Table 1 lists the International Electrical Symbols. Some or all of these symbols
may be used on the instrument or in this manual.
Table 1 International Electrical Symbols
SymbolDescription
AC (Alternating Current)
AC-DC
1 Before You Start
Symbols Used
Battery
CE Complies with European Union Directives
DC
Double Insulated
Electric Shock
Fuse
PE Ground
Hot Surface (Burn Hazard)
Read the User’s Manual (Important Information)
Off
On
1
Page 10
6020 Series Calibration Bath
User’s Guide
SymbolDescription
Canadian Standards Association
OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 re
fers to the level of Impulse Withstand Voltage protection provided. Equipment of
OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from
the fixed installation. Examples include household, office, and laboratory appliances.
C-TIC Australian EMC Mark
The European Waste Electrical and Electronic Equipment (WEEE) Directive
(2002/96/EC) mark.
1.2Safety Information
Use this instrument only as specified in this manual. Otherwise, the protection
provided by the instrument may be impaired.
The following definitions apply to the terms “Warning” and “Caution”.
• “Warning” identifies conditions and actions that may pose hazards to the
user.
• “Caution” identifies conditions and actions that may damage the instru-
ment being used.
-
1.2.1
Warnings
To avoid personal injury, follow these guidelines.
GENERAL
•
Appropriate personal safety protection should be worn by the operator at
all times while using the bath.
•
DO NOT use the instrument for any application other than calibration
work. The instrument was designed for temperature calibration. Any other
use of the unit may cause unknown hazards to the user.
•
DO NOT use the unit in environments other than those listed in the user’s
guide.
•
DO NOT overfill the bath. Overflowing extremely hot fluid may be harm
ful to the operator. See Section 5.3, Bath Preparation and Filling, for spe
cific instructions.
•
Follow all safety guidelines listed in the user’s manual.
•
Calibration Equipment should only be used by Trained Personnel.
•
If this equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment may be impaired.
-
-
2
Page 11
1 Before You Start
Safety Information
Before initial use, or after transport, or after storage in humid or semi-hu
•
mid environments, or anytime the instrument has not been energized for
more than 10 days, the instrument needs to be energized for a “dry-out”
period of 2 hours before it can be assumed to meet all of the safety re
quirements of the IEC 1010-1. If the product is wet or has been in a wet
environment, take necessary measures to remove moisture prior to apply
ing power such as storage in a low humidity temperature chamber operat
ing at 50°C for 4 hours or more.
DO NOT operate high temperature baths near flammable materials. Ex
•
treme temperatures could ignite the flammable material.
Overhead clearance is required. Do not place the instrument under a cabi
•
net or other structure. Always leave enough clearance to allow for safe
and easy insertion and removal of probes.
The instrument is intended for indoor use only.
•
• Ensure that you check the appropriate OSHA and local fire code regula-
tions for proper equipment required to suppress a fire for the fluid utilized
in the bath.
-
-
-
-
-
BURN HAZARD
• High temperatures may be present in this equipment. Fires and severe
burns may result if personnel fail to observe safety precautions.
• Ensure the bath DOES NOT contain any water and has been completely
dried prior to filling with fluid. Any trapped water can cause a steam explosion resulting in personal injury. If the bath has recently been filled
with water, ensure the inside of the drain tube is dry prior to filling the
bath with fluid.
• When immersing any object in the bath, ensure that you are not introduc-
ing anything into the bath that will react with the bath fluid. Ensure that
probes are DRY and free of contaminants. Read the MSDS (Material
Safety Data Sheet) for the bath fluid used.
•
The bath is provided with an access cover. The access cover can be modi
fied to allow the probes being calibrated to be inserted through the access
cover. The bath should be operated with the access cover on at all times.
Operating the bath without the access cover reduces stability, increases
out gassing of fluids, and increases the possibility of personal injury or
fire hazard.
•
When removing probes from the bath DO NOT wipe probes down with a
paper towel. If the bath setpoint is high, the paper towel can ignite in your
hand. Provide a safe surface and situation for the probes to cool prior to
cleaning them.
-
•
The bath generates extreme temperatures. Precautions must be taken to
prevent personal injury or damage to objects. Probes may be extremely
hot when removed from the bath. Cautiously handle probes to prevent
personal injury. Carefully place probes on a heat resistant surface or rack
until they are at room temperature.
3
Page 12
6020 Series Calibration Bath
User’s Guide
Be extremely careful when filling the bath and bringing it up to tempera
•
ture. Most fluids expand with temperature. It is easy to add fluid, but ex
tremely dangerous to remove hot fluid when the bath is over filled. DONOT fill the bath above the indicated fill line when the stirring is on. It
will overflow.
ELECTRICAL HAZARD
These guidelines must be followed to ensure that the safety mechanisms
•
in this instrument will operate properly. The power cord of the instrument
is equipped with a three-pronged grounding plug for your protection
against electrical shock hazards. It must be plugged directly into a prop
erly grounded three-prong receptacle. The receptacle must be installed in
accordance with local codes and ordinances. Consult a qualified electri
cian. DO NOT use an extension cord or adapter plug.
DO use a ground fault interrupt device. This unit contains a liquid. A
•
ground fault device is advised in case liquid is present in the electrical
system and could cause an electrical shock.
-
-
-
-
• Always replace the power cord with an approved cord of the correct rat-
ing and type. If you have questions, contact an Authorized Service Center
(see Section 1.3).
• High voltage is used in the operation of this equipment. Severe injury or
death may result if personnel fail to observe the safety precautions. Before
working inside the equipment, turn off the power and disconnect the
power cord.
• Keep all combustible materials away from the bath. Operate the bath on a
heatproof surface such as concrete. Provide a means of safety for containing any spill, which may occur.
BATH FLUIDS
•
Fluids used in this unit may produce noxious or toxic fumes under certain
circumstances. Consult the fluid manufacturer’s MSDS (Material Safety
Data Sheet). Proper ventilation and safety precautions must be observed.
•
The unit is equipped with a soft cutout (user settable firmware) and a hard
cutout (set at the factory). Check the flash point, boiling point, or other
fluid characteristic applicable to the circumstances of the unit operation.
Ensure that the soft cutout is adjusted to the fluid characteristics of the ap
plication. Failing to set the cutout to the limits of the bath fluid can result
in fire hazards and personal injury.
-
1.2.2
Cautions
To avoid possible damage to the instrument, follow these guidelines.
•
Always operate this instrument at room temperature between 41°F and
104°F (5°C to 40°C). Allow sufficient air circulation by leaving at least 6
inches (15 cm) of clearance around the instrument.
4
Page 13
1 Before You Start
Safety Information
DO NOT overfill the bath. Overflowing liquid may damage the electrical
•
system. Be sure to allow for thermal expansion of the fluid as the bath
temperature increases. See Section 5.3, Bath Preparation and Filling, for
specific instructions.
Read Section 6, Bath Use, before placing the unit into service.
•
DO NOT change the values of the bath calibration constants from the fac
•
tory set values. The correct setting of these parameters is important to the
safety and proper operation of the unit.
Only authorized personnel should perform the Factory Reset Sequence if
•
no other action is successful in correcting a malfunction. You must have a
copy of the most recent Report of Test to restore the test parameters.
DO NOT operate this instrument in an excessively wet, oily, dusty, or
•
dirty environment.
• The bath is a precision instrument. Although it has been designed for op-
timum durability and trouble free operation, it must be handled with care.
Position the bath before the tank is filled with fluid.
-
• Most probes have handle temperature limits. Be sure that the probe handle
temperature limit is not exceeded in the air above the instrument.
• The instrument and any thermometer probes used with it are sensitive in-
struments that can be easily damaged. Always handle these devices with
care. Do not allow them to be dropped, struck, stressed, or overheated.
• DO ensure the fluid is cleaned from the probes prior to immersing the
probe in the next bath. Clean your probe between each bath to avoid contamination between bath fluids.
• Under filling the bath may reduce the bath performance and may possibly
damage the bath.
• When calibrating PRTs always follow correct calibration procedure and
calibrate from high temperatures to low temperatures with the appropriate
triple point of water checks. Never immerse a wet or cold PRT into a bath
filled with hot fluid. Severe damage to the PRT may result as well as per
-
sonal injury to the calibration technician.
•
This bath is not designed to be portable. Therefore, moving the bath once
it has been installed should be kept to a minimum. NEVER MOVE ABATH THAT IS FULL OF “HOT” FLUID. This action could be ex
tremely dangerous and could result in personal injury to the person mov
-
ing the bath. If the bath is going to be placed in an area where it may need
to be moved frequently, Hart Scientific sells a special cart designed to ac
commodate the bath. However, even with a cart the bath should not be
moved full of “hot” fluid. The fluid can splash causing injury or if the
bath and cart tip, the “hot” fluid could cause damage to the surrounding
area and personnel.
•
If the bath must be moved, be sure to drain the fluid to prevent any injury.
To safely move the bath, two people are required. One person should lift
5
Page 14
6020 Series Calibration Bath
User’s Guide
the bath by the tower, while the second person lifts from underneath the
bath tank.
The control probe must be inserted through the lid into the bath fluid and
•
plugged into the socket at the back of the bath. DO NOT operate the bath
without the control probe properly inserted and attached. The bath will
not operate correctly without the control probe. Injury to operating per
sonnel and permanent damage to the bath could occur.
1.3Authorized Service Centers
Please contact one of the following authorized Service Centers to coordinate
service on your Hart product:
Fluke Corporation, Hart Scientific Division
799 E. Utah Valley Drive
American Fork, UT 84003-9775
-
USA
Phone: +1.801.763.1600
Telefax: +1.801.763.1010
E-mail: support@hartscientific.com
Fluke Nederland B.V.
Customer Support Services
Science Park Eindhoven 5108
5692 EC Son
NETHERLANDS
Phone: +31-402-675300
Telefax: +31-402-675321
E-mail: ServiceDesk@fluke.nl
Fluke Int'l Corporation
Service Center - Instrimpex
Room 2301 Sciteck Tower
22 Jianguomenwai Dajie
Chao Yang District
Beijing 100004, PRC
CHINA
6
Page 15
Phone: +86-10-6-512-3436
Telefax: +86-10-6-512-3437
E-mail: xingye.han@fluke.com.cn
Fluke South East Asia Pte Ltd.
Fluke ASEAN Regional Office
Service Center
60 Alexandra Terrace #03-16
The Comtech (Lobby D)
118502
SINGAPORE
Phone: +65 6799-5588
Telefax: +65 6799-5588
1 Before You Start
Authorized Service Centers
E-mail: antng@singa.fluke.com
When contacting these Service Centers for support, please have the following
information available:
• Model Number
• Serial Number
• Voltage
• Complete description of the problem
7
Page 16
Page 17
2Introduction
2 Introduction
The Hart Scientific 6020 series hot baths are constant temperature baths in
tended mainly for the application of temperature calibration. However, their
high stability and the availability of factory modifications, such as a circulation
pump, make these baths suitable for other uses as well.
-
Figure 1 6020 Series of Calibration Baths
9
Page 18
Page 19
3 Specifications and Environmental Conditions
Specifications
3Specifications and Environmental
Conditions
3.1Specifications
6020602160226024
Range40°C to 300°C40°C to 200°C40°C to 300°C40°C to 300°C
Stability±0.001°C at 40°C (water)
±0.003°C at 100°C (oil)
±0.005°C at 300°C (oil)
Uniformity±0.002°C at 40°C (water)
±0.004°C at 100°C (oil)
±0.012°C at 300°C (oil)
Temperature SettingDigital display with push-button data entry
software and RS-232 computer interface are available for setting bath temperature via remote com
puter. For IEEE-488, add the 2001-IEEE to the automation package.
115/230 VAC (±10%),
50/60 Hz, 1518 W
1150 W
230 VAC (±10%), 5 A,
1150 W
115 VAC: 20 A, 250 V, F
230 VAC: 10 A, 250 V, F
115 VAC: 15 A, 250 V, F
230 VAC: 8 A, 250 V, F
32” H x 16” W x 20” D
(813 x 406 x 508 mm)
115/230 VAC (±10%),
50/60 Hz, 1350 W
NANA
115 VAC: 15 A, 250 V, F
230 VAC: 8 A, 250 V, F
115 VAC: 12 A, 250 V, F
230 VAC: 6 A, 250 V, F
11.2 gallons
(42 liters)
32" H x 16" W x 20" D
(813 x 406 x 508 mm)
115/230 VAC (±10%),
50/60 Hz, 1350 W
115 VAC: 15 A, 250 V, F
230 VAC: 8 A, 250 V, F
115 VAC: 12 A, 250 V, F
230 VAC: 6 A, 250 V, F
27.5" H x 19" W x 23" D
(699 x 483 x 587 mm)
-
11
Page 20
6020 Series Calibration Bath
User’s Guide
3.2Environmental Conditions
These instruments have been designed for optimum durability and trouble-free
operation, however, they must be handled with care. The instrument should not
be operated in an excessively dusty or dirty environment. Maintenance and
cleaning recommendations can be found in the Maintenance Section of this
manual.
