Fluke 9150 User Manual

Hart Scientific
9150 Portable
Furnace
User’s Guide
Rev. 5B1601
Fluke Corporation, Hart Scientific Division
799 E. Utah Valley Drive • American Fork, UT 84003-9775 • USA Phone: +1.801.763.1600 • Telefax: +1.801.763.1010 E-mail: support@hartscientific.com
www.hartscientific.com
Subject to change without notice. • Copyright © 2005 • Printed in USA
Rev. 5B1601
Table of Contents
1 Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2.1 WARNINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2.2 CAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3 Authorized Service Centers. . . . . . . . . . . . . . . . . . . . . . 4
2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3 Specifications and Environmental Conditions . . . . . . . . . . 9
3.1 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.2 Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . 9
3.3 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4 Safety Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . 11
5 Quick Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.2 Set-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.3 Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.4 Setting the Temperature . . . . . . . . . . . . . . . . . . . . . . . 14
6 Parts and Controls . . . . . . . . . . . . . . . . . . . . . . . . 15
6.1 Back Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
6.2 Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
6.3 Constant Temperature Block Assembly . . . . . . . . . . . . . . . 17
6.3.1 Constant Temperature Block . . . . . . . . . . . . . . . . . . . . . . . . . . 17
6.3.2 Probe Sleeves and Tongs . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
7 General Operation . . . . . . . . . . . . . . . . . . . . . . . . 19
7.1 Changing Display Units . . . . . . . . . . . . . . . . . . . . . . . 19
7.2 Switching to 230V Operation . . . . . . . . . . . . . . . . . . . . 19
8 Controller Operation . . . . . . . . . . . . . . . . . . . . . . . 21
8.1 Well Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . 21
8.2 Temperature Set-point . . . . . . . . . . . . . . . . . . . . . . . . 21
i
8.2.1 Programmable Set-points . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
8.2.2 Set-point Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
8.2.3 Temperature Scale Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
8.3 Scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
8.3.1 Scan Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
8.3.2 Scan Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
8.4 Ramp and Soak Program . . . . . . . . . . . . . . . . . . . . . . 25
8.4.1 Program Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
8.4.2 Number of Program Points . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
8.4.3 Program Set-Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
8.4.4 Program Function Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
8.4.5 Program Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
8.5 Secondary Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . 28
8.6 Heater Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
8.7 Set-point Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . 29
8.8 Proportional Band . . . . . . . . . . . . . . . . . . . . . . . . . . 29
8.9 Controller Configuration . . . . . . . . . . . . . . . . . . . . . . 30
8.10 Operating Parameters . . . . . . . . . . . . . . . . . . . . . . . . 31
8.10.1 High Limit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
8.10.2 Soft Cutout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
8.10.3 Cutout Reset Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
8.11 Serial Interface Parameters . . . . . . . . . . . . . . . . . . . . . 32
8.11.1 BAUD Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
8.11.2 Sample Period. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
8.11.3 Duplex Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
8.11.4 Linefeed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
8.12 Calibration Parameters . . . . . . . . . . . . . . . . . . . . . . . 34
8.12.1 Hard Cutout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
8.12.2 CT1, CT2, and CT3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
8.12.3 CE1, CE2, and CE3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
9 Digital Communication Interface . . . . . . . . . . . . . . . . 37
9.1 Serial Communications . . . . . . . . . . . . . . . . . . . . . . . 37
9.1.1 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
9.1.2 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
9.1.2.1 BAUD Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
9.1.2.2 Sample Period. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
9.1.2.3 Duplex Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
9.1.2.4 Linefeed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
9.1.3 Serial Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
9.2 Interface Commands . . . . . . . . . . . . . . . . . . . . . . . . 39
10 Test Probe Calibration . . . . . . . . . . . . . . . . . . . . . . 43
10.1 Calibrating a Single Probe. . . . . . . . . . . . . . . . . . . . . . 43
10.2 Furnace Characteristics . . . . . . . . . . . . . . . . . . . . . . . 43
ii
10.2.1 Stabilization and Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
11 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . 45
11.1 Calibration Points . . . . . . . . . . . . . . . . . . . . . . . . . . 45
11.2 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . 45
12 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
13 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . 49
13.1 Troubleshooting Problems, Possible Causes, and Solutions . . . . 49
13.2 Comments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
13.2.1 EMC Directive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
13.2.2 Low Voltage Directive (Safety) . . . . . . . . . . . . . . . . . . . . . . . . . 50
iii
1 Before You Start
1.1 Symbols Used
Table 1 lists the International Electrical Symbols. Some or all of these symbols may be used on the instrument or in this manual.
Tabl e 1 International Electrical Symbols
Symbol Description
AC (Alternating Current)
AC-DC
Battery
CE Complies with European Union Directives
1 Before You Start
Symbols Used
DC
Double Insulated
Electric Shock
Fuse
PE Ground
Hot Surface (Burn Hazard)
Read the User’s Manual (Important Information)
Off
On
1
9150 Portable Furnace
User’s Guide
Symbol Description
Canadian Standards Association
OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 re fers to the level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from the fixed installation. Examples include household, office, and laboratory appliances.
C-TIC Australian EMC Mark
The European Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/EC) mark.
1.2 Safety Information
Use this instrument only as specified in this manual. Otherwise, the protection provided by the instrument may be impaired.
The following definitions apply to the terms “Warning” and “Caution”.
“Warning” identifies conditions and actions that may pose hazards to the user.
“Caution” identifies conditions and actions that may damage the instru­ment being used.
1.2.1
WARNINGS
To avoid personal injury, follow these guidelines.
BURN HAZARD - DO NOT touch the well access surface of the unit.
The temperature of the well access is the same as the actual temperature shown on the display. If the unit is set at 600°C and the display reads 600°C, the well is at 600°C.
The top sheet metal of the furnace may exhibit extreme temperatures for areas close to the well access.
The air over the well can reach temperatures greater than 200°C. Probes should only be inserted and removed from the unit when the unit is set at temperatures less than 200°C.
-
DO NOT turn off the unit at temperatures higher than 100°C. This could create a hazardous situation. Select a set-point less than 100°C and allow the unit to cool before turning it off.
DO NOT remove inserts and insulators at high temperatures. Inserts and insu
-
lators are the same temperature as the display temperature.
2
1 Before You Start
Safety Information
DO NOT operate this unit without a properly grounded, properly polarized power cord.
DO NOT connect this unit to a non-grounded, non-polarized outlet.
HIGH VOLTAGE is used in the operation of this equipment. SEVERE IN
-
JURY OR DEATH may result if personnel fail to observe safety precautions.
Before working inside the equipment, turn the power off and disconnect the power cord.
Always replace the fuse with one of the same rating, voltage, and type.
Overhead clearance is required. DO NOT place this instrument under a cabinet or other structure.
DO NOT use this unit for any application other than calibration work.
DO NOT use this unit in environments other than those listed in the user’s
guide.
DO NOT turn the unit upside down with the inserts in place; the inserts will fall out of the unit.
DO NOT operate near flammable materials.
Use of this instrument at HIGH TEMPERATURES for extended periods of time requires caution.
Completely unattended high temperature operation is not recommended for safety reasons.
Before initial use, after transport, and anytime the furnace has not been ener­gized for more than 10 days, the calibrator must be energized for a dry-out pe­riod of 1 to 2 hours before it can be assumed to meet all of the safety requirements of the IEC1010-1.
Materials used in this furnace may be irritating to skin, eyes, and respiratory tract. Consult the material manufacturer’s MSDS (Material Safety Data Sheet). Proper ventilation and safety precautions must be observed.
Follow all safety guidelines listed in the user’s guide.
Calibration Equipment should only be used by Trained Personnel.
1.2.2
CAUTIONS
To avoid possible damage to the instrument, follow these guidelines.
