Fluke Biomedical
6920 Seaway Blvd
Everett, WA 98203 USA
775-883-3400
800-648-7952
Trademarks
®
IBM
, PC® and PC/AT® are registered trademarks of International Business Machines
Corporation.
Microsoft
®
and MS-DOS® are registered trademarks of Microsoft Corporation.
Restrictions and Liabilities
Information in this document is subject to change and does not represent a commitment by
Fluke Biomedical. Changes made to the information in this document will be incorporated in
new editions of the publication.
No responsibility is assumed by Fluke Biomedical Corporation for the use or reliability of
software or equipment that is not supplied by Fluke Biomedical Corporation or its affiliated
dealers.
iii
601PRO SERIES
Warnings
XL
Use of this instrument is restricted to qualified personnel who recognize shock hazards
and are familiar with safety precautions used when operating electrical equipment. Read
the manual carefully before operating the 601PRO.
☛ The following warning and informational symbols can be found on the 601PRO:
Symbol Description
Caution: Risk of electric shock
Direct / Alternating Current
Protective Earth (PE)
Caution: Refer to accompanying
documentation
Off (power: disconnection from Mains)
On (Power: connection to Mains)
Equipotential/Functional Earth (FE)
iv
NOTICES, WARNINGS AND CONTENT
☛
Exercise extreme caution when a shock hazard is present at the instrument's
measurement terminals during the following tests:
• Mains on Applied Part
• Mains on Applied Part Calibration
• Protective Earth Resistance
• Protective Earth Resistance Calibration
• Equivalent Patient Leakage
• Equivalent Device Leakage
• Equivalent Device/Patient Leakage Calibration
FOR CORRECT OPERATION, ALL GROUND-REFERENCED PERIPHERALS, SUCH AS
PRINTERS AND PCs, MUST BE DISCONNECTED.
Do not discharge a defibrillator while it is plugged into the 601PRO.
☛
☛ Only use Fluke Biomedical-supplied test leads or leads rated for 32 Amps/1000
Volts with the Protective Earth Resistance Test.
☛ Inspect the lead ends for possible wear, cracks or breaks before each use.
☛ Take leakage current measurements only after earth resistance is measured and
found to be compliant with the applied safety limit.
☛ External devices, such as printers and computers, attached to the 601PRO, may
affect the 601PRO's ability to sense Open Earth conditions on the Mains input. If
Mains voltage readings are in error, remove all external devices.
☛ If the DUT fails the Earth Resistance test, the operator must discontinue testing and
label the DUT defective.
☛ If any single test fails, the test must be immediately discontinued and the DUT
labeled defective.
☛ Prior to performing an ECG Simulation test, perform an Applied Part Leakage Test.
If the Applied Part Leakage Test yields an instrument-under-test failure, then do not
perform an ECG Simulation Test, as damage to the instrument may occur.
☛ If operating the 601PRO with a variable AC Supply (Variac), it is important to
perform a calibration after changing the Mains Voltage level. Calibration can be
performed from within the Mains On Applied Part, Equivalent Device Leakage, or
Equivalent Patient Leakage tests.
v
601PRO SERIES
Nomenclature
Comparable Terminology: International and United States
INTERNATIONAL/IEC U.S./AAMI
L1 Hot
L2 Neutral
Earth Ground
Mains Line Voltage
Applied Parts Patient Leads
Enclosure/Case Chassis
Protective Earth Ground Wire
Earth Leakage Current Leakage in Ground Wire
Enclosure Leakage Chassis Leakage
XL
Patient Leakage Lead Leakage
Patient Auxiliary Leakage between Patient Leads
Mains on Applied Parts Lead Isolation
Insulation Resistance Dielectric Strength or
Insulation Resistance between
Hot and Neutral to Ground
Earth Resistance Ground Wire Resistance
vi
NOTICES, WARNINGS AND CONTENT
Hazard Warnings
☛ Warning! Power Rating. The 601PRO's mains power input must be connected to a
power receptacle that provides voltage within the specified rating for the system.
