Fluke 9150 User Manual

Hart Scientific
9150 Portable
Furnace
User’s Guide
Rev. 5B1601
Fluke Corporation, Hart Scientific Division
799 E. Utah Valley Drive • American Fork, UT 84003-9775 • USA Phone: +1.801.763.1600 • Telefax: +1.801.763.1010 E-mail: support@hartscientific.com
www.hartscientific.com
Subject to change without notice. • Copyright © 2005 • Printed in USA
Rev. 5B1601
Table of Contents
1 Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.1 Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2.1 WARNINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2.2 CAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3 Authorized Service Centers. . . . . . . . . . . . . . . . . . . . . . 4
2 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3 Specifications and Environmental Conditions . . . . . . . . . . 9
3.1 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.2 Environmental Conditions . . . . . . . . . . . . . . . . . . . . . . 9
3.3 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4 Safety Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . 11
5 Quick Start . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.2 Set-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.3 Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.4 Setting the Temperature . . . . . . . . . . . . . . . . . . . . . . . 14
6 Parts and Controls . . . . . . . . . . . . . . . . . . . . . . . . 15
6.1 Back Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
6.2 Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
6.3 Constant Temperature Block Assembly . . . . . . . . . . . . . . . 17
6.3.1 Constant Temperature Block . . . . . . . . . . . . . . . . . . . . . . . . . . 17
6.3.2 Probe Sleeves and Tongs . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
7 General Operation . . . . . . . . . . . . . . . . . . . . . . . . 19
7.1 Changing Display Units . . . . . . . . . . . . . . . . . . . . . . . 19
7.2 Switching to 230V Operation . . . . . . . . . . . . . . . . . . . . 19
8 Controller Operation . . . . . . . . . . . . . . . . . . . . . . . 21
8.1 Well Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . 21
8.2 Temperature Set-point . . . . . . . . . . . . . . . . . . . . . . . . 21
i
8.2.1 Programmable Set-points . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
8.2.2 Set-point Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
8.2.3 Temperature Scale Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
8.3 Scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
8.3.1 Scan Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
8.3.2 Scan Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
8.4 Ramp and Soak Program . . . . . . . . . . . . . . . . . . . . . . 25
8.4.1 Program Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
8.4.2 Number of Program Points . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
8.4.3 Program Set-Points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
8.4.4 Program Function Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
8.4.5 Program Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
8.5 Secondary Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . 28
8.6 Heater Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
8.7 Set-point Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . 29
8.8 Proportional Band . . . . . . . . . . . . . . . . . . . . . . . . . . 29
8.9 Controller Configuration . . . . . . . . . . . . . . . . . . . . . . 30
8.10 Operating Parameters . . . . . . . . . . . . . . . . . . . . . . . . 31
8.10.1 High Limit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
8.10.2 Soft Cutout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
8.10.3 Cutout Reset Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
8.11 Serial Interface Parameters . . . . . . . . . . . . . . . . . . . . . 32
8.11.1 BAUD Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
8.11.2 Sample Period. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
8.11.3 Duplex Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
8.11.4 Linefeed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
8.12 Calibration Parameters . . . . . . . . . . . . . . . . . . . . . . . 34
8.12.1 Hard Cutout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
8.12.2 CT1, CT2, and CT3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
8.12.3 CE1, CE2, and CE3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
9 Digital Communication Interface . . . . . . . . . . . . . . . . 37
9.1 Serial Communications . . . . . . . . . . . . . . . . . . . . . . . 37
9.1.1 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
9.1.2 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
9.1.2.1 BAUD Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
9.1.2.2 Sample Period. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
9.1.2.3 Duplex Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
9.1.2.4 Linefeed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
9.1.3 Serial Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
9.2 Interface Commands . . . . . . . . . . . . . . . . . . . . . . . . 39
10 Test Probe Calibration . . . . . . . . . . . . . . . . . . . . . . 43
10.1 Calibrating a Single Probe. . . . . . . . . . . . . . . . . . . . . . 43
10.2 Furnace Characteristics . . . . . . . . . . . . . . . . . . . . . . . 43
ii
10.2.1 Stabilization and Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
11 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . 45
11.1 Calibration Points . . . . . . . . . . . . . . . . . . . . . . . . . . 45
11.2 Calibration Procedure . . . . . . . . . . . . . . . . . . . . . . . . 45
12 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
13 Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . 49
13.1 Troubleshooting Problems, Possible Causes, and Solutions . . . . 49
13.2 Comments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
13.2.1 EMC Directive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
13.2.2 Low Voltage Directive (Safety) . . . . . . . . . . . . . . . . . . . . . . . . . 50
iii
1 Before You Start
1.1 Symbols Used
Table 1 lists the International Electrical Symbols. Some or all of these symbols may be used on the instrument or in this manual.
