ools], [Options], [molbox]
Set up COMPASS for my volume or actual flow units?
From the [Setup], [DUT], [Range] tab, select the <Unit> drop down under the “Flow Range” section of the window. Scroll through this list to find the units you desire. If they
COMPASS for molbox™ User’s Manual
do not exist they can be added using the <Edit Units> selection or the [T
low Unit Editor] option (see Section 8.5.4). When selecting volume flow units,
[F
ools], [Options],
COMPASS requires the actual operating conditions (temperature and absolute pressure)
at the DUT. The temperature and pressure can be either acquired automatically or entered manually by the user. The [S acquisition method for pressure and temperature (see Section 10.2). The [T
ptions], [Run Test] tab controls manual entry (see Section 11.6).
[O
etup], [System] selections determine the data
ools],
Test one specific DUT without having to enter its serial number and ID during test initialization?
Set the DUT up as an individual DUT (see Section 8.5.2). Individual DUTs define unique DUTs and include the DUT’s serial and ID number. When a DUT definition is run, it is assumed there is only one DUT with the specified serial number and ID.
Repeat a point in a test? Use the [Skip Back] and [Skip Forward] keys from the Run Test Toolba r (see Se ction 6. 5.1). Check what DUT or Test Definition is being run and check DUT and Test Definition
details while a test is running?
Click on the [View DUT] or [View Test] Run Test Toolbar keys to cause the DUT or Test Definition screen for the current DUT or test to appear without affecting test execution. Close or move the display to view the previous run screens (see Section 6.5.1).
Repeat the same test several times without having to pause or restart the test?
Specify multiple test cycles in the Test Definition. In the [S eque nce ] tab of the Test Definition (see Section 9.5.4) specify a number >1 under <Number of Test Cycles>. The complete flow point sequence will run for the number of times specified without initializing between runs. If all test operation aspects are automated, and [T
ools], [Options], [ End Test], <Prompt
to repeat last test point> is NOT checked, the cycles will run without interruption
(see Section 11.7).
Create a custom plot of data logged in the data file?
There are two ways of going about this. First, COMPASS provides the ability to create custom plots within the [Plot Properties] toolbar option (see Section 12.4.2). A label for the plot as well as the X and Y data can be set up. The second method requires the use of a third party spreadsheet application, MS Excel for example. Import the delimited COMPASS data file and use plot functions within the tool to create a plot.
Avoid having to go through undesired initialization steps each time a test is started?
Use the [Tools], [Options], [Initialize Test] tab to set preferences to include or exclude the various test initialization steps (see Section 11.7).
Abort a test that is running? Press the [Abort] key on the Run Test Toolbar (see Section 6.5.1) or use the [Run],
bort Run Mode] menu option while the test is running.
[A Change the default RS-232 port for reading DUTs so that I don’t have to re-select the
port manually every time in the <Setup DUT> screen of test initialization?
Use [Setup], [System], <Default DUT RS232 Port> to edit the default DUT RS-232 port (see Section 10.2).
© 1998-2003 DH Instruments, Inc. Page 10
COMPASS for molbox™ User’s Manual
Add or remove a test point while editing a test definition?
In the Test Definition <Points> Table, double clicking any test point automatically removes it from the test point list. All subsequent test points are shifted up. To add a test point press the <Insert> button and drag the pencil icon to the point location to insert. All test points will shift down leaving an opening to enter a new test point (see Section 9.5.2).
Change the resolution with which the reference and/or DUT readings are displayed while running and recorded in the Data File?
DUT resolution is determined in the DUT Definition (see Section 9.5.8). To change the DUT resolution, edit the DUT <Range> tab. Select both the <Flow> and <Output> resolution that are appropriate for the DUT. By default reference resolution is one digit greater than the specified DUT resolution.
Run an automated test if my computer has only one COM port and the molbox reference and DUT (or DMM) are both RS-232?
Connect the computer COM port to COM1 of the molbox . Connect the DUT (or DMM) to COM2 of the molbox and set the DUT or DMM to be read on Ref COM2 (see Section 10.3.4 and/or 13.3.1). This solution is only possible with DHI molbox references that have a COM2.
Troubleshoot communications with remote devices (references, DMMs, DUTs, etc.)?
Use [Tools], [Remote Communications] (see Section 13.6) to interface directly with the device, send command strings and display the responses. Be sure to refer to the device’s User’s Manual for assistance specific to the device. Changes to interface settings made in
ools], [Remote Communications] are temporary. If changes were necessary to
[T
establish communications, these changes must be made in the device’s Hardware Definition (see Section 10.3).
Reposition the main program toolbars?
Right click the <Main Toolbar> or the <Control Toolbar> to get a popup menu of options to reposition the toolbars (see Sections 6.5 and 6.6).
Determine the “characters to strip” from the response of a remote DUT or DMM?
Use [Tools], [Remote Communications] (see Section 10.3) to query the device and observe its response string.
