Fluke 6100 User Manual

4.5 (2)
DataSim 6100
Patient Simulator
Operators Manual
PN 2242959 February 2008
© 2008 Fluke Corporation, All rights reserved. Printed in USA. Specifications subject to change without notice. All product names are trademarks of their respective companies.

Warranty and Product Support

Fluke Biomedical warrants this instrument against defects in materials and workmanship for one year from the date of original purchase. During the war­ranty period, we will repair or at our option replace, at no charge, a product that proves to be defective, provided you return the product, shipping prepaid, to Fluke Biomedical. This warranty covers the original purchaser only and is not transferable. The warranty does not apply if the product has been damaged by accident or misuse or has been serviced or modified by anyone other than an authorized Fluke Biomedical service facility. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, IN­DIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THE­ORY.
This warranty covers only serialized products and their accessory items that bear a distinct serial number tag. Recalibration of instruments is not covered under the warranty
This warranty gives you specific legal rights and you may also have other rights that vary in different jurisdictions. Since some jurisdictions do not allow the exclusion or limitation of an implied warranty or of incidental or conse­quential damages, this limitation of liability may not apply to you. If any pro­vision of this warranty is held invalid or unenforceable by a court or other de­cision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
07/07
Notices
All Rights Reserved
© Copyright 2008, Fluke Biomedical. No part of this publication may be reproduced, transmit­ted, transcribed, stored in a retrieval system, or translated into any language without the written permission of Fluke Biomedical.
Copyright Release
Fluke Biomedical agrees to a limited copyright release that allows you to reproduce manuals and other printed materials for use in service training programs and other technical publications. If you would like other reproductions or distributions, submit a written request to Fluke Biomedical.
Unpacking and Inspection
Follow standard receiving practices upon receipt of the instrument. Check the shipping carton for damage. If damage is found, stop unpacking the instrument. Notify the carrier and ask for an agent to be present while the instrument is unpacked. There are no special unpacking instructions, but be careful not to damage the instrument when unpacking it. Inspect the instrument for physi­cal damage such as bent or broken parts, dents, or scratches.
Technical Support
For application support or answers to technical questions, either email
techservices@flukebiomedical.com
1-425-446-6945.
Claims
Our routine method of shipment is via common carrier, FOB origin. Upon delivery, if physical damage is found, retain all packing materials in their original condition and contact the carrier immediately to file a claim. If the instrument is delivered in good physical condition but does not operate within specifications, or if there are any other problems not caused by shipping damage, please contact Fluke Biomedical or your local sales representative.
or call 1-800- 648-7952 or
Standard Terms and Conditions
Refunds and Credits
Please note that only serialized products and their accessory items (i.e., products and items bearing a distinct serial number tag) are eligible for partial refund and/or credit. Nonserialized parts and accessory items (e.g., cables, carrying cases, auxiliary modules, etc.) are not eligible for return or refund. Only products returned within 90 days from the date of original purchase are eligible for refund/credit. In order to receive a partial refund/credit of a product purchase price on a serialized product, the product must not have been damaged by the customer or by the carrier chosen by the customer to return the goods, and the product must be returned complete (meaning with all manuals, cables, accessories, etc.) and in “as new” and re­salable condition. Products not returned within 90 days of purchase, or products which are not in “as new” and resalable condition, are not eligible for credit return and will be returned to the cus­tomer. The Return Procedure (see below) must be followed to assure prompt refund/credit.
Restocking Charges
Products returned within 30 days of original purchase are subject to a minimum restocking fee of 15 %. Products returned in excess of 30 days after purchase, but prior to 90 days, are subject to a minimum restocking fee of 20 %. Additional charges for damage and/or missing parts and acces­sories will be applied to all returns.
Return Procedure
All items being returned (including all warranty-claim shipments) must be sent freight-prepaid to our factory location. When you return an instrument to Fluke Biomedical, we recommend using United Parcel Service, Federal Express, or Air Parcel Post. We also recommend that you insure your shipment for its actual replacement cost. Fluke Biomedical will not be responsible for lost shipments or instruments that are received in damaged condition due to improper packaging or handling.
Use the original carton and packaging material for shipment. If they are not available, we rec­ommend the following guide for repackaging:
Use a double-walled carton of sufficient strength for the weight being shipped. Use heavy paper or cardboard to protect all instrument surfaces. Use nonabrasive
material around all projecting parts.
Use at least four inches of tightly packed, industry-approved, shock-absorbent
Returns for partial refund/credit:
Every product returned for refund/credit must be accompanied by a Return Material Authoriza­tion (RMA) number, obtained from our Order Entry Group at 1-800-648-7952 or 1-425-446-
6945.
Repair and calibration:
To find the nearest service center, go to www.flukebiomedical.com/service
In the U.S.A.: Cleveland Calibration Lab Tel: 1-800-850-4606 Email: globalcal@flukebiomedical.com
Everett Calibration Lab Tel: 1-888-993-5853 Email: service.status@fluke.com
In Europe, Middle East, and Africa: Eindhoven Calibration Lab Tel: +31-402-675300 Email: ServiceDesk@fluke.com
In Asia: Everett Calibration Lab Tel: +425-446-6945 Email: service.international@fluke.com
material around the instrument.
, or
Certification
This instrument was thoroughly tested and inspected. It was found to meet Fluke Biomedical’s manufacturing specifications when it was shipped from the factory. Calibration measurements are traceable to the National Institute of Standards and Technology (NIST). Devices for which there are no NIST calibration standards are measured against in-house performance standards us­ing accepted test procedures.
WARNING
Unauthorized user modifications or application beyond the published specifications may result in electrical shock hazards or improper operation. Fluke Biomedical will not be re­sponsible for any injuries sustained due to unauthorized equipment modifications.
Restrictions and Liabilities
Information in this document is subject to change and does not represent a commitment by Fluke Biomedical. Changes made to the information in this document will be incorpo­rated in new editions of the publication. No responsibility is assumed by Fluke Biomedi­cal for the use or reliability of software or equipment that is not supplied by Fluke Bio­medical, or by its affiliated dealers.
Manufacturing Location
The DataSim 6100 Patient Simulator is manufactured in Everett, Washington by Fluke Biomedical, 6920 Seaway Blvd., Everett, WA, U.S.A.

