Fluke 5720A II, 5700A II User Manual

Page 1
5700A/5720A
Series II Multi-Function Calibrator
Operator Guide
PN 601648 May 1996
© 1996 Fluke Corporation, Inc. All rights reserved. Printed in U.S.A.
®
Page 2
Page 3

Contents

What is in this Guide? ................................ 2
Safety Summary......................................... 2
P Fuse and Line Voltage .......................... 4
Basic Operation.......................................... 5
A Short Calibration Exercise....................... 5
Warming up the Calibrator ..................... 5
Exercise: Running dc Zero Calibration........ 6
Connecting a Meter................................ 6
Applying a dc Voltage ............................ 7
Checking the Calibrator Uncertainty.......7
Exercise: Activating Error Mode ............. 8
Checking the Meter’s Higher and Lower
Exercise: Applying an ac Voltage........... 9
Front Panel Features.................................. 10
Output Display (Left Side)...................... 10
Control Display (Right Side)................... 10
Display Screen Saver............................. 11
Front Panel Keys ................................... 12
Front Panel Connectors ......................... 21
Rear Panel Features .................................. 23
Rear Panel Connectors.......................... 23
Switches and Fuse Holder...................... 26
Cable Recommendations............................ 27
Connecting to a UUT (Unit Under Test) ...... 28
Using Error Mode....................................... 35
Operating in Remote .................................. 37
RS-232-C Interface Parameters.................. 38
Ranges..................................... 8
Page 4

What is in this Guide?

The
Operator Guide
information, a short meter calibration exercise for new users, then continues with a condensation of information from the
complete information about the Calibrator’s features, functions, and operating procedures, refer to the
Operator Manual
begins with safety
Operator Manual
.
. For

Safety Summary

Warning
[
Lethal voltage may be present on the terminals. Observe all safety precautions in this guide.
To avoid electrical shock hazard, the operator should not electrically contact the OUTPUT HI or SENSE HI binding posts. During operation, lethal voltages of up to 1100V ac or dc may be present on these terminals.
Whenever possible, keep one hand away from the equipment to reduce the hazard of current flowing through vital organs of the body.
Symbols Marked on Equipment
[
P
To avoid fire hazard, use only the fuse specified on the line voltage selection switch label.
Warning - Risk of electric shock.
Q
Ground Protective ground (earth) terminal. Attention Refer to the manual for
information about this feature. This symbol appears on the rear panel ground terminal and by the fuse holder.
Use the Proper Fuse
Page 5
Safety Summary (cont.)
[Grounding the Calibrator
The Calibrator is a Safety Class I (grounded enclosure) instrument. The enclosure is grounded through the grounding conductor of the power cord. To avoid electrical shock, plug the power cord into a properly wired earth grounded receptacle before making any connections to the Calibrator terminals. A protective ground connection by way of the grounding conductor in the power cord is essential for safe operation.
[Do Not Remove the Cover
To avoid personal injury, do not remove the Calibrator cover. Do not operate the Calibrator without the cover properly installed. There are no user-serviceable parts inside the Calibrator, so there is no need to ever remove the cover.
Page 6
P

Fuse and Line Voltage

The correct fuse type for each voltage selection is indicated on the rear panel as shown in the following figure.
CAUTION
FOR FIRE PROTECTION REPLACE ONLY
WITH A 250V FUSE OF INDICATED RATING.
VOLTAGE
SELECTION
S2 S3 S4
100V 110V 115V 120V
FUSE-F1
T 125A
250V
(SB)
VOLTAGE
SELECTION
S2 S3 S4
200V 220V 230V 240V
FUSE-F1
T 125A
250V
(SB)
CAUTION
FOR FIRE PROTECTION REPLACE ONLY
WITH A 250V FUSE OF INDICATED RATING.
VOLTAGE
VOLTAGE
FUSE-F1
SELECTION
S2 S3 S4 100V 110V 115V 120V
FUSE-F1
SELECTION
S2 S3 S4
200V
T 125A
T 125A
220V
250V
250V
230V
(SB)
(SB)
240V
Line Power Label and Switch Locations
Page 7

