This manual provides information on the X-RAY kVp
measurement function of the Fluke 199XRAY Medical
ScopeMeter. It also describes user interface differences
between the Fluke 190B/C and the Fluke 199XRAY.
For other information on the Fluke 199XRAY please
consult the Fluke 190B/C Series Getting Started manual,
the Fluke 190B/C Series Users Manual (on CD-ROM),
and the Fluke 190B/C Medical Functions Users Manual
Supplement (on CD-ROM) that are included in the
shipment.
Safety Information
Read and comply with the safety instructions that you find
in the ScopeMeter 190B/C Series Getting Started manual
and in the Fluke 190B/C Users Manual .
Limited Warranty, Limitation of Liability
Read the Warranty and Limitation of Liability statements
that you find in the ScopeMeter 190B/C Users Manual
that is included in the test tool kit shipment (on CD ROM).
Shipment Contents
In addition to the shipment contents shown in the Fluke
192B-196B/C-199B/C Users Manual, the 199XRAY
shipment includes the following parts:
Part Ordering code
Power Adapter Plug UK 2726493
Power Adapter Plug EURO 2726507
Power Adapter Plug Australia 2726518
Power Adapter Plug USA 2726529
MA190 Accessory Set
See 190B/C Medical
Functions Users
Manual Supplement.
1-1
Fluke 199XRAY
Users Manual
About the X-RAY kVp Function
The 199XRAY Medical Scopemeter, when used with Fluke
Biomedical kVp Divider, automatically measure kVp
average and kV peak. In the radiographic measurement
mode, exposure time is also measured in milliseconds or
pulses.
The Fluke Biomedical Non-Invasive kVp Divider (Model
35080/A/B/M) is an alternative to high-voltage divider
tanks. It provides safe, non-invasive, fast measurement of
kVp without endangering personnel and/or equipment. Its
use is primarily intended for X-ray machine service and
calibration. The kVp Divider provides an analog output
voltage proportional to the instantaneous voltage applied
to the X-ray tube.
The X-RAY kVp function provides oscilloscope setups for
measurements using the kVp Divider and the following
filter packs on Input A :
• Wide Range Filter Pack
• K-edge Mammo Filter Pack
• Linear Mammo Filter Pack
• CT Filter Pack
• Mobile Pack Plus Filter Pack
• Low Range Filter Pack
For details on the kVp Divider refer to the 35080/A/B/M
Operators Manual.
1-2
General190B/C vs. 199XRAY User Interface
1
190B/C vs. 199XRAY User Interface
The 190B/C and the 199XRAY operating instructions are
equal, except for the following:
• The 199XRAY user interface text is available in English
only, language selection is not possible. See ‘USER
Button Bar ’ below.
• The 190B/C RECORDER functions can be selected by
pressing the RECORDER key.
The 199XRAY RECORDER functions can be selected
by pressing the MEDICAL key.
See ‘190B/C RECORDER Key’ and ‘199XRAY
MEDICAL Key’ below.
USER Button Bar
Pressing the 190B/C USER key will open
the button bar:
Pressing the 199XRAY USER key will
open the button bar without the
LANGUAGE selection:
1-3