Fluke 700P2X Series, 700P22, 700P23, 700P24 Instruction Sheet

Page 1
Certifications
P Conforms to European Standard EN61010-1, EN61326.
®
Limited Warranty
Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year and begins on the date of shipment. Parts, product repairs and services are warranted for 90 days. This warranty extends only to the original buyer or end user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries or to any product which, in Fluke’s opinion, has been misused, altered, neglected or damaged by accident or abnormal conditions of operation or handling. Fluke warrants that software will operate substantially in accordance with its functional specifications for 90 days and that it has been properly recorded on nondefective media. Fluke does not warrant that software will be error free or operate without interruption.
Fluke authorized resellers shall extend this warranty on new and unused products to enduser customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price. Fluke reserves the right to invoice Buyer for importation costs of repair/replacement parts when product purchased in one country is submitted for repair in another country.
Fluke’s warranty obligation is limited, at Fluke’s option, to refund of the purchase price, free of charge repair, or replacement of a defective product which is returned to a Fluke authorized service center within the warranty period.
To obtain warranty service, contact your nearest Fluke authorized service center or send the product, with a description of the difficulty, postage and insurance prepaid (FOB Destination), to the nearest Fluke authorized service center. Fluke assumes no risk for damage in transit. Following warranty repair, the product will be returned to Buyer, transportation prepaid (FOB Destination). If Fluke determines that the failure was caused by misuse, alteration, accident or abnormal condition of operation or handling, Fluke will provide an estimate of repair costs and obtain authorization before commencing the work.
Following repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges (FOB Shipping Point).
THIS WARRANTY IS BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCI DENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHER ARISING FROM BREACH OF WARRANTY OR BASED ON CONTRACT , TORT, RELIANCE OR ANY OTHER THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
700P2X Series Pressure Modules
Instruction Sheet
Low
Introduction
The Fluke 700P2X Series Pressure Modules allow you to measure pressure with the Fluke 700 Series Documenting Process Calibrators. Read this sheet before you use the pressure module. This sheet contains specifications and information about how to avoid damaging the pressure module. See the 700 Series Users Manual for operating instructions.
Fluke-701 or Fluke-702 software V1.3 or higher is required. See specification footnote 1.
The pressure module measures pressure using an internal microprocessor. It receives operating power from and sends digital information to the 700 Series calibrator.
Gage pressure modules have one pressure fitting and measure pressure with respect to atmospheric pressure. Differential pressure modules have two pressure fittings and measure the difference between the applied pressure on the high fitting versus the low fitting. A differential pressure module functions like a gage module when the low fitting is open.
Differential
RANGE
700
PRESSURE MODULE
P
04
15 PSID/G
100 kPa
1 bar
Note
High
In Case of Difficulty
For service or calibration, call your nearest authorized Fluke Service Center.
For application or operation assistance or information on Fluke products, call:
USA: 1-888-99-FLUKE (1-888-993-5853) Canada: 1-800-36-FLUKE (1-800-363-5853) Europe: +31 402-678-200 Japan: +81-3-3434-0181 Singapore: +65-738-5655 Anywhere in the world: +1-425-446-5500
Or, visit Fluke’s Web site at www.fluke.com Fluke Corporation Fluke Europe B.V.
P.O. Box 9090 P.O. Box 1186 Everett, WA 98206-9090 5602 B.D. Eindhoven USA The Netherlands
Box Contents
Pressure module, strap, 1/4 NPT to 1/4 ISO metric adapters, instruction sheet.
Protecting Yourself from Pressure Releases
To avoid a violent release of pressure in a pressurized system, shut off the isolation valve and slowly bleed off the pressure before you attach or remove the pressure module from the pressure line.
Avoiding Mechanical Damage
To avoid damaging the pressure module, never apply more than 10 ft.-lbs. of torque between the pressure module fittings or between the fittings and the body of the module. Always apply appropriate torque between the pressure module fitting and connecting fittings or adapters. Figure 1 shows the correct way and incorrect ways to use a wrench when applying torque to the pressure module fitting.
