Fluke 312, 318, 316 User Manual

Page 1
312, 316, 318
Clamp Meter
®
PN 1989445 July 2002 Rev.2, 2/06
© 2002, 2006 Fluke Corporation. All rights reserved. Printed in China. All product names are trademarks of their respective companies.
Page 2
LIMITED WARRANTY AND LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and workmanship for one year from the date of purchase. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, con­tamination, or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Fluke’s behalf. To obtain service during the warranty period, return the unit to point of purchase with a descrip­tion of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDEN­TAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential dam­ages, this limitation of liability may not apply to you.
Fluke Corporation P.O. Box 9090 Everett, WA 98206-9090 U.S.A.
Service Center Address: Fluke Corporation Beijing Service Center Room 2301 Scite Tower 22 Jianguomenwai Dajie, Chao Yang District Beijing 100004, PRC Tel: 010-65123435
Shanghai ShiLu Instrument Co., Ltd. No. 139, Lane 2638, Hongmei South Road Shanghai 201108 Standard No. Q/SXAV 2-2002
Page 3
Table of Contents
Title Page
Safety Information – Read first.........................1
General Specifications .....................................4
Electrical Specifications....................................5
Instrument Overview Feature Locations...........6
Symbol Definition...........................................7
Making Measurements .....................................7
Measuring AC Current ...................................8
Measuring DC Current (316, 318 only) ..........9
Measuring AC and DC Voltage ......................9
Measuring Resistance ...................................12
Testing Continuity ..........................................13
Diode Test......................................................14
Data Hold.......................................................15
Peak...............................................................15
MIN MAX (316, 318 only)...............................15
REL (S) (316, 318 only) ................................16
Range (312 only) ...........................................16
Changing the Battery........................................16
Cleaning the Meter...........................................16
Maintenance.....................................................16
i
Page 4
ii
Page 5
312, 316, 318
Safety Information – Read first
WWarnings: Read Before Using
To avoid possible electric shock or personal injury:
• Use the Meter only as specified in this manual or the
protection provided by the Meter might be impaired.
• Do not use the Meter in wet environments.
• Inspect the Meter before using it. Do not use the Meter
if it appears damaged. Look for cracks or missing
plastic. Pay particular attention to the insulation
around the connectors.
• Inspect the test leads before use. Do not use them if
insulation is damaged or metal is exposed. Check the
test leads for continuity. Replace damaged test leads
before using the Meter.
• Have the Meter serviced only by qualified service
personnel.
• Do not apply more than the rated voltage, as marked
on the Meter, between the terminals or between any
terminal and earth ground.
• Remove test leads from the Meter before opening the
Meter case.
• Never operate the Meter with the back cover removed
or the case open.
• Never remove the back cover or open the case of an
instrument without first removing the power source.
• Use caution when working with voltages above 33 V ac
rms, 46.7 V ac peak, or 70 V dc. These voltages pose a
shock hazard.
• Use the proper terminals, function, and range for your
measurements.
• Do not operate the Meter around explosive gas, vapor,
or dust.
1
Page 6
312, 316, 318
• When using probes, keep your fingers behind the
finger guards.
• When making electrical connections, connect the
common test lead before connecting the live test lead; when disconnecting, disconnect the live test lead before disconnecting the common test lead.
• Disconnect circuit power and discharge all high-
voltage capacitors before testing resistance, continuity, or diodes.
• Use only a single 9 V battery, properly installed in the
Meter case, to power the Meter.
• Replace the battery as soon as the low battery
indicator ( lead to electric shock and injury.
• Check Meter operation on a known source before and
after use.
• When servicing, use only specified replacement parts.
• Adhere to local and national safety codes. Industrial
protective equipment must be used to prevent shock and arc blast injury where hazardous live conductors were exposed.
• Do not hold the Current Probe anywhere beyond the
tactile indicator (see Figure 1).
B) appears to avoid false readings that can
Environmental Conditions:
Altitude: up to 2000 meters. Operating temperature: 0 °C ~ 30 °C, up to 90 % RH, non-
condensing 30 °C ~ 40 °C, up to 80% RH, non-condensing
Storage temperature: -10 °C ~ 60 °C, battery removed Pollution Degree: 2 Measurement Categories: CAT III (600 V), CAT II (1000 V)
Electromagnetic Compatibility: 3 V/m, performance criterion B
2
Page 7
Explanation of Symbols
Users Manual
W X
,
T
B
J B F
CAT II
CAT III
Risk of danger. Important information. Refer to operation instructions.