Tabl e 1 International Electrical Symbols
Symbol Description
AC (Alternating Current)
AC-DC
Battery
CE Complies with European Union Directives
1 Before You Start
Symbols Used
DC
Double Insulated
Electric Shock
Fuse
PE Ground
Hot Surface (Burn Hazard)
Read the User’s Manual (Important Information)
Off
On
1
9150 Portable Furnace
User’s Guide
Symbol Description
Canadian Standards Association
OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 re fers to the level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from the fixed installation. Examples include household, office, and laboratory appliances.
C-TIC Australian EMC Mark
The European Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/EC) mark.
1.2 Safety Information
Use this instrument only as specified in this manual. Otherwise, the protection provided by the instrument may be impaired.
The following definitions apply to the terms “Warning” and “Caution”.
“Warning” identifies conditions and actions that may pose hazards to the user.
“Caution” identifies conditions and actions that may damage the instru­ment being used.
1.2.1
WARNINGS
To avoid personal injury, follow these guidelines.
BURN HAZARD - DO NOT touch the well access surface of the unit.
The temperature of the well access is the same as the actual temperature shown on the display. If the unit is set at 600°C and the display reads 600°C, the well is at 600°C.
The top sheet metal of the furnace may exhibit extreme temperatures for areas close to the well access.
The air over the well can reach temperatures greater than 200°C. Probes should only be inserted and removed from the unit when the unit is set at temperatures less than 200°C.
-
DO NOT turn off the unit at temperatures higher than 100°C. This could create a hazardous situation. Select a set-point less than 100°C and allow the unit to cool before turning it off.
DO NOT remove inserts and insulators at high temperatures. Inserts and insu
-
lators are the same temperature as the display temperature.
2
1 Before You Start
Safety Information
DO NOT operate this unit without a properly grounded, properly polarized power cord.
DO NOT connect this unit to a non-grounded, non-polarized outlet.
HIGH VOLTAGE is used in the operation of this equipment. SEVERE IN
-
JURY OR DEATH may result if personnel fail to observe safety precautions.
Before working inside the equipment, turn the power off and disconnect the power cord.
Always replace the fuse with one of the same rating, voltage, and type.
Overhead clearance is required. DO NOT place this instrument under a cabinet or other structure.
DO NOT use this unit for any application other than calibration work.
DO NOT use this unit in environments other than those listed in the user’s
guide.
DO NOT turn the unit upside down with the inserts in place; the inserts will fall out of the unit.
DO NOT operate near flammable materials.
Use of this instrument at HIGH TEMPERATURES for extended periods of time requires caution.
Completely unattended high temperature operation is not recommended for safety reasons.
Before initial use, after transport, and anytime the furnace has not been ener­gized for more than 10 days, the calibrator must be energized for a dry-out pe­riod of 1 to 2 hours before it can be assumed to meet all of the safety requirements of the IEC1010-1.
Materials used in this furnace may be irritating to skin, eyes, and respiratory tract. Consult the material manufacturer’s MSDS (Material Safety Data Sheet). Proper ventilation and safety precautions must be observed.
Follow all safety guidelines listed in the user’s guide.
Calibration Equipment should only be used by Trained Personnel.
1.2.2
CAUTIONS
To avoid possible damage to the instrument, follow these guidelines.
DO NOT plug the unit into 230 V if the heater switches and fuse holder read 115 V. This action will cause the fuses to blow and may damage the instrument.
Components and heater lifetime can be shortened by continuous high tempera ture operation.
Most probes have handle temperature limits. Be sure that the probe hand tem perature limit is not exceeded in the air above the unit.
-
-
3
9150 Portable Furnace
User’s Guide
Allow for test probe expansion inside the well as the furnace heats.
DO NOT use fluids to clean out the well.
Never introduce foreign material into the probe hole of the insert. Fluids, etc. can leak into the calibrator causing damage.
DO NOT change the values of the calibration constants from the factory set values. The correct setting of these parameters is important to the safety and proper operation of the calibrator.
DO NOT drop or force the probe stems into the well. This type of action can cause a shock to the sensor and affect the calibration.
DO use a ground fault interrupt device.
1.3 Authorized Service Centers
Please contact one of the following authorized Service Centers to coordinate service on your Hart product:
Fluke Corporation, Hart Scientific Division
799 E. Utah Valley Drive
American Fork, UT 84003-9775
USA
Phone: +1.801.763.1600
Telefax: +1.801.763.1010
E-mail: support@hartscientific.com
Fluke Nederland B.V.
