Fluke 80PK-26 Operating Manual

80PK-26
SureGripTM Tapered
Temperature Probe
Instruction Sheet
XWWarning
To avoid electrical shock, do not use this probe when voltages exceeding 24 V ac rms or 60 V dc are present. The probe tip is electronically connected to the output terminals.
Introduction
The 80PK-26 SureGripTM Tapered Temperature Probe is essentially a bead probe with a handle that can be used as a surface probe. The stainless steel sheath allows the bead to measure temperatures higher than bead probes with flexible leads. The handle also allows you to apply more focused surface pressure without having your hand in close proximity to the source. The 40-inch (1-meter) cable is terminated with a Type K miniature thermocouple connector with 0.792-mm (.312-in) pin spacing. The 80PK-26 can be used with any temperature-measuring instrument that is designed to accept Type K thermocouples and has miniature connector input.
®
Specifications
Type
K Standard Grade Ni-Cr vs Ni-Al (Chromel vs Alumel)
Measurement Range
-40 °C to 816 °C (-40 °F to 1500 °F)
Accuracy
(With respect to ANSI MC96.1-1982-Standard Limits of Error):
Note
All error calculations should be done in scaled to °F.
Range
-40 °C to 293 °C (-40 °F to 559.4 °F)
293 °C to 816 °C (559.4 °F to 1500 °F)
November 2004 Rev.2, 9/09
© 2004-2009 Fluke Corporation. All rights reserved. All product names are trademarks of their respective companies
±2.2 °C
±0.75 %
°
C, then
Accuracy
(% of reading)
Output
@ 25 °C (77 °F) = 1.00 mV (reference junction @ 0 °C)
Seebeck Coefficient
25 °C (77 °F) = 40.50 μV / °C
Measurement Time
(Time Constant): 3.0 seconds for air moving at 3.33 meters / second (10.9 feet / second) at sea level pressure (5 time constants = 1 complete step change, i.e., 15 seconds)
Maximum Voltage
24 V ac rms or 60 V dc
Maximum Temperature of Tip
816 °C (1500 °F)
Sheath
Material: 304 Stainless Steel Dimensions
Diameter: 4.8 mm (.19 in) Length: 20.32 cm (8 in)
Bead (sensor) Sheath
Material: Inconel 600 Dimensions
Diameter: 1.5748 mm (.062 in) Length: 11.049 mm (.435 in)
Grounding
Junction Welded to Sheath.
Cable
Length: 1 meter (40 in) Insulation
Material: PVC Maximum Temperature: 105 °C (220 °F) Jacket Color: Gray
Conductors
Type: K Size: AWG #24 stranded (7 strands of #32)
Handle
Material: Hytrel Maximum Temperature: 125 °C (257 °F)
Connector
Type: Mini-thermocouple connector with .792 mm (.0312 in) pin spacing
Material: Yellow Hytrel Maximum Temperature: 1.25 °C (257 °F)
Overall Probe Length
31.9 cm (12.55 in) from tip to end of cable strain relief.
Protection
Class 3. Relates solely to insulation and grounding properties defined in IEC 348.