700PMP
Precision Pressure Pump
Instruction Sheet
®
Introduction
The Fluke 700PMP is a tube type, cylindrical
shaped hand pump with a "T" handle at the
compressor end and a round knob at the volume
adjust end. The overall length is between 11.5"
(full extension) and 8.5" (fully collapsed). The
pump incorporates a needle valve for venting
and a volume adjust vernier for precision
adjustment of pressure.
Pressure connections are made through one 1/8"
NPT internally threaded fitting. The pump is
small in size, lightweight and ruggedly
constructed to withstand typical field use.
Operating Instructions
Producing Pressure
1. Connect the pump's port to the instrument to
be calibrated or checked. Use small-diameter
tubing as short in length as possible (this will
maximize the pressure adjustment range).
2. Set the FINE ADJUST knob to the full
counterclockwise position.
3. Turn the bleed valve knob fully
counterclockwise to relieve all system
pressure and zero any measuring devices.
4. Turn the bleed valve knob fully clockwise to
close.
5. Repeatedly move the "T" handle in and out to
generate the desired pressure.
6. Use the FINE ADJUST knob to bring up the
pressure to the precise level.
7. Use the BLEED VALVE to lower the pressure
from the pressure generated. Opening the
BLEED VALVE 1/4 turn will lower the
pressure very gradually. Opening it 1/2 turn
will release the pressure faster and opening it
3/4 turn will quickly and safely release all the
pressure in the system.
9Warning
It is imperative that all system
pressure is relieved prior to making
any connections or disconnections.
Failure to relieve system pressure
could result in serious personal
injury or equipment damage. Even
nominal pressure values can
generate extreme force if fitting or
tubing failure occurs due to improper
installation or usage. Since the pump
is capable of generating pressures
exceeding 100 psig/70 bar, it is
important that all pressure
connections and test procedures be
done by qualified service personnel,
according to standard engineering
practices, to prevent possible
personal injury or equipment
damage.
Connections
To install a pressure fitting in the pump:
1. Turn the BLEED VALVE counterclockwise to
bleed any pressure.
2. Use a 5/8" open-end wrench on the input port
to prevent it from rotating while tightening the
supply fitting with a 5/8" open-end wrench.
Leak Prevention and Detection
In order to obtain maximum pressure indication
stability, leaks must be avoided. It is strongly
recommended that either Teflon® tape or
commercial pipe sealant is used at all tapered
fittings and connections. If Teflon® tape is used,
care must be taken that the proper amount is
applied. Excessive tape may fray and cause
plugging of relief valves orifices, nozzles, etc.
Overuse of pipe sealant may cause similar
problems.
PN 2094740 May 2003 ©2003 Fluke Corporation. All rights reserved. Printed in U.S.A.
700PMP
Instruction Sheet
External equipment should also be checked
carefully for leaks. Process connections, flange
bolts, and vents must be tightly closed. Defective
gaskets, leaking valves, and damaged
diaphragms are all potential sources of leaks.
For detection of very small system leaks, the
traditional soap bubble method may not be
sufficient. Halogen leak detection devices may
be required with using highly sensitive pressure
calibration equipment.
Temperature Considerations
Since the pressure change of a contained
volume of gas is directly proportional to absolute
temperature, temperature control is critical when
using the pump with any high-resolution
measuring device. Tubing should be kept away
from heat sources (i.e., lamps, operating
electronic equipment, excessive hand contact,
etc.) as well as from heat-dissipating structures
(i.e., open windows, air conditioning vents, etc.)
to minimize temperature variations that might
induce errors, Air is compressed by the pump.
This compression causes some heating of the air
as it is forced into the system. Consequently, a
noticeable decrease in pressure - caused by the
cooling of the newly compressed air - may occur
immediately after cessation of pumping.
Specifications
Pressure range: 145 psig/10 bar
Pressure connections: Single 1/8" NPT female
fitting
Size:
Body diameter: 1.5" (3.8 cm)
Length: 8.5" (21.6 cm) collapsed
Contacting Fluke
To contact Fluke, call one of the following
numbers:
USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
Or, visit Fluke's Web site at www.fluke.com
To register your product, visit register.fluke.com
LIMITED WARRANTY AND LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and
workmanship for one year from the date of purchase. This
warranty does not cover fuses, disposable batteries, or
damage from accident, neglect, misuse, alteration,
contamination, or abnormal conditions of operation or
handling. Resellers are not authorized to extend any other
warranty on Fluke’s behalf. To obtain service during the
warranty period, contact your nearest Fluke authorized
service center to obtain return authorization information,
then send the product to that Service Center with a
description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER
WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR
PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS
NOT LIABLE FOR ANY SPECIAL, INDIRECT,
INCIDENTAL OR CONSEQUENTIAL DAMAGES OR
LOSSES, ARISING FROM ANY CAUSE OR THEORY.
Since some states or countries do not allow the exclusion or
limitation of an implied warranty or of incidental or
consequential damages, this limitation of liability may not
apply to you.
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
11/99
Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
11.5" (29.2 cm) extended
Weight: 0.5 lbs. (0.68 kg)
Construction materials:
Body and piston: Acetal
O-Rings: Buna N
Other wetted parts: Brass or nickel-plated
brass