Risk of electric shock.
Application and removal from hazardous live conductors permitted.
Double insulated.
Battery
Earth Ground
Alternating current
Direct current
Equipment is designed to protect against transients from energy-consuming equipment supplied from the fixed installation, such as televisions, personal computers, portable tools, and other household appliances.
Equipment is designed to protect against transients in equipment in fixed­equipment installations, such as distribution panels, feeders and short branch circuits, and lighting systems in large buildings.
P
Complies with European standards EN/IEC 61010-1 2nd Edition and EN/IEC 61010-02-032
3
Page 8
312, 316, 318
General Specifications
Digital Display: 3-3/4 digits with maximum reading of 3999 Analog Display: Fast 42-segment analog bar graph display Overload: Displays OL when input signal exceeds measuring limit
in 40 A, 400 A, resistance, and continuity functions.
Note
In Volts and Amps, the LCD may continue to display values above the indicated measuring limit (up to 4000 counts). Refer to specifications for measuring limits.
Sample Rate: 2 times/sec for digital display; 20 times/sec for analog bar graph
Low Battery Indication: Displays required voltage
Power Source: 9 V battery, NEDA1604 or 6F22 or 006P Battery Life:
(Alkaline): 33 hrs typical in A 65 hrs typical in GR 100 hrs typical in V, e
Auto Power Off: The Meter will power itself off if there is no push button or rotary function switch operation for 30 minutes. Turn the rotary switch to “OFF” before turning the Meter on again.
To deactivate this feature, press the Peak button and keep it pressed down while powering up the Meter.
Clamp Opening Size: 312: 36 mm (1.42 inches) 316, 318: 40 mm (1.57 inches)
Dimensions (L x W x H): 254 x 66 x 37 mm (10 x 2.60 x 1.45 inches)
Weight: 422 g (14.9 ounces) (battery included)
B when the battery is below its
Included Accessories: Users manual, carrying case, test leads, 9 V battery
Approvals:
P, EN61010 600 V CAT III, 1000 V CAT II, CMC
4
Page 9
Users Manual
Electrical Specifications
The accuracy specification is defined as + (% reading + digits) at
o
23
C + 5 oC1.
Function
AC Amps
2
(50 ~ 500 Hz)
DC Amps
AC Volts
2
(50 ~ 500 Hz)
DC Volts
Resistance (ohms)
Continuity
Diode Test
Range
40.00 A
400.0 A 1000 A
40.00 A
400.0 A 1000 A
400.0 V 750 V
400.0 V 1000 V
400.0 Ω
Ω
4000
On at < 50 Ω
Up to 2 V
Resolution
0.01 A
0.1 A 1 A
0.01 A
0.1 A 1 A
0.1 V 1 V
0.1 V 1 V
0.1
Ω
Ω
1
Accuracy (% reading + digits)
312
1.9 %
±
5
1.9 % ± 5 (50 ~ 60 Hz)
2.5 %
- 2.5 %
1.2 %
±
5 1.5 % ± 5
0.75 %
1 %
±
2 1 % ± 2
±
3 1 % ± 3
316 & 318
±
5 (60 - 500 Hz)
±
10
Min Max
Input
500 ms acquisition time
10 MΩ
Impedance
Auto Shut-off
Auto range
Overload Protection
After 30 min ± 2 min. Disabled in Min Max. Can be disabled as a power up option (hold down the peak button)
Available in the following measurement functions: Volts & Ohms (Amps for 312 only)
600 Vrms per EN61010 CAT III 600 V
1. Temperature Coefficient: 0.1 x (specified accuracy) / oC (< 18 oC or > 28 oC)
2. True rms for ac voltage and ac amps accuracy is specified from 5 % to 100 % of range. Crest factor (50 ~ 60 Hz) is 3.0 max in the 40 A and 400 A ranges and 1.4 max in the 1000 A range (318 only).
