The Hart Scientific model 7005 and 7015 are very precise constant temperature
baths. An innovative state of the art solid-state temperature controller has been
incorporated which maintains the bath temperature with extreme stability. The
controller uses a microcontroller to execute the many operating functions.
User interface is provided by the 8-digit LED display and four key-switches.
Digital remote communications is optionally available with a RS-232 or
IEEE-488 interface.
1.2Symbols Used
Table 1 lists the International Electrical Symbols. Some or all of these symbols
may be used on the instrument or in this manual.
1 Before You Start
Introduction
Table 1 International Electrical Symbols
SymbolDescription
AC (Alternating Current)
AC-DC
Battery
CE Complies with European Union Directives
DC
Double Insulated
Electric Shock
Fuse
PE Ground
1
Page 10
7005/7015 Calibration Bath
User’s Guide
SymbolDescription
Hot Surface (Burn Hazard)
Read the User’s Manual (Important Information)
Off
On
Canadian Standards Association
OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 re
fers to the level of Impulse Withstand Voltage protection provided. Equipment of
OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from
the fixed installation. Examples include household, office, and laboratory appliances.
C-TIC Australian EMC Mark
The European Waste Electrical and Electronic Equipment (WEEE) Directive
(2002/96/EC) mark.
1.3Safety Information
Use this instrument only as specified in this manual. Otherwise, the protection
provided by the instrument may be impaired.
The following definitions apply to the terms “Warning” and “Caution”.
•
“WARNING” identifies conditions and actions that may pose hazards to
the user.
•
“caution” identifies conditions and actions that may damage the instru
ment being used.
-
-
1.3.1
WARNINGS
BURN HAZARD
Extremely cold temperatures may be present in this equipment. Freezer burns
and frostbite may result if personnel fail to observe safety precautions.
High temperatures may be present in this equipment. Fires and severe burns
may result if personnel fail to observe safety precautions.
2
Page 11
1 Before You Start
Safety Information
ELECTRICAL HAZARD
These guidelines must be followed to ensure that the safety mechanisms in this
instrument will operate properly. This mains supply for the instrument must be
capable of the power requirements for the instrument as listed in Section 2.1,
Specifications. The power cord of the instrument is equipped with a
three-pronged grounding plug for your protection against electrical shock haz
ards. It must be plugged directly into a properly grounded three-prong recepta
cle. The receptacle must be installed in accordance with local codes and
ordinances. Consult a qualified electrician.
DO NOT use an extension cord or adapter plug.
DO use a ground fault interrupt device. This instrument contains a fluid. A
ground fault device is advised in case fluid is present in the electrical system
and could cause an electrical shock.
Always replace the power cord with an approved cord of the correct rating and
type. If you have questions, contact a Authorized Service Center (see section
1.4).
-
-
1.3.2
High voltage is used in the operation of this equipment. Severe injury or death
may result if personnel fail to observe the safety precautions. Before working
inside the equipment, turn off the power and disconnect the power cord.
BATH FLUIDS
Fluids used in this bath may produce noxious or toxic fumes under certain circumstances. Consult the fluid manufacturer’s MSDS (Material Safety Data
Sheet). Proper ventilation and safety precautions must be observed.
The instrument is equipped with a soft cutout (user settable firmware) and a
hard cutout (set at the factory). Check the flash point, boiling point, or other
fluid characteristic applicable to the circumstances of the bath operation. Ensure that the soft cutout is adjusted to the fluid characteristics of the application. As a guideline, the soft cutout should be set 10°C to 15°C below the flash
point of the bath fluid. See section 7.1, Bath Fluid, for specific information on
bath fluids and section 8.8, Cutout.
CAUTIONS
To avoid possible damage to the instrument, follow these guidelines.
Always operate the instrument in room temperatures listed in Section 2.2, Envi
ronmental Conditions. Allow sufficient air circulation by leaving at least 15 cm
(6 inches) of clearance around the instrument.
-
DO NOT overfill the bath. Overflowing fluid may damage the electrical sys
tem. See section 4.3, Bath Preparation and Filling, for specific instructions.
Read section 5, Bath Use, before placing the bath into service.
DO NOT turn the bath on without fluid in the tank.
-
3
Page 12
7005/7015 Calibration Bath
User’s Guide
DO NOT change the values of the bath calibration constants from the factory
set values. The correct setting of these parameters is important to the safety and
proper operation of the bath.
The refrigeration may be damaged or the lifetime shortened if the set-point
temperature is set above 60°C for more than one hour with the refrigeration
manually on. Ensure that the refrigeration is off when the bath is used above
60°C.
The Factory Reset Sequence should be performed only by authorized person
nel if no other action is successful in correcting a malfunction. You must have a
copy of the most recent Report of Test to restore the test parameters.
DO NOT operate this instrument in an excessively wet, oily, dusty, or dirty
environment.
Most probes have handle temperature limits. Be sure that the probe handle tem
perature limit is not exceeded in the air above the instrument.
The instrument and any thermometer probes used with it are sensitive instruments that can be easily damaged. Always handle these devices with care. Do
not allow them to be dropped, struck, stressed, or overheated.
COLD BATHS
Refrigerated baths require that the condensing coil be cleaned periodically. Accumulation of dust and dirt on the condenser will result in premature failure of
the compressor.
This bath has been equipped with a brownout and over voltage protection device as a safety feature to protect the system components.
Mode of Operation: This bath needs to be plugged into the line voltage for at
least 2 minutes before operation. This is only necessary for the first time the
bath is energized or when it is moved from one location to another. Turning the
bath ON or OFF does not trigger the delay.
-
If a High/Low voltage condition exists for longer than 5 seconds, the bath
de-energizes. An amber indicator on the back panel lights when this condition
exists.
Re-energization is automatic upon correction of the fault condition and after a
delay cycle of about 2 minutes. If a fault condition exists upon application of
power, the bath will not energize.
Under and Over Voltage Protection at 115 VAC
•
Voltage Cutout: ±12.5% (101 - 129 VAC)
•
Voltage Cut In: ±7.5% (106 - 124 VAC)
Under and Over Voltage Protection at 230 VAC
•
Voltage Cutout: ±12.5% (203 - 257 VAC)
•
Voltage Cut In: ±7.5% (213 - 247 VAC)
4
Page 13
1.4Authorized Service Centers
Please contact one of the following authorized Service Centers to coordinate
service on your Hart product:
Hart Scientific, Inc.
799 E. Utah Valley Drive
American Fork, UT 84003-9775
USA
Phone: +1.801.763.1600
Telefax: +1.801.763.1010
E-mail: support@hartscientific.com
Fluke Nederland B.V.
1 Before You Start
Authorized Service Centers
Customer Support Services
Science Park Eindhoven 5108
5692 EC Son
NETHERLANDS
Phone: +31-402-675300
Telefax: +31-402-675321
E-mail: ServiceDesk@fluke.nl
Fluke Int'l Corporation
Service Center - Instrimpex
Room 2301 Sciteck Tower
22 Jianguomenwai Dajie
Chao Yang District
Beijing 100004, PRC
CHINA
Phone: +86-10-6-512-3436
Telefax: +86-10-6-512-3437
E-mail: xingye.han@fluke.com.cn
Fluke South East Asia Pte Ltd.
Fluke ASEAN Regional Office
Service Center
5
Page 14
7005/7015 Calibration Bath
User’s Guide
60 Alexandra Terrace #03-16
The Comtech (Lobby D)
118502
SINGAPORE
Phone: +65 6799-5588
Telefax: +65 6799-5588
E-mail: antng@singa.fluke.com
When contacting these Service Centers for support, please have the following
information available:
Model Number
•
• Serial Number
Voltage
•
• Complete description of the problem
6
Page 15
2 Specifications and Environmental Conditions
2Specifications and Environmental
Conditions
2.1Specifications
Table 2 Specifications
Specifications
Range
Stability at 25°C
Uniformity at 25°C
Temperature
Setting
Set-point
Resolution
Display
Resolution
Digital Setting
Accuracy
Digital Setting
Repeatability
Heaters
Cooling Capacity
Safety Cutout
Access Opening
0°C to 110°C (32°F to 230°F)
±0.0007°C (water)
±0.001°C (mineral oil 5011)
±0.003°C (water)
±0.005°C (mineral oil 5011)
Digital display with push-button data entry
0.002°C; high-resolution mode, 0.00003°C
0.01°C
±1°C
±0.01°C
500 and 1000 Watts
100 to 200 Watts
Factory-set high temperature
699 mm X 279 mm (27.5” X 11”)
Bath Chamber
Dimensions
(unobstructed
space)
spnumDepth
Wetted Parts
Volume
Weight
Size
Power
279 mm H X 699 mm W X 331 mm D
(11” X 27.5” X 13”)
331 mm (13”)
304 stainless steel
95 liters (25 gallons)
141 kg (310 lb.)
1219 mm H X 1118 mm W X 559 mm D
(48” X 44” X 22”)
115 VAC (±10%), 60 Hz, 1810 W
230 VAC (±10%), 60 Hz, 1810 W (50 Hz Optional)
NOTE: If voltage is outside ±10% VAC, the compressor may be damaged.
Check the back panel label for the correct voltage and frequency prior to energizing
the instrument.
7
Page 16
7005/7015 Calibration Bath
User’s Guide
System Fuse
Heater Fuse
Safety
115 VAC: 20 A 250 V slow blow
230 VAC: 10 A 250 V slow blow
115 VAC: 10 A 250 V fast blow
230 VAC: 6 A 250 V fast blow
OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC 61010-1
2.2Environmental Conditions
Although the instrument has been designed for optimum durability and trou
ble-free operation, it must be handled with care. The instrument should not be
operated in an excessively dusty or dirty environment. Maintenance and clean
ing recommendations can be found in the Maintenance Section of this manual.
The instrument operates safely under the following conditions:
• ambient temperature range: 5–50°C (41–122°F)
• ambient relative humidity: maximum 80% for temperature <31°C, de-
creasing linearly to 50% at 40°C
• pressure: 75kPa–106kPa
-
-
• mains voltage within ±10% of nominal
• vibrations in the calibration environment should be minimized
• altitude does not effect the performance or safety of the unit
• indoor use only
8
Page 17
3Quick Start
CAUTION: READ SECTION 5 ENTITLED BATH USE before placing the
bath in service. Incorrect handling can damage the bath and void the war
ranty.
This section gives a brief summary of the steps required to set up and operate
the 7005/7015 bath. This should be used as a general overview and reference
and not as a substitute for the remainder of the manual. Please read Section 4
through 6 carefully before operating the bath.
3.1Unpacking
Unpack the bath carefully and inspect it for any damage that may have occurred during shipment. If there is shipping damage, notify the carrier
immediately.
3 Quick Start
Unpacking
-
Verify that all components are present:
• 7005/7015 Bath
• Controller Probe
• Access Hole Cover
• User’s Guide
• Report of Test
• Drain Elbow
If you are missing any item, please contact an Authorized Service Center.
3.2Set Up
Set up of the bath requires careful unpacking and placement of the bath, filling
the bath with fluid, installing the probe and connecting power. Consult Section
4 for detailed instructions for proper installation of the bath. Be sure to place
the bath in a safe, clean and level location.
