Each Fluke product is warranted to be free from defects in material and workmanship under normal use and
service. The warranty period for the 712, 713 30G, 713 100G, 714, 715, 716, 717 1G, 717 30G, 717 100G,
717 300G, 717 500G, 717 1000G, 717 1500G, 717 3000G , 717 5000G, 718 1G, 718 30G, 718 100G, and
718 300G Calibrators is three years and begins on the date of shipment. The warranty period for the 718Ex
Pressure Calibrator and the 718 Pump assembly is one year and begins on the date of shipment. Parts,
product repairs and services are warranted for 90 days. This warranty extends only to the original buyer or
end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries or to any
product which, in Fluke's opinion, has been misused, altered, neglected or damaged by accident or
abnormal conditions of operation or handling. Fluke warrants that software will operate substantially in
accordance with its functional specifications for 90 days and that it has been properly recorded on nondefective media. Fluke does not warrant that software will be error free or operate without interruption.
Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers
only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is
available if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable
international price. Fluke reserves the right to invoice Buyer for importation costs of repair/replacement parts
when product purchased in one country is submitted for repair in another country.
Fluke's warranty obligation is limited, at Fluke's option, to refund of the purchase price, free of charge repair,
or replacement of a defective product which is returned to a Fluke authorized service center within the
warranty period.
To obtain warranty service, contact your nearest Fluke authorized service center or send the product, with a
description of the difficulty, postage and insurance prepaid (FOB Destination), to the nearest Fluke
authorized service center. Fluke assumes no risk for damage in transit. Following warranty repair, the
product will be returned to Buyer, transportation prepaid (FOB Destination). If Fluke determines that the
failure was caused by misuse, alteration, accident or abnormal condition of operation or handling, Fluke will
provide an estimate of repair costs and obtain authorization before commencing the work. Following repair,
the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and
return transportation charges (FOB Shipping Point).
THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER
WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY
OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE
FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES,
INCLUDING LOSS OF DATA, WHETHER ARISING FROM BREACH OF WARRANTY OR BASED ON
CONTRACT, TORT, RELIANCE OR ANY OTHER THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or
limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not
apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court of
competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
To locate an authorized service center, visit us on the World Wide Web at
using the phone numbers listed in this manual.
www.fluke.com or call Fluke
Fluke Corporation Fluke Europe B.V.
P.O. Box 9090 P.O. Box 1186
Everett WA 98206-9090 5602 B.D.
5. Replacing the Fuses (715 shown)........................................................................... 24
6. Replacing the Fuse (718 shown) ............................................................................ 25
7. Replacement Parts (718 shown)............................................................................. 74
ix
Page 10
71X Series
Calibration Manual
x
Page 11
Introduction
The information provided in this document is for the use of
qualified personnel only. Do not perform the verification tests
or calibration procedures described in this manual unless you
are qualified to do so.
The information in this manual deals with the 71X Series Process Calibrators (hereafter
referred to as “the Calibrator” or the “71X Calibrator”). The 71X Series includes the 712,
713 30G, 713 100G, 714, 715, 716, 717 1G, 717 30G, 717 100G, 717 300G, 717 500G,
717 1000G, 717 1500G, 717 3000G, 717 5000G, 718 1G, 718 30G, 718 100G, 718
300G, and the 718Ex 30G and 718Ex 100G models.
This manual provides the following information:
• Precautions and safety information
• Specifications
• Basic maintenance (cleaning, replacing the battery and fuses)
• Verification test procedures
• Calibration and calibration adjustment procedures
• Accessories and replaceable parts
To contact Fluke, call:
XW Warning
USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
Or, visit Fluke's Web site at www.fluke.com
To register this product, visit register.fluke.com
.
1
Page 12
71X Series
Calibration Manual
Precautions and Safety Information
Use the Calibrator only as specified in this manual. Otherwise, the protection provided by
the Calibrator may be impaired.
A Warning statement identifies conditions and actions that pose hazard(s) to the user; a
Caution statement identifies conditions and actions that may damage the calibrator. The
following Warning and Caution statement applies to all of the 71X Calibrators unless
noted:
XW Warning
To avoid possible electric shock or personal injury:
• Use the 718Ex Calibrator only as described in the User
Manual and the Fluke 718Ex CCD (Concept Control
Drawing) or the protection provided by the calibrator
may be impaired.
• Follow all equipment safety procedures.
• Inspect the Calibrator before use. Do not use it if it
appears damaged.
• Check the test leads for continuity, damaged insulation,
or exposed metal. Replace damaged test leads.
• When using probes, keep fingers behind the finger
guards on the probes.
• Make sure the battery door is closed before using the
Calibrator.
• Never apply more than 30.0 V between the input
terminals, or between any terminal and earth ground.
• Applying more than 30.0 V to the input terminals
invalidates the 718Ex Calibrator’s Ex Approval and may
result in permanent damage to the unit so it can no
longer be used.
• Use the proper terminals, mode, and range for the
measuring or sourcing application.
• When making connections, connect the COM test probe
before the live test probe. When disconnecting,
disconnect the live probe before the COM probe.
• Never use the 718Ex Calibrator with the red holster
removed.
• Never open the 718Ex Calibrator case. Opening the case
invalidates the Calibrator’s Ex Approval.
2
• Replace the battery as soon as the B(low battery)
symbol appears to avoid false readings that can lead to
electric shock. Remove the 718Ex Calibrator from the Exhazardous area before opening the battery door.
• Use only type 9 V batteries, properly installed in the
meter case, to power the meter. For the 718Ex, refer to
“718Ex Approved Batteries”.
Page 13
Process Calibrators
Precautions and Safety Information
• Remove test leads from the Calibrator before opening
the battery door.
• When servicing the Calibrator, use only specified
replacement parts.
• Do not allow water inside the case.
• When using the Calibrator’s internal pressure sensor, do
not connect a pressure module at the Calibrator to avoid
misleading readings. If both a pressure module and the
internal pressure sensor are connected, the Calibrator
displays ONLY the pressure module measurement. To
avoid misleading readings, disconnect the pressure
module connector at the Calibrator.
• Remove test leads or attached thermocouple miniplug
(714 only) from the calibrator before opening the battery
door.
• Do not operate the calibrator around explosive gas,
vapor or dust.
• 713, 716, 717 30G, 717 100G, 718 30G and 718 100G, and
718Ex 30G and 718Ex 100G only: To avoid a violent
release of pressure in a pressurized system, shut off the
valve and slowly bleed off the pressure before you attach
or detach the calibrator pressure fitting to the pressure
line.
• For 718 (non-Ex) only: Use only two 9 V batteries,
properly installed in the calibrator case, to power the
calibrator.
• For 717 and 718: Turn off circuit power before
connecting the calibrator mA and COM terminals in the
circuit. Place calibrator in series with the circuit.
WCaution
• To prevent damage to the unit under test, be sure the
Calibrator is in the correct mode before connecting the
test leads.
• The 71X Series Calibrators contain parts that can be
damaged by static discharge. If you open the case,
follow the standard practices for handling static
sensitive devices. Refer to “Static Awareness”.
• Models 713, 717 and 718 only: To avoid mechanically
damaging the calibrator, do not apply torque between
the pressure fitting and the calibrator case. See Figure 1
for the proper use of tools.
3
Page 14
71X Series
Calibration Manual
• To avoid overpressure damage, do not apply pressure
that exceeds limits listed in the Users Manual for the
specific product.
• 713, 717 and 718 only: To avoid corrosion in the
pressure sensor, use the calibrator only with media
compatible with glass, ceramic, silicon, RTV, nitrile,
(Buna -N) type 303 stainless steel, and nickel.
• 718 and 718Ex only: To avoid damage to the pump, use
with dry air and non-corrosive gases only. Use of the
optional Fluke 700-ILF In-Line Filter may help isolate the
pump from contaminates.
Hold in
fixed
position
Figure 1. Proper Use of Tools (713, 717, 718, and 718Ex Models)
kf01f.eps
4
Page 15
Process Calibrators
Explanation of International Symbols
Explanation of International Symbols
The following symbols are used on the calibrator or in this calibration manual. Table 1
explains their meaning.
Table 1. International Symbols
Symbol Meaning
+
J
I
M
X
W
T
)
f
P
(
Power ON/OFF
Earth ground
Fuse
Battery
Hazardous Voltage
Refer to the instrument instruction sheet or Users Manual for information about
this feature
Double insulated
Conforms to relevant Canadian and US Standards.
Pressure
Conforms to European Union directives
Conforms to ATEX requirements.
5
Page 16
71X Series
Calibration Manual
Specifications
Maximum voltage applied between any terminal
and earth ground or between any two terminals: 30 V
Storage temperature: 712: -20 °C to 60 °C, 718Ex: -40 °C to 71 °C
(718Ex) Pressure Sensor Media: Non-corrosive gasses only
Operating temperature: -10 °C to 55 °C
Operating altitude: 3000 meters maximum
Relative humidity: 95 % up to 30 °C
Vibration: Random 2 g, 5 Hz to 500 Hz
Shock: 1 meter drop test
Safety: 712, 713, 714, 715, 716, 717
Safety: 718
Safety: 718Ex
Specifications for the 71X Calibrators are based on a one-year calibration cycle and apply
for ambient temperatures from + 18 °C to + 28 °C unless stated otherwise. “Counts” are
the number of increments or decrements of the least significant digit. General
specifications for all models are in Table 2.
Table 2. General Specifications
All other models: -40 °C to 60 °C
75 % up to 40 °C
45 % up to 50 °C
35 % up to 55 °C
Certified as compliant to CAN/CSA C22.2 No. 1010.1:1992 NRTL
Complies with ANSI/ISA S82.01-1994
Certified as compliant to CAN/CSA C22.2 No. 1010.2:1995
Complies with ANSI/ISA S82.01-1995
Certified as compliant to CAN/CSA C22.2 No. 1010.2:1995
Complies with ANSI/ISA S82.01-1995. Complies with IEC 61010-195 CAT I, 30 V;
• P ( II 1 G EEx ia IIC T4
0344 KEMA 04ATEX1061 X
•
) Class I Div. 1 Groups A-D T4
LR110460 AEx ia IIC T4
• Ta = -10 °C… +55 °C
• CE: Complies with EN61010-1 and EN61326
Power requirements: 712, 713, 714, 715, 716, 717 Single 9 V battery (ANSI/NEDA 1604A or IEC 6LR61)
Power requirements: 718 Two 9 V batteries (ANSI/NEDA 1604A or IEC 6LR61)
Power requirements: 718Ex See “718Ex Approved Batteries”.
Size: 712, 713, 714, 715, 716, 717
Size: 718, 718Ex
Weight:
712
713
714
715
716, 717 30G, 717 100G
718 30G and 718 100G and
718Ex 30G and 718Ex 100G
34.9 mm H x 87 mm W x 187 mm L;
With holster and Flex-Stand:
52 mm H x 98 mm W x 201 mm L
60 mm H x 87 mm W x 210 mm L;
With holster: 66 mm H x 94 mm W x 216 mm L
337 g; With holster and Flex-Stand: 587 g
369 g; With holster and Flex-Stand: 624 g
332 g; With holster and Flex-Stand: 584 g
349 g; With holster and Flex-Stand: 601 g
369 g; With holster and Flex-Stand: 624 g
737 g; With holster: 992G
6
Page 17
Process Calibrators
Specifications
712 Specifications
712 Calibrator specifications vary based on the version of the instrument. To display the
firmware version for your instrument, start with the 712 off, press and hold {, then
press the power button. Find the section heading below for the displayed version and use
the specification tables in that section to test and calibrate the instrument.
Firmware V1.1 and Earlier
Table 3. 712 Supported RTD Types
Temperature Range and Resolution Allowable Excitation
Type
Ni 120 -80.0 to 260.0 0.15 to 2.00
Pt 100 385 -200.0 to 800.0 0.15 to 2.00
Pt 200 385 -200.0 to 630.0 0.15 to 2.00
Pt 500 385 -200.0 to 630.0 0.05 to 0.80
Pt 1000 385 -200.0 to 630.0 0.05 to 0.40
Pt 100 392 -200.0 to 630.0 0.15 to 2.00
Pt 100 JIS -200.0 to 630.0 0.15 to 2.00
Range and Resolution for
2
R
R
R
Ohms Simulate and Measure
400.0 Ω to 1500.0 Ω
1500.0 Ω to 3200.0 Ω
°C
15.0 Ω to 400.0 Ω
mA
0.15 to 2.00
0.05 to 0.80
0.05 to 0.40
1
RTD
Addresses pulsed transmitters and PICs with pulses ≥ 100 ms.
1: This column is for simulate mode only. It shows the allowable excitation current from an ohmmeter or
RTD measurement device connected to the calibrator.
2: The R annunciator signifies “resistance,” not an RTD type. Select it the same way as an RTD type.
Resolution
RTD: 0.1 °C
Ohms: 0.1 Ω
Temperature Coefficient
0.005 % of ohms range per °C for temperature ranges –10 °C to 18 °C and 28 °C to
55 °C. Ohms ranges are 400 Ω, 1.5 kΩ, and 3.2 kΩ.
Table 4. 712 RTD and Ohms Simulation
Ohms Range Excitation Current from Measurement Device
15 Ω to 400 Ω
15 Ω to 400 Ω
400 Ω to 1.5 kΩ
1.5 kΩ to 3.2 kΩ
Maximum input voltage: 30 V
0.15 mA to 0.5 mA 0.15
0.5 mA to 2 mA 0.1
0.05 mA to 0.8 mA 0.5
0.05 mA to 0.4 mA 1
Accuracy, ±Ω
7
Page 18
71X Series
Calibration Manual
Table 5. 712 RTD and Ohms Measurement
Ohms Range
15 Ω to 400 Ω
400 Ω to 1.5 kΩ
1.5 kΩ to 3.2 kΩ
Maximum input voltage: 30 V
Excitation current from 712: 0.3 mA
Firmware V1.2 through V1.9
RTD
Type
Ni 120 -80.0 to 260.0
Pt 100 385 -200.0 to 800.0
Pt 200 385 -200.0 to 250.0
Range ˚C (˚F)
(-112.0 to 500.0)
(-328.0 to 1472.0)
(-328.0 to 482.0)
250.0 to 630.0
(482.0 to 1166.0)
Accuracy, Four-Wire ± Ω
0.1
0.5
1
Table 6. RTD Specifications
Accuracy (˚C)
Input
4-Wire 2-Wire & 3-Wire
0.2 0.3 0.2 0.1 to 3.0
0.33 0.5 0.33 0.1 to 3.0
0.2
0.8
0.3
1.6
Output
0.2
0.8
Allowable
Excitation (mA)
0.1 to 3.0
Pt 500 385 -200.0 to 500.0
(-328.0 to 932.0)
500.0 to 630.0
(932.0 to 1166.0)
Pt 1000 385 -200.0 to 100.0
(-328.0 to 212.0)
100.0 to 630.0
(212.0 to 1166.0)
Pt 100 392
(3926)
Pt 100 JIS
(3916)
Addresses pulsed transmitters and PLCs with pulses as short as 5 ms.
Allowable Excitation is for Output mode only. It shows the allowable excitation current from an ohmmeter or
RTD measurement device connected to the calibrator.
Excitation current from 712: 0.2 mA.