3. The measurement time for current above 400 A is limited to 5 minutes followed by a cooling off time of 15 minutes.
5
Page 10
312, 316, 318
Instrument Overview Feature Locations
13
14
12
400
40
A
11
OFF
OFF
1000
A
PEAK HOLD
PEAK HOLD
A
1
2
3
RANGE
V
4
5
10
9
A
Current Sensing Clamp
B
Peak Button
C
Hold Button
D
Range Button (312)
E
AC/DC Button (316, 318)
F
MIN MAX Button (316, 318)
G
LCD
AC/DC MIN MAXREL
318
TRUE RMS CLAMP METER
H
V e
Input Terminal
I
COM
Terminal
J
REL Button (316, 318)
K
Rotary Function Switch
L
Clamp-opening Trigger
M
Conductor Alignment Marks
N
Tactile Indicator
6
7
8
aca11f.eps
Figure 1. Feature Locations (318 shown)
6
Page 11
Symbol Definition
1 14 13 12 11
Users Manual
2
3
4
5
MAX
A
B
C
D
E
F
Maximum reading displayed (316, 318 only)
MIN
Minimum reading displayed (316, 318 only)
Relative (S) mode is active
Negative reading
Battery is low and should be changed
Bar graph
Manual range mode
10
6
H
AC mode
V
I
J
K
L
M
Volts
A
Amps
e
Ohms
Hold function is selected
Diode/continuity test function is selected
Peak- function is selected
9
8
7
aca02f.eps
G
DC mode
Figure 2. LCD Explanation
Making Measurements
XWarning
• When measuring current, center the conductor in the
clamp using the alignment marks on the clamp.
• To avoid electrical shock when making current
measurements, disconnect the test leads from the Meter.
• The bargraph is for reference only; it does not show
the actual measured value or range. Use the main display to read the actual measured value and range.
Peak+ function is
N
selected
7
Page 12
312, 316, 318
Measuring AC Current
1. Turn the rotary function switch to the proper current range.
2. Press
A(316, 318 only) to measure ac current if necessary.
3. Open the clamp by pressing the clamp-opening trigger and insert the conductor to be measured into the clamp.
4. Close the clamp and center the conductor using the alignment marks.
5. View the reading on the LCD.
400
PEAK
PEAK
40
1000
A
A
OFF
OFF
A
HOLD
M
A
X
HOLD
V
Tactile
Barrier
PEAK HOLD
PEAK HOLD
1000
A
A
400
V
A
40
OFF
OFF
AC/DC MIN MAXREL
TRUE RMS
318
CLAMP METER
R
E
L
A C
/D
C
M
IN
318
TRUE RMS CLAMP METER
Figure 3. Connecting the Clamp Meter (318 shown)
aca13f.eps
8
Page 13
Users Manual
Measuring DC Current (316, 318 only)
1. Turn the rotary function switch to the proper current range.
2. Press
3. To ensure accuracy, wait for the reading to stabilize then press
4. Open the clamp by pressing the clamp-opening trigger and insert the conductor to be measured into the clamp.
5. Close the clamp and center the conductor using the alignment marks.
6. View the reading on the LCD.
A to measure dc current.
Q to zero the reading.
Measuring AC and DC Voltage
1. Turn the rotary function switch to V.
2. Choose ac or dc by pressing A(316, 318 only).
When using a 312 Meter:
• turn the rotary function switch to K when measuring ac
voltage
• turn the rotary function switch to L when measuring dc
voltage
3. Connect the black test lead to the COM terminal and the red test lead to the Ve terminal.
4. Measure the voltage by touching the probes to the desired test points of the circuit.
5. View the reading on the LCD.
9
Page 14
312, 316, 318
1000
A
400
A
40
OFF
OFF
PEAK HOLD
PEAK HOLD
A
AC/DC MIN MAXREL
V
Figure 4. AC Voltage Measurement (318 shown)
aca14f.eps
10
Page 15
40
OFF
OFF
400
A
1000
A
PEAK HOLD
PEAK HOLD
A
AC/DC MIN MAXREL
Users Manual
V
TRUE RMS
318
CLAMP METER
Figure 5. DC Voltage Measurement (318 shown)
aca15f.eps
11
Page 16
312, 316, 318
Measuring Resistance
XWarning
To avoid electrical shock when measuring resistance in a circuit, make sure the power to the circuit is turned off and all capacitors are discharged.
1. Turn the rotary function switch to e. Remove power from the circuit being tested.
2. Connect the black test lead to the COM terminal and the red test lead to the Ve terminal.
3. Measure the resistance by touching the probes to the desired test points of the circuit.
4. View the reading on the LCD.
PEAK HOLD
PEAK HOLD
1000
A
400
A
A
40
OFF
OFF
AC/DC MIN MAXREL
318
TRUE RMS CLAMP METER
V
Figure 6. Measuring Resistance (318 shown)
aca16f.eps
12
Page 17
Users Manual
Testing Continuity
XWarning
To avoid electrical shock when testing continuity in a circuit, make sure the power to the circuit is turned off and all capacitors are discharged.