Use the same device for more than one task in the Hardware Definition (e.g., DMM also measure frequency)?
Set up the instrument in both Hardware Definitions and select it for both functions in the System Configuration (see Section 10). Make sure the same remote interface selection is used in both setups.
Log data not already logged by COMPASS?
Use [Setup][Hardware]<Auxiliary> (see Section 10.3.7) to create the interface for the device that will be used to measure the desired output. An example Auxiliary Data setup of the valve voltage measuring ability of the molbox was installed with COMPASS. After the device has been setup, select it in a Test Definition (see Section 9.5.10) or select the device in initialization of [Run DUT] or [Run molbox].
Log data without running a test or take data over an extended period of time?
Use [Run],[Run DUT] or [Run molbox] (see Section 7.3) to begin data acquisition with the molbox, (and DUT) and other support devices. Select the [Create data file] option on the <Run Toolbar>. Then use the data acquisition options to log discrete data points or take points over time (see Section 6.5).
Page 11 © 1998-2003 DH Instruments, Inc.
Customize the test report created by COMPASS to my own requirements?
Modify the report template to reflect your requirements (see Section 16.3.1.3). It is best to select an existing template, then use the <Save As> option to create a copy of the template under a different name. Then edit the template as required to get the desired report output.
Remove items from drop down lists? Certain COMPASS drop down lists are built automatically from operator entries. Items can
COMPASS for molbox™ User’s Manual
be removed from these lists if desired. Use the [T
ools], [Options], [Maintain Lists] tab
(see Section 11.4).
Use the molbox1 MFC control option to control flow and an external DMM to measure the DUT output?
Set up a DUT Definition with a <Voltage/Current> <Output Type> (see Section 8.5.8). Set up the DMM in the Hardware Definition and specify the remote commands necessary to read the output (see Section 10.3.4). In [S
etup], [System], select the DMM in the System Configuration as the <DUT Output DMM>. On the same form, select the <Reference molbox> item as the <Set Point Power Supply> (see Section 10.2). When a test is run or [Run D
UT] is selected, the DMM specified in the System Configuration will be used to measure the DUT output. This example requires that a molbox1 with the MFC control optional installed be selected as the <Reference> in the Hardware Configuration.
Update my existing DUT and Test Files to work with this version of COMPASS?
Select [Tools], [File Upgrade Wizard] (see Section 2.3.3). Select the type of file to upgrade, either DUT or Test File. In either case you will be prompted to select whether you would like to convert all files in a directory or an individual file. Choose the target directory or the target file(s) to convert. The old file name will automatically be saved as the <Record Label> of the new DUT or Test Definition.
Import COMPASS Data Files into Excel?
From within Excel select [File], [Open], locate and select the Data File that you would like to import. Excel will open the Text Import Wizard. Select the “Delimited” option, and the “Next” button. Choose the “semicolon” delimiter (or other selected delimiter see Section 12) and complete the action by pressing the “F
inish” button. Be careful that you do not overwrite your COMPASS Data File with an Excel written file, it will no longer be compatible with COMPASS.
© 1998-2003 DH Instruments, Inc. Page 12
COMPASS for molbox™ User’s Manual
55..
A
PPPPLLIICCAATTIIOONN
A
E
XXAAMMPPLLEES
E
S
5.1 OVERVIEW
This section contains several generic application examples that cover many common COMPASS setups. Using these examples and the How Do I? tips in Section 4.1 will greatly simplify the learning process to get a test started. Although each example may not include explicit instructions on all test steps, there are three required steps that must always be followed before a test can be run:
Create (or select) a DUT Definition. Create (or select) a Test Definition. Set up the Hardware Configuration to support the DUT and Test Definitions.
Each of the examples in Sections 5.1.1 to 5.1.6 uses generic DUT command information. Refer to the actual DUT documentation when trying to run a test following any of these examples.
5.1.1 EXAMPLE #1
“How do I use a molbox1 reference to automatically calibrate a DUT that has a range of 0 to 50 sccm and uses an RS485 or digital interface?”
Most manufacturers that use RS485 and digital interfaces also supply a converter box that transforms the communications interface into RS-232. If this converter is available, set up the DUT Definition with an RS-232 as its <Output Type> on the DUT Definition [Read] tab. Without this converter box, COMPASS cannot communicate with this type of DUT.
For this example, assume the Test Definition “Test1” was previously set up for DUT control and a sequence of test points that fit the requirements of the DUT has been set up (see Section 9). The DUT in this example outputs in the format “N+24.55” as 24.55 % (correspon ding to 12.28 sccm) when passed the command (02H)“01RFX.” The (02H) is a non printable ASCII control code that has a hexidecimal value of 2 (see Section 8.5.8). These codes are sometimes required by DUTs when using an RS-232 converter. The DUT set command is “(02H)”01STD###” where” ###” represents the numerical target flow value as a percentage of full scale. The objective of this example is to demonstrate the basic requirements of a direct flow output DUT using a remote command and how to use [S The DUT in this example uses an RS-232 interface, however changing the DUT <Output Type> on the DUT Definition [Read] tab to IEEE-488 is all that is required to use IEEE-488.
etup], [System] to configure the system hardware.
