Each Fluke product is warranted to be free from defects in material and workmanship under normal use and
service. The warranty period is one year and begins on the date of shipment. Parts, product repairs, and
services are warranted for 90 days. This warranty extends only to the original buyer or end-user customer of
a Fluke authorized reseller, and does not apply to fuses, disposable batteries, or to any product which, in
Fluke's opinion, has been misused, altered, neglected, contaminated, or damaged by accident or abnormal
conditions of operation or handling. Fluke warrants that software will operate substantially in accordance
with its functional specifications for 90 days and that it has been properly recorded on non-defective media.
Fluke does not warrant that software will be error free or operate without interruption.
Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers
only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is
available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the
applicable international price. Fluke reserves the right to invoice Buyer for importation costs of
repair/replacement parts when product purchased in one country is submitted for repair in another country.
Fluke's warranty obligation is limited, at Fluke's option, to refund of the purchase price, free of charge repair,
or replacement of a defective product which is returned to a Fluke authorized service center within the
warranty period.
To obtain warranty service, contact your nearest Fluke authorized service center to obtain return
authorization information, then send the product to that service center, with a description of the difficulty,
postage and insurance prepaid (FOB Destination). Fluke assumes no risk for damage in transit. Following
warranty repair, the product will be returned to Buyer, transportation prepaid (FOB Destination). If Fluke
determines that failure was caused by neglect, misuse, contamination, alteration, accident, or abnormal
condition of operation or handling, including overvoltage failures caused by use outside the product’s
specified rating, or normal wear and tear of mechanical components, Fluke will provide an estimate of repair
costs and obtain authorization before commencing the work. Following repair, the product will be returned to
the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges
(FOB Shipping Point).
THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER
WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY
OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE
FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL DAMAGES OR LOSSES,
INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or
limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not
apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court or other
decision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other
provision.
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
11/99
To register your product online, visit register.fluke.com
The Fluke Calibration 9190A Ultra-Cool Drywell (the Product or Calibrator) is a benchtop temperature calibrator that can calibrate precision temperature instruments from
−95 °C to 140 °C.
This manual includes information on how to set up and turn on the Product for the first
time. For instructions on how to operate the Calibrator, see the 9190A Operators Manual
on the CD-ROM.
Contact Fluke Calibration
To contact Fluke Calibration, call one of the following telephone numbers:
Getting Started
• Technical Support USA: 1-877-355-3225
• Calibration/Repair USA: 1-877-355-3225
• Canada: 1-800-36-FLUKE (1-800-363-5853)
• Europe: +31-40-2675-200
• Japan: +81-3-6714-3114
• Singapore: +65-6799-5566
• China: +86-400-810-3435
• Brazil: +55-11-3759-7600
• Anywhere in the world: +1-425-446-6110
To see product information and download the latest manual supplements, visit Fluke
Calibration’s website at
To register your product, visit http://flukecal.com/register-product.
www.flukecal.com.
9190A
Getting Started
Safety Information
A Warning identifies conditions and procedures that are dangerous to the user. A
Caution identifies conditions and procedures that can cause damage to the Product or the
equipment under test.
Warning
To prevent possible electrical shock, fire, or personal injury:
• Read all safety Information before you use the Product.
• Use the Product only as specified, or the protection
supplied by the Product can be compromised.
• Use this Product indoors only.
• Do not use the Product around explosive gas, vapor, or in
damp or wet environments.
• Do not use and disable the Product if it is damaged.
• Use only the mains power cord and connector approved for
the voltage and plug configuration in your country and rated
for the Product.
• Replace the mains power cord if the insulation is damaged
or if the insulation shows signs of wear.
• Make sure the ground conductor in the mains power cord is
connected to a protective earth ground. Disruption of the
protective earth could put voltage on the chassis that could
cause death.
• Do not put the Product where access to the mains power
cord is blocked.
• Use caution when you install and remove probes and
inserts from the Product. They can be hot.
• Do not touch voltages >30 V ac rms, 42 V ac peak, or
60 V dc.
• Do not apply more than the rated voltage, between the
terminals or between each terminal and earth ground.
• Do not touch the well access surface of the instrument.
• Do not turn off the product at block temperatures higher
than 100 °C. Select a SETPOINT less than 100 °C and let the
instrument to cool before turning it off.
• Use the correct terminals, function, and range for
measurements.
2
• Do not use test leads if they are damaged. Examine the test
leads for damaged insulation, exposed metal, or if the wear
indicator shows. Check test lead continuity.
