Read Safety Information before using the mA Calibrator.
Contacting Fluke
707Ex
mA Calibrator
®
The Calibration Information for the 707Ex mA
Calibrator (hereafter referred to as the mA Calibrator)
provides the information necessary to verify the
performance and adjust the calibration of the mA
Calibrator.
This document includes the following:
• Contacting Fluke
• Specifications
• Safety Information
• Maintenance
• User-Replaceable Parts
• Recommended Test Equipment
• Performance Verification
• Calibration Adjustment
For complete operating instructions, refer to the
707ExUsers Manual located on the CD that came
with the mA Calibrator.
Warranty
This Fluke product comes with a 3-year warranty for
material and workmanship. For warranty information,
refer to the 707Ex Users Manual that came with the
mA Calibrator.
To contact Fluke, call one of the following telephone
numbers:
• Technical Support USA: 1-800-44-FLUKE (1-
800-443-5853)
• Calibration/Repair USA: 1-888-99-FLUKE (1-
888-993-5853)
• Canada: 1-800-36-FLUKE (1-800-363-5853)
• Europe: +31 402-675-200
• Japan: +81-3-3434-0181
• Singapore: +65-738-5655
• Anywhere in the world: +1-425-446-5500
Or, visit Fluke's website at www.fluke.com
To register your product, visit
http://register.fluke.com.
To view, print, or download the latest manual
supplement, visit
http://us.fluke.com/usen/support/manuals
Address correspondence to:
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
Accuracy is specified for 1-year after calibration at
operating temperatures of 18 °C to 28 °C and is given
as:
± ( [ % of reading ] + [ counts ] )
MEASURE V dc:
Range: + 28 V max
Resolution: 1 mV
Input Impedance: 1 MΩ
Accuracy: ± (0.015 % of reading + 2 counts)
MEASURE mA dc:
Range: 20 mA (24 mA max)
Resolution: 1 μA
Accuracy: ± (0.015 % of reading + 2 counts)
SOURCE / SIMULATE mA dc:
Range: 0 mA to 20 mA (24 mA max)
Resolution: 1 μA
Accuracy: ± (0.015 % of reading + 2 counts)
Source Mode:
Compliance: To 700 Ω at 20 mA
Simulate Mode:
External loop voltage requirement: 24 V nominal,
28 V maximum, 12 V minimum
Loop Power:
Maximum Load: 700 Ω
Percent Display:
− 25 % to 125 % (4 to 20 mA mode)
Input/Output Protection:
Fused; not replaceable
General Specifications
Maximum voltage between any terminal and earth
ground or between any two terminals:
Temperature
Coefficient:
± 0.005 % of range per °C for temperatures of
−10 °C to 18 °C and 28 °C to 50 °C
Relative Humidity:
95 % up to 30 °C
75 % up to 40 °C
45 % up to 50 °C
Vibration:
Random 2 g, 5 to 500 Hz
Shock:
1 meter drop test
Safety Compliance:
Complies with IEC 61010-1-95 CAT I, 28 V; CSA
C22.2 No. 1010-92 NRTL; ANSI/ISA S82.02.011994;
( Directive 94/9/EG and p
NEC 500: Uo = 27.6 V, Io = 96.13 mA, Co = 76 nF,
Lo = 2.5 mH, Ui = 30 V, Ii = 24 mA, Ci = 10 nF, Li =
0 mH
CE:
Complies with EN61010-1 and EN61326
Power Requirements:
Single 9 V battery
(See Table 2 in the User-Replaceable Parts section)
Battery Life (typical):
SOURCE mode: 18 hours; 12 mA into 500 Ω;
MEASURE / SIMULATE mode: 50 hours
Size:
69.85 mm (W) x 142.87 mm (L) x 50.80 mm (H)
[ 2.75 in (W) x 5.625 in (L) x 2.00 in (H) ]
With red holster and Flex-Stand:
76.20 mm (W) x 158.75 mm (L) x 54.61mm (H)
[ 3.00 in (W) x 6.25 in (L) x 2.15 in (H) ]
Weight:
0.28 kg (0.62 lb)
With red holster and Stand:
28 V
Storage Temperature:
− 30 °C to 60 °C
Operating Temperature:
− 10 °C to 50 °C
Operating Altitude:
3000 meters maximum
Pollution Degree 2
2
0.42 kg (0.93 lb)
Safety Information
In this manual a Warning identifies conditions and
actions that pose hazards to the user. A Caution
identifies conditions and actions that may damage the
mA Calibrator or equipment under test. International
symbols used in this manual are identified in the
Symbols section.