Customer Support Services
Science Park Eindhoven 5108
5692 EC Son
NETHERLANDS
Phone: +31-402-675300
Telefax: +31-402-675321
E-mail: ServiceDesk@fluke.nl
Fluke Int'l Corporation
Service Center - Instrimpex
Room 2301 Sciteck Tower
22 Jianguomenwai Dajie
4
1 Before You Start
Authorized Service Centers
Chao Yang District
Beijing 100004, PRC
CHINA
Phone: +86-10-6-512-3436
Telefax: +86-10-6-512-3437
E-mail: xingye.han@fluke.com.cn
Fluke South East Asia Pte Ltd.
Fluke ASEAN Regional Office
Service Center
60 Alexandra Terrace #03-16
The Comtech (Lobby D)
118502
SINGAPORE
Phone: +65 6799-5588
Telefax: +65 6799-5588
E-mail: antng@singa.fluke.com
When contacting these Service Centers for support, please have the following
information available:
Model Number
Serial Number
Voltage
Complete description of the problem
5
2 Introduction
2 Introduction
The Hart Scientific 9150 thermocouple furnace can be used for calibrating ther mocouple and RTD temperature probes. Calibrations may be done over a range of 150°C to 1200°C (302°F to 2192°F). Temperature display of the 9150 is 0.1 degrees below 1000°and 1 degrees above 1000°.
The furnace features:
Rapid heating and cooling
Interchangeable multiple hole probe sleeves
Convenient handle
RS-232 interface
Built in programmable features include:
Temperature scan rate control
Eight set-point memory
Adjustable readout in °C or °F
The temperature is accurately controlled by Hart’s hybrid analog/digital con­troller. The controller uses a Type S thermocouple as a sensor and controls the well temperature with a solid state relay (triac) driven heater. The furnace should be calibrated yearly by trained personnel at Hart Scientific.
The LED front panel continuously shows the current well temperature. The temperature may be easily set with the control buttons to any desired tempera­ture within the specified range. The calibrator’s multiple fault protection de­vices insure user and instrument safety and protection.
The 9150 calibrator furnace was designed for portability, low cost, and ease of operation. Through proper use, the instrument will provide continued accurate calibration of temperature sensors and devices. The user should be familiar with the safety guidelines and operating procedures of the furnace as described in this manual.
-
7
3 Specifications and Environmental Conditions
3 Specifications and Environmental
Conditions
3.1 Specifications
Temperature Range 150–1200°C (302–2192°F)
Display Resolution 0.1° to 999.9°, 1° above 1000°
Stability ±0.5°C
Display Accuracy ±5.0°C
Well Diameter 1.25" (32 mm)
Well Depth 4" (102 mm)
Heating Time 35 minutes to 1200°C
Cooling Time 140 minutes with block
Well to Well Uniformity ±0.5°C to ±1.0°C (insert “C” at 1200°C)
Stabilization 20 minutes
Power 115 VAC (±10%), 10.5 amps [230 VAC (±10%), 5.2 amps] switchable,
50/60 Hz, 1200 W
Size 12.4" H x 8.2" W x 12.4"D (315 x 208 x 315 mm)
Weight 28 lb. (13 kg)
Fault Protection Sensor burnout and short protection, over temperature thermal cutout
Specifications
3.2 Environmental Conditions
Although the instrument has been designed for optimum durability and trou ble-free operation, it must be handled with care. The instrument should not be operated in an excessively dusty or dirty environment. Maintenance and clean ing recommendations can be found in the Maintenance Section of this manual.
The instrument operates safely under the following conditions:
temperature range: 5 - 50°C (41 - 122°F)
ambient relative humidity: 15 - 50%
pressure: 75kPa - 106kPa
mains voltage within ± 10% of nominal
vibrations in the calibration environment should be minimized
altitude less than 2000 meters
-
-
9
9150 Portable Furnace
User’s Guide
3.3 Warranty
Fluke Corporation, Hart Scientific Division (Hart) warrants this product to be free from defects in material and workmanship under normal use and service for a period as stated in our current product catalog from the date of shipment. This warranty extends only to the original purchaser and shall not apply to any product which, in Hart's sole opinion, has been subject to misuse, alteration, abuse or abnormal conditions of operation or handling.
Software is warranted to operate in accordance with its programmed instruc tions on appropriate Hart products. It is not warranted to be error free.
Hart's obligation under this warranty is limited to repair or replacement of a product which is returned to Hart within the warranty period and is determined, upon examination by Hart, to be defective. If Hart determines that the defect or malfunction has been caused by misuse, alteration, abuse or abnormal condi tions or operation or handling, Hart will repair the product and bill the pur chaser for the reasonable cost of repair.
