Fluke 199C User Manual

Fluke 196C/199C
4822 872 30482 October 2001
© 2001 Fluke Corporation. All rights reserved. Printed in the Netherlands. All product names are trademarks of their respective companies.
ScopeMeter
Users Manual
LIMITED WARRANTY & LIMITATION OF LIABILITY
Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price. Fluke reserves the right to invoice Buyer for importation costs of repair/replacement parts when product purchased in one country is submitted for repair in another country.
Fluke's warranty obligation is limited, at Fluke's option, to refund of the purchase price, free of charge repair, or replacement of a defective product which is returned to a Fluke authorized service center within the warranty period.
To obtain warranty service, contact your nearest Fluke authorized service center or send the product, with a description of the difficulty, postage and insurance prepaid (FOB Destination), to the nearest Fluke authorized service center. Fluke assumes no risk for damage in transit. Following warranty repair, the product will be returned to Buyer, transportation prepaid (FOB Destination). If Fluke determines that the failure was caused by misuse, alteration, accident or abnormal condition of operation or handling, Fluke will provide an estimate of repair costs and obtain authorization before commencing the work. Following repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges (FOB Shipping Point).
THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHER ARISING FROM BREACH OF WARRANTY OR BASED ON CONTRACT, TORT, RELIANCE OR ANY OTHER THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision.
Fluke Corporation, P.O. Box 9090, Everett, WA 98206-9090 USA, or
Fluke Industrial B.V., P.O. Box 90, 7600 AB, Almelo, The Netherlands
SERVICE CENTERS
To locate an authorized service center, visit us on the W orld Wide Web:
http://www.fluke.com
or call Fluke using any of the phone numbers listed below:
+1-888-993-5853 in U.S.A. and Canada
+31-40-2675200 in Europe
+1-425-446-5500 from other countries
Table of Contents
Chapter Title Page
Unpacking the Test Tool Kit........................................................................................ 2
Safety Information: Read First ....................................................................................... 4
1 Using The Scope .......................................................................................................... 7
Powering the Test Tool .................................................................................................. 7
Resetting the Test Tool.................................................................................................. 8
Navigating a Menu ......................................................................................................... 9
Hiding Key Labels and Menus ....................................................................................... 10
Input Connections .......................................................................................................... 10
Making Scope Connections ........................................................................................... 11
Displaying an Unknown Signal with Connect-and-View™ ............................................. 12
Making Automatic Scope Measurements....................................................................... 13
Freezing the Screen....................................................................................................... 14
Using Average, Persistence and Glitch Capture............................................................ 15
Acquiring Waveforms ..................................................................................................... 18
i
Fluke 196C/199C
Users Manual
Analyzing Waveforms .................................................................................................... 22
2 Using The Multimeter .................................................................................................. 23
Making Meter Connections ............................................................................................ 23
Making Multimeter Measurements................................................................................. 24
Freezing the Readings................................................................................................... 27
Selecting Auto/Manual Ranges...................................................................................... 27
Making Relative Measurements..................................................................................... 28
3 Using The Recorder Functions................................................................................... 29
Opening the Recorder Main Menu................................................................................. 29
Plotting Measurements Over Time (TrendPlot™).......................................................... 30
Recording Scope Waveforms In Deep Memory (Scope Record) .................................. 33
Analyzing a TrendPlot or Scope Record........................................................................ 36
4 Using Replay, Zoom and Cursors............................................................................... 37
Replaying the 100 Most Recent Scope Screens ........................................................... 37
Zooming in on a Waveform............................................................................................ 40
Making Cursor Measurements....................................................................................... 42
5 Triggering on Waveforms............................................................................................ 45
Setting Trigger Level and Slope..................................................................................... 46
Using Trigger Delay or Pre-trigger ................................................................................. 47
Automatic Trigger Options ............................................................................................. 48
Triggering on Edges....................................................................................................... 49
Triggering on External Waveforms ................................................................................ 52
ii
Contents (continued)
Triggering on Video Signals........................................................................................... 53
Triggering on Pulses ...................................................................................................... 55
6 Using Memory, PC and Printer.................................................................................... 59
Saving and Recalling ..................................................................................................... 59
Documenting Screens.................................................................................................... 63
7 Tips................................................................................................................................ 67