Connection must be made via the Mains Power cord provided by Fluke Biomedical.
Use of an incompatible power receptacle or incorrect Mains Power cord may produce
electrical shock and fire hazards. Acceptable Mains Voltage ranges are 90~VAC to
132~VAC, and 180~VAC to 240~VAC 50/60 Hz. The current ratings for the 601PRO are
as follows:
Europe: <=15A (Fused by 15A circuit breaker)
United Kingdom: <=13A (Fused by 13A fused Mains Power Cord)
Australia: <=10A (Fused by 10A circuit breaker)
☛ Warning! Internal Voltage. Always turn off the power switch and unplug the power
cord before cleaning the 601PRO's outer surface.
☛ Warning! Liquids. Avoid spilling liquids on the analyzer; fluid seepage into internal
components creates a potential shock hazard. Do not operate the instrument if internal
components are exposed to fluid.
Precautions
The following precautions are provided to help you avoid damaging the system:
☛ To power up the 601PRO, place the index finger on the rocker switch and use a
☛ Caution:Service. The 601PRO should be serviced by authorized service personnel.
☛ Caution: Environmental Conditions. Do not expose the system to temperature
☛ Caution: Do Not Immerse. Clean only with a mild detergent, and wipe down
rolling motion to push from "OFF" to "ON." Do NOT forcefully push or snap the
rocker switch. This may cause the unit to shut off.
Only qualified technical personnel should perform troubleshooting and service
procedures on internal components.
extremes. Ambient temperatures should remain between 18-40°C. System
performance may be adversely affected if temperatures fluctuate above or below
this range.
with a gentle cloth.
vii
601PRO SERIES
XL
Electromagnetic Interference and Susceptibility
☛ USA FCC CLASS A
Warning: Changes or modifications to this unit not expressly approved by the
manufacturer could void the user's authority to operate the equipment.
This equipment has been tested and found to comply with the limits for a Class A digital
device, pursuant to Part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful interference
when the equipment is operated in a commercial environment. Like all similar equipment,
this equipment generates, uses, and can radiate radio frequency energy and, if not installed
and used in accordance with the instruction manual, may cause harmful interference to
radio communications. Operation of this equipment in a residential area is likely to cause
interference, in which case the user will be required to correct the interference at his own
expense.
☛ Canadian Department of Communications Class A
User Safety
This digital apparatus does not exceed Class A limits for radio emissions from digital
apparatus set out in the Radio Interference Regulations of the Canadian Department of
Communications.
Le present appareil numerique n'met pas du bruits radioelectriques depassant les limites
applicables aux appareils numerique de la Class A prescrites dans le Reglement sur le
brouillage radioelectrique edicte par le ministere des Communications du Canada.
This device has been type tested by an independent laboratory and found to meet the
requirements of the following:
Canadian Standards Association CAN/CSA
C22.2 No.1010.1-1992, “Safety Requirements for Electrical Equipment for
Measurement, Control and Laboratory Use, Part 1: General Requirements”.
UL 3101-1
“Electrical Equipment for Laboratory Use, Part 1: General Requirements”.
viii
NOTICES, WARNINGS AND CONTENT
Based on the testing described below, this instrument bears the CE mark.
