FLUKE 199C, 199B, 196C, 196B, 192B User Manual

ScopeMeter®190C/190B Series
®
and ScopeMeter
120 Series
Technical Data
ScopeMeter 190C and 190B Series: Speed, performance and analysis power
E
G
M
A
E
N
N
A
T
M
S
Y
T
I
L
A
U
Q
C
E
R
T
I
F
I
E
I
D
T
T
E
O
E
M
Now expanded with 3K memory and FFT Frequency Spectrum Analysis
For demanding applications, the
Y
S
T
E
ScopeMeter 190C and 190B Series
M
high-performance oscilloscopes offer specifications usually found on
1
0
0
9
top-end bench instruments. With
O
S
up to 200 MHz bandwidth,
2.5 GS/s real time sampling and a deep memory of 27,500 points per
R
input they’re ideal for engineers who need the full capabilities of a high-performance scope in a hand­held, battery powered instrument.
Dual input - 200, 100 or
60 MHz bandwidth
Up to 2.5 GS/s real time
sampling per input High Waveform Resolution of
3000 datapoints per channel Frequency Spectrum using FFT
analysis (190C) Connect-and-View™automatic
triggering, a full range of man­ual trigger modes plus external triggering
Digital Persistence for analyzing
complex, dynamic signals like on an analog oscilloscope (190C)
Fast display update rate for
seeing dynamic behavior instantaneously
Automatic capture and replay of
100 screens 27,500 points and more per
input record length using ScopeRecord
1,000 V CAT II and 600 V CAT
III safety certified Up to 1,000 V independently
floating isolated inputs Four hours rechargeable
NiMH battery pack
mode
ScopeMeter 120 Series: Three-in-one simplicity
The compact ScopeMeter 120 Series is the rugged solution for industrial troubleshooting and installation applications. It’s a truly integrated test tool, with oscilloscope, multi­meter and “paperless” recorder in one affordable, easy-to-use instru­ment. Find fast answers to problems in machinery, instrumentation, control and power systems.
A dual input 40 MHz or 20 MHz
digital oscilloscope Two 5,000 counts true-rms
digital multimeters Cursor measurements (Fluke 124)
A dual input TrendPlot recorder
Connect-and-View trigger
simplicity for hands-off operation Shielded test leads for oscillo-
scope, resistance, continuity and capacitance measurements
Full bandw
40 MHz probe included standard with Fluke 124
Up to seven hours battery
operation 600 V CAT III safety certified
Optically isolated RS-232
interface Rugged, compact case
idth, VPS40 10:1
Technical Specifications 190C and 190B Series
Oscilloscope Mode
Vertical Deflection
Fluke 199C Fluke 196C
Bandwidth 200 MHz 100 MHz 60 MHz Rise time 1.7 ns 3.5 ns 5.8 ns
Bandwidth limiter: User selectable: 10 kHz, 20 MHz or off Number of inputs: 2 plus external trigger. All inputs isolated from each other and ground.
Input coupling: AC or DC, with ground level indicator Input sensitivity: 2 mV/div to 100 V/div (Fluke 190C
Series); 5 mV/div to 100 V/div (Fluke 190B Series) Normal/Invert: On both input channels; switched separately
Variable attenuator: Variable gain on input channel A Input voltage: 1000 V CAT II, 600 V CAT III rated -
See “general specifications” for further details.