Using the Standard Accessories.................................................................................... 67
Using the Independently Floating Isolated Inputs .......................................................... 70
Using the Tilt Stand........................................................................................................ 72
Resetting the Test Tool.................................................................................................. 72
Suppressing Key Labels and Menu’s............................................................................. 72
Changing the Information Language.............................................................................. 73
Adjusting the Contrast and Brightness........................................................................... 73
Changing the Display Color ........................................................................................... 74
To set the display to color or Black and White, do the following:................................... 74
Changing Date and Time ............................................................................................... 74
Saving Battery Life ......................................................................................................... 75
Changing the Auto Set Options...................................................................................... 76
8 Maintaining the Test Tool............................................................................................ 77
Cleaning the Test Tool................................................................................................... 77
Storing the Test Tool...................................................................................................... 77
Charging the Batteries ................................................................................................... 78
Extending Battery Operation Time ................................................................................. 79
Replacing the NiMH Battery Pack BP190 ...................................................................... 80
Calibrating the Voltage Probes ...................................................................................... 80
iii
Fluke 196C/199C
Users Manual
Displaying Calibration Information ................................................................................. 82
Parts and Accessories ................................................................................................... 82
Troubleshooting ............................................................................................................. 87
9 Specifications............................................................................................................... 89
Introduction ....................................................................................................................89
Dual Input Oscilloscope ................................................................................................. 90
Automatic Scope Measurements................................................................................... 92
Meter.............................................................................................................................. 96
DMM Measurements on Meter Inputs............................................................................ 96
Recorder ........................................................................................................................98
Zoom, Replay and Cursors ............................................................................................ 99
Miscellaneous ................................................................................................................ 99
Environmental................................................................................................................ 101
Safety ..................................................................................................................... 102
10:1 Probe ..................................................................................................................... 103
Electromagnetic Immunity.............................................................................................. 105
iv
Unpacking the Test Tool Kit
Declaration of Conformity
for
Fluke 196C/199C
®
ScopeMeter
Manufacturer
Fluke Industrial B.V.
7602 EA Almelo
The Netherlands
Statement of Conformity
Based on test results using appropriate standards,
the product is in conformity with
Electromagnetic Compatibility Directive 89/336/EEC
Low Voltage Directive 73/23/EEC
test tools
Lelyweg 1
Sample tests
Standards used:
EN 61010.1 (1993)
Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use
EN-IEC61326-1 (1997)
Electrical equipment for
measurements and laboratory
use -EMC requirements-
The tests have been performed in a
typical configuration.
This Conformity is indicated by the symbol
i.e. “Conformité Européenne”.
,
1
Fluke 196C/199C
Users Manual
Unpacking the Test Tool Kit
The following items are included in your test tool kit:
Note
When new, the rechargeable NiMH battery is not fully charged. See Chapter 8.
Figure 1. ScopeMeter Test Tool Kit
2
Unpacking the Test Tool Kit
# Description
1 ScopeMeter Test Tool
2 Battery Charger (country dependent)
3 10:1 Voltage Probe Set (red)
a) 10:1 Voltage Probe (red) b) Hook Clip for Probe Tip (red) c) Ground Lead with Hook Clip (red) d) Ground Lead with Mini Alligator Clip (black) e) 4-mm Test Probe for Probe Tip (red) f) Ground Spring for Probe Tip (black)
4 10:1 Voltage Probe Set (gray)
a) 10:1 Voltage Probe (gray) b) Hook Clip for Probe Tip (gray) c) Ground Lead with Hook Clip (gray) d) Ground Lead with Mini Alligator Clip (black)
e) 4-mm Test Probe for Probe Tip (gray)
5 Test Leads (red and black)
6 Accessory Set
a) 2-mm Test Probe for Probe Tip (red) b) Industrial Alligator for Probe Tip (red) c) 2-mm Test Probe for Banana Jack (red) d) Industrial Alligator for Banana Jack (red) e) Ground Lead with 4-mm Banana Jack (black)
# Description
7 Accessory Set
a) 2-mm Test Probe for Probe Tip (gray) b) Industrial Alligator for Probe Tip (gray) c) 2-mm Test Probe for Banana Jack (gray) d) Industrial Alligator for Banana Jack (gray) e) Ground Lead with 4-mm Banana Jack (black)
8 Getting Started Manual
9 CD ROM with Users Manual (multi-language)
10 Shipment box (basic version only)
Fluke 196C-S and 199C-S versions include also the following items:
# Description
11 Optically Isolated RS-232 Adapter/Cable
12 FlukeView® ScopeMeter® Software for
Windows
®
13 Hard Case
3
Fluke 196C/199C
Users Manual
Safety Information: Read First
Carefully read the following safety information before using the test tool.
Specific warning and caution statements, where they apply, appear throughout the manual.
A “Warning” identifies conditions and actions that pose hazard(s) to the user.
A “Caution” identifies conditions and actions that may damage the test tool.