Read the entire 707Ex Users Manual and the 707Ex mA Calibrator CCD (Concept Control Drawing)
before using the mA Calibrator.
Page 3
mA Calibrator
Safety Information
XW Warnings and Cautions
To avoid electric shock, injury, or damage to the
mA Calibrator:
• Use the mA Calibrator only as described in this
Calibration Information, the Users Manual, and
the Fluke 707Ex mA Calibrator CCD (Concept
Control Drawing) or the protection provided by
the mA Calibrator may be impaired.
• Inspect the mA Calibrator before use. Do not
use it if it appears damaged.
• Check the test leads for continuity, damaged
insulation, or exposed metal. Replace damaged
test leads.
• Never apply more than 28 V between the input
terminals, or between any terminal and earth
ground.
• Applying more than 28 V to the input terminals
invalidates the mA Calibrator’s Ex Approval
and may result in permanent damage to the unit
so it can no longer be used.
• Use the proper terminals, mode, and range for
your measuring or sourcing application.
• To prevent damage to the unit under test, put
the mA Calibrator in the correct mode before
connecting the test leads.
• When making connections, connect the COM
test probe before the live test probe. When
disconnecting, disconnect the live probe before
the COM probe.
• Never use the mA Calibrator with the red
holster removed.
• Never use the mA Calibrator with the case
open. Opening the case violates Ex Approval.
• Make sure the battery door is closed before
using the mA Calibrator.
• Replace the battery as soon as the M (low
battery) symbol appears to avoid false readings
that can lead to electric shock.
• Remove test leads from the mA Calibrator
before opening the battery door.
•This equipment is specified for use in
measurement category I (CAT I) pollution
degree 2 environments and should not be used
in CAT II, CAT III, or CAT IV environments.
Voltage transients should not exceed 300 V for
the CAT I applications where this product is
used. Measurement transients are defined in
IEC1010-1 as 2 µs rise time with a 50 µs
duration at 50 % of the maximum amplitude
height.
•Measurement Category I (CAT I) is defined for
measurements performed on circuits not
directly connected to the mains.
•Do not use in a damp or wet environment.
Faults and Damage
XW Warning
Applying a voltage greater than 28 V to the
input of the mA Calibrator invalidates its
Ex Approval and may impair its safe
operation in an Ex-hazardous area.
If there is any reason to suspect that the safe operation
of the mA Calibrator has been affected, it must be
immediately withdrawn from use, and precautionary
measures must be taken to prevent any further use of
the mA Calibrator in an Ex-hazardous area.
The safety features and integrity of the unit may be
compromised by any of the following:
• External damage to the housing
• Internal damage to the mA Calibrator
• Exposure to excessive loads
• Incorrect storage of the unit
• Damage sustained in transit
• Correct certification is illegible
• Using the product with the red holster removed
• Functioning errors occur
• Permitted limitations are exceeded
• Functioning errors or obvious measurement
inaccuracies occur which prevent further
measurement by the mA Calibrator
Safety Regulations
The use of this 707Ex mA Calibrator meets the
requirements of the regulations providing that the user
observes and applies the requirements as laid down in
the regulations and that improper and incorrect use of
the unit is avoided.
• Use must be restricted to the specified application
parameters.
• Do not open the mA Calibrator.
• Do not remove or install the battery within the
Ex-hazardous area.
• Do not carry additional batteries within the
Ex-hazardous area.
• Use only type-tested batteries. The use of any
other battery will invalidate the Ex-certification
and present a safety risk. Refer to Table 2.
• Do not use the mA Calibrator in an Ex-hazardous
area unless it is completely and securely fitted in
its accompanying red holster.
3
Page 4
Fluke 707Ex
Calibration Information
• After using the mA Calibrator in a non-
intrinsically safe protected circuit, a rest time of 3
minutes minimum duration must occur before the
mA Calibrator is taken into and/or used in an
Ex-hazardous area.
Ex-Certification Data
• ATEX Certificate of Conformity: ZELM 02
ATEX 0120 X
• Certification: ( II 2 G EEx ia IIC T4
• Permitted for: Zone1, Equipment Group II, gas
group C hazardous gases, vapor or mist,
Temperature class T4.
International symbols used on the mA Calibrator and
in this manual are explained below.
Symbol Meaning
+
J
W
(
ON / OFF button.
Earth ground
Caution: Important information. Refer to
instruction sheet
Conforms to ATEX requirements
Maintenance
XW Warning
To avoid electric shock, personal injury, or
damage to the mA Calibrator:
•Do not service this product. To
maintain the integrity of the mA
Calibrator in explosive atmospheres,
return the mA Calibrator to Fluke for
all repairs.
