Fluke Model 713 30G and 713 100G Pressure Calibrators
(hereafter called “calibrator”) calibrate P/I (pressure to current)
transmitters, or measure pressure or current. The calibrator
simultaneously displays pressure and current measurements.
Full scale pressure input is as follows:
•Model 713 30G: 30 psi (206.85 kPa, 2.0685 bar)
®
•Model 713 100G: 100 psi (689.5 kPa, 6.895 bar)
The calibrator measures and displays current up to 24 mA, and
the pressure applied to the 1/8-inch NPT pressure fitting. The
calibrator measures pressure in the units shown in the next table.
Your calibrator is supplied with a Flex-Stand holster, an
installed 9 V alkaline battery, one set of TL75 test leads, one set
of AC70A alligator clips, and this instruction sheet.
If the calibrator is damaged or something is missing, contact the
place of purchase immediately. Contact your Fluke distributor for
information about accessories. See “How to Contact Fluke.” To
order replacement parts or spares, see “Replacement Parts.”
PN 650298 August 1997 Rev. 1, 10/97
1997 Fluke Corporation. All rights reserved. Printed in U.S.A .
All product names are trademarks of t hei r respect iv e com pani es.
Page 2
Pressure Range and Resolution
Displayed
Pressure Units
psi30.000 psi100.00 psi
inH2O at 4°C830.4 inH2O2768.0 inH2O
inH2O at 20°C831.9 inH2O2772.9 inH2O
cmH2O at 4°C,2109.0 cmH2O7030 cmH2O
cmH2O at 20°C2113.0 cmH2O7043 cmH2O
bar2.0685 bar6.895 bar
mbar2068.5 mbar6895 mbar
kPa206.85 kPa689.5 kPa
inHg61.080 inHg203.6 inHg
Model 713 30G
Range and
Resolution
Model 713 100G
Range and
Resolution
mmHg1551.3 mmHg5171 mmHg
kg/cm
2
2.1090 kg/cm
2
7.030 kg/cm
2
Safety Information
Use the calibrator only as specified in this instruction sheet,
otherwise the protection provided by the calibrator may be
impaired.
A Warning identifies conditions and actions that pose hazard(s)
to the user; a Caution identifies conditions and actions that may
damage the calibrator or the equipment under test.
W Warning
To avoid possible electric shock or personal injury:
•Never apply more than 30 V between the mA
terminals, or between either of the mA terminals
and earth ground.
Page 3
•Remove the test leads from the calibrator
before you open the battery door.
•Make sure the battery door is closed and
latched before you operate the calibrator.
• Do not operate the calibrator if it is damaged.
• Do not operate the calibrator around explosive
gas, vapor, or dust.
•To avoid a violent release of pressure in a
pressurized system, shut off the valve and
slowly bleed off the pressure before you attach
or detach the calibrator pressure fitting to the
pressure line.
•When servicing the calibrator, use only
specified replacement parts
Caution
To avoid overpressure damage, do not apply
pressure that exceeds the following:
•Model 713 30G: 3X top of range (90 psi, 620 kPa,
6.2 bar)
•Model 713 100G: 2X top of range (200 psi, 1380
kPa, 13.8 bar)
To avoid corrosion in the pressure sensor, use the
calibrator only with media compatible with glass,
ceramic, silicon, RTV, type 316 stainless steel, and
nickel.
To avoid mechanically damaging the calibrator, do
not apply torque between the pressure fitting and
the calibrator case. See the figure for the proper
use of tools.
Page 4
Hold in
fixed
position
kf01f.eps
Explanation of International Symbols
The symbols in the table below are used on the calibrator or in
this instruction sheet.
Internatio na l Symbols
SymbolMeaning
J
I
M
W
T
P
Earth ground
Fuse
Battery
Refer to this instructi on sheet for information about
this feature.
Double insulated
Conforms to relevant Canadian Standards
Association directi ves.
Conforms to European Union directiv es
Page 5
Getting Acquainted with the Calibrator
Press the green O pushbutton to turn the calibrator on and off.
The calibrator displays pressure and current measurements
simultaneously. See the figure below.
The upper part of the display shows the gage pressure applied to
the pressure fitting. Press
to select a different pressure unit.
U
When you cycle the power off and on, th e calibrator retains the
pressure unit you last used.