The following international symbols are used on the test tool and in this manual:
See explanation in manual
Disposal information Earth
Recycling information
Safety Approval Safety Approval
Direct Current Alternating Current
Double Insulation (Protection Class)
Conformité Européenne
Warning
To avoid electrical shock or fire:
Use only the Fluke power supply, Model
BC190 (Battery Charger / Power Adapter).
Before use check that the selected/indicated
range on the BC190 matches the local line power voltage and frequency.
For the BC190/808 universal Battery Charger /
Power Adapter) only use line cords that comply with the local safety regulations.
Note:
To accomodate connection to various line power sockets, the BC190/808 universal Battery Charger / Power Adapter is equipped with a male plug that must be connected to a line cord appropriate for local use. Since the adapter is isolated, the line cord does not need to be equipped with a terminal for connection to protective ground. Since line cords with a protective grounding terminal are more commonly available you might consider using these anyhow.
4
Safety Information: Read First
Warning
To avoid electrical shock or fire if a test tool input is connected to more than 42 V peak (30 Vrms) or on circuits of more than 4800 VA:
Use only insulated voltage probes, test leads
and adapters supplied with the test tool, or indicated by Fluke as suitable for the Fluke190 ScopeMeter series.
Before use, inspect voltage probes, test leads
and accessories for mechanical damage and replace when damaged.
Remove all probes, test leads and accessories
that are not in use.
Always connect the battery charger first to the
ac outlet before connecting it to the test tool.
Do not connect the ground spring (figure 1,
item f) to voltages higher than 42 V peak (30 Vrms) from earth ground.
Do not apply voltages that differ more than 600
V from earth ground to any input when measuring in a CAT III environment. Do not apply voltages that differ more than 1000 V from earth ground to any input when measuring in a CAT II environment.
Do not apply voltages that differ more than 600
V from each other to the isolated inputs when measuring in a CAT III environment. Do not apply voltages that differ more than 1000 V from each other to the isolated inputs when measuring in a CAT II environment.
Do not apply input voltages above the rating of
the instrument. Use caution when using 1:1 test leads because the probe tip voltage will be directly transmitted to the test tool.
Do not use exposed metal BNC or banana plug
connectors.
Do not insert metal objects into connectors.
Always use the test tool only in the manner
specified.
Voltage ratings that are mentioned in the warnings, are given as limits for “working voltage”. They represent V ac rms (50-60 Hz) for ac sinewave applications and as V dc for dc applications.
Overvoltage Category III refers to distribution level and fixed installation circuits inside a building. Overvoltage Category II refers to local level, which is applicable for appliances and portable equipment.
5
Fluke 196C/199C
Users Manual
The terms ‘Isolated’ or ‘Electrically floating’ are used in this manual to indicate a measurement in which the test tool input BNC or banana jack is connected to a voltage different from earth ground.
The isolated input connectors have no exposed metal and are fully insulated to protect against electrical shock.
The red and gray BNC jacks, and the red and black 4-mm banana jacks can independently be connected to a voltage above earth ground for isolated (electrically floating) measurements and are rated up to 1000 Vrms CAT II and 600 Vrms CAT III above earth ground.
If Safety Features are Impaired
Use of the test tool in a manner not specified may impair the protection provided by the equipment.
Before use, inspect the test leads for mechanical damage and replace damaged test leads!
Whenever it is likely that safety has been impaired, the test tool must be turned off and disconnected from the line power. The matter should then be referred to qualified personnel. Safety is likely to be impaired if, for example, the test tool fails to perform the intended measurements or shows visible damage.
6
About this Chapter
This chapter provides a step-by-step introduction to the scope functions of the test tool. The introduction does not cover all of the capabilities of the scope functions but gives basic examples to show how to use the menus and perform basic operations.
Powering the Test Tool
Follow the procedure (steps 1 through 3) in Figure 2 to power the test tool from a standard ac outlet. See Chapter 8 for instructions on using battery power.
Turn the test tool on with the on/off key.
Chapter 1

Using The Scope

The test tool powers up in its last setup configuration.
Figure 2. Powering the Test Tool
7
Fluke 196C/199C
Users Manual
Resetting the Test Tool
If you want to reset the test tool to the factory settings, do the following:
1 Turn the test tool off.
2 Press and hold the USER key.
3 Press and release.
The test tool turns on, and you should hear a double beep, indicating the reset was successful.
4 Release the USER key.
Now look at the display; you will see a screen that looks like Figure 3.
8
Figure 3. The Screen After Reset
Navigating a Menu
The following example shows how to use the test tool's menus to select a function. Subsequently follow steps 1 through 4 to open the scope menu and to choose an item.
1 Press the SCOPE key to display
the labels that define the present use for the four blue function keys at the bottom of the screen.