Basic Operation

To set the output, simply press the following sequence of keys to select an output function and amplitude:
[numeric keys] [multiplier] [function] E O For example, to set the output to 10 mV dc, press:
1 0 m V E O
To set an ac output, press the following additional keys:
[numeric keys] [multiplier] H E For example, to change the present 10 mV dc to
10 mV ac @ 1.8 kHz, press:
1 . 8 K H E
To change the output back to dc, press:
0 H E
or
+ E

A Short Calibration Exercise

If you are not familiar with the Calibrator, you will find that the following exercise quickly teaches you some important front panel operations.

Warming up the Calibrator

When you first turn the Calibrator on, you should let it warm up for 30 minutes. This ensures that the Calibrator meets or exceeds its specifications.
Page 8

Exercise: Running dc Zero Calibration

DC Zero is a brief internal process that removes offset errors. The specifications require that you run dc Zero at least every 30 days.
To run dc Zero, press the following sequence of softkeys:
Setup Menus Cal Zero Press any key, then press P twice to return to
normal operation.

Connecting a Meter

Typical connections for handheld and benchtop meters are shown on the next page. The diagrams show the use of Fluke Model 5440-7002 Low Thermal Cables. Other types of cable may be used for this exercise. To connect either type of meter, proceed as follows:
1. Verify that the Calibrator is on, and is in standby mode (STANDBY indicator lit). If the Calibrator is not in standby, press r.
2. Set the meter to the lowest range that reads 10V dc.
3. Connect the shield lead (if your cables are shielded) to V GUARD on the Calibrator.
4. Verify that both X and x are off. Press them if they are lit.
5. Connect the meter input LO to the Calibrator OUTPUT LO.
6. Connect the meter input HI to the Calibrator OUTPUT HI.
Page 9
Exercise: Connecting a Meter (cont.)
INPUT
DMM
10A
300mA
V
COM
NC
DMM
EX SNS
: OFF
EX SNS
NC
EX GRD
: OFF
: OFF
EX GRD
Calibrator
OUTPUT
V A
HI
LO
HI
AUX
CURRENT
: OFF
OUTPUT
HI
LO
HI
CURRENT
SENSE
Calibrator
ΩΩ
V A
AUX

Applying a dc Voltage

Apply 10V dc to the meter by pressing the following sequence of keys:
1 0 V E O
V
GUARD
SENSE
GUARD
WIDEBAND
HI
LO
GROUND
V
WIDEBAND
HI
LO
GROUND

Checking the Calibrator Uncertainty

Press s. The total uncertainty for the 10V you are applying to the meter is displayed on the Control Display (the large dot-matrix display on the right side of the front panel). This number depends on the setting of the calibration cycle and specification confidence level in a setup menu.
Press P to clear the uncertainty display.
Page 10

Exercise: Activating Error Mode

Chances are the meter reads something other than exactly 10V. To quickly check the error of the meter, turn the rotary knob to obtain a meter reading of exactly 10V, and read the error off the Control Display.
Turning the knob is all it takes to activate error mode. When you turn the knob, the least significant digit is highlighted. This digit changes as you turn the knob. To change a higher order digit, thereby speeding up the adjustment rate, press <.
The error on the Control Display is a combination of offset, scale, and linearity errors. You can view component parts of the error by pressing o and S.
Checking the Meter’s Higher and Lower Ranges
To check the meter error at 100V dc, first set the meter range (if applicable) to the lowest range that reads 100V. Press Y O. This sets a new reference equal to ten times the previous reference of 10V dc. (Pressing O was necessary because the Calibrator goes into standby when it crosses the safety threshold from below 22V to over 22V.) Turn the rotary knob to re-activate error mode and determine the meter error at 100V.
Now press Z Z. Set the meter range switch (if applicable) to the lowest range that reads 1V.
To end the error mode session, press E. This recalls the reference value and exits error mode.
Press r
to set the output to 0 mV dc in standby.
Page 11