•Remove any input signals from the
test probes prior to removing test
leads from the mA Calibrator.
•When servicing the mA Calibrator, use
only specified replacement parts
described in Table 1.
•Use only the battery specified in Table
2.
•Do not allow water inside the case.
Contact a Fluke Service Center before performing any
maintenance procedure not described in the 707Ex Users Manual. Refer to the Contacting Fluke section.
Cleaning
XW Warning
To avoid electrical shock, remove test
leads and input signals before cleaning.
To clean the case, wipe it with a cloth lightly
dampened with water and mild detergent.
4
p
T
M
)
P
F
~
;
Conforms to Factory Mutual requirements
Double insulated
Battery
Conforms to relevant Canadian
Standards Association directives.
Certification # LR110460-2.
Conforms to European Union
requirements
Direct current
Do not dispose of this product as
unsorted municipal waste. Go to Fluke’s
website for recycling information.
Conforms to relevant Australian
standards
WCaution
Do not use abrasives, solvents, or alcohol.
These materials will damage the case.
Page 5
static awareness
A Message From
Fluke Corporation
Some semiconductors and custom IC's can be
damaged by electrostatic discharge during
handling. This notice explains how you can
minimize the chances of destroying such devices
by:
1. Knowing that there is a problem.
2. Learning the guidelines for handling them.
3. Using the procedures, packaging, and
bench techniques that are recommended.
The following practices should be followed to minimize damage to S.S. (static sensitive) devices.
1. MINIMIZE HANDLING
2. KEEP PARTS IN ORIGINAL CONTAINERS
UNTIL READY FOR USE.
3. DISCHARGE PERSONAL STATIC BEFORE
HANDLING DEVICES. USE A HIGH RESIS TANCE GROUNDING WRIST STRAP.
4. HANDLE S.S. DEVICES BY THE BODY.
Page 6
5. USE STATIC SHIELDING CONTAINERS FOR
HANDLING AND TRANSPORT.
6. DO NOT SLIDE S.S. DEVICES OVER
ANY SURFACE.
8. WHEN REMOVING PLUG-IN ASSEMBLIES
HANDLE ONLY BY NON-CONDUCTIVE
EDGES AND NEVER TOUCH OPEN EDGE
CONNECTOR EXCEPT AT STATIC-FREE
WORK STATION. PLACING SHORTING
STRIPS ON EDGE CONNECTOR HELPS
PROTECT INSTALLED S.S. DEVICES.
9. HANDLE S.S. DEVICES ONLY AT A
STATIC-FREE WORK STATION.
7. AVOID PLASTIC,VINYL AND STYROFOAM
IN WORK AREA.
PORTIONS REPRINTED
WITH PERMISSION FROM TEKTRONIX INC.
AND GERNER DYNAMICS, POMONA DIV.
Dow Chemical
10. ONLY ANTI-STATIC TYPE SOLDER SUCKERS SHOULD BE USED.
11. ONLY GROUNDED-TIP SOLDERING
IRONS SHOULD BE USED.
Page 7
mA Calibrator
User-Replaceable Parts
Replacing the Battery
XW Warning
To avoid false readings, which could lead to
electric shock or injury, replace the battery
as soon as M (low battery indicator)
appears on the display.
Do not remove or install the battery within
the Ex-hazardous area.
Use only a single 9 V alkaline battery,
properly installed, to power the
mA Calibrator. See Table 2.
To replace the battery (refer to Figure 1 and Table 2):
1. Remove the test probes from the input signal.
2. Press D to turn the mA Calibrator OFF.
3. Remove the test leads from the input terminals.
4. Remove the red holster.
5. Lift off the battery door on the back of the mA
Calibrator as shown.
6. Remove the battery.
7. Insert the replacement battery and re-install the
battery door. Make sure it is securely in place.
User-Replaceable Parts
User-replaceable parts and accessories are shown in
Figure 2 and Table 1.
1
2
7
6
4
5
Figure 2. User-Replaceable Parts
Table 1. User-Replaceable Parts
ItemPart Description Fluke Part No.Qty.