DO NOT plug the unit into 230 V if the heater switches and fuse holder read 115 V. This action will cause the fuses to blow and may damage the instrument.
Components and heater lifetime can be shortened by continuous high tempera ture operation.
Most probes have handle temperature limits. Be sure that the probe hand tem perature limit is not exceeded in the air above the unit.
-
-
3
9150 Portable Furnace
User’s Guide
Allow for test probe expansion inside the well as the furnace heats.
DO NOT use fluids to clean out the well.
Never introduce foreign material into the probe hole of the insert. Fluids, etc. can leak into the calibrator causing damage.
DO NOT change the values of the calibration constants from the factory set values. The correct setting of these parameters is important to the safety and proper operation of the calibrator.
DO NOT drop or force the probe stems into the well. This type of action can cause a shock to the sensor and affect the calibration.
DO use a ground fault interrupt device.
1.3 Authorized Service Centers
Please contact one of the following authorized Service Centers to coordinate service on your Hart product:
Fluke Corporation, Hart Scientific Division
799 E. Utah Valley Drive
American Fork, UT 84003-9775
USA
Phone: +1.801.763.1600
Telefax: +1.801.763.1010
E-mail: support@hartscientific.com
Fluke Nederland B.V.
Customer Support Services
Science Park Eindhoven 5108
5692 EC Son
NETHERLANDS
Phone: +31-402-675300
Telefax: +31-402-675321
E-mail: ServiceDesk@fluke.nl
Fluke Int'l Corporation
Service Center - Instrimpex
Room 2301 Sciteck Tower
22 Jianguomenwai Dajie
4
1 Before You Start
Authorized Service Centers
Chao Yang District
Beijing 100004, PRC
CHINA
Phone: +86-10-6-512-3436
Telefax: +86-10-6-512-3437
E-mail: xingye.han@fluke.com.cn
Fluke South East Asia Pte Ltd.
Fluke ASEAN Regional Office
Service Center
60 Alexandra Terrace #03-16
The Comtech (Lobby D)
118502
SINGAPORE
Phone: +65 6799-5588
Telefax: +65 6799-5588
E-mail: antng@singa.fluke.com
When contacting these Service Centers for support, please have the following
information available:
Model Number
Serial Number
Voltage
Complete description of the problem
5
2 Introduction
2 Introduction
The Hart Scientific 9150 thermocouple furnace can be used for calibrating ther mocouple and RTD temperature probes. Calibrations may be done over a range of 150°C to 1200°C (302°F to 2192°F). Temperature display of the 9150 is 0.1 degrees below 1000°and 1 degrees above 1000°.
The furnace features:
Rapid heating and cooling
Interchangeable multiple hole probe sleeves
Convenient handle
RS-232 interface
Built in programmable features include:
Temperature scan rate control
Eight set-point memory
Adjustable readout in °C or °F
The temperature is accurately controlled by Hart’s hybrid analog/digital con­troller. The controller uses a Type S thermocouple as a sensor and controls the well temperature with a solid state relay (triac) driven heater. The furnace should be calibrated yearly by trained personnel at Hart Scientific.
The LED front panel continuously shows the current well temperature. The temperature may be easily set with the control buttons to any desired tempera­ture within the specified range. The calibrator’s multiple fault protection de­vices insure user and instrument safety and protection.
The 9150 calibrator furnace was designed for portability, low cost, and ease of operation. Through proper use, the instrument will provide continued accurate calibration of temperature sensors and devices. The user should be familiar with the safety guidelines and operating procedures of the furnace as described in this manual.
-
7
3 Specifications and Environmental Conditions
3 Specifications and Environmental
Conditions
3.1 Specifications
Temperature Range 150–1200°C (302–2192°F)
Display Resolution 0.1° to 999.9°, 1° above 1000°
Stability ±0.5°C
Display Accuracy ±5.0°C
Well Diameter 1.25" (32 mm)
Well Depth 4" (102 mm)
Heating Time 35 minutes to 1200°C
Cooling Time 140 minutes with block
Well to Well Uniformity ±0.5°C to ±1.0°C (insert “C” at 1200°C)
Stabilization 20 minutes
Power 115 VAC (±10%), 10.5 amps [230 VAC (±10%), 5.2 amps] switchable,
50/60 Hz, 1200 W
Size 12.4" H x 8.2" W x 12.4"D (315 x 208 x 315 mm)
Weight 28 lb. (13 kg)
Fault Protection Sensor burnout and short protection, over temperature thermal cutout
Specifications
3.2 Environmental Conditions
Although the instrument has been designed for optimum durability and trou ble-free operation, it must be handled with care. The instrument should not be operated in an excessively dusty or dirty environment. Maintenance and clean ing recommendations can be found in the Maintenance Section of this manual.
The instrument operates safely under the following conditions:
temperature range: 5 - 50°C (41 - 122°F)
ambient relative humidity: 15 - 50%
pressure: 75kPa - 106kPa
mains voltage within ± 10% of nominal
vibrations in the calibration environment should be minimized
altitude less than 2000 meters
-
-
9
9150 Portable Furnace
User’s Guide
3.3 Warranty
Fluke Corporation, Hart Scientific Division (Hart) warrants this product to be free from defects in material and workmanship under normal use and service for a period as stated in our current product catalog from the date of shipment. This warranty extends only to the original purchaser and shall not apply to any product which, in Hart's sole opinion, has been subject to misuse, alteration, abuse or abnormal conditions of operation or handling.
Software is warranted to operate in accordance with its programmed instruc tions on appropriate Hart products. It is not warranted to be error free.
Hart's obligation under this warranty is limited to repair or replacement of a product which is returned to Hart within the warranty period and is determined, upon examination by Hart, to be defective. If Hart determines that the defect or malfunction has been caused by misuse, alteration, abuse or abnormal condi tions or operation or handling, Hart will repair the product and bill the pur chaser for the reasonable cost of repair.
To exercise this warranty, the purchaser must forward the product after calling or writing an Authorized Service Center (see Section 1.3). The service centers assume NO risk for in-transit damage.
THE FOREGOING WARRANTY IS PURCHASER'S SOLE AND EXCLU­SIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EX­PRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OR MERCHANTABILITY, OR FITNESS FOR ANY PARTICULAR PURPOSE OR USE. HART SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL DAM­AGES OR LOSS WHETHER IN CONTRACT, TORT, OR OTHERWISE.
-
-
-
10
4 Safety Guidelines
Operate the instrument in room temperatures between 5-50°C (41-122°F).
Allow sufficient air circulation by leaving at least 6 inches of space be tween the instrument and nearby objects. Overhead clearance needs to al low for safe and easy insertion and removal of probes for calibration.
The furnace is a precision instrument. Although it has been designed for
optimum durability and trouble free operation, it must be handled with care. Always carry the unit in an upright position to prevent the probe sleeves from dropping out. The convenient fold-up handle allows one hand carrying. The instrument should not be operated in excessively wet, oily, dusty, or dirty environments. It is important to keep the well of the instrument clean and clear of any foreign matter. Do not operate near flammable materials.
DO NOT use fluids to clean out the well.
The instrument can generate extreme temperatures. Precautions must be
taken to prevent personal injury or damage to objects. Probes may be ex­tremely hot or cold when removed from the instrument. Cautiously handle probes to prevent personal injury. Always use the special sleeve tongs that are supplied with the calibrator to remove the sleeve. Carefully place probes on a heat/cold resistant surface or rack until they are at room tem­perature. Never place any objects other than the special probe sleeves sup­plied with the calibrator into the well.
Use only a grounded AC mains supply of the appropriate voltage to
power the instrument. Refer to Section 3.1, Specifications, for power de­tails.
Before initial use, after transport, and anytime the furnace has not been energized for more than 10 days, the instrument needs to be energized for a “dry-out” period of 1-2 hours before it can be assumed to meet all of the safety requirements of the IEC 1010-1.
