Fluke 80PK-22 CAL User manual

Page 1
80PK-22
SureGripTM Immersion
Temperature Probe
Instruction Sheet
XWWarning
To avoid electrical shock, do not use this probe when voltages exceeding 24 V ac rms or 60 V dc are present. The probe tip is electronically connected to the output terminals.
Introduction
The 80PK-22 SureGripTM Immersion Temperature Probe is designed for use in liquids or gels. The probe can also be used as a general purpose temperature probe. The thermocouple junction is protected from tip to handle by an Inconel sheath. The 40-inch (1 meter) cable is terminated with a Type K miniature thermocouple connector with 0.792-mm (0.312-in) pin spacing. The 80PK-22 can be used with any temperature­measuring instrument that is designed to accept Type K thermocouples and has a miniature connector input.
Note
The 80PK-22 is not recommended for use in food or beverage preparation. Use 80PK-25 for food preparation.
®
Specifications
Type
K Standard grade Ni-Cr vs. Ni-Al (Chromel vs. Alumel)
Measurement Range
-40 °C to 1090 °C (-40 °F to 1994 °F) Accuracy
(With respect to ANSI MC96.1-1982-Standard Limits of Error)
Note
All error calculations should be done in °C, then scaled to °F.
Range
-40 °C to 293 °C (-40 °F to 559.4 °F)
293 °C to 1090 °C (559.4 °F to 1994 °F)
November 2004 Rev.2, 9/09
© 2004-2009 Fluke Corporation. All rights reserved All product names are trademarks of their respective companies.
±2.2 °C
±0.75 %
Accuracy
(% of reading)
www.GlobalTestSupply.com
Page 2
Output
@ 25 °C (77 °F) =1.00 mV (reference junction @ 0 °C)
Seebeck Coefficient
@ 25 °C (77 °F) = 40.50 μV / °C
Measurement Time
(Time Constant): 3.0 seconds in 100 °C still water at sea level pressure. Complete step change equals 5 time constants (3.75 seconds).
Maximum Voltage
24 V ac rms or 60 V dc
Maximum Temperature of Tip
1090 °C (1994 °F)
Sheath
Material: Alloy 600 Dimensions: Diameter: 3.175 mm (1/8 in) Length: 212.725 mm (8.375 in)
Grounding
Junction welded to sheath.
Cable
Length: 40 inches (1 meter) Insulation Material: PVC Maximum Temperature: 105 °C (220 °F) Jacket Color: Gray
Conductors
Type: K Size: AWG #24 stranded (7 strands of #32)
Handle
Material: Hytrel Maximum Temperature: 125 °C (257 °F)
Connector
Type: Mini-thermocouple connector with .792 mm (0.312 in) pin spacing
Material: Yellow Hytrel Maximum Temperature: 125 °C (257 °F)
Overall Length
31.75 cm (12.5 in) from probe tip to end of cable strain relief.
Protection
Class 3. Relates solely to insulation and grounding properties defined in IEC 348.
www.GlobalTestSupply.com
Page 3
Measurement Considerations
Instrument Compatibility
The 80PK-22 is compatible with any temperature-measuring instrument that accepts Type K thermocouples, has a miniature thermocouple connector, and has cold reference junction compensation. Accuracy of the temperature­measuring instrument must be considered along with the 80PK-22 accuracy specification to determine the overall accuracy of the combination.
Temperature Limitations
The probe tip of the 80PK-22 has a continuous temperature rating of 1090 °C (1994 °F). However the opposite end of the sheath nearest the handle should not be subjected to temperatures greater than 125 °C (257 °F). This is the maximum temperature limitation of the handle.
Media Limitations
The sheath material of the 80PK-22 is Inconel, an alloy of chromium and nickel. It has excellent resistance to oxidation and corrosion at high temperatures, but it should not be used in the presence of sulfur above 537 °C (1000 °F).
Operation
Use the 80PK-22 as follows:
1. Connect the 80PK-22 to a compatible Type K temperature measuring instrument using the miniature (0.312-inch pin spacing) thermocouple connector.
2. Turn on the measuring instrument, and select the appropriate range and scale.
3. Check the readout on the measuring instrument. With no heat or cold source applied to the tip of the probe, the measuring instrument should display the ambient (room) temperature. If the instrument does not read out properly, refer to “Troubleshooting”.
Minimizing Thermal Shunting
The 80PK-22 Immersion Probe should be inserted at least 6.35 cm (2.5 in) into the environment to be measured to minimize the shunting effect of the sheath.
Troubleshooting
With no heat or cold applied to the probe, the measuring instrument should display the ambient temperature. If the measuring instrument does not read out properly, try the following:
1. Verify that the temperature-measuring instrument is designed to be used with Type K thermocouples. The temperature-measuring instrument should have a yellow input connector and / or be marked with a “K” either on the case or on the display.
2. Check for an open circuit indicator on the measuring instrument. Some temperature measuring instruments have a built-in circuit to indicate if the connected probe is open. (All Fluke Temperature-measuring instruments have this feature.) Refer to the owners manual accompanying the measuring instrument to see if this feature is available.
Short the two input pins of the measuring instrument with a piece of wire. If the instrument is functioning, it should indicate the ambient temperature.
www.GlobalTestSupply.com
Page 4
3. If you suspect a broken connection, use an ordinary ohmmeter to read the continuity of the probe from pin to pin. The ohmmeter should read 10 ohms or less if there is continuity.
Scale Conversions
Use the following equation to convert °C to °F:
(°C x 1.8) + 32 = °F
Use the following equation to convert °F to °C:
(°F –32) x 0.5556 = °C
8.375"
(212.725 mm)
Diameter: .125" (3.175 mm)
www.GlobalTestSupply.com
Loading...