The instrument operates safely under the following conditions:
ambient temperature range: 5 - 40°C (41 - 104°F)
•
ambient relative humidity: maximum 80% for temperature < 31°C, de
•
creasing linearly to 50% at 40°C
pressure: 75kPa - 106kPa
•
mains voltage within ± 10% of nominal
•
vibrations in the calibration environment should be minimized
•
• altitude less than 2,000 meters
-
• indoor use only
12
Page 21
4Quick Start
CAUTION: READ SECTION 6 ENTITLED BATH USE before placing the
bath in service. Incorrect handling can damage the bath and void the war
ranty.
This section gives a brief summary of the steps required to set up and operate
the bath. This should be used as a general overview and reference and not as a
substitute for the remainder of the manual. Please read Sections 5 through 8
carefully before operating the bath.
4.1Unpacking
Unpack the bath carefully and inspect it for any damage that may have occurred during shipment. If there is shipping damage, notify the carrier
immediately.
4 Quick Start
Unpacking
-
Verify that all components are present:
• Bath
• Access Hole Cover
• Control Probe
• Manual
• Report of Test
If you are missing any item, please call an Authorized Service Center (see Section 1.3) or assistance.
4.2Set Up
Set up of the bath requires careful unpacking and placement of the bath, filling
the bath with fluid, installing the probe and connecting power. Consult Section
5 for detailed instructions for proper installation of the bath. Be sure to place
the bath in a safe, clean and level location.
Fill the bath tank with an appropriate liquid. For operation at warm bath tem
peratures, clean distilled water works well. Carefully pour the fluid into the
bath tank through the large rectangular access hole above the tank avoiding
spilling any fluid. The fluid must not exceed a height of 1/2 inch below the bath
lid.
-
The control probe must be inserted through the lid into the bath and plugged
into the socket at the back of the bath. DO NOT operate the bath without the
control probe properly installed.
13
Page 22
6020 Series Calibration Bath
User’s Guide
4.3Power
Plug the bath power cord into a mains outlet of the proper voltage, frequency,
and current capability. See Section 3.1, Specifications. Set the “HEATER”
switch on the front panel to position “LOW” and turn the bath on using the
front panel “POWER” switch. The bath will turn on and begin to heat or cool to
reach the previously programmed temperature set-point. The front panel LED
display will indicate the actual bath temperature.
4.4Setting the Temperature
In the following discussion and throughout this manual a solid box around the
word SET, UP, EXIT or DOWN indicates the panel button while the dotted box
indicates the display reading. Explanation of the button or display reading are
to the right of each button or display value.
To view or set the bath temperature set-point proceed as follows. The front
panel LED display normally shows the actual bath temperature.
24.68
When “SET” is pressed the display will show the set-point memory that is currently being used and its value. Eight set-point memories are available.
S
C
Access set-point selection
Bath temperature display
1. 25.0Set-point 1, 25.0°C currently used
Press “SET” to select this memory and access the set-point value.
S
Access set-point value
C 25.00Current value of set-point 1, 25.00°C
Press “UP” or “DOWN” to change the set-point value.
U
Increment display
C 30.00New set-point value
14
Press “SET” to accept the new value and display the vernier value. The bath be
gins heating or cooling to the new set-point.
S
Store new set-point, access vernier
0.00000Current vernier value
-
Page 23
Setting the Temperature
Press “EXIT” and the bath temperature will be displayed again.
4 Quick Start
E
Return to the temperature display
24.73 CBath temperature display
The bath will heat or cool until it reaches the new set-point temperature. Set the
heater switch to position “HIGH” to allow the bath to more quickly reach a
higher temperature. The “HIGH” setting may be necessary to reach higher tem
peratures and control at high temperatures.
When setting the set-point temperature be careful not to exceed the temperature
limit of the bath fluid. The over-temperature cutout should be correctly set to
prevent this from happening. See Section 9.8.
To obtain optimum control stability adjust the proportional band as discussed in
Section 9.7.
-
15
Page 24
Page 25
5Installation
CAUTION: READ SECTION 6 ENTITLED BATH USE
before placing the bath in service. Incorrect handling can damage the
bath and void the warranty.
This bath is not designed to be portable. Therefore, moving the bath once it has
been installed should be kept to a minimum.
5 Installation
Bath Environment
Never move a bath that is full of “hot” fluid. This action could be ex
tremely dangerous and could result in personal injury to the person mov
ing the bath.
If the bath is going to be placed in an area where it will need to be moved fre
quently, a special cart can be designed to accommodate the bath. Thus making
the bath much more portable. Hart sells carts designed for these baths. How
ever, even with a cart the bath should not be moved full of “hot” fluid. The
fluid can splash causing injury or if the bath and cart tip, the “hot” fluid could
cause damage to the surrounding area and personal injury to personnel.
If the bath must be moved, be sure to drain the fluid to prevent any injury. To
safely move the bath, two people are required. One person should lift the bath
by the tower, while the second person slides their hands underneath the bath
tank and lifts the bath.
Refer to the Appendix specific to your bath model for additional installation
instructions.
5.1Bath Environment
CAUTION: DO NOT place under a cabinet or other structure. Allow for
overhead clearance.
-
-
-
-
The bath is a precision instrument which should be located in an appropriate
environment. The location should be free of drafts, extreme temperatures and
temperature changes, dirt, etc. The surface where the bath is placed must be
level.
Because the bath is designed for operation at high temperatures, keep all flam
mable and meltable materials away from the bath. Although the bath is well in
sulated, top surfaces do become hot. Beware of the danger of accidental fluid
spills.
If used at high temperatures, a fume hood should be used to remove any va
pors given off by hot bath fluid.
Allow for sufficient air circulation by leaving at least six inches of space be
tween the bath and nearby objects.
-
-
-
-
17
Page 26
6020 Series Calibration Bath
User’s Guide
5.2“Dry-out” Period
Before initial use, after transport, and any time the instrument has not been en
ergized for more than 10 days, the bath needs to be energized for a “dry-out”
period of 1-2 hours before it can be assumed to meet all of the safety require
ments of the IEC 1010-1.
5.3Bath Preparation and Filling
Baths are not provided with fluid. Various fluids are available from Hart Scien
tific and other sources. Depending on the desired temperature range, any of the
following fluids, as well as others, may be used in the bath:
Water
•
• Ethylene Glycol/Water
Mineral oil
•
• Dow 200.10 silicone oil
• Dow 200.20 silicone oil
• Dow 710 silicone oil
Fluids are discussed in detail in Section 8.1.
Remove the access hole cover from the bath and check the tank for foreign
matter (dirt, remnant packing material, etc.). Use clean unpolluted fluid. Carefully fill the bath through the large square access hole to a level that will allow
for stirring and thermal expansion. The fluid should never exceed a height of
1/2" below the top of the tank. Carefully monitor the bath fluid level as the bath
temperature rises to prevent overflow or splashing. Remove excess hot fluid if
necessary with caution.
-
-
-
18
Use caution to prevent bath fluid from spilling on the stirring motor while
filling.
NOTE: Under filling may reduce bath performance and may possibly
damage the bath heater.
5.4Probe
Inspect the bath control probe. It should not be bent or damaged in any way.
Reasonable caution should be used in handling this probe as it contains a preci
sion platinum sensor and is mechanically shock sensitive. Dropping, striking,
or other physical shock may cause a shift in resistance in the probe resulting in
diminished bath accuracy. If damaged, the probe can be replaced. Contact an
Authorized Service Center (see Section 1.3) for assistance.
-
Page 27
5 Installation
Insert the probe into the 1/4 inch probe hole at the top left side of the bath lid.
The tip of the probe must be well immersed in the fluid. The probe connector is
plugged into the rear of the bath into the socket labeled “PROBE”.
5.5Power
With the bath power switch off, plug the bath into an AC mains outlet of the
appropriate voltage, frequency, and current capacity. See Section 3.1,
Specifications.
Be sure the stirring motor power cord is plugged into the “STIRRER” socket at
the back of the bath.
5.6Tipping Prevention Brackets Installation (6024
with model 2072 cart only)
Power
WARNING: Hart Scientific recommends installing the bath on a heat-re-
sistant non-flammable surface.
In order to create the safest possible conditions under use, the bath comes
equipped with Tipping Prevention Brackets. The installation of the brackets is
required for compliance with the International Safety Standard IEC 61010-1,Section 7.3, Stability, which applies to the stability of the bath under normal
operating conditions.
5.6.1Installation on a Concrete Floor
Usingaconcretedrillandconcretedrillbit,drillthree1/4inchx11/4inch
deep (approximately 6.5 mm x 32 mm) holes in the concrete floor using the
bracket to mark the hole placement. Drop the flare anchor bolt into the hole.
Tightening the screw expands the anchor in the drilled hole and secures the
bracket. Ensure that the bracket is installed in such a way as to ensure the bath
will have a minimum of 6 inches of clearance for air circulation (see Figure 2).
Screw the brackets securely to the floor.
5.6.2Installation of the Bath
Slide the back casters of the bath completely into the bracket caster slots (see
Figure 2). Lock the front casters of the bath. Check that the bath is securely
locked into the brackets by gently pushing on the bath. Proceed to fill the bath
with the applicable bath fluid after reading the entire User's Guide. If you have
any questions concerning installation of the tipping prevention brackets, please
contact a Hart Authorized Service Center (see Section 1.3).
19
Page 28
6020 Series Calibration Bath
User’s Guide
Caster Slots
16.029” (40.714cm)
8.5” (21.6cm)
Front
Figure 2 Tipping Prevention Bracket Installation
8.5” (21.6cm)
Wall
20
Page 29
6Bath Use
CAUTION: Read before placing the bath in service.
6 Bath Use
General
The information in this section is for general information only. It is not de
signed to be the basis for calibration laboratory procedures. Each laboratory
will need to write their own specific procedures.
6.1General
Be sure to select the correct fluid for the temperature range of the calibration.
Bath fluids should be selected to operate safely with adequate thermal proper
ties to meet the application requirements. Also, be aware that some fluids ex
pand and could overflow the bath if not watched. Refer to General Operation,
section 8, for information specific to fluid selection and to the MSDS sheet specific to the fluid selected. Generally, baths are set to one temperature and used
to calibrate probes only at that single temperature. This means that the type of
bath fluid does not have to change. Additionally, the bath can be left energized
reducing the stress on the system.
The bath generates extreme temperatures. Precautions must be taken to prevent
personal injury or damage to objects. Probes may be extremely hot or cold
when removed from the bath. Cautiously handle probes to prevent personal injury. Carefully place probes on a heat/cold resistant surface or rack until they
are at room temperature. It is advisable to wipe the probe with a clean soft cloth
or paper towel before inserting it into another bath. This prevents the mixing of
fluids from one bath to another. If the probe has been calibrated in liquid salt,
carefully wash the probe in warm water and dry completely before transferring
it to another fluid. Always be sure that the probe is completely dry before in
serting it into a hot fluid. Some of the high temperature fluids react violently to
water or other liquid mediums. Be aware that cleaning the probe can be danger
ous if the probe has not cooled to room temperature. Additionally, high temper
ature fluids may ignite the paper towels if the probe has not been cooled.
-
-
-
-
-
-
For optimum accuracy and stability, allow the bath adequate stabilization time
after reaching the set-point temperature.
6.2Comparison Calibration
Comparison calibration involves testing a probe (unit under test, UUT) against
a reference probe. After inserting the probes to be calibrated into the bath, al
low sufficient time for the probes to settle and the temperature of the bath to
stabilize.
One of the significant dividends of using a bath rather than a dry-well to cali
brate multiple probes is that the probes do not need to be identical in construc
tion. The fluid in the bath allows different types of probes to be calibrated at
-
-
-
21
Page 30
6020 Series Calibration Bath
User’s Guide
the same time. However, stem effect from different types of probes is not to
tally eliminated. Even though all baths have horizontal and vertical gradients,
these gradients are minimized inside the bath work area. See Section 7.2 enti
tled Bath Tank and Lid. Nevertheless, probes should be inserted to the same
depth in the bath liquid. Be sure that all probes are inserted deep enough to pre
vent stem effect. From research at Hart Scientific, we suggest a general
rule-of-thumb for immersion depth to reduce the stem effect to a minimum: 15
x the diameter of the UUT + the sensor length. Do not submerge the probehandles. If the probe handles get too warm during calibration at high tempera
tures, a heat shield could be used just below the probe handle. This heat shield
could be as simple as aluminum foil slid over the probe before inserting it in
the bath or as complicated as a specially designed reflective metal apparatus.