Fill the bath tank with an appropriate liquid. For operation at moderate bath
temperatures, clean distilled water works well. Carefully pour the fluid into the
bath tank through the large rectangular access hole above the tank avoiding
spilling any fluid. The fluid must not exceed a height of 1/2 inch below the bath
lid.
The control probe must be inserted through the lid into the bath and plugged
into the socket at the back of the bath.
9
Page 18
7005/7015 Calibration Bath
User’s Guide
3.3Power
Plug the bath power cord into a mains outlet of the proper voltage, frequency,
and current capability. Refer to Section 2.1, Specifications, for power
requirements. Set the “HEATER” switch on the front panel to position “LOW”
and turn the bath on using the front panel “POWER” switch. The bath will turn
on and begin to heat or cool to reach the previously programmed temperature
set-point. The front panel LED display will indicate the actual bath
temperature.
3.4Setting the Temperature
In the following discussion a solid box around the word SET, UP, EXIT or
DOWN indicates the panel button while the dotted box indicates the display
reading. Explanation of the button or display reading are to the right of each
button or display value.
To view or set the bath temperature set-point proceed as follows. The front
panel LED display normally shows the actual bath temperature.
24.68 CBath temperature display
When “SET” is pressed the display shows the set-point memory that is currently being used and its value. Eight set-point memories are available.
S
Access set-point selection
1. 25.0Set-point 1, 25.0°C currently used
Press “SET” to select this memory and access the set-point value.
S
Access set-point value
C 25.00Current value of set-point 1, 25.00°C
Press “UP” or “DOWN” to change the set-point value.
U
Increment display
C 30.00New set-point value
10
Press “SET” to accept the new value and display the vernier value. The bath be
gins heating or cooling to the new set-point.
S
Store new set-point, access vernier
-
Page 19
0.00000Current vernier value
Press “EXIT” and the bath temperature is displayed again.
3 Quick Start
Setting the Temperature
E
Return to the temperature display
24.73 CBath temperature display
The bath heats or cools until it reaches the new set-point temperature. Set the
heater switch to position “HIGH” to allow the bath to more quickly reach a
higher temperature. The “HIGH” setting may be necessary to reach higher tem
peratures and control at high temperatures.
When setting the set-point temperature be careful not to exceed the temperature
limit of the bath fluid. The over-temperature cut-out should be correctly set to
prevent this from happening. See Section 8.8.
If operating the bath below 45 °C set the COOLING power switch to ON. The
cooling temperature may require adjustment to provide the proper amount of
cooling. See Section 7.5.
To obtain optimum control stability adjust the proportional band as discussed in
Section 8.7.
-
11
Page 20
4Installation
CAUTION: READ SECTION 5 ENTITLED BATH USE before placing the
bath in service. Incorrect handling can damage the bath and void the war
ranty.
This bath is not designed to be portable. Therefore, moving the bath once it has
been installed should be kept to a minimum.
WARNING: Never move a bath that is full of fluid. This action could be
extremely dangerous and could result in personal injury to the person
moving the bath.
If the bath is going to be placed in an area where it will need to be moved frequently, a special cart can be designed to accommodate the bath making the
bath much more portable. Hart sells carts designed for these baths. However,
even with a cart the bath should not be moved full of fluid. The fluid can splash
causing injury or if the bath and cart tip, the fluid could cause damage to the
surrounding area and personal injury to personnel.
4 Installation
Bath Environment
-
If the bath must be moved, be sure to drain the fluid to prevent any injury. The
side of the bath with the compressor is heavier than the tank side. To safely
move the bath, two people are required.
4.1Bath Environment
The 7005 and 7015 baths are precision instruments which should be located in
an appropriate environment. The location should be free of drafts, extreme temperatures and temperature changes, dirt, etc. The surface where the bath is
placed must be level.
If used at higher temperatures where fluid vaporization is significant, a fume
hood should be used.
4.2“Dry-out” Period
Before initial use, after transport, and any time the instrument has not been en
ergized for more than 10 days, the bath will need to be energized for a
“dry-out” period of 1-2 hours before it can be assumed to meet all of the safety
requirements of the IEC 1010-1.
-
4.3Bath Preparation and Filling
The 7005 and 7015 baths are designed for use with standard resistors, there
fore, the soft cutout of the instrument has been set at the factory to 50°C to pro
tect standard resistors placed in the bath. In order to use the bath for other
-
-
13
Page 21
7005/7015 Calibration Bath
User’s Guide
calibration purposes, the user should insure that the soft cutout is adjusted ap
propriately for the application.
The 7005 and 7015 baths are not provided with a fluid. Various fluids are avail
able from Hart Scientific and other sources. Depending on the desired tempera
ture range, any of the following fluids, as well as others, may be used in the
bath:
Water
•
Ethylene Glycol/Water
•
Methanol
•
Mineral oil
•
Silicone oil
•
Fluids are discussed in detail in Section 7.1.
Remove any access hole cover from the bath and check the tank for foreign
matter (dirt, remnant packing material, etc.). Use clean unpolluted fluid. Carefully fill the bath through the large square access hole to a level that will allow
for stirring and thermal expansion. The fluid should never exceed a height of
1/2" below the top of the tank. Carefully monitor the bath fluid level as the bath
temperature rises to prevent overflow or splashing. Remove excess fluid if necessary and with caution if the fluid is hot.
-
-
-
CAUTION: Be careful to prevent bath fluid from spilling on the stirring
motor while filling.
NOTE: Underfilling may reduce bath performance and may possibly
damage the bath heater.
4.4Probe
Inspect the bath controller probe. This probe should not be bent or damaged in
any way. Reasonable caution should be used in handling this probe as it con
tains a precision thermistor sensor. If damaged, the probe can be replaced. Con
tact an Authorized Service Center for assistance.
Insert the probe into the 1/4 inch probe hole at the top left side of the bath lid.
The tip of the probe must be well immersed in the fluid. The probe connector is
plugged into the rear of the bath into the socket labelled “PROBE”.
4.5Power
With the bath power switch off, plug the bath into an AC mains outlet of the
appropriate voltage, frequency, and current capacity. Refer to Section 2.1,
Specifications, for power requirements.
-
-
14
Page 22
4 Installation
Power
Be sure the stirring motor power cord is plugged into the “STIRRER” socket at
the back of the bath.
15
Page 23
5Bath Use
READ BEFORE PLACING THE BATH IN SERVICE
5 Bath Use
General
The information in this section is for general information only. It is not de
signed to be the basis for calibration laboratory procedures. Each laboratory
will need to write their own specific procedures.
5.1General
Be sure to select the correct fluid for the temperature range of the calibration.
Bath fluids should be selected to operate safely with adequate thermal proper
ties to meet the application requirements. Also, be aware that some fluids ex
pand and could overflow the bath if not watched. Refer to General Operation,
section 8, for information specific to fluid selection and to the MSDS sheet specific to the fluid selected. Generally, baths are set to one temperature and used
to calibrate probes only at that single temperature. This means that the type of
bath fluid does not have to change. Additionally, the bath can be left energized
reducing the stress on the system.
The bath generates extreme temperatures. Precautions must be taken to prevent
personal injury or damage to objects. Probes may be extremely hot or cold
when removed from the bath. Cautiously handle probes to prevent personal injury. Carefully place probes on a heat/cold resistant surface or rack until they
are at room temperature. It is advisable to wipe the probe with a clean soft cloth
or paper towel before inserting it into another bath. This prevents the mixing of
fluids from one bath to another. If the probe has been calibrated in liquid salt,
carefully wash the probe in warm water and dry completely before transferring
it to another fluid. Always be sure that the probe is completely dry before in
serting it into a hot fluid. Some of the high temperature fluids react violently to
water or other liquid mediums. Be aware that cleaning the probe can be danger
ous if the probe has not cooled to room temperature. Additionally, high temper
ature fluids may ignite the paper towels if the probe has not been cooled.
-
-
-
-
-
-
For optimum accuracy and stability, allow the bath adequate stabilization time
after reaching the set-point temperature.
5.2Comparison Calibration
Comparison calibration involves testing a probe (unit under test, UUT) against
a reference probe. After inserting the probes to be calibrated into the bath, al
low sufficient time for the probes to settle and the temperature of the bath to
stabilize.
One of the significant dividends of using a bath rather than a dry-well to cali
brate multiple probes is that the probes do not need to be identical in construc
tion. The fluid in the bath allows different types of probes to be calibrated at
-
-
-
17
Page 24
7005/7015 Calibration Bath
User’s Guide
the same time. However, stem effect from different types of probes is not to
tally eliminated. Even though all baths have horizontal and vertical gradients,
these gradients are minimized inside the bath work area. Nevertheless, probes
should be inserted to the same depth in the bath liquid. Be sure that all probes
are inserted deep enough to prevent stem effect. From research at Hart Scien
tific, we suggest a general rule-of-thumb for immersion depth to reduce the
stem effect to a minimum: 15 x the diameter of the UUT + the sensor length.
Do not submerge the probe handles. If the probe handles get too warm dur
ing calibration at high temperatures, a heat shield could be used just below the
probe handle. This heat shield could be as simple as aluminum foil slid over the
probe before inserting it in the bath or as complicated as a specially designed
reflective metal apparatus.
When calibrating over a wide temperature range, better results can generally be
achieved by starting at the highest temperature and progressing down to the
lowest temperature.
Probes can be held in place in the bath by using probe clamps or drilling holes
in the access cover. Other fixtures to hold the probes can be designed. The object is to keep the reference probe and the probe(s) to be calibrated as closely
grouped as possible in the working area of the bath. Bath stability is maximized
when the bath working area is kept covered.
In preparing to use the bath for calibration start by:
• Placing the reference probe in the bath working area.
• Placing the probe to be calibrated, the UUT, in the bath working area as
close as feasibly possible to the reference probe.
5.3Calibration of Multiple Probes
Fully loading the bath with probes increases the time required for the temperature to stabilize after inserting the probes. Using the reference probe as the
guide, be sure that the temperature has stabilized before starting the calibration.
18
Page 25
Page 26
7005/7015 Calibration Bath
User’s Guide
the heater is ON. When the indicator is green the heater is OFF and the bath is
cooling.
(5) The heater power switch is used to select the appropriate heater power lev
els for heating and controlling the bath at various temperatures.
(6) The cooling power switch controls power to the cooling compressor and
cooling fan.
6.2Side Panel
The side panel has three features (see Figure 2): 1) the back pressure valve, 2)
the cooling temperature regulating valve, and 3) the cooling temp gauge. With
the interface option an extra cooling valve (HIGH) is provided.
-
20
Figure 2. Refrigeration Controls - Side Panel
The back pressure valve adjustment is used to control the amount of
1)
cooling supplied to the system. This valve reduces the cooling capacity
by restricting the flow of refrigerant to the bath, allowing the adjustment
of the heating to cooling percentage. Under normal operation the valve
should be fully open (counter clockwise).
Page 27
6 Parts and Controls
Back Panel
2)The cooling temperature regulating valve is used to adjust the tempera
ture at which the refrigerant evaporates, which determines cooling effi
ciency. The ideal temperature for operation is about 5-10 degrees Celsius
below the desired bath temperature. Refer to the label below the gauge
for approximate psi and evaporative temperature settings. The following
table is reproduced from the label.
