Fluke 712B RTD, 714B Data Sheet

Fluke 712B and
714B Temperature
Calibrators
Accuracy and Simplicity
For the temperature calibration professional that wants a highly accurate, easy-to-use, single function temperature calibrator the 712B and 714B are ideal test tools.
Technical Data
Product Highlights
The 712B can measure and simulate (13) different RTD types and resistance
The 714B can measure and simulate (17) different thermocouple types and millivolts
Measure 4 to 20 mA signals while simultaneously sourcing a temperature signal
Hanging tool designed in and included with every unit
Configurable 0 source settings for quick 25 linearity checks
Linear ramp and 25 ramp based on 0 settings
Dual inputs and backlit display for easy interpretation of measurements
Power down settings remembered at power up for easy restart of tests
1-year and 2-year specifications and traceable certificate of calibration
% and 100 %
% step auto
% and 100 %
%
Specifications
Specifications are based on a one year calibration cycle and apply from +18 °C to +28 °C unless stated otherwise. All specifications assume a 5-minute warmup period.
General specifications
Maximum voltage applied
30 V
between any terminal and earth ground or between any two terminals:
Operating temperature -10 °C to 50 °C
Storage temperature - 30 °C to 60 °C
Operating altitude 2,000 meters
Storage altitude 12,000 meters
Relative humidity (% RH operating without condensation)
Non condensing 90 % (10 °C to 30 °C) 75 % (30 °C to 40 °C) 45 % (40 °C to 50 °C) (Without condensation)
Vibration requirements MIL-T-28800E, Class 2
Drop text requirements 1 meter
IP Rating IEC 60529: IP52
Electromagnetic
IEC 61326-1, Portable
environment
Safety IEC 61010-1, Max 30 V to
earth, Pollution Degree 2
Power supply 4 AA NEDA 1.5A IEC LR6
batteries
Size (H x W x L) 52.5 mm x 84 mm x 188.5 mm
Weight 515 g
DC mA measurement
Resolution Range Accuracy
0-24 mA 0.001 mA 0.010 % + 2 µA
Temperature coefficient: ± (0.002 % of reading + 0.002 % of range) /°C ( <18 °C or >28 °C)
(% of reading + counts)
Ohms measurement (Fluke 712B)
Ohms range Accuracy
0.00 Ω to 400.00 Ω 0.015 % + 0.05 Ω
400.0 Ω to 4000.0 Ω 0.015 % + 0.5 Ω
Note: Read accuracy is based on 4-wire input. For 3-wire ohm measurements, assuming all three leads are matched, add 0.05 Ω (0.00 Ω~400.00 Ω),
0.2 Ω (400.0 Ω~4000.0 Ω) to the specifications. Temperature coefficient:
± (0.002 % of reading + 0.002 % of range) /°C ( <18 °C or >28 °C)
(% of reading +counts)
Millivolt measurement and source (Fluke 714B)
Resolution Range Accuracy