The instrument is equipped with operator accessible fuses. If a fuse blows, it may be due to a power surge or failure of a component. Replace the fuse once. If the fuse blows a second time, it is likely caused by fail ure of a component part. If this occurs, contact Hart Scientific Customer Service. Always replace the fuse with one of the same rating, voltage, and type. Never replace the fuse with one of a higher current rating.
If a mains supply power fluctuation occurs, immediately turn off the in strument. Power bumps from brownouts and blackouts could damage the instrument. Wait until the power has stabilized before re-energizing the instrument.
Air circulated through the unit keeps the chassis cool. DO NOT SHUT OFF THE FURNACE WHILE AT HIGH TEMPERATURES.
4 Safety Guidelines
-
-
-
-
11
5 Quick Start
5.1 Unpacking
5 Quick Start
Unpacking
Unpack the furnace carefully and inspect it for any damage that may have oc curred during shipment. If there is shipping damage, notify the carrier immediately.
Verify that the following components are present:
9150 Furnace
3150, Insert
Insert Insulator
Power Cord
Manual
RS-232 Cable
5.2 Set-Up
Place the calibrator on a flat surface with at least 6 inches of free space around the instrument. Plug the power cord into a grounded mains outlet. Observe that the nominal voltage corresponds to that indicated on the back of the calibrator.
Carefully insert the probe sleeve into the well. (DO NOT drop the sleeve in the well.) Probe sleeve holes should be of the smallest diameter possible while still allowing the probe to slide in and out easily. Sleeves with various hole sizes are available from Hart Scientific. The well must be clear of any foreign objects, dirt and grit before the sleeve is inserted. The sleeve is inserted with the two small tong holes positioned upward.
Turn on the power to the calibrator by toggling the switch on the power entry module. The fan should begin quietly blowing air through the instrument and the controller display should illuminate after 3 seconds. After a brief self test the controller should begin normal operation. If the unit fails to operate please check the power connection.
The display begins to show the well temperature and the well heater starts oper ating to bring the temperature of the well to the set-point temperature.
-
-
5.3 Power
Plug the furnace power cord into a mains outlet of the proper voltage, fre quency, and current capability. Refer to Section 3.1, Specifications, for power details. Turn the furnace on using the rear panel “POWER” switch. The furnace turns on and begins to heat to the previously programmed temperature set-point. The front panel LED display indicates the actual furnace temperature.
-
13
9150 Portable Furnace
User’s Guide
5.4 Setting the Temperature
Section 8 explains in detail how to set the temperature set-point on the calibra tor using the front panel keys. The procedure is summarized here.
(1) Press “SET” twice to access the set-point value.
(2) Press “SET” to move the cursor to the units that need changing.
(3) Press “UP” or “DOWN” to change the set-point value.
(4) Press “SET” until the display exits to store the new set-point.
(5) Press “EXIT” to return to the temperature display without saving the
new set-point value.
When the set-point temperature is changed the controller switches the well heater on or off to raise or lower the temperature. The displayed well tempera ture gradually changes until it reaches the set-point temperature. The well may require 35 minutes to reach the set-point depending on the span. Another 10 to 20 minutes is required to stabilize within ±0.5°C of the set-point. Ultimate sta­bility may take 15 to 20 minutes more of stabilization time.
-
-
14
6 Parts and Controls
The user should become familiar with the furnace back panel, front panel, and constant temperature block assembly.
6.1 Back Panel
6 Parts and Controls
Back Panel
The back panel (Figure 1) features the Power Entry Module (PEM) that con tains the power cord socket, the power switch, and the heater voltage switch, the serial port, and the fan.
Figure 1 Back Panel
-
Power Cord - On the back of the calibrator is the removable power cord inlet that plugs into an IEC grounded socket.
Power Switch - The power switch is located on the power entry module (PEM). The PEM also houses the fuses and the dual voltage selector. The PEM and Heater Voltage Switch allow the unit to be field switchable for 115 VAC (±10%) or 230 VAC (±10%) operation.
Heater Voltage Switch - To be used only when changing the input voltage. (See Section 7.2 for instructions.)
Serial Port - A DB-9 male connector is present for interfacing the calibrator to a computer or terminal with serial RS-232 communications.
15
9150 Portable Furnace
User’s Guide
Fan - The fan inside the calibrator runs continuously when the unit is being op erated to provide cooling for the instrument. Slots at the top and sides of the calibrator are provided for airflow. The area around the calibrator must be kept clear to allow adequate ventilation. The airflow is directed upward and can be
extremely hot.
6.2 Front Panel
The front panel consists of the controller display and the controller keypad as showninFigure2.
Figure 2 Front Panel
-
16
Controller Display - The digital display is an important part of the temperature controller because it not only displays set and actual temperatures but also dis
­plays various calibrator functions, settings, and constants. The display shows temperatures in units according to the selected scale °C or °F.
Controller Keypad - The four button keypad allows easy setting of the set-point temperature. The control buttons (SET, DOWN, UP, and EXIT) are used to set the calibrator temperature set-point, access and set other operating parameters, and access and set calibration parameters.
Setting the control temperature is done directly in degrees of the current scale. It can be set to one-tenth of a degree Celsius or Fahrenheit.
The functions of the buttons are as follows:
SET – Used to display the next parameter in the menu and to store parameters to the displayed value.
DOWN – Used to decrement the displayed value of parameters.
UP –Usedtoincrementthedisplayedvalue.
Constant Temperature Block Assembly
EXIT – Used to exit from a menu. When EXIT is pressed any changes made to the displayed value are ignored.
6 Parts and Controls
6.3 Constant Temperature Block Assembly
The constant temperature block assembly is shown in Figure 3 and consists of removable inserts.
Figure 3 Removable Inserts
6.3.1 Constant Temperature Block
The “Block” is made of aluminum-oxide and provides a relatively constant and accurate temperature environment for the sensor that is to be calibrated. A
1.25-inch diameter well is provided that may be used for sensors of that size or may be sleeved down with various sized multi-hole probe sleeves. Heaters sur round the block assembly and provide even heat to the sensor. A Type S ther mocouple is used to sense and control the temperature of the block. The entire assembly is surrounded by a heat reflector to thermally isolate the chassis and electronics.
6.3.2 Probe Sleeves and Tongs
The calibrator is supplied with a multi-hole aluminum-oxide probe sleeve and insulator sleeve for insertion into the calibrator well and tongs for removing sleeves. Probe sleeves of various hole sizes are available to allow the user’s probe to fit snugly into the well whatever the diameter of the probe.
Insert A and an insulator is provided unless otherwise indicated . The inserts are:
-
-
17
9150 Portable Furnace
User’s Guide
or
Insert A Model 3150-2 (variety block): 1/2”, 1/4”, 3/8”, 3/16”, 1/8”, and
1/16” holes
Insert B Model 3150-3 (comparison block): two each 3/8”, 1/4”, and
3/16” holes
Insert C Model 3150-4 (1/4” comparison block): six 1/4” holes
18
7 General Operation
7.1 Changing Display Units
The 9150 can display temperature in Celsius or Fahrenheit. The temperature units are shipped from the factory set to Celsius. To change to Fahrenheit or back to Celsius there are two ways:
1 - Press the “SET” and “DOWN” simultaneously.
or
1 - Press the “SET” key three times from the temperature display and then “EXIT” to show the units.
Un= C
2 - Press the “UP” or “DOWN” key to change units.
3 - Press “SET” to store changes and then “EXIT” to display the temperature.
7.2 Switching to 230V Operation
7 General Operation
Changing Display Units
The 9150 is switchable from 115 VAC to 230 VAC 50/60 Hz. Switching the voltage can change the calibration, so the unit should be calibrated after
changing the input voltage.
