Fluke 712 User Manual

712 RTD Calibrator
Compact, lightweight, and easy to carry
RTD CALIBRATOR
712
°C/°F
TYPE
2W 3W 4W
INPUT
OUTPUT
MAX
INPUT
30V
30V
30V
MAX
MAX
2W 3W 4W
OUTPUT
4W 3W
712 RTD Calibrator
Measure temperature from RTD
output Simulate RTD output
Rosemount pulsed RTD transmitter
compatible Operates with seven types
of RTDs Measure additional RTDs using Ohms
measurement function Simulate additional RTDs using Ohms
source function °F or °C selectable
Four shrouded banana jacks
General features
Compact and lightweight
Simple controls
Clear, 5-digit, liquid crystal display
Operating temperature:
-10°C to 55°C Safety: CSA C22.2 No. 1010.1:1992
Three-year warranty
Included:
– Fluke 712 RTD Calibrator – Protective Yellow Holster with Flex-Stand – Test leads and alligator clips – Single 9V alkaline battery – Instructions (14 languages)
General specifications
Maximum voltage: 30 volts Non-operating temperature:
-40°C to 60°C
Operating temperature: -10°C to 55°C Relative humidity: 95% (10°C to 30°C);
75% (30°C to 40°C); 45% (40°C to 50°C); 35% (50°C to 55°C)
Operating altitude: 3,000m max Shock: 1m drop test Vibration: random, 2g, 5-500 Hz Safety: CSA C22.2 No. 1010.1:1992 EMC: EN50082-1:1992 and
EN55022:1994 Class B Size/weight: 187 mm L x 87 mm W x 32 mm D (7.35 in L x 3.41 in W x
1.25 in D) 330g (12 oz)
Size/weight, (with holster and FlexStand™): 201mm L x 98 mm W x
52 mm D (7.93 in L x 3.86 in W x
2.06 in D) 600g (21 oz) Power: 9V battery ANSI/NEDA 1604A or IEC 6LR619V alkaline Battery Life: 4 to 20 hours, typical, depending on functions used Warranty: Three years
Ordering Information
Fluke-712 RTD Calibrator
Need technical assistance? Call:
1-800-44-FLUKE
Toll-free in the U.S. (905) 890-7600 in Canada
(425) 356-5500 other countries
For more information to be sent to you by fax, call 1-800-FLUKE-FAX in the U.S. and Canada. Or, visit our Website at http://www.fluke.com
Specifications ( 18° to 28° C, one year)
Function Range Resolution Accuracy Notes
Measure/simulate -200 to 800°C0.1°C, 0.1°F 0.33°C, 0.6°F Pt; 100 200 500 1000 RTD (Pt 100) (Pt 100) (385); Pt 100 (392);
Measure/simulate 15 to 3200 0.1 0.1 to 1 resistance
Pt 100 (392) JIS; Ni 120 (672)
Fluke. Keeping your world
up and running.
C0199UEN 8/98 Rev A