Accessories and Ordering Information
Included Accessories:
FLUKE 705: C10 Protective Yellow Holster, TL75 Test Leads, AC70 Test Clips, Instruction Sheet (14 languages).
FLUKE 715: C81Y Protective Yellow Holster with Flex-Stand, Test Leads and Alligator Clips, Single 9V alkaline battery, Instruction Sheet (14 languages).
FLUKE 787: C81Y Protective Yellow Holster with Flex-Stand, TL75 SafetyDesigned Test Lead Set plus AC70A Alligator Clips, one 9V alkaline battery, Users Manual.
Ordering information:
FLUKE 705 Loop Calibrator FLUKE 715 Volt/mA Calibrator FLUKE 787 ProcessMeter
Pub_ID: 10034-eng
Rev. 02
®
Loop Calibrators
Low Cost Calibrator Muscle
Fluke 705
●A large display and simple, pushbutton interface for ease of use.
●Simultaneous mA and % readout for quick, easy, interpretation of readings.
●mA accuracy of 0.02%.
●Push button 25% steps for fast, easy linearity checks. Save time over traditional knob type controls.
●"Span Check" for fast confirmation of zero and span.
●Selectable slow ramp, fast ramp, and step to provide smooth outputs for valve slewing and loop functional tests.
●24 volt internal loop supply, so you can power and read a transmitter at the same time without carrying a DMM.
●0 - 20 mA or 4 - 20 mA default start up modes.
715 Volt/mA Calibrator
●Measure loop current (0-20 mA, 4-20 mA) signals with 0.015% accuracy and 0.001 mA resolution
●Measure voltage output process signals from PLCs, transmitters
●Source or simulate 24 mA loop current
●Source voltage to 100 mV or 10V
●24V loop supply with simultaneous current measurement
Fluke 787
●DMM and loop calibrator in one tool
●Precision 1000V, 440 mA digital multimeter
●DC current source/loop calibrator
●Clear LCD with backlight; 4000 counts (30 000 counts for DC current)
●Externally accessible battery for easy battery changes
●1000V overload protection on V, ohms, frequency
●Meets 1000 volt EN61010-1 CAT III standards
●Min/Max/Average/Hold/Relative modes
Features
Functions |
705 |
715 |
787 |
|
Voltage |
|
|
|
|
measurement |
|
|
|
|
Range |
0-28 V DC |
0-100 mV 0-10 V |
0-1000 V AC or DC |
|
Resolution |
1 mV |
10 µV 1 mV |
0.1 mV to 1.0 V |
|
Accuracy |
0.025% Rdg + 1 LSD |
0.02% Rdg + 2 LSD |
0.1% Rgd + 1 LSD (VDC) |
|
Current |
|
|
|
|
measurement |
|
|
|
|
Range |
0-24 mA |
0-24 mA |
0-1 A |
0-30 mA |
Resolution |
0.001 mA |
0.001 mA |
1 mA |
0.001 mA |
Accuracy |
0.02% + 2 LSD |
0.015% |
0.2%+2 LSD |
0.05%+2 LSD |
Current sourcing |
|
|
|
|
Range |
0-20 mA or 4-20 mA1 |
0-20 mA or 4-20 mA1 |
0-20 mA or 4-20 mA1 |
|
Accuracy |
0.025% Rdg + 2 LSD |
0.015% + 2 LSD |
0.05% of span |
|
Drive capability |
1000 Ω @ 24 mA |
1000 Ω @ 24 mA |
500 ΩΩ @ 24 mA |
|
Loop power while |
|
|
|
|
measuring mA |
24 V |
24 V |
N/A |
|
Voltage sourcing |
N/A |
0-100 mV or 0-10 V |
N/A |
|
Resistance |
|
|
to 40 MΩ , 0.2% + 1 LSD |
|
measurement |
N/A |
N/A |
||
Frequency |
N/A |
N/A |
to 19.999 kHz, |
|
|
|
|
0.005% + 1 LSD |
|
Diode test |
N/A |
N/A |
2.4 V shows |
|
|
|
|
diode voltage drop |
|
Continuity |
N/A |
N/A |
Beeps for |
|
|
|
|
resistance <100 ohms |
|
Display current and |
|
|
|
|
% of span |
Yes |
mA or % |
Yes |
|
Auto step, auto ramp |
Yes |
No |
Yes |
|
Span Check |
Yes |
No |
No |
|
Battery life |
18 hours typical, |
18 hours typical, |
50 hours typical, |
|
|
at 12 mA |
at 12 mA |
12 hours sourcing 12 mA |
|
|
|
|
|
|
1 will over-range to 24 mA