Fluke 707 User Manual

Introduction
Calibration Information
XW Warning
Read Safety Information before using the mA Calibrator.
Contacting Fluke
mA Calibrator
®
The Calibration Information for the 707Ex mA Calibrator (hereafter referred to as the mA Calibrator) provides the information necessary to verify the performance and adjust the calibration of the mA Calibrator.
This document includes the following:
Contacting Fluke
Specifications
Safety Information
Maintenance
User-Replaceable Parts
Recommended Test Equipment
Performance Verification
Calibration Adjustment
For complete operating instructions, refer to the 707Ex Users Manual located on the CD that came with the mA Calibrator.
Warranty
This Fluke product comes with a 3-year warranty for material and workmanship. For warranty information, refer to the 707Ex Users Manual that came with the mA Calibrator.
To contact Fluke, call one of the following telephone numbers:
Technical Support USA: 1-800-44-FLUKE (1-
800-443-5853)
Calibration/Repair USA: 1-888-99-FLUKE (1-
888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
Or, visit Fluke's website at www.fluke.com
To register your product, visit
http://register.fluke.com.
To view, print, or download the latest manual supplement, visit
http://us.fluke.com/usen/support/manuals
Address correspondence to:
Fluke Corporation P.O. Box 9090 Everett, WA 98206-9090 U.S.A.
Fluke Europe B.V. P.O. Box 1186 5602 BD Eindhoven The Netherlands
.
.
PN 2086601 May 2003 Rev. 2, 5/09
© 2003, 2009 Fluke Corporation, All rights reserved. Printed in U.S.A. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
1
Fluke 707Ex Calibration Information
Accuracy Specifications
Accuracy is specified for 1-year after calibration at operating temperatures of 18 °C to 28 °C and is given as:
± ( [ % of reading ] + [ counts ] )
MEASURE V dc:
Range: + 28 V max Resolution: 1 mV Input Impedance: 1 MΩ Accuracy: ± (0.015 % of reading + 2 counts)
MEASURE mA dc:
Range: 20 mA (24 mA max) Resolution: 1 μA Accuracy: ± (0.015 % of reading + 2 counts)
SOURCE / SIMULATE mA dc:
Range: 0 mA to 20 mA (24 mA max) Resolution: 1 μA Accuracy: ± (0.015 % of reading + 2 counts)
Source Mode:
Compliance: To 700 Ω at 20 mA
Simulate Mode:
External loop voltage requirement: 24 V nominal, 28 V maximum, 12 V minimum
Loop Power:
Maximum Load: 700 Ω
Percent Display:
25 % to 125 % (4 to 20 mA mode)
Input/Output Protection:
Fused; not replaceable
General Specifications
Maximum voltage between any terminal and earth ground or between any two terminals:
Temperature
Coefficient:
± 0.005 % of range per °C for temperatures of
10 °C to 18 °C and 28 °C to 50 °C
Relative Humidity:
95 % up to 30 °C 75 % up to 40 °C 45 % up to 50 °C
Vibration:
Random 2 g, 5 to 500 Hz
Shock:
1 meter drop test
Safety Compliance:
Complies with IEC 61010-1-95 CAT I, 28 V; CSA C22.2 No. 1010-92 NRTL; ANSI/ISA S82.02.01­1994;
( Directive 94/9/EG and p
NEC 500: Uo = 27.6 V, Io = 96.13 mA, Co = 76 nF, Lo = 2.5 mH, Ui = 30 V, Ii = 24 mA, Ci = 10 nF, Li = 0 mH
CE:
Complies with EN61010-1 and EN61326
Power Requirements:
Single 9 V battery (See Table 2 in the User-Replaceable Parts section)
Battery Life (typical):
SOURCE mode: 18 hours; 12 mA into 500 Ω; MEASURE / SIMULATE mode: 50 hours
Size:
69.85 mm (W) x 142.87 mm (L) x 50.80 mm (H) [ 2.75 in (W) x 5.625 in (L) x 2.00 in (H) ]
With red holster and Flex-Stand:
76.20 mm (W) x 158.75 mm (L) x 54.61mm (H) [ 3.00 in (W) x 6.25 in (L) x 2.15 in (H) ]
Weight:
0.28 kg (0.62 lb)
With red holster and Stand:
28 V
Storage Temperature:
30 °C to 60 °C
Operating Temperature:
10 °C to 50 °C
Operating Altitude:
3000 meters maximum
Pollution Degree 2
2
0.42 kg (0.93 lb)
Safety Information
In this manual a Warning identifies conditions and actions that pose hazards to the user. A Caution identifies conditions and actions that may damage the mA Calibrator or equipment under test. International symbols used in this manual are identified in the Symbols section.