When calibrating over a wide temperature range, better results can generally be
achieved by starting at the highest temperature and progressing down to the
lowest temperature.
Probes can be held in place in the bath by using probe clamps or drilling holes
in the access cover. Other fixtures to hold the probes can be designed. The object is to keep the reference probe and the probe(s) to be calibrated as closely
grouped as possible in the working area of the bath. Bath stability is maximized
when the bath working area is kept covered.
-
In preparing to use the bath for calibration start by:
• Placing the reference probe in the bath working area.
• Placing the probe to be calibrated, the UUT, in the bath working area as
close as feasibly possible to the reference probe.
6.3Calibration of Multiple Probes
Fully loading the bath with probes increases the time required for the temperature to stabilize after inserting the probes. Using the reference probe as the
guide, be sure that the temperature has stabilized before starting the calibration.
22
Page 31
7Parts and Controls
Refer to the Appendix specific to your bath model for additional information on
parts and controls.
7.1Front Panel
The following controls and indicators are present on the controller front panel
(see Figure 3 below): (1) the digital LED display, (2) the control buttons, (3)
the bath on/off power switch, (4) the control indicator light, and (5) the heater
power switch.
(1) The digital display is an important part of the temperature controller because it not only displays set and actual temperatures but also various bath
functions, settings, and constants. The display shows temperatures in values according to the selected scale units °C or °F.
(2) The control buttons (SET, DOWN, UP, and EXIT) are used to set the bath
temperature set-point, access and set other operating parameters, and access
and set bath calibration parameters.
7 Parts and Controls
Front Panel
4
HEATER
MODE
HIGH PRECISION BATH 20°C to 300°C
6020
1
Smoked DisplayWindow
SET
A brief description of the functions of the buttons follows:
SET – Used to display the next parameter in a menu and to set parameters to
the displayed value.
DOWN – Used to decrement the displayed value of setable parameters.
UP – Used to in crement the displayed value.
EXIT – Used to exit from a menu. When EXIT is pressed any changes made to
the displayed value will be ignored.
(3) The on/off switch controls power to the entire bath assembly. It powers the
stirring motor and the bath controller/heater circuit.
DOWN
HEATING
5
UP
EXIT
2
HIGH
LOW
POWER
3
I
O
Figure 3 Front Panel Features
23
Page 32
6020 Series Calibration Bath
User’s Guide
(4) The control indicator is a two color light emitting diode. This indicator lets
the user visually see the ratio of heating to cooling. When the indicator is red
the heater is on, and when it is green the heater is off and the bath is cooling.
High PrecisionBath
2
4
Hart
Scientific
5
3
6021, 6022
only
1
Figure 4 Typical 6020 Series Calibration Bath
24
6020, 6021,
6022, 6024
6
Page 33
7 Parts and Controls
Bath Tank and Lid
(5) The heater power switch is used to select the appropriate heater power lev
els for heating and controlling temperatures.
7.2Bath Tank and Lid
The bath tank and lid assembly includes: (1) the tank, (2) the control probe, (3)
the stir motor, (4) the access hole, (5) the access hole cover, and (6) the bath
drain. (See Figure 4 on page 24.)
(1) The bath tank is constructed of stainless steel and is very resistant to oxida
tion in the presence of most chemicals and over a wide range of temperatures.
(2) The control probe provides temperature feedback signal to the controller al
lowing the controller to maintain a constant temperature. The control probe is a
precision platinum resistance thermometer (PRT). The control probe is delicate
and must be handled carefully. The probe is placed in the small hole in the top
of the bath so that the probe tip is fully immersed in the bath fluid. The probe
cable connects to the bath at the probe connector on the back panel.
(3) The stir motor is mounted on the bath tank lid. It drives the stirring propeller to provide mixing of the bath fluid. Proper mixing of the fluid is important
for good constant temperature stability. The stir motor power cord plugs into
the back of the bath at the power socket labeled “STIRRER”.
-
-
-
(4) On the bath lid is a large rectangular access hole. This hole is used for filling and emptying the bath with fluids and placement of thermometers and devices into the bath. When possible the access hole should be covered. The
working area is defined as one inch from the sides and bottom and three
inches from the top of the fluid.
(5) An access hole cover is recommended to cover the large square access
opening in the top of the bath. This improves bath temperature stability, prevents excess fluid evaporation or fumes and increases safety with hot fluid. The
user may drill or cut holes in the cover to accommodate the instruments to be
calibrated or immersed in the bath. Spare covers are available from Hart
Scientific.
(6) A drain valve is provided for ease of removing the fluid media from the
bath. Always use a container of adequate size to hold the full load of fluid.
Some oils are more easily drained at higher temperatures. (See caution note
below.)
CAUTION: Extreme caution must be maintained to prevent harm to the
user or the surrounding environment. Do not exceed a 120°C fluid temper
ature for draining. The valve could be damaged if 120°C is exceeded. In
-
sulate the container from the floor or other objects that may be damaged
by high temperatures as required.
25
Page 34
6020 Series Calibration Bath
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
User’s Guide
7.3Back Panel
The back panel of the bath contains the following features: (1) the bath power
cord, (2) the stir motor power socket, (3) the system fuses (located internally),
(4) the probe socket, (5) the RS-232 interface connector (optional), (6) the
IEEE-488 interface connector (optional), (7) the cooling tubes, and (8) the bath
drain. (See Figure 5 on page 26.)
(1) The bath power cord extends from the back of the bath. It provides power
for the bath temperature controller, the heaters, and the stir motor. The cord is
plugged into an AC mains socket of appropriate voltage and current. See Sec
tion 3.1, Specifications for specific details.
(Optional)(Optional)
IEEE-488RS-232PROBESTIRRER
-
Figure 5 Back Panel
134562
7
COOLING TUBES
8
(2) The stir motor power cord plugs into the socket labeled “STIRRER”. Power
to the stir motor is switched on by the POWER switch on the control panel.
(3) The system fuses are located internally. Never replace the fuse with one of a
rating greater than that specified.
(4) The control probe plugs into the bath at the socket on the back of the bath
labeled “PROBE”.
(5) If the bath is supplied with a serial RS-232 interface, the interface cable is
attached to the back of the bath at the connector labeled “RS-232”.
26
(6) If the bath is supplied with a GPIB IEEE-488 interface, the interface cable
is attached to the back of the bath at the connector labeled “IEEE-488”.
Page 35
7 Parts and Controls
Back Panel
(7) The cooling tubes protrude from the bottom of the back of the bath. Cooling
fluid may be circulated around the bath by pumping fluid through the tubes.
(8) A drain valve is provided for ease of removing the fluid media from the
bath. Always use a container of adequate size to hold the full load of fluid.
Some oils are more easily drained at higher temperatures. (See caution note
above.)
27
Page 36
Page 37
8General Operation
8.1Bath Fluid
8 General Operation
Bath Fluid
Many fluids work with the 6020 series baths. Choosing a fluid requires consid
eration of many important characteristics of the fluid. Among these are temper
ature range, viscosity, specific heat, thermal conductivity, thermal expansion,
electrical resistivity, fluid lifetime, safety, and cost.
8.1.1Temperature Range
One of the most important characteristics to consider is the temperature range
of the fluid. Few fluids work well throughout the complete temperature range
of the bath. The temperature at which the bath is operated must always be
within the safe and useful temperature range of the fluid. The lower temperature range of the fluid is determined by the freeze point of the fluid or the temperature at which the viscosity becomes too great. The upper temperature is
usually limited by vaporization, flammability, or chemical breakdown of the
fluid. Vaporization of the fluid at higher temperatures may affect temperature
stability because of cool condensed fluid dripping into the bath from the lid.
The bath temperature should be limited by setting the safety cutout so that the
bath temperature cannot exceed the safe operating temperature limit of the
fluid.
8.1.2Viscosity
-
-
Viscosity is a measure of the thickness of a fluid, how easily it can be poured
and mixed. Viscosity affects the temperature stability of the bath. With low viscosity, fluid mixing is better which creates a more uniform temperature
throughout the bath. This improves the bath response time which allows it to
maintain a more constant temperature. For good control the viscosity should be
less than 10 centistokes. 50 centistokes is about the upper limit of allowable
viscosity. Viscosities greater than this cause very poor control stability and may
also overheat or damage the stirring motor. With oils viscosity may vary greatly
with temperature.
When using fluids with higher viscosities the controller proportional band may
need to be increased to compensate for the reduced response time. Otherwise
the temperature may begin to oscillate.
8.1.3Specific Heat
Specific heat is the measure of the heat storage ability of the fluid. Specific
heat, to a small degree, affects the control stability. It also affects the heating
and cooling rates. Generally, a lower specific heat means quicker heating and
cooling. The proportional band may require some adjustment depending on the
specific heat of the fluid.
29
Page 38
6020 Series Calibration Bath
User’s Guide
8.1.4Thermal Conductivity
Thermal conductivity measures how easily heat flows through the fluid. Ther
mal conductivity of the fluid affects the control stability, temperature unifor
mity, and probe temperature settling time. Fluids with higher conductivity
distribute heat more quickly and evenly improving bath performance.
8.1.5Thermal Expansion
Thermal expansion describes how the volume of the fluid changes with temper
ature. Thermal expansion of the fluid used must be considered since the in
crease in fluid volume as the bath temperature changes may cause overflow.
Excessive thermal expansion may also be undesirable in applications where
constant liquid level is important. Many fluids including oils have significant
thermal expansion.
8.1.6Electrical Resistivity
Electrical resistivity describes how well the fluid insulates against the flow of
electric current. In some applications, such as measuring the resistance of bare
temperature sensors, it may be important that little or no electrical leakage occur through the fluid. In such conditions choose a fluid with very high
resistivity.
8.1.7Fluid Lifetime
-
-
-
-
Many fluids degrade over time because of evaporization, water absorption, gelling, or chemical breakdown. Often the degradation becomes significant near
the upper temperature limit of the fluid, substantially reducing the fluid’s
lifetime.
8.1.8Safety
When choosing a fluid always consider the safety issues associated. Obviously
where there are conditions of extreme hot or cold there can be danger to people
and equipment. Fluids may also be hazardous for other reasons. Some fluids
may be considered toxic. Contact with eyes, skin, or inhalation of vapors may
cause injury. A proper fume hood must be used if hazardous or bothersome va
pors are produced.
WARNING: Fluids at high temperatures may pose danger from BURNS,
FIRE, and TOXIC FUMES. Use appropriate caution and safety equip
ment.
Fluids may be flammable and require special fire safety equipment and proce
dures. An important characteristic of the fluid to consider is the flash point. The
flash point is the temperature at which there is sufficient vapor given off so that
when there is sufficient oxygen present and an ignition source is applied the va
por will ignite. This does not necessarily mean that fire will be sustained at the
-
-
-
-
30
Page 39
8 General Operation
Bath Fluid
flash point. The flash point may be either of the open cup or closed cup type.
Either condition may occur in a bath situation. The closed cup temperature is
always the lower of the two. The closed cup represents the contained vapors in
side the tank and the open cup represents the vapors escaping the tank. Oxygen
and an ignition source will be less available inside the tank.
-
Environmentally hazardous fluids require special disposal according to applica
ble federal or local laws after use.
8.1.9Cost
Cost of bath fluids may vary greatly, from cents per gallon for water to hun
dreds of dollars per gallon for synthetic oils. Cost may be an important consid
eration when choosing a fluid.
8.1.10Commonly Used Fluids
Below is a description of some of the more commonly used fluids and their
characteristics.
8.1.10.1Water
Water is often used because of its very low cost, availability, and excellent temperature control characteristics. Water has very low viscosity and good thermal
conductivity and heat capacity which makes it among the best fluids for control
stability at low temperatures. Temperature stability is much poorer at higher
temperatures because water condenses on the lid, cools and drips into the bath.
Water is safe and relatively inert. The electrical conductivity of water may prevent its use in some applications. Water has a limited temperature range, from a
few degrees above 0°C to a few degrees below 100°C. At higher temperatures
evaporation becomes significant. Water used in the bath should be distilled or
softened to prevent mineral deposits. Consider using an algicide chemical in the
water to prevent contamination.
-
-
-
8.1.10.2Ethylene Glycol
The temperature range of water may be extended by using a solution of 1 part
water and 1 part ethylene glycol (antifreeze). The characteristics of the ethylene
glycol-water solution are similar to water. Use caution with ethylene glycol
since the fluid is very toxic. Ethylene glycol must be disposed of properly.