To change from 115 VAC to 230 VAC
1. Unplug the unit.
2. With a small straight slot screwdriver remove the fuse holder from the PEM located on the back of the unit.
3. Replace the two 12A F 250V fuses with two 6.3A F 250V fuses.
4. Replace the fuse holder with the “230V” in the display window.
5. Lay the unit on its side and with a small straight slot screwdriver, move the heater switches located on the bottom of the unit to display “230V”.
Note: If the heater switches and the fuse holder do not all read “230V when complete, the unit will either not heat or only heat at a fraction of its capacity. If not done properly, the unit could become damaged and void the calibration and warranty. Refer to Section 3.1, Specifications, for correct fuse usage.
CAUTION: Do not plug the unit into 230V if the heater switches and fuse
holder read 115V. This will cause the fuses to blow and may damage the instrument.
19
8 Controller Operation
8 Controller Operation
Well Temperature
This section discusses in detail how to operate the furnace temperature control ler using the front control panel. By using the front panel key-switches and LED display the user may monitor the well temperature, adjust the set-point temperature in degrees C or F, monitor the heater output power, adjust the con troller proportional band, and program the probe calibration parameters, operat ing parameters, serial interface configuration, and controller calibration parameters. Operation of the functions and parameters are shown in the flowchart in Figure 4 on page 22. This chart may be copied for reference.
In the following discussion a button with the word SET, UP, DOWN, or EXIT inside indicates the panel button while the dotted box indicates the display reading. Explanation of the button or display reading are to the right of each button or display value.
8.1 Well Temperature
The digital LED display on the front panel allows direct viewing of the actual well temperature. This temperature value is what is normally shown on the dis­play. The units, C or F, of the temperature value are displayed at the right. For example,
962.3 C Well temperature in degrees Celsius
The temperature displayed function may be accessed from any other function by pressing the “EXIT” button.
-
-
-
8.2 Temperature Set-point
The temperature set-point can be set to any value within the range and resolu tion as given in the specifications. Be careful not to exceed the safe upper tem perature limit of any device inserted into the well.
Setting the temperature involves two steps: (1) select the set-point memory and (2) adjust the set-point value.
8.2.1 Programmable Set-points
The controller stores 8 set-point temperatures in memory. The set-points can be quickly recalled to conveniently set the calibrator to a previously programmed temperature set-point.
To set the temperature one must first select the set-point memory. This function is accessed from the temperature display function by pressing “SET”. The number of the set-point memory currently being used is shown at the left on the display followed by the current set-point value.
-
-
21
9150 Portable Furnace
User’s Guide
Figure 4 Controller Operation Flowchart
22
200.0 C Well temperature in degrees Celsius
8 Controller Operation
Temperature Set-point
S
Access set-point memory
1. 200. Set-point memory 1, 200.0°C currently used
To change the set-point memory press “UP” or “DOWN”.
5. 900. New set-point memory 5, 900.0°C
Press “SET” to accept the new selection and access the set-point value.
S
Accept selected set-point memory
8.2.2 Set-point Value
The set-point value may be adjusted after selecting the set-point memory and pressing “SET”.
0900.0 Set-point value in °C
The first digit flashes on and off. If the set-point value is correct, press “EXIT” to display the temperature scale units.
If the first digit is correct press “SET”.
To adjust the first digit, press “UP” or “DOWN”. Then, press “SET” to go to the second digit. The second digit of the temperature should now be flashing. Adjust this digit by pressing “UP” or “DOWN”.
Press “SET” to accept the second digit and repeat until the last digit has been adjusted.
0962.7 New set-point value
Press “SET” to accept the new set-point. If “EXIT” is pressed all changes made to the set-point are discarded.
S
Accept new set-point value
8.2.3 Temperature Scale Units
Temperature Scale Units of the controller are set by the user to degrees Celsius (°C) or Fahrenheit (°F). The units are used in displaying the well temperature, set-point, and proportional band.
Press “SET” after adjusting the set-point value to change display units.
Un= C
23
9150 Portable Furnace
User’s Guide
Press “UP” or “DOWN” to change the units.
Press “EXIT” to display the well temperature or press “SET” to access the scan control.
8.3 Scan
The scan rate can be set and enabled so that when the set-point is changed the furnace heats or cools at a specified rate (degrees per minute) until it reaches the new set-point. With the scan disabled the furnace heats or cools at the maxi mum possible rate.
8.3.1 Scan Control
The scan is controlled with the scan on/off function that appears in the main menu after the temperature scale units.
Scale units currently selected
Un= F New units selected
-
Sc=OFF Scan function off
Press “UP” or “DOWN” to toggle the scan on or off.
Sc=On Scan function on
Press “SET” to accept the present setting and access the scan rate.
S
Accept scan setting
8.3.2 Scan Rate
The next function in the main menu is the scan rate. The scan rate can be set from .1 to 99.9°C/minute. The maximum scan rate however is actually limited by the natural heating or cooling rate of the instrument. This rate is often less than 100°C/minute, especially when cooling.
The scan rate function appears in the main menu after the scan control function. The scan rate units are in degrees per minute, degrees C or F depending on the selected units.
Sr=10.0 Scan rate in °C/min.
Press“UP”or“DOWN”tochangethescanrate.
Sr= 5.8 New scan rate
24
Press “SET” to accept the new scan rate and continue.
8 Controller Operation
Ramp and Soak Program
S
Accept scan rate
8.4 Ramp and Soak Program
The ramp and soak program feature for the 9150 allows the user to program a number of set-points, cycle the furnace automatically between the temperatures at a scan rate set by the user, and hold the furnace at each temperature for a pe riod of time set by the user. The user can select one of four different cycle func tions. The Ramp and Soak Menu is accessed by pressing “SET” and “UP” simultaneously.
8.4.1 Program Points
The 9150 contains eight “program points”. Each program point contains a set-point, scan rate, and soak time. When the unit is in program mode the unit heats or cools to the current program set-point at the current program scan rate. Once the program set-point is reached the unit waits for the program soak time before heating or cooling to the next program set-point. To access the Ramp and Soak Program Menu press “SET” and “UP” simultaneously.
962.4 Well temperature
S+U
ProG Ramp and Soak Program
Access Ramp and Soak Program Menu
-
-
Press “SET” to access the number of program points.
8.4.2 Number of Program Points
The first parameter in the program menu is the number of program points to cy cle through. Up to 8 set-points can be used in a ramp and soak program.
Pn= 8 Number of program points to cycle through
Use the “UP” and “DOWN” buttons to change the program points. The valid range is from 2 to 8.
Pn= 4 New number of program points
Press “SET” to continue. Pressing “EXIT” causes any changes made to the pa rameter to be discarded.
S
Accept the new number of program points.
-
-
25
9150 Portable Furnace
User’s Guide
8.4.3 Program Set-Points
The controller allows the user to adjust up to eight program points. These are accessed by pressing “SET” after setting the number of program points as de scribed in Section 8.4.2. Each program point has three associated parameters: the program set-point, the program scan rate, and the program hold (or soak) time. After adjusting the number of program points press “SET”.
Use the “UP” or “DOWN” buttons to select any of the program points. The controller only allows the user to edit program points that are less than or equal to the number of programs points selected as explained in Section 8.4.2. For example, if the user has selected 4 program points program points 5, 6, 7, and 8 cannot be edited.
Press “SET” to edit a program point.
-
SP 1 Program point 1
SP 4 Program point 4
S
The first value to edit is the program set-point.
Edit program point
0962.7 Program set-point value in °C
Use “UP”, “DOWN”, and “SET” to adjust the set-point as each digit flashes.
0970.0 New program set-point value for program point 4
Press “SET” to save the new set-point value or “EXIT” to discard changes.
S
The next value to edit is the program soak time.
Accept the program point set-point
Pt 4 Program point 4 soak time
Press “SET” to edit the program soak time.