Maximum input voltage: 30 V
-200.0 to 630.0
(-328.0 to 1166.0)
-200.0 to 630.0
(-328.0 to 1166.0)
0.3
0.4
0.2
0.2
0.3 0.5 0.3 0.1 to 3.0
0.3 0.5 0.3 0.1 to 3.0
0.6
0.9
0.4
0.5
0.3
0.4
0.2
0.2
0.05 to 0.8
0.05 to 0.4
8
Page 19
Process Calibrators
Table 7. Ohms Specifications
Specifications
Ohms Range
400 Ω to 1.5 kΩ0.5 0.5 0.05 to 0.8
1.5 kΩ to 3.2 kΩ1 1 0.05 to 0.4
Allowable Excitation is for Output mode only. It shows the allowable excitation current from an ohmmeter or
RTD measurement device connected to the calibrator.
Excitation current from 712: 0.2 mA.
Maximum input voltage: 30 V
Input Accuracy
4-Wire
±Ω
Output
Accuracy
±Ω
0.15 0.1 to 0.5 0 Ω to 400 Ω 0.1
0.1 0.5 to 3.0
Allowable Excitation (mA)
Resolution
RTD: 0.1 °C
Ohms: 0.1 Ω
Rev 1.3 or Later: < 400 Ω 0.01 Ω
> 400 Ω 0.1 Ω
Temperature Coefficient
0.005 % of ohms range per °C for temperature ranges –10 °C to 18 °C and 28 °C to
55 °C. Ohms ranges are 400 Ω, 1.5 kΩ, and 3.2 kΩ.
Firmware V2.0 and Later
Table 8. RTD Specifications
Accuracy (˚C) *
RTD
Type
Ni 120 -80.0 to 260.0 0.20 0.25 0.2 0.1 to 3.0
Pt 100 385
Pt 200 385
Pt 500 385
Range ˚C
4-Wire 2-Wire & 3-Wire
-200.0 to 100.0 0.20 0.28 0.2
100.0 to 300.0 0.30 0.40 0.3
300.0 to 600.0 0.40 0.52 0.4
600.0 to 800.0 0.50 0.65 0.5
-200.0 to 100.0 0.80 1.00 0.8
100.0 to 300.0 0.90 1.15 0.9
300.0 to 630.0 1.00 1.20 1.0
-200.0 to 100.0 0.40 0.60 0.4
100.0 to 300.0 0.50 0.75 0.5
300.0 to 630.0 0.60 0.90 0.6
Input
Source
Allowable
Excitation (mA)
0.1 to 3.0
0.05 to 0.8
0.05 to 0.8
Table 8. RTD Specifications (cont.)
9
Page 20
71X Series
Calibration Manual
RTD
Type
Range ˚C
Accuracy (˚C) *
Input
4-Wire 2-Wire & 3-Wire
Source
Allowable
Excitation (mA)
Pt 1000 385
Pt 100 392
(3926)
Pt 100 JIS
3916
Addresses pulsed transmitters and PLCs with pulses as short as 5 ms.
Excitation current from 712: 0.2 mA.
Maximum input voltage: 30 V
*2-wire: Does not include lead resistance
3-wire: Assumes matched leads
Excitation current: 0.2 mA.
Maximum input voltage: 30 V
*2-wire: Does not include lead resistance
5 to 400 Ω0.1 to 0.5 mA 0.025 % ±0.1 Ω
5 to 400 Ω0.5 to 3.0 mA 0.025 % ±0.05 Ω
400 to 1500 Ω0.05 to 0.8 mA 0.025 % ±0.5 Ω
1500 to 4000 Ω0.05 to 0.4 mA 0.025 % ±0.5 Ω
Resolution
RTD: 0.1 °C
Ohms: 0.1 Ω
Rev 1.3 or Later: < 400 Ω 0.01 Ω
> 400 Ω 0.1 Ω
Table 10. Ohms Source Specifications
Excitation Current from
Measurement Device
Accuracy
Page 21
Process Calibrators
Specifications
Temperature Coefficient
0.005 % of ohms range per °C for temperature ranges –10 °C to 18 °C and 28 °C to
55 °C. Ohms ranges are 400 Ω, 1.5 kΩ, and 4.0 kΩ.
713 Specifications
Table 11. 713 30G Pressure Input
Firmware Version* Range Accuracy
V1.2 or earlier 0 to 30 psi (206.85 kPa) ± 0.05 % of range
V1.3 or later -12 psi (82.74 kPa) to 30 psi (206.85 kPa) ± 0.05 % of range
Maximum nondestructive pressure: 3X top of range (90 psi, 620 kPa, 6.2 bar)
Temperature coefficient: 0.01 % of range per °C for temperature ranges -10 °C to 18 °C and 28 °C to 55°C
* To display the firmware version, start with the 713 off, press and hold M, then press the power button.
Table 12. 713 100G Pressure Input
Firmware Version* Range Accuracy
V1.2 or earlier 0 to 100 psi (689.5 kPa) ± 0.05 % of range
V1.3 or later -12 psi (-82.8 kPa) to 100 psi (689.5 kPa) ± 0.05 % of range
Maximum nondestructive pressure: 2X top of range (200 psi, 1380 kPa, 13.8 bar)
Temperature coefficient: 0.01% of range per °C for temperature ranges -10 °C to 18 °C and 28 °C to 55 °C
* To display the firmware version, start with the 713 off, press and hold M, then press the power button.
11
Page 22
71X Series
Calibration Manual
Table 13. 713 Pressure Range and Resolution
Displayed
Pressure Units
psi 30.000 psi 100.00 psi
inH2O at 4 °C 830.4 inH2O 2768.0 inH2O
inH2O at 20 °C 831.9 inH2O 2772.9 inH2O
cmH2O at 4 °C, 2109.0 cmH2O 7030 cmH2O
cmH2O at 20 °C 2113.0 cmH2O 7043 cmH2O
bar 2.0685 bar 6.895 bar
mbar 2068.5 mbar 6895 mbar
kPa 206.85 kPa 689.5 kPa
inHg 61.080 inHg 203.6 inHg
mmHg 1551.3 mmHg 5171 mmHg
kg/cm2 2.1090 kg/cm2 7.030 kg/cm2
Table 14. 713 30G and 713 100G DC mA Input
Range Resolution
Model 713 30G
Range and Resolution
Model 713 100G
Range and Resolution
Accuracy,
±(% of Reading + Counts)
24 mA 0.001 mA 0.025 + 1
Overload protection: 125 mA, 250 V fast acting fuse
Temperature coefficient: 0.005 % of range per °C for temperature ranges – 10 °C to 18 °C and 28 °C to
55 °C
714 Specifications
714 Calibrator specifications vary based on the version of the instrument. To display the
firmware version for your instrument, start with the 714 off, press and hold {, then
press the power button. Find the section heading below for the displayed version and use
the specification tables in that section to test and calibrate the instrument.
Firmware Earlier than v2.0
Table 15. 714 Temperature Measure and Thermocouple Simulate
TC Type Resolution Error Reference Junction Error
J, K, T, E, L, U 0.1 °C ± (0.3 °C + 10 µV) ± 0.2 °C
B, R, S 1 °C ± (0.3 °C + 10 µV) ± 0.2 °C
Maximum input voltage: 30 V
Temperature Coefficient: 0.05 x specified accuracy per °C for temperature ranges – 10 °C to 18 °C and
28 °C to 55 °C
12
Page 23
Process Calibrators
Table 16. 714 Millivolt Measure and Source
Range Resolution Accuracy
-10 mV to 75 mV 0.01 mV ± (0.025 % of range (75 mV) + 1 count)
Maximum input voltage: 30 V
Specifications
Firmware V2.0 and Later
Table 17. 714 Temperature Measure and Thermocouple Simulate
TC Type Range °C Accuracy °C *
J
K
T
E
R
S
B
L
U
Maximum input voltage: 30 V
* Includes 0.2 °C cold junction compensation (CJC) error
-210.0 to 0.0 0.6
0.0 to 800.0 0.4
800.0 to 1200.0 0.5
-200.0 to 0.0 0.8
0.0 to 1000.0 0.5
1000.0 to 1372.0 0.7
-250.0 to 0.0 0.8
0.0 to 400.0 0.4
-250.0 to -100.0 0.8
-100.0 to 1000.0 0.4
-20.0 to 0.0 2.0
0.0 to 1787.0 1.4
-20.0 to 0.0 2.0
0.0 to 1767.0 1.4
600.0 to 800.0 1.4
800.0 to 1000.0 1.5
1000.0 to 1820.0 1.7
-200.0 to 0.0 0.45
0.0 to 900.0 0.4
-200.0 to 0.0 0.7
0.0 to 600.0 0.45
Table 18. 714 Millivolt Measure and Source
Range Resolution Accuracy
-10 mV to 75 mV 0.001 mV 0.015 % ± 10 µV
Maximum input voltage: 30 V
Maximum source current is 1.0 mA
13
Page 24
71X Series
Calibration Manual
715 Specifications
Firmware Earlier than V2.0
715 Calibrator specifications vary based on the version of the instrument. To display the
firmware version for your instrument, start with the 715 off, press and hold {, then
press the power button. Find the section heading below for the displayed version and use
the specification tables in that section to test and calibrate the instrument.
Table 19. 715 DC V Input and Output
Range Resolution
100 mV 0.01 mV 0.02 % + 2
10 V 0.001 V 0.02 % + 2
Input impedance: 2 MΩ (nominal), < 100 pF
Overvoltage protection: 30 V
Voltage drive capability: 1 mA
Table 20. 715 DC mA Input and Output
Range Resolution
24 mA 0.001 mA 0.02 % + 2
Overload protection: 125 mA, 250 V fast acting fuse
mA Output: 0 % = 4 mA, 100 % = 20 mA
Temperature Coefficient
0.005 % of ohms range per °C for temperature ranges – 10 °C to 18 °C and 28 °C to
55 °C
Source mode
Compliance: 1000 Ω at 20 mA for battery voltage ≥ 6.8 V (700 Ω at 20 mA for battery
voltage 5.8 to 6.8 V)
Accuracy,
± (% of Reading + Counts)
Accuracy,
± (% of Reading + Counts)
14
Simulate mode
External loop voltage requirement: 24 V nominal, 30 V maximum, 12 V minimum
Loop Power
24 V ± 10 %
Page 25
Process Calibrators
Specifications
Firmware 2.0 and Later
Table 21. 715 DC V Input and Output
Range Resolution
200 mV 0.01 mV 0.015 % + 2
20 V output
25 V input
Input impedance: 1 MΩ (nominal), < 100 pF
Overvoltage protection: fuseless
Voltage drive capability: 1 mA
Table 22. 715 DC mA Input and Output
Range Resolution
24 mA 0.001 mA 0.01 % + 2
Overload protection: fuseless
Temperature Coefficient
0.005 % of ohms range per °C for temperature ranges – 10 °C to 18 °C and 28 °C to
55 °C
Source mode
Compliance: 1000 Ω at 20 mA for battery voltage ≥ 6.8 V (700 Ω at 20 mA for battery
voltage 5.8 to 6.8 V)
Accuracy,
± (% of Reading + Counts)
0.001 V 0.01 % + 2
Accuracy,
± (% of Reading + Counts)
Simulate mode
External loop voltage requirement: 24 V nominal, 30 V maximum, 12 V minimum
Overload protection: 125 mA, 250 V fast acting fuse
Temperature coefficient: 0.005 % of range per °C for temperature ranges -10 °C to 18 °C and 28 °C to
55 °C
± (% of Reading + Counts)
Accuracy,
Loop Supply
24 V ±10%
16
Page 27
Process Calibrators
Specifications
717 Specifications
Accuracy is specified for 1 year after calibration at operating temperatures of -10 °C to +
55 °C. To display firmware version, start with the unit off and press and hold M, then
press O.
Pressure
Table 25. 717 Pressure Specifications
Version 1.2 or Lower
Model Range SI Range Metric Max SI Max Metric
717-30 G (0 to 30) PSI 0 to 206.85 kPa 90 PSI 620 kPa
Version 1.3 or Higher
Model Range SI Range Metric Max SI Max Metric
717-1G (-1 to 1) PSI (-7 to 7) kPa 5 PSI 34.5 kPa
717-30G (-12 to 30) PSI (-83 to 207) kPa 60 PSI 413 kPa
717-100G (-12 to 100) PSI (-83 to 690) kPa 200 PSI 1379 kPa or 1.4 mPa
(-83 to 2068) kPa or
717-300G (-12 to 300) PSI
717-500G (0 to 500) PSI 3447 kPa or 3.4 mPa 1000 PSI 6895 kPa or 6.9 mPa
717-1000G (0-1000) PSI 6895 kPa or 6.9 mPa 2000 PSI 13790 kPa or 13.8 mPa
717-1500G (0-1500) PSI 10342 kPa or 10.3 mPa 3000 PSI 20684 kPa or 20.7 mPa
717-3000G (0-3000) PSI 20684 kpa or 20.7 mPa 6000 PSI 41369 kPa or 41.4 mPa
717-5000G (0-5000) PSI 34474 kPa or 34.5 mPa 10000 PSI 68948 kPa or 69 mPa
Accuracy: Pressure Accuracy is +/- 0.05% of range
Temperature coefficient: 0.01 % of range per °C for temperature ranges of between −10 °C to18 °C and 28 °C to 55 °C.
Table 26. Pressure Display, Pressure Module Input
Range Resolution Accuracy
Refer to the Instruction Sheet for the pressure module
Range Resolution Accuracy,± ( % of Reading + Counts)
0.005 % of range per °C for temperature ranges of between −10 °C to 18 °C and 28 °C to
55 °C.
Loop Supply
24 V dc nominal
Pressure Module Input
Determined by pressure module
18
Page 29
Process Calibrators
Specifications
718 and 718Ex Specifications
Accuracy is specified for 1 year after calibration at operating temperatures of -10 °C to +
55 °C.
To display firmware version, start with the unit off, press and hold M, then press O.
Table 29. Pressure Specifications
For version 1.3 or higher
Model Range SI Range Metric Max SI Max Metric
718-1G (-1 to 1) PSI (-7 to 7) kPa 5 PSI 34.5 kPa
718-30G (-12 to 30) PSI (-83 to 207) kPa 60 PSI 413 kPa
718-100G (-12 to 100) PSI (-83 to 690) kPa 200 PSI 1379 kPa or 1.4 mPa
718-300G (-12 to 300) PSI (-83 to 2068) kPa or 2.1 mPa 375 PSI 2586 kPa or 2.6 mPa
Accuracy: Pressure Accuracy is +/- 0.05% of range
Temperature coefficient: .01 % of range per °C for temperature ranges of between −10 °C to 18 °C and 28 °C to 55 °C.