8.1.10.3Mineral Oil
Mineral oil or paraffin oil is often used at moderate temperatures above the
range of water. Mineral oil is relatively inexpensive. At lower temperatures
mineral oil is quite viscous and control may be poor. At higher temperatures
vapor emission becomes significant. The vapors may be dangerous and use of a
fume hood is highly recommended. As with most oils, mineral oil expands as
temperature increases so be careful not to fill the bath too full that it overflows
when heated. The viscosity and thermal characteristics of mineral oil is poorer
31
Page 40
6020 Series Calibration Bath
User’s Guide
than water so temperature stability is not be as good. Mineral oil has very low
electrical conductivity. Use caution with mineral oil since it is flammable and
may also cause serious injury if inhaled or ingested.
8.1.10.4Silicone Oil
Silicone oils are available which offer a much wider operating temperature
range than mineral oil. Like most oils, silicone oils have temperature control
characteristics which are somewhat poorer than water. The viscosity changes
significantly with temperature and thermal expansion also occurs. These oils
have very high electrical resistivity. Silicone oils are fairly safe and non-toxic.
Silicone oils are fairly expensive.
8.1.11Fluid Characteristics Charts
Table 2 and Figure 6 on pages 33 and 34 have been created to provide help in
selecting a heat exchange fluid media for your constant temperature bath.
These charts provide both a visual and numerical representation of most of the
physical qualities important in making a selection. The list is not all inclusive.
There may be other useful fluids not shown in this listing.
The charts include information on a variety of fluids which are often used as
heat transfer fluid in baths. Because of the temperature range some fluids may
not be useful with your bath.
8.1.11.1Limitations and Disclaimer
The information given in this manual regarding fluids is intended only to be
used as a general guide in choosing a fluid. Though every effort has been made
to provide correct information we cannot guarantee accuracy of data or assure
suitablility of a fluid for a particular application. Specifications may change
and sources sometimes offer differing information. Hart Scientific cannot be liable for any personal injury or damage to equipment, product or facilities resulting from the use of these fluids. The user of the bath is responsible for
collecting correct information, exercising proper judgement, and insuring safe
operation. Operating near the limits of certain properties such as the flash point
or viscosity can compromise safety or performance. Your company’s safety
policies regarding flash points, toxicity, and such issues must be considered.
You are responsible for reading the MSDS (material safety data sheets) and act
ing accordingly.
8.1.11.2About the Graph
The fluid graph visually illustrates some of the important qualities of the fluids
shown.
-
32
Temperature Range: The temperature scale is shown in degrees Celsius. The
fluids’ general range of application is indicated by the shaded bands. Qualities
including pour point, freeze point, important viscosity points, flash point, boil
ing point and others may be shown.
-
Page 41
Table 2 Table of Various Fluids and Their Properties
8 General Operation
Bath Fluid
Fluid
(# = Hart Part
No.)
Halocarbon 0.8
#5019
Methanol–96°C (fr)60°C (b)54°C1.3 @ –35°C
Water0°C (fr)95°C (b)NONE1 @ 25°C
Ethylene
Glycol—50%
#5020
Mineral Oil
#5011
Dow Corning
200.5
Silicone Oil
Dow Corning
200.10
#5012
Dow Corning
200.20
#5013
Dow Corning
200.50
Silicone Oil #5014
Dow Corning 550
#5016
Dow Corning 710
#5017
Dow Corning
210-H
Silicone Oil
Heat Transfer Salt
#5001
*Limiting Factors — b - boiling point e - high evaporation fl - flash point fr - freeze point v - viscosity — Flash point test oc = open cup cc = closed cup
**Very low water solubility, ice will form as a slush from condensation below freezing.
Freezing Point: The freezing point of a fluid is an obvious limitation to stir
ring. As the freezing point is approached high viscosity may also limit
performance.
Pour Point: This represents a handling limit for the fluid.
Viscosity: Points shown are at 50 and 10 centistokes viscosity. When viscosity
is greater than 50 centistokes stirring is very poor and the fluid is unsatisfactory
for bath applications. Optimum stirring generally occurs at 10 centistokes and
below.
33
Page 42
6020 Series Calibration Bath
User’s Guide
–100°C0°C100°C200°C300°C400°C500°C600°C
Silicone Oil
5017
Silicone Oil
5014
Silicone Oil
5013
Silicone Oil
5012
Silicone Oil
5010
HFE 7500
5023
Halocarbon
5019
Mineral Oil
5011
Water
Methanol
Ethanol
Ethylene Glycol
(50/50 with H O)
Dynalene HF/LO
2
5020
5022
PP
PP
PP
PP
10 CS
FR (Pure)
10 CS
10 CS
FR
10 CS
FR
10 CS
FL 12°C
FL 16°C
10 CS
10 CS
10 CS
EP
10 CS
BP
FL 60°C
FL 133°C
EP 100°C
BP
BP
10 CS
10 CS
FL 302°CPP
FL 280°C
FL 232°C
FL 211°C
FL 177°C
Bath Salt
5001
Legend
Figure 6 Chart of Various Bath Fluids
Fume Point: A fume hood should be used. This point is very subjective in na
ture and is impacted by individual tolerance to different fumes and smells, how
well the bath is covered, the surface area of the fluid in the bath, the size and
ventilation of the facility where the bath is located and other conditions. We as
sume the bath is well covered at this point. This is also subject to company
policy.
Flash Point: The point at which ignition may occur. The point shown may be
either the open or closed cup flash point. Refer to the flash point discussion in
Section 8.1.8.
FR
Shaded area represents usable range of fluid starting at
50 centistokes. Lighter shading represents decreasing
viscosity, while vaporization and decomposition increase.
Black area represents liquid range with excessive
viscosity.
Range over which a fume hood is recommended.
Decomposition Starts
BP - Boiling Point
CS - Centistokes
EP - Evaporation Point
(fluid loss
due to evaporation)
FL - Flash Point
FR - Freeze Point
PP - Pour Point
-
-
34
Page 43
8 General Operation
Stirring
Boiling Point: At or near the boiling point of the fluid, the temperature stabil
ity is difficult to maintain. Fuming or evaporation is excessive. Large amounts
of heater power may be required because of the heat of vaporization.
Decomposition: The temperature may reach a point at which decomposition of
the fluid begins. Further increasing the temperature may accelerate decomposi
tion to the point of danger or impracticality.
8.2Stirring
Stirring of the bath fluid is very important for stable temperature control. The
fluid must be mixed well for good temperature uniformity and fast controller
response. The stirrer is precisely adjusted for optimum performance.
8.3Power
Power to the bath is provided by an AC mains supply. See Section 3.1, Specifi
cations, for additional details. Power to the bath passes through a filter to prevent switching spikes from being transmitted to other equipment.
To turn on the bath switch the control panel power switch to the ON position.
The stir motor will turn on, the LED display will begin to show the bath temperature, and the heater will turn on or off until the bath temperature reaches
the programmed set-point.
-
-
-
When powered on the control panel display will briefly show a four digit number. This number indicates the number of times power has been applied to the
bath. Also briefly displayed is data which indicates the controller hardware
configuration. This data is used in some circumstances for diagnostic purposes.
8.4Heater
The power to the bath heater is precisely controlled by the temperature control
ler to maintain a constant bath temperature. Power is controlled by periodically
switching the heater on for a certain amount of time using a solid-state relay.
The front panel red/green control indicator shows the state of the heater. The
control indicator glows red when the heater is on and glows green when the
heater is off. The indicator will pulse constantly when the bath is maintaining a
stable temperature.
The heater has two power level settings. The “HIGH” heater power setting is
used to quickly heat up the bath fluid to the desired operating temperature. The
“HIGH” heater power setting may also be required for control at high tempera
tures. The “LOW” setting is used for control at lower temperatures and for
scanning at slower rates. When controlling at the “HIGH” heater power setting
instead of “LOW” the proportional band may need to be increased (typically by
a factor of four) to compensate for the increase in power gain. Otherwise the
temperature may oscillate.
-
-
35
Page 44
6020 Series Calibration Bath
User’s Guide
Heater fuses are located internally. See Section 13, Troubleshooting, for more
information
8.5Cooling
WARNING: Do not use cooling water at temperatures approaching
100°C. When operating the bath at these high temperatures make sure the
hoses are disconnected and completely drained, otherwise there may be
danger of steam explosion.
Cooling water or other external coolant circulates beneath the bottom of the
bath tank to provide cooling. The cooling fluid must be of constant temperature
and flow rate for good temperature stability. For optimum control stability the
temperature difference between the cooling water and bath fluid should be
about 20°C. Temperatures for this fluid below 15°C are not recommended.
Attach
1
inch I.D. hose lines to the tubes protruding from the back of the bath.
4
Fasten the hoses tightly with hose clamps. Cooling fluid pressure should not
exceed the capability of the lines attached. Connect a valve in line with the
cooling water to regulate the flow rate. Run the outlet line to a drain or tank.
The cooling fluid must match the properties required for the bath temperature.
Fluids should have the same safety margin and careful consideration as any
fluidusedinthebath.
8.6Temperature Controller
The bath temperature is controlled by Hart Scientific’s unique hybrid digital/analog temperature controller. The controller offers the tight control stability
of an analog temperature controller as well as the flexibility and programmabil
ity of a digital controller.
The bath temperature is monitored with a platinum resistance sensor in the con
trol probe. The signal is electronically compared with the programmable refer
ence signal, amplified, and then fed to a pulse-width modulator circuit which
controls the amount of power applied to the bath heater.
The bath is operable within the temperature range given in the specifications.
For protection against solid-state relay failure or other circuit failure, the mi
cro-controller will automatically turn off the heater with a second mechanical
relay anytime the bath temperature is more than a certain amount above the
set-point temperature. As a second protection device, the controller is also
equipped with a separate thermocouple temperature monitoring circuit which
will shut off the heater if the temperature exceeds the cutout set-point.
-
-
-
-
36
The controller allows the operator to set the bath temperature with high resolu
tion, set the cutout, adjust the proportional band, monitor the heater output
power, and program the controller configuration and calibration parameters.
The controller may be operated in temperature units of degrees Celsius or Fahr
-
-
Page 45
8 General Operation
Temperature Controller
enheit. The controller is operated and programmed from the front control panel
using the four key switches and digital LED display. The controller may also be
optionally equipped with an serial RS-232 or IEEE-488 GPIB digital interface
for remote operation. Operation of the controller using the front control panel is
discussed in Section 9. Operation using the digital interfaces is discussed in
Section 10.
When the controller is set to a new set-point the bath will heat or cool to the
new temperature. Once the new temperature is reached the bath usually takes
10-15 minutes for the temperature to settle and stabilize. There may be a small
overshoot or undershoot of about 0.5°C.
37
Page 46
Page 47
9Controller Operation
This section discusses in detail how to operate the bath temperature controller
using the front control panel. Using the front panel key switches and LED dis
play the user may monitor the bath temperature, set the temperature set-point in
degrees C or F, monitor the heater output power, adjust the controller propor
tional band, set the cutout set-point, and program the probe calibration parame
ters, operating parameters, serial and IEEE-488 interface configuration, and
controller calibration parameters. operation is summarized in Figure 7 on page
40.
9.1Bath temperature
The digital LED display on the front panel allows direct viewing of the actual
bath temperature. This temperature value is what is normally shown on the display. The units, C or F, of the temperature value are displayed at the right. For
example,
9 Controller Operation
Bath temperature
-
-
-
25.00 C
The temperature display function may be accessed from any other function by
pressing the “EXIT” button.
9.2Reset Cutout
If the over-temperature cutout has been triggered then the temperature display
will alternately flash,
cutout
The message continues to flash until the temperature is reduced and the cutout
is reset.
The cutout has two modes — automatic reset and manual reset. The mode de
termines how the cutout is reset which allows the bath to heat up again. When
in automatic mode, the cutout resets itself as soon as the temperature is lowered
below the cutout set-point. With manual reset mode, the cutout must be reset by
the operator after the temperature falls below the set-point.
When the cutout is active and the cutout mode is set to manual “reset”,the
display flashs “cutout” until the user resets the cutout. To access the reset
cutout function press the “SET” button.
Bath temperature in degrees Celsius
Indicates cutout condition
-
S
The display indicates the reset function.
Access cutout reset function
39
Page 48
6020 Series Calibration Bath
User’s Guide
EXIT
Cutout Active
EXIT
EXIT
Display
Temperature
SET
Reset Cutout
SET
Select Setpoint
SET
Adjust Setpoint
EXIT
EXIT
EXIT
Secondary Functions
+
SET
EXIT
Display Power
SET
Set Proportional Band
SET
Set Cutout Temp.