Vertical resolution: 8 bit Accuracy: ± (1.5% of reading + 0.04 x range/div) Input impedance: 1 M± 1% // 15 pF ± 2 pF
Horizontal
Maximum 2.5 GS/s 1 GS/s 500 MS/s real-time sample rate
Number of 2 2 2 digitizers
Time base 5 ns/div 10 ns/div range to 5 s/div to 5 s/div
Maximum record length: 3000 points per input in Scope-mode; 27,500 points per input in ScopeRecord
Accuracy: ± (0.01 % of reading + 1 pixel) Glitch capture: 50 nsec (at 5 µsec/div to 1 min/div);
250 nsec (at 2 min/div)
Display and Acquisition
Display 144 mm 144 mm
Display modes Input A, Input B, dual, average, Replay Persistence Digital Persistence: Persistence
modes short/medium/long/infinite on / off
Visible screen width: 12 divisions Waveform mathematics: A + B, A - B, A * B, all
with user selectable scaling of resultant; A versus B (X - Y-mode); Frequency Spectrum using FFT analysis (190C only) Acquisition modes: Normal, auto, single shot, ScopeRecord compare, waveform compare with automatic “Pass / Fail testing” (in 199C and 196C only)
Trigger and Delay
Source: Input A, input B, external trigger input. All input references isolated from each other and from ground Modes: Automatic Connect-and-View, single shot, edge, delay, video, video line, selectable pulsewidth, dual slope (190C only), N-cycle (190C only)
Fluke 199B Fluke 196B Fluke 192B
Fluke 199C Fluke 196C Fluke 199B Fluke 196B Fluke 192B
roll mode (5 ms/div to 2 min/div)
Fluke 190C Fluke 190B
Full Color LCD Monochrome LCD
roll, glitch capture, waveform
free run,
Connect-and-View
: Advanced automatic triggering
that recognizes signal patterns, automatically sets up and continuously adjusts triggering, time base and amplitude. Automatically displays stable waveforms of complex and dynamic signals like motor drive and control signals. Video triggering: NTSC, PAL, PAL+, SECAM. Includes field 1, field 2 and line select. Pulse width triggering: Pulse width qualified by time. Allows for triggering <t, >t, =t, t, where t is selec­table in minimal steps of 0.01 div. or 50 nsec. Time delay: One full screen of pre-trigger view or up to 100 screens (= 1200 divisions) of post-trigger delay Dual slope triggering: Both rising and falling transi­tions, when crossing the trigger level, initiate an acquisition (190C only) N-cycle triggering: Triggers on N-th occurrence of a trigger event; N to be set in the range 2 to 99 (190C only)
Automatic Capture of 100 Screens
The instrument ALWAYS memorizes the last 100 screens (no user setup required). When an anomaly occurs on screen, the REPLAY button can be pressed to review the full screen sequence over and over.
Instrument can be set up for triggering on glitches or intermittent anomalies and will operate in “baby-sit” mode and will capture 100 events.
Alternatively, the 199C and 196C can be set up in waveform compare mode to store only matching (“Pass”) or only non-matching (“Fail”) acquired wave­forms in the replay memory bank for further analysis. Replay: Manual or continuous replay. Displays the captured 100 screens as a “live” animation or under manual control. Each screen has date- and time-stamp. Replay storage: Up to 2 sets of 100 screens each can be saved for later recall and analysis
FFT - Frequency Spectrum Analysis (190C only)
Shows frequency content of oscilloscope waveform using Fast Fourier Transform
Window: Automatic, Hamming, Hanning or None Automatic window: Digitally re-samples acquired
waveform to get optimum frequency resolution in FFT resultant
Vertical scale: Linear / Logarithmic, in volts Frequency axis: Logarithmic; frequency range auto-
matically set as function of timebase range of oscillo­scope
Waveform Compare and Pass/Fail Testing
Waveform compare: Provides storage and display of a reference waveform for visual comparison with newly acquired waveforms. Reference is derived from an acquired waveform and can be modified in the ScopeMeter or externally using FlukeView Pass/Fail Testing (199C, 196C): In waveform compare mode, the Color Scopemeter can be set up to store only matching (“Pass”) or only non-matching (“Fail”) acquired waveforms in the replay memory bank for further analysis.
®
Software.