• Do not touch the probes to a voltage source when the test
leads are connected to the current terminals.
Ultra-Cool Drywell
Safety Information
• Keep fingers behind the finger guards on the probes.
• Do not exceed the Measurement Category (CAT) rating of
the lowest rated individual component of a Product, probe,
or accessory.
See Table 1 for a list of symbols used in this manual and on the Calibrator.
Table 1. Symbols
Symbol Description Symbol Description
Conforms to European Union
directives
Risk of Danger. Important
information. See manual.
Conforms to relevant North
American Safety Standards.
Conforms to relevant Australian
EMC requirements
Earth ground Hazardous voltage
This product complies with the WEEE Directive (2002/96/EC) marking requirements.
The affixed label indicates that you must not discard this electrical/electronic product
in domestic household waste. Product Category: With reference to the equipment
types in the WEEE Directive Annex I, this product is classed as category 9 "Monitoring
and Control Instrumentation” product. Do not dispose of this product as unsorted
municipal waste. Go to Fluke’s website for recycling information.
3
9190A
Getting Started
Calibrator Features
Table 2 identifies and describes the panels on the front of the Calibrator.
Table 2. Front Panel
9190A Model
9190A
1
3
9190A
9190A-P Model
1
2
3
4
gzs046.eps
Item Name Function
Display and Control Panel Control panel and display. See page 5.
-P Option Panel (Input Panel)
Input panel used to connect to external sensors and probes.
Panel is only available on the “-P” model. See page 6.
Power and Remote Interface Panel Power module and Remote Interface Panel. See page 7.
gzs001.eps
Ultra-Cool Drywell
Calibrator Features
Display and Control Panel
Table 3 shows and describes the function of each button on the Control Panel.
Table 3. Display and Control Panel
45
1
9190A
9
8
Item Name Function
Shows block temperature, measurements, status information, operating
Display
Arrow Keys ▲▼◄►
Enter Key
parameters, and softkey functions. The contrast of the display is
adjustable. To adjust the contrast, push ▲ to increase contrast or ▼ to
decrease contrast while the Main screen is shown.
Navigates through menu selections, increases or decreases numbers, and
scrolls menus up or down.
Selects menus and sets new values.
67
2
3
gzs002.eps
SET Point Key
°C/°F Key
Menu Key
Exit Key
Softkeys
Block Temperature
Indicator
Set a SETPOINT temperature to heat or cool to.
Switches the displayed temperature units between °C and °F. Key is
enabled only when the Main screen is shown.
Note
This key is disabled in some regions of the world.
Opens the Main menu.
Cancels all changes and navigates back to the previous menu.
Navigates the menus on the display. The functions of the softkeys are
shown on the display above the buttons.
Visual safety indicator that illuminates when the block temperature is
unsafe and extinguishes when the block temperature is safe. If the block
temperature is unsafe and the Calibrator is turned off or the mains power
cord is disconnected, the indicator flashes until the block temperature
cools to a safe temperature. Do not transport or remove inserts until the
indicator is off.
Warning
For safe operation and maintenance of the product, do not
remove Inserts when the Block Temperature indicator is
illuminated.
5
9190A
Getting Started
-P Option Panel (Input Panel)
Table 4 shows and describes the connectors and ports on the -P Option Panel. The
optional process version -P Option Panel is also referred to as the Input Panel.
Table 4. -P Option Panel (Input Panel)
2
1
4
3
5
Item Name Function
Reference
Thermometer Input
(REF PRT)
4-20 mA Connectors
Connect a Reference PRT probe to the Calibrator for use with the
reference thermometer function. See the 9190A Operators Manual for
more information.
Connect a 4-20 mA transmitter to the Calibrator. The 4-20 mA
Connectors can supply a low voltage (24 V) to power a transmitter.
gzs003.eps
6
4-Wire PRT/RTD
Connector
Thermocouple (TC)
Connector
Connect a 4-wire, 3-wire, and 2-wire PRT/RTDs to the readout to be
calibrated.
Connect a subminiature thermocouple (TC) connector.
Fuse Fuse for the 4-20 mA circuit.
Ultra-Cool Drywell
Calibrator Features
Power and Remote Interface Panel
Table 5 shows and describes the connectors and ports on the Power and Remote Interface
Panel.
Table 5. Power and Remote Interface Panel
3
4
1
2
5
Item Name Function
Mains Power Cord
Receptacle
Receptacle for the mains power cord. Use an AC mains supply
appropriate for the voltage range and region of use.