For this bath:
Control Temperature
°C°FNominal Cooling PSIG
–10144.7
–5239.2
03214.5
54120.7
105027.9
155936.1
206845.5
257756.2
Set the Cooling Temp. Valve to this pressure
-
-
308668.4
359582.1
3798.686
DO NOT set the Cooling Pressure above 90 PSIG
3)The cooling temp gauge is used to indicate the temperature at which the
refrigerant is evaporating. The cooling temp regulating valve is used to
set and then control this temperature.
6.3Back Panel
The back panel has six standard features and two optional features. (See Figure
3.) 1) the probe connector, 2) the stirrer power outlet, 3) system fuses (inter
nal), 4) the power cord, 5) the drain valve, 6) serial number notation, 7) the
RS-232 interface conector (optional), and 8) the IEEE-488 interface connector
(optional).
1)The probe connector in the back panel is used for the temperature con
troller probe.
2)The stirrer power is provided for the stirring motor.
-
-
3)The system fuses (internal).
4)The power cord.
21
Page 28
7005/7015 Calibration Bath
User’s Guide
5)A drain valve is provided for ease of removing the fluid media from the
bath. Always use a container of adequate size to hold the FULL LOAD
of fluid. Some oils are more easily drained at higher temperatures. (See
caution note in next section.)
6)The serial number is located on the right side of the back panel. When
contacting an Authorized Service Center, use the serial number and
model number.
7)If the bath is supplied with a serial RS-232 interface, the interface cable
attaches to the back of the bath at the connector labeled “RS-232”.
8)If the bath is supplied with a GPIB IEEE-488 interface, the interface ca
ble attaches to the back of the bath at the connector labeled “IEEE-488”.
WARNING: Extreme caution must be maintained to prevent harm to the
user or the surrounding environment. Do not exceed a 120°C fluid temperature for draining. The valve could be damaged if 120°C is exceeded.
Insulate the container from the floor or other objects that may be damaged
by high temperatures as required.
-
22
Page 29
6 Parts and Controls
Back Panel
Figure 3 Back Panel
23
Page 30
7General Operation
7.1Bath Fluid
7 General Operation
Bath Fluid
Many fluids will work with 7005/7015 bath. Choosing a fluid requires consid
eration of many important characteristics of the fluid. Among these are temper
ature range, viscosity, specific heat, thermal conductivity, thermal expansion,
electrical resistivity, fluid lifetime, safety, and cost.
7.1.1Temperature Range
One of the most important characteristics to consider is the temperature range
of the fluid. Few fluids work well throughout the complete temperature range
of the bath. The temperature at which the bath is operated must always be
within the safe and useful temperature range of the fluid. The lower temperature range of the fluid is determined by the freeze point of the fluid or the temperature at which the viscosity becomes too great. The upper temperature is
usually limited by vaporization, flammability, or chemical breakdown of the
fluid. Vaporization of the fluid at higher temperatures may affect temperature
stability because of cool condensed fluid dripping into the bath from the lid.
The bath temperature should be limited by setting the safety cut-out so that the
bath temperature cannot exceed the safe operating temperature limit of the
fluid.
7.1.2Viscosity
-
-
Viscosity is a measure of the thickness of a fluid, how easily it can be poured
and mixed. Viscosity affects the temperature stability of the bath. With low viscosity, fluid mixing is better which creates a more uniform temperature
throughout the bath. This improves the bath response time which allows it to
maintain a more constant temperature. For good control the viscosity should be
less than 10 centistokes. 50 centistokes is about the upper limit of allowable
viscosity. Viscosities greater than this cause very poor control stability and may
also overheat or damage the stirring motor. With oils viscosity may vary greatly
with temperature.
When using fluids with higher viscosities the controller proportional band may
need to be increased to compensate for the reduced response time. Otherwise
the temperature may begin to oscillate.
7.1.3Specific Heat
Specific heat is the measure of the heat storage ability of the fluid. Specific
heat, though to a lesser degree, also affects the control stability and the heating
and cooling rates. Generally, a lower specific heat causes slightly better control
stability and quicker heating and cooling. With fluids with higher specific heat
the controller may require a decreased proportional band to compensate for the
decrease in sensitivity of the bath temperature to heat input.
25
Page 31
7005/7015 Calibration Bath
User’s Guide
7.1.4Thermal Conductivity
Thermal conductivity measures how easily heat flows through the fluid. Ther
mal conductivity of the fluid affects the control stability, temperature unifor
mity, and probe temperature settling time. Fluids with higher conductivity
distribute heat more quickly and evenly improving bath performance.
7.1.5Thermal Expansion
Thermal expansion describes how the volume of the fluid changes with temper
ature. Thermal expansion of the fluid used must be considered since the in
crease in fluid volume as the bath temperature changes may cause overflow.
Excessive thermal expansion may also be undesirable in applications where
constant liquid level is important. Oils typically have significant thermal
expansion.
7.1.6Electrical Resistivity
Electrical resistivity describes how well the fluid insulates against the flow of
electric current. In some applications, such as measuring the resistance of bare
temperature sensors, it may be important that little or no electrical leakage occur through the fluid. In this case consider a fluid with very high resistivity.
7.1.7Fluid Lifetime
-
-
-
-
Many fluids degrade over time because of evaporization, water absorption, gelling, or chemical breakdown. Often the degradation becomes significant near
the upper temperature limit of the fluid.
7.1.8Safety
When choosing a fluid always consider the safety issues associated. Obviously,
where there are extreme temperatures there can be danger to personnel and
equipment. Fluids may also be hazardous for other reasons. Some fluids may
be considered toxic. Contact with eyes, skin, or inhalation of vapors may cause
injury. A proper fume hood must be used if hazardous or bothersome vapors
are produced.
WARNING: Fluids at high temperatures. May pose danger from BURNS,
FIRE, and TOXIC fumes. Use appropriate caution and safety equipment.
Fluids may be flammable and require special fire safety equipment and proce
dures. An important characteristic of the fluid to consider is the flash point. The
flash point is the temperature at which there is sufficient vapor given off so that
when there is sufficient oxygen present and an ignition source is applied the va
por will ignite. This does not necessarily mean that fire will be sustained at the
flash point. The flash point may be either of the open cup or closed cup type.
Either condition may occur in a bath situation. The closed cup temperature is
always the lower of the two. The closed cup represents the contained vapors in
-
-
-
26
Page 32
7 General Operation
Bath Fluid
side the tank and the open cup represents the vapors escaping the tank. Oxygen
and an ignition source will be less available inside the tank.
Environmentally hazardous fluids require special disposal according to applica
ble federal or local laws after use.
7.1.9Cost
Cost of bath fluids may vary greatly, from cents per gallon for water to hun
dreds of dollars per gallon for synthetic oils. Cost may be an important consid
eration when choosing a fluid.
7.1.10Commonly Used Fluids
Below is a description of some of the more commonly used fluids and their
characteristics
7.1.10.1Water
Water is often used because of its very low cost, availability, and excellent temperature control characteristics. Water has very low viscosity and good thermal
conductivity and heat capacity which makes it among the best fluids for control
stability at low temperatures. Temperature stability is much poorer at higher
temperatures because water condenses on the lid, cools and drips into the bath.
Water is safe and relatively inert. The electrical conductivity of water may prevent its use in some applications. Water has a limited temperature range, from a
few degrees above 0°C to a few degrees below 100°C. At higher temperatures
evaporation becomes significant. Water used in the bath should be distilled or
softened to prevent mineral deposits. Consider using an algicide chemical in the
water to prevent contamination.
-
-
-
7.1.10.2Ethylene Glycol
The temperature range of water may be extended by using a solution of 1 part
water and 1 part ethylene glycol (antifreeze). The characteristics of the ethylene
glycol-water solution are similar to water. Use caution with ethylene glycol
since the fluid is very toxic. Ethylene glycol must be disposed of properly.
7.1.10.3Methanol
Methanol or methyl alchohol is often used at low temperatures below 0°C.
Methanol is relatively inexpensive, has good control characteristics, and has a
low freeze point. Methanol is very toxic so care must be taken when using and
disposing of this fluid.
7.1.10.4Mineral Oil
Mineral oil or paraffin oil is often used at moderate temperatures above the
range of water. Mineral oil is relatively inexpensive. At lower temperatures
mineral oil is quite viscous and control may be poor. At higher temperatures
vapor emission becomes significant. The vapors may be dangerous and a fume
27
Page 33
7005/7015 Calibration Bath
User’s Guide
hood should be used. As with most oils mineral oil will expand as temperature
increases so be careful not to fill the bath too full that it overflows when heated.
The viscosity and thermal characteristics of mineral oil is poorer than water so
temperature stability will not be as good. Mineral oil has very low electrical
conductivity. Use caution with mineral oil since it is flammable and may also
cause serious injury if inhaled or ingested.
7.1.10.5Silicone Oils
Silicone oils are available which offer a much wider operating temperature
range than mineral oil. Like most oils, silicone oils have temperature control
characteristics which are somewhat poorer than water. The viscosity changes
significantly with temperature and thermal expansion also occurs. These oils
have very high electrical resistivity. Silicon oils are fairly safe. These oils are
relatively expensive.
7.1.11Fluid Characteristics Charts
Table 3 and Figure 4 have been created to provide help in selecting a heat exchange fluid media for your constant temperature bath. The charts provide both
a visual and numerical representation of most of the physical qualities important in making a selection. The list is not all inclusive, many usable fluids may
not have been shown in this listing.
7.1.11.1Limitations and Disclaimer
Every effort has gone into making these charts accurate, however, the data here
does not imply any guarantee of fitness of use for a particular application.
Working near the limits of a property such as the flash point or viscosity limit
can compromise safety or serviceability. Sources of information sometimes
vary for particular properties. Your company’s safety policies as well as personal judgment regarding flash points, toxicity, etc. must also be considered.
You are responsible for reading the Material Safety Data Sheets and making a
judgment here. Cost may require some compromises as well. Hart Scientific
cannot be liable for the suitability of application or for any personal injury,
damage to equipment, product or facilities in using these fluids.
The charts include information on a variety of fluids which are often used as
heat transfer fluid in baths. Because of the temperature range some fluids may
not be useful with your bath.
7.1.11.2About the Graph
The fluid graph visually illustrates some of the important qualities of the fluids
shown.
Temperature Range: The temperature scale is shown in degrees Celsius. A
sense of the fluid’s general range of application is indicated. Qualities including
pour point, freeze point, important viscosity points, flash point, boiling point
and others may be shown.
28
Page 34
7 General Operation
Bath Fluid
–100°C0°C100°C200°C300°C400°C500°C600°C
Silicone Oil
5017
Silicone Oil
5014
Silicone Oil
5013
Silicone Oil
5012
Silicone Oil
5010
HFE 7500
5023
Halocarbon
5019
Mineral Oil
5011
Water
Methanol
Ethanol
Ethylene Glycol
(50/50 with H O)
Dynalene HF/LO
2
5020
5022
PP
PP
PP
PP
10 CS
FR (Pure)
10 CS
10 CS
FR
10 CS
FR
10 CS
FL 12°C
FL 16°C
10 CS
10 CS
10 CS
EP
10 CS
BP
FL 60°C
FL 133°C
EP 100°C
BP
BP
10 CS
10 CS
FL 302°CPP
FL 280°C
FL 232°C
FL 211°C
FL 177°C
Bath Salt
5001
FR
Legend
Shaded area represents usable range of fluid starting at
50 centistokes. Lighter shading represents decreasing
viscosity, while vaporization and decomposition increase.