To exercise this warranty, the purchaser must forward the product after calling or writing an Authorized Service Center (see Section 1.3). The service centers assume NO risk for in-transit damage.
THE FOREGOING WARRANTY IS PURCHASER'S SOLE AND EXCLU­SIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EX­PRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OR MERCHANTABILITY, OR FITNESS FOR ANY PARTICULAR PURPOSE OR USE. HART SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL DAM­AGES OR LOSS WHETHER IN CONTRACT, TORT, OR OTHERWISE.
-
-
-
10
4 Safety Guidelines
Operate the instrument in room temperatures between 5-50°C (41-122°F).
Allow sufficient air circulation by leaving at least 6 inches of space be tween the instrument and nearby objects. Overhead clearance needs to al low for safe and easy insertion and removal of probes for calibration.
The furnace is a precision instrument. Although it has been designed for
optimum durability and trouble free operation, it must be handled with care. Always carry the unit in an upright position to prevent the probe sleeves from dropping out. The convenient fold-up handle allows one hand carrying. The instrument should not be operated in excessively wet, oily, dusty, or dirty environments. It is important to keep the well of the instrument clean and clear of any foreign matter. Do not operate near flammable materials.
DO NOT use fluids to clean out the well.
The instrument can generate extreme temperatures. Precautions must be
taken to prevent personal injury or damage to objects. Probes may be ex­tremely hot or cold when removed from the instrument. Cautiously handle probes to prevent personal injury. Always use the special sleeve tongs that are supplied with the calibrator to remove the sleeve. Carefully place probes on a heat/cold resistant surface or rack until they are at room tem­perature. Never place any objects other than the special probe sleeves sup­plied with the calibrator into the well.
Use only a grounded AC mains supply of the appropriate voltage to
power the instrument. Refer to Section 3.1, Specifications, for power de­tails.
Before initial use, after transport, and anytime the furnace has not been energized for more than 10 days, the instrument needs to be energized for a “dry-out” period of 1-2 hours before it can be assumed to meet all of the safety requirements of the IEC 1010-1.
The instrument is equipped with operator accessible fuses. If a fuse blows, it may be due to a power surge or failure of a component. Replace the fuse once. If the fuse blows a second time, it is likely caused by fail ure of a component part. If this occurs, contact Hart Scientific Customer Service. Always replace the fuse with one of the same rating, voltage, and type. Never replace the fuse with one of a higher current rating.
If a mains supply power fluctuation occurs, immediately turn off the in strument. Power bumps from brownouts and blackouts could damage the instrument. Wait until the power has stabilized before re-energizing the instrument.
Air circulated through the unit keeps the chassis cool. DO NOT SHUT OFF THE FURNACE WHILE AT HIGH TEMPERATURES.
4 Safety Guidelines
-
-
-
-
11
5 Quick Start
5.1 Unpacking
5 Quick Start
Unpacking
Unpack the furnace carefully and inspect it for any damage that may have oc curred during shipment. If there is shipping damage, notify the carrier immediately.
Verify that the following components are present:
9150 Furnace
3150, Insert
Insert Insulator
Power Cord
Manual
RS-232 Cable
5.2 Set-Up
Place the calibrator on a flat surface with at least 6 inches of free space around the instrument. Plug the power cord into a grounded mains outlet. Observe that the nominal voltage corresponds to that indicated on the back of the calibrator.
Carefully insert the probe sleeve into the well. (DO NOT drop the sleeve in the well.) Probe sleeve holes should be of the smallest diameter possible while still allowing the probe to slide in and out easily. Sleeves with various hole sizes are available from Hart Scientific. The well must be clear of any foreign objects, dirt and grit before the sleeve is inserted. The sleeve is inserted with the two small tong holes positioned upward.
Turn on the power to the calibrator by toggling the switch on the power entry module. The fan should begin quietly blowing air through the instrument and the controller display should illuminate after 3 seconds. After a brief self test the controller should begin normal operation. If the unit fails to operate please check the power connection.
The display begins to show the well temperature and the well heater starts oper ating to bring the temperature of the well to the set-point temperature.
-
-
5.3 Power
Plug the furnace power cord into a mains outlet of the proper voltage, fre quency, and current capability. Refer to Section 3.1, Specifications, for power details. Turn the furnace on using the rear panel “POWER” switch. The furnace turns on and begins to heat to the previously programmed temperature set-point. The front panel LED display indicates the actual furnace temperature.
-
13
Loading...
+ 34 hidden pages