Exercise: Applying an ac Voltage

There is no “ac mode” switch on the Calibrator. You change a dc output to ac by entering a frequency through the keypad and pressing E.
To test the meter at 10V at 1 kHz, set the meter to read 10V ac, press 1 0 V, then 1 K HEO. Adjust the rotary knob for a reading on the meter of exactly 10V.
To test the mete r’s f lat ne ss, pre ss a. The 0.1 Hz digit is highlighted. Press < four times. Turn the knob to change the 1 kHz digit until the Calibrator is set to 10 kHz. P ress a and turn the knob again to check the meter’s accuracy at 10 kHz. Each time you change the frequency, you need to wait for the Calibrator output to settle (the “u” annunciator will go out).
This ends the front panel exercise. To zero the output and go to standby, press r.
Page 12

Front Panel Features

Following is a brief description of the Calibrator’s front panel features.

Output Display (Left Side)

Shows output amplitude and frequency. The top line shows the active output value in up to eight digits, plus a polarity sign. The bottom line shows output frequency in five digits. Annunciators below the amplitude line on the output display indicate the following active conditions:
OPERATE Lit when an output is active at the
STANDBY Lit when the Calibrator is in standby. ADDR Lit when the Calibrator is addressed
ØLCK Lit when the Calibrator output is
ØSHF Lit when the Calibrator output has a
u (Unsettled) Lights when the output is
binding posts or auxiliary amplifier.
over the IEEE-488 interface.
phase locked to a signal at the rear panel PHASE LOCK connector.
programmed phase difference with a signal at the real panel VARIABLE PHASE OUT connector.
changed, and remains lit until the output settles to within specification.

Control Display (Right Side)

Shows data entries, UUT error adjustments, softkey labels, and other prompts and messages. Each softkey label identifies the functions of the softkey that appears directly below it. All of the softkey labels that appear on the display at once are referred to as a menu. When you access the function provided through the softkeys, you open up other menus, containing new softkey labels.
10
Page 13

Display Screen Saver

Both the Output and Control displays blank automatically after 30 minutes of inactivity, unless the Calibrator is in one of the operating states that overrides the screen saver.
If the screen saver has activated, you can restore the display by pressing C. Pressing another key or turning the knob restores the display and performs the command sent by the key or knob.
The screen saver does not operate:
if any part of the Setup menu is displayed.
during remote operation.
during calibration or diagnostics.
when an error message is displayed.
11
Page 14