The lower part of the display shows the current (up to 24 mA)
applied to the current (mA) inputs. The current inputs are fused
with a 0.125 A, 250 V fast fuse (Littelfuse type 2AG).
PRESSURE
713 100G
CALIBRATOR
HOLD
Pressure
Current mA
measurement
ZERO
MIN
MAX
DAMPUNITS
CLR
HOLD
measurement
Pressure
fitting
RANGE
0-100psi 0-689kPa
200psi 1378kPa
MAX
Current input
kf05f.eps
Page 6
Pushbutton Functions
PushbuttonFunction
U
D
Z
m
c
M
Selects a different press ure unit.
Turns pressure reading damping on and off. With
damping on, the display does not update as
quickly.
Press this to zero the pressure displ ay. Vent the
pressure fitting on the cali brator to atmosphere
before you press this pushbutton.
Press and hold to read the minimum pressure and
current readings since the power was turned on or
was pressed.
c
Press to clear the MIN and MAX memories.
Press and hold to read the maximum pressure and
current readings the power was turned on or
was pressed.
c
h
Press
g
again to resume normal operation.
to freeze the display. The
h
symbol appears on the display. Press
h
Calibrating a P/I Transmitter
To calibrate a P/I (pressure to current) transmitter, you apply a
pressure to the transmitter and measure the transmitter’s current
loop output. Connect the calibrator to the transmitter as shown in
the figure.
Page 7
SIGNAL
713 100G
PRESSURE
CALIBRATOR
DAMPUNITS
+
TEST
–
ZERO
MIN
MAX
CLR
HOLD
kf02f.eps
Page 8
Maintenance
For maintenance procedures not described in this instruction
sheet, or if the calibrator needs repair, contact a Fluke Service
Center.
In Case of Difficulty
•Check the battery, test leads, and pressure tubing. Replace
as necessary.
•Review this instruction sheet to make sure you are using the
calibrator correctly.
If the calibrator needs repair, and the calibrator is under
warranty, see the warranty statement for terms. If the warranty
has lapsed, the calibrator will be repaired and returned for a fixed
fee.
Cleaning
Periodically wipe the case with a damp cloth and detergent; do
not use abrasives or solvents.
Calibration
Calibrate your calibrator once a year to ensure that it performs
according to its specifications. A calibration manual is available
(PN 686540). Call 1-800-526-4731 from the USA and Canada. In
other countries, contact a Fluke Service Center.
Page 9
Replacing the Battery
When the Bsymbol appears on the display, replace the battery
with a 9 V alkaline battery.
SIGNAL
713 100G
PRESSURE
CALIBRATOR
DAMPUNITS
+
TEST
–
ZERO
MIN
MAX
CLR
HOLD
it07f.eps
Page 10
Replacing the Fuse
! Warning
To avoid personal injury or damage to the
calibrator, use only a 0.125A 250V fast fuse,
Littelfuse
Fuse F1 is probably blown if the mA measurement display
reading does not respond to current applied to the current (mA)
inputs.
Replace the fuse as follows:
1. Remove the test leads and turn the calibrator off.
2. Remove the battery door.
3. Remove the three Phillips-head screws from the case
bottom and turn the case over.
4. Gently lift the top cover from the end nearest the current
2AG.
(mA) inputs until it unsnaps from the bottom cover.
5. Replace the fuse with a 0.125 A 250 V fast fuse, Littelfuse
type 2AG.
6. Carefully fit the case top and circuit board assembly
together, making sure that the O-ring is properly seated
between the pressure sensor and the pressure fitting on the
case top. Fit the case bottom onto the case top, engaging
the two snaps near the display end of the case. Reinstall the
three screws.
7. Replace the battery door.
Page 11
F1
HOLD
CLR
ZERO
MAX
MIN
UNITS
DAMP
Snaps
kf03f.eps
Page 12
Replacement Parts and Accessories
Replacement Parts
Part or
ItemDescription
model
no.
Qty.
BT19V battery, ANSI/NE DA 1604A or
IEC 6LR61
CG81YHolster, YellowCG81Y1
!