Table 30. Pressure Display, Pressure Module Input
Range Resolution Accuracy
Refer to the Instruction Sheet for the pressure module
Table 31. DC mA Input
Range Resolution Accuracy, ± ( % of Reading + Counts)
718: Version 1.2 24 mA 0.001 mA 0.025 + 1
718: Version 1.3 24 mA 0.001 mA 0.015 + 2
718 EX 24 mA 0.001 mA 0.02 + 2
19
Page 30
71X Series
Calibration Manual
Range -1 to 1 -12 to 30 -12 to 100 -12 to 300
psi 1.0000 30.000 100.00 300.00
bar 0.0689 2.0684 6.8947 20.684
mbar 68.948 2068.4 6894.8 20684
kPa 6.8948 206.84 689.48 2068.4
kg/cm2 0.0703 2.1092 7.0307 21.092
cmH2O@ 4 °C 70.309 2109.3 7030.9 21093
cmH2O@ 20 °C 70.434 2113.0 7043.4 21130
inH2O@ 4 °C 27.681 830.42 2768.1 8304.2
inH2O@ 20 °C 27.730 831.89 2773.0 8318.9
Table 32. 718 Range and Resolution
1 30 100 300
mmHg@
0 °C
inHg@
0 °C
51.715 1551.5 5171.5 15515
2.0360 61.081 203.60 610.81
Overload Protection
Fuseless overvoltage protection
Temperature Coefficient
0.005 % of range per °C for temperature ranges of between −10 °C to 18 °C and 28 °C to
55 °C.
Loop Supply
718: 24 V dc nominal
718 EX: No loop Supply
Pressure Module Input
Determined by pressure module
20
Page 31
static awareness
A Message From
Fluke Corporation
Some semiconductors and custom IC's can be
damaged by electrostatic discharge during
handling. This notice explains how you can
minimize the chances of destroying such devices
by:
1. Knowing that there is a problem.
2. Learning the guidelines for handling them.
3. Using the procedures, packaging, and
bench techniques that are recommended.
The following practices should be followed to minimize damage to S.S. (static sensitive) devices.
1. MINIMIZE HANDLING
2. KEEP PARTS IN ORIGINAL CONTAINERS
UNTIL READY FOR USE.
3. DISCHARGE PERSONAL STATIC BEFORE
HANDLING DEVICES. USE A HIGH RESIS TANCE GROUNDING WRIST STRAP.
4. HANDLE S.S. DEVICES BY THE BODY.
Page 32
5. USE STATIC SHIELDING CONTAINERS FOR
HANDLING AND TRANSPORT.
6. DO NOT SLIDE S.S. DEVICES OVER
ANY SURFACE.
8. WHEN REMOVING PLUG-IN ASSEMBLIES
HANDLE ONLY BY NON-CONDUCTIVE
EDGES AND NEVER TOUCH OPEN EDGE
CONNECTOR EXCEPT AT STATIC-FREE
WORK STATION. PLACING SHORTING
STRIPS ON EDGE CONNECTOR HELPS
PROTECT INSTALLED S.S. DEVICES.
9. HANDLE S.S. DEVICES ONLY AT A
STATIC-FREE WORK STATION.
7. AVOID PLASTIC,VINYL AND STYROFOAM
IN WORK AREA.
PORTIONS REPRINTED
WITH PERMISSION FROM TEKTRONIX INC.
AND GERNER DYNAMICS, POMONA DIV.
Dow Chemical
10. ONLY ANTI-STATIC TYPE SOLDER SUCKERS SHOULD BE USED.
11. ONLY GROUNDED-TIP SOLDERING
IRONS SHOULD BE USED.
Page 33
Process Calibrators
Basic Maintenance
Basic Maintenance
For maintenance procedures not described in this manual, contact an authorized service
center.
Cleaning
Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or
solvents.
Replacing the Battery
XWWarning
To avoid false readings, which could lead to possible electric
shock or personal injury, replace the battery as soon as the
battery indicator (B) appears.
To replace the battery in models 712,713,714,715,716 and 717, refer to Figure 2.
To replace batteries in model 718 models, refer to Figure 3.
To replace the battery in model 718Ex, refer to Figure 4 and “718Ex Approved
Batteries”.
Figure 2. Replacing the Battery
21
it07i.eps
Page 34
71X Series
Calibration Manual
Figure 3. Replacing the Battery (718 only)
Hex Head
Wrench
Wh008f.eps
22
wh009f.eps
Figure 4. 718Ex Battery Replacement
Page 35
Process Calibrators
Basic Maintenance
718Ex Approved Batteries
Battery Manufacturer Type
Alkaline, 9 volt Duracell 6LR61/MN1604
Alkaline Ultra, 9 volt Duracell 6LR61/MX1604
Alkaline Energizer, 9 volt Eveready 6LR61/522
Alkaline Power Line Industrial Battery, 9 volt Panasonic 6LR61.9V
Replacing the Fuse
XW Warning
To avoid personal injury or damage to the calibrator, use only a
0.125A 250V fast fuse, Littelfuse
To check if the fuse(s) are blown, refer to Table 33.
Table 33. Verifying a Blown Fuse
Calibrator Model Determining a Blown Fuse
712 V1.1 Put the calibrator in simulate mode (OUTPUT on the display), and check for proper
resistance at the calibrator’s OUTPUT terminals. An open or very high impedance
suggests fuse F1 is blown.
2AG.
712 V1.2 Fuse F1 is probably blown if no current comes from the center jacks when in input
2W mode. Fuse F2 is probably blown when output mode will not work.
712 ≥V2.0 Fuseless overvoltage protection.
713 Series
716
717 Series <V1.9
718 Series <V1.9
717 Series ≥V2.0
718 Series ≥V2.0
714 <V1.9 Fuse F1 is probably blown if, in the input mode, the calibrator always reads OL,
714 ≥V2.0 Fuseless overvoltage protection.
715 Fuse F1 is probably blown if:
715 ≥V2.0 Fuseless overvoltage protection.
Fuse F1 is probably blown if the mA measurement display reading does not
respond to current applied to the current (mA) inputs.
Fuseless overvoltage protection.
even with a thermocouple connected.
•in the mA input mode, the calibrator always reads 0.000, even with a signal
applied.
•in the mA output mode, with a short across the mA OUTPUT jacks, OL is
flashing on the display.
Fuse F2 is probably blown if:
•in the V output mode, with the test leads removed from the calibrator, the
display flashes OL.
•in the V input mode, the calibrator always displays OL, even with a signal
applied that is within the measurement range.
718Ex No user serviceable fuse. An open fuse voids safety certification. The unit should
be returned to Fluke for repair.
23
Page 36
71X Series
Calibration Manual
Replace the fuse(s) as follows, refer to Figures 5 and 6:
1. Remove the test leads and turn the calibrator off.
2. Remove the battery door.
3. Remove the three Phillips-head screws from the case bottom and turn the case over.
4. Gently lift the top cover from the end nearest the input jacks until it unsnaps from the
bottom cover.
5. Replace the fuse(s) with a 0.125 A 250 V fast fuse, Littelfuse
type 2AG.
F1 and F2 are the same type on the 715 and 712 V1.2.
6. (712, 714, and 715 only)
Fit the top and bottom covers together, engaging the two snaps. Make sure that the
gasket is properly seated. Reinstall the three screws.
(713, 716, 717, and 718 only)
Carefully fit the case top and circuit board assembly together, making sure that the Oring is properly seated between the pressure sensor and the pressure fitting on the
case top. Fit the case bottom onto the case top, engaging the two snaps near the
display end of the case. Reinstall the three screws.
7. Replace the battery door.
mV
V
%
mA
F2
F1
Figure 5. Replacing the Fuses (715 shown)
Snaps
it08f.eps
24
Page 37
Process Calibrators
F1
Snaps
Required Equipment
Figure 6. Replacing the Fuse (718 shown)
Required Equipment
The equipment required to perform the verification and calibration procedures in this
manual is identified in Table 34.
Table 34. Required Calibration Equipment
Calibrator
Model
712 DC Calibrator 5 to 3000 Ω
Reference Multimeter 5 to 3000 Ω
Four Test Leads 5500A/LEADS
713 DC Calibrator 0 to 24 mA Fluke 5520A
Reference Multimeter 0 to 24 mA
Dead Weight Tester -12 to 100 PSIG
Equipment Minimum Specifications
Accuracy: 70 ppm + 0.025 Ω
Accuracy: 70 ppm + 0.037 Ω
Accuracy: 0.006 % + 0.25 mA
Accuracy: 0.012 % of range
Recommended
Model
Fluke 5520A
Fluke 8508A
Fluke 8508A
Wh003f.eps
Teflon Tape
Two Test Leads 5500A/LEADS
25
Page 38
71X Series
Calibration Manual
Table 34. Required Calibration Equipment (cont.)
Calibrator
Model
714 DC Calibrator -10 to 75 mV
TC to Banana Test
Type J TC Test
715 DC Calibrator 0 to 10 Volts
Two Test Leads
716 DC Calibrator 0 to 24 mA Fluke 5520A
Two Test Leads 5500A/LEADS
700P Pressure
Load Resistors 1 kΩ & 2 kΩ 10% 0.5 W
717 DC Calibrator 0 to 24 mA Fluke 5520A
Equipment Minimum Specifications
Accuracy: 0.006 % + 0.25 mV
Reference
Multimeter
Leads
Lead
Reference
Multimeter
Reference
Multimeter
Module
Reference
Multimeter
-10 to 75 mV
Accuracy: 0.006 % + 0.25 mV
Copper Mini-Jack to Copper Wire
5500A/LEADS
Accuracy: 0.005 % + 0.5 mV
0 to 10 Volts
Accuracy: 0.005 % + 0.5 mV
0 to 24 mA
Accuracy: 0.006 % + 0.25 mA
0 to 24 mA
Accuracy: 0.006 % + 0.25 mA
Recommended
Model
Fluke 5520A
Fluke 8508A
Fluke 5520A
Fluke 8508A
Fluke 8508A
Fluke 8508A
Dead Weight Tester -12 to 100 PSIG
Accuracy: 0.012 % of range
Two Test Leads 5500A/LEADS
700P Pressure
Module
Load Resistors 1 kΩ & 2 kΩ 10% 0.5W
718/718Ex DC Calibrator 0 to 24 mA Fluke 5520A
Dead Weight Tester -12 to 100 PSIG
Two Test Leads 5500A/LEADS
700P Pressure
Load Resistors 1 kΩ & 2 kΩ 10% 0.5 W
Reference
Multimeter
Module
0 to 24 mA
Accuracy: 0.006 % + 0.25 mA
Accuracy: 0.012 % of range
Fluke 8508A
26
Page 39
Process Calibrators
Verification
Verification
The following verification tests check the accuracy of each calibrator function against the
calibrator’s specifications. If the calibrator fails any of these tests, calibration adjustment
or repair is required. Fluke recommends that you calibrate your 71X Calibrator once a
year to ensure that it performs according to its specifications.
To perform the verification tests, it is not necessary to open the case or adjust the
calibrator. Simply make the required connections, apply the designated source stimulus
and determine if the measurements fall within the acceptable range indicated.
Note
Throughout this section, the 71X Calibrator may be referred to as the
“UUT” (unit under test).
Preparing for Verification
To prepare for verification, do the following:
1. Make sure fuse(s) in the UUT are intact. By referring to Table 33, you should not
have to open the unit.
2. Make sure you have the required test equipment available. (Refer to Table 34.)
3. Turn on and warm up the test equipment for the time required.
4. Allow UUT to come to ambient temperature. Turn it on and allow 5 minutes for
warm-up of the UUT.
712 Verification (V1.1 and Earlier)
Resistance Measure Verification
1. Press the green
O key to turn on the 712. Press the I key, the R key, so that the
display indicates:
INPUT TYPE R 4W
2. Configure the 712 Calibrator into a 4-wire ohms measurement:
• Connect two test leads from the NORMAL jacks of the 5520A to the INPUT
jacks (two middle jacks) on the 712 Calibrator (black to black, red to red).
• Connect two additional test leads, also from the NORMAL jacks of the 5520A,
to the 712 Calibrator (black wire to the 712 red jack labeled [3W NC] and red
wire to the red jack labeled [4W NC]). All four input jacks of the 712 should now
be configured in a 4-wire ohms measurement.
3. Set the Fluke 5520A to the settings in Table 35, and verify the reading as displayed
on the 712 Calibrator:
Table 35. 712 Resistance Measure Verification
Fluke 5520A Fluke 712
207.5 Ω207.4 to 207.6 Ω
950.0 Ω949.5 to 950.5 Ω
2350.0 Ω2349.0 to 2351.0 Ω
4. Disconnect the test leads.
27
Page 40
71X Series
Calibration Manual
Resistance Source Verification
1. Using four test leads, connect the four jacks of Fluke 8508A (Ω Sense 'Hi & Lo' and
Input 'Hi & Lo') to the two middle jacks of the 712 Calibrator (black to black; red to
red). This puts the Fluke 8508A in a 4-wire configuration.
2. On the Fluke 8508A , select four-wire ohms measurement and up-range to the 1 kΩ
range. Use the 1 kΩ range for the first test point, and use the 10 kΩ range for the last
two test points. Correct ranging is important in supplying the correct excitation
current back into the 712 Calibrator. Otherwise the specifications may change, or the
measurements may be incorrect.
3. Press the
OUTPUT
Using the large ([,{), and small (],}), scroll keys, source the resistance
settings in Table 36, verifying the readings on the Fluke 8508A display:
Keypad Test
1. Press and hold the large { to source 950.0 ohms.
2. Press and hold the small
changes in 0.1 increments, then changes in 1.0 increments, then in 10.0 increments.
3. Press and hold the small
changes in 0.1 decrements, then changes in 1.0 decrements, then in 10.0 decrements.
4. Disconnect all test leads from the 712 Calibrator and press the
calibrator off.
I key on the 712 Calibrator so that the display indicates:
Table 36. 712 Resistance Source Verification
Fluke 712 Fluke 8508A
207.5 Ω.2074 to .2076 kΩ
950.0 Ω.9495 to .9505 kΩ
2350.0 Ω2.349 to 2.351 kΩ
] scroll key. Verify the numbers scrolling on the display
} scroll key. Verify the numbers scrolling on the display
O key to turn the
28
Display Verification
1. Press and hold the
the green O key. This locks the 712 in a mode where all display segments are on. All
segments will stay on until the
2. Check to see that all segments of the LCD are displayed.
3. Turn the 712 off.
The 712 verification test is complete.
I key and then turn the 712 back on by pressing and releasing
I key is released.
Page 41
Process Calibrators
Verification
712 Verification (V1.2 through V1.9)
1. Press I button until Input comes up on the display.
2. Press R until PT100 392 JIS is on the display.
3. Press [ until 4W is on the display. Set temperature standard to C.
4. Set the 5520A to 2-wire output with 2-wire compensation turned off; then make 2wire connections on the 5520A to 4-wire connections on the 712. Set the 5520A to
PT 3916 (ITS-90) mode.
5. Set the 5520A to output the RTD (resistance) values in Table 37. Verify that the
temperature readings are within the limits shown for 4-wire UUT.
Table 37. 712 Verification RTD Values
5520A Outputs (C) 3-wire 712 Readings ( C) 4-wire 712 Readings ( C)
-180 (25.799 Ohms) -179.5 to -180.5 -179.7 to -180.3
100 (139.171 Ohms) 99.5 to 100.5 99.7 to 100.3
550 (300.822 Ohms) 549.5 to 550.5 549.7 to 550.3
6. Remove the 4-wire connection on the 712 (rightmost connection). Press { once
and verify that 3W is on the display and that the temperature readings are within the
limits shown for 3-wire readings. Press [ to return to 4W mode.
7. Restore the 4-wire connection to the 712 and maintain 2-wire connection on the
5520A.
