Fluke 707 Data Sheet

Fluke 707Ex Intrinsically
Safe Loop Calibrator
The fast, one-handed tool for
loop checks in Ex zones
The 707Ex is an intrinsically safe loop calibra­tor for use in explosion endangered areas. The Fluke 707Ex is certified in accordance with the ATEX directive (Ex II 2 G Ex ia IIC T4) in Zones 1 and 2. The 707Ex is also certified to operate in accordance with Factory Mutual N.I. Class 1, Division 2 areas Group A-D. The 707Ex features:
Large display and simple, quick-click push
rotary button for easy one-handed operation.
Simultaneous mA and % readout for quick,
easy, interpretation of readings.
mA accuracy of 0.015 %, superior to other
loop calibrators.
Push button with 25 % steps for fast, easy
linearity checks.
0-100 % “span check” for fast confirmation of
zero and span.
Selectable slow ramp, fast ramp, and step
ramp to provide smooth outputs for valve slewing and loop functional tests.
Internal loop supply, so you can power and
read a transmitter at the same time without carrying a DMM.
Measures V dc to 28 V.
Innovative output adjustment dial with
.001 mA and 0.1 mA resolution.
HART® compatible resistance is connected in
series with the loop supply for compatibility with HART communicators.
Technical Data
ATEX and Factory Mutual Certified
Summary Specifications (18 ºC to 28 ºC, one year)
Function Range Resolution Accuracy Voltage measure 0 to 28 V 0.001 V ± (0.015 % Rdg + 2 counts) mA measure 0 to 24 mA 0.001 mA ± (0.015 % Rdg + 2 LSD) mA source 0 to 24 mA1 0.001 mA ± (0.015 % Rdg + 2 LSD) mA simulate 0 to 24 mA2 0.001 mA ± (0.015 % Rdg + 2 LSD) Loop supply 24 V dc n/a 24 V ± 1 V dc, no load
Temperature Coefficient, -10 to 18 ºC, 28 to 55 ºC, ± .005 % of range per ºC;
1
Max load, 700 Ohms at 20 mA
2
Max applied voltage for simulation, 28 V
General Specifications
Maximum voltage: 28 volts Non-operating temperature: -30 to 60 ºC Operating temperature: -10 to 50 ºC Relative humidity: 0 to 95 % (0 to
30 ºC); 0 to 75 % (30 to 40 ºC); 45 % up to 50 ºC
Operating altitude: 3,000 meters max. Shock and vibration: Per MIL-T-28800 for a
Class 2 Instrument
Safety: CSA C22.2 No. 1010.1: 1992 EMC: EN50082-1:1992 and EN55022:1994
Class B ATEX certification: Ex II 2 G Ex ia IIC T4 Zones 1 and 2, ZELM 02 ATEX 0120 X FM certification: N.I. Class 1, Division 2, Groups A-D
CE certification: EN61010-1 and EN61326 Size, without holster: (H x W x D)
6.0 in x 2.8 in x 1.6 in (152 mm x 70 mm x 41 mm) Size, with holster: (H x W x D)
6.4 in x 2.9 in x 1.9 in (164 mm x 75 mm x 47 mm) Weight: 0.28 kg (0.6 lb), with holster and stand: .92 lb (0.42 kg)
Battery: One 9 V alkaline Battery life: 18 hours typical, at 12 mA into
500 ohms
Warranty: 1 year
Operating Modes
Measure current Measure current/Loop power Measure voltage Simulate current; manual adjust; scroll up/down Simulate current; 25% steps; 0 - 100% steps Simulate current; Auto ramp (smooth/step) Source current; manual adjust; scroll up/down Source current; 25% steps; 0 - 100% steps Source current; Auto ramp (smooth/step)
Ordering information
Model
Fluke-707Ex Intrinsically Safe Loop
Calibrator
Each calibrator includes:
Safety designed test leads, alligator test clips, hol­ster, instruction manual on CD (English, French, German, Spanish, Italian, Dutch, Norwegian, Danish, Swedish, Finnish, Portuguese, Korean, Chinese, Japanese), 707Ex CCD Content Control Drawing, CE, CSA, FM and ATEX markings.
Fluke. Keeping your world up and running.
Fluke Corporation
PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or Fax (425) 446-5116 In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222 In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116 Web access: http://www.fluke.com
©2003-2010 Fluke Corporation. Specifications subject to change without notice. Printed in U.S.A. 7/2010 2072174B D-EN-N
Modification of this document is not permitted without written permission from Fluke Corporation.
®
Loading...