Using The Scope
Navigating a Menu
Figure 4. Basic Navigation
1
Note
To hide the labels for full screen view, press the
SCOPE key again. This toggling enables you to
check the labels without affecting your settings.
2 Open the Waveform Options
menu. This menu is displayed at the bottom of the screen.
3a
3b
4 Press the ENTER key until you exit
Repeatedly pressing through a menu without changing the settings.
Use the blue arrow keys to highlight the item. Press the blue ENTER key to accept the selection.
the menu.
Note
lets you to step
9
Fluke 196C/199C
Users Manual
Hiding Key Labels and Menus
You can hide a menu or key label at any time:
Press the CLEAR MENU key to hide any key label or menu.
To display menus or key labels, press one of the yellow menu keys, e.g. the
SCOPE key.
Input Connections
Look at the top of the test tool. The test tool has four signal inputs: two safety BNC jack inputs (red input A and gray input B) and two safety 4-mm banana jack inputs (red and black). Use the two BNC jack inputs for scope measurements, and the two banana jack inputs for meter measurements.
Isolated input architecture allows independent floating measurements with each input.
Figure 5. Measurement Connections
10
Making Scope Connections
To make dual input scope measurements, connect the red voltage probe to input A, and the gray voltage probe to input B. Connect the short ground leads of each voltage probe to its own reference potential. (See Figure
6.)
Using The Scope
Making Scope Connections
1
Note
To maximally benefit from having independently isolated floating inputs and to avoid problems caused by improper use, read Chapter 7: “Tips”.
Figure 6. Scope Connections
11
Fluke 196C/199C
Users Manual
Displaying an Unknown Signal with Connect-and-View™
The Connect-and-View feature lets the test tool display complex, unknown signals automatically. This function optimizes the position, range, time base, and triggering and assures a stable display of virtually any waveform. If the signal changes, the setup is automatically adjusted to maintain the best display result. This feature is especially useful for quickly checking several signals.
To enable the Connect-and-View feature, do the following:
1 Perform an Auto Set. AUTO appears at
the top right of the screen.
The bottom line shows the range, the time base, and the trigger information.
The waveform identifier (A) is visible on the bottom right side of the screen, as shown in Figure 7. The input A zero
) at the left side of the screen identifies the ground
icon (
-
level of the waveform.
2 Press a second time to select the
manual range again. MANUAL appears at the top right of the screen.
12
Figure 7. The Screen After an Auto Set
Use the light-gray RANGE, TIME and MOVE keys at the bottom of the keypad to change the view of the waveform manually.
Using The Scope
Making Automatic Scope Measurements
1
Making Automatic Scope Measurements
The test tool offers a wide range of automatic scope measurements. You can display two numeric readings:
EADING 1 and READING 2. These readings are selectable
R
independently, and the measurements can be done on the input A or input B waveform
To choose a frequency measurement for input A, do the following:
1 Display the SCOPE key labels.
2 Open the Reading 1 menu.
3 Select on A. Observe that the
highlight jumps to the present measurement.
4 Select the Hz measurement.
Observe that the top left of the screen displays the Hz measurement. (See Figure 8.)
To choose also a Peak-Peak measurement for Input B as second reading, do the following:
1 Display the SCOPE key labels.
2
3
4
5
Open the Reading 2 menu.
Select on B. The highlight jumps to the measurements field.
Open the PEAK menu.
Select the Peak-Peak measurement.
13
Fluke 196C/199C
Users Manual
Figure 8 shows an example of the screen. Note that the Peak-Peak reading for input B appears next to the input A frequency reading at the top of the screen.
Figure 8. Hz and V peak-peak as Scope Readings
Freezing the Screen
You can freeze the screen (all readings and waveforms) at any time.
1 Freeze the screen. HOLD appears
at the right of the reading area.
2 Resume your measurement.
14
Using The Scope
Using Average, Persistence and Glitch Capture
1
Using Average, Persistence and Glitch Capture
Using Average for Smoothing Waveforms
To smooth the waveform, do the following:
1 Display the SCOPE key labels.
2 Open the Waveform Options
menu.
3 Jump to Average:
4 Select On... to open the Average
Factors menu
5 Select Average 64.This averages
the outcomes of 64 acquisitions.
6 Exit the menu.
You can use the average functions to suppress random or uncorrelated noise in the waveform without loss of bandwidth. Waveform samples with and without smoothing are shown in Figure 9.
Figure 9. Smoothing a Waveform
15
Fluke 196C/199C
Users Manual
Using Persistence to Display Waveforms
You can use Persistence to observe dynamic signals.
1 Display the SCOPE key labels.
2 Open the Waveform Options
menu.
3 Jump to Waveform: and open the
Persistence... menu.