SET
SET
EXIT
Adjust Vernier
EXIT
SET
EXIT
Set Scale °C/°F
SET
Configuration Menu
SET
EXITEXITEXITEXITEXIT
UP
Probe
Menu
SET
R0
SETSETSET
Adj. R0
SET/EXITSET/EXITSET/EXIT
ALPHA
SET
Adj. ALPHA
SET/EXIT
DOWN
EXITEXITEXIT
EXIT
Operating
Parameters
Menu
SETSETSETSET
Cutout
Reset Mode
SETSET
Adj. Cutout
Reset Mode
SET/EXIT
UPUPUP
DOWNDOWNDOWN
EXITEXIT
Serial
Interface
Menu
BAUD
Rate
Adjust
BAUD Rate
Sample
Period
SET
Adj. Sample
Period
SET/EXIT
Duplex
Mode
SET
Adj. Duplex
Mode
SET/EXIT
EXIT
EXIT
IEEE-488
Interface
Menu
Device
Address
Adj. Device
Address
SET/EXIT
IEEE-488
Option Installed
Calibration
Adjust CTO
SET
Adjust C0
SET/EXIT
SET
DO NOT CHANGE THESE VALUES. SEE MANUAL
Adjust CG
SET/EXIT
Menu
X 5
CTO
C0
EXIT
CG
EXIT
Figure 7 Controller Operation Flowchart
40
Linefeed
SET
EXIT
Adjust
Linefeed
SET/EXIT
Interface
Option Installed
H
SET
Adjust H
SET/EXIT
L
SET
Adjust L
SET/EXIT
EXIT
EXIT
Page 49
rESEt ?Cutout reset function
Press “SET” once more to reset the cutout.
9 Controller Operation
Temperature Set-point
S
This also switchs the display to the set temperature function. To return to dis
playing the temperature press the “EXIT” button. If the cutout is still in the
over-temperature fault condition the display continues to flash “cutout”.The
bath temperature must drop a few degrees below the cutout set-point before the
cutout can be reset.
Reset cutout
9.3Temperature Set-point
The bath temperature can be set to any value within the range and with resolu
tion as given in the specifications. The temperature range of the particular fluid
used in the bath must be known by the operator and the bath should only be op
erated well below the upper temperature limit of the liquid. In addition, the
cutout temperature should also be set below the upper limit of the fluid.
Setting the bath temperature involves three steps: (1) select the set-point memory, (2) adjust the set-point value, and (3) adjust the vernier if desired.
9.3.1Programmable Set-points
The controller stores 8 set-point temperatures in memory. The set-points can be
quickly recalled to conveniently set the bath to a previously programmed temperature set-point.
-
-
-
To set the bath temperature one must first select the set-point memory. This
function is accessed from the temperature display function by pressing “SET”.
The number of the set-point memory currently being used is shown at the left
on the display followed by the current set-point value.
25.00 C
S
Access set-point memory
1. 25.0
To change the set-point memory press “UP” or “DOWN”.
4. 40.0
Press “SET” to accept the new selection and access the set-point value.
S
Accept selected set-point memory
Bath temperature in degrees Celsius
Set-point memory 1, 25.0°C currently used
New set-point memory 4, 40.0°C
41
Page 50
6020 Series Calibration Bath
User’s Guide
9.3.2Set-point Value
The set-point value may be adjusted after selecting the set-point memory and
pressing “SET”. The set-point value is displayed with the units, C or F, at the
left.
C 40.00
If the set-point value need not be changed then press “EXIT” to resume dis
playing the bath temperature. Press “UP” or “DOWN” to adjust the set-point
value.
C 42.50
When the desired set-point value is reached press “SET” to accept the new
value and access the set-point vernier. If “EXIT” is pressed instead then any
changes made to the set-point will be ignored.
S
Accept new set-point value
Set-point 4 value in °C
New set-point value
9.3.3Set-point Vernier
The set-point value can only be set with a resolution of 0.01°C. The user may
want to adjust the set-point slightly to achieve a precise bath temperature. The
set-point vernier allows one to adjust the temperature below or above the
set-point by a small amount with very high resolution. Each of the 8 stored
set-points has an associated vernier setting. The vernier is accessed from the
set-point by pressing “SET”. The vernier setting is displayed as a 6 digit number with five digits after the decimal point. This is a temperature offset in degrees of the selected units, C or F.
-
42
0.00000
To adjust the vernier press “UP” or “DOWN”. Unlike most functions the ver
nier setting has immediate effect as the vernier is adjusted. “SET” need not be
pressed. This allows one to continually adjust the bath temperature with the
vernierasitisdisplayed.
0.00090
Next press “EXIT” to return to the temperature display or “SET” to access the
temperature scale units selection.
S
Access scale units
Current vernier value in °C
-
New vernier setting
Page 51
9.4Temperature Scale Units
The temperature scale units of the controller may be set by the user to degrees
Celsius (°C) or Fahrenheit (°F). The units are used in displaying the bath tem
perature, set-point, vernier, proportional band, and cutout set-point.
The temperature scale units selection is accessed after the vernier adjustment
function by pressing “SET”. From the temperature display function access the
units selection by pressing “SET” 4 times.
9 Controller Operation
Temperature Scale Units
-
25.00 C
S
1. 25.0
S
C 40.00
S
0.00000
S
Un= C
Press “UP” or “DOWN” to change the units.
Access set-point memory
Access set-point value
Access vernier
Access scale units selection
Bath temperature
Set-point memory
Set-point value
Vernier setting
Scale units currently selected
Un= F
Press “SET” to accept the new selection and resume displaying the bath
temperature.
S
Set the new units and resume temperature display
New units selected
9.5Secondary Menu
Functions which are used less often are accessed within the secondary menu.
The secondary menu is accessed by pressing “SET” and “EXIT” simulta
neously and then releasing. The first function in the secondary menu is the
heater power display.
-
43
Page 52
6020 Series Calibration Bath
User’s Guide
9.6Heater Power
The temperature controller controls the temperature of the bath by pulsing the
heater on and off. The total power being applied to the heater is determined by
the duty cycle or the ratio of heater on time to the pulse cycle time. This value
may be estimated by watching the red/green control indicator light or read di
rectly from the digital display. By knowing the amount of heating to the bath
the user can tell if the bath is heating up to the set-point, cooling down, or con
trolling at a constant temperature. Monitoring the percent heater power will let
the user know how stable the bath temperature is. With good control stability
the percent heating power should not fluctuate more than ±1% within one
minute.
The heater power display is accessed in the secondary menu. Press “SET” and
“EXIT” simultaneously and release. The heater power is displayed as a percent
age of full power.
-
-
-
S+E
12 Pct
To exit out of the secondary menu press “EXIT”. To continue on to the proportional band setting function press “SET”.
Access heater power in secondary menu
Heater power in percent
9.7Proportional Band
In a proportional controller such as this the heater output power is proportional
to the bath temperature over a limited range of temperatures around the
set-point. This range of temperature is called the proportional band. At the bottom of the proportional band the heater output is 100%. At the top of the proportional band the heater output is 0. Thus as the bath temperature rises the
heater power is reduced, which consequently tends to lower the temperature
back down. In this way the temperature is maintained at a fairly constant
temperature.
The temperature stability of the bath depends on the width of the proportional
band. See Figure 8. If the band is too wide the bath temperature will deviate ex
cessively from the set-point due to varying external conditions. This is because
the power output changes very little with temperature and the controller cannot
respond very well to changing conditions or noise in the system. If the propor
tional band is too narrow the bath temperature may swing back and forth be
-
-
-
44
Page 53
9 Controller Operation
Proportional Band
cause the controller overreacts to temperature variations. For best control
stability the proportional band must be set for the optimum width.
Figure 8 Bath Temperature Fluctuation at Various Proportional Band Settings
The optimum proportional band width depends on several factors among which
are fluid volume, fluid characteristics (viscosity, specific heat, thermal conductivity), heater power setting, operating temperature, and stirring. Thus the proportional band width may require adjustment for best bath stability when any of
these conditions change. Of these, the most significant factors affecting the optimum proportional band width are heater power setting and fluid viscosity.
The proportional band should be wider when the higher power setting is used
so that the change in output power per change in temperature remains the same.
The proportional band should also be wider when the fluid viscosity is higher
because of the increased response time.
The proportional band width is easily adjusted from the bath front panel. The
width may be set to discrete values in degrees C or F depending on the selected
units. The optimum proportional band width setting may be determined by
monitoring the stability with a high resolution thermometer or with the control
ler percent output power display. Narrow the proportional band width to the
point at which the bath temperature begins to oscillate and then increase the
band width from this point to 3 or 4 times wider. Table 3 lists typical propor
-
-
45
Page 54
6020 Series Calibration Bath
User’s Guide
tional band settings for optimum performance with a variety of fluids at se
lected temperatures.
Table 3 Proportional Band — Fluid Table
Proportional Band
FluidTemperatureHeater Setting
Water30.0°CLow0.04°C±0.0004°C
Water60.0°CLow0.04°C±0.001°C
Eth-Gly 50%35.0°CLow0.05°C±0.0005°C
Eth-Gly 50%60.0°CLow0.05°C±0.001°C
Eth-Gly 50%100.0°CHigh0.4°C±0.007°C
Oil35.0°CLow0.1°C±0.003°C
Oil60.0°CLow0.2°C±0.002°C
Oil100°CLow0.2°C±0.003°C
Oil150°CLow0.2°C±0.004°C
Oil160°C+High0.8°C±0.01°C
Stability
-
The proportional band adjustment may be accessed within the secondary menu.
Press “SET” and “EXIT” to enter the secondary menu and show the heater
power. Then press “SET” to access the proportional band.
S+E
12 Pct
S
Pb=0.101C
To change the proportional band press “UP” or “DOWN”.
Pb=0.060C
To accept the new setting and access the cutout set-point press “SET”. Pressing
“EXIT” will exit the secondary menu ignoring any changes just made to the
proportional band value.
S
Access proportional band
Accept the new proportional band setting
Access heater power in secondary menu
Heater power in percent
Proportional band setting
New proportional band setting
46
Page 55
9.8Cutout
As a protection against software or hardware fault, shorted heater triac, or user
error, the bath is equipped with an adjustable heater cutout device that will shut
off power to the heater if the bath temperature exceeds a set value. This protects
the heater and bath materials from excessive temperatures and, most impor
tantly, protects the bath fluids from being heated beyond the safe operating
temperature preventing hazardous vaporization, breakdown, or ignition of the
liquid. The cutout temperature is programmable by the operator from the front
panel of the controller. It must always be set below the upper temperature limit
of the fluid and no more than 10 degrees above the upper temperature limit of
the bath.
If the cutout is activated because of excessive bath temperature then power to
the heater will be shut off and the bath will cool. The bath will cool until it
reaches a few degrees below the cutout set-point temperature. At this point the
action of the cutout is determined by the setting of the cutout mode parameter.
The cutout has two selectable modes — automatic reset or manual reset. If the
mode is set to automatic, then the cutout will automatically reset itself when the
bath temperature falls below the reset temperature allowing the bath to heat up
again. If the mode is set to manual, then the heater will remain disabled until
the user manually resets the cutout.
9 Controller Operation
Cutout
-
The cutout set-point may be accessed within the secondary menu. Press “SET”
and “EXIT” to enter the secondary menu and show the heater power. Then
press “SET” twice to access the cutout set-point.
S+E
12 Pct
S
Pb=0.101C
S
CO= 210C
To change the cutout set-point press “UP” or “DOWN”.
CO= 95C
Access proportional band
Access cutout set-point
Access heater power in secondary menu
Heater power in percent
Proportional band setting
Cutout set-point
New cutout set-point
To accept the new cutout set-point press “SET”.
S
Accept cutout set-point
47
Page 56
6020 Series Calibration Bath
User’s Guide
The next function is the configuration menu. Press “EXIT” to resume display
ing the bath temperature.
9.9Controller Configuration
The controller has a number of configuration and operating options and calibra
tion parameters which are programmable via the front panel. These are ac
cessed from the secondary menu after the cutout set-point function by pressing
“SET.” There are 5 sets of configuration parameters — probe parameters, oper
ating parameters, serial interface parameters, IEEE-488 interface parameters,
and controller calibration parameters. The menus are selected using the “UP”
and “DOWN” keys and then pressing “SET”.
9.10Probe Parameters
The probe parameter menu is indicated by,
PrObE
Press “SET” to enter the menu. The probe parameters menu contains the parameters, R0 and ALPHA, which characterize the resistance-temperature relationship of the platinum control probe. These parameters may be adjusted to
improve the accuracy of the bath. This procedure is explained in detail in Sec-
tion 11.