PN 601093 (English) August 1995 Rev.2, 11/01
1995-2001 Fluke Corporation. All rights reserved. Printed in USA
Page 2
Pressure Calibration Kit
RANGE
70
PRESS
0
P
0 4
U RE M
15 PSID/G
OD
10
U
0 kP
L E
a
1 b a r
The Fluke-700PCK Pressure Calibration Kit makes it possible to calibrate pressure modules at ambient temperature with a precision pressure calibrator better than the module specification. A 386 or better PC and Windows® 3.1 or later are required. The kit is an optional accessory available from your distributor or Fluke.
Performance Test
If you need to check that the pressure module meets its accuracy specification, use a dead weight tester or suitable pressure calibrator. The accuracy of the dead weight tester or pressure calibrator should be significantly better than the
Figure 1.
Avoiding Overpressure Damage
Applying pressure in excess of the BURST PRESSURE specified on the pressure module can destroy the pressure module. Burst pressure is 3X full scale.
Avoiding Corrosion Damage
To avoid damaging the pressure module from corrosion, use it only with specified media as shown below:
High: any medium that is compatible with type 316 stainless steel.
Low: dry, noncorrosive gasses only.
Recommended Measurement Technique
For best results, it is recommended that the module be pressurized to full scale and then vented to zero pressure (atmosphere) prior to zeroing and making measurements.
Note
Low range pressure modules may be sensitive to gravity. For best results, pressure modules 30 psi and below should be held at the same physical orientation from the time they are zeroed until the measurement is complete.
Table 1. Specifications
Model Range
2
Fluke-700P22 0 to 1.0000 psi
0 to 6.8900 kPa 0 to 6.89 E-2 bar
Fluke-700P23 0 to 5.0000 psi
0 to 34.000 kPa 0 to 0.3400 bar
Fluke-700P24 0 to 15.000 psi
0 to 100.00 kPa 0 to 1.0000 bar
1. Use of pressure zero funct i on i s required to achieve these specifications. Contact your Fluke Service Cent er for upgrade of your Fluke 701 or Fluke 702 V1.0, V1.1, or V1. 2 Cal i brator.
2. Available pressure units (i nHg, kg/cm
3. Accuracy specifications apply for 1 y ear for 0 to 100% of full span from 0 t o 50 °C. Typical uncertainty i s 1% of full span from
-10 °C to 0 °C.
4. Maximum non-destructi ve pressure: 3X maximum rated pressure, including comm on m ode pressure.
5. Maximum common mode pressure: 3X maximum rated pressure.
6. Specifications reflect a confidence interval of 95%.
Gage or
Differential
Isolated or
Nonisolated
Differential Isolated 0.100% 0.020% 0.030% 0.150%
Differential Isolated 0.025% 0.010% 0.015% 0.050%
Differential Isolated 0.025% 0.010% 0.015% 0.050%
2
, mmH20) are determined by the calibrator being used.
700P2X Series pressure specification. Proceed as follows to verify that a pressure module is operating within specification:
1. Read the pressure value with no externally applied pressure to make sure the 0% of scale is correct. When reading the pressure, press the ZERO key to remove any zero offset.
Note
The pressure ZERO function is available on Fluke­701 and Fluke-702 calibrators with V1.3 or higher software. Contact your Fluke Service Center for upgrade of earlier calibrators.
2. Connect the pressure module to a dead weight tester.
3. Set the dead weight tester to 20% of the pressure modules full scale value.
4. Make sure the reading agrees with the dead weight tester value within the specification in Table 1.
5. Set the dead weight tester to 40, 60, 80, and 100% of full scale and compare the respective readings.
6. If temperature sensitivity is of concern, repeat steps 1 through 5 at various controlled temperatures.
1
(% of full span)
Reference
Uncertainty
Stability
(1 Year)
Temp
(0 to 50 °C)
Total
Uncertainty
(23 ° ± 3 °C)
3