3
aqv10f.eps
8. Return the mA Calibrator to its red holster.
Figure 1. Replacing the Battery
aqv07f.eps
A Non-skid foot 885884 1
Battery, 9 volt,
B
Alkaline
C Battery door 665106 1
D Red holster, with bail 2040228 1
E Alligator clips AC72 1
F Test lead set TL75 1
G Users Manual on CD 2053979 1
Table 2. Approved Batteries
Battery Description Manufacturer Type
Alkaline, 9 volt Duracell 6LR61
Alkaline Ultra, 9 volt Duracell 6LR61
Professional Alkaline
Battery Procell, 9 volt
Alkaline Energizer, 9 volt Eveready 6LR61
822270 or
see Table 2
Duracell 6LR61
1
Alkaline Power Line
Industrial Battery, 9 volt
Alkaline, 9 volt Daimon 6LR61
Panasonic 6LR61
5
Page 8
Fluke 707Ex
Calibration Information
Recommended Test Equipment
A list of recommended equipment for the
performance verification tests and calibration
adjustment procedure is shown in Table 3.
Performance Verification
XW Warning
To avoid electric shock, do not perform
the verification tests unless the mA
Calibrator is fully assembled.
Perform the following tests to check the accuracy of
each of the mA Calibrator’s functions against its
specifications.
If the mA Calibrator fails any of these tests,
calibration adjustment or repair is required.
Performance Verification Procedures
DC Current Source Mode
Test 1:
1. Turn the mA Calibrator on.
2. After the P.S. power timeout displays, the mA
Calibrator should display 4.000 mA. If it is
displaying 0.000 mA, turn the mA Calibrator off.
Hold down A while turning the mA Calibrator
on. Continue to hold A for 2 seconds. When the
timeout display ends, the mA Calibrator should
display 4.000 mA.
3. Connect the mA Calibrator’s
3458A I input, and the COM terminal to the HP
3458A LO input.
4. Verify that the mA Calibrator's display shows
SOURCE in the upper-left corner.
5. Set the HP 3458A to measure DC Amps (DCI).
+ terminal to the HP
6. Refer to Table 4 to verify the readings on the HP
3458A for the following tests. (No adjustment is
necessary for test 1.)
Test 2:
1. Press B twice to select 12.000 mA.
2. Verify the readings.
Test 3:
1. Press B twice to select 20.000 mA.
2. Press and rotate the knob until 24.000 mA is
displayed.
3. Verify the readings.
DC Current Measurement Mode
Tests 4-6:
1. Press the A key twice to put the mA Calibrator
into the mA measurement mode.
2. The mA Calibrator’s display should show
MEASURE in the top right corner.
3. Disconnect the HP 3458A from the mA
Calibrator.
4. Connect the test leads from the
red AUX terminal
of the Fluke 5520A to the HP 3458A I input.
5. Connect a test lead from the
black AUX terminal
of the Fluke 5520A to the COM terminal on the
mA Calibrator.
6. Connect a test lead from the HP 3458A
LO input
to the mA Calibrator's + terminal.
7. Adjust the Fluke 5520A to output the value of test
4 in Table 4 as measured by the HP 3458A.
8. Verify the display readings on the mA Calibrator.
9. Repeat adjustment and reading for steps 5 and 6.
Table 3. Recommended Equipment
Equipment Recommended Model or Equivalent
DC Calibrator DC Voltage: 0 to 30 V
Accuracy: ± 0.004 % +0.5 mV
DC Current: 0 to 24 mA
Accuracy: ± 0.004% +0.5 μA
Digital Multimeter DC Current: 0 to 26 mA
Accuracy: ± 0.004% +0.5 μA
Flexible Test Leads Fluke TL24
6
Fluke 5520A Multi-Product Calibrator
HP 3458A
Page 9
mA Calibrator
Calibration Adjustment
Table 4. DC Current Source / Current Measure / Voltage Measure Mode Tests
Test No. 707Ex mA Calibrator Output Mode HP 3458A Minimum HP 3458A Maximum
1
2
3
Test No. 5520A Calibrator Output Mode Fluke 707Ex Minimum Fluke 707Ex Maximum
4
5
6
Test No. 5520A Calibrator Output Mode Fluke 707Ex Minimum Fluke 707Ex Maximum
7
8
9
DC Voltage Measurement Mode
Tests 7-9:
1. Press A twice to get the mA Calibrator to
V MEASURE. The mA Calibrator’s display
should read "V".
2. Connect the test leads from the
terminals of the Fluke 5520A to the input
terminals on the mA Calibrator (black to COM
and red to +).
3. Set the Fluke 5520A to test 7 in Table 4 and
verify the display reading on the mA Calibrator.
4. Repeat for tests 8 and 9.
5. The reading on the display should be within the
minimum and maximum values shown in
Table 4.