Read the entire 707Ex Users Manual and the 707Ex mA Calibrator CCD (Concept Control Drawing) before using the mA Calibrator.
mA Calibrator
Safety Information
XW Warnings and Cautions
To avoid electric shock, injury, or damage to the mA Calibrator:
Use the mA Calibrator only as described in this
Calibration Information, the Users Manual, and the Fluke 707Ex mA Calibrator CCD (Concept Control Drawing) or the protection provided by the mA Calibrator may be impaired.
Inspect the mA Calibrator before use. Do not
use it if it appears damaged.
Check the test leads for continuity, damaged
insulation, or exposed metal. Replace damaged test leads.
Never apply more than 28 V between the input
terminals, or between any terminal and earth ground.
Applying more than 28 V to the input terminals
invalidates the mA Calibrator’s Ex Approval and may result in permanent damage to the unit so it can no longer be used.
Use the proper terminals, mode, and range for
your measuring or sourcing application.
To prevent damage to the unit under test, put
the mA Calibrator in the correct mode before connecting the test leads.
When making connections, connect the COM
test probe before the live test probe. When disconnecting, disconnect the live probe before the COM probe.
Never use the mA Calibrator with the red
holster removed.
Never use the mA Calibrator with the case
open. Opening the case violates Ex Approval.
Make sure the battery door is closed before
using the mA Calibrator.
Replace the battery as soon as the M (low
battery) symbol appears to avoid false readings that can lead to electric shock.
Remove test leads from the mA Calibrator
before opening the battery door.
This equipment is specified for use in
measurement category I (CAT I) pollution degree 2 environments and should not be used in CAT II, CAT III, or CAT IV environments. Voltage transients should not exceed 300 V for the CAT I applications where this product is used. Measurement transients are defined in IEC1010-1 as 2 µs rise time with a 50 µs duration at 50 % of the maximum amplitude height.
Measurement Category I (CAT I) is defined for
measurements performed on circuits not directly connected to the mains.
Do not use in a damp or wet environment.
Faults and Damage
XW Warning
Applying a voltage greater than 28 V to the input of the mA Calibrator invalidates its Ex Approval and may impair its safe operation in an Ex-hazardous area.
If there is any reason to suspect that the safe operation of the mA Calibrator has been affected, it must be immediately withdrawn from use, and precautionary measures must be taken to prevent any further use of the mA Calibrator in an Ex-hazardous area.
The safety features and integrity of the unit may be compromised by any of the following:
External damage to the housing
Internal damage to the mA Calibrator
Exposure to excessive loads
Incorrect storage of the unit
Damage sustained in transit
Correct certification is illegible
Using the product with the red holster removed
Functioning errors occur
Permitted limitations are exceeded
Functioning errors or obvious measurement
inaccuracies occur which prevent further measurement by the mA Calibrator
Safety Regulations
The use of this 707Ex mA Calibrator meets the requirements of the regulations providing that the user observes and applies the requirements as laid down in the regulations and that improper and incorrect use of the unit is avoided.
Use must be restricted to the specified application
parameters.
Do not open the mA Calibrator.
Do not remove or install the battery within the
Ex-hazardous area.
Do not carry additional batteries within the
Ex-hazardous area.
Use only type-tested batteries. The use of any
other battery will invalidate the Ex-certification and present a safety risk. Refer to Table 2.
Do not use the mA Calibrator in an Ex-hazardous
area unless it is completely and securely fitted in its accompanying red holster.
3
Loading...
+ 7 hidden pages