-10 mV to 75 mV 0.01 mV 0.015 % + 10 µA
Temperature coefficient: ± (0.002 % of reading + 0.002 % of range) /°C ( <18 °C or >28 °C)
(% of reading +counts)
Ohms source (Fluke 712B)
Ohms range Excitation
current from measurement device
1.0 Ω to 400.0 Ω
1.00 Ω to 400.00 Ω
400.0 Ω to 1500.0 Ω
1500.0 Ω to 4000.0 Ω
0.1 mA to 0.5 mA
0.5 mA to 3 mA
0.05 mA to 0.8 mA
0.05 mA to 0.4 mA
Resolution
0.00 Ω to 400.00 Ω 0.01 Ω
400.0 Ω to 4000.0 Ω 0.1 Ω
Temperature coefficient: ± (0.002 % of reading + 0.002 % of range) /°C ( <18 °C or >28 °C)
Supports pulsed transmitters and PLCs with pulse times as short as 5 ms
Accuracy (% of reading + counts)
0.015 % + 0.1 Ω
0.015 % + 0.05 Ω
0.015 % + 0.5 Ω
0.015 % + 0.5 Ω
2 Fluke Corporation Fluke 71XB Temperature Calibrators
RTD input and output (Fluke 712B)
Measure (°C) Source (°C)
RTD Type (α) Range (°C)
10Ω
Pt(385)
50Ω
Pt(385)
100 Ω
Pt(385)
200 Ω
Pt(385)
500 Ω
Pt(385)
1000 Ω Pt(385)
100 Ω
Pt(3916)
100 Ω
Pt(3926)
10 Ω
Cu(427)
120 Ω
Ni(672)
50 Ω
Cu(427)
100 Ω
Cu(427)
-200 to 100 °C 1.5 °C 3 °C 1 mA 1.5 °C 3 °C
100 to 800 °C 1.8 °C 3.6 °C 1 mA 1.8 °C 3.6 °C
-200 to 100 °C 0.4 °C 0.7 °C 1 mA 0.4 °C 0.7 °C
100 to 800 °C 0.5 °C 0.8 °C 1 mA 0.5 °C 0.8 °C
-200 to 100 °C 0.2 °C 0.4 °C
100 to 800 °C 0.015 %+0.18 °C 0.03 %+0.36 °C 0.015 %+0.18 °C 0.03 %+0.36 °C
-200 to 100 °C 0.2 °C 0.4 °C
100 to 630 °C 0.015 %+0.18 °C 0.03 %+0.36 °C 0.015 %+0.18 °C 0.03 %+0.36 °C
-200 to 100 °C 0.3 °C 0.6 °C
100 to 630 °C 0.015 %+0.28 °C 0.03 %+0.56 °C 0.015 %+0.28 °C 0.03 %+0.56 °C
-200 to 100 °C 0.2 °C 0.4 °C
100 to 630 °C 0.015 %+0.18 °C 0.03 %+0.36 °C 0.015 %+0.18 °C 0.03 %+0.36 °C
-200 to 100 °C 0.2 °C 0.4 °C
100 to 630 °C 0.015 %+0.18 °C 0.03 %+0.36 °C 0.015 %+0.18 °C 0.03 %+0.36 °C
-200 to 100 °C 0.2 °C 0.4 °C
100 to 630 °C 0.015 %+0.18 °C 0.03 %+0.36 °C 0.015 %+0.18 °C 0.03 %+0.36 °C
-100 to 260 °C 1.5 °C 3 °C 1 mA 1.5 °C 3 °C
-80 to 260 °C 0.15 °C 0.3 °C 1 mA 0.15 °C 0.3 °C
-180 to 200 °C 0.4 °C 0.7 °C 1 mA 0.4 °C 0.7 °C
-180 to 200 °C 0.2 °C 0.4 °C 1 mA 0.2 °C 0.4 °C
1-year 2-year
YSI400 15 to 50 °C 0.2 °C 0.4 °C 250 μA 0.2 °C 0.4 °C
1. Sensor inaccuracies not included.
2. Resolution: 0.1 °C.
3. Read accuracy is based on 4-wire input. For 3-wire RTD measurements, assuming all three RTD leads are matched, add 1.0 °C (Pt10 and Cu10), 0.6 °C (Pt50 and Cu50), 0.4 °C (Other RTD types) to the specifications.
4. Source Accuracy in source mode is based on 0.5 mA~3 mA (1.00 Ω~400.00 Ω), 0.05 mA~0.8 mA (400.0 Ω~1500.0 Ω), 0.05 mA~0.4 mA (1500.0 Ω ~4000.0 Ω), excitation current (0.25 mA for Pt1000 range).