Black area represents liquid range with excessive
viscosity.
Range over which a fume hood is recommended.
Figure 4 Chart of various bath fluids and their properties
Freezing Point: The freezing point of a fluid is an obvious limitation to stir
ring. As the freezing point is approached high viscosity may also limit good
stirring.
Pour Point: This represents a handling limit for the fluid.
Viscosity: Points shown are at 50 and 10 centistokes. Greater than 50 centi
stokes stirring is very poor and unsatisfactory for bath applications. At 10
Decomposition Starts
BP - Boiling Point
CS - Centistokes
EP - Evaporation Point
(fluid loss
due to evaporation)
FL - Flash Point
FR - Freeze Point
PP - Pour Point
-
-
29
Page 35
7005/7015 Calibration Bath
User’s Guide
Table 3 Table of various bath fluids and their properties
Fluid
(# = Hart Model
No.)
Halocarbon 0.8
#5019
Methanol–96°C (fr)60°C (b)54°C1.3 @ –35°C
Water0°C (fr)95°C (b)NONE1 @ 25°C
Ethylene
Glycol—50%
#5020
Mineral Oil40°C (v)190°C (fl)190°C15 @ 75°C
Dow Corning
200.5
Silicone Oil
Dow Corning
200.10
#5012
Dow Corning
200.20
#5013
Dow Corning
200.50
Silicone Oil
Dow Corning 550
#5016
Dow Corning 710
#5017
Dow Corning
210-H
Silicone Oil
Heat Transfer
Salt
#5001
*Limiting Factors — b - boiling point e - high evaporation fl - flash point fr - freeze point v - viscosity — Flash point test oc = open cup cc = closed cup
**Very low water solubility, ice will form as a slush from condensation below freezing.
centistokes and below optimum stirring can occur. These are rules of thumb
which have been useful for most applications.
Fume Point: The point at which a fume hood should be used. This point is
very subjective in nature and is impacted by individual tolerance to different
fumes and smells, how well the bath is covered, the surface area of the fluid in
the bath, the size and ventilation of the facility where the bath is located and
others. We assume the bath is well covered at this point. This is also subject to
company policy.
Page 36
7 General Operation
Stirring
Flash Point: The point at which ignition may occur. See flash point discussion
in Section 7.1.8. The point shown may be either the open or closed cup flash
point.
Boiling Point: At the boiling point of the fluid the temperature stability is diffi
cult to maintain. Fuming is excessive. Excessive amounts of heater power may
be required because of the heat of vaporization.
Decomposition: All high temperature fluids may be reach a temperature point
at which decomposition of some form will begin. While it always begins slowly
at some lower temperature, the rate can increase to the point of danger or im
practicality at a higher temperature.
7.2Stirring
Stirring of the bath fluid is very important for stable temperature control. The
fluid must be mixed well for good temperature uniformity and fast controller
response. The stirrer is precisely adjusted for optimum performance.
7.3Power
Power to the bath is provided by an AC mains supply. Refer to Section 2.1,
Specifications, for power requirements. Power to the bath passes through a filter to prevent switching spikes from being transmitted to other equipment.
To turn on the bath switch the control panel power switch to the ON position.
The stir motor will turn on, the LED display will begin to show the bath temperature, and the heater will turn on or off until the bath temperature reaches
the programmed set-point.
-
-
When powered on the control panel display briefly shows a four digit number.
This number indicates the number of times power has been applied to the bath.
Also briefly displayed is data which indicates the controller hardware configu
ration. This data is used in some circumstances for diagnostic purposes.
7.4Heater
The power to the bath heater is precisely controlled by the temperature control
ler to maintain a constant bath temperature. Power is controlled by periodically
switching the heater on for a certain amount of time using a solid-state relay.
The front panel red/green control indicator shows the state of the heater. The
control indicator glows red when the heater is on and glows green when the
heater is off. The indicator will pulse constantly when the bath is maintaining a
stable temperature.
The heater has two power level settings. The “HIGH” heater power setting is
used to quickly heat up the bath fluid to the desired operating temperature. The
“HIGH” heater power setting may also be required for control at high tempera
tures. The “LOW” setting is used for control at lower temperatures and for
-
-
-
31
Page 37
7005/7015 Calibration Bath
User’s Guide
scanning at slower rates. When controlling at the “HIGH” heater power setting
instead of “LOW” the proportional band may need to be increased (typically by
a factor of four) to compensate for the increase in power gain. Otherwise the
temperature may oscillate.
7.5Cooling
The back pressure control valve limits the cooling capacity of the unit. It will
normally be open all the way (full CCW) for temperature slewing and opera
tion. If during operation the front panel meter indicates excessive cooling, this
valve is closed partially (turn CW) until the percentage of heating to cooling is
brought into line. It is necessary to wait a few minutes after each adjustment
until the system settles.
-
Set the COOLING TEMPERATURE to 5-10°C below the SET TEMPERA
TURE for near ambient and below operation. For temperature above 16°C set
the COOLING TEMPERATURE at 7°C. The cooling temperature should NOT
be set above 7°C or 90 psig. About 45°C the refrigeration will not be required
as there is sufficient cooling to the room. Wait a few minutes after each adjustment until the system establishes itself. Readjust if required after settling for a
while.
For maximum cooling for slewing to lower temperatures, the BACK PRESSURE valve should be fully open and the COOLING TEMPERATURE set to
0°C. Readjust higher or lower as required after set temperature is reached.
7.6Fluid Drain
The drain at the back of the bath (see Figure 3 and Figure on pages 23 and )
may be used to remove fluid from the bath. During operation of the bath the
drain must be closed.
7.7Temperature Controller
The bath temperature is controlled by Hart Scientific’s unique hybrid digi
tal/analog temperature controller. The controller offers the tight control stability
of an analog temperature controller as well as the flexibility and programmabil
ity of a digital controller.
-
-
-
32
The bath temperature is monitored with a thermistor resistance sensor in the
control probe. The signal is electronically compared with the programmable
reference signal, amplified, and then fed to a pulse-width modulator circuit
which controls the amount of power applied to the bath heater.
The bath is operable within the temperature range given in the specifications.
For protection against solid-state relay failure or other circuit failure, the
microcontroller automatically turns off the heater with a second mechanical re
lay anytime the bath temperature is more than a certain amount above the
set-point temperature. As a second protection device, the controller is also
-
Page 38
7 General Operation
Temperature Controller
equipped with a separate thermocouple temperature monitoring circuit which
shuts off the heater if the temperature exceeds the cut-out set-point.
The controller allows the operator to set the bath temperature with high resolu
tion, set the cut-out, adjust the proportional band, monitor the heater output
power, and program the controller configuration and calibration parameters.
The controller may be operated in temperature units of degrees Celsius or Fahr
enheit. The controller is operated and programmed from the front control panel
using the four key switches and digital LED display. The controller may also be
optionally equipped with a serial RS-232 or IEEE-488 GPIB digital interface
for remote operation. Operation of the controller using the front control panel is
discussed in Section 8. Operation using the digital interfaces is discussed in
Section 9.
When the controller is set to a new set-point the bath heats or cool to the new
temperature. Once the new temperature is reached the bath usually takes 10-15
minutes for the temperature to settle and stabilize. There may be a small over
-
shoot or undershoot of about 0.5°C.
-
33
Page 39
8Controller Operation
This section discusses in detail how to operate the bath temperature controller
using the front control panel. Using the front panel key switches and LED dis
play the user may monitor the bath temperature, set the temperature set-point in
degrees C or F, monitor the heater output power, adjust the controller propor
tional band, set the cutout set-point, and program the probe calibration parame
ters, operating parameters, serial and IEEE-488 interface configuration, and
controller calibration parameters. Operation is summarized in Figure 5.
8.1Bath Temperature
The digital LED display on the front panel allows direct viewing of the actual
bath temperature. This temperature value is what is normally shown on the dis
play. The units, C or F, of the temperature value are displayed at the right. For
example,
8 Controller Operation
Bath Temperature
-
-
-
-
25.00 CBath temperature in degrees Celsius
The temperature display function may be accessed from any other function by
pressing the “EXIT” button.
8.2Reset Cutout
If the over-temperature cutout has been triggered then the temperature display
will alternately flash,
Cut-outIndicates cutout condition
The message continues to flash until the temperature is reduced and the cutout
is reset.
The cutout has two modes — automatic reset and manual reset. The mode de
termines how the cutout is reset which allows the bath to heat up again. When
in automatic mode, the cutout resets itself as soon as the temperature is lowered
below the cutout set-point. With manual reset mode the cutout must be reset by
the operator after the temperature falls below the set-point.
When the cutout is active and the cutout mode is set to manual (“reset”) then
the display will flash “Cut-out” until the user resets the cutout. To access the
reset cutout function press the “SET” button.
-
S
The display will indicate the reset function.
Access cutout reset function
rESEt ?Cutout reset function
35
Page 40
7005/7015 Calibration Bath
User’s Guide
Figure 5 Controller Operation Flowchart
36
Page 41
Press “SET” once more to reset the cutout.
8 Controller Operation
Temperature Set-point
S
This switchs the display to the set temperature function. To return to displaying
the temperature press the “EXIT” button. If the cutout is still in the over-tem
perature fault condition the display continues to flash “Cut-out”. The bath
temperature must drop a few degrees below the cutout set-point before the cut
out can be reset.
Reset cutout
8.3Temperature Set-point
The bath temperature can be set to any value within the range and with resolu
tion as given in the specifications. The temperature range of the particular fluid
used in the bath must be known by the operator and the bath should only be op
erated well below the upper temperature limit of the liquid. In addition, the
cutout temperature should also be set below the upper limit of the fluid.
Setting the bath temperature involves three steps: (1) select the set-point memory, (2) adjust the set-point value, and (3) adjust the vernier if desired.
8.3.1Programmable Set-points
The controller stores 8 set-point temperatures in memory. The set-points can be
quickly recalled to conveniently set the bath to a previously programmed temperature set-point.
-
-
-
-
To set the bath temperature one must first select the set-point memory. This
function is accessed from the temperature display function by pressing “SET”.
The number of the set-point memory currently being used is shown at the left
on the display followed by the current set-point value.
25.00 CBath temperature in degrees Celsius
S
Access set-point memory
1. 25.0Set-point memory 1, 25.0°C currently used
To change the set-point memory press “UP” or “DOWN”.
4. 40.0New set-point memory 4, 40.0°C
Press “SET” to accept the new selection and access the set-point value.
S
Accept selected set-point memory
37
Page 42
7005/7015 Calibration Bath
User’s Guide
8.3.2Set-point Value
The set-point value may be adjusted after selecting the set-point memory and
pressing “SET”. The set-point value is displayed with the units, C or F, at the
left.
C 40.00Set-point 4 value in °C
If the set-point value need not be changed then press “EXIT” to resume dis
playing the bath temperature. Press “UP” or “DOWN” to adjust the set-point
value.
C 42.50New set-point value
When the desired set-point value is reached press “SET” to accept the new
value and access the set-point vernier. If “EXIT” is pressed instead then any
changes made to the set-point will be ignored.