F1Fuse, 125 mA, 250V fast6865271
MP85713 30G case top6202181
MP85713 100G case top6202261
MP86Case bottom6201681
H2, 3, 4Case screw8322463
MP89, 90Non-skid foot8244662
OR1O-ring for pressure input1466881
SP1Spacer for pressure input687449
MP92Battery door6199471
6144871
H5, 6Battery door fasteners9486092
S1Keypad6870681
TL75Test lead setTL751
AC70AAlligator clipsAC70A1
−713 Instruction Sheet6502981
TL20Industrial test lead setTL20Option
−71X Series Calibration Manual686540Option
Page 13
MP85
BT1
MP86
S1
F1
OR1
SP1
H2, 3, 4
H5, 6
Instruction
Sheet
MP89, 90
MP92
AC70A
Alligator Clips
TL75
Test Lead Set
Holster
kf04c.eps
Page 14
Specifications
Specificatio ns are based on a one year calibrati on cycle and
apply for ambient temperature from +18 °C to +28 °C unles s
stated otherwise. “Counts” are the number of increments or
decrements of the least significant digit.
Pressure Input, 713 30G
RangeAccuracy
30 psi (206.85 kPa)±0.05% of range
Maximum nondestructive press ure: 3X top of range (90 psi, 620 kPa,
6.2 bar)
Temperature coefficient: 0.01% of range per
-10
°
C to 18 °C and 28 °C to 55 °C
°
C for temperature ranges
Pressure Input, 713 100G
RangeAccuracy
100 psi (689.5 kPa)±0.05% of range
Maximum nondestructive press ure: 2X top of range (200 psi, 1380 kPa,
13.8 bar)
Temperature coefficient: 0.01% of range per
-10
°
C to 18 °C and 28 °C to 55 °C
°
C for temperature ranges
DC mA Input, 713 30G and 713 100G
RangeResolutionAccuracy,
±(% of Reading + Counts)
24 mA0.001 mA0.025 + 1
Overload protection: 125 mA, 250V fast acting fuse
Temperature coefficient: 0.005% of range per
ranges -10
°
C to 18 °C and 28 °C to 55 °C
°
C for temperature
Page 15
General Specifications
Maximum voltage applied between either mA terminal and
earth ground or between the mA terminals: 30 V
Storage temperature: -40°C to 60°C
Operating temperature: -10°C to 55°C
Operating altitude: 3000 meters maximum
Relative humidity: 95% up to 30°C, 75% up to 40°C, 45% up to
50°C, and 35% up to 55°C
Vibration: Random 2 g, 5 Hz to 500 Hz
Shock: 1 meter drop test
Safety: Certified as compliant to CAN/CSA C22.2 No.
1010.1:1992. Complies with ANSI/ISA S82.01-1994.
Power requirements: Single 9 V battery (ANSI/NEDA 1604A or
IEC 6LR61)
Size: 32 mm H x 87 mm W x 187 mm L (1.25 in H x 3.41 in W x
7.35 in L);
With holster and Flex-Stand: 52 mm H x 98 mm W x 201 mm L
(2.06 in H x 3.86 in W x 7.93 in L)
Weight: 369 g (13 oz);
With holster and Flex-Stand: 624 g (22 oz)
Page 16
How to Contact Fluke
To order accessories, receive operating assistance, or get the
location of the nearest Fluke distributor or Service Center, call:
1-800-44FLUKE (1-800-443-5853) in U.S.A. and Canada
+31-402-678-200 in Europe
+1-425-356-5500 from other countries
Address correspondence to:
Fluke CorporationFluke Europe B.V.
P.O. Box 9090,P.O. Box 1186,
Everett, WA 98206-9090 5602 BD Eindhoven
USAThe Netherlands
Visit us on the World Wide Web at: www.fluke.com
LIMITED WARRANTY & LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and workmanship
for three years from the date of purchase. This warranty does not cover
fuses, disposable batteries or damage from accident, neglect, misuse or
abnormal conditions of operation or handling. Resell ers are not
authorized to extend any other warranty on Fluke’s behalf. To obtain
service during the warranty peri od, send your defect ive calibrator to the
nearest Fluke Authorized Serv ice Center wi th a description of the
problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER
WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE,
ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY
SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES
OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some
states or countries do not allow the exclusi on or limitation of an implied
warranty or of incidental or consequent ial damages, this limitation of
liability may not apply to you.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.