Fluke 712 Instruction Manual

Page 1
712
Instruction Sheet
Introduction
Fluke Model 712 RTD Calibrator is a handheld tool for calibrating RTD (Resistance Temperature Detector) transmitters, including most pulsed transmitters. It simulates and measures seven different types of RTDs, in units of °C or °F. It also simulates and measures resistance in units of ohms. It does not source and measure simultaneously.
Your calibrator is supplied with a Flex-Stand holster, an installed 9 V alkaline battery, two sets of test leads, and this instruction sheet.
If the calibrator is damaged or something is missing, contact the place of purchase immediately. Contact your Fluke distributor for information about accessories. To order replacement parts or spares, see “Replacement Parts.”
The next table lists the RTD types supported by the calibrator, with their ranges, resolution, and the allowable excitation current from an RTD measurement device under test. All RTD types use ITS-90 curves. Full calibrator specifications are listed at the end of this instruction sheet.
®
PN 650280 August 1997, Rev.2, 5/04 1997-2004 Fluke Corporation. All rights reserved. Printed in U.S.A.
All product names are trademarks of their respective companies.
Page 2
Safety Information
W Warning
To avoid possible electric shock or personal injury:
Never apply more than 30 V between any two
terminals, or between any terminal and earth ground.
Make sure the battery door is closed and
latched before you operate the calibrator.
Remove test leads from the calibrator before
you open the battery door.
Do not operate the calibrator if it is damaged.
Do not operate the calibrator around explosive
gas, vapor, or dust.
When servicing the calibrator, use only specified replacement parts.
International Symbols
Symbol Meaning
J
I
Earth ground
Fuse
M
W
T
$
P
Battery
Refer to this instruction sheet for information about this feature.
Double insulated
Conforms to relevant Canadian Standards Association directives
Conforms to European Union directives
Page 3
Getting Acquainted with the Calibrator
Press the green O pushbutton to turn the calibrator on and off. Press the INPUT/OUTPUT pushbutton to select either INPUT (measuring) or OUTPUT (simulating).
1
2
3
4
8
7
6
5
Display Elements
Element Meaning
A INPUT Lit when measuring an RTD or resistance
B OUTPUT Lit when simulating an RTD or resistance
C
D °C, °F, When an RTD type is selected, one of these is lit
When simulating an RTD or ohms, the excitation
i H
current from the measuring device under test is too high. The calibrator output is unspecified.
to show the selected scale.
kg03f.eps
Page 4
Display Elements (cont.)
Element Meaning
E 2W, 3W, 4W
F RTD TYPES The RTD type annunciators (example, Ni 120)
G B Lit when the battery is low.
H Numerals Show the measured or simulated value in degrees
Pushbutton Function
R
C
o
2W 3W 4W
[ {
] }
When measuring an RTD, one of these is lit to indicate a two-wire, three-wire, or four-wire configuration. These annunciators are not used when simulating an RTD or resistance (output).
show the selected RTD type.
or ohms. When OL appears, the value is out of range.
Pushbutton Functions
Press to select a different RTD type. When you select the e type (ohms), displayed units are ohms, not degrees.
Press to toggle temperature scales between Celsius and Fahrenheit.
Press to select input (measure) or output (simulate) mode.
In simulate mode, press to step up or down 50° or 50 .
In measurement mode, press a two-wire, three-wire, or four-wire RTD input configuration. You will see the corresponding 2W, 3W, or 4W on the display.
Press to scroll up or down display. Hold down to scroll faster.
or { to select
[
Page 5
Simulating an RTD
To simulate an RTD, proceed as follows:
1. Press O to turn on the power.
2. If the calibrator is in input mode (INPUT on the display), press INPUT OUTPUT once. Make sure the display shows OUTPUT.
3. Press
R to select the desired RTD type.
4. Connect test leads to the terminals of the RTD measuring device as shown. Use only the two center outputs (labeled 2W 3W 4W OUTPUT).
INPUT
4W 3W
Sense connection (if needed)
30V
MAX
NC NC
30V
MAX
2W 3W 4W
OUTPUT
RTD
measurement
device input
30V
MAX
Sense connection (if needed)
kg01f.eps
Page 6
Measuring an RTD
To measure an RTD, proceed as follows:
1. Press O to turn on the power.
2. If the calibrator is in simulate mode (OUTPUT on the display), press INPUT OUTPUT once. Make sure the display shows INPUT.