The system has been type tested by an independent, accredited testing laboratory and
found to meet the requirements of EN 61326-1:1998 for Radiated Emissions and Line
Conducted Emissions. Verification of compliance was conducted to the limits and
methods of the following:
CISPR 16-1:1993 and CISPR 16-2:1996
☛ Immunity
The system has been type tested by an independent, accredited testing laboratory and
found to meet the requirements of EN 61326-1:1998 for Immunity. Verification of
compliance was conducted to the limits and methods of the following:
EN 61000-4-2 (1991) Electrostatic Discharge
EN 61000-4-3 (1995) Radiated EM Fields
EN 61000-4-4 (1995) Electrical Fast Transient/Burst
EN 61000-4-5 (1995) Surge Immunity
EN 61000-4-6 (1996) Conducted Disturbances
EN 61000-4-11 (1994) Voltage Dips, Short Interruptions and Variations
EC Directive 73/23/EEC Low Voltage (Safety)
The system has been type tested by an independent testing laboratory and found to meet
the requirements of EC Directive 73/23/EEC for Low Voltage. Verification of compliance
was conducted to the limits and methods of the following:
EN 61010-1 (1993) & IEC 1010-1
“Safety Requirements for Electrical Equipment for Measurement, Control and
Laboratory Use, Part 1: General requirements” (including amendments 1 & 2).
ix
601PRO SERIES
Warranty
XL
This Warranty is limited and applies only to new products, except for computerbased software, which is covered under a separate Warranty Policy,
manufactured by Fluke Biomedical. Fluke Biomedical makes no warranty
statement regarding the condition of used products.
Fluke Biomedical warrants the instrument (hereinafter collectively referred to as
“Products” or “Product”) for a period of one (1) year from the original purchase
date against defective materials or workmanship. This Warranty is limited to the
original purchaser (the “Purchaser”) and cannot be assigned or transferred. All
claims under this Limited Warranty must be made in writing to Fluke Biomedical,
Attention: Service Department. Purchaser must ship the Product to Fluke
Biomedical, postage pre-paid. Fluke Biomedical shall either repair or replace with
new or like new, at its option and without cost to the Purchaser, any Product
which in Fluke Biomedical’s sole judgment is defective by reason of defects in the
materials or workmanship.
This Warranty is VOID if the Product has been damaged by accident or misuse, or
has been damaged by abuse or negligence in the operation or maintenance of the
Product, including without limitation unsafe operation, operation by untrained
personnel, and failure to perform routine maintenance. This Warranty is VOID if
the Product has been repaired or altered by persons not authorized by Fluke
Biomedical, or if the Product has had the serial number altered, effaced, or
removed. This Warranty is VOID if any of the Products has not been connected,
installed or adjusted strictly in accordance with written directions furnished by
Fluke Biomedical. Batteries, fuses, light bulbs, and other “consumable” items used
in any of the Products are not covered by this Warranty. Software utilized in
conjunction with any of the Products is not covered by the terms of this Warranty
but may be covered under a separate Fluke Biomedical software warranty.
We will continue to stock parts for a maximum period of five (5) years after the
manufacture of any equipment has been discontinued. Parts shall include all
materials, charts, instructions, diagrams, and accessories that were furnished with
the standard models.
THIS WARRANTY CONTAINS THE ENTIRE OBLIGATION OF FLUKE
BIOMEDICAL, AND NO OTHER WARRANTIES, EXPRESSED, IMPLIED, OR
STATUTORY ARE GIVEN. PURCHASER AGREES TO ASSUME ALL LIABILITY
FOR ANY DAMAGES AND/OR BODILY INJURY OR DEATH THAT MAY
RESULT FROM THE USE OR MISUSE OF ANY EQUIPMENT OR
INSTRUMENT BY THE PURCHASER, HIS EMPLOYEES, AGENTS, OR
CUSTOMERS, OTHER THAN THE EXPRESS WARRANTY CONTAINED
HEREIN. WE SHALL NOT BE RESPONSIBLE FOR ANY DIRECT OR
CONSEQUENTIAL DAMAGES OF ANY KIND. THIS WARRANTY SHALL NOT
BE CHANGED OR MODIFIED IN ANY WAY WITHOUT THE EXPRESS
WRITTEN PERMISSION OF AN OFFICER OF FLUKE BIOMEDICAL.
THIS WARRANTY IS VOID UNLESS THE PURCHASE REGISTRATION CARD
HAS BEEN COMPLETED AND MAILED TO US WITHIN TEN (10) DAYS OF
PURCHASE.
