Fluke 65 User Manual

Page 1

Introduction

This calibration information for the Fluke 65 Infrared Thermometer (also referred to as “the UUT”) provides the following information:
®
Model 65
Infrared Thermometer
Calibration Information
Specifications
Basic maintenance
Equipment required for performance tests and calibration
Performance test procedures
Calibration adjustment procedures
List of replaceable parts
For complete operating instructions, refer to the Model 65 Infrared Thermometer Instruction Sheet.

Contacting Fluke

To locate an authorized service center, visit us on the World Wide Web at www.fluke.com. Or, call Fluke as follows:
USA: 1-888-99-FLUKE (1-888-993-5853)
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-678-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-446-5500
PN 1554518 January 2000 ©2000 Fluke Corporation. All rights reserved. Printed in U.S.A. 1
Page 2
Model 65
Calibration Information

Safety Information

To avoid possible personal injury:
The information provided in this manual is for the use of qualified personnel
only. Do not perform the verification tests or calibration procedures described in this manual unless you are qualified to do so.
Performance of procedures other than those specified herein may result in
hazardous light exposure.
Do not stare into the laser beam or direct it toward the eyes.
To avoid possible damage to the thermometer and to ensure correct measurements:
The Fluke 65 Infrared Thermometer contains parts that can be damaged by
static discharge. No procedure in this document requires the case to be opened. If you do so, follow the standard practices for handling static sensitive devices.
A Warning
Caution
Do not drop the thermometer or subject it to violent shocks.
Do not operate the thermometer near large electrical or magnetic fields.
Do not touch the object with the thermometer.
Keep the thermometer away from direct sunlight or strong sources of light,
hot objects (70 °C / 158 °F), high temperatures, high humidity, or dust during use and storage.
Do not apply sudden temperature changes to the thermometer. Before
taking measurements, wait for the thermometer to return to a stable temperature.
During performance tests and calibration, expose the thermometer to the
blackbody heat source only long enough to complete the measurement (< 15 seconds). Between measurements, move the thermometer to an area not affected by the heat source.
Do not touch the focal lens.
Condensation may form on the focal lens if the thermometer moves quickly
from a cold to a hot environment. Before taking measurements, wait for the condensation to dissipate.
2
Page 3

Specifications

Measurement Range 40 °C to +500 °C (40 °F to +932 °F) Operating Temperature (Ambient) 0 °C to +50 °C (+32 °F to +122 °F) Storage Temperature 20 °C to +70 °C (4 °F to +158 °F), without battery
Specifications
Display Resolution Below 200 °: 0.1°
Above 200
Measurement Accuracy Below 0 °C: ±5 °C (32 °F: ±9 °F)
Above 0 °C: ±2 °C (32 °F: ±4 °F) Above 100 °C: ±2 % of reading [Ambient: 25 °C ±3 °C (77 °F ±5 °F)]
Temperature Coefficient > 400 °C (752 °F): ±0.06 % reading
400 °C (752 °F): ±0.24 % reading [Ambient: < 22 °C (72 °F); > 28 °C (82 °F)]
Response Time 0.8 second (for 95 % response) Spectral Response 8 µm to 14 µm nominal Field of View/Target Size 8:1; 25 mm (1 in) minimum spot size Laser beam divergence < 0.01 radian Repeatability ±1 % of reading or ±1 °, whichever is greater Emissivity Fixed at 0.95 Humidity 10 % to 90 % RH non-condensing Altitude Storage: 0 km to 12 km (40,000 ft)
°: 1°
Operating: 0 km to 3 km (10,000 ft)
Sensor Element Thermopile Power Supply 2 AA alkaline batteries, not installed Battery Life > 15 hours with laser and backlight activated (> 4000 individual measurements under
typical conditions)
Backlight Auto-on under low-light conditions Hold Temperature value is held on the screen for 20 seconds Dimensions 38.1 mm H × 63.5 mm W × 185.4 mm L
(1.5 in H × 2.5 in W × 7.3 in L)
Weight 283.5 g (10 oz) Safety/Regulatory Compliance
Certification
Approval
US 21 CFR Subchapter J Part 1040.10 IEC 60825-1 (1998-01) Edition 1.1 EN 60825-1:1994/A11:1996
3
Page 4
Model 65
Calibration Information

Maintenance

Installing the Batteries

Remove the holster to access the battery door. Figure 1 shows how to replace the batteries.
Match the + and polarities of the battery with the battery case.
Do not attempt to recharge the batteries.
Do not throw batteries into a fire.
Follow local laws or regulations when disposing of batteries.
The
B indicates battery life remaining. Replace both batteries when I starts blinking.
Caution

Cleaning the Focal Lens

Do not use solvents to clean the focal lens. Blow off loose particles using clean, dry, compressed air. Gently brush remaining debris away with a camel hair brush or high-quality lens tissue. The tissue may be moistened with clean water.

Cleaning the Case and Holster

Use soap and water or a mild commercial cleaner. Wipe with a damp sponge or soft rag.
4
Figure 1. Replacing the Batteries
acd02f.eps
Page 5

Required Equipment

Table 1 shows the equipment required for the performance tests and calibration adjustments.
Equipment Minimum Specifications Recommended Model
Required Equipment
Table 1. Required Equipment
Blackbody Calibration Source Temperature range: Ambient to 400 °C
Minimum aperture diameter: 76 mm (3.0 in) Maximum temperature: 400 °C Accuracy: ±0.5 °C to 100 °C
±0.5 % of reading above 100 °C

Setting Up for Performance Tests and Ca libration

1. Turn on the blackbody. Set it to room temperature and let it stabilize as recommended by the
manufacturer.
2. Let the UUT stabilize in the calibration area environment (25 °C ±3 °C) for a minimum of 20 minutes
before proceeding.
3. Verify that the UUT’s laser light is activated. When you press G, A appears in the display and the laser
spot is visible. If not, turn on the laser switch located under the battery hatch.
4. Place the UUT 8” from the front of the blackbody. Center the UUT’s thermopile on the blackbody’s
aperture. Note that the laser is offset from the thermopile.