Front Panel Keys

O
Toggles the Calibrator between operate and standby modes. In standby mode, the OUTPUT binding posts are internally disconnected from the Calibrator. The Calibrator powers up in standby mode.
The Calibrator automatically switches to standby mode when:
r is pressed.
the Calibrator is sourcing an ac or dc voltage
below 22V, and a voltage greater than 22V is selected.
the output function is changed, e.g., from dc current to ac current.
the output location is changed.
The Calibrator does not switch to standby when the voltage is switched from any ac or dc voltage to an ac or dc voltage less than 22V.
12
Page 15
Front Panel Keys (cont.)
X
Opens and closes an internal connection between the SENSE and OUTPUT binding posts. The Calibrator powers up with these two binding posts connected internally (the SENSE binding posts are open circuited), and the X indicator off. Toggling X on (so it lights) disconnects the sense lines from the OUTPUT binding posts.
External sensing should be used in the following function under the stated conditions.
In the dc voltage function when the UUT draws enough current to produce a significant voltage drop in the cables.
In the resistance function when the UUT has a four-wire input and the Calibrator is set to 100 k or less.
External sensing can also be used in the two- wire ohms function to activate two-wire compensation circuitry to the UUT terminals.
13
Page 16
Front Panel Keys (cont.)
x
Opens and closes an internal connection between the V GUARD (voltage guard) and OUTPUT LO. The Calibrator powers up with the V GUARD internally connected to OUTPUT LO and the x indicator off. Toggling x on (so it lights) disconnects OUTPUT LO from the V GUARD.
The V GUARD binding post provides an external connection point for the voltage internal guard. For a UUT with ungrounded inputs, the V GUARD should be connected to LO internally. (The EX GRD key is toggled off.) For a UUT with a grounded input, the GUARD may be externally connected to the grounded UUT input. (The EX GRD key is toggled on.)
w
Enables and disables the -03 Wideband AC option and sets the Calibrator to standby mode. When enabled, the ac voltage output over the range of 10 Hz to 30 MHz is available at the front panel WIDEBAND connector. The wideband function is disabled whenever w is toggled off, or when another function (such as current) is selected.
14
Page 17
Front Panel Keys (cont.)
B
Enables or disables output from an amplifier, when it would not otherwise be automatically selected. Sets the Calibrator to standby if this selection moves the output location.
When available, an amplifier is automatically selected for output settings that exceed Calibrator capabilities, but fall within the limits of the selected amplifier. The B key is only needed to activate an amplifier for an output setting that is available from either the Calibrator or from the amplifier. This allows you to take advantage of amplifier capabilities other than extended range (i.e. higher compliance voltage).
For both voltage and current, the amplifier is assumed to be a 5725A unless another model is designated in the Setup menus.
P
Aborts the current operation of the Calibrator and recalls the previous menu. Some menus display a
more specific label for this key, such as “DONE setting up.”
Softkeys The functions of the five unlabeled softkeys are
provided through each menu in the Control Display, with each function located directly above the softkey that activates that function. Many of these functions access new menus that contain a new set of functions for the softkeys, providing access to many capabilities and configuration options through the menu tree.
POWER switch Turns the power on and off.
u
15
Page 18
Front Panel Keys (cont.)
<a>
These keys are the output adjustment controls. If any of these keys are pressed, or if the knob is turned, a digit on the Output Display becomes highlighted. Turning the knob adjusts the output value. An error display appears on the Control Display, showing the difference between the original (reference) output and the new (adjusted) output.
The < and > keys move the highlight to the digit you wish to change. In the ac functions, the a key allows you to move amongst voltage, current, and frequency. In practice, for voltage and current outputs, the knob and arrow keys are used to adjust output until the UUT reads correctly. The error display shows the UUT deviation from the reference.
r
During local operation, pressing r aborts the current operating state of the Calibrator and returns the Calibrator to the power-up default state. (This key is not effective during remote operation.)
S
Identifies a UUT full-scale endpoint for checking linearity, and does not change the output. If the output was adjusted with the rotary knob, subsequent keyed-in output values are multiplied by a scale factor. Scaling is deactivated by pressing S again, or by selecting another function. Scaling is not available for resistance outputs.
16
Page 19
Front Panel Keys (cont.)
L
Calls up a menu that allows you to specify limits beyond which the Calibrator will not operate. This can be used to set limits that protect both operators and equipment.
Z
Changes the output to one tenth the reference value (which is not necessarily the same as the present output value). The value only changes if the new value is within performance limits.
Y
Changes the output to ten times the reference value (which is not necessarily the same as the present output value). The value only changes if the new value is within performance limits. This key sets the Calibrator to standby if this change is made from a value less than 22V to one that is greater than or equal to 22V.
s
Causes the Calibrator to compute and display its uncertainty for the present output setting for the calibration interval and specification confidence level selected in the setup menus.
o
Identifies a UUT zero-scale endpoint and does not change the output. Subsequent keyed-in output values have the offset value (the Calibrator output value when o was pressed) added to them. Offset mode is deactivated by pressing o again, or by selecting another function. Offsets are available for dc outputs only.
17
Page 20
Front Panel Keys (cont.)
C
Clears a partially completed keypad entry from the Control Display, or clears an error message that required acknowledgment, If there is a partially completed entry when C is pressed, the output is unaffected.
N
This key is only active during error mode operation. It establishes the present output value as a new reference for meter error computation.
D
When in the ac volts, or wideband function, and if no entry is in progress, the D key shows the equivalent dBm output on the Control Display. For the ac voltage function, dBm is calculated for a 600 load. (For ac outputs less than 220 mV, the Calibrator output impedance is always 50because the output is resistively divided.) For the wideband function, dBm is calculated for a 50 resistive termination at the end of a three-foot 50 coaxial cable.
Note
The formula to compute dBm is:
dBm = 10 log (power in mW)
Examples:
For 3.0V into a 600 load: dBm = 10 log (15.000) = 11.7609 dBm
for 3.0V into a 50 load: dBm = 10 log (180.000) = 22.5527 dBm
18
Page 21
Front Panel Keys (cont.)
E
Loads a newly keyed-in output value showing on the Control Display into the Calibrator. If you press E without identifying the units for the entry, the Calibrator keeps the units (except Hz) that were last used. This allows you to alter values for a function without keying in the units every time. The multipliers are not retained, for example, you can enter 1 mV, and then later simply enter 10 to obtain 10V.
In addition to these functions, the E key recalls the currently programmed reference value during error mode operation.
D V A Q H
(Output Function Keys) These keys select the output function as follows:
D Decibels relative to 1 mW V Voltage A Current Q Resistance H Frequency
When a value, then H are selected, the Calibrator automatically switches to ac. To change back to dc, enter 0 H or enter a signed (+ or ­amplitude).
U m K M
(Multiplier Keys) These keys select output value multipliers. For
example, if you enter 33, then m, then V, then ENTER, the Calibrator output value is 33 mV.
U micro (10 m milli (10 K kilo (10 M Mega(10
-6
or 0.000001)
-3
or 0.001
3
or 1000)
6
or 1,000,000)
19
Page 22
Front Panel Keys (cont.)
0 - 9 (Numeric Keypad)
Used to enter the digits of the output amplitude and frequency, as well as other data such as the time and date. The proper sequence for entering a value is to enter the digits, then any necessary multipliers, an output function key, then E. For example, to obtain an output of 20 mV, you would press the following sequence of keys: 2 0 m V E. The . key is available for decimal places.
+
If the output function is dc voltage, current, ac voltage entered in dBm, or a wideband output entered in dBm, pressing + E toggles the polarity of the output. If the output function is ac voltage or current, pressing + E changes the output to dc.
20
Page 23