8. Press R until Ω is displayed.
9. Set the 5520A to source resistance, to a 2-wire output, with 2-wire compensation
turned off.
10. Set the 5520A to source the resistance values in Table 38. Verify that the resistance
values on the 712 are within the limits shown.
Table 38. 712 Verification Resistance Values
5520A Outputs (Ω) 4-wire 712 Readings (Ω)
5.00 4.90 to 5.10
300.00 299.90 to 300.10
1500.0 1499.5 to 1500.5
30000.0 2999.0 to 3001.0
11. Make 2-wire connections on the 712 to 4-wire connections on the Fluke 8508A . Set
the Fluke 8508A to measure 4-wire resistance.
12. Press I until OUTPUT is displayed.
13. Set the 712 to output the resistance values in Table 39. Verify that the Fluke 8508A
readings are within the limits shown.
29
Page 42
71X Series
Calibration Manual
712 Verification (V2.0 and Later)
Table 39. 712 Verification Outputs
712 Outputs (Ohms) Fluke 8508A Readings (Ohms)
5.00 4.90 to 5.10
300.00 299.90 to 300.10
1500.0 1499.5 to 1500.5
30000.0 2999.0 to 3001.0
14. Disconnect all connections to the 712. The 712 verification test is complete.
1. Press I button until Input comes up on the display.
2. Press R until PT100 392 JIS is on the display.
3. Press [ until 4W is on the display. Set temperature standard to C.
4. Set the 5520A to 2-wire output with 2-wire compensation turned off; then make 2wire connections on the 5520A to 4-wire connections on the 712. Set the 5520A to
PT 3916 (ITS-90) mode.
5. Set the 5520A to output the RTD (resistance) values in Table 40. Verify that the
temperature readings are within the limits shown for 4-wire UUT.
Table 40. 712 Verification RTD Values
5520A Outputs (C) 3-wire 712 Readings ( C) 4-wire 712 Readings ( C)
-180 (25.799 Ohms) -179.72 to -180.28 -179.8 to -180.2
100 (139.171 Ohms) 99.6 to 100.4 99.7 to 100.3
300 (213.957 Ohms) 299.7 to 300.3 299.6 to 300.4
550 (300.822 Ohms) 549.48 to 550.52 549.6 to 550.4
6. Remove the 4-wire connection on the 712 (rightmost connection). Press { once
and verify that 3W is on the display and that the temperature readings are within the
limits shown for 3-wire readings. Press [ to return to 4W mode.
7. Restore the 4-wire connection to the 712 and maintain 2-wire connection on the
5520A.
8. Press R until Ω is displayed.
9. Set the 5520A to source resistance, to a 2-wire output, with 2-wire compensation
turned off.
10. Set the 5520A to source the resistance values in Table 41. Verify that the resistance
values on the 712 are within the limits shown.
30
Page 43
Process Calibrators
Table 41. 712 Verification Resistance Values
5520A Outputs (Ω) 4-wire 712 Readings (Ω)
5.00 4.899 to 5.101
300.00 299.825 to 300.175
1500.0 1499.525 to 1500.475
30000.0 2999.15 to 3000.85
Verification
11. Make 2-wire connections on the 712 to 4-wire connections on the Fluke 8508A. Set
the Fluke 8508A to measure 4-wire resistance.
12. Press I until OUTPUT is displayed.
13. Set the 712 to output the resistance values in Table 42. Verify that the Fluke 8508A
readings are within the limits shown.
Table 42. 712 Verification Outputs
712 Outputs (Ohms) Fluke 8508A Readings (Ohms)
5.00 4.899 to 5.101
300.00 299.825 to 300.175
1500.0 1499.525 to 1500.475
30000.0 2999.15 to 3000.85
14. Disconnect all connections to the 712. The 712 verification test is complete.
713 Verification
Pressure Verification
1. Carefully attach the pressure fitting of the deadweight tester to the pressure port of
the 713 Calibrator.
The use of TEFLON tape at the pressure fitting strengthens the seal.
2. Press the green O key to turn the 713 on.
3. The display should read 0.00 PSI with the deadweight tester opened up to ambient
air. If not, press the
4. Depending if the instrument being tested is a 713 30G or a 713 100G, set up the
deadweight tester for the sequence of PSI inputs from Table 43 to be injected into the
pressure port of the 713 Calibrator.
5. Ensure the pressure has stabilized at each input before verifying the display reading.
6. Carefully vent all pressure and disconnect the 713 from the deadweight tester.
-12 psi * -12.015 to -11.985 -12 psi* -12.05 to -11.95
* Vacuum is only specified for V1.3 and later. To display the firmware version, start with the 713 off, press and
hold M, then press the power button.
mA Measure Verification
The display should indicate:
0.000mA 0.00 PSI
1. Connect a test lead from Red AUX terminal of the 5520A to the Fluke 8508A I+
terminal.
2. Connect a test lead from Black AUX terminal of the 5520A to the 713 Calibrator
Com terminal.
3. Connect a test lead from I- terminal of the Fluke 8508A to the 713 Calibrator mA
terminal.
4. Adjust the 5520A if necessary so that the current shown on the Fluke 8508A is the
same as the 5520A values shown in Table 44.
5. Verify that the display readings on the 713 calibrator are within the limits.
Table 44. 713 mA Measure Verification
32
Fluke 5520A Fluke 713
4.0000 mA 3.998 mA to 4.002 mA
12.0000 mA 11.996 mA to 12.004 mA
24.0000 mA 23.993 mA to 24.007 mA
6. Disconnect the test leads and turn off the 713 Calibrator.
Page 45
Process Calibrators
Verification
Display Verification
1. Press and hold the I key and then press and release the green O key. This locks
the 713 in a mode where all display segments are on. All segments will stay on until
the I key is released.
2. Check to see that all segments of the display are showing.
3. Turn the 713 off.
The 713 verification test is now complete.
714 Verification (Earlier than V2.0)
Thermocouple Measure Verification
1. Press the
O key to turn on the 714 Calibrator. Press the I key and the C key so
that the display indicates:
INPUT xx.x °C
where xx.x is some variable number; or OL (overload) may be indicated.
2. Press the
T key to measure Type J thermocouple and connect the Type J
thermocouple test lead from the TC jack of the Fluke 5520A to the TC jack on the
714 Calibrator, observing correct polarity.
3. Set the Fluke 5520A to output in Type J thermocouple, press the Okey, and set
the 5520A to the settings in Table 45, verifying the display readings on the 714:
Table 45. 714 Thermocouple Measure Verification
Fluke 5520A Fluke 714
-200.00 °C -200.9 °C to -199.1 °C
0.00 °C -0.7 °C to +0.7 °C
1200.00 °C 1199.3 °C to 1200.7 °C
4. Disconnect the Type J thermocouple test lead and install the two-wire copper test
lead from the Fluke 5520A NORMAL jacks to the 714 Calibrator TC jack, observing
correct polarity.
5. On the 714 press the
T key until "mV" is shown on the display.
6. Set the Fluke 5520A to the settings in Table 46, and verify the display readings on
the 714 Calibrator.
7. Disconnect the copper test lead from the Fluke 5520A.
33
Page 46
71X Series
Calibration Manual
Thermocouple Source Verification
1. Set the Fluke 8508A to VDC and 200 mV range. Connect the two-wire copper test
lead from the TC jack on the 714 Calibrator to the INPUT jacks of the Fluke 8508A.
2. Press the
Ikey on the 714 Calibrator to indicate:
OUPUT 0.00 mV
3. On the 714, press the large [ and { scroll keys to source the voltages in Table
47 while verifying the readings on the Fluke 8508A.
800.00 °C 799.6 °C to 800.4 °C
1200 °C 1199.5 °C to 1200.5 °C
The 714 verification test is now complete.
715 Verification (Earlier than V2.0)
DC Voltage Source Verification
1. On the 715 Calibrator turn the green O key on.
2. Press the I and the V key so that the display indicates:
OUTPUT 0.000 V
3. Connect test leads from the Fluke 8508A input HI & LO jacks to the voltage jacks on
the 715 Calibrator (black to COM and red to V).
4. Using the large [ scroll button of the 715 Calibrator, press to step to the voltages
in Table 53, verifying the output on the Fluke 8508A:
36
Page 49
Process Calibrators
Table 53. 715 DC Voltage Source Verification (0.000 to 10.000 V)
Fluke 715 Fluke 8508A
0.000 V -0.002 V to + 0.002 V
5.000 V 4.997 V to 5.003 V
10.000 V 9.996 V to 10.004 V
Verification
5. Press the V key on the 715 Calibrator. Display should change to:
OUTPUT 0.00 mV
6. Using the large [ key scroll button of the 715 Calibrator, press to step to the
voltages in Table 54, verifying the output on the Fluke 8508A.
Table 54. 715 DC Voltage Source Verification (0.00 to 100.00 V)
Fluke 715 Fluke 8508A
0.00 mV -0.02 mV to + 0.02 mV
50.00 mV 49.97 mV to 50.03 mV
100.00 mV 99.96 mV to 100.04 mV
37
Page 50
71X Series
Calibration Manual
DC Current Source Verification
1. Disconnect the Fluke 8508A. Press the A key on the 715 Calibrator. The display
should change to:
OUTPUT 0.000 mA
2. Connect test leads from the 715 Calibrator's mA output jacks (black to V jack and red
to +LOOP jack) to the Fluke 8508A input mA jacks (black to I- and red to I+).
3. Set the Fluke 8508A function to [
the 715 Calibrator, press to step to the current outputs in
Table 55, verifying the readings on the Fluke 8508A.
Keypad Test
1. Using the large { scroll key, press to step down to 12.000 mA.
2. Using the small ] scroll key, press to verify that the numbers scrolling on the
display change in .001 increments; then change in .01 increments; then change in 0.1
increments.
3. Using the small } scroll key, press to verify that the numbers scrolling on the
display change in .001 decrements; then change in .01 decrements; then change in 0.1
decrements.
4. Scroll to 12.000 mA.
5. Press the A key on the 715 Calibrator. Display should change to:
DC CURRENT]. Using the large [ scroll button of
Table 55. 715 DC Current Source Verification
Fluke 715 Fluke 8508A
4.000 mA 3.9972 mA to 4.0028 mA
12.000 mA 11.9956 mA to 12.0044 mA
24.000 mA 23.9932 mA to 24.0068 mA
38
OUTPUT 50.00 mA %
DC Current Measure Verification
1. Press I key on the 715 Calibrator. Display should indicate closely to:
INPUT -25.00 mA %
2. Press the A key and the display should change to:
INPUT 0.000 mA
3. Connect a test lead from Red AUX terminal of the 5520A to the Fluke 8508A I+
terminal.
4. Connect a test lead from Black AUX terminal of the 5520A to the 715 Calibrator
Com terminal.
5. Connect a test lead from I- terminal of the Fluke 8508A to the 715 Calibrator mA
terminal.
6. Adjust the 5520A if necessary so that the current shown on the Fluke 8508A is the
same as the 5520A values shown in Table 56.
7. Verify that the display readings on the 715 Calibrator are within the limits.
Page 51
Process Calibrators
Table 56. 715 DC Current Measure Verification
Fluke 5520A Fluke 715
24.0000 mA 23.993 mA to 24.007 mA
12.0000 mA 11.996 mA to 12.004 mA
4.0000 mA 3.998 mA to 4.002 mA
Verification
DC Voltage Measure Verification
1. Press the V key on the 715 Calibrator. Display should change to:
INPUT 0.000 V
2. Connect test leads from the output NORMAL jacks of the Fluke 5520A to the voltage
jacks on the 715 Calibrator (black to COM jack and red to the V jack).
3. Set the Fluke 5520A for the voltage settings in Table 57, and verify the display
readings on the 715 Calibrator.
Table 57. 715 DC Voltage Measure Verification (10.0000 to 0.0000 V)
Fluke 5520A Fluke 715
10.00000 V 9.996 V to 10.004 V
5.00000 V 4.997 V to 5.003 V
0.0000 V -0.002 V to +0.002 V
4. Press the V key on the 715 Calibrator. Display should indicate closely to:
INPUT 0.00 mV
5. Set the Fluke 5520A for the mV settings in Table 58, and verify display readings on
the 715 Calibrator.
Table 58. 715 DC Voltage Measure Verification (0.0000 mV to 100.0000 mV)
Fluke 5520A Fluke 715
50.0000 mV 49.97 mV to 50.03 mV
100.0000 mV 99.96 mV to 100.04 mV
6. Disconnect the test leads from 715 Calibrator and turn the green O key off.
Display Verification
1. Press and hold the I key and then turn the 715 back on by pressing and releasing
the green O key. This locks the 715 in a mode where all display segments are on. All
segments will stay on until the I key is released.
0.0000 mV -0.02 mV to + 0.02 mV
2. Check to see that all segments of the LCD are displayed.
3. Turn the 715 off.
The 715 verification test is complete.
39
Page 52
71X Series
Calibration Manual
715 Verification (V2.0 and Later)
DC Voltage Source Verification
1. On the 715 Calibrator turn the green O key on.
2. Press the I and the V key so that the display indicates:
OUTPUT 0.000 V
3. Connect test leads from the Fluke 8508A input HI & LO jacks to the voltage jacks on
the 715 Calibrator (black to COM and red to V).
4. Using the large [ scroll button of the 715 Calibrator, press to step to the voltages
in Table 59, verifying the output on the Fluke 8508A:
Table 59. 715 DC Voltage Source Verification (0.000 to 20.000 V)
Fluke 715 Fluke 8508A
0.000 V -0.002 V to + 0.002 V
5.000 V 4.998 V to 5.002 V
15.000 V 14.997 V to 15.003 V
5. Press the V key on the 715 Calibrator. Display should change to:
OUTPUT 0.00 mV
6. Using the large [ key scroll button of the 715 Calibrator, press to step to the
voltages in Table 60, verifying the output on the Fluke 8508A.
Table 60. 715 DC Voltage Source Verification (0.00 to 200.00 mV)
Fluke 715 Fluke 8508A
0.00 mV -0.02 mV to + 0.02 mV
50.00 mV 49.98 mV to 50.02 mV
150.00 mV 149.96 mV to 150.04 mV
40
Page 53
Process Calibrators
Verification
DC Current Source Verification
1. Disconnect the Fluke 8508A. Press the A key on the 715 Calibrator. The display
should change to:
OUTPUT 0.000 mA
2. Connect test leads from the 715 Calibrator's mA output jacks (black to V jack and red
to +LOOP jack) to the Fluke 8508A input mA jacks (black to I- and red to I+).
3. Set the Fluke 8508A function to [
the 715 Calibrator, press to step to the current outputs in
Table 61, verifying the readings on the Fluke 8508A.
Table 61. 715 DC Current Source Verification
Fluke 715 Fluke 8508A
4.000 mA 3.9976 mA to 4.0024 mA
12.000 mA 11.9968 mA to 12.0032 mA
24.000 mA 23.9956 mA to 24.0044 mA
DC Current Measure Verification
1. Press I key on the 715 Calibrator. Display should indicate closely to:
INPUT -25.00 mA %
2. Press the A key and the display should change to:
INPUT 0.000 mA
3. Connect a test lead from Red AUX terminal of the 5520A to the Fluke 8508A I+
terminal.