2 Fluke Corporation ScopeMeter 190C/190B Series and ScopeMeter 120 Series
Automatic Scope Measurements
Vdc, Vac rms, Vac+dc, Vpeak max, Vpeak min, Vpeak to peak, Aac, Adc, Aac+dc, frequency (Hz), risetime, fall­time, power factor, Watts, VA, VA reactive, phase, pulse width (pos/neg), dutycycle (pos/neg), temperature °C, temperature °F, dBV, dBm into 50 and 600
Vpwm ac, Vpwm ac+dc for measurement on pulse
width modulated motordrives and frequency inverters
Cursor Measurements
Source: Input A, input B or the Mathematical Result trace (excl. A vs B curve) Dual horizontal lines: Voltage at cursor 1 and 2, voltage between cursors Dual vertical lines: Time between cursors, 1/T between cursors (in Hz), voltage between markers, risetime with markers, falltime with markers; Vrms between cursors (190C only), watts between cursors (190C only) Single vertical line: Min-Max and Average voltage at cursor position; frequency and rms-value of individ­ual frequency component in FFT result (190C only) Zoom: Up to 16x horizontal zoom
Recorder Mode
ScopeRecord-Roll Mode
Dual input waveform storage mode
Source and display: Input A, Input B, Dual Memory depth: 27,500 points per input. Each point
consists of Min-Max pair Min-Max values: Min-Max values are measured at high sample rates ensuring capture and display of glitches
Time base range 5 ms/div to 1 min/div 2 min/div
Recorded timespan 6 sec to 24 hr 48 hr Glitch capture 50 ns 250 ns Sample rate 20 MS/s 4 MS/s Resolution 200 µsec to 2 sec 4.8 sec
Recording modes: Single sweep, continuous roll, Start-on-Trigger, Stop-on-Trigger Stop-on-Trigger (through external): ScopeRecord mode can be stopped by an individual trigger event, or by an interruption of a repetitive trigger signal
Horizontal scale: Time from start, time of day Zoom: Up to 100x Memory: Up to 2 dual input ScopeRecord waveforms
can be saved for later recall and analysis
TrendPlot™recording
Single or dual input electronic paperless chart recorder. Plots, displays and stores meter and scope measure­ments.
Source and display: Input A, Input B or DMM input
Memory depth: 18,000 points recording. Per record
point a minimum, a maximum and an average value, plus a date and timestamp are recorded. Ranges: 5 s/div to 30 min/div in normal view mode; 5 min/div to 48 hr/div in view all mode, giving overview of total record Recorded timespan: Up to 22 days with a resolution of 1 minute Recording mode: Continuous roll for the duration of the full recordable timespan
Measurement speed: 5 measurements per second Horizontal scale: Time from start, time of day Zoom: Up to 64x zoom Memory: Up to 2 TrendPlot recordings can be saved
for later recall and analysis
Cursor measurements - all recorder modes
Source: Input A, B or DMM input Dual vertical lines: Min-Max or Average
voltage. Time between cursors Single vertical line: Min-Max or Average voltage. Absolute date and time or time from start
Meter Mode
Via 4 mm banana inputs. Fully isolated from scope inputs and scope ground. The specified accuracy is valid over the temperature range 18 °C to 28 °C (65 °F to 82 °F). Add 10 % of specified accuracy for each degree C below 18 °C or above 28 °C.
Maximum Resolution: 5,000 counts Voltmeter Ranges: 500 mV, 5 V, 50 V, 500 V, 1,000 V Accuracy:
Vdc ± (0.5 % + 5 counts) Vac true rms
15 Hz...60 Hz: ± (1 % + 10 counts) 60 Hz...1 kHz: ± (2.5 % + 15 counts)
Vac+dc true rms
dc...60 Hz: ± (1 % + 10 counts) 60 Hz...1 kHz: ± (2.5 % + 15 counts)
Ohms: Ranges: 500 , 5 k, 50 k, 500 k, 5 M, 30 M Accuracy: ± (0.6 % + 5 counts)
Other Meter Functions
Continuity: Beeper on < 50 (± 30 Ω) Diode test: Up to 2.8 V Amps: Adc, Aac, Aac+dc using an optional current
clamp or shunt. Scaling factors: 0.1 mV/A ... 100 V/A Temperature (°C, °F): With optional accessories. Scale factors 1 mV/°C or 1 mV/°F
Input impedance: 1 M± 1% // 10 pF ± 2 pF Advanced meter functions: Auto/manual ranging,
relative measurements (Zero reference), TrendPlot recording
ScopeMeter 190C/190B Series and ScopeMeter 120 Series Fluke Corporation 3