S
Accept new set-point value
8.3.3Set-point Vernier
The set-point value can only be set with a resolution of 0.002°C. The user may
want to adjust the set-point slightly to achieve a precise bath temperature. The
set-point vernier allows one to adjust the temperature below or above the
set-point by a small amount with very high resolution. Each of the 8 stored
set-points has an associated vernier setting. The vernier is accessed from the
set-point by pressing “SET”. The vernier setting is displayed as a 6 digit number with five digits after the decimal point. This is a temperature offset in degrees of the selected units, C or F.
-
38
0.00000Current vernier value in °C
To adjust the vernier press “UP” or “DOWN”. Unlike most functions the ver
nier setting has immediate effect as the vernier is adjusted. “SET” need not be
pressed. This allows one to continually adjust the bath temperature with the
vernierasitisdisplayed.
-
0.00090New vernier setting
Next press “EXIT” to return to the temperature display or “SET” to access the
temperature scale units selection.
S
Access scale units
Page 43
8.4Temperature Scale Units
The temperature scale units of the controller may be set by the user to degrees
Celsius (°C) or Fahrenheit (°F). These units are used in displaying the bath
temperature, set-point, vernier, proportional band, and cutout set-point.
The temperature scale units selection is accessed after the vernier adjustment
function by pressing “SET”. From the temperature display function access the
units selection by pressing “SET” 4 times.
25.00 CBath temperature
8 Controller Operation
Temperature Scale Units
S
1. 25.0Set-point memory
S
C 25.00Set-point value
S
0.00000Vernier setting
S
Un= CScale units currently selected
Press “UP” or “DOWN” to change the units.
Un= FNew units selected
Access set-point memory
Access set-point value
Access vernier
Access scale units selection
Press “SET” to accept the new selection and resume displaying the bath
temperature.
S
Set the new units and resume temperature display
8.5Secondary Menu
Functions which are used less often are accessed within the secondary menu.
The secondary menu is accessed by pressing “SET” and “EXIT” simulta
neously and then releasing. The first function in the secondary menu is the
heater power display.
-
39
Page 44
7005/7015 Calibration Bath
User’s Guide
8.6Heater Power
The temperature controller controls the temperature of the bath by pulsing the
heater on and off. The total power being applied to the heater is determined by
the duty cycle or the ratio of heater on time to the pulse cycle time. This value
may be estimated by watching the red/green control indicator light or read di
rectly from the digital display. By knowing the amount of heating to the bath
the user can tell if the bath is heating up to the set-point, cooling down, or con
trolling at a constant temperature. Monitoring the percent heater power will let
the user know how stable the bath temperature is. With good control stability
the percent heating power should not fluctuate more than ±1% within one
minute.
The heater power display is accessed in the secondary menu. Press “SET” and
“EXIT” simultaneously and release. The heater power will be displayed as a
percentage of full power.
-
-
S+E
Access heater power in secondary menu
12 PctHeater power in percent
To exit out of the secondary menu press “EXIT”. To continue on to the proportional band setting function press “SET”.
8.7Proportional Band
In a proportional controller such as this the heater output power is proportional
to the bath temperature over a limited range of temperatures around the
set-point. This range of temperature is called the proportional band. At the bottom of the proportional band the heater output is 100%. At the top of the proportional band the heater output is 0. Thus as the bath temperature rises the
heater power is reduced, which consequently tends to lower the temperature
40
Figure 6 Bath temperature fluctuation at various proportional band settings
Page 45
8 Controller Operation
Proportional Band
back down. In this way the temperature is maintained at a fairly constant
temperature.
The temperature stability of the bath depends on the width of the proportional
band. See Figure 6. If the band is too wide the bath temperature will deviate ex
cessively from the set-point due to varying external conditions. This is because
the power output changes very little with temperature and the controller cannot
respond very well to changing conditions or noise in the system. If the propor
tional band is too narrow the bath temperature may swing back and forth be
-
cause the controller overreacts to temperature variations. For best control
stability the proportional band must be set for the optimum width.
The optimum proportional band width depends on several factors among which
are fluid volume, fluid characteristics (viscosity, specific heat, thermal conduc
tivity), heater power setting, operating temperature, and stirring. Thus the pro
-
portional band width may require adjustment for best bath stability when any of
these conditions change. Of these, the most significant factors affecting the op
timum proportional band width are heater power setting and fluid viscosity.
The proportional band should be wider when the higher power setting is used
so that the change in output power per change in temperature remains the same.
The proportional band should also be wider when the fluid viscosity is higher
because of the increased response time.
-
-
The proportional band width is easily adjusted from the bath front panel. The
width may be set to discrete values in degrees C or F depending on the selected
units. The optimum proportional band width setting may be determined by
monitoring the stability with a high resolution thermometer or with the controller percent output power display. Narrow the proportional band width to the
point at which the bath temperature begins to oscillate and then increase the
Table 4 Proportional Band — Fluid Table
Heater SettingProportional
FluidTemperature
Water30.0°CLow0.04°C
Water60.0°CLow0.04°C
Eth-Gly 50%35.0°CLow0.05°C
Eth-Gly 50%60.0°CLow0.05°C
Eth-Gly 50%100.0°CHigh0.4°C
Oil35.0°CLow0.1°C
BandStability
±
0.001°C
±
0.002°C
±
0.001°C
±
0.002°C
±
0.007°C
±
0.003°C
Oil60.0°CLow0.2°C
Oil100°CLow0.2°C
±
0.002°C
±
0.003°C
41
Page 46
7005/7015 Calibration Bath
User’s Guide
band width from this point to 3 or 4 times wider. Table 4 lists typical propor
tional band settings for optimum performance with a variety of fluids at se
lected temperatures.
The proportional band adjustment may be accessed within the secondary menu.
Press “SET” and “EXIT” to enter the secondary menu and show the heater
power. Then press “SET” to access the proportional band.
S+E
Access heater power in secondary menu
-
-
12 PctHeater power in percent
S
Access proportional band
Pb=0.101CProportional band setting
To change the proportional band press “UP” or “DOWN”.
Pb=0.060CNew proportional band setting
To accept the new setting and access the cutout set-point press “SET”. Pressing
“EXIT” will exit the secondary menu ignoring any changes just made to the
proportional band value.
S
8.8Cutout
As a protection against software or hardware fault, shorted heater triac, or user
error, the bath is equipped with an adjustable heater cutout device that shuts off
power to the heater if the bath temperature exceeds a set value. This protects
the heater and bath materials from excessive temperatures and, most impor
tantly, protects the bath fluids from being heated beyond the safe operating
temperature preventing hazardous vaporization, breakdown, or ignition of the
liquid. The cutout temperature is programmable by the operator from the front
panel of the controller. It must always be set below the upper temperature limit
of the fluid and no more than 10 degrees above the upper temperature limit of
the bath.
If the cutout is activated because of excessive bath temperature, power to the
heater shuts off and the bath cools. The bath cools until it reaches a few degrees
below the cutout set-point temperature. At this point the action of the cutout is
determined by the setting of the cutout mode parameter. The cutout has two
selectable modes — automatic reset or manual reset. If the mode is set to auto
matic, the cutout automatically resets itself when the bath temperature falls be
low the reset temperature allowing the bath to heat up again. If the mode is set
to manual, the heater remains disabled until the user manually resets the cutout.
Accept the new proportional band setting
-
-
-
42
Page 47
8 Controller Operation
Controller Configuration
The cutout set-point may be accessed within the secondary menu. Press “SET”
and “EXIT” to enter the secondary menu and show the heater power. Then
press “SET” twice to access the cutout set-point.
S+E
Access heater power in secondary menu
12 PctHeater power in percent
S
Access proportional band
Pb=0.101CProportional band setting
S
Access cutout set-point
CO= 110Ccutout set-point
To change the cutout set-point press “UP” or “DOWN”.
CO= 75CNew cutout set-point
To accept the new cutout set-point press “SET”.
S
Accept cutout set-point
The next function is the configuration menu. Press “EXIT” to resume displaying the bath temperature.
8.9Controller Configuration
The controller has a number of configuration and operating options and calibra
tion parameters which are programmable via the front panel. These are ac
cessed from the secondary menu after the cutout set-point function by pressing
“SET.” There are 5 sets of configuration parameters — probe parameters, oper
ating parameters, serial interface parameters, IEEE-488 interface parameters,
and controller calibration parameters. The menus are selected using the “UP”
and “DOWN” keys and then pressing “SET”.
8.10Probe Parameters
The probe parameter menu is indicated by,
PrObEProbe parameters menu
Press “SET” to enter the menu. The probe parameters menu contains the pa
rameters, D0 and Dg, which characterize the resistance-temperature relation
-
-
-
-
-
43
Page 48
7005/7015 Calibration Bath
User’s Guide
ship of the thermistor control probe. These parameters may be adjusted to
improve the accuracy of the bath. This procedure is explained in detail in
Section10.
The probe parameters are accessed by pressing “SET” after the name of the pa
rameter is displayed. The value of the parameter may be changed using the
“UP” and “DOWN” buttons. After the desired value is reached press “SET” to
set the parameter to the new value. Pressing “EXIT” will cause the parameter to
be skipped ignoring any changes that may have been made.
8.10.1D0
This probe parameter refers to the resistance of the control probe at 0°C.
Normally this is set for –25.2290 ohms.
8.10.2DG
This probe parameter refers to the average sensitivity of the probe between 0
and 100°C. Normally this is set for 186.9740
8.11Operating Parameters
The operating parameters menu is indicated by,
PArOperating parameters menu
-
Press “SET” to enter the menu. The operating parameters menu contains the
cutout reset mode setting.
8.11.1Cutout Reset Mode
The cutout reset mode determines whether the cutout resets automatically when
the bath temperature drops to a safe value or must be manually reset by the
operator.
The parameter is indicated by,
CtorStCutout reset mode parameter
Press “SET” to access the parameter setting. Normally the cutout is set for
manual mode.
Cto=rStCutout set for manual reset
To change to automatic reset mode press “UP” and then “SET”.
Cto=AutoCutout set for automatic reset
44
Page 49
8.12Serial Interface Parameters
8 Controller Operation
Serial Interface Parameters
The 7005/7015 bath may optionally be fitted with an RS-232 interface. The se
rial RS-232 interface parameters menu is indicated by,
SErIALSerial RS-232 interface parameters menu
The serial interface parameters menu contains parameters which determine the
operation of the serial interface. These controls only apply to baths fitted with
the serial interface. The parameters in the menu are — BAUD rate, sample pe
riod, duplex mode, and linefeed.
8.12.1Baud Rate
The baud rate is the first parameter in the menu. The baud rate setting deter
mines the serial communications transmission rate.
The baud rate parameter is indicated by,
BAUdSerial baud rate parameter
Press “SET” to choose to set the baud rate. The current baud rate value will
then be displayed.
-
-
-
1200 bCurrent baud rate
The baud rate of the bath serial communications may be programmed to
300,600,1200, or 2400 baud. Use “UP” or “DOWN” to change the baud rate
value.
2400 bNew baud rate
Press “SET” to set the baud rate to the new value or “EXIT” to abort the opera
tion and skip to the next parameter in the menu.
8.12.2Sample Period
The sample period is the next parameter in the serial interface parameter menu.