4. Connect a test lead from Black AUX terminal of the 5520A to the 715 Calibrator
Com terminal.
5. Connect a test lead from I- terminal of the Fluke 8508A to the 715 Calibrator mA
terminal.
DC CURRENT]. Using the large [ scroll button of
6. Adjust the 5520A if necessary so that the current shown on the Fluke 8508A is the
same as the 5520A values shown in Table 62.
7. Verify that the display readings on the 715 Calibrator are within the limits.
Table 62. 715 DC Current Measure Verification
Fluke 5520A Fluke 715
24.0000 mA 23.993 mA to 24.007 mA
12.0000 mA 11.996 mA to 12.004 mA
4.0000 mA 3.998 mA to 4.002 mA
DC Voltage Measure Verification
1. Press the V key on the 715 Calibrator. Display should change to:
INPUT 0.000 V
2. Connect test leads from the output NORMAL jacks of the Fluke 5520A to the voltage
jacks on the 715 Calibrator (black to COM jack and red to the V jack).
3. Set the Fluke 5520A for the voltage settings in Table 63, and verify the display
readings on the 715 Calibrator.
41
Page 54
71X Series
Calibration Manual
Table 63. 715 DC Voltage Measure Verification (25.0000 to 0.0000 V)
Fluke 5520A Fluke 715
20.00000 V 19.996 V to 20.004 V
10.00000 V 9.997 V to 10.003 V
0.0000 V -0.002 V to +0.002 V
4. Press the V key on the 715 Calibrator. Display should indicate closely to:
INPUT 0.00 mV
5. Set the Fluke 5520A for the mV settings in Table 64, and verify display readings on
the 715 Calibrator.
Table 64. 715 DC Voltage Measure Verification (0.0000 mV to 200.0000 mV)
Fluke 5520A Fluke 715
0.0000 mV -0.02 mV to + 0.02 mV
50.0000 mV 49.98 mV to 50.02 mV
150.0000 mV 149.96 mV to 150.04 mV
6. Disconnect the test leads from 715 Calibrator and turn the green O key off.
The 715 verification test is complete.
42
Page 55
Process Calibrators
Verification
716 Verification
mA Measure Verification
1. Press the green O button to turn the calibrator on. The display should read:
----------- mA
then change to:
0.000 mA
2. Connect a test lead from Red AUX terminal of the 5520A to the Fluke 8508A I+
terminal.
3. Connect a test lead from Black AUX terminal of the 5520A to the 716 Calibrator
Com terminal.
4. Connect a test lead from I- terminal of the Fluke 8508A to the 716 Calibrator mA
terminal.
5. Adjust the 5520A if necessary so that the current shown on the Fluke 8508A is the
same as the 5520A values shown in Table 65.
6. Verify that the display readings on the 716 Calibrator are within the limits.
Table 65. 716 mA Measure Verification
Fluke 5520A Fluke 716
4.0000 mA 3.998 mA to 4.002 mA
12.0000 mA 11.996 mA to 12.004 mA
24.0000 mA 23.993 mA to 24.007 mA
7. Disconnect the test leads and press O to turn the power off.
mA Loop Power Verification
1. Hold down both LOOP POWER keys ( U and O) at the same time until the
screen reads:
---------- LOOP mA
then release the keys.
2. Apply 1 kΩ to the mA jacks from the decade box or other resistor source. The
display should read over 21 mA but less than OL.
3. Apply 2 kΩ to the mA jacks from the decade box or other resistor source. The
display should read between 11 mA and 13 mA.
4. Press
Oto turn the calibrator off.
43
Page 56
71X Series
Calibration Manual
Sensor Jack Verification
---------- mA
---------- mA
717 Verification
Pressure Verification
1. Make sure nothing is connected to the sensor port of the Fluke 716. Push
Oto turn
the calibrator on. The display should read:
then change to:
OL 0.000 mA
2. Plug the Fluke Pressure Module into the sensor port of the Fluke 716. The display
should change to:
A pressure value should appear on the screen after all of the calibration constants
have been downloaded.
3. Disconnect the Fluke Pressure Module and push O to turn the Fluke 716 off.
1. Carefully attach the pressure fitting of the deadweight tester to the pressure port of
the 717 30G Calibrator.
Note
Use Plenty of TEFLON tape when attaching pressure fitting.
The display should read 0.00 PSI with the deadweight tester opened up to ambient
air. If not, press the
Zkey until display reads 0.00 PSI.
2. Set up the deadweight tester for the sequence of PSI inputs from Table 66 to be
injected into the pressure port of the 717 Calibrator.
3. Ensure the pressure has stabilized at each input before verifying the display reading.
44
Page 57
Process Calibrators
Table 66. 717 Pressure Verification
717 1G 717 30G
Input
Pressure
0 psi 0.0005 to -0.00050 psi-0.015 to 0.015
0.2 psi 0.2005 to 0.19956 psi5.985 to 6.015
0.4 psi 0.4005 to 0.399512 psi11.985 to 12.015
0.6 psi 0.6005 to 0.599518 psi17.985 to 18.015
0.8 psi 0.8005 to 0.799524 psi23.985 to 24.015
1 psi 1.0005 to 0.999530 psi29.985 to 30.015
0.8 psi 0.8005 to 0.799524 psi23.985 to 24.015
0.6 psi 0.6005 to 0.599518 psi17.985 to 18.015
0.4 psi 0.4005 to 0.399512 psi11.985 to 12.015
0.2 psi 0.2005 to 0.19956 psi5.985 to 6.015
-1 psi -0.9995 to -1.0005-12 psi *-12.015 to -11.985
717 100G 717 300G
Input
Pressure
0 psi -0.05 to 0.050 psi0.15 to -0.15
20 psi 19.95 to 20.0560 psi60.15 to 59.85
40 psi 39.95 to 40.05120 psi120.15 to 119.85
60 psi 59.95 to 60.05180 psi180.15 to 179.85
80 psi 79.95 to 80.05240 psi240.15 to 239.85
100 psi 99.95 to 100.05300 psi300.15 to 299.85
80 psi 79.95 to 80.05240 psi240.15 to 239.85
60 psi 59.95 to 60.05180 psi180.15 to 179.85
40 psi 39.95 to 40.05120 psi120.15 to 119.85
20 psi 19.95 to 20.0560 psi60.15 to 59.85
* -12 psi -12.05 to -11.95-12 psi-11.85 to -12.15
*717 500G 717 1000G
Input
Pressure
Display
Reading
Display
Reading
Display
Reading
Input
Pressure
Input
Pressure
Input
Pressure
Display
Reading
Display
Reading
Display
Reading
Verification
0 psi 0.25 to -0.250 psi0.5 to -0.5
100 psi 100.25 to 99.75200 psi200.5 to 199.5
200 psi 200.25 to 199.75400 psi400.5 to 399.5
300 psi 300.25 to 299.75600 psi600.5 to 599.5
400 psi 400.25 to 399.75800 psi800.5 to 799.5
500 psi 500.25 to 499.751000 psi1000.5 to 999.5
400 psi 400.25 to 399.75800 psi800.5 to 799.5
300 psi 300.25 to 299.75600 psi600.5 to 599.5
200 psi 200.25 to 199.75400 psi400.5 to 399.5
100 psi 100.25 to 99.75200 psi200.5 to 199.5
45
Page 58
71X Series
Calibration Manual
717 1500G 717 3000G
Input
Pressure
0 psi 0.8 to -0.80 psi1.5 to -1.5
300 psi 300.8 to 299.3600.0 psi601.5 to 598.5
600 psi 600.8 to 599.31200.0 psi1201.5 to 1198.5
900 psi 900.8 to 899.31800.0 psi1801.5 to 1798.5
1200 psi 1200.8 to 1199.32400.0 psi2401.5 to 2398.5
1500 psi 1500.8 to 1499.33000.0 psi3001.5 to 2998.5
1200 psi 1200.8 to 1199.32400.0 psi2401.5 to 2398.5
900 psi 900.8 to 899.31800.0 psi1801.5 to 1798.5
600 psi 600.8 to 599.31200.0 psi1201.5 to 1198.5
300 psi 300.8 to 299.3600.0 psi601.5 to 598.5
Input
Pressure
0 psi 2.5 to -2.5
Display
Reading
717 5000G
Input
Pressure
Display
Reading
Display
Reading
1000 psi 1002.5 to 997.5
2000 psi 2002.5 to 1997.5
3000 psi 3002.5 to 2997.5
4000 psi 4002.5 to 3997.5
5000 psi 5002.5 to 4997.5
4000 psi 4002.5 to 3997.5
3000 psi 3002.5 to 2997.5
2000 psi 2002.5 to 1997.5
1000 psi 1002.5 to 997.5
*Vacuum is only specified for V1.3 and later. To display the firmware version, start with the 717 off, press
and hold M, then press the power button.
4. Carefully vent all pressure and disconnect the 717 from the deadweight tester.
mA Measure Verification
1. Press
Oto turn power on. The display should read:
---------- mA
then change to:
0.000 mA 0.00PSI
2. Connect a test lead from Red AUX terminal of the 5520A to the Fluke 8508A I+
terminal.
46
3. Connect a test lead from Black AUX terminal of the 5520A to the 717 Calibrator
Com terminal.
4. Connect a test lead from I- terminal of the Fluke 8508A to the 717 Calibrator mA
terminal.
Page 59
Process Calibrators
Verification
5. Adjust the 5520A if necessary so that the current shown on the Fluke 8508A is the
same as the 5520A values shown in Table 67.
6. Verify that the display readings on the 717 Calibrator are within the limits.
Table 67. 717 mA Measure Verification
Fluke 5520A Fluke 717
4.0000 mA 3.998 mA to 4.002 mA
12.0000 mA 11.996 mA to 12.004 mA
24.0000 mA 23.993 mA to 24.007 mA
7. Disconnect the test leads and press O to turn the power off.
mA Loop Power Verification
1. Hold down both LOOP POWER keys ( U and O) at the same until the screen
reads:
---------- LOOP mA
then release the keys.
2. Apply 1 kΩ to the mA jacks from the decade box or other resistor sources. The
display should read over 21 mA but less than OL.
3. Apply 2 kΩ to the mA jacks from the decade box or other resistor source. The
display should read between 11 mA and 13 mA.
4. Press
Oto turn the power off.
Sensor Jack Verification
1. Make sure nothing is connected to the sensor port of the Fluke 717.
2. Turn power on. The display should read:
---------- mA
then change to:
0.000 PSI
3. Plug the Fluke Pressure Module into the sensor port on the Fluke 717. The display
should change to:
---------- mA
A pressure value should appear on the screen after all of the calibration constants
have been down loaded.
4. Disconnect the Fluke Pressure Module and press
718 and 718Ex Verification
This procedure is appropriate for the 718 and 718Ex. When the steps vary, the step will
be labeled (718 Only) or (718Ex Only).
Oto turn the Fluke 717 off.
Pressure Verification
XWWarning
To avoid a violent release of pressure or vacuum, always
depressurize the system slowly using the pressure/vacuum
release control before detaching any pressure line.
47
Page 60
71X Series
Calibration Manual
1. Carefully attach the pressure fitting of the deadweight tester to the pressure port of
the Calibrator.
Note
The use of TEFLON tape at the pressure fitting strengthens the seal.
2. Press the green O key to turn the calibrator on.
3. The display should read 0.00 PSI with the deadweight tester opened up to ambient
air. If not, press the
Z key until display does read 0.00 PSI.
4. Depending if the instrument being tested is a 718(Ex) 1G, 30G, 100G, 300G, set up
the deadweight tester for the sequence of PSI inputs from Table 68 to be injected into
the pressure port of the Calibrator.
5. Ensure the pressure has stabilized at each input before verifying the display reading.
6. Carefully vent all pressure and disconnect the calibrator from the deadweight tester.
Note
When verifying vacuum pressure make sure the pressure/vacuum switch is
in the vacuum position. Forward (clockwise) is for pressure and backward
(counter-clockwise) is for vacuum.
48
Page 61
Process Calibrators
Table 68. 718 and 718Ex Pressure Verification
718 and 718Ex 1G 718 and 718Ex 30G
Input
Pressure
0 psi 0.0005 to -0.0005 0 psi -0.015 to 0.015
0.2 psi 0.2005 to 0.1995 6 psi 5.985 to 6.015
0.4 psi 0.4005 to 0.3995 12 psi 11.985 to 12.015
0.6 psi 0.6005 to 0.5995 18 psi 17.985 to 18.015
0.8 psi 0.8005 to 0.7995 24 psi 23.985 to 24.015
1 psi 1.0005 to 0.9995 30 psi 29.985 to 30.015
0.8 psi 0.8005 to 0.7995 24 psi 23.985 to 24.015
0.6 psi 0.6005 to 0.5995 18 psi 17.985 to 18.015
0.4 psi 0.4005 to 0.3995 12 psi 11.985 to 12.015
0.2 psi 0.2005 to 0.1995 6 psi 5.985 to 6.015
-1 psi -0.9995 to -1.0005 -12 psi * -12.015 to -11.985
718 and 718Ex 100G 718 and 718Ex 300G
Input
Pressure
0 psi -0.05 to 0.05 0 psi 0.15 to -0.15
20 psi 19.95 to 20.05 60 psi 60.15 to 59.85
40 psi 39.95 to 40.05 120 psi 120.15 to 119.85
60 psi 59.95 to 60.05 180 psi 180.15 to 179.85
80 psi 79.95 to 80.05 240 psi 240.15 to 239.85
100 psi 99.95 to 100.05 300 psi 300.15 to 299.85
80 psi 79.95 to 80.05 240 psi 240.15 to 239.85
60 psi 59.95 to 60.05 180 psi 180.15 to 179.85
40 psi 39.95 to 40.05 120 psi 120.15 to 119.85
20 psi 19.95 to 20.05 60 psi 60.15 to 59.85
* -12 psi -12.05 to -11.95 -12 psi -11.85 to -12.15
*Vacuum is only specified for V1.3 and later. To display the firmware version, start with the 717 off, press
and hold M, then press the power button.
Display
Reading
Display
Reading
Input
Pressure
Input
Pressure
Display
Reading
Display
Reading
Verification
Leak Test Verification
1. Seal off the pressure sensor input port.
2. Using the fine adjust knob, pressure-vacuum switch and internal pump, set the
calibrator to (approximately) the values in Table 69. Let the unit sit for one minute,
then record the displayed value.
3. Wait for one more minute, then record a second reading. The difference between the
first and the second reading is the leak rate.
49
Page 62
71X Series
Calibration Manual
Adjusted Pressure Maximum Leak Rate Adjusted Pressure Maximum Leak Rate
Table 69. 718 Leak Test Verification
718 1G 718 30G
-1.000 psi -10.000 psi
0.005 psi/min
718 100G 718 300G
Adjusted Pressure Maximum Leak Rate Adjusted Pressure Maximum Leak Rate
-10.00 psi -10.00 psi
3.00 psi 3.00 psi
100.00 psi
mA Measure Verification
0.10 psi/min
3.000 psi 1.000 psi
30.000 psi
100.00 psi
0.05 psi/min
0.3 psi/min
The display should indicate:
0.000mA 0.00 PSI
1. Connect a test lead from Red AUX terminal of the 5520A to the Fluke 8508A I+
terminal.
