Fluke 713 Instruction Sheet

Page 1
713
Pressure Calibrator
Instruction Sheet
Introduction
(hereafter called “calibrator”) calibrate P/I (pressure to current) transmitters, or measure pressure or current. The calibrator simultaneously displays pressure and current measurements. Full scale pressure input is as follows:
Model 713 30G: 30 psi (206.85 kPa, 2.0685 bar)
®
Model 713 100G: 100 psi (689.5 kPa, 6.895 bar) The calibrator measures and displays current up to 24 mA, and
the pressure applied to the 1/8-inch NPT pressure fitting. The calibrator measures pressure in the units shown in the next table.
Your calibrator is supplied with a Flex-Stand holster, an installed 9 V alkaline battery, one set of TL75 test leads, one set of AC70A alligator clips, and this instruction sheet.
If the calibrator is damaged or something is missing, contact the place of purchase immediately. Contact your Fluke distributor for information about accessories. See “How to Contact Fluke.” To order replacement parts or spares, see “Replacement Parts.”
PN 650298 August 1997 Rev. 1, 10/97 1997 Fluke Corporation. All rights reserved. Printed in U.S.A . All product names are trademarks of t hei r respect iv e com pani es.
Page 2
Pressure Range and Resolution
Displayed
Pressure Units
psi 30.000 psi 100.00 psi inH2O at 4°C 830.4 inH2O 2768.0 inH2O inH2O at 20°C 831.9 inH2O 2772.9 inH2O cmH2O at 4°C, 2109.0 cmH2O 7030 cmH2O cmH2O at 20°C 2113.0 cmH2O 7043 cmH2O bar 2.0685 bar 6.895 bar mbar 2068.5 mbar 6895 mbar kPa 206.85 kPa 689.5 kPa inHg 61.080 inHg 203.6 inHg
Model 713 30G
Range and Resolution
Model 713 100G
Range and Resolution
mmHg 1551.3 mmHg 5171 mmHg kg/cm
2
2.1090 kg/cm
2
7.030 kg/cm
2
Safety Information
Use the calibrator only as specified in this instruction sheet, otherwise the protection provided by the calibrator may be impaired.
A Warning identifies conditions and actions that pose hazard(s) to the user; a Caution identifies conditions and actions that may damage the calibrator or the equipment under test.
W Warning
To avoid possible electric shock or personal injury:
Never apply more than 30 V between the mA
terminals, or between either of the mA terminals and earth ground.
Page 3
Remove the test leads from the calibrator
before you open the battery door.
Make sure the battery door is closed and
latched before you operate the calibrator.
Do not operate the calibrator if it is damaged.
Do not operate the calibrator around explosive
gas, vapor, or dust.
To avoid a violent release of pressure in a
pressurized system, shut off the valve and slowly bleed off the pressure before you attach or detach the calibrator pressure fitting to the pressure line.
When servicing the calibrator, use only
specified replacement parts
Caution
To avoid overpressure damage, do not apply pressure that exceeds the following:
Model 713 30G: 3X top of range (90 psi, 620 kPa,
6.2 bar)
Model 713 100G: 2X top of range (200 psi, 1380
kPa, 13.8 bar)
To avoid corrosion in the pressure sensor, use the calibrator only with media compatible with glass, ceramic, silicon, RTV, type 316 stainless steel, and nickel.
To avoid mechanically damaging the calibrator, do not apply torque between the pressure fitting and the calibrator case. See the figure for the proper use of tools.
Page 4
Hold in
fixed
position
kf01f.eps
Explanation of International Symbols
The symbols in the table below are used on the calibrator or in this instruction sheet.
Internatio na l Symbols
Symbol Meaning
J
I
M
W
T
P
Earth ground Fuse
Battery Refer to this instructi on sheet for information about
this feature. Double insulated
Conforms to relevant Canadian Standards Association directi ves.
Conforms to European Union directiv es
Page 5
Getting Acquainted with the Calibrator
Press the green O pushbutton to turn the calibrator on and off. The calibrator displays pressure and current measurements simultaneously. See the figure below.
The upper part of the display shows the gage pressure applied to the pressure fitting. Press
to select a different pressure unit.