5. Temperature Coefficient: ±0.05 °C /°C for measure, ± 0.05 °C /°C (<18 °C or >28 °C) for source.
6. Supports pulsed transmitters and PLCs with pulse times as short as 5 ms.
Source current
1 mA
500 μA
250 μA
250 μA
1 mA
1 mA
1-year 2-year
0.2 °C 0.4 °C
0.2 °C 0.4 °C
0.3 °C 0.6 °C
0.2 °C 0.4 °C
0.2 °C 0.4 °C
0.2 °C 0.4 °C
3 Fluke Corporation Fluke 71XB Temperature Calibrators
Thermocouple input and output (Fluke 714B)
TC Type Range (°C)
-250 to 200 °C 1.3 2.0 0.6 0.9
E
N
J
K
T
B
R
S
C
L
U
BP
XK
G
D
P
M
-200 to -100 °C 0.5 0.8 0.3 0.4
-100 to 600 °C 0.3 0.4 0.3 0.4 600 to 1000 °C 0.4 0.6 0.2 0.3
-200 to -100 °C 1.0 1.5 0.6 0.9
-100 to 900 °C 0.5 0.8 0.5 0.8 900 to 1300 °C 0.6 0.9 0.3 0.4
-210 to -100 °C 0.6 0.9 0.3 0.4
-100 to 800 °C 0.3 0.4 0.2 0.3 800 to 1200 °C 0.5 0.8 0.3 0.3
-200 to -100 °C 0.7 1.0 0.4 0.6
-100 to 400 °C 0.3 0.4 0.3 0.4 400 to 1200 °C 0.5 0.8 0.3 0.4
1200 to 1372 °C 0.7 1.0 0.3 0.4
-250 to -200 °C 1.7 2.5 0.9 1.4
-200 to 0 °C 0.6 0.9 0.4 0.6 0 to 400 °C 0.3 0.4 0.3 0.4
600 to 800 °C 1.3 2.0 1.0 1.5
800 to 1000 °C 1.0 1.5 0.8 1.2
1000 to 1820 °C 0.9 1.3 0.8 1.2
-20 to 0 °C 2.3 2.8 1.2 1.8 0 to 100 °C 1.5 2.2 1.1 1.7
100 to 1767 °C 1.0 1.5 0.9 1.4
-20 to 0 °C 2.3 2.8 1.2 1.8 0 to 200 °C 1.5 2.1 1.1 1.7
200 to 1400 °C 0.9 1.4 0.9 1.4
1400 to 1767 °C 1.1 1.7 1.0 1.5
0 to 800 °C 0.6 0.9 0.6 0.9
800 to 1200 °C 0.8 1.2 0.7 1.0 1200 to 1800 °C 1.1 1.6 0.9 1.4 1800 to 2316 °C 2.0 3.0 1.3 2.0
-200 to -100 °C 0.6 0.9 0.3 0.4
-100 to 800 °C 0.3 0.4 0.2 0.3 800 to 900 °C 0.5 0.8 0.2 0.3
-200 to 0 °C 0.6 0.9. 0.4 0.6 0 to 600 °C 0.3 0.4 0.3 0.4
0 to 1000 °C 1.0 1.5 0.4 0.6 1000 to 2000 °C 1.6 2.4 0.6 0.9 2000 to 2500 °C 2.0 3.0 0.8 1.2
-200 to 300 °C 0.2 0.3 0.2 0.5 300 to 800 °C 0.4 0.6 0.3 0.6 100 to 300 °C 1.6 2.4 1.2 1.8
300 to 1500 °C 1.0 1.5 1.0 1.5
1500 to 2320 °C 2.0 3.0 1.6 2.4
0 to 300 °C 1.6 2.4 1.2 1.8
300 to 1500 °C 1.0 1.5 1.0 1.5
1500 to 2315 °C 2.0 3.0 1.6 2.4
0 to 1000 °C 1.6 2.4 0.6 0.9
1000 to 1395 °C 2.0 3.0 0.8 1.2
-50 to 100 °C 1.0 1.5 0.4 0.6
100 to 1000 °C 1.6 2.4 0.6 0.9
1000 to 1410 °C 2.0 3.0 0.8 1.2
Measure (°C) Source (°C)
1 year 2 year 1 year 2 year
4 Fluke Corporation Fluke 71XB Temperature Calibrators
Thermocouple Probes
Bead H VAC Immersion Surface Air Piercing
General purpose
Industrial
surface
80PK-1 80PJ-1
Lowest temperature
Highest temperature
Probe material
Probe length 1 m lead wire
Cable length 1 m 1.3 m 1 m Connection Molded thermocouple plug SureGrip handle No No Yes No Yes Yes Yes Ye s
Key feature
Thermocouple types
Typical use General purpose HVAC Food service Industrial Residential Commercial
-40 °C -30 °C -40 °C 0 °C -40 °C
260 °C 105 °C 1090 °C 260 °C 816 °C 350 °C 816 °C 600 °C
Type K wire
with
teflon
insulation
Ideal for initial
troubleshoot-
ing. Can be
secured in
place with a
magnet.