S
Edit program point soak time
00001 Current program point soak time
Use “UP”, “DOWN”, and “SET” to adjust the program soak time. This value can be any integer from 0 to 14400. This time is the minutes the program set-point maintains after the temperature of the furnace has settled and before proceeding to the next set-point. Each digit flashes individually to indicate that
26
8 Controller Operation
Ramp and Soak Program
it can be adjusted.
00200 Program point 4 soak time set for 200 minutes
Press “SET” to save the new soak-time value or “EXIT” to discard changes
S
The next value to edit is the program scan rate. This value is ignored if scan is not enabled for the unit (See Section 8.3.1).
Accept the program point soak time
Sr 4 Program point 4 scan rate
Press “SET” to edit the program scan rate.
S
Edit the program point scan rate
10.0 Current program point 4 scan rate
Use “UP” and “DOWN” to adjust the program scan rate.
11.3 New program point 4 scan rate
Press “SET” to save the new scan rate value or “EXIT” to discard changes.
S
After “SET” is pressed the controller advances to the next program point or, if there are no more program points to edit, exits to the Program Function Menu. Repeat the above steps to edit any program point.
Accept the program point scan rate
8.4.4 Program Function Mode
The next parameter is the program function or cycle mode. There are four pos sible modes which determine whether the program scans up (from set-point 1 to n) only or both up and down (from set-point n to 1), and also whether the program stops after one cycle or repeats the cycle indefinitely. The table below shows the action of each of the four program mode settings.
-
Function Action
1 up-stop
2 up-down-stop
3 up-repeat
4 up-down-repeat
Pf=1
27
9150 Portable Furnace
User’s Guide
Press “SET” to adjust the program mode and the “UP” or “DOWN” buttons to change the mode.
Program mode
Pf=4
Press “SET” to continue or “EXIT” to continue without saving the new value.
S
Save new setting
New mode
8.4.5 Program Control
The final parameter in the program menu is the control parameter. You may choose between three options to either start the program from the beginning, continue the program from where it was when it was stopped, or stop the program.
P=OFF
Use the “UP” or “DOWN” buttons to change the status.
P=Go
Press “SET” to activate the new program control command and return to the temperature display.
S
Activate new command.
Program presently off
Start cycle from beginning
8.5 Secondary Menu
28
Functions which are used less often are accessed within the secondary menu. The secondary menu is accessed by pressing “SET” and “EXIT” simulta neously and then releasing. The first function in the secondary menu is the heater power display. (See Figure 4 on page 22.)
8.6 Heater Power
The temperature controller controls the temperature of the furnace by pulsing the heater on and off. The total power being applied to the heater is determined by the duty cycle or the ratio of heater on time to the pulse cycle time. By knowing the amount of heating the user can tell if the calibrator is heating up to the set-point, cooling down, or controlling at a constant temperature. Monitor ing the percent heater power lets the user know how stable the well temperature is. With good control stability the percent heating power should not fluctuate more than ±1% within one minute.
The heater power display is accessed in the secondary menu. Press “SET” and
-
-
8 Controller Operation
Set-point Voltage
”EXIT” simultaneously and release. The heater power is displayed as a percent age of full power.
962.4 Well temperature
S+E
Access heater power in secondary menu
100.0 P Heater power in percent
To exit out of the secondary menu press “EXIT”. To continue on to the set-point voltage setting function press “SET”.
8.7 Set-point Voltage
The set-point voltage is displayed for informational purposes and is used to cal ibrate the instrument.
The value of the set-point voltage changes when the set-point temperature is changed and also when DC1 and DC2 are adjusted.
S+E
100.0 P Heater power in percent
S
Access heater power in secondary menu
Access set-point voltage
-
-
11.934U Set-point voltage in millivolts
To exit out of the secondary menu, press “EXIT”. To continue on to the propor tional band setting function press “SET”.
8.8 Proportional Band
In a proportional controller such a this the heater output power is proportional to the well temperature over a limited range of temperatures around the set-point. This range of temperature is called proportional band. At the bottom of the proportional band the heater output is 100%. At the top of the propor tional band the heater output is 0. Thus as the temperature rises the heater power is reduced, which consequently tends to lower the temperature back down. In this way the temperature is maintained at a fairly constant temperature.
The temperature stability of the well and response time depend on the width of the proportional band. If the band is too wide the well temperature deviates ex cessively from the set-point due to varying external conditions. This deviation is because the power output changes very little with temperature and the con
-
-
-
-
29
9150 Portable Furnace
User’s Guide
troller cannot respond very well to changing conditions or noise in the system. If the proportional band is too narrow the temperature may swing back and forth because the controller overreacts to temperature variations. For best con trol stability the proportional band must be set for the optimum width.
The proportional band width is set at the factory to about 30.0°C. The propor tional band width may be altered by the user if he desires to optimize the con trol characteristics for a particular application.
The proportional band width is easily adjusted from the front panel. The width may be set to discrete values in degrees C or F depending on the selected units. The proportional band adjustment can be accessed within the secondary menu. Press “SET” and “EXIT” to enter the secondary menu and show the heater power. Then press “SET” twice to access the proportional band.
-
-
-
S+E
Access heater power in secondary menu
100.0 P Heater power in percent
S
Access set-point voltage
11.934 U Set-point voltage in millivolts
S
Access proportional band
ProP Flashes and then displays the setting
30.0 Proportional band
To change the proportional band, press “UP” or “DOWN.”
30.5 New proportional band setting
To store the new setting press “SET”. Press “EXIT” to continue without storing the new value.
S
Accept the new proportional band setting
30
8.9 Controller Configuration
The controller has a number of configuration and operating options and calibra tion parameters which are programmable via the front panel. These are ac cessed from the secondary menu after the proportional band function by pressing “SET”. Pressing “SET” again enters the first of three groups of config uration parameters—operating parameters, serial interface parameters and cali
-
-
-
-
bration parameters. The groups are selected using the “UP” and “DOWN” keys andthenpressing“SET”.
8.10 Operating Parameters
The operating parameters menu is indicated by:
Par Operating parameters menu
Press “SET” to enter the menu. The operating parameters menu contains the HL (High Limit) parameter, the Soft Cutout parameter, and the Cutout Reset Mode parameter.
8.10.1 High Limit
The HL parameter adjusts the upper set-point temperature. The factory default and maximum are set to 1200. For safety, a user can adjust the HL down so the maximum temperature set-point is restricted.
HL High Limit parameter
Press “SET” to enable adjustment of HL.
8 Controller Operation
Operating Parameters
S
Access High Limit
1200 Flashes the current value and then displays the value for
1200.0 Current HL setting
Adjust the HL parameter digit by digit using “UP”, “DOWN”, and “SET” as each digit flashes.
1005.9 New HL setting
Press “SET” to accept the new temperature limit.
8.10.2 Soft Cutout
The Soft Cutout parameter is used by the controller to shut the unit down dur ing over-temperature conditions.
SoFtCo Soft Cutout parameter
Press “SET” to enable adjustments of the Soft Cutout.
S
Access Soft Cutout
adjustment
-
31
9150 Portable Furnace
User’s Guide
Adjust this parameter by using “UP”, “DOWN”, and “SET” as each digit flashes.
Press “SET” to accept the new temperature limit.
If the temperature of the unit is ever greater than the Soft Cutout temperature the controller shuts itself down and displays, alternately, “SCtOut”and“Err 8”.
8.10.3 Cutout Reset Mode
The cutout reset mode determines whether the cutout resets automatically when the well temperature drops to a safe value or must be manually reset by the operator.
The parameter is indicated by,
1225 Flashes the current value and then displays the value for
adjustment
1225.0 Current Soft Cutout setting
1200.0 New Soft Cutout setting
32
CtorSt Cutout reset mode parameter
Press “SET” to access the parameter setting. Normally the cutout is set for au­tomatic mode.