Performance Tests

Use the following procedure to verify the performance of the UUT. If the UUT fails any of the tests, it needs calibration or repair.
Mikron M315 Blackbody Calibration Source with 0.95 emissivity er ror correction table
Caution
To avoid inaccurate readings during performance tests, expose the UUT to the blackbody heat source only long enough to complete the measurement (< 15 seconds). Between measurements, move the UUT to an area not affected by the heat source.
1. Set up the UUT and the blackbody as described under “Setting Up for Performance Tests and
Calibration”.
2. The emissivity of the blackbody will be different than the emissivity of the UUT. Use the emissivity
error correction table provided for the blackbody to determine the proper set point for each temperature in the first column of Table 2 at 0.95 emissivity.
3. For each temperature you determined in step 2, do the following: a. Set the blackbody to the temperature. Let the blackbody stabilize. b. Use the UUT to measure the temperature of the blackbody. Verify that the UUT reading is within
the limits given.
5
Page 6
Model 65
Calibration Information
Table 2. Performance Test Values
Temperature at 0.95 Emissivity UUT Reading Limits
Room temperature Room temperature ±2 °C (±3.6 °F)
200 °C (392 °F) 196 °C to 204 °C (384.8 °F to 399.2 °F) 300 °C (572 °F) 294 °C to 306 °C (561.2 °F to 582.8 °F) 400 °C (752 °F) 392 °C to 408 °C (737.6 °F to 766.4 °F)
1. Room temperature must be 25 °C ±3 °C (77 °F ±5 °F).
2. Limits given are ±2 % of temperature at 0.95 emissivity.

Calibration Adjustments

The Model 65 Infrared Thermometer should be calibrated and, if necessary, adjusted annually using the following procedure:
To avoid inaccurate readings during calibration, expose the UUT to the blackbody heat source only long enough to complete the measurement (< 15 seconds). Between measurements, move the UUT to an area not affected by the heat source.
1, 2
Caution
Caution
Do not make any calibration adjustments when UUT is in factory calibration mode (refer to Figure 2). Doing so will cause the UUT to give inaccurate readings, and you will need to return the UUT to a Fluke service center for recalibration.
1. Turn on the UUT.
2. Press and hold Fand D until all display segments are lit (about 5 seconds); then also press and hold Eand G until the UUT displays 1_L (about 5 seconds), as shown in Figure 2. To scroll the calibration menu, press F or D.
3. Set up the UUT and the blackbody as described under “Setting Up for Performance Tests and
Calibration”.
4. In the room-temperature adjustment mode (display shows 1_L), press G. The default room temperature appears on the UUT display.
5. Verify that the UUT thermopile is centered on the blackbody target; then use F or D to adjust the UUT reading down or up, respectively, to match the room temperature reading at 0.95 emissitivity.
6. When you have set the UUT reading to room temperature at 0.95 emissivity, press G. The UUT display flashes briefly, then returns to the calibration selection mode.
7. Use the emissivity error correction table to adjust the blackbody for an output of 350 °C at 0.95 emissivity.
6
Page 7
Room temperature calibration mode
High temperature calibration mode
MIN
MAX
°
C
°
F
Factory Calibration mode.
Caution
Do not change the factory calibration setting
Calibration Adjustm ent s
End calibration mode Press to exit
calibration mode
Figure 2. Calibration Menu
acd01f.eps
7
Page 8
Model 65
Calibration Information
8. Press D to enter the high-temperature adjustment mode (display shows 2_H), as shown in Figure 2.
9. Press G. The default high-temperature appears on the UUT display.
10. Aim the UUT at the blackbody target; then use F or D to adjust the UUT reading down or up,
respectively, to read 350 °C.
11. When you have set the UUT reading to 350 °C, press G. The UUT display flashes briefly, then returns to the calibration selection mode.
If the high temperature you enter during calibration varies greatly from the measured temperature, the UUT display will flash Err after you press G. In this case, press G again to exit the error mode. Verify that the blackbody is working properly; then try the calibration again.
12. Press D to enter the end-of-calibration mode (display shows E), as shown in Figure 2. Press G to exit the calibration menu.
13. Test the UUT’s operation by running the performance tests given earlier under “Performance Tests”. If the UUT fails the performance tests, recalibrate the UUT and try the tests again. If the UUT continues to fail, return it to a Fluke Service Center for repair or replacement.
Note

Replacement Parts

Table 3 shows the replacement parts available from Fluke for the Model 65 Infrared Thermometer.
Table 3. Replacement Parts
Description Fluke Part Number
Top case 803566 Bottom case 803509 Battery door 803517 LCD lens 803558 End cap 803525 Keypad 803541 Softcase 670676 Holster 670612 AA battery, 1.5 V, carbon-zinc (2 required) 650181 AA battery, 1.5 V, alkaline (2 required) 376756
Model 65 Infrared Thermometer Instruction Sheet
Americas (English, Spanish, French, Portuguese)
801826 European I (English, French, German, Italian, Finnish, Dutch, ) European II (English, Norwegian, Spanish, Portuguese, Swedish, Danish) Asian (English, Korean, Japanese, Simplified Chinese, Traditional Chinese, Thai)
Model 65 Infrared Thermometer Quick Reference Card
8
(English only) 802261
801842
801859 801834
Loading...