1. Turn the rotary function switch to D. Remove power from the
circuit being tested.
2. Connect the black test lead to the COM terminal and the red test lead to the Ve terminal.
3. Connect the probes across the circuit or component to be tested.
If the resistance is under 50 Ω, the beeper will sound continuously, designating a short circuit. If the Meter reads OL, the circuit is open.
PEAK HOLD
PEAK HOLD
1000
A
400
A
40
A
OFF
OFF
V
AC/DC MIN MAXREL
TRUE RMS
318
CLAMP METER
PEAK HOLD
PEAK HOLD
A
1000
400
A
40
A
OFF
OFF
V
AC/DC MIN MAXREL
TRUE RMS
318
CLAMP METER
Short Circuit Open Circuit
aca17f.eps
Figure 7. Measuring Continuity (318 shown)
13
Page 18
312, 316, 318
Diode Test
XWarning
To avoid electrical shock when testing diodes in a circuit, make sure the power to the circuit is turned off and all capacitors are discharged.
1. Turn the rotary function switch to D. Remove power from the
circuit being tested.
2. Connect the black test lead to the COM terminal and the red test lead to the Ve terminal.
3. Connect the black test lead to the cathode side and the red test lead to the anode side of the diode being tested.
4. Read forward bias voltage value on the LCD.
5. If the polarity of the test leads is the reverse of the diode polarity, the LCD reading shows “OL”. This can be used for distinguishing the anode side and cathode side of a diode.
PEAK HOLD
PEAK HOLD
A
1000
400
A
40
A
OFF
OFF
AC/DC MIN MAXREL
318
TRUE RMS CLAMP METER
V
Figure 8. Diode Test (318 shown)
aca18f.eps
14
Page 19
Users Manual
XWarning
Dangerous voltages may be present at the input terminals and may not be displayed.
Data Hold
To hold the present reading, press the HOLD button. Press the button again to return to the live reading.
Peak
The Meter has a 1 ms peak feature for use with ACA, DCA (316,
318), ACV, and DCV functions. Before making a peak hold measurement, the peak feature must be calibrated as follows:
1. Press and hold the peak button for 2 seconds. The CAL indicator appears on the LCD to show that the offset has been calculated and stored by the Meter.
2. After calibration, press the peak button once to enter Peak mode and begin displaying Peak+ values.
3. Press the Peak button again to enter Peak displaying Peak
4. Press the Peak button for 2 seconds to return to normal operation.
- values.
- mode and begin
+
If the rotary function switch is changed following calibration, the calibration steps must be repeated before making another peak measurement.
Note
The peak mode continuously updates the LCD to show maximum magnitudes. Smaller magnitudes are not displayed.
MIN MAX (316, 318 only)
1. Press M once to set the Meter to MAX mode.
2. Press
3. Press while keeping track of the MAX and MIN value. This mode is also indicated by the blinking MAXMIN symbol on the LCD.
4. Pressing to normal operation.
M again to set the Meter to MIN mode. M again and the Meter displays the present reading
M for more than 2 seconds sets the Meter back
15
Page 20
312, 316, 318
REL (S) (316, 318 only)
1. Press Q to set the zero to the present reading. The
difference between the reference reading and subsequent readings is displayed.
2. Press reading. The blinking S on the LCD also indicates this mode.
3. Press normal operation.
Q again and the Meter displays the reference
Q for more than 2 seconds to return the Meter to
Range (312 only)
Press R to select the manual range mode. Manual shows on the LCD, and the Meter remains in the same range. To exit manual-range mode and return to the auto-range mode, press and hold
R for 1 second.
Changing the Battery
When the battery voltage drops below the value required for proper operation, the battery symbol (B) appears on the LCD.
1. Before changing the battery, switch the Meter off and disconnect the test leads.
2. Open the back cover using a screwdriver.
3. Replace the battery with a new 9 V battery.
4. Close the back cover and fasten the screws.
Cleaning the Meter
Periodically clean the Meter by wiping it with a dry cloth. Do not use abrasives or solvents on this instrument.
Maintenance
XWarning
• To avoid electrical shock, disconnect test leads from
the Meter before removing its back cover. Never use the Meter with the back cover removed.
• Repairs or servicing not covered in this manual should
be performed only by qualified personnel.
Caution
• To avoid contamination or static damage, do not touch
the circuit board without proper static protection.
• If the Meter is not going to be used for a long time,
remove the battery. Do not store the Meter in a high temperature or a high humidity environment.
16