Appendix D: Test Data Record ASCII Character Formats
xiv
Introduction and Description
Chapter
1
1. Introduction to the 601PRO SeriesXL
2. Accessories
3. Optional Accessories
4. Menu Structure
5. System Characteristics
6. Statement of Compatibility
1. Introduction to the 601PRO SeriesXL
The 601PRO SeriesXL (601PRO) is an automated electrical safety analyzer that
meets stringent international standards for electrical safety testing of hospital
and laboratory electromedical equipment.
The 601PRO conducts electrical safety testing in accordance with IEC 601-1,
VDE 751, VDE 701, HEI 95, IEC 1010, AAMI, and AS/NZS 3551
requirements, flags failures, and simulates performance, ECG, and arrhythmia
waveforms. The 601PRO stores 1000 device records. Test results, which are
automatically analyzed and saved in non-volatile memory, can be printed using
the internal ZY column thermal printer or an attached external printer, or
uploaded to a PC using the serial port.
The 601PRO offers automatic, manual, computer control, or step mode
operation.
The 601PRO will accept device information that is input using an external
keyboard, integrated keypad, or barcode keyboard wedge.
1-1
601PRO SERIES
XL
Available electrical safety tests include:
♦ Mains Voltage
♦ Dual Lead Voltage
♦ Dual Lead Leakage
♦ Current Consumption
♦ Insulation Resistance
♦ Protective Earth Resistance
♦ Earth Leakage Current
♦ Enclosure Leakage Current
♦ Patient Leakage Current
♦ Mains on Applied Part Leakage
♦ Patient Auxiliary Current
♦ Accessible Voltage
♦ Accessible Leakage
♦ Equivalent Device Leakage
♦ Equivalent Patient Leakage
Available ECG performance waveforms include:
♦ Square wave: 0.125, 2 Hz
♦ Sine wave: 10, 40, 50, 60, 100 Hz
♦ Triangle wave: 2 Hz
♦ ECG complex: 30, 60, 120, 180, 240 BPM
♦ Pulse: 30, 60 BPM
♦ A-Fib, A-Flutter, A-Tach, Idioventricular,
PVC1, R-on-T, Run, V-Fib, V-Tach
1-2
INTRODUCTION AND DESCRIPTION
2.Accessories
The following accessories are shipped standard with the 601PRO. To order
additional quantities, contact your Fluke Biomedical equipment dealer, and use
the Fluke Biomedical Part Numbers provided.
Description Quantity
Supplied
Probe/Safety Lead, Red 1 2213252
Probe/Safety Lead, Black 1 2213241
Adapter, Banana/Alligator 5 2212941
Operators Manual 1 2234222
Large Clamp, Red 1 2004125
Warranty Card 1 2241856
Printer Paper Roll (original) 1 2248719
Printer Paper Roll (new style) 1 2248743
Part #
3. Optional Accessories
The following optional accessories are available for the 601PROXL. To order,
contact a Fluke Biomedical equipment dealer and use the Fluke Biomedical
part numbers provided.
Description Part #
Carry Case 2234065
RS232 Cable (9M-9F) 2238659
Printer Cable 2238072
Barcode, Keyboard, Wedge 2248255
Adapter, Banana, ECG 2212657
Keyboard English 2213184
Powercord Set Australian 2238603
Powercord Set Schuko 2238015
Powercord Set US 120 V 2286644
Powercord Set UK 2238570
4.Menu Structure
A layout showing various system functions is provided on the following page.