Front Panel Connectors

WIDEBAND Connector
A type “N” connector that provides a connection point for output from the -03 Wideband AC option. Wideband output specifications are stated for output levels present at the end of its three-foot 50 coaxial cable terminated into a 50 purely resistive load. The connector shell is connected to chassis ground.
GND Binding Post If the Calibrator is the location of the ground
reference point in a system, the GND binding post can be used for connecting other instruments to earth ground. (The chassis is normally connected to earth ground through the three-conductor line cord instead of through the earth ground binding post.)
V GUARD Binding Post Provides an external connection point for the
internal voltage guard. For a UUT with floating (ungrounded) inputs, the C GUARD should be connected to LO internally (x key toggled off). For a UUT with a grounded input, the V GUARD must be externally connected to the grounded UUT input (x key toggled on). The maximum allowable potential between the V GUARD connector and chassis ground is 20V peak.
21
Page 24
Front Panel Connectors (cont.)
SENSE Binding Posts Used in the resistance and voltage functions for
sensing at the UUT after you have selected external sense by toggling the X key on, or by remote command.
External sensing should be used in the dc voltage function when the UUT draws enough current to produce a significant voltage drop in the cables, and in the resistance function when the UUT has a four-wire ohms input and the Calibrator is set to 100 k or less. External sensing can also be used in the two-wire ohms function to allow the two-wire compensation circuitry to the UUT terminals. Refer to the connection diagrams for examples of usage.
OUTPUT Binding Posts Provide connection points for ac and dc current
and voltage output, as well as for resistance. The function of each OUTPUT binding post is defined below:
LO The common binding post for all
HI The active binding post for all output
AUX CURRENT OUTPUT
output functions including Calibrator amplified voltage output, but not for Option -03 Wideband AC or auxiliary amplifier output.
functions, including 5725A amplified voltage output; does not include Option -03 Wideband AC or auxiliary amplifier output.
An optional active binding post for current. This output is convenient for calibrating UUTs that have a separate current input terminal.
22
Page 25

Rear Panel Features

The following list provides descriptions of the
connectors and switches on the Calibrator’s rear panel.