Probe parameters menu
-
-
-
-
The probe parameters are accessed by pressing “SET” after the name of the parameter is displayed. The value of the parameter may be changed using the
“UP” and “DOWN” buttons. After the desired value is reached press “SET” to
set the parameter to the new value. Pressing “EXIT” will cause the parameter to
be skipped ignoring any changes that may have been made.
9.10.1R
0
This probe parameter refers to the resistance of the control probe at 0°C. Nor
mally this is set for 100.000 ohms.
9.10.2ALPHA
This probe parameter refers to the average sensitivity of the probe between 0
and 100°C. Normally this is set for 0.00385°C
9.11Operating Parameters
The operating parameters menu is indicated by,
PAr
Operating parameters menu
-
-1
.
48
Page 57
Press “UP” to enter the menu. The operating parameters menu contains the
cutout reset mode setting.
9.11.1Cutout Reset Mode
The cutout reset mode determines whether the cutout resets automatically when
the bath temperature drops to a safe value or must be manually reset by the
operator.
The parameter is indicated by,
Cutout reset mode parameter
Press “SET” to access the parameter setting. Normally the cutout is set for
manual mode.
Cto=rStCutout set for manual reset
To change to automatic reset mode press “UP” and then “SET”.
9 Controller Operation
Serial Interface Parameters
Cto=AutoCutout set for automatic reset
9.12Serial Interface Parameters
The serial RS-232 interface parameters menu is indicated by,
SErIALSerial RS-232 interface parameters menu
The serial interface parameters menu contains parameters which determine the
operation of the serial interface. These controls only apply to baths fitted with
the serial interface. The parameters in the menu are — BAUD rate, sample period, duplex mode, and linefeed.
9.12.1Baud Rate
The baud rate is the first parameter in the menu. The baud rate setting deter
mines the serial communications transmission rate.
The baud rate parameter is indicated by,
BAUdSerial BAUD rate parameter
-
Press “SET” to choose to set the baud rate. The current baud rate value will
then be displayed.
1200 bCurrent baud rate
49
Page 58
6020 Series Calibration Bath
User’s Guide
The baud rate of the bath serial communications may be programmed to
300,600,1200, or 2400 BAUD. Use “UP” or “DOWN” to change the baud rate
value.
2400 bNew baud rate
Press “SET” to set the baud rate to the new value or “EXIT” to abort the opera
tion and skip to the next parameter in the menu.
9.12.2Sample Period
The sample period is the next parameter in the serial interface parameter menu.
The sample period is the time period in seconds between temperature measure
ments transmitted from the serial interface. If the sample rate is set to 5, the
bath transmits the current measurement over the serial interface approximately
every five seconds. The automatic sampling is disabled with a sample period of
0. The sample period is indicated by,
SAmPLESerial sample period parameter
Press “SET” to choose to set the sample period. The current sample period
value is displayed.
SA= 1Current sample period (seconds)
Adjust the value with “UP” or “DOWN” and then use “SET” to set the sample
rate to the displayed value.
-
-
50
SA= 60New sample period
9.12.3Duplex Mode
The next parameter is the duplex mode. The duplex mode may be set to full du
plex or half duplex. With full duplex any commands received by the bath via
the serial interface are immediately echoed or transmitted back to the device of
origin. With half duplex the commands are executed but not echoed. The du
plex mode parameter is indicated by,
dUPLSerial duplex mode parameter
Press “SET” to access the mode setting.
dUP=FULLCurrent duplex mode setting
The mode may be changed using “UP” or “DOWN” and pressing “SET”.
dUP=HALFNew duplex mode setting
-
-
Page 59
9.12.4Linefeed
9 Controller Operation
IEEE-488 Parameters
The final parameter in the serial interface menu is the linefeed mode. This pa
rameter enables (on) or disables (off) transmission of a linefeed character (LF,
ASCII 10) after transmission of any carriage-return. The linefeed parameter is
indicated by,
LFSerial linefeed parameter
Press “SET” to access the linefeed parameter.
LF= OnCurrent linefeed setting
The mode may be changed using “UP” or “DOWN” and pressing “SET”.
LF= OFFNew linefeed setting
9.13IEEE-488 Parameters
Baths may optionally be fitted with an IEEE-488 GPIB interface. In this case
the user may set the interface address within the IEEE-488 parameter menu.
This menu does not appear on baths not fitted with the interface. The menu is
indicated by,
-
IEEEIEEE-488 parameters menu
Press “SET” to enter the menu.
9.13.1IEEE-488 Address
The IEEE-488 interface must be configured to use the same address as the external communicating device. The address is indicated by,
AddrESSIEEE-488 interface address
Press “SET” to access the address setting.
Add= 22Current IEEE-488 interface address
Adjust the value with “UP” or “DOWN” and then use “SET” to set the address
to the displayed value.
Add= 15New IEEE-488 interface address
51
Page 60
6020 Series Calibration Bath
User’s Guide
9.14Calibration Parameters
The operator of the bath controller has access to a number of the bath calibra
tion constants namely CTO, C0, CG, H, and L. These values are set at the fac
tory and must not be altered. The correct values are important to the accuracy
and proper and safe operation of the bath. Access to these parameters is avail
able to the user only so that in the event that the controller’s memory fails the
user may restore these values to the factory settings. The user should have a list
of these constants and their settings with the manual.
CAUTION: DO NOT change the values of the bath calibration constants
from the factory set values. The correct setting of these parameters is im
portant to the safety and proper operation of the bath.
The calibration parameters menu is indicated by,
Press “SET” five times to enter the menu.
9.14.1CTO
Parameter CTO sets the limit of the over-temperature cutout. This is not adjustable by software but is adjusted with an internal potentiometer. For the 6020
bath series this parameter should read as follows.
CAL
-
-
-
-
Calibration parameters menu
CTO310°C210°C310°C310°C
9.14.2CO and CG
These parameters calibrate the accuracy of the bath set-point. These are pro
grammed at the factory when the bath is calibrated. Do not alter the value of
these parameters. If the user desires to calibrate the bath for improved accuracy
then calibrate R0 and ALPHA according to the procedure given in Section 11.
9.14.3H and L
These parameters set the upper and lower set-point limits of the bath. DO NOT
change the values of these parameters from the factory set values. To do so may
present danger of the bath overheating and causing damage or fire.
6020602160226024
-
52
Page 61
10 Digital Communication Interface
10Digital Communication Interface
Serial Communications
If supplied with the option, the bath is capable of communicating with and be
ing controlled by other equipment through the digital interface. Two types of
digital interface are available — the RS-232 serial interface and the IEEE-488
GPIB interface.
With a digital interface the bath may be connected to a computer or other
equipment. This allows the user to set the bath temperature, monitor the tem
perature, and access any of the other controller functions, all using remote com
munications equipment. In addition the heater power may be switched from
low to high using the interface. To enable the heater to be switched to high us
ing the interface the “HEATER” switch must be set to the “LOW” position.
10.1Serial Communications
The bath may be installed with an RS-232 serial interface that allows serial digital communications over fairly long distances. With the serial interface the user
may access any of the functions, parameters and settings discussed in Section 9
with the exception of the BAUD rate setting. The serial interface uses 8 data
bits, 1 stop bit, and no parity.
-
-
-
-
53
Page 62
6020 Series Calibration Bath
User’s Guide
10.1.1Wiring
The serial communications ca
ble attaches to the bath through
the DB-9 connector at the back
of the instrument. Figure 9
shows the pin-out of this con
nector and suggested cable wir
ing. To eliminate noise, the
serial cable should be shielded
with low resistance between the
connector (DB-9) and the
shield.
10.1.2Setup
Before operation the serial interface of the bath must first be
set up by programming the
baud rate and other configuration parameters. These parameters are programmed within the
serial interface menu.
To enter the serial parameter
programming mode first press
“EXIT” while pressing “SET”
and release to enter the secondary menu. Press “SET” repeatedly until the display reads
“ProbE” . This is the menu selection. Press “UP” repeatedly
until the serial interface menu
is indicated with “SErIAL”. Finally press “SET” to enter the serial parameter
menu. In the serial interface parameters menu are the BAUD rate, the sample
rate, the duplex mode, and the linefeed parameter.
-
-
-
Figure 9 Serial Communications Cable Wiring
54
10.1.2.1Baud Rate
The baud rate is the first parameter in the menu. The display prompts with the
baud rate parameter by showing “BAUd”. Press “SET” to choose to set the
baud rate. The current baud rate value is displayed. The baud rate of the bath
serial communications may be programmed to 300,600,1200, or 2400 baud.
The baud rate is pre-programmed to 1200 baud. Use “UP” or “DOWN” to
change the baud rate value. Press “SET” to set the baud rate to the new value or
“EXIT” to abort the operation and skip to the next parameter in the menu.
10.1.2.2Sample Period
The sample period is the next parameter in the menu and prompted with “SAm
-
Page 63
PLE”. The sample period is the time period in seconds between temperature
measurements transmitted from the serial interface. If the sample rate is set to
5, the bath transmits the current measurement over the serial interface approxi
mately every five seconds. The automatic sampling is disabled with a sample
period of 0. Press “SET” to choose to set the sample period. Adjust the period
with “UP” or “DOWN” and then use “SET” to set the sample rate to the dis
played value.
10.1.2.3Duplex Mode
The next parameter is the duplex mode indicated with “dUPL”. The duplex
mode may be set to half duplex (“HALF”) or full duplex (“FULL” ). With full
duplex any commands received by the thermometer via the serial interface are
immediately echoed or transmitted back to the device of origin. With half du
plex the commands are executed but not echoed. The default setting is full du
plex. The mode may be changed using “UP” or “DOWN” and pressing “SET”.
10.1.2.4Linefeed
10 Digital Communication Interface
IEEE-488 Communication (optional)
-
-
-
-
The final parameter in the serial interface menu is the linefeed mode. This parameter enables (“On”)ordisables(“OFF”) transmission of a linefeed charac-
ter (LF, ASCII 10) after transmission of any carriage-return. The default setting
is with linefeed on. The mode may be changed using “UP” or “DOWN” and
pressing “SET”.
10.1.3Serial Operation
Once the cable has been attached and the interface set up properly the controller immediately begins transmitting temperature readings at the programmed
rate. The set-point and other commands may be sent to the bath via the serial
interface to set the bath and view or program the various parameters. The inter
face commands are discussed in Section 10.3. All commands are ASCII charac
ter strings terminated with a carriage-return character (CR, ASCII 13).
10.2IEEE-488 Communication (optional)
The IEEE-488 interface is available as an option. Baths supplied with this op
tion may be connected to a GPIB type communication bus which allows many
instruments to be connected and controlled simultaneously. To eliminate noise,
the GPIB cable should be shielded.
-
-
-
10.2.1Setup and Address Selection
To use the IEEE-488 interface first connect an IEEE-488 standard cable to the
back of the bath.
Next set the device address. This parameter is programmed within the
IEEE-488 interface menu.
To enter the IEEE-488 parameter programming menu first press “EXIT” while
55
Page 64
6020 Series Calibration Bath
User’s Guide
pressing “SET” and release to enter the secondary menu. Press “SET” repeat
edly until the display reaches “PrObE”. This is the menu selection. Press “UP”
repeatedly until the IEEE-488 interface menu is indicated with “IEEE”.Press
“SET” to enter the IEEE-488 parameter menu. The IEEE-488 menu contains
the IEEE-488 address parameter.
The IEEE-488 address is prompted with “AddrESS” . Press “SET” to program
the address. The default address is 22. Change the device address of the bath if
necessary to match the address used by the communication equipment by
pressing “UP” or “DOWN” and then “SET”.
10.2.2IEEE-488 Operation
Commands may now be sent via the IEEE-488 interface to read or set the tem
perature or access other controller functions. All commands are ASCII charac
ter strings and are terminated with a carriage-return (CR, ASCII 13). Interface
commands are listed below.
10.3Interface Commands
The various commands for accessing the bath controller functions via the digital interfaces are listed in this section (see Table 5). These commands are used
with both the RS-232 serial interface and the IEEE-488 GPIB interface. In either case the commands are terminated with a carriage-return character. The interface makes no distinction between upper and lower case letters, hence either
may be used. Commands may be abbreviated to the minimum number of letters
which determines a unique command. A command may be used to either set a
parameter or display a parameter depending on whether or not a value is sent
with the command following a “=” character. For example “s”<CR> returns the
current set-point and “s=50.00”<CR> sets the set-point to 50.00 degrees.