4 Select
Digital Persistence: Infinite, Envelope: Off.
Start monitoring the waveform.
Figure 10. Using Persistence to Observe Dynamic
Signals
When you select Envelope: On, the test tool displays the upper and lower boundaries of dynamic waveforms.
16
Using The Scope
Using Average, Persistence and Glitch Capture
1
Displaying Glitches
To capture glitches on a waveform, do the following:
1 Display the SCOPE key labels.
2 Open the Waveform Options
menu.
3 Select Glitch Detect: On
4 Exit the menu.
You can use this function to display events (glitches or other asynchronous waveforms) of 50 ns (nanoseconds) or wider, or you can display HF modulated waveforms.
Suppressing High Frequency Noise
Switching Glitch Detect to Off will suppress the high frequency noise on a waveform. Averaging will suppress the noise even more.
1 Display the SCOPE key labels.
2 Open the Waveform Options
menu.
3 Select Glitch Detect: Off, then
select Average: On to open the
Average menu
4 Select Factor : 8x
Tip
Glitch capture and average do not affect bandwidth. Further noise suppression is possible with bandwidth limiting filters. See Chapter 1: “Working with Noisy Waveforms”.
17
Fluke 196C/199C
Users Manual
Acquiring Waveforms
Selecting AC-Coupling
After a reset, the test tool is dc-coupled so that ac and dc voltages appear on the screen.
Use ac-coupling when you wish to observe a small ac signal that rides on a dc signal. To select ac-coupling, do the following:
1 Display the INPUT A key labels.
2 Highlight AC.
Observe that the bottom left of the screen displays the ac-coupling icon:
.
Reversing the Polarity of the Displayed Waveform
To invert the input A waveform, do the following:
1 Display the INPUT A key labels.
2 Open the Input A menu.
3 Select Inverted and accept
inverted waveform display.
4 Exit the menu.
For example, a negative-going waveform is displayed as positive-going waveform which may provide a more meaningful view. An inverted display is identified by an inversed trace identifier (
) at the right of the waveform.
18
Using The Scope
Acquiring Waveforms
1
Variable Input Sensitivity
The variable input sensitivity allows you to adjust the sensitivity continuously, for example to set the amplitude of a reference signal to exactly 6 divisions.
The input sensitivity of a range can be increased up to 2.5 times, for example between 10 mV/div and 4 mV/div in the 10 mV/div range.
To use the variable input sensitivity, do the following:
1 Apply the input signal
2 Perform an Auto Set (AUTO must
appear at the top of the screen)
An Auto Set will turn off the variable input sensitivity. You can now select the required input range. Keep in mind that the sensitivity will increase when you start adjusting the variable sensitivity (the displayed trace amplitude will increase).
3 Display the INPUT A key labels.
4 Open the Input A Options...
menu.
5 Select and accept Variable.
6 Exit the menu.
At the bottom left of the screen the text A Var is displayed.
Selecting Variable will turn off cursors and automatic input ranging.
7 Press mV to increase the
sensitivity, press V to decrease the sensitivity.
19
Fluke 196C/199C
Users Manual
Working with Noisy Waveforms
To suppress high frequency noise on waveforms, you can limit the working bandwidth to 10 kHz or 20 MHz. This function smoothes the displayed waveform. For the same reason, it improves triggering on the waveform.
To choose HF reject, do the following:
1 Display the INPUT A key labels.
2 Open the Input A menu.
3 Jump to Bandwidth.
4 Select 10kHz (HF reject) to
accept the bandwidth limitation.
Tip
To suppress noise without loss of bandwidth, use the average function or turn off Display
Glitches.
Using Waveform Mathematics Functions
When adding (A+B), subtracting (A-B), or multiplying (A*B) the input A and input B waveform, the test tool will display the mathematical result waveform and the input A and input B waveforms.
A versus B provides a plot with input A on the vertical axis and input B on the horizontal axis.
The Mathematics functions perform a point-to-point operation on waveforms A and B.
To use a Mathematics function, do the following:
1 Display the SCOPE key labels.
2 Open the Waveform Options
menu.
20
Using The Scope
Acquiring Waveforms
1
3 Jump to Waveform: and Select
Mathematics... to open the Mathematics menu.
4 Select Function: A+B, A-B, A*B or
A vs B.
5 Select a scale factor to fit the
mathematical result waveform onto the display, and return.
The sensitivity range of the mathematical result is equal to the sensitivity range of the least sensitive input divided by the scale factor.
Comparing Waveforms
You can display a fixed reference waveform with the actual waveform for comparison.
To create a reference waveform and to display it with the actual waveform, do the following:
1 Display the SCOPE key labels.
2 Open the Waveform Options
menu.