Auto Cutout set for automatic reset
To change to manual reset mode press “UP” or “DOWN” and then “SET”.
rSt Cutout set for manual reset
8.11 Serial Interface Parameters
The serial RS-232 interface parameters menu is indicated by,
SEriAL Serial RS-232 interface parameters menu
The serial interface parameters menu contains parameters which determine the operation of the serial interface. The parameters in the menu are: BAUD rate, sample period, duplex mode, and linefeed.
8.11.1 BAUD Rate
8 Controller Operation
Serial Interface Parameters
The BAUD rate is the first parameter in the menu. The BAUD rate setting de termines the serial communications transmission rate.
The BAUD rate parameter is indicated by,
bAUd Serial BAUD rate parameter
Press “SET” to choose to set the BAUD rate. The current BAUD rate value is then be displayed.
2400 b Current BAUD rate
The BAUD rate of the serial communications may be programmed to 300 600, 1200, 2400, 4800, or 9600 BAUD. 2400 BAUD is the default setting. Use “UP” or “DOWN” to change the BAUD rate value.
4800 b New BAUD rate
Press “SET” to set the BAUD rate to the new value or “EXIT” to abort the op­eration and skip to the next parameter in the menu.
8.11.2 Sample Period
The sample period is the next parameter in the serial interface parameter menu. The sample period is the time period in seconds between temperature measure­ments transmitted from the serial interface. If the sample rate is set to 5, the in­strument transmits the current measurement over the serial interface approximately every five seconds. The automatic sampling is disabled with a sample period of 0. The sample period is indicated by,
-
SPEr Serial sample period parameter
Press “SET” to choose to set the sample period. The current sample period value is displayed.
SP=1 Current sample period (seconds)
Adjust the value with “UP” or “DOWN” and then use “SET” to set the sample rate to the displayed value.
SP=60 New sample period
8.11.3 Duplex Mode
The next parameter is the duplex mode. The duplex mode may be set to full du plex or half duplex. With full duplex any commands received by the calibrator
-
33
9150 Portable Furnace
User’s Guide
via the serial interface are immediately echoed or transmitted back to the device of origin. With half duplex the commands are executed but not echoed. The du plex mode parameter is indicated by,
Press “SET” to access the mode setting
The mode may be changed using “UP” or DOWN” and pressing “SET”.
8.11.4 Linefeed
-
dUPL Serial duplex mode parameter
d=FULL Current duplex mode setting
d=HALF New duplex mode setting
The final parameter in the serial interface menu is the linefeed mode. This pa rameter enables (on) or disables (off) transmission of a linefeed character (LF, ASCII 10) after transmission of any carriage-return. The linefeed parameter is indicated by,
LF Serial linefeed parameter
Press “SET” to access the linefeed parameter.
LF=On Current linefeed setting
The mode may be changed using “UP” or “DOWN” and pressing “SET”.
LF=OFF New linefeed setting
8.12 Calibration Parameters
The operator of the 9150 has access to the furnace calibration constants. These values are set at the factory and must not be altered. The correct values are im portant to the accuracy and proper and safe operation of the furnace. Access to these parameters is available to the user only so that in the event the controller memory fails, the user may restore these values to the factory settings. The user should have a list of these constants and their settings with the manual.
CAUTION: DO NOT change the values of the furnace calibration con
stants from the factory set values. The correct settings of these parameters is important to the safety and proper operation of the furnace.
-
-
-
34
The calibration parameters menu is indicated by,
CAL Calibration parameters menu
Press “SET” five times to enter the menu. The calibration parameters menu contains the parameters Hard Cutout, CT1, CE1, CT2, CE2, CT3, and CE3.
8.12.1 Hard Cutout
8 Controller Operation
Calibration Parameters
This parameter is the temperature above which the unit shuts down automati cally. The parameter is set at the factory to approximately 1260°C and cannot be changed by the user.
8.12.2 CT1, CT2, and CT3
The calibration parameters CT1, CT2, and CT3 are the calibration temperatures.
8.12.3 CE1, CE2, and CE3
The calibration parameters CE1, CE2 and CE3 are the calibration errors corre­sponding to the calibration temperatures.
-
35
9 Digital Communication Interface
9 Digital Communication Interface
The furnace is capable of communicatingwithandbeingcontrolledbyother equipment through the digital serial interface.
With a digital interface the instrument may be connected to a computer or other equipment. This allows the user to set the set-point temperature, monitor the temperature, and access any of the other controller functions, all using remote communications equipment. Communications commands are summarized in Table 2 on page 40.
9.1 Serial Communications
Serial Communications
The calibrator is installed with an RS-232 serial interface that allows serial dig ital communications over fairly long distances. With the serial interface the user may access any of the functions, parameters and settings discussed in Section 8 with the exception of the BAUD rate setting.
9.1.1 Wiring
The serial communications ca­ble attaches to the calibrator through the DB-9 connector at the back of the instrument. Fig­ure 5 shows the pin-out of this connector and suggested cable wiring. The serial cable should be shielded. If the unit is used in a heavy industrial setting, the serial cable must be limited to ONE meter.
9.1.2 Setup
Before operation the serial in terface must first be set up by programming the BAUD rate and other configuration param eters. These parameters are programmed within the serial interface menu. The serial in terface parameters menu is out lined in Figure 4 on page 22.
To enter the serial parameter programming mode first press “EXIT” while pressing “SET” and release to enter the second
-
-
-
-
-
Figure 5 Serial Cable Wiring Diagram
-
37
9150 Portable Furnace
User’s Guide
ary menu. Press “SET” repeatedly until the display reads “CAL”. Press “UP” until the serial interface menu is indicated with “SErIAL”. Finally press “SET” to enter the serial parameter menu. In the serial interface parameters menu are the BAUD rate, the sample rate, the duplex mode, and the linefeed parameter.
9.1.2.1 BAUD Rate
The BAUD rate is the first parameter in the menu. The display prompts with the BAUD rate parameter by showing “bAUd”. Press “SET” to choose to set the BAUD rate. The current BAUD rate value is displayed. The BAUD rate of the 9150 serial communications may be programmed to 300, 600, 1200, 2400, 4800, or 9600 baud. The BAUD rate is preprogrammed to 2400 BAUD. Use “UP” or “DOWN” to change the BAUD rate value. Press “SET” to set the BAUD rate to the new value or “EXIT” to abort the operation and skip to the next parameter in the menu.
9.1.2.2 Sample Period
The sample period is the next parameter in the menu and prompted with “SPEr”. The sample period is the time period in seconds between temperature measurements transmitted from the serial interface. If the sample rate is set to 5, the instrument transmits the current measurement over the serial interface ap­proximately every five seconds. The automatic sampling is disabled with a sample period of 0. Press “SET” to choose to set the sample period. Adjust the period with “UP” or “DOWN” and then use “SET” to set the sample rate to the displayed value.
38
9.1.2.3 Duplex Mode
The next parameter is the duplex mode indicated with “dUPL”. The duplex mode may be set to half duplex (“HALF”) or full duplex (“FULL”). With full duplex any commands received by the thermometer via the serial interface are immediately echoed or transmitted back to the device of origin. With half du plex the commands are executed but not echoed. The default setting is full du plex. The mode may be changed using “UP” or “DOWN” and pressing “SET”.
9.1.2.4 Linefeed
The final parameter in the serial interface menu is the linefeed mode. This pa rameter enables (“On”) or disables (“OFF”) transmission of a linefeed charac ter (LF, ASCII 10) after transmission of any carriage-return. The default setting is with linefeed on. The mode may be changed using “UP” or “DOWN” and pressing “SET”.
