Fluke 715 User Manual

715 V olt/mA Calibrator
Compact, lightweight, and easy to carry
+
LOOP
INPUT
OUTPUT
30V
MAX
VOLT/mA CALIBRATOR
715
V
mV
INPUT
COM
mA
30V
MAX
- mA SIMULATE, +
+ mA -
OUTPUT(30V MAX)
mA
30V
MAX
VOLTS
715 Volt/mA Calibrator
Measure loop current
(0-20 mA, 4-20 mA) signals with 0.015% accuracy and
0.001mA resolution Measure voltage output process
signals from PLCs, transmitters Source or simulate 24 mA
loop current Source voltage to 100 mV or 10V
24V loop supply with simultaneous
%
current measurement
General features
Compact and lightweight
Simple controls
Clear, 5-digit, liquid crystal display
Operating temperature:
-10°C to 55°C Safety: CSA C22.2 No. 1010.1:1992
Three-year warranty
Included:
V
– Fluke 715 Volt/mA Calibrator – Protective Yellow Holster with Flex-Stand – Test leads and alligator clips
General specifications
Maximum voltage: 30 volts: Non-operating temperature:
-40°C to 60°C
Operating temperature: -10°C to 55°C Relative humidity: 95% (10°C to 30°C);
75% (30°C to 40°C); 45% (40°C to 50°C); 35% (50°C to 55°C)
Operating altitude: 3,000m max Shock: 1m drop test Vibration: random, 2g, 5-500 Hz Safety: CSA C22.2 No. 1010.1:1992 EMC: EN50082-1:1992 and
EN55022:1994 Class B Size/weight: 187 mm L x 87 mm W x 32 mm D (7.35 in L x 3.41 in W x
1.25 in D) 330g (12 oz)
Size/weight (with holster and FlexStand™): 201mm L x 98 mm W x
52 mm D (7.93 in L x 3.86 in W x
2.06 in D) 600g (21 oz) Power: 9V battery ANSI/NEDA 1604A or IEC 6LR619V alkaline Battery Life: 4 to 20 hours, typical, depending on functions used
Warranty: Three years – Single 9V alkaline battery – Instructions (14 languages)
Ordering Information
Fluke-715 Volt/mA Calibrator
1-800-44-FLUKE
Toll-free in the U.S. (905) 890-7600 in Canada
(425) 356-5500 other countries
For more information to be sent to you by fax, call 1-800-FLUKE-FAX in the U.S. and Canada. Or, visit our Website at http://www.fluke.com
Specifications ( 18° to 28°C, one year)
Function Range Resolution Accuracy Notes
Loop supply 24 V dc N/A ±10% Drive; 20 mA into 1000
for battery >6.8V; 700 for battery 5.8 to 6.8V
Measure/simulate 0 to 24 mA 0.001 mA 0.015% + 2 counts Sink (simulate transmitter) mA/% of span (-25% to 125%) display mA or % of span
Measure/simulate 0 to 100 mV 0.01 mV 0.02% + 2 counts mV
Measure/simulate 0 to 10 V 0.001 V 0.02% + 2 counts V
Fluke. Keeping your world
up and running.
C0202UEN 8/98 Rev A
Loading...