U
When you cycle the power off and on, th e calibrator retains the pressure unit you last used.
The lower part of the display shows the current (up to 24 mA) applied to the current (mA) inputs. The current inputs are fused with a 0.125 A, 250 V fast fuse (Littelfuse type 2AG).
PRESSURE
713 100G
CALIBRATOR
HOLD
Pressure
Current mA
measurement
ZERO
MIN
MAX
DAMPUNITS
CLR
HOLD
measurement
Pressure
fitting
RANGE
0-100psi 0-689kPa
200psi 1378kPa
MAX
Current input
kf05f.eps
Page 6
Pushbutton Functions
Pushbutton Function
U
D
Z
m
c
M
Selects a different press ure unit.
Turns pressure reading damping on and off. With damping on, the display does not update as quickly.
Press this to zero the pressure displ ay. Vent the pressure fitting on the cali brator to atmosphere before you press this pushbutton.
Press and hold to read the minimum pressure and current readings since the power was turned on or
was pressed.
c
Press to clear the MIN and MAX memories.
Press and hold to read the maximum pressure and current readings the power was turned on or was pressed.
c
h
Press
g
again to resume normal operation.
to freeze the display. The
h
symbol appears on the display. Press
h
Calibrating a P/I Transmitter
To calibrate a P/I (pressure to current) transmitter, you apply a
pressure to the transmitter and measure the transmitter’s current loop output. Connect the calibrator to the transmitter as shown in the figure.
Page 7
SIGNAL
713 100G
PRESSURE CALIBRATOR
DAMPUNITS
+
TEST
ZERO
MIN
MAX
CLR
HOLD
kf02f.eps
Page 8
Maintenance
For maintenance procedures not described in this instruction sheet, or if the calibrator needs repair, contact a Fluke Service Center.
In Case of Difficulty
Check the battery, test leads, and pressure tubing. Replace as necessary.
Review this instruction sheet to make sure you are using the calibrator correctly.
If the calibrator needs repair, and the calibrator is under warranty, see the warranty statement for terms. If the warranty has lapsed, the calibrator will be repaired and returned for a fixed fee.
Cleaning
Periodically wipe the case with a damp cloth and detergent; do not use abrasives or solvents.
Calibration
Calibrate your calibrator once a year to ensure that it performs according to its specifications. A calibration manual is available (PN 686540). Call 1-800-526-4731 from the USA and Canada. In other countries, contact a Fluke Service Center.
Page 9
Replacing the Battery
When the Bsymbol appears on the display, replace the battery with a 9 V alkaline battery.
SIGNAL
713 100G
PRESSURE CALIBRATOR
DAMPUNITS
+
TEST
ZERO
MIN
MAX
CLR
HOLD
it07f.eps
Page 10
Replacing the Fuse
! Warning
To avoid personal injury or damage to the calibrator, use only a 0.125A 250V fast fuse, Littelfuse
Fuse F1 is probably blown if the mA measurement display reading does not respond to current applied to the current (mA) inputs.
Replace the fuse as follows:
1. Remove the test leads and turn the calibrator off.
2. Remove the battery door.
3. Remove the three Phillips-head screws from the case bottom and turn the case over.
4. Gently lift the top cover from the end nearest the current
2AG.
(mA) inputs until it unsnaps from the bottom cover.
5. Replace the fuse with a 0.125 A 250 V fast fuse, Littelfuse type 2AG.
6. Carefully fit the case top and circuit board assembly together, making sure that the O-ring is properly seated between the pressure sensor and the pressure fitting on the case top. Fit the case bottom onto the case top, engaging the two snaps near the display end of the case. Reinstall the three screws.
7. Replace the battery door.
Page 11
F1
HOLD
CLR
ZERO
MAX
MIN
UNITS
DAMP
Snaps
kf03f.eps
Page 12
Replacement Parts and Accessories
Replacement Parts
Part or
Item Description
model
no.
Qty.
BT1 9V battery, ANSI/NE DA 1604A or
IEC 6LR61
CG81Y Holster, Yellow CG81Y 1
!