Rear Panel Connectors

5725A AMPLIFIER Connector Provides the analog and digital interface for the
Fluke 5725A Amplifier. 5205A AMPLIFIER Connector Provides the analog and control interface for either
the Fluke 5205A or 5215A Precision Power Amplifier.
5220A AMPLIFIER Connector Provides the analog and control interface for the
Fluke 5220A Transconductance Amplifier. VARIABLE PHASE OUT BNC Connector Provides access to a variable-phase nominal 2.5V
rms sine wave signal. The phase of this signal can be adjusted using the arrow keys and rotary knob (or by remote commands) to lead or lag the main Calibrator output signal by up to 180°. The connector shell is not connected directly to chassis ground; it is connected internally to the OUTPUT LO binding post. The maximum allowable potential between the connector shell and chassis ground is 20V peak.
23
Page 26
Rear Panel Connectors (cont.)
PHASE LOCK IN BNC Connector Provides the input for an external signal onto
which the Calibrator can be phase locked (1V to 10V rms, 10 k input impedance). The connector shell is not connected directly to chassis ground; it is connected internally to the OUTPUT LO binding post. The maximum allowable potential between the connector shell and chassis ground is 20V peak.
IEEE-488 CONNECTOR This is a standard interface connector for
operating the Calibrator in remote control as either a Talker or a Listener of the IEEE-488 Bus.
RS 232C Connector A male (DTE) serial port connector for transmitting
internal calibration constant data to a printer, monitor, or host computer. This connector is also used for remote control of the Calibrator.
24
Page 27
Rear Panel Connectors (cont.)
The OUTPUT, SENSE, and V GUARD binding posts on the rear panel are alternative connections to the UUT. An internal cable enables either the front or rear binding posts. The procedure to disable the front panel binding posts and enable the rear panel binding posts involves opening the cover of the Calibrator, which should only be done by authorized service personnel.
The I GUARD binding post provides an external connection point for the internal current guard. The current guard is used when the Calibrator is supplying low-level ac current through a long cable to remove errors introduced by leakage through the cable capacitance.
Refer to the connection diagrams for examples of usage.
AUX CURRENT OUTPUT This output is not available on the rear panel
binding posts. CHASSIS GROUND Binding Post A binding post that is internally grounded to the
chassis. If the Calibrator is the location of the ground reference point in a system, this binding post can be used for connecting other instruments to earth ground. (The chassis is normally connected to earth ground through the three­conductor line cord instead of through the earth ground binding post.)
25
Page 28

Switches and Fuse Holder

CALIBRATION SWITCH A slide switch that enables and disables the
capability to write to nonvolatile memory. This switch is used for recording and then storing calibration constants, dates, and parameter settings. When this switch is in the ENABLE position, you can write to memory; when in the NORMAL position, the data in the memory is protected from being overwritten. The switch is recessed in order to allow the metrologist to cover it with a calibration sticker to guarantee calibrator integrity.
F1 Fuse Holder
This holds the line power fuse. See “Fuse and Line Voltage” at the beginning of this guide for the location of the correct ratings.
Line Voltage Selection Switches Select the operating line voltage. See “Fuse and
Line Voltage” at the beginning of this guide for the location of the label that shows the correct ratings.
26
Page 29

Cable Recommendations

Output Function Cable Recommendations
DC Voltage AC Voltage 10 kHz AC Current ≤2A, ≤10
kHz DC Current ≤2A Resistance AC Voltage > 10 kHz SENSE/GUARD: Triaxial cable or
AC Current with Guard Triaxial cable Wideband AC 3 feet (1m) 50 coaxial cable with
Voltage-Boosted Output, 5205A or 5215A
Voltage-Boosted Output, 5725A
Current-Boosted Output, 5725A, 5220A
Low Thermal EMF Test Leads (5440A-7002) or twisted shielded pair
Twinax (e.g., Alpha 2829/2). OUTPUT: Coaxial
Or: SENSE: Coaxial, OUTPUT: Coaxial GUARD Lead: Separate wire
Type “N” male connector supplied with the option. A 50 feedthrough terminator is also supplied for connecting to meters with impedance higher than 50Ω.
Use the cable supplied with the amplifier.
Low Thermal EMF Test Leads (5440A-7002). (Output is at the Calibrator front panel.)
16-gauge or heavier twisted-pair insulated wire, as short as possible to minimize resistance and inductance. (Output is at the amplifier terminals.)
27
Page 30