3. Press
4. Press
R to select the desired RTD type. [ or { to select a two-wire, three-wire, or four-
wire RTD input configuration. Look for the 2W, 3W, or 4W annunciator on the display to verify that the configuration is set correctly.
5. Connect test leads to the RTD as shown below. Use two, three, or four inputs, depending on the setting of 2W, 3W, or 4W on the display.
INPUT
4W 3W
30V
MAX
NC NC
30V
MAX
2W 3W 4W
OUTPUT
30V
MAX
3W4W
RTD
kg02f.eps
Page 7
Maintenance
For maintenance procedures not described in this sheet, contact a Fluke Service Center.
In Case of Difficulty
Check the battery and test leads. Replace as necessary.
Review this sheet to make sure you are using the calibrator
correctly.
If the calibrator needs repair, contact a Fluke Service Center. If the calibrator is under warranty, see the warranty statement for terms. If the warranty has lapsed, the calibrator will be repaired and returned for a fixed fee. Contact a Fluke Service Center for information and price.
Cleaning
Periodically wipe the case with a damp cloth and detergent; do not use abrasives or solvents.
Calibration
Calibrate your calibrator once a year to ensure that it performs according to its specifications. A calibration manual is available (PN 686540). Call 1-800-526-4731 from the USA and Canada. In other countries, contact a Fluke Service Center.
Page 8
Replacing the Battery
When the B symbol appears on the display, replace the battery with a 9 V alkaline battery.
it07f.eps
Page 9
Testing and Replacing the Fuses
WWarning
To avoid personal injury or damage to the calibrator, use only a 0.125A 250V fast fuse, Littelfuse 2AG.
Fuse F1 protects the input circuit. Fuse F2 protects the output circuit. Test and replace the fuses using the following procedure:
1. Remove the test leads from the calibrator terminals and turn the calibrator off.
2. Remove the battery door.
3. Remove the three Phillips-head screws from the case bottom and turn the case over.
4. Gently lift the top cover from the end nearest the input/output terminals until it unsnaps from the bottom cover.
5. Gently remove the fuse from its mounting bracket.
6. Measure the resistance of the fuse. An open or high resistance suggests that the fuse is blown.
7. Replace the blown fuse with a 0.125 A 250 V fast fuse, Littelfuse 2AG.
8. Fit the top and bottom covers together, engaging the two snaps. Make sure that the keypad and the input/output terminal gasket are properly seated.
9. Reinstall the three screws.
10. Replace the battery door.
Page 10
F1
F2
Snaps
kg08f.eps
Page 11
Replacement Parts and Accessories
Replacement Parts
Item Description PN or
Model
no.
Qty.
BT1 9V battery, ANSI/NEDA 1604A or
IEC 6LR61
CG81Y Holster, Yellow CG81Y 1
W F1,F2 Fuse, 125 mA, 250V fast 686527 2
MP85 Case top 620192 1
MP86 Case bottom 620168 1
H2, 3, 4 Case screw 832246 3
MP89, 90 Non-skid foot 824466 2
MP8 O-ring for input/output receptacle 831933 1
MP92 Battery door 609930 1
H5, 6 Battery door fasteners 948609 2
S1 Keypad 687084 1
- 712 Instruction Sheet 650280 1
- Test lead, red 688051 2
- Test lead, black 688066 2
- 71X Series Calibration Manual 686540 Option
614487 1
Page 12
MP8
BT1
MP86
H2, 3, 4
MP85
S1
F2
F1
H5, 6
Instruction
Sheet
Test Lead Set
MP92
Holster
MP89, 90
kg10c.eps
Page 13
Specifications
Specifications are based on a one year calibration cycle and apply for ambient temperature from +18°C to +28°C unless stated otherwise.
Note
Specifications on this Instruction Sheet apply to the 712 RTD Calibrators with serial number 7676001 or above.
Ohms Specifications
Ohms Range
400.00
400.0 to
1500.0
1500.0 to
3200.0
Allowable Excitation is for Output mode only. It shows the allowable excitation current from an ohmmeter or RTD measurement device connected to the calibrator.
Excitation current from 712: 0.2 mA.
