Technical Data
Fluke 718 Pressure Calibrator
The Fluke 718 pressure calibrator family (4 models)
provide a total pressure calibration solution for
transmitters, gauges and switches. This compact
pressure calibration solution is about a third of the
kilogram (two pounds). The unique, easy to clean
pump from fl uids and provides the ability to service
the pump without disassembly.
Pressure Source and Milliamp Measurement to
Integrated pump is easily cleaned when
– reduces cost of ownership and repairs
– service the pump in the fi eld
Error % calculation for quicker pass/fail decision
making in the fi eld
Best in class mA accuracy of 0.015 % for better
measurement confi dence and workload coverage
Min/Max/Hold to capture changing measurements
Switch test feature for quicker and more accurate
pressure switch testing
New 1 and 300 PSI pressure ranges mean fewer
1, 30, 100 & 300 PSI ranges now available
Maximum voltage:
30 volts: Non-operating
Temperature:
-40 °C to 60 °C
Operating temperature:
-10 °C to 55° C
Relative humidity:
95 % (10 °C to 30 °C); 75 % (30 °C to 40 °C);
Operating altitude:
3,000 m max
1 m drop test
Vibration:
Random, 2 g, 5 Hz – 500 Hz
CSA C22.2 No. 1010.1: 1992
EMC:
EN50082-1: 1992 and EN55022: 1994 Class B
201 mm x 98 mm x 52 mm
216 mm x 94 mm x 66 mm
Power:
9V battery ANSI/NEDA 1604A or IEC 6LR619V alkaline; two
batteries in 718
Battery Life:
4 to 20 hours typical, depending on functions used
Warranty:
Three years (one year on pressure pump in Fluke 718)
Display:
LCD, 5 digit pressure and current simultaneous
Accuracy:
0.05 %
Note:
For use with non-corrosive gasses
2 Fluke Pressure Calibrators
Functional Pressure Specifi cations
Model
Range
Resolution
Over Pressure
Functions
718 1G
-1 PSI to +1 PSI, -68.9 mbar to
0.0001 psi, 0.001 mbar
Zero, Min, Max, Hold, Damp
718 30G
-12 PSI to 30 PSI, (-850 mbar to
0.001 psi, 0.1 mbar
Zero, Min, Max, Hold, Damp
718 100G
-12 PSI to 100 PSI, (-850 mbar to
0.01 psi, 1 mbar
Zero, Min, Max, Hold, Damp
718 300G
-12 PSI to 300 PSI, (-850 mbar to
0.01 psi, 1 mbar
Zero, Min, Max, Hold, Damp
717 1G
-1 PSI to 1 PSI, (-68.9 mbar to
0.001 psi, 0.001 mbar
Zero, Min, Max, Hold, Damp
717 30G
-12 PSI to 30 PSI, (-850 mbar to
0.001 psi, 0.1 mbar
Zero, Min, Max, Hold, Damp
717 100G
-12 PSI to 100 PSI, (-850 mbar to
0.01 psi, 1 mbar
Zero, Min, Max, Hold, Damp
717 300G
-12 PSI to 300 PSI, (-850 mbar to
0.01 psi, 1 mbar
Zero, Min, Max, Hold, Damp
717 500G
0 PSI to 500 PSI, (0 mbar to
34.47 bar, 0 to 3447.4 kPa )
0.01 psi, 1 mbar
Zero, Min, Max, Hold, Damp
717 1000G
0 PSI to 1000 PSI, (0 mbar to
0.1 psi, 1 mbar
Zero, Min, Max, Hold, Damp
717 1500G
0 PSI to 1500 PSI, (0 mbar to
0.1 psi, 0.01 bar
Zero, Min, Max, Hold, Damp
717 3000G
0 PSI to 3000 PSI, (0 mbar to
0.1 psi, 0.01 bar
Zero, Min, Max, Hold, Damp
717 5000G
0 PSI to 5000 PSI, (0 mbar to
344.74 bar, 0 to 34474 kPa)
0.1 psi, 0.01 bar
Zero, Min, Max, Hold, Damp
Functional Specifi cations
The New 718 Calibrator Pump
The new Fluke 718 is a uniquely protected pneu-
matic pressure calibrator and pump. The new pump
pass through the pump for calibration but inhibits
fl uids from entering the pump and causing damage.
These two check valves allow the bi-directional
fl ow of air into or out of the 718. Even after drain-
ing the process fl uid from the calibration port of the
may often still be lurking behind. In the event these
fl uids enter the 718 pump cavities, the new pump
Pump valve assembly cleaning instructions
retention caps located in the oval shaped opening
the spring and o-ring assembly.
residue.
retention caps with IPA and inspect the o-rings
needed.
They should be approximately 8.6 mm long in the
relaxed state. If
they are shorter
than this, they
may not allow
the o-ring to
Replace if
needed.
parts have
been cleaned
reinstall the o-
ring and spring
the valve body.
retention caps
tighten the cap.
the unit to at least 50 % its rated pressure.
The Calibrator is now ready for use.
Cleaning the fl uid trap in the Fluke-718
Calibrator is quick, easy and only requires a
slotted screwdriver and cotton swab
3 Fluke Pressure Calibrators
Fluke 718 series via using 700 Series Pressure Modules
Range:
2
0 to 10,000 psi, 2.5 mbar to
Resolution:
Per pressure module specs
Accuracy:
To 0.025 % of full span, per pressure module specs
Functions:
Note: media compatibility per pressure module specs
Fluke 718 series via built in pump
Range:
-12 PSI or -850 mbar to full scale, Supported Pressure Units; psi, in H
2
0 (4 °C), in
2
0 (20 °C), cm H
2
0 (4 °C), cm H
2
0 (20 °C), bar, mbar, kPa, inHg, mmHg, kg/cm
Loop Power:
Range:
Accuracy:
Note: Drive; 20 mA into 1000
for battery > 6.8V; 700
for battery 5.8 to 6.8 V
Measure Accuracy:
Range:
0 mA to 24 mA
Resolution:
0.001 mA
Accuracy:
0.015 % + 1 count
Fluke 717 Series via pressure module connector using optional Pressure Modules
Range:
0 to 10,000 PSI, 2.5 mbar to
Resolution:
Per pressure module specs
Accuracy:
0.025 % of full span, per pressure module specs
Functions:
Note: Media compatibility per pressure module specs
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa (31 40) 2 675 200 or
Fax (31 40) 2 675 222
In Canada (800) 36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fl uke.com
Printed in U.S.A. 2/2006 2577864 D-EN-N Rev A
Keeping your world
up and running.
For additional detail on the 718, see the application note "Pressure calibration with the 718."
Literature code: 2577485 at
www.fl uke.com/library
Loading...