1-3
601PRO SERIES
XL
SETUP
SYSTEM
(CHAPTER 2)
PRINT
(CHAPTER 2)
PRINTDELETE
(CHAPTER 9)
SUMMARY
LEAKAGE
ENCLOSURE
CONFIGURE
ENCLOSURE
LEAKAGE FOR
USE DURING
AUTO MODE
SEQUENCE
TEST
TEST
ENABLE
STANDARDS
EDIT
CUSTOM
STANDARD
MAIN MENU
CLASS/TYPEUTILITIES
TESTS/
AUTOMODES
TEST
DEVICE
SYSTEM
SELECT
ALL VALID
CLASS/TYPES
MANUAL
STEP
RECORDS
OR
RECORDS
TEMPLATES
TEST
FOR THIS
PRESENT
STANDARD
(CHAPTER 3)
(CHAPTER 4)
(CHAPTER 5)
(CHAPTER 6)
CREATE
(CHAPTER 3)
(CHAPTER 3)
SELECT
TEST AND
ADVANCE TO
SELECTED
SCREEN
ENTER
CONTROL
NO. TO
STEP
THROUGH
TRANSMIT
(CHAPTER 4)
RECEIVE
EDITDELETE ALLDELETE
TRANSMIT
CHAPTER 4
EACH TEST
DATE
FORMAT
TIME
FORMAT
BEEPER
RS232
OUTPUT
PRINTER
TEST
STANDARD
SET DATE
TO EITHER
MM-DD-YYORDD-MM-YY
SET CLOCK
TO EITHER
12-HOUR OR
24-HOUR
FORMAT
ACTIVATE OR
DEACTIVATE
BEEPER
CONFIGURE
RS232 SERIAL
PORT FOR:
- EXTERNAL
RS232
SELECT AND
SET UP
PRINTER
INTERFACE
SELECT FROM
- IEC 601-1
- VDE 701-1
- VDE 751-1
- HEI 95
FORMAT
DEVICE
- IEC 1010
- AAMI
- AS/NZS 3551
- CUSTOM
DEV RECS/
TEMPLATES
ON
STOP
FAILURE
SELECT
AUTO/STEP
SEQUENCES
DC
LANGUAGE
CONFIGURE
FOR DEVICE
RECORDS OR
TEMPLATES
AUTOMATICALLY
STOPS AN AUTO
MODE SEQUENCE
WHEN A TEST
FAILS
SELECT
CONTROLLED
POWER
SEQUENCE
OR 601CE
ENABLE DC
READINGS
SELECT
KEYBOARD
OR DISPLAY
LANGUAGE:
CONVENTIONAL
TEST
SEQUENCE
- ENGLISH
- FRENCH
- GERMAN
- ITALIAN
1-4
AUTO
(CHAPTER 4)
ENTER
CONTROL
NO. TO
PERFORM
COMPLETE
TEST
baw007f.eps
INTRODUCTION AND DESCRIPTION
5. System Characteristics
The 601PRO uses a membrane keypad for selection of tests or menu options.
The keys are grouped by color and functionality. The red keys below the
display are used to access menu options. These include the
SOFT KEYS, and the enter key. The black keys allow the operator to
four
gain access to additional functions. These include the
present settings
key, the print header key, and the print data key.
previous key, the
esc/stop key, the view
The red keys numbered 0 through
– are used to enter information and can also
be used to gain quick access to the manual tests.
AUDIO FEEDBACK
• A one-beep signal indicates that a key has been pressed.
• A two-beep-per-second signal indicates high voltage or current generated
by the 601PRO is present.
