712 RTD Calibrator
Compact, lightweight, and easy to carry
712 RTD CALIBRATOR
RTD |
°C/°F |
TYPE |
2W 3W 4W
INPUT
OUTPUT
|
|
INPUT |
|
4W |
30V |
30V |
30V |
|
MAX |
MAX |
MAX |
2W 3W 4W OUTPUT
712 RTD Calibrator
• |
Measure temperature from RTD |
• |
output |
Simulate RTD output |
|
• |
Rosemount pulsed RTD transmitter |
• |
compatible |
Operates with seven types |
|
• |
of RTDs |
Measure additional RTDs using Ohms |
|
• |
measurement function |
Simulate additional RTDs using Ohms |
|
• |
source function |
°F or °C selectable |
|
• |
Four shrouded banana jacks |
General features |
|
• |
Compact and lightweight |
• |
Simple controls |
• |
Clear, 5-digit, liquid crystal display |
• |
Operating temperature: |
3W |
-10°C to 55°C |
• |
Safety: CSA C22.2 No. 1010.1:1992 |
• |
Three-year warranty |
• |
Included: |
– Fluke 712 RTD Calibrator
– Protective Yellow Holster with Flex-Stand™
– Test leads and alligator clips
– Single 9V alkaline battery
– Instructions (14 languages)
General specifications
Maximum voltage: 30 volts Non-operating temperature:
-40°C to 60°C
Operating temperature: -10°C to 55°C Relative humidity: 95% (10°C to 30°C); 75% (30°C to 40°C); 45% (40°C to 50°C); 35% (50°C to 55°C)
Operating altitude: 3,000m max
Shock: 1m drop test
Vibration: random, 2g, 5-500 Hz Safety: CSA C22.2 No. 1010.1:1992 EMC: EN50082-1:1992 and EN55022:1994 Class B
Size/weight: 187 mm L x 87 mm W x 32 mm D (7.35 in L x 3.41 in W x 1.25 in D) 330g (12 oz)
Size/weight, (with holster and FlexStand™): 201mm L x 98 mm W x 52 mm D (7.93 in L x 3.86 in W x 2.06 in D) 600g (21 oz)
Power: 9V battery ANSI/NEDA 1604A or IEC 6LR619V alkaline
Battery Life: 4 to 20 hours, typical, depending on functions used Warranty: Three years
Specifications (18° to 28°C, one year)
Function |
Range |
Resolution |
Accuracy |
Notes |
|
||||
|
|
|
|
|
Measure/simulate |
-200 to 800°C |
0.1°C, 0.1°F |
0.33°C, 0.6°F |
Pt; 100 200 500 1000 |
RTD |
(Pt 100) |
|
(Pt 100) |
(385); Pt 100 (392); |
|
|
|
|
Pt 100 (392) JIS; |
|
|
|
|
Ni 120 (672) |
Measure/simulate |
15 Ω to 3200 Ω |
0.1 Ω |
0.1Ω to 1 Ω |
|
resistance |
|
|
|
|
|
|
|
|
|
Ordering
Information
Fluke-712 RTD Calibrator
Need technical assistance? Call:
1-800-44-FLUKE
Toll-free in the U.S.
(905) 890-7600 in Canada
(425) 356-5500 other countries
For more information to be sent to you by fax, call 1-800-FLUKE-FAX
in the U.S. and Canada. Or, visit our Website at http://www.fluke.com
Fluke. Keeping your world up and running.
C0199UEN 8/98 Rev A