The Fluke 568 Ex
Intrinsically Safe
Infrared Thermometer
Intrinsically safe
temperature measurements .
Anywhere in the world .
Technical Data
The Fluke 568 Ex Intrinsically Safe Infrared
Thermometer is the one product you can use in
Class I Div 1 and Div 2 or Zone 1 and 2 hazardous
environments anywhere in the world. Whether
you work in petroleum, chemical, oil & gas or
pharmaceutical environments, the new 568 Ex
allows you to carry the most trusted name in test
tools into most Ex rated areas all around the globe.
With a straight-forward user interface and softkey menus, the Fluke 568 Ex makes even complex
measurements easy. Quickly navigate and adjust
emissivity, save data or turn on and off alarms,
with just a few pushes of a button. All in a single
intrinsically safe tool certified by major rating
bodies from around the world.
Product Highlights
With a rugged, easy-to-use, ergonomic design,
the Fluke 568 Ex can stand up to tough industrial,
electrical, and mechanical environments.
• Meets intrinsically safe certifications in Class I
Div 1 and Div 2 or Zone 1 and 2 hazardous
environments from recognized safety agencies
around the world
• Measure -40 °C to 800 °C (-40 °F to 1472 °F)
• Conductive Case for carrying the IR thermometer
safely into hazardous area
• Easily access advanced features with the soft-key
buttons and graphical display
• Measure small objects from further away, with
a distance-to-spot ratio of 50:1
• Compatible with mini-connector K-type thermo-
couple (KTC) probe
• Confidently measure a wide variety of surfaces
with the adjustable emissivity feature, including
a built-in material table
• Capture up to 99 points of data
• Confidently troubleshoot equipment with ± 1 %
measurement accuracy
• Versatile interface with five languages from which
to choose
• Two-year warranty
Specifications
Fluke 568 Ex Infrared Thermometer
Infrared temperature
range
Infrared accuracy < 0 °C (32 °F): ± (1.0 °C (± 2.0 °F) + 0.1°/1 °C or °F);
Display resolution 0.1 °C/0.1 °F
Infrared spectral response 8 µm to 14 µm
Infrared response time < 500 ms
Thermocouple Type-K
input temperature range
Thermocouple Type-K
input accuracy
D:S (distance to measurement spot size)
Laser sighting Single-point laser
Minimum spot size 19 mm (0.75 in)
Emissivity adjustment By built-in table of common materials or digitally
Data storage 99 points
Hi/Low alarms Audible and two-color visual
Min/Max/Avg/Dif Yes
Display Dot matrix with function menus
Backlight Two levels, normal and extra bright for darker
Trigger lock Yes
Switchable Celsius and
Fahrenheit
Power 2 AAA/LR03 type-approved Batteries (For a list of
Battery life 4 hours with laser and backlight on; 100 hours with
Operating temperature 0 °C to 50 °C (32 °F to 122 °F)
Storage temperature -20 °C to 60 °C (-4 °F to 140 °F)
Bead thermocouple
Type-K range
Bead thermocouple
Type-K accuracy
-40 °C to 800 °C (-40 °F to 1472 °F)
≥ 0 °C (32 °F): ± 1 % or ± 1.0 °C (± 2.0 °F),
whichever is greater
-270 °C to 1372 °C (-454 °F to 2501 °F)
<-40 °C: ±(1 °C + 0.2 °/1 °C)
≥-40 °C: ±1 % or 1 °C, whichever is greater
<-40 °F: ±(2 °F + 0.2 °/1 °F)
≥-40 °F: ±1 % or 2 °F, whichever is greater
50:1
adjustable from 0.10 to 1.00 by 0.01
environments
Yes
type-approved batteries, refer to Product Safety
Instructions)
laser and backlight off, at 100 % duty cycle
-40 °C to 260 °C (-40 °F to 500 °F)
± 1.1 °C (2.0 °F) from 0 °C to 260 °C
(32 °F to 500 °F), typically within 1.1 °C (2.0 °F)
from -40 °C to 0 °C (-40 °F to 32 °F)
Ordering information
FLUKE-568 Ex Intrinsically Safe Infrared Thermometer
Included equipment
• K-type thermocouple bead probe
• Conductive IS hard carrying case
• User’s manual
Safety Certifications
Agency Safety rating
ATEX/IECEx
NEC-500/NEC-505
GOST
PCEC
INMETRO
Fluke. The Most Trusted Tools
in the World.
Fluke Corporation
PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2014 Fluke Corporation. Specifications subject to change
without notice. Printed in U.S.A. 9/2014 6000203C_EN
Modification of this document is not permitted without
written permission from Fluke Corporation.
Zone 1 and 2
IECEx EPS 13.0006X
Ex ia IIC T4 Gb
0 °C ≤ Ta ≤ 50 °C
EPS 13 ATEX 1.525 X
II 2G Ex ia IIC T4 Gb
Class I Division 1 and 2
Class I, Division 1, Groups
ABCD T4
Class I, Division 2, Groups
ABCD T4
Class I, Zone 1, AEx ia IIC T4
Ex ia IIC T4
0 °C ≤ Ta ≤ 50 °C
Zone 1 and 2
POCC DE.
Ex ia IIC T4 Gb X
ОТ 0 °C ДО +50 °C
EPS 13 ATEX 1 525 X
II 2G Ex ia IIC T4 Gb
0 °C ≤ Ta ≤ 50 °C
Zone 1 and 2
PCEC
Ex ia IIC T4 Gb
CE13.
EPS 13 ATEX 1 525 X
II 2G Ex ia IIC T4 Gb
0 °C ≤ Ta ≤ 50 °C
Zone 1 and 2
IEx 13.0122X
Ex ia IIC T4 Gb
EPS 13 ATEX 1 525 X
II 2G Ex ia IIC T4 Gb
0 °C ≤ Ta ≤ 50 °C
ГБ05.B