K, J K K K, T K
80PK-11 80PK-22 80PK-3A 80PK-24
Velcro Inconel 600
48.26 cm
velcro cuff
Velcro probe allows hands free tempera-
ture measure-
ment.
21.27 cm 9.525 cm 21.59 cm 10.16 cm 21.57 cm 20.32 cm
For use in
liquids or in
gels.
Type K sensor
with teflon body
Exposed
junction for
direct contact
with flat or
slightly convex
surfaces.
Inconel
Perforated
baffle for air
and non-
caustic gas
measurements.
80PK-25 80PT-25
K Type:
-40 °C
T Type:
-196 °C
316 Stainless
Steel
Probe mate-
rial safe for
use in foods.
Sharp tip
pierces solid
surfaces.
80PK-26 80PK-27
-40 °C -127 °C
304 Stainless Steel
Use for general
purpose air
or surface
measurements.
Low conductivity
stainless steel
minimizes ther-
mal shunting.
Extra rugged.
5 Fluke Corporation Fluke 71XB Temperature Calibrators
Thermocouple kits and accessories
Thermocouple Plug Kits
80PK-8, 80PK-10, Pipe Clamp Temperature Probes
80CK-M and 80CJ-M Type K and J Male Mini-Connectors
80PJ-EXT, 80PK­EXT, 80PT-EXT Thermocouple Wire Extension Kits
5627A-6 -J, 5627A­9-J and 5627A-12-J Industrial RTD probes
700T C1
A kit of ten mini-plug connectors. One each of the following: Type J (black) Type K (yellow) Type T (blue) Type E (purple)
700TC2
A kit of seven mini-plug connectors. Type J (black), two Type K (yellow), two Type E (purple), one Type T (blue), one
Type R/S (green), one Type R/S (green) Type B or Cu (white) Type L (J-DIN) (blue) Type U (T-DIN) (brown) Type C (red) Type N (orange)
Type K thermocouples clamp securely onto pipes for fast temperature
and superheat measurements
Durable ribbon sensors
1 m lead
Measure from -29 °C to 149 °C
80PK-8 for 6.4 mm to 34.9 mm
80PK-10 for 32 mm to 64 mm
Isothermal screw terminal for K or J wire
Suitable for up to 20 gauge thermocouple wire
Color coded to industry standards (K-yellow, J-black)
Two per package
For extending and repairing type J, K or T thermocouple wires.
Kit includes
3 m of thermocouple wire and 1 pair of male/female mini-connectors
Maximum continuous exposure temperature: 260 °C
80PK-EXT is compatible with type K thermometers; 80PJ-EXT is
designed for type J thermometers, and 80PT-EXT is designed for type T thermometers
5627A-6-J, 5627A-9-J and 5627A-12-J Industrial RTD probes for Flu k e -712B
15.24 cm and 22.86 cm (6 and 9 inch) models measure to 300 °C,
30.48 cm model (12 inch) measures up to 420 °C
Accuracy to ± 0.025 °C.
NVLAP-accredited calibration included
Uses IEC standard PT-100-385 RTD curve
Each probe is individually calibrated and includes a NVLAP-
accredited report of calibration
Terminated with (4) Banana plugs for 4-wire temperature
measurement with the 712B
Add 2601 (22.86 cm, 9 in) or 2609 (63.5 cm , 25 in) protective case to
protect the probe
Ordering information
FLUKE-712B Temperature Calibrator FLUKE-714B Temperature Calibrator
Included equipment
Magnetic hanging tool, batteries, manual, traceable calibration certificate, and test leads
6 Fluke Corporation Fluke 71XB Temperature Calibrators
Fluke. The Most Trusted Tools in the World.
Fluke Corporation PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V. PO Box 1186, 5602 BD Eindhoven, The Netherlands
For more information call: In the U.S.A. (800) 443-5853 or Fax (425) 446-5116 In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222 In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116 Web access: http://www.fluke.com
©2014 Fluke Corporation. Specifications subject to change without notice. Printed in U.S.A. 1/2014 6001702B_EN
Modification of this document is not permitted without written permission from Fluke Corporation.
Loading...