9.1.3 Serial Operation
Once the cable has been attached and the interface set up properly, the control ler immediately begins transmitting temperature readings at the programmed rate. The serial communications uses 8 data bits, one stop bit, and no parity. The set-point and other commands may be sent via the serial interface to set the
-
-
-
-
-
9 Digital Communication Interface
Interface Commands
temperature set-point and view or program the various parameters. The inter face commands are discussed in Section 9.2. All commands are ASCII charac ter strings terminated with a carriage-return character (CR, ASCII 13).
9.2 Interface Commands
The various commands for accessing the calibrator functions via the digital in terfaces are listed in this section (see Table 2). These commands are used with the RS-232 serial interface. The commands are terminated with a carriage-re turn character. The interface makes no distinction between upper and lower case letters, hence either may be used. Commands may be abbreviated to the minimum number of letters which determines a unique command. A command may be used to either set a parameter or display a parameter depending on whether or not a value is sent with the command following a “=” character. For example “s”<CR> returns the current set-point and “s=150.0”<CR> sets the set-point to 150.0 degrees.
In the following list of commands, characters or data within brackets, “[” and “]”, are optional for the command. A slash, “/”, denotes alternate characters or data. Numeric data, denoted by “n”, may be entered in decimal or exponential notation. Characters are shown in lower case although upper case may be used. Spaces may be added within command strings and are ignored. Backspace (BS, ASCII 8) may be used to erase the previous character. A terminating CR is im­plied with all commands.
-
-
-
-
39
9150 Portable Furnace
User’s Guide
Tabl e 2 Communications Command Summary
Command Description
Display Temperature
Read current set-point s[etpoint] s set: 9999.99 {C or F} set: 150.0 C
Set current set-point to n s[etpoint]=n s=450 Instrument
Set temperature units: u[nits]=c/f
Set temperature units to Celsius u[nits]=c u=c
Set temperature units to Fahrenheit
Read scan function sc[an] sc scan: {ON or OFF} scan: ON
Set scan function: sc[an]=on/of[f]
Turn scan function on sc[an]=on sc=on
Turn scan function off sc[an]=of[f] sc=of
Read scan rate sr[ate] sr srat: 999.99 {C or F}/min srat: 10.0 C/min
Setscanrateto minute
Secondary Menu
Read proportional band setting pr[op-band] pr pb: 999.9 pb: 15.9
Set proportional band to
Read heater power
(duty cycle)
Ramp and Soak Menu
Read number of programmable set-points
Set number of programmable set-points to
Read programmable set-point number
Set programmable set-point num ber
nton
Read program set-point soak time
Set program set-point soak time to
n
minutes
Read program scan rate px
Set program scan rate px
Read program control mode pc pc prog: {OFF or ON} prog: OFF
Set program control mode: pc=g[o]/s[top]/c[ont]
Start program pc=g[o] pc=g
n
degrees per
n
n
n
Command Format
u[nits]=f u=f
sr[ate]=
n
pr[op-band]=n pr=8.83 0.1 to 100
po[wer] po po: 999.9 po: 1
pn pn pn: 9 pn: 2
pn=
n
ps
n
-
psn=
n
pt
n
n=n
pt
n
n=n
Command Example Returned
sr=5 .1 to 99.9
pn=4 1 to 8
ps3 psn: 9999.99 {C or F} ps1: 50.00 C
ps3=50 1 to 8, Instru
pt3 tin: 999 ti1: 5
pt3=5 0 to 14400
px3 srn: 99.9 sr3: 11.3
px3=10 .1 to 99.9
Returned Example
Acceptable Values
Range
CorF
ON or OFF
ment Range
GO or STOP or CONT
-
40
Tabl e 3 Communications Commands Summary continued
9 Digital Communication Interface
Interface Commands
Command Description
Stop program pc=s[top] pc=s
Continue program pc=c[ont] pc=c
Read program function pf pf pf: 9 pf: 3
Set program function to
Configuration Menu
Operating Parameters Menu
Read high limit hl hl hl: 9999 hl: 925
Set high limit hl=n hl=900 100–1200
Read soft cutout cuto cuto c: 9999.9 c: 1150.0
Set soft cutout setting: cu[tout]=n
Set soft cut-out to
Read cut-out mode cm[ode] cm cm:{xxxx} cm: AUTO
Set cut-out mode cm[ode]=r[eset]/a[uto] Reset or Auto
Set cut-out to be reset manually cm[ode]=r[eset] cm=r
Set cut-out to be reset automatically
Serial Interface Menu
Read serial sample setting sa[mple] sa sa: 9 sa: 1
Set serial sampling setting to seconds
Set serial duplex mode: du[plex]=f[ull]/h[alf]
Set serial duplex mode to full du[plex]=f[ull] du=f
Set serial duplex mode to half du[plex]=h[alf] du=h
Set serial linefeed mode: lf[eed]=on/of[f]
Set serial linefeed mode to on lf[eed]=on lf=on
Set serial linefeed mode to off lf[eed]=of[f] lf=of
Cal Menu
Read CT
n
calibration parameter ct
Set CT
n
calibration parameter tonctn=
n
n
degrees cu[tout]=n cu=500 0.0 to 1150.0
Command Format
pf=
n
cm[ode]=a[uto] cm=a
n
sa[mple]=
n
n
n
Command Example Returned
pf=2 1to4
sa=0 0 to 4000
ct1 ctn: 99.9C ct1: –10.0C
ct1=–10.0 0 to 12.00
Returned Example
Acceptable Values
FULL or HALF
ON or OFF
Read CE
n
calibration parameter ce
Set CE
n
calibration parameter toncen=
n
n
ce1 cen: 99.9C ce1: -10.1C
ce1=–10 –99.9 to 99.9
41
9150 Portable Furnace
User’s Guide
Tabl e 4 Communications Commands Summary continued
Command Description
These commands are only used for factory testing.
Miscellaneous (not on menus)
Read firmware version number *ver[sion] *ver ver.9999,9.99 ver.9150,2.20
Read structure of all commands h[elp] h list of commands
Legend: [] Optional Command data
Note: When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed by a
Command Format
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
carriage return and linefeed. Then the value is returned as indicated in the RETURNED column.
Command Example Returned
Returned Example
Acceptable Values
42
10 Test Probe Calibration
For optimum accuracy and stability, allow the calibrator to warm up for 10 minutes after power-up and then allow adequate stabilization time after reach ing the set-point temperature. After completing operation of the calibrator, al low the well to cool by setting the temperature to 150°C or less for one-half hour before switching the power off.
10.1 Calibrating a Single Probe
Insert the probe to be calibrated into the well of the furnace calibrator. The probe should fit snugly into the calibrator probe sleeve yet should not be so tight that it cannot be easily removed. Avoid any dirt or grit that may cause the probe to jam into the sleeve. Best results are obtained with the probe inserted to the full depth of the well. Once the probe is inserted into the well, allow ade quate stabilization time to allow the test probe temperature to settle as de scribed above. Once the probe has settled to the temperature of the well, it may be compared to the calibrator display temperature. The display temperature should be stable to within 1°C degree for best results.
CAUTION: Never introduce any foreign material into the probe hole of
the insert. Fluids etc. can leak into the calibrator causing damage to the calibrator or binding and damage to your probe.
10 Test Probe Calibration
Calibrating a Single Probe
-
-
-
-
10.2 Furnace Characteristics
There is a temperature gradient vertically in the test well. The heater has been applied to the block in such a way as to compensate for nominal heat losses out of the top of the furnace. However, actual heat losses will vary with design of the thermometer probes inserted into the calibrator and the temperature. For best results, insert probe to full depth of well.
CAUTION: Do not remove inserts and insulators at high temperatures.
Inserts and insulators are the same temperature as the display tempera ture. Use extreme care when removing hot inserts and insulators.