3 Jump to Reference: and accept
New. The momentary waveform is
stored and permanently displayed. The display also shows the actual waveform.
To recall a saved waveform from memory and use it as a reference waveform refer to Chapter 6 Recalling Screens with Associated Setups.
21
Fluke 196C/199C
Users Manual
Analyzing Waveforms
You can use the analysis functions CURSOR, ZOOM and
REPLAY to perform detailed waveform analysis. These
functions are described in Chapter 4: “Using Cursors, Zoom and Replay”.
22
About this Chapter
This chapter provides a step-by-step introduction to the multimeter functions of the test tool (hereafter called “meter”). The introduction gives basic examples to show how to use the menus and perform basic operations.
Making Meter Connections
Use the two 4-mm safety red ( ) and black (COM) banana jack inputs for the Meter functions. (See Figure 11.)
Note
Typical use of the Meter test leads and accessories is shown in Chapter 7.
Chapter 2

Using The Multimeter

Figure 11. Meter Connections
23
Fluke 196C/199C
Users Manual
Making Multimeter Measurements
The screen displays the numeric readings of the measurements on the meter input.
Measuring Resistance Values
To measure a resistance, do the following:
1 Connect the red and black test leads from the
4-mm banana jack inputs to the resistor.
The resistor value is displayed in ohms. Observe also that the bargraph is displayed. (See Figure 12.)
24
2
Display the METER key labels.
3
Open the Measurement menu.
Figure 12. Resistor Value Readings
4
Highlight Ohms.
5
Select Ohms measurement.
Making a Current Measurement
You can measure current in both Scope mode and Meter mode. Scope mode has the advantage of two waveforms being displayed while you perform measurements. Meter mode has the advantage of high measurement resolution.
The next example explains a typical current measurement
in Meter mode.
Using The Multimeter
Making Multimeter Measurements
Figure 13. Measurement Setup
2
Warning
Carefully read the instructions about the current probe you are using.
To set up the test tool, do the following:
1 Connect a current probe (e.g. i400, optional)
from the 4-mm banana jack outputs to the conductor to be measured.
Ensure that the red and black probe connectors correspond to the red and black banana jack inputs. (See Figure 13.)
2
Display the METER key labels.
3 Open the Measurement menu.
4
5
Highlight A ac....
Open the Current Probe submenu.
25
Fluke 196C/199C
Users Manual
6
7
Now, you will see a screen like in Figure 14.
Observe the sensitivity of the current probe. Highlight the corresponding sensitivity in the menu, e.g. 10 mV/A.
Accept the current measurement.
Figure 14. Ampere Measurement Readings
26
Using The Multimeter
Freezing the Readings
2
Freezing the Readings
You can freeze the displayed readings at any time.
1 Freeze the screen. HOLD appears
at the top right of the reading area.
2 Resume your measurement.
You can use this function to hold accurate readings for later examination.
Note
For saving screens into memory, see Chapter 6.
Selecting Auto/Manual Ranges
To activate manual ranging, do the following during any Meter measurement:
1 Activate manual ranging.
2 Increase or decrease the range.
Observe how the bargraph sensitivity changes.
Use manual ranging to set a fixed bargraph sensitivity and decimal point.
3 Choose auto ranging again.
When in auto ranging, the bargraph sensitivity and decimal point are automatically adjusted while checking different signals.
27
Fluke 196C/199C
Users Manual
Making Relative Measurements
A relative measurement displays the present measurement result relative to a defined reference value.
The following example shows how to perform a relative voltage measurement. First obtain a reference value:
1 Display the METER key labels.
2 Measure a voltage to be used as
reference value.
3 Set RELATIVE to ON. (ON is
highlighted.)
This stores the reference value as reference for subsequent measurements. The stored reference value is displayed in small digits at the bottom right side of the screen after the word
4 Measure the voltage to be
REFERENCE.
compared to the reference.
Observe that the main reading is displayed as variations from the reference value. The actual reading with its bargraph is displayed beneath these readings. (See Figure 15.)
Figure 15. Making a Relative Measurement
You can use this feature when, for example, you need to monitor input activity (voltage, resistance, temperature) in relation to a known good value.
28
About this Chapter
This chapter provides a step-by-step introduction to the recorder functions of the test tool. The introduction gives examples to show how to use the menus and perform basic operations.
Opening the Recorder Main Menu
First choose a measurement in scope or meter mode. Now you can choose the recorder functions from the recorder main menu. To open the main menu, do the following:
1 Open the RECORDER main menu.
(See Figure 16.)