Table of Contents

Chapter Title Page
1 Introduction and Specifications.............................................. 1-1
Introduction .......................................................................................... 1-3
Features................................................................................................. 1-3
General Safety Considerations.............................................................. 1-4
Symbols ............................................................................................ 1-4
Warnings and Cautions..................................................................... 1-4
Instrument Familiarity .......................................................................... 1-5
Front Panel........................................................................................ 1-5
Pendant Keypad Controller............................................................... 1-9
Direct Functions............................................................................ 1-10
Special Functions.......................................................................... 1-10
PAC and PVC Insertion ................................................................ 1-17
Modifier/Cal Mode ....................................................................... 1-18
Specifications........................................................................................ 1-19
Accessories ........................................................................................... 1-21
2 Operation .................................................................................. 2-1
Introduction .......................................................................................... 2-3
Size Adjustments .................................................................................. 2-3
Calibration ............................................................................................ 2-5
Heart Rate Adjustment.......................................................................... 2-6
Artifact Simulation ............................................................................... 2-7
Respiration............................................................................................ 2-8
Defibrillation Training.......................................................................... 2-9
Blood Pressure Simulation.................................................................... 2-11
Personality Modules (Optional Special Functions)............................... 2-12
Interactive IABP Personality Modules.............................................. 2-16
Cardiac Output Personality Modules ................................................ 2-18
Blood Pressure Cables .......................................................................... 2-19
Connector Interface Pin Definitions...................................................... 2-21
3 Maintenance, Service, and Calibration................................... 3-1
Maintenance.......................................................................................... 3-3
Avoiding Damage............................................................................. 3-3
Cleaning............................................................................................ 3-3
i
DataSim 6100
Operators Manual
Battery Charging and Replacement...................................................... 3-4
Service and Calibration ........................................................................ 3-5
ii