2. Connect a test lead from Black AUX terminal of the 5520A to the Calibrator Com
terminal.
3. Connect a test lead from I- terminal of the Fluke 8508A to the Calibrator mA
terminal.
4. Adjust the 5520A if necessary so that the current shown on the Fluke 8508A is the
same as the 5520A values shown in Table 70.
5. Verify that the display readings on the Calibrator are within the limits.
Table 70. 718 and 718Ex mA Measure Verification
Fluke 5520A Fluke 718 Fluke 718Ex
4.0000 mA 3.998 mA to 4.002 mA 3.997 mA to 4.003 mA
12.0000 mA 11.996 mA to 12.004 mA 11.996 mA to 12.004 mA
24.0000 mA 23.993 mA to 24.007 mA 23.993 mA to 24.007 mA
6. Disconnect the test leads and turn off the 718 Calibrator.
mA Loop Power Verification (718 Only)
1. Hold down both LOOP POWER keys ( U and O) at the same time until the
screen reads:
---------- LOOP mA
then release the keys.
2. Apply 1 kΩ to the mA jacks from the decade box or other resistor source. The
display should read over 21 mA but less than OL.
3. Apply 2 kΩ to the mA jacks from the decade box or other resistor source. The
display should read between 11 mA and 13 mA.
50
Page 63
Process Calibrators
4. PressOto turn the calibrator off.
Sensor Jack Verification
1. Make sure nothing is connected to the sensor port of the calibrator. Push
the calibrator on. The display should read:
Oto turn
Calibration
----------
then change to:
0.000
2. Plug the Fluke Pressure Module into the sensor port of the Fluke 718. The display
should change to:
----------
A pressure value should appear on the screen after all of the calibration constants
have been downloaded.
3. Disconnect the Fluke Pressure Module and push O to turn the calibrator off.
Display Verification
1. Press and hold the
the 718 in a mode where all display segments are on. All segments will stay on until
Z key is released.
the
2. Check to see that all segments of the display are showing.
3. Turn the calibrator off. The 718 and 718Ex verification test is now complete.
Z key and then press and release the green O key. This locks
Calibration
Re-calibration (obtaining new cal-constants) is required only if a calibrator fails
verification. Always re-verify after a re-calibration.
Throughout this section, the 71X Calibrator may be referred to as the
“UUT” (unit under test).
Preparing for Calibration
To prepare for calibration, do the following:
1. Make sure fuse(s) in the UUT are intact. By referring to Table 33, you should not
have to open the unit.
2. Make sure you have the required test equipment available. (Refer to Table 34.)
3. Turn on and warm up the test equipment for the time required.
4. Allow UUT to come to ambient temperature. Turn it on and allow 5 minutes for
warm-up of the UUT.
712 Calibration (V1.1 and Earlier)
Millivolts Measure
1. Hold down the
key.
R and the C key at the same time, then press and release the O
Note
2. When the display momentarily shows CAL , release both the
This puts the 712 into Cal Mode. The 712 should display:
INPUT 0.00 mV
51
R and the C keys.
Page 64
71X Series
Calibration Manual
3. Connect the test lead from the NORMAL jacks of the Fluke 5520A to the INPUT
jacks (two middle jacks) on the 712 Calibrator (black to black, red to red).
4. Set the Fluke 5520A to 0.0000 mV and press O. Press any key on the 712
Calibrator. The display should read:
INPUT ----- mV
then change to:
INPUT 0.00 mV
5. Press any key on the 712 Calibrator. The display should read:
INPUT ----- mV
then change to:
INPUT 160.00 mV
6. Set the Fluke 5520A to 160.0000 mV.
7. Press any key on the 712 Calibrator. The display should read:
INPUT ----- mV
then change to:
INPUT 320.00 mV
8. Set the Fluke 5520A to 320.0000 mV.
9. Press any key on the 712 Calibrator. The display should read:
INPUT ----- mV
then change to:
INPUT 640.00 mV
10. Set the Fluke 5520A to 640.000 mV.
11. Press any key on the 712 Calibrator. The display should read:
INPUT ----- mV
then change to:
INPUT 1280.00 mV
12. Set the Fluke 5520A to 1280.000 mV.
13. Press any key on the 712 Calibrator. The display should read:
INPUT ----- mV
then change to:
INPUT 400.00 R 4W
Resistance Measure
1. Connect two additional test leads from the NORMAL jacks of the 5520A to the 712
Calibrator (black wire to the 712 red jack labeled [3W NC] and red wire to the red
jack labeled [4W NC]. All four input jacks of the 712 should now be in a 4-wire
ohms measurement configuration.
52
2. Set the Fluke 5520A to 400.00 Ω and press O.
3. Press any key on the 712 Calibrator. The display should read:
INPUT ----- R 4W
then change to:
INPUT -----
Page 65
Process Calibrators
Calibration
mA Measure
1. Disconnect all four test leads from the Fluke 5520A and the 712 Calibrator.
2. Connect the two test leads from the AUX jacks on the Fluke 5520A to the middle
jacks on the 712 Calibrator (black to black; red to red).
3. Press any key on the 712 Calibrator. The display will change to:
INPUT 0.00 mA
4. Set the Fluke 5520A to 0.00000 mA and press O.
5. Press any key on the 712 Calibrator. The display should read:
INPUT ----- mA
then change to:
INPUT 0.00 mA
6. Again, press any key on the 712 Calibrator, still with a 0.00000 mA input from the
Fluke 5520A. The display should read:
INPUT ----- mA
then change to:
INPUT 2.000 mA
7. Set the Fluke 5520A to 2.00000 mA.
8. Press any key on the 712 Calibrator. The display should read:
INPUT ----- mA
then change to:
INPUT 0.500 mA
9. Set the Fluke 5520A to 0.50000 mA.
10. Press any key on the 712 Calibrator. The display should read:
INPUT ----- mA
11. Press any key on the 712 Calibrator and the unit will automatically reset itself, with
all segments momentarily being displayed. Calibrator is now out of CAL mode. Press
O key to turn the calibrator off and disconnect all test leads.
the
712 Calibration (V1.2 through V1.9)
1. Hold down R and C while turning the power on.
2. Hold the two keys until CAL is displayed, then release the two keys. The display
should read:
INPUT 15 Ω
3. Connect the two test leads from the NORMAL jacks of the 5520A to the INPUT
jacks (two middle jacks) on the 712 Calibrator (black to black, red to red).
4. Connect two additional test leads, also from the NORMAL jacks of the 5520A, to the
712 Calibrator (black wire to the 712 red jack labeled [3W NC] and red wire to the
red jack labeled [4W NC]). All four input jacks of the 712 should now be configured
in a 4-wire ohm measurement.
5. Set the Fluke 5520A to 15Ω and press O.
6. Press C. The display should read:
53
Page 66
71X Series
Calibration Manual
INPUT ----- Ω,
then change to:
INPUT 350 Ω
7. Set the Fluke 5520A to 350 Ω.
8. Press C. The display should read:
INPUT ----- Ω
then change to:
INPUT 500 Ω.
9. Set the Fluke 5520A to 500 Ω.
10. Press C. The display should read:
INPUT ----- Ω
then change to:
INPUT 3200 Ω
11. Set the Fluke 5520A to 3200 Ω.
12. Press C. The display should read:
INPUT ----- Ω.
then change to:
OUTPUT 17-50 Ω.
13. Using four test leads, connect the four jacks of Fluke 8508A (Ω Sense 'Hi & Lo' and
Input 'Hi & Lo') to the two middle jacks of the 712 Calibrator (black to black; red to
red). This puts the Fluke 8508A in a 4-wire configuration.
14. Set the 3458A to the 1000 Ω range.
15. Press ] and } until the 712 display matches the Fluke 8508A.
16. Press C. The display should read:
OUTPUT ----- Ω
then change to:
OUTPUT 350.00 0.
17. Press ] and } until the 712 display matches the Fluke 8508A.
18. Press C. The display should read:
OUTPUT ----- Ω
then change to:
OUTPUT 126.00 Ω
19. Put the Fluke 8508A in the 10 kΩ range.
54
20. Press ] and } until the 712 display matches the Fluke 8508A.
21. Press C. The display should read:
OUTPUT ----- Ω
then change to:
OUTPUT 2535.0 Ω
22. Press ] and } until the 712 display matches the Fluke 8508A.
23. Press C. The display should read:
Page 67
Process Calibrators
OUTPUT ----- Ω
then the unit will reset.
24. Press the power key to turn the unit off and disconnect the test leads.
712 Calibration (V2.0 and Later)
1. Hold down R and C while turning the power on.
2. The display shows CAL for 1 second and then xxx, where xxx is the number the
times the unit has been through a calibration adjustment. When the two keys are
released, PASS is displayed. One second later a zero is displayed for the most
significant of the three digit password. Use the two small arrows to change the
number. After setting each digit, press
712 will go into calibration. If the password was wrong, the 712 will go into normal
operation mode. The password is 217.
3. The calibration will be performed in the following order:
OHMS INPUT
OHMS OUTPUT low range
OHMS OUTPUT high range
OHMS INPUT
1. In OHMS INPUT mode, the unit will display “C 15” for 1 second and then display
the OHMS input using the default calibration.
C. If the password entry was correct, the
Calibration
2. Connect the test leads from the 5520 main Jacks to the 712 jacks in 4 wire
configuration (2 wires at 5520, 4 wires at 712). The unit should read 15.0 Ohms +/5%.
3. Press
4. Source 400 Ohms from the 5520. The unit should read 400.0 Ohms +/- 5%.
5. Press
6. Source 3200 Ohms from the 5520. The unit should read 3200.0Ohms +/- 5%.
7. Press
8. Press
OHMS OUTPUT Low Range
1. In OHMS OUTPUT low range mode, the unit will display “ 15.00” and will source
2. Connect the 712 to the DMM in OHMS read using 4 wire mode connection (2 wires
3. Adjust the value for OHMS OUTPUT using the arrows on the 712 keypad to source
C to go to the next calibration point. “C 400” will be displayed for 1 second
and then it will display the ohms input.
C to go to the next calibration point. “C3200” will be displayed for 1 second
and then it will display the ohms input.
C to complete OHMS INPUT calibration. “CHEC” will be displayed for 1
second. The new calibration constants are applied and the OHMS INPUT calibration
can be checked.
C to go to the next mode.
OHMS using the default calibration.
at 712, 4 wires at DMM). The DMM should read 15.00 Ohms +/- 5%.
exactly 15.00 Ohms.
4. Press
5. Adjust using the arrows on the keypad to source exactly 400.00 Ohms.
55
C to go to the next calibration point. “400.00” will be displayed
Page 68
71X Series
Calibration Manual
6. Press C to complete OHMS OUTPUT low range calibration. “CHEC” will be
displayed for 1 second. The new calibration constants are applied and the calibration
can be checked.
7. Press
C to go to the next mode.
OHMS OUTPUT High Range
1. In OHMS OUTPUT high range mode, the unit will display “ 20.0” and will source
OHMS using the default calibration. Keep the connections from the previous mode.
The DMM should read 20.0 Ohms +/- 5%.
2. Adjust the value for OHMS OUTPUT using the arrows on the 712 keypad to source
exactly 20.0 Ohms.
3. Press
C to go to the next calibration point. “4000.0” will be displayed.
4. Adjust using the arrows on the keypad to source exactly 4000.0 Ohms.
5. Press
C to complete OHMS OUTPUT high range calibration. “CHEC” will be
displayed for 1 second. The new calibration constants are applied and the calibration
can be checked.
6. Press
C to complete calibration.
713 Calibration
mA Measure Measure
1. On the 713 Pressure Calibrator hold down both the
release the O key.
2. When the display momentarily shows CAL, release the
puts the 713 into Cal Mode. The 713 should display:
U and D keys. Press and
U and the D keys. This
CAL 0.000 mA
3. Connect the test leads from the AUX jacks of the Fluke 5520A to the mA jacks on
the 713 Calibrator (black to black and red to red).
4. Set the Fluke 5520A to 0.00000 mA and press O.
5. Press any key on the 713 Calibrator. The display should read:
----- 0.000 mA
then change to:
CAL 24.000 mA
6. Set the Fluke 5520A to 24.0000 mA.
7. Press any key on the 713 Calibrator. The display should read:
----- 24.00 mA
then change to:
rAnGE
Pressure Measure
The 713 Pressure Calibrators have built-in temperature compensation. Instruments being
calibrated should be in a stable temperature environment for several minutes before
calibration. Calibration facilities should be maintained near 23 °C nominal.
Re-calibration (re-characterization) is performed in terms of ‘psi’ pressure units. Inputs using
other pressure units (ie. kPa and bar) must be mathematically converted.
56
Page 69
Process Calibrators
Calibration
1. Remove two-wire test lead from the 713 Calibrator and carefully attach the pressure
fitting of the deadweight tester to the pressure jack on the 713 Calibrator.
Note
The use of TEFLON tape at the pressure fittings strengthens the seal.
At this point, the display should be indicating “rAnGE”. This is an indication to
choose the correct model of the 713 Calibrator.
2. For 713 30G, press the
m key. For 713 100G, press the M key. The display will
flash for a short time:
------ rAnGE
then display:
CAL 0.00 PSI
3. Vent the system to ambient air to ensure 0.00 PSI.
4. Press any key on the 713 Calibrator. The display will change to read:
CAL ------ PSI
then change to:
CAL 30.000 PSI
or
CAL 100.00 PSI
depending on which model /range was selected in step 2.
5. Set up the deadweight tester for either 30 PSI or 100 PSI to be injected into the
pressure port of the 713 Calibrator.
6. Once the pressure has stabilized, press any key on the 713 Calibrator. The display
should read:
CAL ------ PSI
then change to:
CAL 15.000 PSI
or
CAL 50.00 PSI
depending on which model /range was previously selected.
7. Set the deadweight tester for either 15 PSI or 50 PSI.
8. Once the pressure has stabilized, press any key on the 713 Calibrator. The display
should read:
CAL ------ PSI
then change to:
0.000mA XX.xx PSI (where XX.xx is the applied pressure)
If the display reads
CALU, continue to step 9. Otherwise, the 713 Calibrator is no
longer in CAL mode, skip to step 11.
9. Press m to go to vacuum calibration. (Pressing any other key will end the
calibration.) If you chose to calibrate vacuum, the display should read:
CAL -12.000 PSI
10. Apply -12 PSI, wait for the vacuum standard to stabilize, then press any key. The
display should show:
-------------------
57
Page 70
71X Series
Calibration Manual
714 Calibration (Earlier than V2.0)
Temperature Measure
11. The unit will then reset power. The 713 is now out of the CAL mode and in the
normal mode of operation. Vent all pressure/vacuum.
1. On the 714 Calibrator press and hold the
T and C keys, then turn on the 714 by
pressing and releasing the O key.
2. When the display momentarily shows CAL, release both the
puts the 714 into the Cal Mode. The 714 should display:
INPUT 0.00 mV
3. Connect the two-wire copper test lead (banana jacks to mini-connector) from the
NORMAL jacks of the Fluke 5520A to the TC jack on the 714 Calibrator, observing
correct polarity.