Chapter 3

Using The Recorder Functions

Figure 16. Recorder Main Menu
29
Fluke 196C/199C
Users Manual
Plotting Measurements Over Time (TrendPlot™)
Use the TrendPlot function to plot a graph of Scope or Meter measurements as function of time.
Note
Because the navigations for the dual input TrendPlot (Scope) and the single input TrendPlot (Meter) are identical, only TrendPlot (Scope) is explained in the next sections.
Starting a TrendPlot Function
To start plotting a graph of the reading over time, do the following:
1 Apply a signal to the red BNC input A and turn
on Reading 1 in scope mode
2 Open the RECORDER main menu.
3 Highlight Trend Plot (Scope).
4 Start the TrendPlot recording.
The test tool continuously records the digital readings of the input A measurements and displays these as a graph. The TrendPlot graph rolls from right to left like a paper chart recorder.
Observe that the recorded time from start appears at the bottom of the screen. The present reading appears on top of the screen. (See Figure 17.)
Note
When simultaneously TrendPlotting two readings, the screen area is split into two sections of four divisions each.
30
Using The Recorder Functions
Plotting Measurements Over Time (TrendPlot™)
Displaying Recorded Data
When in normal view (NORMAL), only the nine most recently recorded divisions are displayed on screen. All previous recordings are stored in memory.
VIEW ALL shows all data in memory:
7 Display an overview of the full
waveform.
Press repeatedly to toggle between normal view
NORMAL) and overview (VIEW ALL)
(
3
Figure 17. TrendPlot Reading
When the Scope is in automatic mode, automatic vertical scaling is used to fit the TrendPlot graph on the screen.
5 Set RECORDER to STOP to freeze
the recorder function.
6 Set RECORDER to RUN to continue.
When the recorder memory is full, an automatic compression algorithm is used to compress all samples into half of the memory without loss of transients. The other half of the recorder memory is free again to continue recording.
31
Fluke 196C/199C
Users Manual
Changing the Recorder Options
At the right bottom of the display you can choose to display the time elapsed from start and the actual time of the day.
To change the time reference, proceed from step 6 as follows:
7 Open the Recorder Options
menu.
8 Select Time of Day and jump to
the next field.
Now the recorded time and the current time appear at the bottom of the screen.
The options Reading 1 and Reading 2 allow you to select the scope readings to be recorded. (Or one meter measurement when in TrendPlot meter mode.)
To continue without making further changes:
9 Close.
Turning Off the TrendPlot Display
10 Exit the recorder function.
32
Recording Scope Waveforms In Deep Memory (Scope Record)
Recording Scope Waveforms In Deep Memory (Scope Record)
The SCOPE RECORD function is a roll mode that logs one or two long waveforms. This function can be used to monitor waveforms like motion control signals or the power-on event of an Uninterruptable Power Supply (UPS). During recording, fast transients are captured. Because of the deep memory, recording can be done for more than one day. This function is similar to the roll mode in many DSO’s but has deeper memory and better functionality.
Using The Recorder Functions
3
Starting a Scope Record Function
1 Apply a signal to the red BNC input A.
2 From the Recorder main menu,
highlight Scope Record.
3 Start the recording.
The waveform moves across the screen from right to left like a normal chart recorder. (See Figure 18.)
Figure 18. Recording Waveforms
Observe that the top of the screen displays the following:
Time from start at the top of the screen.
The status at the bottom of the screen which includes
the time/div setting as well as the total timespan that fits the memory.
Note
For accurate recordings it is advised to let the instrument first warm up for five minutes.
33
Fluke 196C/199C
Users Manual
Displaying Recorded Data
In Normal view, the samples that roll off the screen are stored in deep memory. When the memory is full, recording continues by shifting the data in memory and deleting the first samples out of memory.
In View All mode, the complete memory contents are displayed on the screen.
4 Press to toggle between VIEW ALL
(overview of all recorded samples) and NORMAL view.
You can analyze the recorded waveforms using the Cursors and Zoom functions. See Chapter 4: “Using Replay, Zoom and Cursors”.
Using ScopeRecord in Single Sweep Mode
Use the recorder Single Sweep function to automatically stop recording when the deep memory is full.
Continue from step 3 of the previous section:
4 Open the Recorder options
menu.
5
6 Select Single Sweep and accept
Jump to the Mode field.
(2x)
the recorder options.
34
Recording Scope Waveforms In Deep Memory (Scope Record)
Using External Triggering to Start or Stop Scope Record
To record an electrical event that causes a fault, it might be useful to start or stop recording on an external trigger signal:
Start on Trigger to start recording; recording stops when the deep memory is full
Stop on Trigger to stop recording.
Run When Triggered to continue recording as long as a
next trigger comes within 1 division in view all mode.
