Fluke 67-MAX Specifications

TECHNICAL DATA
Fluke 67 MAX Clinical Infrared Thermometer
FEATURES
Adjustable alarm visually alerts the user when the temperature exceeds a programmed limit.
Withstands a 6.5 foot (2 meter) drop
Powered by a single, standard AA battery
Comes with a three-year warranty
Offers small, lightweight form factor to slip in a
coat pocket
Clinical infrared thermometer: ± 0.5 °F from
71.6 °F to 109 °F (± 0.3 °C from 22 °C to 43 °C)
Non-contact thermal scanner device for measurement of human body temperature.
Body temperature results in seconds.
The Fluke 67 MAX Clinical Infrared Thermometer* is a handheld non-contact thermometer optimized for measuring human tem­perature. The Fluke 67 MAX is accurate within +/- 0.5 °F over a measurement range of 71.6 °F to 109 °F (+/- 0.3 °C accuracy within range from 22 °C to 43 °C).
The Fluke 67 MAX has been tested to IEC standard 60601-1-2 for electromagnetic compatibility. This tool has been tested to remain accurate when near other electronic devices such as portable RF communication equipment (two-way radios).
The Fluke 67 MAX, with its small size and versatile features, is extremely easy to use. Plus, it’s rugged enough to take a 2-meter drop, helpful if you work in busy, crowded work areas. You can also easily clean the surface of the Fluke 67 MAX with disinfecting
wipes.
*The F luke 67 M AX accuracy complies with and accuracy tested in accordance with ASTM
E1965-98 and ISO 80601-2-56 standard but is not FDA 510(k) cleared or approved
Specifications
General specifications
Display resolution 0.1 °F (0.1 °C)
Power AA battery
Battery life
Weight 8.8 oz (250 g)
Size 6.88 x 3.34 x 2.95 in (175 x 85 x 75 mm)
Operating temperature 59 °F to 104 °F (15 °C to 40 °C)
Storage temperature -4 °F to 122 °F (-20 °C to 50 °C), without batter y
Operating humidity 10 % to 95 % RH non-condensing
Operating altitude 2000 meters above mean sea level
Storage altitude 12,000 meters above mean sea level
Drop test
Vibration and shock IEC 68-2-6 2.5 g, 10 to 200 Hz, IEC 68-2-27, 50 g, 11 ms
Electromagnetic compliance IEC 60601-1-2: Portable electromagnetic equipment
Standard compliance IEC 60601-1: Pollution degree 2
ASTM E1965: 22 to 40.0 °C
IEC 80601-2-56: 34.0 °C to 43.0 °C
100 hours
6.5 feet (2 meters)
Ordering information
Fluke 67 MAX Clinical Infrared Thermometer
Included
Non-contact infrared thermometer Manual (1) AA battery Registration card Statement of calibration
2 Fluke Corporation Fluke 67 MAX Clinical Infrared Thermometer
Fluke. Keeping your world up and running.
Fluke Corporation
PO Box 9090, Everett, WA 98206 U.S.A.
For more in formation ca ll:
In the U.S.A. (800) 443-5853 In Canada (800) 36-FLUKE From other countries +1 (425) 446-5500 www.fluke.com
©2020 Fluke Corporation. Specifications subject to change without notice. 09/2020 6013435b-en
Modification of this document is not permitted without written permission from Fluke Corporation.
®