4. Set the Fluke 5520A to 0.0000 mV and press O.
5. Press any key on the 714 Calibrator. The display should read:
INPUT -------- mV
then change to:
INPUT 77.00 mV
6. Set the Fluke 5520A to 77.0000 mV.
7. Press any key on the 714 Calibrator. The display should read:
INPUT -------- mV
then change to:
OUTPUT mV
Temperature Source
1. Remove the copper test lead from the 714 Calibrator.
T and C keys. This
Note
Copper wire must be used during this step.
58
2. Press any key on the 714 Calibrator. The display should read:
OUTPUT ------ mV
After a few seconds, the display should change to:
TYPE J 0.0 °C
Thermocouple Block Calibration
1. Connect the Type-J thermocouple test lead from the TC jack on the Fluke 5520A to
the TC jack on the 714 Calibrator, observing correct polarity.
Note
Type-J thermocouple wire (not copper) must be used.
2. Set the Fluke 5520A to source Type-J thermocouple at 0.0 °C. Press O.
3. Press any key on the 714 Calibrator. The display will very quickly flash:
TYPE J ------ °C
then read:
TYPE J HOLD °C
Page 71
Process Calibrators
Calibration
4. Wait for three to five minutes. The longer the waiting period, the more accurate the
Thermo-block calibration will be.
5. Press any key on the 714 Calibrator. The display should read:
TYPE J ------ °C
then change to:
TYPE J 0.xx °C
This number represents the absolute error of the "CJC" sensor and should be below
1°C.
6. Press any key on the 714 Calibrator. The 714 should reset itself, displaying all
segments momentarily. The display should then change to:
INPUT xx.x °C
where xx.x is some variable number; or OL (overload) may be indicated.
7. Press the O key to turn the calibrator off and remove the Type J thermocouple test
lead.
714 Calibration (V2.0 and Later)
1. On the 714 Calibrator press and hold the
T and C keys, then turn on the 714 by
pressing and releasing the O key.
2. The display shows CAL for 1 second and then xxx, where xxx is the number the
times the unit has been through a calibration adjustment. When the two keys are
released, PASS is displayed. One second later a zero is displayed for the most
significant of the three digit password. Use the two small arrows to change the
number. After setting each digit, press the °C/°F key. If the password entry was
correct, the 714 will go into calibration. If the password was wrong, the 714 will go
into normal operation mode. The password is 417.
3. The calibration will be performed in the following order:
mV OUTPUT
1. In mV OUTPUT mode, the unit will display “C -10” for 1 second followed by “ –
10.000”. The calibrator is now sourcing mV using the default calibration. The 714
should source approximately –10.000mV through copper wire.
2. Adjust the source value using the up and down big arrows for gross and small arrows
for fine, until you read -10.000mV.
3. Press
second and then “ 75.000”. The 714 should source approximately 75.000mV.
4. Adjust the source value until you read 75.000 mV on the DMM.
5. Press
mV OUTPUT
mV INPUT
CJC
C key to go to the next calibration point. “C 75” will be displayed for 1
C to complete mV OUTPUT calibration.
mV INPUT
1. In mV INPUT mode, the unit will display “C -10” for 1 second and then will display
the input voltage using the default calibration.
2. Source -10.000 mV from the 5520. The unit should read approximately –10.000 mV
through copper wire.
59
Page 72
71X Series
Calibration Manual
3. Press C to go to the next calibration point. “C 75” will be displayed for 1 second
and then the input voltage.
4. Source 75.000 mV from the 5520. The unit should read approximately 75.000mV .
5. Press
C to complete mV INPUT calibration. “CHEC” will be displayed for 1
second. The new calibration constants are applied and the mV INPUT calibration can
be checked.
6. Press
C to go to the next mode.
CJC
1. In CJC Cal Mode the unit will display “C CJC” for 1 second followed by IN 0°C for
another second. Next it will display the difference between the temperature as
measured by the CJC sensor and the temperature measured by a J type thermocouple.
2. Use the 5520 with J Type thermocouple wire to input a J type thermocouple signal of
0°C with CJC on.
3. When the number displayed on the 714 is stable, press the
calibration.
715 Calibration (Earlier than V2.0)
mA/Volts Measure
1. On the 715 Calibrator, hold down the
press and release the
O key.
2. When the display momentarily shows CAL , release the
CAL shows momentarily, the 715 will display:
INPUT 0.000 mA
3. Connect two test leads from the AUX jacks of the Fluke 5520A to the middle mA
input jacks on the 715 Calibrator (black to COM jack and red to mA jack).
C to complete the
V and the A keys at the same time, then
V and the A keys. After
4. Set the Fluke 5520A to 0.00000 mA and press O.
5. Press any key on the 715 Calibrator. The display should read:
INPUT ----- mA
then change to:
INPUT 24.000 mA
6. Set the Fluke 5520A to 24.0000 mA.
7. Press any key on the 715 Calibrator. The display should read:
INPUT ----- mA
then change to:
INPUT 0.000 V
8. Connect test lead from the NORMAL output jacks of the Fluke 5520A to the voltage
jacks on the 715 Calibrator (black to COM jack and red to the V jack).
9. Set the Fluke 5520A to 0.0000 mV and press O.
10. Press any key on the 715. A relay will click and the display should read:
INPUT ----- V
then change to:
INPUT 10.000 V
11. Set the Fluke 5520A to 10.00000 V.
60
Page 73
Process Calibrators
Calibration
12. Press any key on the 715 Calibrator. The display should read:
INPUT ----- V
then change to:
INPUT 0.00 mV
13. Set the Fluke 5520A to 0.0000 mV.
14. Press any key on the 715 Calibrator. The display should read:
INPUT ----- mV
then change to:
INPUT 100.00 mV
15. Set the Fluke 5520A to 100.0000 mV.
16. Press any key on the 715 Calibrator. The display should read:
INPUT ----- mV
then change to:
OUTPUT mV
mA/Volts Source Measure
1. Remove all test leads from the 715 Calibrator.
2. Press any key on the 715 Calibrator. Display should change to:
OUTPUT ----- mV
After a couple of seconds, display should change to:
OUTPUT ----- V
After a couple more seconds, display should change to:
OUTPUT mA
3.Take one single test lead and plug one end into the V jack and the other end into the
+LOOP jack of the 715 Calibrator (places a short between the two jacks).
4. Press any key on the 715 Calibrator. Display should read:
OUTPUT ----- mA
5. The 715 will then reset itself with all segments momentarily being displayed, then
change to:
OUTPUT 0.0000 V
The 715 is now out of the CAL mode. Turn the green O key to turn the calibrator off.
715 Calibration (V2.0 and Later)
1. On the 715 Calibrator, hold down the
press and release the
O key.
2. The display shows CAL for 1 second and then xxx, where xxx is the number the
times the unit has been through a calibration adjustment. When the two keys are
released, PASS is displayed. One second later, a zero is displayed for the most
significant of the three digit password. Use the two small arrows to change the
number. After setting each digit, press
will go into calibration mode and the letter “K” will appear on the display. If the
password was wrong, 715 will go into normal operation mode.
V and the A keys at the same time, then
A. If the password entry was correct, 715
61
Page 74
71X Series
Calibration Manual
V INPUT
3. The calibration will be performed in the following order:
V INPUT
mV INPUT
mA INPUT
V OUTPUT
mV OUTPUT
mA OUTPUT
1. In V INPUT mode, the unit displays “CAL 0” for 1 second followed by the input
voltage using the default calibration.
2. Connect the test leads from the 5520 main Jacks to the V and COM and source 0.0 V.
The unit should read 0.000V +/- 0.010V.
3. Press
4. Source 25V from the 5520. The unit should read 25.000V with 5% tolerance.
5. Press
6. Set the 5520 to 0.0 and press mA_% key to go to the next mode.
mV INPUT
1. In mV INPUT mode, the unit displays “CAL 0” for 1 second followed by the input
2. Source 0.0 mV from 5520. The unit should read 0.00mV +/- 0.10mV.
3. Press
4. Source 200mV from the 5520. The unit should read 200.00V with 5% tolerance.
5. Press
6. Press
mA INPUT
1. In mA INPUT mode, the unit displays “CAL 0” for 1 second followed by the input
A to go to the next calibration point. “CAL 25” will be displayed for 1
second and then the input voltage.
A to complete V INPUT calibration. “CHEC” will be displayed for 1
second. The new calibration constants are applied and the V INPUT calibration can
be checked.
voltage using the default calibration.
A to go to the next calibration point. “C 200” will be displayed for 1 second
and then the input voltage.
A to complete mV INPUT calibration. “CHEC” will be displayed for 1
second. The new calibration constants are applied and the mV INPUT calibration can
be checked.
A to go to the next mode
mA using the default calibration.
62
2. Move the red test lead to mA jack.
3. Source 0.0 mA from 5520 using the AUX jacks. The unit should read 0.000mA +/-
0.010mA.
4. Press
5. Source 24mA from the 5520. The unit should read 24.000mA with 5% tolerance.
6. Press
7. Set the 5520 to STBY mode and press
A to go to the next calibration point. “CAL24” will be displayed for 1
second and then the input mA.
A to complete mA INPUT calibration. “CHEC” will be displayed for 1
second. The new calibration constants are applied and the mA INPUT calibration can
be checked.
A to go to the next mode.
Page 75
Process Calibrators
V OUTPUT
1. In V OUTPUT mode, the unit displays “C 0” for 1 second followed by “ .0000”.
The Calibrator is now sourcing V using the default calibration.
2. Connect the test leads from V and COM to V and COM of ….DMM. The 715 should
source 0.000V +/- 0.010V.
3. Adjust the source value using the up and down big arrows for gross and small arrows
for fine adjustment, until you read 0.000V on the DMM.
Calibration
4. Press
and then “ .0000”. The 715 should source 20V with 5% tolerance.
5. Adjust the source value until you read 20.000 on the DMM.
6. Press
second. The new calibration constants are applied and the V OUTPUT calibration
can be checked.
7. Press
mV OUTPUT
1. In mV OUTPUT mode, the unit displays “C 0” for 1 second followed by “ 0.000”.
The calibrator is now sourcing mV using the default calibration. 715 should source
0.00mV +/- 0.10mV.
2. Adjust the source value using the up and down big arrows for gross and small arrows
for fine, until you read 0.00mV on the DMM.
3. Press
and then “ 0.000”. The 715 should source 200.00mV with 5% tolerance.
4. Adjust the source value until you read 200.00 mV on the DMM.
5. Press
second. The new calibration constants are applied and the mV OUTPUT calibration
can be checked.
6. Press
A to go to the next calibration point. “C 20” will be displayed for 1 second
A to complete V OUTPUT calibration. “CHEC” will be displayed for 1
A to go to the next mode.
A to go to the next calibration point. “C 200” will be displayed for 1 second
A to complete mV OUTPUT calibration. “CHEC” will be displayed for 1
A to go to the next mode.
mA OUTPUT
1. In mA OUTPUT mode, the unit displays “C 1” for 1 second followe by “ .0000”.
The calibrator is now sourcing mA using the default calibration.
2. Connect the test leads from LOOP (red) and V (blak) to mA and COM jacks of
….DMM. The 715 should source 1.000mA +/- 0.010mA.
3. Adjust the source value using the up and down big arrows for gross and small arrows
for fine adjustment, until you read 1.000mA on the DMM.
4. Press
5. Adjust the source value until you read 21.000mA on the DMM.
6. Press
7. Press
63
A to go to the next calibration point. “C 21” will be displayed for 1 second
and then “ .0000”. The 715 should source 21mA with 5% tolerance.
A to complete mA OUTPUT calibration. “CHEC” will be displayed for 1
second. The new calibration constants are applied and the mA OUTPUT calibration
can be checked.
A to complete the calibration and go to the normal mode.
Page 76
71X Series
Calibration Manual
716 Calibration
mA Measure
1.Hold down the U and the D keys at the same time, then press the green
O key.
When the display shows CAL momentarily, release the U and the D keys
immediately. After CAL shows, the display will automatically read:
CAL 0.000mA
2. Connect the test leads from the AUX jacks of the Fluke 5520A to the mA jacks on
the 716 Calibrator (Black to Black and Red to Red).
3. Set the Fluke 5520A to 0.00000 mA and press the O key.
4. Press any key on the 716 Calibrator. The display should change to:
------- 0.000mA
then to:
CAL 24.000mA
5. Set the Fluke 5520A to 24.0000 mA.
6. Press any key on the 716 Calibrator. Display should change to:
------- 24.000mA
The unit will reset itself. All segments will be displayed momentarily and change to:
------- mA
7. Turn the green O power key OFF and push the Y key on the Fluke 5520A.
717 Calibration (Earlier than V2.0)
mA Measure
1. Hold down the U and the D keys at the same time, then press the green
When the display shows CAL momentarily, release the U and the D keys
immediately. After CAL shows, the display will automatically read:
CAL 0.000mA
2. Connect the test leads from the AUX jacks of the Fluke 5520A to the mA jacks on
the 717 Calibrator (Black to Black and Red to Red).
3. Set the Fluke 5520A to 0.00000 mA and press O.
4. Press any key on the 717 Calibrator. The display should change to:
------- 0.000mA
then to:
CAL 24.000mA
5. Set the Fluke 5520A to 24.0000 mA.
6. Press any key on the 717 Calibrator. Display should change from:
------- 24.000mA
to:
O key.
64
rAnGE
Pressure Measure
1. Remove test leads from the 717 Calibrator and carefully attach the pressure fitting to
the pressure jack on the 717 Calibrator.
Page 77
Process Calibrators
Calibration
Note
The use of TEFLON tape at the pressure fittings strengthens the seal.
2. For the 717 100G press the M key. For the 717 30G, press the
display will flash for a short time:
------ rAnGE
then display:
CAL 0.00 PSI
3. Vent the system to ambient air to ensure 0.00 PSI.
4. Press any key on the 717 Calibrator. The display will change to read:
CAL ------ PSI
then depending on the model selected in step 2, change to either:
CAL 30.000 PSI
CAL 100.00 PSI
or
5. Set up the deadweight tester for 30 PSI (or 100 PSI) to be injected into the pressure
port of the 717 Calibrator.
6. Once the pressure has stabilized, press any key on the 717 Calibrator. The display
should read:
CAL ------ PSI
then depending on the model selected in step 2, change to either:
CAL 15.000 PSI
CAL 50.000 PSI
or
7. Set the deadweight tester for 15 PSI (or 50 PSI) to be injected into the pressure port
of the 717 calibrator.
m key. The
8. Once the pressure has stabilized, press any key on the 717 Calibrator. The display
should read:
CAL ------ PSI
then the calibrator will be out of CAL mode and in OPERATE mode. The display
should read:
0.000mA XX.xx PSI (where XX.xx is the applied pressure)
If the display reads
CALU, continue to step 9. If the 717 Calibrator is no longer in
CAL mode, skip to step 11.
9. Press m to go to vacuum calibration. (Pressing any other key will end the
calibration.) If you chose to calibrate vacuum, the display should read:
CAL -12.000 PSI
10. Apply -12 PSI, wait for the vacuum standard to stabilize, then press any key. The
display should show:
-------------------
11. The unit will then reset power. The 717 is now out of the CAL mode and in the
normal mode of operation. Vent all pressure/vacuum.
65
Page 78
71X Series
Calibration Manual
717 Calibration (V2.0 and Later)
mA Measure
Note
For versions 3.00 and later, chas changed to a.
