Accessories and Ordering Information

Included Accessories:

FLUKE 705: C10 Protective Yellow Holster, TL75 Test Leads, AC70 Test Clips, Instruction Sheet (14 languages).

FLUKE 715: C81Y Protective Yellow Holster with Flex-Stand, Test Leads and Alligator Clips, Single 9V alkaline battery, Instruction Sheet (14 languages).

FLUKE 787: C81Y Protective Yellow Holster with Flex-Stand, TL75 SafetyDesigned Test Lead Set plus AC70A Alligator Clips, one 9V alkaline battery, Users Manual.

Ordering information:

FLUKE 705 Loop Calibrator FLUKE 715 Volt/mA Calibrator FLUKE 787 ProcessMeter

Pub_ID: 10034-eng

Rev. 02

®

Loop Calibrators

Low Cost Calibrator Muscle

Fluke 705

A large display and simple, pushbutton interface for ease of use.

Simultaneous mA and % readout for quick, easy, interpretation of readings.

mA accuracy of 0.02%.

Push button 25% steps for fast, easy linearity checks. Save time over traditional knob type controls.

"Span Check" for fast confirmation of zero and span.

Selectable slow ramp, fast ramp, and step to provide smooth outputs for valve slewing and loop functional tests.

24 volt internal loop supply, so you can power and read a transmitter at the same time without carrying a DMM.

0 - 20 mA or 4 - 20 mA default start up modes.

715 Volt/mA Calibrator

Measure loop current (0-20 mA, 4-20 mA) signals with 0.015% accuracy and 0.001 mA resolution

Measure voltage output process signals from PLCs, transmitters

Source or simulate 24 mA loop current

Source voltage to 100 mV or 10V

24V loop supply with simultaneous current measurement

Fluke 787

DMM and loop calibrator in one tool

Precision 1000V, 440 mA digital multimeter

DC current source/loop calibrator

Clear LCD with backlight; 4000 counts (30 000 counts for DC current)

Externally accessible battery for easy battery changes

1000V overload protection on V, ohms, frequency

Meets 1000 volt EN61010-1 CAT III standards

Min/Max/Average/Hold/Relative modes

Features

Functions

705

715

787

 

Voltage

 

 

 

 

measurement

 

 

 

 

Range

0-28 V DC

0-100 mV 0-10 V

0-1000 V AC or DC

Resolution

1 mV

10 µV 1 mV

0.1 mV to 1.0 V

Accuracy

0.025% Rdg + 1 LSD

0.02% Rdg + 2 LSD

0.1% Rgd + 1 LSD (VDC)

Current

 

 

 

 

measurement

 

 

 

 

Range

0-24 mA

0-24 mA

0-1 A

0-30 mA

Resolution

0.001 mA

0.001 mA

1 mA

0.001 mA

Accuracy

0.02% + 2 LSD

0.015%

0.2%+2 LSD

0.05%+2 LSD

Current sourcing

 

 

 

 

Range

0-20 mA or 4-20 mA1

0-20 mA or 4-20 mA1

0-20 mA or 4-20 mA1

Accuracy

0.025% Rdg + 2 LSD

0.015% + 2 LSD

0.05% of span

Drive capability

1000 @ 24 mA

1000 @ 24 mA

500 ΩΩ @ 24 mA

Loop power while

 

 

 

 

measuring mA

24 V

24 V

N/A

 

Voltage sourcing

N/A

0-100 mV or 0-10 V

N/A

 

Resistance

 

 

to 40 M, 0.2% + 1 LSD

measurement

N/A

N/A

Frequency

N/A

N/A

to 19.999 kHz,

 

 

 

0.005% + 1 LSD

Diode test

N/A

N/A

2.4 V shows

 

 

 

 

diode voltage drop

Continuity

N/A

N/A

Beeps for

 

 

 

 

resistance <100 ohms

Display current and

 

 

 

 

% of span

Yes

mA or %

Yes

 

Auto step, auto ramp

Yes

No

Yes

 

Span Check

Yes

No

No

 

Battery life

18 hours typical,

18 hours typical,

50 hours typical,

 

at 12 mA

at 12 mA

12 hours sourcing 12 mA

 

 

 

 

 

1 will over-range to 24 mA