1-5
601PRO SERIES
XL
A
601 Pro Series
XL
INTERNATIONAL SAFETY ANALYZER
CLASS IPROTECTIVE EARTH
CLASS II
CLASS IP
TYPE BNON-ISOLATED APPLIED PART
TYPE BFISOLATED APPLIED PART
TYPE CF
TYPE F
TYPE T
DOUBLE INSULATION
INTERNAL POWER SUPPLY
ISOLATED APPLIED PART
DIRECT CARDIAC APPLICATION
FIXED DEVICE
TRANSPORTABLE DEVICE
E
F
Figure 1-1: 601PRO Top Panel Illustration
B
GH
PREVIOUS
ESC/STOP
VIEW PRESENT
SETTINGS
PRINT HEADER
PRINT DATA
C
VOLTS-V
PROTECTIVE EARTH
RESISTANCE-
34
PATIENT
LEAKAGE-µA
IEC 1010
ACCESSIBLE-V/µA
9
TESTS
TESTS
CURRENT-A
10
EARTH
LEAKAGE-µA
MAINS ON AP
LEAKAGE-µA
76
VDE EQUIVALENT
DEVICE LEAK-µA
/
INSULATION-M
2
ENCLOSURE
LEAKAGE-µA
5
PATIENT AUXILIARY
CURRENT-µA
8
VDE EQUIVALENT
PATIENT LEAK-µA
ENT
D
Baw152f.eps
Top Panel: Identifying 601PRO Components
Use the drawing of the 601PRO top panel, displayed above, to locate the
following components.
A Printer
Optional 24-character printer for
immediate hard copy of results.
B Previous Key
C
Soft Keys 1-4
Returns user to the previous screen.
Make dynamic assignments based on
current screen.
Enter Key
D
Advances to the next menu or
saves/selects options.
E
Esc/Stop Key
Discontinues current test and returns
operator to the
the DUT outlet.
F
View Present
Settings Key
Advances to View Present Settings
screen when at the
Displays current settings and allows
operator to edit test standard, class/type,
and lead assignments (see page 3-7).
G
Print Header Key
Sends device information fields to
enabled printer.
H
Print Data Key
Sends displayed test data to printer.
MAIN MENU. Turns off
MAIN MENU.
1-6
INTRODUCTION AND DESCRIPTION
Keys Used to Enter Device Control Numbers
The keys described below (0 through –) can be used to enter test control
numbers in the auto/step modes of operation. These keys are also referred to as
test-shortcut keys and can be used to initiate manual tests.
0Volts In single lead mode, displays mains voltage.
In dual lead mode, displays voltage between
RED and BLACK test leads.
1 CurrentMeasures the current consumption (in
amperes) of the device under test.
2
3 Protective Earth
4 Earth Leakage Is measured between the DUT “Protective
5 Enclosure Leakage In single-lead mode, measures the enclosure
6
Patient
7 Mains on
8 Patient Auxiliary
9 IEC 1010
Insulation Tests insulation resistance (mains to case or
applied parts to case).
Measures the earth resistance using a 1A test
Resistance
Leakage
Applied Part
Leakage
Current
Accessible
Voltage/Leakage
current (unless 10A or 25A is selected).
Earth” terminal and the “Protective Earth”
terminal of the 601PRO.
leakage (RED test lead to DUT protective
earth on the 601PRO).
Measures the patient leakage current (applied
part to earth.
Applies 110% mains voltage to selected
applied part and measures leakage to earth in
both normal and reverse polarity. Does not
apply to patient auxiliary selections.
Measures the leakage and biasing current
between applied parts.
Selects the IEC 1010 Test Load. Measures
accessible voltage through the RED jack to
601PRO earth. Allows access to the
accessible leakage test.
/ VDE Equivalent
Device Leakage
– VDE Equivalent
Patient Leakage
Selects the IEC 601 test load. Applies 110%
of mains voltage between L1/L2 and earth.
Selects the IEC 601 test load. Applies 110%
of mains voltage to the selected applied part
and measures the leakage.
1-7
601PRO SERIES
XL
B
A
Front Panel
A Applied Part
B Red Input
C Black Input
D Green Input
E Power Outlet Allows standard power plug connection
C
D
Figure 1-2: 601PRO Front Panel Illustration
The jacks allow direct connection to
Terminal
banana jacks, or 4mm-to-alligator
adapters provided.
Single test lead connection.
Terminal
Used for dual lead testing in combination
Terminal
with red test lead.
Protective Earth of Device Under Test.
Terminal
of the Device Under Test. 120V @ 15A
or 240V @ 15A maximum.