1. Hold down both U and D on the 717 Pressure Calibrator. Press and release O.
2. The display shows CAL xxx, as long as Uand D are pressed. Where xxx is a
number, indicating the number of times the 717 has been through the calibration
adjustment. When Uand D are released,
PASS is displayed. The password is
required before you can enter the calibration mode. The password is 817. Enter the
most significant digit first. The password appears below PASS. Use c to increase
the displayed number, h to decrease it. When the required digit appears, press D
to proceed to the next digit, or the password check. If the password entry was
correct, the 717 will go to the CAL mode. If the password is incorrect, BADPASS
shows on the display and then the 717 returns to the normal operation mode. In the
CAL mode, the 717 should display:
0.000 mA
in the upper display and
CAL 0 mA
on the lower line of the display.
The upper display shows an uncalibrated reading. This reading is approximately
+/- 10 % of full scale of the applied reading. When D is pressed, the calibration
value is calculated relative to this reading. The lower display indicates what function
and point is being calibrated.
3. Connect the test leads from the AUX jacks of the Fluke 5520A to the mA jacks on
the 717 Calibrator (black to black and red to red).
4. Set the Fluke 5520A to 0.00000 mA and press O.
5. When 0.0 mA is sourced from the 5520A the upper display indicates about
"0.000 mA". When D is pushed, the unit establishes the calibration constant for
that point and the lower display changes to "CAL 24 mA". The upper display still
reflects the applied current so it does not change until after step 6 when 24 mA is
applied."
6. Set the Fluke 5520A to 24.0000 mA.
7. The upper display reads approximately 24.000 mA. Press D on the 717 Calibrator.
CHEC appears on the lower display.
8. The CHEC mode allows for a quick check of the calibration constants that have just
been stored. Changes to the applied current are reflected in the display. Press D to
continue. The lower display changes to:
rAnGE
66
Pressure Measure
The 717 Pressure Calibrator has built-in temperature compensation. Instruments being
calibrated should be in a stable temperature environment for several minutes before
calibration. Calibration facilities should be maintained near 23 °C nominal.
Re-calibration (re-characterization) is performed in terms of ‘PSI’ pressure units. Inputs
using other pressure units (ie. kPa and bar) must be mathematically converted.
Note
For versions 3.00 and later, chas changed to a.
1. Remove two-wire test lead from the 717 Calibrator and carefully attach the pressure
fitting of the deadweight tester to the pressure jack on the 717 Calibrator.
Page 79
Process Calibrators
Note
The use of TEFLON tape at the pressure fittings strengthens the seal.
2. The lower display shows rAnGE. This is an indication to choose the correct model of
the 717 Calibrator. Choose the correct range. The appropriate range for the 717 is as
follows:
Calibration
717 1G (1)
717 30G (30)
717 100G (100)
The upper display shows 30, which is the default pressure range. Pressing
2.0) changes the sensor range to 100 PSI, pressing
version 3.0 press a to uprange, hold to down range. Match the displayed range to the
measurement range of the 717.
3. After selecting the proper range, press
4. Vent the system to ambient air to ensure 0.00 PSI.
5. The lower display indicates C 0 PSI and the upper display indicates the current
reading. When the unit is vented (step 5) the upper display will read 0.00 +/approximately 10 % of full scale. When
the next pressure to apply. Apply the requested pressure that shows on the lower
display and when the upper reading is stable, press
appears on the display.
6. The CHEC mode allows for a quick check of the calibration constants that have just
been stored. Changes to the applied pressure reflect on the display. Press
continue. The unit will then reset power.
7. The 717 Calibrator is now out of the CAL mode and into the normal mode of
operation. Carefully vent all pressure, press O to turn the calibrator off, and
disconnect all pressure fittings.
717 300G (300)
717 500G (500)
717 1000G (1000)
h changes it back to 30 PSI. For a ≥
D on the 717 Calibrator.
D is pressed, the lower display indicates
717 1500G (1500)
717 3000G (3000)
717 5000G (5000)
c(≤ version
D. Repeat this until CHEC
D to
718 Calibration (Earlier than V2.0)
mA Measure
1. On the 718 Pressure Calibrator hold down both the
release the O key.
2. When the display momentarily shows CAL, release the
puts the 718 into Cal Mode. The 718 should display:
CAL 0.000 mA
3. Connect the test leads from the AUX jacks of the Fluke 5520A to the mA jacks on
the 718 Calibrator (black to black and red to red).
4. Set the Fluke 5520A to 0.00000 mA and press O.
5. Press any key on the 718 Calibrator. The display should read:
----- 0.000 mA
then change to:
CAL 24.000 mA
6. Set the Fluke 5520A to 24.0000 mA.
7. Press any key on the 718 Calibrator. The display should read:
----- 24.00 mA
67
U and D keys. Press and
U and the D keys. This
Page 80
71X Series
Calibration Manual
Pressure Measure
then change to:
rAnGE
The 718 Pressure Calibrators have built-in temperature compensation. Instruments being
calibrated should be in a stable temperature environment for several minutes before
calibration. Calibration facilities should be maintained near 23 °C nominal.
Re-calibration (re-characterization) is performed in terms of ‘psi’ pressure units. Inputs using
other pressure units (ie. kPa and bar) must be mathematically converted.
1. Remove two-wire test lead from the 718 Calibrator and carefully attach the pressure
fitting of the deadweight tester to the pressure jack on the 718 Calibrator.
Note
The use of TEFLON tape at the pressure fittings strengthens the seal.
2. At this point, the display should be indicating “rAnGE”. This is an indication to
choose the correct model of the 718 Calibrator.
For 718 30G, press the
m key. For 718 100G, press the M key. The display will
flash for a short time:
------ rAnGE
then display:
CAL 0.00 PSI
3. Vent the system to ambient air to ensure 0.00 PSI.
Press any key on the 718 Calibrator. The display will change to read:
CAL ------ PSI
then change to:
CAL 30.000 PSI
or
CAL 100.00 PSI
depending on which model /range was selected in step 2.
5. Set up the deadweight tester for either 30 PSI or 100 PSI to be injected into the
pressure port of the 718 Calibrator.
Once the pressure has stabilized, press any key on the 718 Calibrator. The display
should read:
CAL ------ PSI
then change to:
68
CAL 15.000 PSI
or
CAL 50.00 PSI
depending on which model /range was previously selected.
6. Set the deadweight tester for either 15 PSI or 50 PSI.
7. Once the pressure has stabilized, press any key on the 718 Calibrator. The display
should read:
CAL ------ PSI
then change to:
Page 81
Process Calibrators
CALU
Calibration
8. Release pressure by venting the system.
9. Press the
m key to go to vacuum calibration. Otherwise, press any key to finish
calibration. If calibrating vacuum, the display should show:
CAL -12.000 PSI
10. Make sure the pressure/vacuum switch is in the vacuum position. Forward
(clockwise) is for pressure and backward (counter-clockwise) is for vacuum.
11. Set up the deadweight tester to apply -12.000 PSI.
12. Wait for the pressure standard to stabilize, then press any key.
13. The display should show:
---------------------------
14. The unit will then reset power.
15. The 718 Calibrator is now out of the CAL mode and into the normal mode of
operation. Carefully vent all pressure, press the O key to turn the calibrator off, and
disconnect all pressure fittings.
718 Calibration (V2.0 and Later)
mA Measure
1.Hold down both U and D on the 718 Pressure Calibrator. Press and release O.
For versions 3.00 and later, c has changed to a.
2. The display shows CAL xxx, as long as Uand D are pressed. Where xxx is a
number, indicating the number of times the 718Ex has been through the calibration
adjustment. When Uand D are released,
required before you can enter the calibration mode. The password is 817. Enter the
most significant digit first. The password appears below
the displayed number, h to decrease it. When the required digit appears, press D
to proceed to the next digit, or the password check. If the password entry was
correct, the 718 will go to the CAL mode. If the password is incorrect,
shows on the display and then the 718Ex returns to the normal operation mode. In the
CAL mode, the 718 should display:
Note
PASS is displayed. The password is
PASS. Use c to increase
BAD PASS
0.000 mA
in the upper display and
CAL 0 mA
on the lower line of the display.
The upper display shows an uncalibrated reading. This reading is approximately
+/- 10 % of full scale of the applied reading. When D is pressed, the calibration
value is calculated relative to this reading. The lower display indicates what function
and point is being calibrated.
3. Connect the test leads from the AUX jacks of the Fluke 5520A to the mA jacks on
the 718 Calibrator (black to black and red to red).
4. Set the Fluke 5520A to 0.00000 mA and press O.
5. When 0.0 mA is sourced from the 5520A the upper display indicates about
"0.000 mA". When D is pushed, the unit establishes the calibration constant for
that point and the lower display changes to "CAL 24 mA". The upper display still
69
Page 82
71X Series
Calibration Manual
Pressure Measure
reflects the applied current so it does not change until after step 6 when 24 mA is
applied."
6. Set the Fluke 5520A to 24.0000 mA.
7. The upper display reads approximately 24.000 mA. Press D on the 718 Calibrator.
CHEC appears on the lower display.
8. The CHEC mode allows for a quick check of the calibration constants that have just
been stored. Changes to the applied current are reflected in the display. Press D to
continue. The lower display changes to:
rAnGE
The 718 Pressure Calibrator has built-in temperature compensation. Instruments being
calibrated should be in a stable temperature environment for several minutes before
calibration. Calibration facilities should be maintained near 23 °C nominal.
Re-calibration (re-characterization) is performed in terms of ‘PSI’ pressure units. Inputs
using other pressure units (ie. kPa and bar) must be mathematically converted.
Note
For versions 3.00 and later, chas changed to a.
1. Remove two-wire test lead from the 718 Calibrator and carefully attach the pressure
fitting of the deadweight tester to the pressure jack on the 718 Calibrator.
Note
The use of TEFLON tape at the pressure fittings strengthens the seal.
2. The lower display shows rAnGE. This is an indication to choose the correct model of
the 718 Calibrator. Choose the correct range. The appropriate range for the 718 is as
follows:
718 1G (1)
718 30G (30)
718 100G (100)
718 300G (300)
For unit with V2.0 or less
range. Pressing
cchanges the sensor range to 100 PSI, pressing h changes it back
, the upper display shows 30, which is the default pressure
to 30 PSI. For a≥ version 3.0 press ato uprange, hold to down range. Match the
displayed range to the measurement range of the 718.
3. After selecting the proper range, press D on the 718Ex Calibrator.
4. Vent the system to ambient air to ensure 0.00 PSI.
5. The lower display indicates C 0 PSI and the upper display indicates the current
reading. When the unit is vented (step 5) the upper display will read 0.00 +/approximately 10 % of full scale. When
D is pressed, the lower display indicates
the next pressure to apply. Apply the requested pressure that shows on the lower
display and when the upper reading is stable, press
D. Repeat this until CHEC
appears on the display.
70
6. The CHEC mode allows for a quick check of the calibration constants that have just
been stored. Changes to the applied pressure reflect on the display. Press
D to
continue. The unit will then reset power.
Page 83
Process Calibrators
Calibration
7. The 718 Calibrator is now out of the CAL mode and into the normal mode of
operation. Carefully vent all pressure, press O to turn the calibrator off, and
disconnect all pressure fittings.
718Ex Calibration
mA Measure
1. Hold down both U and D on the 718Ex Pressure Calibrator. Press and release O.
2. The display shows CAL xxx, as long as Uand D are pressed. Where xxx is a
number, indicating the number of times the 718Ex has been through the calibration
adjustment. When Uand D are released,
PASS is displayed. The password is
required before you can enter the calibration mode. The password is 817. Enter the
most significant digit first. The password appears below
PASS. Use c to increase
the displayed number, h to decrease it. When the required digit appears, press D
to proceed to the next digit, or the password check. If the password entry was
correct, the 718Ex will go to the CAL mode. If the password is incorrect,
BAD PASS
shows on the display and then the 718Ex returns to the normal operation mode. In the
CAL mode, the 718Ex should display:
0.000 mA
in the upper display and
CAL 0 mA
on the lower line of the display.
The upper display shows an uncalibrated reading. This reading is approximately
+/- 10 % of full scale of the applied reading. When D is pressed, the calibration
value is calculated relative to this reading. The lower display indicates what function
and point is being calibrated.
3. Connect the test leads from the AUX jacks of the Fluke 5520A to the mA jacks on
the 718Ex Calibrator (black to black and red to red).
4. Set the Fluke 5520A to 0.00000 mA and press O.
5. When 0.0 mA is sourced from the 5520A the upper display indicates about
"0.000 mA". When D is pushed, the unit establishes the calibration constant for
that point and the lower display changes to "CAL 24 mA". The upper display still
reflects the applied current so it does not change until after step 6 when 24 mA is
applied."
6. Set the Fluke 5520A to 24.0000 mA.
7. The upper display reads approximately 24.000 mA. Press D on the 718Ex
Calibrator. CHEC appears on the lower display.
8. The CHEC mode allows for a quick check of the calibration constants that have just
been stored. Changes to the applied current are reflected in the display. Press D to
continue. The lower display changes to:
rAnGE
Pressure Measure
The 718Ex Pressure Calibrator has built-in temperature compensation. Instruments being
calibrated should be in a stable temperature environment for several minutes before
calibration. Calibration facilities should be maintained near 23 °C nominal.
Re-calibration (re-characterization) is performed in terms of ‘PSI’ pressure units. Inputs
using other pressure units (ie. kPa and bar) must be mathematically converted.
71
Page 84
71X Series
Calibration Manual
1. Remove two-wire test lead from the 718Ex Calibrator and carefully attach the
pressure fitting of the deadweight tester to the pressure jack on the 718Ex Calibrator.
Note
The use of TEFLON tape at the pressure fittings strengthens the seal.
2. The lower display shows rAnGE. This is an indication to choose the correct model of
the 718Ex Calibrator. The appropriate range for the 718Ex 30G is 30 and 100 for the
718Ex 100G. Choose the correct range.
3. The upper display shows 30, which is the default pressure range. Pressing
changes the sensor range to 100 PSI, pressing h changes it back to 30 PSI.
4. After selecting the proper range, press
5. Vent the system to ambient air to ensure 0.00 PSI.
6. The lower display indicates C 0 PSI and the upper display indicates the current
reading. When the unit is vented (step 5) the upper display will read 0.00 +/approximately 10 % of full scale. When
the next pressure to apply. Apply the requested pressure that shows on the lower
display and when the upper reading is stable, press
appears on the display.
7. The CHEC mode allows for a quick check of the calibration constants that have just
been stored. Changes to the applied pressure reflect on the display. Press
continue. The unit will then reset power.
8. The 718Ex Calibrator is now out of the CAL mode and into the normal mode of
operation. Carefully vent all pressure, press O to turn the calibrator off, and
disconnect all pressure fittings.
D on the 718Ex Calibrator.
D is pressed, the lower display indicates
D. Repeat this until CHEC
c
D to
72
Page 85
Process Calibrators
Replacement Parts and Accessories
Replacement Parts and Accessories
Table 71 lists the replaceable parts and accessories for the 71X Calibrators. See Figure 7
for an exploded view of the 718 Calibrator. Fluke repair is recommended for the 718Ex
and no 718Ex replacement parts are listed in Table 71.
Note
Not all parts listed in Table 71 are shown in Figure 7.