Connecting to a UUT (Unit Under Test)

EX SNS
: OFF
A.
UUT
SENSE
INPUT
4-WIRE
GUARD
B.
HI
LO
: ON
EX SNS
HI
LO
A
UUT
INPUT
DC
AC
C.
SHIELDED TWISTED
SHIELDED TWISTED
: OFF
EX SNS
PAIR
PAIR
UUT
LOHI
EX GRD
EX SNS
EX GRD
: OFF
OUTPUT
HI
LO
HI
AUX
CURRENT
: ON
OUTPUT
V A
HI
LO
HI
AUX
CURRENT
: ON
OUTPUT
HI
LO
HI
CURRENT
Calibrator
SENSE
V A
V
GUARD
Calibrator
SENSE
V
GUARD
Calibrator
SENSE
V A
AUX
GUARD
HI
LO
V
HI
LO
GROUND
WIDEBAND
GROUND
WIDEBAND
WIDEBAND
HI
LO
GROUND
DC Voltage, AC Voltage 10 kHz
28
Page 31
Connecting to a UUT (cont.)
A.
EX SNS
: ON
EX GRD
: OFF
UUT
HI
LO
INPUT
A
SENSE
4-WIRE
GUARD
B.
HI
LO
EX SNS
TRIAXIAL CABLE
TRIAXIAL CABLE
: OFF
EX GRD
HI
LO
HI
CURRENT
: OFF
UUT
OUTPUT
HI
LO
HI
CURRENT
HI
LO
INPUT
A
SENSE
4-WIRE
GUARD
TRIAXIAL CABLE
HI
LO
NOTE
Keep the SENSE leads as short as possible.
Calibrator
OUTPUT
V A
AUX
V-GUARD
Calibrator
V A
AUX
V-GUARD
SENSE
V
SENSE
V
WIDEBAND HI
LO
GROUND
WIDEBAND
HI
LO
GROUND
Be careful not to exceed the capacitive load limit of 1000 pF up to 220V, 600 pF 220 to 1100V. (1000 pF with 5725A Amplifier.)
AC Voltage >10 kHz
29
Page 32
Connecting to a UUT (cont.)
HI
LO
HI
LO
INPUT
A
INPUT
A
UUT
SENSE
4-WIRE
UUT
SENSE
4-WIRE
GUARD
GUARD
A.
EX SNS
HI
LO
B.
EX SNS
TRIAXIAL CABLE
HI
LO
NC
AC Current ≤2A
: OFF
: OFF
EX GRD
EX GRD
: OFF
Calibrator
OUTPUT
SENSE
V A
HI
LO
HI
AUX
GUARDGROUND
CURRENT
: OFF
Calibrator
OUTPUT
SENSE
V A
HI
LO
I-GUARD
V-GUARD
(FRONT)
V
WIDEBAND
HI
LO
(REAR)
V
HI
LO
30
Page 33
Connecting to a UUT (cont.)
HI
LO
A
INPUT
UUT
10A
300mA
UUT
SENSE
4-WIRE
V
COM
A.
GUARD
UUT
B.
NC
EX SNS
: ON
EX GRD
: OFF
2-WIRE
COMP
OFF
Calibrator
SENSE
OUTPUT
V
V A
HI
LO
SENSE
SOURCE
SOURCE
SENSE
: OFF
EX SNS
EX GRD
HI
LO
HI
AUX
CURRENT
: OFF
WIDEBAND
HI
LO
GUARD
GROUND
Calibrator
2-WIRE COMP
OFF
Calibrator
OUTPUT
SENSE
V A
V
WIDEBAND
GUARD
HI
LO
GROUND
HI
LO
HI
AUX
CURRENT
UUT
Calibrator
Resistance
31
Page 34
Connecting to a UUT (cont.)
EX SNS
: ON
C.
UUT
V
10A
COM
300mA
Use Connections With Exposed Plug Tips For The Ohms Function Only.
Caution
NC
NC
EX GRD
: OFF
OUTPUT
V A
HI
LO
HI
AUX
CURRENT
COMP
2-WIRE
COMP
ON
Calibrator
SENSE
V
WIDEBAND
HI
LO
GUARD
GROUND
UUT
10A
COM
300mA
UUT
D.
EX SNS
: OFF
EX GRD
COMP
: OFF
Calibrator
2-WIRE
COMP
ON
Calibrator
OUTPUT
SENSE
V A
V
WIDEBAND
HI
LO
HI
V
NC
CURRENT
COMP
AUX
UUT
COMP
HI
LO
GUARD
GROUND
Calibrator
Resistance (cont.)
32
Page 35
Connecting to a UUT (cont.)
50 Feedthrough Terminator
Supplied With Option -03
LO
CURRENT
OUTPUT
V A
HI
HI
AUX
Calibrator
SENSE
V
HI
LO
GUARD
GROUND
WIDEBAND
Cable Supplied With
Option -03
Cable Supplied With
Option -03
NOTE
For wideband meters with higher than 50input impedance, use the 50 feedthrough terminator at the meter connection end. For all wideband applications, take care to achieve a good 50impedance match (use cable and connectors with a characteristic input impedance of 50).
Wideband AC Voltage Output
(Option -03)
33
Page 36
Connecting to a UUT (cont.)
5725A
HI
LO
UUT
INPUT
HI
LO
A
SENSE
4-WIRE
GUARD
HI
LO
5725A Amplified Current Output
34
Page 37