Page 13 © 1998-2003 DH Instruments, Inc.
5.1.1.1 SET UP AN RS-232 DUT
Use [Setup], [DUT] to create a new DUT Definition (see Section 8). If a DUT with similar characteristics already exists, it is easier to copy and edit the existing DUT than to start from scratch. Before attempting to set up the RS-232 DUT, refer to the DUT documentation to determine the commands used to measure and control flow as well as the RS-232 interface settings. The first step is to enter all of the DUT header information and decide whether to make this an <Individual DUT> or a <DUT Profile>. If the DUT is unique, use the <Individual DUT> option to avoid the need to enter the DUT serial number and identification before starting a test (see Section 8.5.2). Next, enter the flow range information. In most cases, if a DUT outputs directly in RS-232 or IEEE-488, it will output in flow units. However, this is not the case in this example. Note that <Flow Units> are “sccm @ 0.00C” and <Output Units> are “%” (see Section 8.5.4).
COMPASS for molbox™ User’s Manual
Figure 1. Example 1, DUT Range Setup
Select RS-232 as the DUT <Output Type> on the [Read] tab and enter the RS-232 settings for this DUT: 9600,E,7,1 (see Section 8.5.8). Then click the <Edit> button next to the <Read Commands> list to display the remote command setup form. To create the control code (02H), press the <Insert Special Character> button. Select Hex then scroll to character “02” and press the insert button. The cursor will turn into a pencil. Go to the first <Remote Command> entry field and click the mouse in the left most position in the text display. A square or rectangle will display where the mouse was clicked. Now enter the DUT flow command, “01RFX” next to the control code. The DUT in this example outputs “N+24.55” as
24.55 %. So enter 1 in the <Leading characters to strip>. Press <OK> to save the command setup and return to the DUT Definition Editor.
© 1998-2003 DH Instruments, Inc. Page 14
COMPASS for molbox™ User’s Manual
Setting up the <Set Point Type> is similar to setting up the DUT <Output Type>. On the [Set] tab, first select RS-232 as the <Set Point Type>, then enter the RS-232 settings for the DUT: 9600,E,7,1. Then click the <Edit> button next to the <Set Commands> list to display the remote command setup form. Insert the (02H) control code as previously described. Now enter the DUT Set command, “01STD” next to the control code. Instead of ## for the numerical set value, either type or drag and drop the <Set Unit Variable> at the end of the set command. This tells COMPASS where to place the numerical flow target value.
Figure 2. Example 1, DUT Read Command Setup
Save this DUT using the <Save> toolbar icon. The RS-232 communications port has not been entered. This selectio n occurs as one of the last steps in test initialization, <Setup DUT(s) - Data Acquisition> (see Section 7.2.1.5).
5.1.1.2 SET UP A molbox AS THE REFERENCE
Use [Setup], [Hardware] to open the Hardware Definition screen (see Section 10.3.3). Select the [Reference] tab and check to see if the desired molbox is already in the list. If so, verify that the remote settings are correct and close the display. Otherwise press <New> to create a new reference (see Section 10.3.3). Select DH Instruments as the manufacturer and molbox1 (or RFM) as the model reference. Then enter all other support information. For this example, the molbox will use an IEEE-488 interface. The IEEE-488 address of the molbox can be determined using the molbox’s front panel [Special] key (see the molbox Operation and Maintenance Manual). In this example, 10 is used. Press <OK> to save this Reference Definition.
Page 15 © 1998-2003 DH Instruments, Inc.
COMPASS for molbox™ User’s Manual
Figure 3. Example 1, molbox1 Reference Setup
Now that the molbox is set up, it must be selected in the System Configuration. Select [S the only place the hardware to be used by COMPASS is specified. [S
etup], [System] from the main menu (see Section 10.2). This form is
etup], [Hardware]
creates Hardware Definitions which are lists of supported devices that can be selected to create the current System Configuration. Select the molbox in the <Flow Reference> list. Since the molbox will use an IEEE-488 interface, make sure an <IEEE-488 Card> selection is made. To streamline the setup, the <Default DUT RS232 Port> can also be selected. All other selections have no impact when running this example. System Configuration items are only used as required by the test that is run. When the System Configuration is complete, there is no reason to change it unless a new DUT to be tested requires a different System Configuration. Click the <Close> button to store the current System Configuration and return to the main program.
Figure 4. Example 1, System Configuration
5.1.1.3 RUN THE TEST
Use [Run], [Run Test] to execute a test (see Section 7.2). COMPASS will verify the interface of the reference molbox and determine which features it supports. Select the desired DUT and “Test1” in the test initialization (see Section 7.2.1).
© 1998-2003 DH Instruments, Inc. Page 16
Loading...
+ 196 hidden pages