-
-
-
56
In the following list of commands, characters or data within brackets, “[” and
“]”, are optional for the command. A slash, “/”, denotes alternate characters or
data. Numeric data, denoted by “n”, may be entered in decimal or exponential
notation. Characters are shown in lower case although upper case may be used.
Spaces may be added within command strings and will simply be ignored.
Backspace (BS, ASCII 8) may be used to erase the previous character. A termi
nating CR is implied with all commands.
-
Page 65
10.4Power Control Functions
The digital interface is capable of controlling the heating functions so that the
bath can be remotely operated at any temperature within the range of the bath.
To allow the interface to control the heating, the front panel controls are dis
abled by switching the heater switch to “LOW”. Otherwise, the interface can
not switch the heater functions off. The 6020 series baths have one control
function with the digital interface. This control function is for the heater
settings.
To control the heater with the digital interface, the front panel heater switch
must be set to “LOW”. The heater function is controlled with the “F1" com
mand. This command is either set to ”0" or “1" according to Table 4. The de
fault setting is ”F1=0" which is “LOW”. Sending a command without a
parameter returns the state, “1" for on or ”0" for off.
Table 4 Power Control Functions
10 Digital Communication Interface
Power Control Functions
-
-
-
PowerF1
LOW0
HIGH1
57
Page 66
6020 Series Calibration Bath
User’s Guide
Table 5 Interface Command Summary
Command
Command Description
Display Temperature
Read current set-points[etpoint]sset: 9999.99 {C or F}set: 150.00 C
Set current set-point to ns[etpoint]=ns=450Instrument
Read vernierv[ernier]vv: 9.99999v: 0.00000
Set vernier to nv[ernier]=nv=.00001Depends on
Read temperaturet[emperature]tt: 9999.99 {C or F}t: 55.69 C
Read temperature unitsu[nits]uu: xu: c
Set temperature units:u[nits]=c/f
Set temperature units to Celsiusu[nits]=cu=c
Set temperature units to
Fahrenheit
Secondary Menu
Read proportional band settingpr[op-band]prpb: 999.9pb: 15.9
Set proportional band to
n
Format
u[nits]=fu=f
pr[op-band]=npr=8.83Depends on
Command
ExampleReturned
Returned
Example
Acceptable
Values
Range
Configuration
CorF
Configuration
Read cutout settingc[utout]cc: 9999 {x},{xxx}c: 620 C, in
Read high set-point limit value*th[igh]*thth: 999th: 205
Set high set-point limit to
Miscellaneous (not on menus)
Read firmware version number*ver[sion]*verver.9999,9.99ver.2100,3.56
Read structure of all commandsh[elp]hlist of commands
Read Heaterf1f1f1:9f1:1
Set Heaterf1=1/0 (1=high, 0=low)0or1
Set heater to lowf1=
Set heater to highf1=
Legend:[] Optional Command data
Note:When DUPLEX is set to FULL and a command is sent to READ, the command is returned followed by a
n
*cg=n*cg-156.25–999.9 to 999.9
n
n
*tl[ow]=n*tl=-80–999.9 to 999.9
*th[igh]=n*th=205–999.9 to 999.9
n
n
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
carriage return and linefeed. Then the value is returned as indicated in the RETURNED column.
f1=0
f1=1
59
Page 68
Page 69
11Calibration Procedure
11 Calibration Procedure
Calibration Points
In some instances the user may want to calibrate the bath to improve the tem
perature set-point accuracy. Calibration is done by adjusting the controller
probe calibration constants R0 and ALPHA so that the temperature of the bath
as measured with a standard thermometer agrees more closely with the bath
set-point. The thermometer used must be able to measure the bath fluid temper
ature with higher accuracy than the desired accuracy of the bath. By using a
good thermometer and carefully following procedure the bath can be calibrated
to an accuracy of better than 0.02°C over a range of 100 degrees.
11.1Calibration Points
In calibrating the bath R0 and ALPHA are adjusted to minimize the set-point
error at each of two different bath temperatures. Any two reasonably separated
bath temperatures may be used for the calibration however best results will be
obtained when using bath temperatures which are just within the most useful
operating range of the bath. The farther apart the calibration temperatures the
larger will be the calibrated temperature range but the calibration error will also
be greater over the range. If for instance 50°C and 150°C are chosen as the cali-
bration temperatures then the bath may achieve an accuracy of say ±0.03°C
over the range 40 to 160°C. Choosing 80°C and 120°C may allow the bath to
have a better accuracy of maybe ±0.01°C over the range 75 to 125°C but outside that range the accuracy may be only ±0.05°C.
-
-
11.2Measuring the Set-point Error
The first step in the calibration procedure is to measure the temperature errors
(including sign) at the two calibration temperatures. First set the bath to the
lower set-point, t
. Wait for the bath to reach the set-point and allow 15 minutes
L
to stabilize at that temperature. Check the bath stability with the thermometer.
When both the bath and the thermometer have stabilized, measure the bath tem
perature with the thermometer and compute the temperature error err
the actual bath temperature minus the set-point temperature. For example, if the
bath is set for a lower set-point of t
=50°C and the bath reaches a measured
L
temperature of 49.7°C then the error is –0.3°C.
Next, set the bath for the upper set-point, t
bath temperature and compute the error, err
, and after stabilizing measure the
H
. For our example, suppose the
H
bath was set for 150°C and the thermometer measured 150.1°C giving an error
of +0.1°C.
11.3Computing R0 and ALPHA
Before computing the new values for R0 and ALPHA the current values must
be known. The values may be found by either accessing the probe calibration
menu from the controller panel or by inquiring through the digital interface.
which is
L
-
61
Page 70
6020 Series Calibration Bath
User’s Guide
The user should keep a record of these values in case they may need to be re
-
stored in the future. The new values R0′ and ALPHA′ are computed by entering
the old values for R0 and ALPHA, the calibration temperature set-points t
t
, and the temperature errors errLand errHinto the following equations,
H
⎡
R
err terr t
00
⎢
⎣
ALPHA
−
HLLH
tt
−
HL
⎡
()()11
ALPHA t errALPHA t err
+−+
′=
⎢
ALPHAR
HLLH
⎣
⎤
+
1′=
⎥
⎦
tt
−
HL
⎤
1⎥ALPHA
+
⎦
L
and
For example, if R0 and ALPHA were previously set for 100.000 and 0.0038500
respectively and the data for t
,errL,anderrHwere as given above then the
L,tH
new values R0′ and ALPHA′ would be computed as 100.193 and 0.0038272 respectively. Program the new values R0 and ALPHA into the controller. Check
the calibration by setting the temperature to t
and tHand measuring the errors
L
again. If desired the calibration procedure may be repeated again to further improve the accuracy.
11.4Calibration Example
The bath is to be used between 75 and 125°C and it is desired to calibrate the
bath as accurately as possible for operation within this range. The current values for R0 and ALPHA are 100.000 and 0.0038500 respectively. The calibration points are chosen to be 80.00 and 120.00°C. The measured bath
temperatures are 79.843 and 119.914°C respectively. Refer to Figure 10 for applying equations to the example data and computing the new probe constants.
62
Page 71
R0 = 100.000
ALPHA = 0.0038500
t
= 80.00°C
L
measured t = 79.843°C
t
= 120.00°C
H
measured t = 119.914°C
Compute errors,
err
= 79.843 - 80.00°C = -0.157°C
L
err
= 119.914 - 120.00°C = -0.086°C
H
11 Calibration Procedure
Calibration Example
Compute R0,
⎡
0 08680 00157120 0
−−−
R0
(. ). (. ).
⎢
⎣
xx
120 080 0
−
..
Compute ALPHA,
⎡
ALPHA′=
+−−+(..)(.)(..)(10 00385 120 0015710 0038580 0xx
⎢
⎣
Figure 10 Calibration Example
0 00385′=
.
120 080 0
..
−
⎤
+
1 100 000100115
⎥
=..
⎦
0 086
.)
−
⎤
1 0 003850 0038387
..
+
⎥
=
⎦
63
Page 72
Page 73
12Maintenance
12 Maintenance
The calibration instrument has been designed with the utmost care. Ease of op
eration and simplicity of maintenance have been a central theme in the product
development. Therefore, with proper care the instrument should require very
little maintenance. Avoid operating the instrument in dirty or dusty
environments.
A battery is used to maintain operating parameters in the unit. All operat
•
ing parameters, including calibration parameters should be checked on a
regular basis to insure accuracy and proper operation of the instrument.
See the troubleshooting section for the procedure on checking the status
of the battery.
The bath should be cleaned regularly to prevent a buildup of oil or dust.
•
Use a paint safe cleaning agent on all painted surfaces. Solvents such as
Trichloroethylene or Acetone may dull or dissolve the paint. The stainless
steel surfaces may be cleaned with solvents as necessary to remove oils.
• The stirring motor should be clean to allow proper stirring. Normally only
the outside surfaces require any attention. If the inside of the motor has
become heavily loaded with oily dust, blow it out with compressed air.
Follow normal safety procedures when using pressurized gasses.
• Periodically check the fluid level in the bath to ensure that the level has
not dropped. A drop in the fluid level affects the stability of the bath.
Changes in fluid level are dependent upon several factors specific to the
environment in which the equipment is used. A schedule cannot be outlined to meet each environmental setting. Therefore, the first year the bath
should be checked weekly with notes kept as to changes in bath fluid. After the first year, the user can set up a maintenance schedule based on the
data specific to the application.
-
-
• Heat transfer medium lifetime is dependent upon the type of medium and
the environment. The fluid should be checked at least every two weeks for
the first year and regularly thereafter. This fluid check provides a baseline
for knowledge of bath operation with clean, usable fluid. Once some flu
ids have become compromised, the break down can occur rapidly. Partic
ular attention should be paid to the viscosity of the fluid. A significant
change in the viscosity can indicate that the fluid is contaminated, being
used outside of its temperature limits, contains ice particles, or is close to
a chemical breakdown. Once data has been gathered, a specific mainte
nance schedule can be outline for the instrument. Refer to the General
Operation section (Section 8) for more information about the different
types of fluids used in calibration baths.
•
If a hazardous material is spilt on or inside the equipment, the user is re
sponsible for taking the appropriate decontamination steps as outlined by
the national safety council with respect to the material. MSDS sheets ap
plicable to all fluids used in the baths should be kept in close proximity to
the instrument.
-
-
-
-
-
65
Page 74
6020 Series Calibration Bath
User’s Guide
If the mains supply cord becomes damaged, replace it with a cord with
•
the appropriate gauge wire for the current of the bath. If there are any
questions, call an Authorized Service Center for more information.
Before using any cleaning or decontamination method except those rec
•
ommended by Hart, users should check with an Authorized Service Cen
ter to be sure that the proposed method will not damage the equipment.
If the instrument is used in a manner not in accordance with the equip
•
ment design, the operation of the bath may be impaired or safety hazards
may arise.
The over-temperature cutout should be checked every 6 months to see that
•
it is working properly. In order to check the user selected cutout, follow
the controller directions (Section 8.2) for setting the cutout. Both the man
ual and the auto reset option of the cutout should be checked. Set the bath
temperature higher than the cutout. Check to see if the display flashes
cutout and the temperature is decreasing. Note: When checking the
over-temperature cutout, be sure that the temperature limits of the bath
fluid are not exceeded. Exceeding the temperature limits of the bath fluid
could cause harm to the operator, lab, and instrument.
-
-
-
-
• The constant temperature bath depends upon the certain qualities of the
fluid medium in order to maintain a uniform and stable temperature environment. Some oils change their characteristics or become dirty after a
period of use. Always remove any foreign materials from the bath.
• Silicone oils as well as others may evaporate off their lighter components
over a period of time leaving the very viscous components remaining. In
addition, some decomposition occurs which may impair the temperature
stability of the bath. When this happens, the fluid is generally very dark to
black and viscous. Vegetable oils polymerize (turn plastic like) after they
have been used for a time at high temperatures making them very difficult
to remove. When the oil has become unusable it should be changed. The
instructions for removing the oil is as follows.
12.1Draining the Bath
The drain is located on the back of the bath. See Figure 5 on page 26. Locate
the drain plug on the end of the drain tube. The following information is help
ful when draining the bath. Always use a container capable of holding the en
tire load of fluid. Using an adequate size (about 8 gallons or 1 cubic foot), heat
proof fluid container is extremely important. Use safety equipment as
appropriate.
-
-
66
1.Water and low viscosity fluids - Drain at room temperature. Normal care
must be taken for fluids that may have corrosive or damaging effects on
the surrounding facility or equipment.