Maximum input voltage: 30 V
Input
Accuracy
4-Wire ±Ω
0.1
0.5 0.5 0.05 to 0.8
1 1 0.05 to 0.4
Output
Accuracy
±
0.15 0.1 to 0.5 0.00 to
0.1 0.5 to 3.0
Allowable
Excitation (mA)
Page 14
RTD Specifications
Note
Since ohms input and output units are available, you can use the calibrator for any unsupported RTD type by selecting the ohms range and making manual calculations or referring to tables.
Accuracy (°C)
0.2
0.8
0.3
0.4
0.2
0.2
Allowable Excitation
(mA)
0.1 to 3.0
0.05 to 0.8
0.05 to 0.4
Input
RTD Type Range °C (°F)
4-Wire 2-Wire
Output
&
3-Wire
Ni 120 -80.0 to 260.0
Pt 100 385 -200.0 to 800.0
Pt 200 385 -200.0 to 250.0
Pt 500 385 -200.0 to 500.0
Pt 1000 385 -200.0 to 100.0
Pt 100 392 (3926) Pt 100 JIS (3916)
Addresses pulsed transmitters and PLCs with pulses as short as 5 ms. Allowable Excitation is for Output mode only. It shows the allowable excitation
current from an ohmmeter or RTD measurement device connected to the calibrator.
Excitation current from 712: 0.2 mA. Maximum input voltage: 30 V
(-112.0 to 500.0)
(-328.0 to 1472.0)
(-328.0 to 482.0)
250.0 to 630.0
(482.0 to 1166.0)
(-328.0 to 932.0)
500.0 to 630.0
(932.0 to 1166.0)
(-328.0 to 212.0)
100.0 to 630.0
(212.0 to 1166.0)
-200.0 to 630.0
(-328.0 to 1166.0)
-200.0 to 630.0
(-328.0 to 1166.0)
0.2 0.3 0.2 0.1 to 3.0
0.33 0.5 0.33 0.1 to 3.0
0.2
0.8
0.3
0.4
0.2
0.2
0.3 0.5 0.3 0.1 to 3.0
0.3 0.5 0.3 0.1 to 3.0
0.3
1.6
0.6
0.9
0.4
0.5
Page 15
General Specifications
Resolution: RTD: 0.1°C, 0.1°F. Ohms: 0.1 Maximum voltage applied between any terminal and earth
ground or between any two terminals: 30 V Storage temperature: -20°C to 60°C Operating temperature: -10°C to 55°C Operating altitude: 3000 meters maximum Temperature coefficient: 0.005% of ohms range per °C for
temperature ranges -10°C to 18°C and 28°C to 55°C. Ohms ranges are 400 , 1.5 k, and 3.2 kΩ.
Relative humidity: 95% up to 30°C, 75% up to 40°C, 45% up to 50°C, and 35% up to 55°C
Vibration: Random 2 g, 5 Hz to 500 Hz Shock: 1 meter drop test Safety: Certified as compliant to CAN/CSA C22.2 No.
1010.1:1992. Complies with ANSI/ISA S82.01-1994.
Power requirements: Single 9 V battery (ANSI/NEDA 1604A or IEC 6LR61)
Size: 32 mm H x 87 mm W x 187 mm L (1.25 in H x 3.41 in W x
7.35 in L);
With holster and Flex-Stand: 52 mm H x 98 mm W x 201 mm L (2.06 in H x 3.86 in W x 7.93 in L)
Weight: 337 g (11.9 oz); With holster and Flex-Stand: 587 g (20.7 oz)
Page 16
How to Contact Fluke
To order accessories, receive operating assistance, or get the location of the nearest Fluke distributor or Service Center, call:
To contact Fluke, call one of the following telephone numbers:
USA: 1-888-99-FLUKE (1-888-993-5853) Canada: 1-800-36-FLUKE (1-800-363-5853) Europe: +31 402-675-200 Japan: +81-3-3434-0181 Singapore: +65-738-5655 Anywhere in the world: +1-425-446-5500
Or, visit Fluke's Web site at www.fluke.com. To register your product, visit register.fluke.com Address correspondence to:
Fluke Corporation Fluke Europe B.V. P.O. Box 9090, P.O. Box 1186, Everett, WA 98206-9090 5602 BD Eindhoven U.S.A. The Netherlands
LIMITED WARRANTY & LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and workmanship for three years from the date of purchase. This warranty does not cover fuses, disposable batteries or damage from accident, neglect, misuse or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Fluke’s behalf. To obtain service during the warranty period, send your defective calibrator to the nearest Fluke Authorized Service Center with a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you.
Loading...