4.000 mA
12.000 mA
24.000 mA
4.000 mA
12.000 mA
24.000 mA
0.000 V
14.000 V
28.000 V
mA Source
mA Source
mA Source
mA Measure
mA Measure
mA Measure
Volt Measure
Volt Measure
Volt Measure
NORMAL output
3.9974 mA
11.9962 mA
23.9944 mA
3.997 mA
11.996 mA
23.994 mA
-0.002 V
13.995 V
27.993 V
4.0026 mA
12.0038 mA
24.0056 mA
4.003 mA
12.004 mA
24.006 mA
0.002 V
14.005 V
28.007 V
Calibration Adjustment
Perform the following calibration adjustment
procedure if the mA Calibrator fails the performance
verification test.
Calibration Adjustment Counter
The mA Calibrator contains a calibration adjustment
counter. The value in the calibration adjustment
counter can be recorded and used to show that no
adjustments have been made during a calibration
cycle. When Fluke performs the calibration, the value
of the calibration adjustment counter will be recorded
on the calibration label.
To find the value in the calibration adjustment
counter, hold down B
The software revision will display for 2 seconds
followed by the calibration counter for another 2
seconds.
while turning the unit on.
6. Set the 5520A output to 0 V (zero), and set output
to
STANDBY.
7. Turn the mA Calibrator
OFF and disconnect the
mA Calibrator from the 5520A.
8. The performance verification tests are now
complete.
9. If the mA Calibrator failed any of these tests,
calibration adjustment or repair is required.
Make sure that the mA Calibrator has a new
battery before starting the calibration
procedure. Calibration will not function
properly if the low battery indicator (M) is
on.
Because the mA Calibrator incorporates several key
hold start up features, entering the Adjust mode
through the keypad requires an exact key hold
Note
sequence:
1. With the mA Calibrator off, hold down
B and
C and turn the mA Calibrator on.
"
CAL" appears on the main display and the
calibration adjustment counter is displayed on the
7
Page 10
Fluke 707Ex
Calibration Information
secondary display. If the knob is not turned
within 2 seconds, the unit returns to the normal
operation mode.
2. The mA Calibrator adjustment procedure is
password protected. At this point, a password is
needed to go forward. "
main display and "000" on the secondary display.
Enter the password "
1. Rotate the knob to add a 2 in the LCD of
the secondary display push and release
the knob.
2. Rotate the knob to get 9 into the LCD,
push and release.
3. Rotate the knob to get a 7, push and
release.
If the password is correct, you'll be led through all
the adjust modes (source, mA read, V read),
otherwise the unit returns to normal operation. In
the adjust mode, the main display will show
"
1.000 mA".
3. Connect the mA Calibrator
3458A I input, and the COM terminal to the HP
3458A LO input. The mA Calibrator will be in
mA source cal and "1.000 mA" is displayed. On
the secondary LCD, the last 3 digits from the
knob value are shown. Adjust the mA output by
tuning the knob until "1.000 mA" shows on the
HP 3458A.
4. Press
5. Press
C, "22.000 mA" is displayed. Adjust the
mA output by turning the knob until "22.000 mA"
is shown on the HP 3458A.
A to go to mA measure adjust.
PAS" is displayed on the
297" as follows:
+ terminal to the HP
7. Connect the test leads from the
the Fluke 5520A to the terminals on the mA
Calibrator (black to COM and red to +).
8. The mA Calibrator will display "
0.000 mA is displayed. On the secondary display,
the last 3 digits of the A/D reading are shown. Set
the 5520A to output 0.000 mA into the mA
Calibrator. Press C.
9. After a moment, the mA Calibrator will display
24.000 mA. Set the 5520A to output 24.000 mA
into the mA Calibrator.
10. Press
11. Connect the test leads from the output
12. The mA Calibrator will display "
13. Press
14. Press
The calibration adjustment procedure is complete.
To see the value of the calibration counter without
calibrating the unit, hold down B
software revision number is displayed for 2 seconds
followed by the calibration counter for 2 seconds.
A to go to V measure adjust.
terminals of the Fluke 5520A to the input
terminals on the mA Calibrator.
0.000 V. The Secondary display will show the
last 3 digits of the A/D reading. Set the 5520A to
output 0.000 V into the mA Calibrator.
C. After a moment, 25.000 V will
display on the mA Calibrator. Set the 5520A to
output 25.000 V into the mA Calibrator.
A to complete the adjustment. The
calibration counter is incremented and the
calibration data is stored. The calibration count
will be on the main display and "CAL" on the
secondary display for 2 seconds.
AUX terminals of
MEASURE" and
NORMAL
MEASURE" and
at power up. The
6. Disconnect the HP 3458A from the mA
Calibrator.
8
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.