F1 Fuse, 125 mA, 250V fast 686527 1
MP85 713 30G case top 620218 1 MP85 713 100G case top 620226 1 MP86 Case bottom 620168 1 H2, 3, 4 Case screw 832246 3 MP89, 90 Non-skid foot 824466 2 OR1 O-ring for pressure input 146688 1 SP1 Spacer for pressure input 687449 MP92 Battery door 619947 1
614487 1
H5, 6 Battery door fasteners 948609 2 S1 Keypad 687068 1 TL75 Test lead set TL75 1 AC70A Alligator clips AC70A 1
713 Instruction Sheet 650298 1 TL20 Industrial test lead set TL20 Option
71X Series Calibration Manual 686540 Option
Page 13
MP85
BT1
MP86
S1
F1
OR1
SP1
H2, 3, 4
H5, 6
Instruction
Sheet
MP89, 90
MP92
AC70A
Alligator Clips
TL75
Test Lead Set
Holster
kf04c.eps
Page 14
Specifications
Specificatio ns are based on a one year calibrati on cycle and apply for ambient temperature from +18 °C to +28 °C unles s
stated otherwise. “Counts” are the number of increments or decrements of the least significant digit.
Pressure Input, 713 30G
Range Accuracy
30 psi (206.85 kPa) ±0.05% of range
Maximum nondestructive press ure: 3X top of range (90 psi, 620 kPa,
6.2 bar) Temperature coefficient: 0.01% of range per
-10
°
C to 18 °C and 28 °C to 55 °C
°
C for temperature ranges
Pressure Input, 713 100G
Range Accuracy
100 psi (689.5 kPa) ±0.05% of range
Maximum nondestructive press ure: 2X top of range (200 psi, 1380 kPa,
13.8 bar) Temperature coefficient: 0.01% of range per
-10
°
C to 18 °C and 28 °C to 55 °C
°
C for temperature ranges
DC mA Input, 713 30G and 713 100G
Range Resolution Accuracy,
±(% of Reading + Counts)
24 mA 0.001 mA 0.025 + 1
Overload protection: 125 mA, 250V fast acting fuse Temperature coefficient: 0.005% of range per
ranges -10
°
C to 18 °C and 28 °C to 55 °C
°
C for temperature
Page 15
General Specifications
Maximum voltage applied between either mA terminal and earth ground or between the mA terminals: 30 V
Storage temperature: -40°C to 60°C Operating temperature: -10°C to 55°C Operating altitude: 3000 meters maximum Relative humidity: 95% up to 30°C, 75% up to 40°C, 45% up to
50°C, and 35% up to 55°C
Vibration: Random 2 g, 5 Hz to 500 Hz Shock: 1 meter drop test Safety: Certified as compliant to CAN/CSA C22.2 No.
1010.1:1992. Complies with ANSI/ISA S82.01-1994.
Power requirements: Single 9 V battery (ANSI/NEDA 1604A or IEC 6LR61)
Size: 32 mm H x 87 mm W x 187 mm L (1.25 in H x 3.41 in W x
7.35 in L);
With holster and Flex-Stand: 52 mm H x 98 mm W x 201 mm L (2.06 in H x 3.86 in W x 7.93 in L)
Weight: 369 g (13 oz); With holster and Flex-Stand: 624 g (22 oz)
Page 16
How to Contact Fluke
To order accessories, receive operating assistance, or get the location of the nearest Fluke distributor or Service Center, call:
1-800-44FLUKE (1-800-443-5853) in U.S.A. and Canada +31-402-678-200 in Europe +1-425-356-5500 from other countries Address correspondence to:
Fluke Corporation Fluke Europe B.V. P.O. Box 9090, P.O. Box 1186, Everett, WA 98206-9090 5602 BD Eindhoven USA The Netherlands
Visit us on the World Wide Web at: www.fluke.com
LIMITED WARRANTY & LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and workmanship for three years from the date of purchase. This warranty does not cover fuses, disposable batteries or damage from accident, neglect, misuse or abnormal conditions of operation or handling. Resell ers are not
authorized to extend any other warranty on Fluke’s behalf. To obtain service during the warranty peri od, send your defect ive calibrator to the nearest Fluke Authorized Serv ice Center wi th a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusi on or limitation of an implied warranty or of incidental or consequent ial damages, this limitation of liability may not apply to you.
Loading...