Using Error Mode

Activate error mode by turning the rotary knob, pressing an arrow key, or pressing a. When you enter error mode, the starting value is the reference from which errors are computed. A new reference is established when you exit and reenter error mode. The following table lists the actions that cause the Calibrator to exit error mode.
35
Page 38
Using Error Mode (cont.)
Keys Action
E Returns to the previous reference
+, then E Establishes a new reference.
A keypad entry, then
E N Establishes the present output as a
Y Sets the Calibrator to ten times the
Z Sets the Calibrator to one-tenth the
o Identifies the present output as a zero-
S Identifies the present output as a full-
r Returns to the power-up state.
The “Setup Menus” softkey
value.
Establishes a new reference.
new reference.
reference value and establishes a new reference.
reference value and establishes a new reference.
scale endpoint for scaling, and establishes 0.0 as the new reference.
scale endpoint for scaling and causes the display to show scale error.
Opens the setup menu.
36
Page 39

Operating in Remote

To operate the Calibrator remotely, proceed as follows:
1. Turn off the Calibrator power.
2. Connect a remote interface to either the RS­232C or IEEE-488 connector on the Calibrator rear panel.
3. Turn on the Calibrator power.
4. Choose the correct remote parameters using the softkey menus:
a. Setup Menus Instmt Setup Remote Port
Setup
b. Select Remote Port type (GPIB = IEEE-
488), Remote lang (NORMAL or emulation mode)
c. Select either GPIB Port Setup or RS-232
Port Setup. d. If GPIB: select the address. e. If RS-232: select the parameters in the
next table.
37
Page 40

RS-232-C Interface Parameters

Parameter Choices Default Setting
Data Bits 7 or 8 8 Stop Bits 1 or 2 1 Flow Control Ctrl S/Ctrl Q
(XON/XOFF), RTS,
or none Parity Checking Odd, even, or none none Baud Rate 110,300,600,1200,
2400,4800,9600, or
19200 Time-out Period 0 to 30 seconds 0 (no time-out) EOL (End of Line) CR, LF, or CR LF CR LF EOF (End of File) Any two ASCII
characters
Ctrl S/Ctrl Q
9600
no characters
38
Loading...