List of Tables

Table Title Page
1-1. Symbols.............................................................................................. 1-4
1-2. Simulator Front Panel Controls and Indicators .................................. 1-7
1-3. Lead ECG Amplitude......................................................................... 1-8
1-4. Pendant Keypad Controls and Indicators ........................................... 1-10
1-5. Pendant Keypad Codes and Functions ............................................... 1-11
1-6. Functions Used in the Modifier/Cal Mode......................................... 1-18
1-7. Standard Accessories ......................................................................... 1-21
1-8. Optional Accessories (Series 90) Blood Pressure Cables................... 1-22
2-1. Waveform Amplitudes ....................................................................... 2-3
2-2. Effects of Size Adjustments on Parameters........................................ 2-4
2-3. Calibration Functions ......................................................................... 2-6
2-4. Heart Rates Obtained via Rate Adjust Special Function .................... 2-7
2-5. Artifact Modifier Effects.................................................................... 2-8
2-6. Personality Modules........................................................................... 2-13
2-7. Appropriate Pressure Simulator Interface Cables............................... 2-19
2-8. ECG Connectors ................................................................................ 2-21
2-9. Pressure Connectors ........................................................................... 2-22
iii
DataSim 6100
Operators Manual
iv

List of Figures

Figure Title Page
1-1. Simulator Front Panel Controls and Indicators .................................. 1-6
1-2. Pendant Keypad Controls and Indicators ........................................... 1-9
2-1. Pressure Cable Wiring for Monitor with 5 µV/V mmHg Sensitivity . 2-20 2-2. Pressure Cable Wiring for Monitor with 40 µV/V mmHg Sensitivity 2-21
2-3. Connector Pin Assignments ............................................................... 2-22
v
DataSim 6100
Operators Manual
vi
Chapter 1

Introduction and Specifications

Contents Page
Introduction ................................................................................. 1-3
Features........................................................................................ 1-3
General Safety Considerations..................................................... 1-4
Symbols.................................................................................... 1-4
Warnings and Cautions ............................................................ 1-4
Instrument Familiarity ................................................................. 1-5
Front Panel ............................................................................... 1-5
Pendant Keypad Controller ...................................................... 1-9
Direct Functions ................................................................... 1-10
Special Functions.................................................................. 1-10
PAC and PVC Insertion........................................................ 1-17
Modifier/Cal Mode............................................................... 1-18
Specifications............................................................................... 1-19
Accessories .................................................................................. 1-21
1-1
DataSim 6100
Operators Manual
1-2
Introduction and Specifications Introduction
1

Introduction

The DataSim 6100 Patient Simulator, hereafter referred to as the Simulator, is a battery-operated, six-channel patient simulator and defibrillation trainer.
Critical care nurses, ACLS program instructors, and other clinical educators can generate an extensive range of simulations, from a simple normal sinus rhythm to a complex Swan-Ganz catheter insertion, as well as a wide range of real world ECG, pressure, and respiration artifacts. In the standard defibrillation training mode, the instructor can interface the Simulator with patient manikins and teach students the correct way to defibrillate a patient, plus the desired effect on ECG and BP activity.
All control keys and display prompts are on a hand-held pendant keypad connected to the unit with a flexible, 20-foot, coiled telephone-style cord that gives a wide range of motion. Standard waveforms and sequences, selected via the keypad, are stored in microcomputer memory within the Simulator.
A variety of optional Personality Modules expands the number of available waveform selections and sequences and includes functions for special purpose ECG, Cardiac Output Dilution Curves, Capnography, and IABP. Each personality module plugs into the Simulator front panel.
A video adapter allows display of generated ECG and BP waveforms on any standard TV set. The enlarged format increases visibility and comfort level for a class of students.

Features

The Simulator has the following features:
Training capabilities
Six channels for generating ECG, arrhythmias, blood pressure, and
respiration
Synchronization of hemodynamic waveforms
Manual PAC and PVC insertions
Swan-Ganz procedure
Interface with Resusci-Anne, Arrhythmia Anne, and Chris Clean manikins
Optional Personality Modules
Video adapter interface
1-3
DataSim 6100
Operators Manual

General Safety Considerations

Read the Users Manual before operating the Analyzer.
Note
If calibration or other service of the Simulator is required, only qualified service personnel should be permitted to remove the front panel of the Simulator.