2.High viscosity oils - The fluid should be sufficiently low in viscosity to
drain efficiently. Some oils such as 710 silicone oil may need to be
heated to 80 C to be fluid enough to drain well. The viscosity affects
Page 75
12 Maintenance
Draining the Bath
how rapidly it drains as well as how well it flows off of the walls. Heat
proof containers and appropriate safety equipment such as face shields,
gloves and body covering should be used.
WARNING: Extreme danger of BURNS and FIRE. Use safety equipment.
Use proper equipment, and have fire safety equipment standing by.
67
Page 76
Page 77
13Troubleshooting
13.1Troubleshooting
In the event the bath appears to function abnormally this section may help to
find and solve the problem. Several possible problem conditions are described
along with likely causes and solutions. If a problem arises please read this sec
tion carefully and attempt to understand and solve the problem. If the bath
seems faulty or the problem cannot otherwise be solved, then contact an Autho
rized Service Center (see Section 1.3) for assistance. Opening the unit without
contacting an Authorized Service Center may void the warranty.
ProblemCauses and Solutions
The heater indicator LED
stays red but the temper
ature does not increase
If the display does not show “cutout” and shows the correct bath temperature,
consider the following possibilities:
-
Insufficient heating. Insufficient heating may be caused by the heater power
setting being too low, especially at higher operating temperatures. Switching to
the higher heater power switch setting, if available, may solve the problem.
13 Troubleshooting
Troubleshooting
-
-
The controller display
flashes “Cutout” and the
heater does not operate
No heating. This is caused by blown heater fuses and/or burned out heaters.
Always replace the fuse with one of the same rating, voltage, and type. Never
replace the fuse with one of a higher current rating. Check the heater fuses to
make sure that they are still good. Access the heater fuses by removing the
back panel covering the display electronics. If they are blown, and continue to
blow when replaced, the heaters may be shorted. If you suspect that the heaters are shorted or burned out, contact an Authorized Service Center (see Section 1.3) for assistance.
If the display flashes “Cutout” alternately with the correct process temperature,
check the following:
Wrong cutout setting. The cutout disconnects power to the heaters when the
bath temperature exceeds the cutout set-point. This causes the bath tempera
ture to drop back down to a safe value. If the cutout mode is set to “AUTO”, the
heater switches back on when the temperature drops. If the mode is set to “RE
SET”, the heater only comes on again when the temperature is reduced and the
cutout is manually reset by the operator. (Refer to section 9.2.)
Check that the cutout set-point is adjusted to 10 or 20°C above the desired
maximum bath operating temperature and that the cutout mode is set as
desired.
Bad cutout. If the cutout activates when the bath temperature is well below the
cutout set-point or the cutout does not reset when the bath temperature drops
and it is manually reset, the cutout circuitry may be faulty. Try performing the
Factory Reset Sequence explained below.
-
-
Factory Reset Sequence - Hold the “SET” and “EXIT” keys down at the same
time while powering up the unit. The display shows “init”, the model number,
and the firmware version. Each of the controller parameters and calibration con
stants must be re-programmed. The values can be found on the Report of Cali
bration that was shipped with the instrument.
69
-
-
Page 78
6020 Series Calibration Bath
User’s Guide
ProblemCauses and Solutions
The display flashes
“Cutout” alternately with
an incorrect process
temperature
Low battery. A problem could exist with the memory back-up battery. If the bat
tery voltage is insufficient to maintain the memory, data may become scrambled
causing problems. A nearby large static discharge may also affect data in mem
ory. Access the battery by removing the back panel covering the display
electronics.
-
-
The controller displays
the wrong temperature
and the bath continually
heats or cools regardless
of the set-point value
The controller controls or
attempts to control at an
inaccurate temperature
Corrupt controller memory. If the problem reoccurs after the battery is re
placed, initialize the memory by performing a Factory Reset Sequence (de
scribed in a previous solution).
Bad control probe. The bath control probe may be disconnected, burned out,
or shorted. Check first that the probe is connected properly to the socket in the
rear of the bath labeled “PROBE”.
The probe may be checked with an ohmmeter to see if it is open or shorted.
The probe is a platinum 4-wire Din 43760 type. The resistance should read 0.2
to 2.0 ohms between pins 1 and 2 on the probe connector and 0.2 to 2.0 ohms
between pins 3 and 4. The resistance should read from 100 to 300 ohms be
tween pins 1 and 4 depending on its current temperature.
Corrupt controller memory. Initialize the memory by performing a Factory Re
set Sequence (described in a previous solution).
If the controller appears to operate normally except that the bath’s temperature
does not agree with the temperature measured by the user’s reference thermometer to within the specified accuracy, consider the following:
Erroneous parameters. Check that the calibration parameters are all correct
according to the Report of Calibration. If not, reprogram the constants. If the
controller does not keep the correct parameters, the memory backup battery
may be weak causing errors in data. See “Low Battery” in a previous solution.
Poor uniformity. There may be an actual difference between the bath’s control
probe and the reference thermometer due to excess gradients in the bath.
Check that the bath has an adequate amount of fluid in the tank and that the
stirrer is operating properly. Also check that the reference thermometer and
control probe are both fully inserted into the bath to minimize temperature gradi
ent errors.
-
-
-
-
-
70
Bad control probe. Check that the control probe has not been struck, bent, or
damaged. Refer to the previous solution for how to check the probe’s
resistance.
Page 79
ProblemCauses and Solutions
The controller shows that
it is controlling at the
proper temperature, but
the bath temperature is
unstable
If the bath does not achieve the expected degree of temperature stability when
measured using a thermometer, consider the following:
Wrong proportional band setting. If the proportional band is set too narrow,
the bath will oscillate causing poor stability. In this case, increase the width of
the proportional band.
If the proportional band setting is too wide, the long-term stability of the bath is
affected. In this case decrease the width of the band. (Refer to Section 9.7.)
13 Troubleshooting
Troubleshooting
Bath fluid is too thick. Make sure that the bath fluid used is less than 50 centi
Stokes (10 is ideal) at the temperature at which the bath is controlling. Check
the fluid manufacturer’s specifications.
You should also change the bath fluid regularly and if it changes colors or be
comes too thick.
Bad control probe. Check that the control probe has not been struck, bent, or
damaged. Refer to the previous solution for how to check the probe’s
resistance.
The controller alternately
heats for a while then
cools
The controller does not
maintain controller parameters or parameters
are reset each time the
power to the unit is
removed
Power UpThe instrument is equipped with operator accessible fuses. If a fuse blows, it
Wrong proportional band setting. If the proportional band is set too narrow,
the bath will oscillate between too much heating and too much cooling causing
instability. Increase the width of the proportional band until the temperature stabilizes. (Refer to section 9.7.)
Note: Before performing the memory check, you need to record the controller
calibration parameters (found in the CAL menu of the instrument) and any
user-adjusted parameters that you have changed (such as the programmable
set points and proportional band).
Memory Check
Doing a memory check is the easiest way to verify the ability of the battery to
maintain controller parameters.
1. Power off the instrument.
2. Disconnect the instrument from AC power for 10 seconds.
3. Reconnect the AC power and power on the instrument.
4. If the display shows "InIT" and/or the cycle count shows a low number such
as 0002, the battery is spent and should be replaced. Contact an Authorized
Service Center for assistance.
5. After replacing the battery, you must reprogram the calibration and user-ad
justable parameters into the controller.
may be due to a power surge or failure of a component. Replace the fuse once.
Always replace the fuse with one of the same rating, voltage, and type. Never
replace the fuse with one of a higher current rating. If the fuse blows a second
time, it is likely caused by failure of a component part. If this occurs, contact an
Authorized Hart Scientific Service Center.
-
-
-
71
Page 80
Page 81
14Appendix – Model 6021
The Hart Scientific 6021 is a heated bath adapted for use with a chiller. A
switch on the back of the controller permits the use of the chiller in an auto
mated or manual mode.
14.1Installation
14 Appendix – Model 6021
Installation
-
The information contained here is in addition to that in the Section 5 (Installa
tion ) of this manual.
The bath and chiller must be level and operating in an appropriate environment
prior to installation. Refer to the chiller manual for details on chiller installa
tion, filling and operation. The general installation of a chiller is as follows.
1.Connect the tubing lines between the chiller and the bath. Quick-discon
nect fittings, if installed on chiller, make attaching tubing lines easy. Be
sure the lines fit securely on the chiller so they do not disconnect during
operation. Either cooling tube on the 6021 bath may be used as inlet or
outlet.
If the hoses are not connected to the 6021 bath already, connect them securely with hose clamps. Silicone rubber tubing is typically used, although resistant to high temperatures it is not compatible with silicone
oils.
2.If applicable, fill the chiller with the appropriate fluid to between 1 and
1.5 inches below the lid. If unsure on how to fill the chiller, consult the
chiller manual. Water works for most applications. After operating the
chiller for a short period of time, check the fluid level in the chiller. More
fluid may have to be added to the chiller, due to fluid required to fill the
newly attached system. DO NOT use the chiller above it’s maximum rec
ommended temperature. Some chillers automatically cut out when the
maximum operating temperature is exceeded (consult the chiller manual
for more information).
-
-
-
-
WARNING: Water left in the 6021 bath cooling coils will boil as the bath
temperature increases above the boiling point. This process is slow and
safe as long as the lines to the chiller are unobstructed. Otherwise, a dan
gerous health hazard may exist.
3.Connect the power plug for the chiller to the Auxiliary Chiller Outlet on
the rear of the 6021 controller. The bath also needs to be setup and pow
ered on. Use the appropriate mains voltage with an adequate current ca
pacity available. Refer to Section 3.1, Specifications and Environmental
Conditions for all power requirements and specifications.
If the bath is in high heat mode or switched to high heat mode, the chiller
output is automatically disabled. Therefore, high heat mode will override
-
-
-
73
Page 82
6020 Series Calibration Bath
User’s Guide
the chiller operation, but the chiller mode cannot override the high heat
mode. The chiller option can only be used when the bath is in low heat
mode. The bath operates as such, since it is not capable of supplying the
current required to run a chiller and high heat mode simultaneously. This
functionality applies to both manual and automatic control as long as the
chiller is powered by the 6021 bath.
WARNING: Insure that the chiller fluid is compatible with the operating
temperature of the 6021.
4.Set the chiller temperature to at least 10°C below the bath temperature.
To test the cooling system make sure the 6021 controller is on and then
set the Auxiliary Chiller Outlet switch on the back of the 6021 bath to
On (not Auto). When the chiller power switch is turned on, the chiller
power indicator and in some cases the chilled water solenoid will energize. In the case of a chiller, the instrument will engage and begin to
pump fluid. In the case of a chilled water system, the solenoid valve will
open and begin to circulate fluid through the system. The 6021 temperature should begin to decrease and approach the new set-point, or the
duty-cycle of the bath will increase as the 6021 attempts to maintain the
current set-point.
14.2Chiller Parts and Controls
The information contained here is for the chiller parts and controls and is in addition to that in the Section 7 (Parts and Controls) of this manual. The auxiliary
chiller controls for the 6021 bath are located on the rear of the controller panel.
The controls are illustrated in Figure 11 and include: the On/Auto control
switch, the auxiliary power outlet, and the cooling tube connections.
1.The On/Auto control switch allows the system to switch between manual
and computer control. For manual power, move the switch to the On po
sition. For computer control move the switch to the Auto position. A
computer program may be written to permit control of all 6021 functions
as well as turning the power off and on to the chiller from the 6021 aux
iliary chiller outlet.
2.The Auxiliary Chiller Outlet power connector can supply the current
listed in Section 3.1, Specifications. DO NOT connect a higher load.
3.The cooling tube connections permit the attachment of a chiller, com
pressed air, tap water, or other source cooling. The tubes are 3/8 inch
outside diameter (OD) copper. These tubes connect to two cooling plates
attached to the bottom and back walls of the bath. Use 3/8 inch inside di
ameter (ID) tubing to connect to the cooling source.
-
-
-
-
74
Page 83
14 Appendix – Model 6021
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
aaaaaaaaaaaaaaaaaaaaaaaaa
Chiller Controller Operation
(Optional)(Optional)
IEEE-488RS-232PROBESTIRRER
Figure 11 6021 Back panel
1
2
I
O
3
COOLING TUBES
14.3Chiller Controller Operation
Sections 8 (General Operation), 9 (Controller Operation), and 10 (Digital Communications) of this manual explain the general operation, controller operation,
and digital communications for the 6021 bath. The digital interface is also capable of turning the auxiliary chiller on and off as shown in Table 6. The chiller
must be connected as described in the Installation portion of this Appendix and
the auxiliary switch must be turned to the Auto position. The bath heater
position must be in low.
Table 6 Auxiliary Chiller Control Functions
Chiller StateF2
Off0
On1
75
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.