EF
baw154f.eps
1-8
F On-Off Switch /
Circuit Breaker
Power up the 601PRO, I=ON, 0=OFF;
built-in ISA circuit breaker.
Note: To power up the 601PRO, place
the index finger on the rocker switch and
use a rolling motion to push from "OFF"
to "ON." Do NOT forcefully push or
snap the rocker switch. This may cause
the unit to shut off.
INTRODUCTION AND DESCRIPTION
D
Back Panel
A RS232 Connection Allows bi-directional computer control.
B Keyboard Input Allows the use of an external keyboard
A
BC
E
baw155f.eps
Figure 1-3: 601PRO Back Panel Illustration
Serial D-9 female connector.
for data input. DIN 5 socket.
C Printer Connector Allows for an external parallel printer.
D-25 female connector.
D Power Cord
Connection
E Protective Earth
Connection
120V/15A or 240V/15A power cord
connection.
Provides a direct connection to the
power cord ground.
1-9
601PRO SERIES
6.Statement of Compatibility
XL
The 601PRO is compatible with and has the same base features as the previous
601PRO safety analyzer. It also has some new features that include the
following:
• Because of new features in this version of the 601PRO, the Device Record
Format has been changed. If old format device records are transmitted to
the 601PRO Series
via RS232, they will be accepted. However, when
XL
receiving device records from the 601PRO, the new format will always be
used. Refer to Chapter 6 for details.
• Custom standard names are fixed. Any old-style device record received via
RS232 must be updated to reflect this change. Refer to Chapter 9 for more
details.
• Supports all computer control commands from the previous 601PRO.
Some of the commands no longer have meaning in the new 601PRO and
have been given fixed responses. New commands have been added to
support new features on the instrument. Refer to Chapter 10 for complete
details.
• Due to new features, the RS232 result output format has changed. It is very
similar to the old format but has extra fields for the new features. The transfer utility is no longer required or supported. Refer to Chapter 5
for complete details.
• Because the 601PRO now supports IEC 601, IEC 1010, and AAMI test
loads, some changes in computer control commands have been made. All
existing (previous) commands use the 601 test load. New commands have
been created to support the new loads. Refer to Chapter 10 for details.
• Unlike previous versions of the 601PRO, the 601PRO Series
XL
standard D9M-D9F serial cable.
uses a
1-10
Setting Up the 601PRO
Chapter
2
This chapter describes the factory default settings and how to change them,
and provides instructions for one-time customization of the 601PROXL using
SYSTEM SETUP key and the Select Setup Function menu.
the
1. Using Factory Default Settings
2. Selecting the Test Standard
3. Selecting the Printer Output
4. Selecting the RS232 Baud Rate
5. Activating the Beeper
6. Setting the Time and Date
7. Configuring the Enclosure Leakage for the Auto
Mode Sequence
8. Selecting Language Options
9. Selecting the DC Option
10. Selecting the Auto/Step Tests: Controlled Power
Sequences or 601CE Conventional Test
Sequences
11. Enabling Stop on Failure
12. Configuring for Device Records or Templates
1. Using Factory Default Settings
The factory default settings are shown below.
Description Factory Default Setting
Test Standard IEC 601-1 (Class I, BF)
Number of Applied Parts 0 (To change, refer to Chapter 3)
Printer Output Internal
RS232 9600 Baud
Beeper On
Time Format 24 hour
Date Format MDY
Enclosure Leakage (multiple readings
during the Auto Mode sequence)
Keyboard Language English
Display/Print Language English
DC Readings Off
Auto/Step Sequences: Controlled Power
Sequence
Stop on Failure Off
DEV RECS / TEMPLATES DEVICE REC
Off
On
2-1
601PRO SERIESXL
2. Selecting the Test Standard
The Select Test Standard option allows the operator to choose among
IEC 601-1, VDE 701-1, VDE 751-1, HEI 95, IEC 1010, AAMI,
AS/NZS 3551, or custom standards.