Symbols

Table 1-1 describes the symbols associated with the Analyzer.
Table 1-1. Symbols
Symbol Description
X Hazardous voltage W Important information; refer to manual. P Conforms to European Union directives
~
Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information.

Warnings and Cautions

A Warning identifies hazardous conditions and actions that could cause bodily harm or death.
A Caution identifies conditions and actions that could damage the Analyzer, the equipment under test, or cause permanent loss of data.
Levels of signals available via the output connectors and the Personality Module connector do not exceed 12 volts, and as such, do not constitute a potential danger to the operator. However, observe the following:
1-4
Introduction and Specifications Instrument Familiarity
XW Warning
To avoid possible electrical shock or personal injury, follow these guidelines:
• Use this Simulator only in the manner specified by the manufacturer or the protection provided may be impaired.
• Read the Users Manual before operating the Simulator .
• Do not connect the Simulator to a patient or equipment connected to a patient. The Simulator is intended for equipment evaluation only and should never be used in diagnostics, treatment or in any other capacity where the Simulator would come in contact with a patient
1

Instrument Familiarity

The functions of the Simulator are controlled from the front panel and the pendant keypad.

Front Panel

Figure 1-1 shows the front panel controls and indicators of the Simulator. Table 1-2 lists these components with accompanying descriptions.
1-5
DataSim 6100
Operators Manual
1 2
10
DataSim 6100
Patient Simulator
NSR1TACH2BRAD3ASYS
0
A-FLTR5A-FIB62AVB172AVB2
4
3AVB9PAC
V-TAC
PVC1 PVC2
SPEC
V-FIB
STD
FUNC
8
CPLT
KEYPAD
POWER
ECG
ART
PA
RA
On Off
LL (+)
LA (-)
RA
RL
V1
V2
V3
V4
V5
(-)
3
4
9
AUX
V6
PERSONALITY
8
MODULE
MANKIN
SIMULATE
DEFIB
5
6
1-6
7
Figure 1-1. Simulator Front Panel Controls and Indicators
fdg001.eps
Introduction and Specifications Instrument Familiarity
Table 1-2. Simulator Front Panel Controls and Indicators
Label Component Description
1
A Power Indicator Illuminates when power is ON
B Power Switch Switches power on and off
This battery operated unit is rechargeable with the supplied charger.
C ECG Snap
Connectors
D Pressure Output
Connectors
Outputs low level 12 lead ECG
Attach standard patient ECG cable lead wires to these snap connectors to display the simulated ECG waveforms on the monitor screen.
To minimize 60-cycle artifact, use a reference electrode connection. For example, most five-lead ECG diagnostic recorders require that a Simulator REF signal is connected to the RL input of the patient cable. The Simulator can also output the ECG in a modified chest lead configuration (MCL2) used for bedside monitoring. Simply attach the LL (+) red, RA (-) white, and LA (ref) patient cable snaps to the patient Simulator. Refer to the operators manual of the ECG monitor for hook-up instructions.
Output for arterial, PA, and RA pressures
This output simulates the electrical output of the BP transducer that would be used with the patient monitor. The Simulator generates signals that are compatible with either 5 or 40-microvolt transducers.
1-7
DataSim 6100
Operators Manual
Table 1-2. Simulator Front Panel Controls and Indicators (cont.)
Label Component Description
E Aux Input/Output
Connector
Cable interconnects for optional waveform, such as capnography and IABP
F Manual Defib Switch Manually simulates defibrillation
G Defib Connector Input for Arrhythmia Anne or Chris
Clean manikins equipped with defibrillation option
H Personality Module
Connector
Input for optional modules
Used to add numerous functions and waveforms to the standard set
I Keypad Connector Input for hand-help keypad
J Keypad Rest Storage area for the keypad
Inner
Well
Back Battery Charger Input Connector
Table 1-3. Lead ECG Amplitude
Lead # ECG Amplitude
I +.25 mv (+/- 5%)
II +1.0
III +.75
AVR .60
AVL -.25
AVF +.90
V1 .40 V2 .90 V3 +.60 V4 +1.0 V5 +1.5 V6 + 1.0
Jack used to recharge battery
The unit can be operated while recharging.
1-8
Loading...
+ 44 hidden pages