The sample period is the time period in seconds between temperature measure
ments transmitted from the serial interface. If the sample rate is set to 5, the
bath transmits the current measurement over the serial interface approximately
every five seconds. The automatic sampling is disabled with a sample period of
0. The sample period is indicated by,
SAmPLESerial sample period parameter
-
-
Press “SET” to choose to set the sample period. The current sample period
value will be displayed.
45
Page 50
7005/7015 Calibration Bath
User’s Guide
SA= 1Current sample period (seconds)
Adjust the value with “UP” or “DOWN” and then use “SET” to set the sample
rate to the displayed value.
SA= 60New sample period
8.12.3Duplex Mode
The next parameter is the duplex mode. The duplex mode may be set to full du
plex or half duplex. With full duplex any commands received by the bath via
the serial interface are immediately echoed or transmitted back to the device of
origin. With half duplex the commands are executed but not echoed. The du
plex mode parameter is indicated by,
dUPLSerial duplex mode parameter
Press “SET” to access the mode setting.
dUP=FULLCurrent duplex mode setting
The mode may be changed using “UP” or “DOWN” and pressing “SET”.
dUP=HALFNew duplex mode setting
8.12.4Linefeed
The final parameter in the serial interface menu is the linefeed mode. This pa
rameter enables (on) or disables (off) transmission of a linefeed character (LF,
ASCII 10) after transmission of any carriage-return. The linefeed parameter is
indicated by,
-
-
-
46
LFSerial linefeed parameter
Press “SET” to access the linefeed parameter.
LF= OnCurrent linefeed setting
The mode may be changed using “UP” or “DOWN” and pressing “SET”.
LF= OFFNew linefeed setting
Page 51
8.13IEEE-488 Parameters
Baths may optionally be fitted with an IEEE-488 GPIB interface. In this case
the user may set the interface address within the IEEE-488 parameter menu.
This menu does not appear on baths not fitted with the interface. The menu is
indicated by,
IEEEIEEE-488 parameters menu
Press “SET” to enter the menu.
8.13.1IEEE-488 Address
8 Controller Operation
IEEE-488 Parameters
The IEEE-488 interface must be configured to use the same address as the ex
ternal communicating device. The address is indicated by,
AddrESSIEEE-488 interface address
Press “SET” to access the address setting.
Add= 22Current IEEE-488 interface address
Adjust the value with “UP” or “DOWN” and then use “SET” to set the address
to the displayed value.
Add= 15New IEEE-488 interface address
8.14Calibration Parameters
The operator of the bath controller has access to a number of the bath calibration constants namely CTO, H, and L. These values are set at the factory and
must not be altered. The correct values are important to the accuracy and
proper and safe operation of the bath. Access to these parameters is available to
the user only so that in the event that the controller’s memory fails the user may
restore these values to the factory settings. The user should have a list of these
constants and their settings with the manual.
-
DO NOT change the values of the bath calibration constants from the factory set
values. The correct setting of these parameters is important to the safety
and proper operation of the bath.
The calibration parameters menu is indicated by,
CALCalibration parameters menu
Press “SET” five times to enter the menu.
47
Page 52
7005/7015 Calibration Bath
User’s Guide
8.14.1CTO
Parameter CTO sets the calibration of the over-temperature cutout. This is not
adjustable by software but is adjusted with an internal potentiometer. For the
7005/7015 baths this parameter should read between 120.
8.14.2H and L
These parameters set the upper and lower set-point limits of the bath. DO NOT
change the values of these parameters from the factory set values. To do so may
present danger of the bath exceeding its temperature range causing damage or
fire.
48
Page 53
9 Digital Communication Interface
9Digital Communication Interface
If supplied with the option, the 7005/7015 bath is capable of communicating
with and being controlled by other equipment through the digital interface. Two
types of digital interface are available — the RS-232 serial interface and the
IEEE-488 GPIB interface.
With a digital interface the bath may be connected to a computer or other
equipment. This allows the user to set the bath temperature, monitor the tem
perature, and access any of the other controller functions, all using remote com
munications equipment. In addition the heater power setting and cooling
capacity may be controlled using the interface. To enable the heater to be
switched to high using the interface the “HEATER” switch must be set to the
“LOW” position. The cooling power switch must be set to OFF to enable re
mote control.
9.1Serial Communications
Serial Communications
-
-
-
The bath may be installed with an RS-232 serial interface that allows serial digital communications over fairly long distances. With the serial interface the user
may access any of the functions, parameters and settings discussed in Section 8
with the exception of the BAUD
rate setting. The serial interface
uses 8 data bits, 1 stop bit, and no
parity.
9.1.1Wiring
The serial communications cable
attaches to the bath through the
DB-9 connector at the back of the
instrument. Figure 7 shows the
pin-out of this connector and sug
gested cable wiring. To eliminate
noise, the serial cable should be
shielded with low resistance be
tween the connector (DB-9) and
the shield.
9.1.2Setup
-
-
Before operation the serial inter
face of the bath must first be set
up by programming the BAUD
rate and other configuration pa
rameters. These parameters are
-
-
Figure 7 Serial Communications Cable Wiring
49
Page 54
7005/7015 Calibration Bath
User’s Guide
programmed within the serial interface menu.
To enter the serial parameter programming mode first press “EXIT” while
pressing “SET” and release to enter the secondary menu. Press “SET” repeat
edly until the display reads “ProbE”. This is the menu selection. Press “UP”
repeatedly until the serial interface menu is indicated with “SErIAL”. Finally
press “SET” to enter the serial parameter menu. In the serial interface parame
ters menu are the BAUD rate, the sample rate, the duplex mode, and the line
feed parameter.
9.1.2.1Baud rate
The baud rate is the first parameter in the menu. The display will prompt with
the baud rate parameter by showing “BAUd”. Press “SET” to choose to set the
baud rate. The current baud rate value will then be displayed. The baud rate of
the 7005/7015 serial communications may be programmed to 300,600,1200, or
2400 baud. The baud rate is pre-programmed to 1200 baud. Use “UP” or
“DOWN” to change the baud rate value. Press “SET” to set the baud rate to the
new value or “EXIT” to abort the operation and skip to the next parameter in
the menu.
-
-
-
9.1.2.2Sample Period
The sample period is the next parameter in the menu and prompted with “SAmPLE”. The sample period is the time period in seconds between temperature
measurements transmitted from the serial interface. If the sample rate is set to
5, the 7005/7015 transmits the current measurement over the serial interface approximately every five seconds. The automatic sampling is disabled with a
sample period of 0. Press “SET” to choose to set the sample period. Adjust the
period with “UP” or “DOWN” and then use “SET” to set the sample rate to the
displayed value.
9.1.2.3Duplex Mode
The next parameter is the duplex mode indicated with “dUPL”. The duplex
mode may be set to half duplex (“HALF”) or full duplex (“FULL”). With full
duplex any commands received by the thermometer via the serial interface are
immediately echoed or transmitted back to the device of origin. With half du
plex the commands are executed but not echoed. The default setting is full du
plex. The mode may be changed using “UP” or “DOWN” and pressing “SET”.
9.1.2.4Linefeed
The final parameter in the serial interface menu is the linefeed mode. This pa
rameter enables (“On”) or disables (“OFF”) transmission of a linefeed charac
ter (LF, ASCII 10) after transmission of any carriage-return. The default setting
is with linefeed on. The mode may be changed using “UP” or “DOWN” and
pressing “SET”.
-
-
-
-
50
Page 55
9.1.3Serial Operation
9 Digital Communication Interface
IEEE-488 Communication (optional)
Once the cable has been attached and the interface set up properly the control
ler immediately begins transmitting temperature readings at the programmed
rate. The set-point and other commands may be sent to the bath via the serial
interface to set the bath and view or program the various parameters. The inter
face commands are discussed in Section 9.3. All commands are ASCII charac
ter strings terminated with a carriage-return character (CR, ASCII 13).
9.2IEEE-488 Communication (optional)
The IEEE-488 interface is available as an option. Baths supplied with this op
tion may be connected to a GPIB type communication bus which allows many
instruments to be connected and controlled simultaneously. To eliminate noise,
the GPIB cable should be shielded.
9.2.1Setup
To use the IEEE-488 interface first connect an IEEE-488 standard cable to the
back of the bath. Next set the device address. This parameter is programmed
within the IEEE-488 interface menu.
To enter the IEEE-488 parameter programming menu first press “EXIT” while
pressing “SET” and release to enter the secondary menu. Press “SET” repeatedly until the display reaches “PrObE”. This is the menu selection. Press “UP”
repeatedly until the IEEE-488 interface menu is indicated with “IEEE”.Press
“SET” to enter the IEEE-488 parameter menu. The IEEE-488 menu contains
the IEEE-488 address parameter.
-
-
-
-
9.2.1.1IEEE-488 Interface Address
The IEEE-488 address is prompted with “AddrESS”. Press “SET” to program
the address. The default address is 22. Change the device address of the bath if
necessary to match the address used by the communication equipment by
pressing “UP” or “DOWN” and then “SET”.
9.2.2IEEE-488 Operation
Commands may now be sent via the IEEE-488 interface to read or set the tem
perature or access other controller functions. All commands are ASCII charac
ter strings and are terminated with a carriage-return (CR, ASCII 13). Interface
commands are listed below.
9.3Interface Commands
The various commands for accessing the bath controller functions via the digi
tal interfaces are listed in this section (see Table 5). These commands are used
with both the RS-232 serial interface and the IEEE-488 GPIB interface. In ei
ther case the commands are terminated with a carriage-return character. The in
-
-
-
-
-
51
Page 56
7005/7015 Calibration Bath
User’s Guide
Table 5 Interface Command Summary.
Command
Command Description
Display Temperature
Read current set-points[etpoint]sset: 9999.99 {C or F}set: 100.00 C
Set current set-point to ns[etpoint]=ns=100Instrument
Read vernierv[ernier]vv: 9.99999v: 0.00000
Set vernier to nv[ernier]=nv=.00001Depends on
Read temperaturet[emperature]tt: 9999.99 {C or F}t: 55.69 C
Read temperature unitsu[nits]uu: xu: c
Set temperature units:u[nits]=c/f
Set temperature units to Celsiusu[nits]=cu=c
Set temperature units to
Fahrenheit
Secondary Menu
Read proportional band settingpr[op-band]prpr: 999.9pr: 15.9
Set proportional band to
n
Format
u[nits]=fu=f
pr[op-band]=npr=8.83Depends on
Command
ExampleReturned
Returned
Example
Acceptable
Values
Range
Configuration
CorF
Configuration
Read cut-out settingc[utout]cc: 9999 {x},{xxx}c: 110 C, in
Read high set-point limit value*th[igh]*thth: 999th: 205
Set high set-point limit to
n
n
Format
cm[ode]=a[uto]cm=a
sa[mple]=nsa=00 to 4000
n
*tl[ow]=n*tl=-80–999.9 to 999.9
*th[igh]=n*th=205–999.9 to 999.9
Command
ExampleReturned
Returned
Example
Acceptable
Values
FULL or HALF
ON or OFF
Miscellaneous (not on menus)
Read firmware version number*ver[sion]*verver.9999,9.99ver.2100,3.56
Read structure of all commandsh[elp]hlist of commands
Read Heaterf1f1f1:9f1:1
Set Heaterf1=1/00or1
Set heater to lowf1=
Set heater to highf1=
Read Refrigerationf2f2f2:9f2:0
Set Refrigerationf2=1/00or1
Set Refrigeration to onf2=
Set Refrigeration to offf2=
Read Expansion Valve 1f3f3f3:9f3:1
Set Expansion Valve 1
Set Expansion Valve 1 to onf3=
Set Expansion Valve 1 to offf3=
Read Cooling Powerf4f4f4:9f4:1
n
n
n
n
f3=1/0
n
n
f1=0
f1=1
f2=1
f2=0
0or1
f3=1
f3=0
53
Page 58
7005/7015 Calibration Bath
User’s Guide
Digital Communications Commands continued
Command
Command Description
Set Back Pressure
Set back pressure onf4=
Set back pressure offf4=
Legend:[] Optional Command data
Note:When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed by a
Format
f4=1/0
n
n
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
carriage return and linefeed. Then the value is returned as indicated in the RETURNED column.