10.2.1 Stabilization and Accuracy
The stabilization time of the calibrator depends on the conditions and tempera tures involved. Typically the test well is stable to 0.5°C within 20 minutes of reaching the set-point temperature as indicated by the display. Ultimate stability is achieved 10 to 20 minutes after reaching the set temperature.
Inserting a cold probe into a well requires another period of stabilizing depend ing on the magnitude of the disturbance and the required accuracy. For exam ple, inserting a .25 inch diameter room temperature probe into a sleeve at
-
-
-
-
43
9150 Portable Furnace
User’s Guide
1200°C takes 10 minutes to be within 0.5°C of its settled point and takes 15 minutes to achieve maximum stability.
Speeding up the calibration process can be accomplished by knowing how soon to make the measurement. It is recommended that typical measurements be made at the desired temperatures with the desired test probes to establish these times.
44
11 Calibration Procedure
At times the user may want to calibrate the unit to improve the temperature set-point accuracy. Calibration is done by adjusting the controller probe cali bration constants CE1, CE2, and CE3 so that the temperature of the unit as measured with a standard thermocouple agrees more closely with the set-point. The thermometer used must be able to measure the well temperature with higher accuracy than the desired accuracy of the unit.
11.1 Calibration Points
In calibrating the unit, CE1, CE2, and CE3 are adjusted to minimize the set-point error at each of three different well temperatures. Any three reason ably separated temperatures may be used for the calibration. Improved results can be obtained for shorter ranges when using temperatures that are just within the most useful operating range of the unit. The farther apart the calibration temperatures, the larger the calibration range but the calibration error is also greater over the range. Choosing a range of 150°C to 500°C may allow the cali­brator to have a better accuracy of maybe ±2.0°C but outside that range the ac­curacy may be greater than ±10.0°C.
11 Calibration Procedure
Calibration Points
-
-
11.2 Calibration Procedure
1. Choose three set-points to use in the calibration of CE1, CE2, and CE3 parameters. These set-points are generally CT1 = 150°C, CT2 = 675°C and CT3 = 1200°C but other set-points may be used if desired or neces­sary. Using these three temperature set-points may result in ±4.0°C accuracy.
2. If the normal set-points are not used, initialize CT1, CT2, and CT3 to the desired set points. Where CT1 is the low-set point and CT3 is the high set-point.
3. Set the unit to the low set-point. When the unit reaches the set-point and the display is stable, wait 15 minutes or so and take a reading from the thermometer. Repeat step 3 for the other two set-points recording them as Tm1, Tm2, and Tm3.
4. Retrieve the original calibration errors from the unit or the Report of Calibration.
5. Record the previous values for CE1, CE2, and CE3 and calculate new values for CE1, CE2, and CE3 using the following formula.
Tmn –Tsn +CEn = CEm
Where:
Tmn is the temperature measured, Tsn (CTn) is the set-point tempera
-
45
9150 Portable Furnace
User’s Guide
6. Enter the new CEm value in the calibration parameter menu using either
ture, CEn is the old value for calibration error, and CEm is the new value for calibration error
the keypad or through the serial port.
46
12 Maintenance
The calibration instrument has been designed with the utmost care. Ease
of operation and simplicity of maintenance have been a central theme in the product development. Therefore, with proper care the instrument should require very little maintenance. Avoid operating the instrument in an oily, wet, dirty, or dusty environment.
If the outside of the instrument becomes soiled, it may be wiped clean
with a damp cloth and mild detergent. Do not use harsh chemicals on the surface which may damage the paint.
Be sure that the well of the furnace is kept clean and clear of any foreign
matter. DO NOT use fluids to clean out the well.
If a hazardous material is split on or inside the equipment, the user is re
sponsible for taking the appropriate decontamination steps as outlined by the national safety council with respect to the material.
If the mains supply cord becomes damaged, replace it with a cord of the appropriate gauge wire for the current of the instrument. If there are any questions, call Hart Scientific Customer Service for more information.
Before using any cleaning or decontamination method except those rec­ommended by Hart, users should check with Hart Scientific Customer Service to be sure that the proposed method will not damage the equip­ment.
If the instrument is used in a manner not in accordance with the equip ­ment design, the operation of the furnace may be impaired or safety haz­ards may arise.
The over-temperature cutout should be checked every 6 months to see that it is working properly.
12 Maintenance
-
47
13 Troubleshooting
13 Troubleshooting
This section contains information on troubleshooting, CE Comments, and a wiring diagram.
13.1 Troubleshooting Problems, Possible Causes, and Solutions
In the event that the instrument appears to function abnormally, this section may help to find and solve the problem. Several possible problem conditions are described along with likely causes and solutions. If a problem arises, please read this section carefully and attempt to understand and solve the problem. If the problem cannot otherwise be solved, contact Hart Scientific Customer Ser vice for assistance (1-801-763-1600). Be sure to have the model number and serial number of your instrument available.
Problem Possible Causes and Solutions
Incorrect temperature reading
Incorrect calibration parameters. Find the values for CT1, CT2, CT3, CE1, CE2, and CE3 on the Report of Calibration that was shipped with the instrument (or from subsequent calibrations of the instrument). Reprogram the parameters into the Model 9150 memory (see Section 8.12, Calibration Parameters). Allow the instrument to stabilize and verify the accuracy of the temperature reading.
-
The instrument heats or cools too quickly or too slowly
Memory scrambled. The memory may be scrambled due to a power surge or other aberration. Initialize the system by performing the Factory Reset Sequence.
Factory Reset Sequence. Hold the SET and EXIT buttons down at the same time while powering up the instrument. After the instrument displays lease the buttons. The display shows version. After performing the master reset sequence, all of the configuration pa rameters are reset to their default values. Reprogram the calibration parameters into the Model 9150 memory (see Section 8.12, Calibration Parameters) and any other applicable configuration parameters. Allow the instrument to stabilize and verify the accuracy of the temperature reading.
Incorrect scan and scan rate settings. The scan and scan rate settings may be set to unwanted values. Check the Scan and Scan Rate settings. The scan may be off (if the unit seems to be responding too quickly). The scan may be on with the Scan Rate set low (if unit seems to be responding too slowly).
Incorrect heater switch polarity. The PEM and the heater voltage switch may not be set for the same voltage. Check both settings and correct as needed.
9150
, and then displays the firmware
-init-
,re
49
-
-
9150 Portable Furnace
User’s Guide
Problem Possible Causes and Solutions
The display shows any of the following:
err 2,err 3,err 4 Err 7
Temperature cannot be set above a certain point
The display flashes
Cut-out
13.2 Comments
13.2.1 EMC Directive
Equipment manufactured by Hart Scientific has been tested to meet the Euro­pean Electromagnetic Compatibility Directive (EMC Directive, 89/336/EEC). The Declaration of Conformity for your instrument lists the specific standards to which the unit was tested.
err 1
,
Err 5,Err 6
,or
Err 8
Controller problem. The error messages signify the following problems with the controller.
,
Err 1
,
- a RAM error
Err 2
- a NVRAM error
Err 3
- a RAM error
Err 4
- an ADC setup error
Err 5
- an ADC ready error
Err 6
– a Sensor error
Err 7
– a Heater control error
Err 8
– a Soft cutout error
Initialize the system by performing the Factory Reset Sequence describe above.
Incorrect High Limit parameter. The High Limit parameter may be set below 1200°C. Check this value as described in Section 8.10, Operating Parameters.
Incorrect Soft Cutout parameter. Wait for the instrument to cool and adjust the Soft Cutout parameter as described in Section 8.10, Operating Parameters.
50
13.2.2 Low Voltage Directive (Safety)
In order to comply with the European Low Voltage Directive (73/23/EEC), Hart Scientific equipment has been designed to meet the IEC 1010-1 (EN 61010-1) and the IEC 1010-2-010 (EN 61010-2-010) standards.
Loading...