CO-205
Aspirator Kit
Instruction Sheet
Introduction
The CO-205 Aspirator Kit allows the user to sample the carbon
monoxide (CO) level in the flue of a residential furnace.
The Aspirator Kit works as an extension to the Fluke
CO-210 Carbon Monoxide Probe and the Fluke CO-220
Carbon Monoxide Meter. See Figure 1.
W
Read First: Safety Information
WWarning
To avoid possible personal injury:
• Do not place the Fluke CO-210 Carbon Monoxide
Probe or the CO-220 Carbon Monoxide Meter against
a flue.
• Do not handle the aspirator probe after removing it
from a flue. Allow the probe to cool before storage.
®
Contacting Fluke
To contact Fluke, call:
1-888-993-5853 in USA and Canada
+31 402-678-200 in Europe
+81-3-3434-0181 Japan
+65-738-5655 Singapore
+1-350-5500 in other countries
Visit Fluke’s Web site at: www.fluke.com
PN 691201 (English) June 1998
© 1998 Fluke Corporation. All rights reserved.

Using the Aspirator
To use the aspirator:
1. Snap the aspirator cap to the CO-210 or CO-220.
Downstream from the furnace but before any intersections
2.
to the flue,
accommodate the aspirator probe.
3. Insert about half of the aspirator probe into the flue.
Caution: Do not leave the aspirator probe in the flue for
extended periods of time (longer than 1 hour).
4. Turn on the CO-210 or the CO-220. (If the CO-210 is
connected to a meter, turn on and configure the meter.)
5. Squeeze the aspirator bulb 4 to 5 times or until the
CO-210 or CO-220 shows a constant reading.
6. After sampling is complete, plug the sampling hole using
material appropriate for your flue.
make a 5/16” sampling hole in the flue to
4
Specifications
Hose Dimensions:
Inside diameter
Outside diameter
Length
Filter Pore Size: 1 µm
Operating Temperature:
Probe:
Hose:
Storage Temperature: -30 o C to +60 o C (-22 o F to +140 o F)
3.175 mm (1/8 inch)
6.35 mm (1/4 inch)
122 cm (48 inches)
o
371
C (700 o F)
o
204
C (400 o F)
Replacing the Filter
The PTFE filter on the aspirator hose can clog with use. A
clogged filter reduces the accuracy of the CO reading by
impeding air flow to the measuring device. Examine the filter
periodically and replace as necessary.
If you notice water in the hose on
, the filter has become saturated. Remove the filter, drain
filter
the water, and replace the filter.
the instrument side of the
3
1
2
A
Aspirator probe assembly (Fl uke P/N 691565)
B
PTFE filter (Fluke P/ N 691188. Whatman 7590-002)
C
Aspirator hose (Viton)
D
Aspirator cap (Fluke P/ N 691185)
E
Fluke CO-210 or CO-220 (Not included in 205 Aspirator Kit)
Figure 1. CO-205 Aspirator Kit
5
XKOO1F.EPS
Warranty
Limited Warranty & Limitation Of Liability
This Fluke product will be free from defects in material and workmanship for
one year from the date of purchase. This warranty does not cover fuses,
disposable batteries or damage from accident, neglect, misuse or abnormal
conditions of operation or handling. Resellers are not authorized to extend any
other warranty on Fluke’s behalf. To obtain service during the warranty period,
send your defective tester to the nearest Fluke Authorized Service Center with
a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES,
SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED
OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT,
INCIDENTAL OR CONSEQUENTIAL DAMAGES O R LOSSES, ARISING
FROM ANY CAUSE OR THEORY. Since some states or countries do not
allow the exclusion or limitation of an implied warranty or of incidental or
consequential damages, this limitation of liability may not apply to you.
Fluke Corporation Fluke Europe B.V.
P.O. Box 9090 P.O. Box 1186
Everett, WA. 5602 B.D. Eindhoven
98206-9090 The Netherlands
10/96