Command
ExampleReturned
f4=1
f4=0
Returned
Example
Acceptable
Values
0or1
terface makes no distinction between upper and lower case letters, hence either
may be used. Commands may be abbreviated to the minimum number of letters
which determines a unique command. A command may be used to either set a
parameter or display a parameter depending on whether or not a value is sent
with the command following a “=” character. For example “s”<CR> will return
the current set-point and “s=50.00”<CR> will set the set-point to 50.00
degrees.
In the following list of commands, characters or data within brackets, “[” and
“]”, are optional for the command. A slash, “/”, denotes alternate characters or
data. Numeric data, denoted by “n”, may be entered in decimal or exponential
notation. Characters are shown in lower case although upper case may be used.
Spaces may be added within command strings and will simply be ignored.
Backspace (BS, ASCII 8) may be used to erase the previous character. A termi
nating CR is implied with all commands.
9.4Power Control Functions
The digital interface is capable of controlling the heating and cooling functions
so that the bath can be remotely operated at any temperature within the range of
the bath. To allow the interface to control the heating and the cooling, the front
panel controls are disabled by 1) switching the heater switch to LOW, and 2)
switching the refrigeration switch to OFF. Otherwise, the interface would not
be able to switch these functions off. The 7005/7015 bath has four control
functions with the digital interface. These are heater power high/low, cooling
on/off, expansion valve on/off, and back pressure valve on/off.
-
54
Page 59
9.4.1Heater Control
To control the heater with the digital interface the front panel heater switch
must be set to LOW (500 W). The heater function is controlled with the “F1"
command. Setting the “F1" parameter to 0 sets the heater to LOW (500 W) and
setting it to 1 sets the heater to HIGH (1000 W). Sending “F1" with no value
causes the controller to return a value showing what the heater setting is. When
the heater setting is changed a pop is heard as the heater relay opens or closes.
9.4.2Cooling Control
9 Digital Communication Interface
Power Control Functions
To control the refrigeration power with the digital interface the front panel re
frigeration switch must be off. The refrigeration power function is controlled
with the “F2" command. Setting the “F2" value to 0 turns the refrigeration off
and setting it to 1 turns it on. “F2" alone returns 0 or 1 showing the state of the
refrigeration power control.
The “F3" command controls the “high” and “low” cooling temperature range or
expansion valves. As discussed in Section 6.2 this valve adjusts the cooling
temperature which sets the cooling capacity. Setting ”F3" to 1 opens the valve
(on or high) and 0 closes the valve (off or low). A command with no value returns the current value. With the valve closed the cooling temperature is set
with the cooling temperature knob labeled “LOW”. This should be adjusted for
approximately 9–14 psi if the bath is to be used as low as –10°C. The knob labeled “HIGH” controls the cooling temperature when the expansion valve is
open or on. This should be adjusted for approximately 45–50 psi. These values
are typical but may be adjusted as required for specific applications. The low
range must be set to a lower pressure (temperature) than the high range valve.
The “F4" command controls the back pressure valve. When the valve is open
the cooling capacity is significantly reduced. This valve should normally be left
off/closed throughout the operating range of the bath. Table 6 summarizes the
control functions for heating and cooling. Table 7 shows the recommended
control settings for each operating temperature range. The ranges may need to
be adjusted depending on the bath application.
Table 6 Serial Power Control Functions
FunctionCommand01
HeaterF1lowhigh
RefrigerationF2offon
Expansion valveF3off (low)on (high)
Back pressureF4offon
55
Page 60
7005/7015 Calibration Bath
User’s Guide
Table 7 Temperature Range Control Functions
RangeHeaterRefr.Exp.Back.
–10 to 15°Clowonoffoff
15 to 40°Clowononoff
40 to 110°Chighoffoffoff
56
Page 61
10Calibration Procedure
10 Calibration Procedure
Calibration Points
In some instances the user may want to calibrate the bath to improve the tem
perature set-point accuracy. Calibration is done by adjusting the controller
probe calibration constants DO and DG so that the temperature of the bath as
measured with a standard thermometer agrees more closely with the bath
set-point. The thermometer used must be able to measure the bath fluid temper
ature with higher accuracy than the desired accuracy of the bath. By using a
good thermometer and carefully following the procedure the bath can be cali
brated to an accuracy of better than 0.1°C over a range of 50 degrees.
10.1Calibration Points
In calibrating the bath DO and DG are adjusted to minimize the set-point error
at each of two different bath temperatures. Any two reasonably separated bath
temperatures may be used for the calibration however best results are obtained
when using bath temperatures which are just within the most useful operating
range of the bath. The farther apart the calibration temperatures the larger will
be the calibrated temperature range but the calibration error will also be greater
over the range. If for instance 20°C and 80°C are chosen as the calibration tem-
peratures, the bath may achieve an accuracy of say ±0.2°C over the range 20 to
80°C. Choosing 30°C and 70°C may allow the bath to have a better accuracy of
maybe ±0.05°C over the range 30 to 70°C but outside that range the accuracy
may be only ±0.5°C.
-
-
-
10.2Measuring the Set-point Error
The first step in the calibration procedure is to measure the temperature errors
(including sign) at the two calibration temperatures. First set the bath to the
lower set-point which we will call t
. Wait for the bath to reach the set-point
L
and allow 15 minutes to stabilize at that temperature. Check the bath stability
with the thermometer. When both the bath and the thermometer have stabilized
measure the bath temperature with the thermometer and compute the tempera
ture error err
which is the actual bath temperature minus the set-point temper
L
ature. If for example the bath is set for a lower set-point of t
bath reaches a measured temperature of 19.7°C then the error is -0.3°C.
Next, set the bath for the upper set-point t
bath temperature and compute the error err
and after stabilizing measure the
H
. For our example we will suppose
H
the bath was set for 80°C and the thermometer measured 80.1°C giving an error
of +0.1°C.
10.3Computing D0 and DG
Before computing the new values for D0 and DG the current values must be
known. The values may be found by either accessing the probe calibration
=20°C and the
L
-
-
57
Page 62
7005/7015 Calibration Bath
User’s Guide
menu from the controller panel or by inquiring through the digital interface.
The user should keep a record of these values in case they may need to be re
stored in the future. The new values DO′ and DG′ are computed by entering the
old values for DO and DG, the calibration temperature set-points t
the temperature errors errL and errH into the following equations,
and tH,and
L
-
errtDerrtD
()()
D
DG
0
LHHL
⎡
errerr
′=
HL
⎢
tt
⎣
HL
00
−−−
tt
−
HL
−
−
⎤
DG
+
1
⎥
⎦
If for example D0 and DG were previously set for –25.229 and 186.9740 respectively and the data for t
L,tH
new values D0′ and DG′ would be computed as –25.831 and 188.220 respectively. Program the new values D0 and DG into the controller. The new constants will be used the next time the bath temperature is set. Check the
calibration by setting the temperature to t
again. If desired the calibration procedure may be repeated again to further improve the accuracy.
10.4Calibration Example
The bath is to be used between 25 and 75°C and it is desired to calibrate the
bath as accurately as possible for operation within this range. The current values for D0 and DG are –25.229 and 186.974 respectively. The calibration
points are chosen to be 25.00 and 75.00°C. The measured bath temperatures are
24.869 and 74.901°C respectively. Refer to Figure 8 for applying equations to
the example data and computing the new probe constants.
D
0′=
+
,errL,anderrHwere as given above then the
and tHand measuring the errors
L
58
Page 63
D0 = -25.229
DG = 186.974
t
= 25.00°C
L
measured t = 24.869°C
t
= 75.00°C
H
measured t = 74.901°C
Compute errors,
err
= 24.869 - 25.00°C = -0.131°C
L
err
= 74.901 - 75.00°C = -0.099°C
H
10 Calibration Procedure
Calibration Example
Compute D0,
0131 75 025 2290 099 25 025 22
D0
−−−−− −−(.)(.( .))(.)(.( .
′=
75 025 0
..
Compute DG,
−−−
(. )(.)
0 0990131
DG′=
Figure 8 Sample Calibration Computations
⎡
⎢
⎣
−
..
75 025 0
⎤
+
1 186 974187 094
⎥
⎦
−
=
..
9
))
25 22925392
(.).
+−=−
59
Page 64
11Maintenance
11 Maintenance
The calibration instrument has been designed with the utmost care. Ease of op
eration and simplicity of maintenance have been a central theme in the product
development. Therefore, with proper care the instrument should require very
little maintenance. Avoid operating the instrument in dirty or dusty
environments.
A battery is used to maintain operating parameters in the unit. All operat
•
ing parameters, including calibration parameters should be checked on a
regular basis to insure accuracy and proper operation of the instrument.
See the troubleshooting section for the procedure on checking the status
of the battery.
The bath should be cleaned regularly to prevent a buildup of oil or dust.
•
Use a paint safe cleaning agent on all painted surfaces. Solvents such as
Trichloroethylene or Acetone may dull or dissolve the paint. The stainless
steel surfaces may be cleaned with solvents as necessary to remove oils.
• The stirring motor should be clean to allow proper cooling. Normally
only the outside surfaces require any attention. If the inside of the motor
has become heavily loaded with oily dust, blow it out with compressed
air. Follow normal safety procedures when using pressurized gasses.
• Periodically check the fluid level in the bath to ensure that the level has
not dropped. A drop in the fluid level affects the stability of the bath.
Changes in fluid level are dependent upon several factors specific to the
environment in which the equipment is used. A schedule cannot be outlined to meet each environmental setting. Therefore, the first year the bath
should be checked weekly with notes kept as to changes in bath fluid. After the first year, the user can set up a maintenance schedule based on the
data specific to the application.
-
-
• Heat transfer medium lifetime is dependent upon the type of medium and
the environment. The fluid should be checked at least every month for the
first year and regularly thereafter. This fluid check provides a baseline for
knowledge of bath operation with clean, usable fluid. Once some fluids
have become compromised, the break down can occur rapidly. Particular
attention should be paid to the viscosity of the fluid. A significant change
in the viscosity can indicate that the fluid is contaminated, being used out
side of its temperature limits, contains ice particles, or is close to a chemi
cal breakdown. Once data has been gathered, a specific maintenance
schedule can be outline for the instrument. Refer to the General Operation
section (Section 7) for more information about the different types of flu
ids used in calibration baths.