Power Switch Turn on (I) and turn off (O) the Calibrator.
9-pin Subminiature
Serial Connector (RS-232)
Transmits measurements and remotely controls the operation of
the Calibrator.
gzs004.eps
USB Serial Connector
Transmits measurements and remotely controls the operation of
the Calibrator.
Fuse Product fuse.
7
9190A
Getting Started
Main Screen
When the Calibrator turns on, the system initializes, does a self-check, then shows a
startup screen that shows the model number and firmware version information. If the selfcheck finds an error, the error is shown on the Startup screen. Contact Fluke Calibration
if an error shows on the Startup screen.
After the start-up initialization is complete, the Startup screen disappears and the Main
screen shows on the display. Table 6 shows and describes the indicators on the Main
screen.
Table 6. Main Screen
1
-80.01 °C
2
SETPT: -80.00°C
COOL: 67%
5
3
REF : -79.964 °C
6
TC-T : -79.81 °C
4
Item Name Function
Block Temperature Temperature of the internal temperature block.
SETPOINT
Temperature
Reference
Temperature
UUT Output
Heating/Cooling
Status
Stability Status
Indicator
[-P Only]
[-P Only]
Target SETPOINT temperature. A set temperature value is referred to as
a “SETPOINT”. The Calibrator uses the SETPOINT value to know what
temperature to heat or cool to.
Shows the most recent reference measurement when a Reference PRT
probe is connected and set up.
Shows the most recent UUT output measurement of a probe that is
connected and setup. The value shown depends on the output type
selected in the Input Setup Menu.
Shows the mode the calibrator is in. The modes are: OFF, COOL, HEAT,
and CUTOUT. See the 9190A Operators Manual for more information.
Visually shows if the block temperature is stable and within the stability
limits. See the 9190A Operators Manual for more information.
gzs005.eps
8
Ultra-Cool Drywell
Calibrator Setup
Calibrator Setup
Unpack and Inspect
Unpack the instrument carefully and examine it for any damage that could have occurred
during shipment. If there is shipping damage, notify Fluke Calibration and the carrier
immediately. Table 7 lists the equipment and the accessories that comes with the
Calibrator. Verify that all the equipment and accessories in Table 7 are in the box.
Table 7. Parts and Accessories
Name Quantity
9190-INSX Insert (X=A, B, C, D, E, or F) 1
6-foot Mains Power Cord 1
USB Cable 1
Getting Started Manual 1
Product CD that contains manuals and remote interface driver files 1
9930 Interface-it Calibration Software and Users Guide 1
Report of Calibration and Calibration Label 1
Well Insulator Cap 1
Insert Removal Tool 1
Clamp-on Ferrites (-P model only) 4
6-pin DIN Connector (-P model only) 1
Test Lead Kit (-P model only) 1
Placement
Put the Calibrator on a clean, flat surface. Make sure the Calibrator is 150 mm (6 inches)
away from all objects. For best results, choose a location to set up the Calibrator where
room temperature changes are minimum.
Warning
To prevent possible electrical shock, fire, or personal injury:
• Do not operate Product in orientations other than upright. A
fire hazard can be made if the Product is put on its side.
• Do not remove Inserts when the Product shows
temperatures more than 50 °C.
• Do not operate near flammable materials.
• Do not touch the well access surface of the Product.
• Do not turn off the Product when the temperature is above
100 °C. Set a SETPOINT temperature below 100 °C and let
the Product cool.
9
9190A
Getting Started
Caution
For safe operation and maintenance of the Product:
• Energize the Product for a 2-hour dry-out period before use, if the
Product was:
o In transport
o In a humid or semi-humid storage environment
o Not energized for more than 10 days
If the product is wet or has been in a wet environment, take
necessary measures to remove moisture prior to applying
power.
• Always operate this Product on a flat, level, stable surface.
• Do not store the Product at temperatures above 50 °C. The
Product has a refrigeration system and contains gasses
under pressure.
• Do not turn the Product upside down. The inserts will fall
out.
• To prevent damage to the cooling system, do not tilt the
Product on its side or upside down while the Product is
operating.
Connect to Mains Power
Use the 2-meter (6-foot) mains power cord to connect the Product to a 120 V ac or
230 V ac outlet rated for at least 15 amps.
Turn On the Product
1. Push the “I” side of the power switch on the front panel of the Calibrator.
2. Monitor the Startup screen for errors while the product turns on. If an error shows,
contact Fluke Calibration.