Refer to Chapter 8, Standards and Principles, for a detailed discussion of IEC
601-1, VDE 701-1, VDE 751-1, HEI 95, IEC 1010, AAMI, and AS/NZS 3551
standards and their selection.
• At the 601PRO
Setup Function
• At the
Select Test Standard menu, select a test standard, or press *MORE
for additional standards.
MAIN MENU, press SYSTEM SETUP to access the Select
menu. Press TEST STANDARD:
baw008f.eps
2-2
☛ Note: To enable/disable the test standards that appear at the
Standard
*MORE
Enable Test Standard menu.
menu, press UTILITIES at the MAIN MENU, then
, then ENABLE STANDARDS. Press YES or NO at the
Once the selection is made, the 601PRO will automatically return
MAIN MENU where the selected standard is displayed.
to the
Default Setting: IEC 601-1
baw009f.eps
Select Test
SETTING UP THE 601PRO
3.Selecting the Printer Output
Before testing, choose whether to use the optional internal printer, an external
printer, or no printer.
• From the 601PRO
Select Setup Function menu. Press PRINTER OUTPUT:
• At the
Printer Output menu, use the SOFT KEYS to select a printer
interface:
MAIN MENU, press SYSTEM SETUP to access the
baw026f.eps
baw010f.eps
⇒ The EXTERNALASCII ONLY setting is for older (and some newer)
printers that will interpret and print basic ASCII characters.
⇒ The EXTERNALHP PCL3 printer setting is for any HP printer that
supports PCL3 (Printer Control Language, Version 3). The control
sequence feature configures the printer and then downloads ASCII
characters to it.
☛Note:Follow the printer manufacturer’s instructions for handling page
ejects and printing using form-feed commands.
• Press
Default Setting: Internal
previous or enter to save the settings and return to the MAIN MENU.
2-3
601PRO SERIESXL
4.Selecting the RS232 Baud Rate
To select the baud rate for the RS232 serial port:
• From the 601PRO
Select Setup Function menu. Press RS232:
At the 601PRO
RS232 Settings menu, press a SOFT KEY to select a baud rate:
MAIN MENU, press SYSTEM SETUP to access the
baw027f.eps
Default Setting: 9600
• Press
enter to save the setting and return to the Select Setup Function
menu. Press
previous to return to the MAIN MENU.
baw011f.eps
2-4
SETTING UP THE 601PRO
5.Activating the Beeper
To activate/deactivate the beeper:
• From the 601PRO
Select Setup Function menu. Press *MORE until the following menu is
displayed, then press
MAIN MENU, press SYSTEM SETUP to access the
BEEPER:
• Press a
SOFT KEY to select ON or OFF at the Key and Error Beeper
menu:
baw012f.eps
baw025f.eps
Selecting
an error occurs. Selecting
OFF will silence the beep that sounds when a key is pressed or when
OFF will not disable the beeper during lead
calibration or when performing tests using the beeper as a warning.
Default Setting: ON
• Press enter to save the settings and return to the Select Setup Function
menu. Press
previous to return to the MAIN MENU.
2-5
601PRO SERIESXL
6.Setting the Time and Date
The time and date are used to record testing times and their associated dates.
The clock can be set in either 12-hour or 24-hour format. The date format is
either MM-DD-YY or DD-MM-YY.
To set the time format:
• From the 601PRO
Select Setup Function menu. Press *MORE until the following menu is
displayed, then press
MAIN MENU, press SYSTEM SETUP to access the
TIME FORMAT:
To set the time:
• At the following menu, press a
24 HOUR format. Press the cursor SOFT KEY to change cursor position
and enter the time of day using the
SOFT KEY to select either the 12 HOUR or
test-shortcut keys.
baw028f.eps
2-6
Default Setting: 24 HOUR
• Press
enter to save the settings and return to the Select Setup Function
menu. Press
previous to return to the MAIN MENU.
baw013f.eps
Loading...
+ 156 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.