•
If a hazardous material is spilt on or inside the equipment, the user is re
sponsible for taking the appropriate decontamination steps as outlined by
the national safety council with respect to the material. MSDS sheets ap
plicable to all fluids used in the baths should be kept in close proximity to
the instrument.
-
-
-
-
-
61
Page 65
7005/7015 Calibration Bath
User’s Guide
If the mains supply cord becomes damaged, replace it with a cord with
•
the appropriate gauge wire for the current of the bath. If there are any
questions, contact an Authorized Service Center for more information.
Before using any cleaning or decontamination method except those rec
•
ommended by Hart, users should check with an Authorized Service Cen
ter to be sure that the proposed method will not damage the equipment.
If the instrument is used in a manner not in accordance with the equip
•
ment design, the operation of the bath may be impaired or safety hazards
may arise.
WARNING: When checking the over-temperature cutout, be sure that the
temperature limits of the bath fluid are not exceeded. Exceeding the tem
perature limits of the bath fluid could cause harm to the operator, lab, and
instrument.
• The over-temperature cutout should be checked every 6 months to see that
it is working properly. In order to check the user selected cutout, follow
the controller directions (Section 8.8) for setting the cutout. Both the manual and the auto reset option of the cutout should be checked. Set the bath
temperature higher than the cutout. Check to see if the display flashes cutout and the temperature is decreasing.
-
-
-
-
• The constant temperature bath depends upon the certain qualities of the
fluid medium in order to maintain a uniform and stable temperature environment. Some oils change their characteristics or become dirty after a
period of use. Always remove any foreign materials from the bath.
• Silicone oils as well as others may evaporate off their lighter components
over a period of time leaving the very viscous components remaining. In
addition, some decomposition occurs which may impair the temperature
stability of the bath. When this happens, the fluid is generally very dark to
black and viscous. Vegetable oils polymerize (turn plastic like) after they
have been used for a time at high temperatures making them very difficult
to remove. When the oil has become unusable it should be changed. The
instructions for removing the oil is as follows.
11.1Draining the Bath
The drain is located on the back of the bath. See Figure 3 on page 23. Locate
the drain plug on the end of the drain tube. This drain plug is to be fluid tight
until the time of draining. The following information is helpful when draining
the bath. Always use a container capable of holding the entire load of fluid. Us
ing an adequate size (about 8 gallons or 1 cubic foot), heat proof fluid container
is extremely important. Use safety equipment as appropriate.
-
62
1.Water and low viscosity fluids - Drain at room temperature. Normal care
must be taken for fluids that may have corrosive or damaging effects on
the surrounding facility or equipment.
Page 66
11 Maintenance
Draining the Bath
2.High viscosity oils - The fluid should be sufficiently low in viscosity to
drain efficiently. Some oils such as 710 silicone oil may need to be
heated to 80°C to be fluid enough to drain well. The viscosity affects
how rapidly it drains as well as how well it flows off of the walls. Heat
proof containers and appropriate safety equipment such as face shields,
gloves and body covering are recommended.
WARNING: Extreme danger of BURNS and FIRE. Use safety equipment,
use proper equipment and have fire safety equipment standing by.
63
Page 67
12Troubleshooting
This section contains information on troubleshooting and CE Comments. This
information pertains to a number of bath models and certain specifics may not
pertain to your model.
12.1Troubleshooting
In the event that the instrument appears to function abnormally, this section
may help to find and solve the problem. Several possible problem conditions
are described along with likely causes and solutions. If a problem arises, please
read this section carefully and attempt to understand and solve the problem. If
the probe seems faulty or the problem cannot otherwise be solved, contact an
Authorized Service Center (see Section 1.4, on page 5) for assistance. Be sure
to have the instrument model number, serial number, voltage, and problem description available.
12 Troubleshooting
Troubleshooting
ProblemCauses and Solutions
The heater indicator LED stays
red but the temperature does not
increase
The display does not show “
temperature, and the controller otherwise appears to operate normally. The problem may be insufficient heating, no heating at all, or
too much cooling.
• The heater power setting being too low, especially at
higher operating
• One or more burned out heaters or blown heater fuses
may also cause this problem. If the heaters seem to be
burned out, contact an Authorized Service Center (see
Section 1.4, on page 5) for assistance.
The controller display flashes
“
Cut-out
not operate
” and the heater does
The display flashs “
temperature.
•
If the process temperature displayed seems grossly in
error, consult the following problem:
“
Cut-out
•
Normally, the cutout disconnects power to the heater
when the bath temperature exceeds the cutout set-point
causing the temperature to drop back down to a safe
value. If the cutout mode is set to “AUTO”, the heater
switches back on when the temperature drops. If the
mode is set to “RESET”, the heater only comes on again
when the temperature is reduced and the cutout is man
ually reset by the operator, see Section 8.8, Cutout.
Check that the cutout set-point is adjusted to 10 or 20°C
above the maximum bath operating temperature and
that the cutout mode is set as desired.
•
If the cutout activates when the bath temperature is well
below the cutout set-point or the cutout does not reset
when the bath temperature drops and it is manually re
set, then the cutout circuitry or the cutout thermocouple
sensor may be faulty or disconnected. Contact an Au
thorized Service Center (see Section 1.4, on page 5) for
assistance.
Cut-out
” and an incorrect process temperature’.
Cut-out
” alternately with the process
” nor displays an incorrect bath
‘The display flashes
-
-
-
65
Page 68
7005/7015 Calibration Bath
User’s Guide
ProblemCauses and Solutions
The display flashes “cutout” and
an incorrect process temperature
The displayed process temperature is in error and the controller
remains in the cooling or the
heating state at any set-point
value
The problem may be that the controller’s voltmeter circuit is not func
tioning properly.
A problem could exist with the memory back-up battery.
•
If the battery voltage is insufficient to maintain the mem
ory, data may become scrambled causing problems. A
nearby large static discharge may also affect data in
memory. Verify that the parameters on the Report of
Test are accurate. Cycle the power off, disconnect the
bath from AC, and then restart the bath.
The controller may need to be reset. Perform the follow
•
ing Factory Reset Sequence.
Factory Reset Sequence. Hold the SET and EXIT but
tons down at the same time while powering up the
instrument. The instrument display shows '-init-', the
model number, and the firmware version. Each of the
controller parameters and calibration constants must be
reprogrammed. The values can be found on the Report
of Test that was shipped with the instrument.
If initializing the memory does not remedy the problem,
•
there may be a failed electronic component. Contact an
Authorized Service Center (see Section 1.4, on page 5)
for assistance.
Possible causes may be either a faulty control probe or erroneous
data in memory.
• The probe may be disconnected, burned out, or shorted.
• Check that the probe is connected properly. The probe
may be checked with an ohmmeter to see if it is open or
shorted. The probe is a thermistor. The resistance
should read approximately 3 kohms between pins 3 and
4, 12 kohms between pins 1 and 4, and 9-11 kohms between pins 1 and 3 at room temperature (25°C). If the
probe appears to be defective, contact an Authorized
Service Center (see Section 1.4, on page 5) for assis
tance.
•
If the problem is not the probe, erroneous data in mem
ory may be the cause. Re-initialize the memory as dis
cussed in the problem
an incorrect process temperature’.
mains, the cause may be a defective electronic compo
nent, contact an Authorized Service Center (see Section
1.4, on page 5) for assistance.
‘The display flashes “cutout” and
If the problem re
-
-
-
-
-
-
-
-
-
66
Page 69
ProblemCauses and Solutions
12 Troubleshooting
Troubleshooting
The controller controls or at
tempts to control at an inaccurate
temperature
The controller shows that the output power is steady but the process temperature is unstable
-
The controller operates normally except when controlling at a speci
fied set-point. At this set-point, the temperature displayed does not
agree with the temperature measured by the user’s reference ther
mometer to within the specified accuracy. This problem may be
caused by an actual difference in temperature between the points
where the control probe and thermometer probe measure tempera
ture, by erroneous bath calibration parameters, or by a damaged con
trol probe.
Check that the bath has an adequate amount of fluid in
•
the tank and that the stirrer is operating properly.
Check that the thermometer probe and control probe are
•
both fully inserted into the bath to minimize temperature
gradient errors.
Check that the calibration parameters are all correct ac
•
cording to the Report of Test. If not, re-program the con
stants. The memory backup battery may be weak
causing errors in data as described in the problem:
display flashes “cutout” and an incorrect process tem
-
-
-
‘The
-
perature’.
Check that the control probe has not been struck, bent,
•
or damaged. If the cause of the problem remains unknown, contact an Authorized Service Center (see Section 1.4, on page 5) for assistance.
Possible causes are an improper proportional band setting or the fluid
being used.
• If the bath temperature does not achieve the expected
degree of stability when measured using a thermometer,
try adjusting the proportional band to a narrower width
as discussed in Section 8.7, Proportional Band.
•
Check to ensure the fluid has not deteriorated or is not
too thick.
-
-
-
The controller alternately heats
for a while then cools
The controller erratically heats
then cools, control is unstable
The bath is not stable and the duty cycle is not constant.
•
The proportional band being too narrow typically causes
this oscillation. Increase the width of the proportional
band until the temperature stabilizes as discussed in
Section 8.7, Proportional Band.
If both the bath temperature and output power do not vary periodically
but in a very erratic manner, the problem may be excess noise in the
system. Noise due to the control sensor should be less than 0.001°C.
However, if the probe has been damaged or has developed an inter
mittent short, erratic behavior may exist. The probe is located inside
the stirrer motor cover.
•
Check for a damaged probe or poor connection between
the probe and bath.
•
Intermittent shorts in the heater or controller electronic
circuitry may also be a possible cause. Contact an Au
thorized Service Center (see Section 1.4, on page 5) for
assistance.
-
-
67
Page 70
7005/7015 Calibration Bath
User’s Guide
ProblemCauses and Solutions
The bath does not achieve low
temperatures
The controller does not maintain
controller parameters or parame
ters are reset each time the
power to the unit is removed
12.2Comments
Too much heating or not enough cooling can cause this problem.
Check that the control indicator glows green showing
•
that the controller is attempting to cool. The heaters may
be disabled as a test by temporarily removing the heater
fuses.
Note: Before performing the memory check,you need to record the
-
controller calibration parameters (found in the CAL menu of the instru
ment) and any user-adjusted parameters that you have changed
(such as the programmable set points and proportional band).
Memory Check
Doing a memory check is the easiest way to verify the ability of the
battery to maintain controller parameters.
1. Power off the instrument.
2. Disconnect the instrument from AC power for 10 seconds.
3. Reconnect the AC power and power on the instrument.
4. If the display shows InIT and/or the cycle count shows a low num
ber such as 0002, the battery is spent and should be replaced. Con
tact an Authorized Service Center for assistance.
5. After replacing the battery, you must reprogram the calibration and
user-adjustable parameters into the controller.
-
-
-
12.2.1EMC Directive
Hart Scientifics’ equipment has been tested to meet the European Electromagnetic Compatibility Directive (EMC Directive, 89/336/EEC). The Declaration
of Conformity for your instrument lists the specific standards to which the unit
was tested.
12.2.2Low Voltage Directive (Safety)
In order to comply with the European Low Voltage Directive (73/23/EEC),
Hart Scientific equipment has been designed to meet the IEC 1010-1 (EN
61010-1) and the IEC 1010-2-010 (EN 61010-2-010) standards.
68
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.