Change Language
To change the display language:
1. Push .
2. Push .
3. Push .
4. Push .
5. Push or to highlight a language.
6. Push to set language.
If the incorrect language is set by accident, push softkeys and at
the same time to temporarily switch back to the English language.
Note
10
Set Display Contrast
With the Main screen shown in the display, push to increase or push to decrease
display contrast.
Toggle Key Beep On or Off
With the Main screen shown in the display, push and at the same time to
enable or disable key beep.
Ultra-Cool Drywell
Calibrator Setup
Security and Password
The Calibrator has two user-level access security levels (Low and High) to protect from
undesired changes to the settings (see Table 8). The Calibrator comes from the factory
with the security level set to High and a default password of “1234”.
If the password is not available, the information can still be viewed. To view the
information without the password, push twice or push when prompted for the
password. The information is then shown on the screen, but cannot be changed.
Note
The Calibrator does not have a password reset function. If the password is
lost, contact Fluke Calibration for password reset assistance.
Table 8. Security Levels
Security Level Definition
Low
High
Protects the specific metrological information and calibration information settings.
Protects all operating parameters. It is intended to minimize user choices, for example,
to perform repeated identical calibrations under consistent conditions.
To change the password:
1. Push .
2. Push .
3. Push .
4. Enter the current 4-digit password to open the password screen (the default factory
password is
1234).
5. Push and to highlight a digit then push to increase the digit or push to
decrease the digit.
6. Push to save the password.
To change the security level:
1. Push .
2. Push .
3. Push .
4. Enter the current 4-digit password to open the password screen (the default factory
password is
5. Push and to highlight
1234).
HIGH or LOW.
6. Push to save the selection.
11
9190A
Getting Started
Specifications
Base Unit Specifications
Temperature Range at 23 °C ................................ –95 °C to 140 °C (–139 °F to 284 °F)
Display Accuracy .................................................. ±0.2 °C Full Range
Accuracy with External Reference
Stability .................................................................. ±0.015 °C Full Range
Axial Uniformity at 40 mm (1.6 in) ....................... ±0.05 °C Full Range
Radial Gradient ..................................................... ±0.01 °C Full Range
Loading Effect
±0.10 °C at 140 °C
Operating Conditions ........................................... 0 °C to 35 °C, 0 % to 90 %
Environmental conditions for all
specifications except temperature range ........... 13 °C to 33 °C
Immersion (Well) Depth ........................................ 160 mm (6.3 in)
Well Diameter ........................................................ 30 mm (1.18 in)
............................................ Type J: ±0.70 °C at 140 °C
Type K: ±0.75 °C at 140 °C
Type T: ±0.60 °C at 140 °C
Type E: ±0.60 °C at 140 °C
Type R: ±1.60 °C at 140 °C
Type S: ±1.60 °C at 140 °C
Type M: ±0.65 °C at 140 °C
Type L: ±0.65 °C at 140 °C
Type U: ±0.70 °C at 140 °C
Type N: ±0.75 °C at 140 °C
Type C: ±1.00 °C at 140 °C
TC Millivolt Range ................................................. –10 mV to 75 mV
Voltage Accuracy .................................................. 0.025 % of reading +0.01 mV
Internal Cold Junction
Compensation Accuracy ...................................... ±0.35 °C (ambient of 13 °C to 33 °C)
Built-in mA Readout Accuracy ............................ 0.02 % of reading + 0.002 mA
mA Range .............................................................. Cal 4-22 mA, Spec 4-24 mA
mA Connection ..................................................... 2 terminal input
Loop Power Function ........................................... 24 VDC loop power
Built-in Electronics Temperature Coefficient
(0 °C to 13 °C, 33 °C to 50 °C) ............................... ±0.005 % of range per °C
Notes:
[1] – For ambient temperature of 23 °C.
[2] – Time from when the SETPOINT is reached to when the unit is with in Stability specification.
[3] – The temperature range may be limited by the reference probe connected to the readout. The built-in Reference Accuracy does not
include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors.
[4] – Measurement accuracy specifications apply within the operating range and assume 4 wires for PRTs. With 3-wire RTDs add 0.05 Ω
to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.
[5] – The thermocouple input readout is sensitive to EM fields in the frequency range of 500 MHz to 700 MHz.
[3]
................ ±0.010 °C at –95 °C
[4]
..................... 0 Ω to 42 Ω: ±0.0025 Ω
13
9190A
Getting Started