To set up the test tool, continue from step 3 of the previous section:
4 Apply the signal to be recorded to the red BNC
input A. Apply a trigger signal to the red and black external trigger banana inputs. (See Figure 19.)
5 Open the Recorder Options
menu.
Using The Recorder Functions
3
Figure 19. Scope Record Using External Triggering
6 Jump to Display Glitches:.
7 Jump to Mode:.
8 Select on EXT. ... to open the
Single Sweep on Ext. menu.
9 Select one of the Conditions:
and jump to Slope:.
35
Fluke 196C/199C
Users Manual
10 Select the desired trigger slope,
and jump to Level:
11 Select the 0.12V or 1.2 V trigger
level and accept all recorder options.
During recording samples are continuously saved in deep memory. The last nine recorded divisions are displayed on the screen. Use View All to display the full memory contents.
Note
To learn more about the Single Shot trigger function, see Chapter 5 “Triggering on Waveforms”.
36
Figure 20. Triggered Single Sweep Recording
Analyzing a TrendPlot or Scope Record
From a Scope TrendPlot or Scope Record you can use the analysis functions CURSORS and ZOOM to perform detailed waveform analysis. These functions are described in Chapter 4: “Using Replay, Zoom and Cursors”.
Chapter 4

Using Replay, Zoom and Cursors

About this Chapter
This chapter covers the capabilities of the analysis functions Cursor, Zoom, and Replay. These functions can be used with one or more of the primary functions Scope, TrendPlot or Scope Record.
It is possible to combine two or three analysis functions. A typical application using these functions follows:
First replay the last screens to find the screen of special interest.
Then zoom in on the signal event.
Finally, make measurements using the cursors.
Replaying the 100 Most Recent Scope Screens
When you are in scope mode, the test tool automatically stores the 100 most recent screens. When you press the
HOLD key or the REPLAY key, the memory contents are
frozen. Use the functions in the in time” by stepping through the stored screens to find the screen of your interest. This feature lets you capture and view signals even if you did not press
REPLAY menu to “go back
HOLD.
37
Fluke 196C/199C
Users Manual
Replaying Step-by-Step
To step through the last scope screens, do the following:
1 From scope mode, open the
REPLAY menu.
Observe that the trace is frozen and that REPLAY appears at the top of the screen (see Figure 21).
2 Step through the previous
screens.
3 Step through the next screens.
Observe that the bottom of the waveform area displays the replay bar with a screen number and related time stamp:
38
Figure 21. Replaying a Waveform
The replay bar represents all 100 stored screens in memory. The icon represents the picture being displayed on the screen (in this example:
SCREEN -84). If
the bar is partly white, the memory is not completely filled with 100 screens.
From this point you can use the zoom and cursor functions to study the signal in more detail.
Using Replay, Zoom and Cursors
Replaying the 100 Most Recent Scope Screens
4
Replaying Continuously
You can also replay the stored screens continuously, like playing a video tape.
To replay continuously, do the following:
1 From Scope mode, open the
REPLAY menu.
Observe that the trace is frozen and REPLAY appears at the top of the screen.
2 Continuously replay the stored
screens in ascending order.
Wait until the screen with the signal event of interest appears.
3 Stop the continuous replay.
Turning Off the Replay Function
4 Turn off REPLAY.
Capturing 100 Intermittents Automatically
When you use the test tool in triggered mode, 100 triggered screens are captured. This way you could use Pulse Triggering to trigger and capture 100 intermittent glitches or you could use External Triggering to capture 100 UPS startups.
By combining the trigger possibilities with the capability of capturing 100 screens for later replay, you can leave the test tool unattended to capture intermittent signal anomalies.
For triggering, see Chapter 5: “Triggering on Waveforms”.
39
Fluke 196C/199C
Users Manual
Zooming in on a Waveform
To obtain a more detailed view of a waveform, you can zoom in on a waveform using the
To zoom in on a waveform, do the following:
1 Display the ZOOM key labels.
Observe that the trace is frozen,
HOLD appears at the top of the
screen, and the waveform is magnified.
ZOOM function.
Figure 22. Zooming in a Waveform
2 Enlarge (decrease the time/div)
or shrink (increase the time/div) the waveform.
3 Scroll. A position bar displays the
position of the zoomed part in relation to the total waveform.
Tip
Even when the key labels are not displayed at the bottom of the screen, you can still use the arrow keys to zoom in and out.
40
Observe that the bottom of the waveform area displays the zoom ratio, position bar, and time/div (see Figure 22). The zoom range depends on the amount of data samples stored in memory.
From this point you can use the cursor function for further measurements on the waveform.
Using Replay, Zoom and Cursors
Zooming in on a Waveform
4
Displaying the Zoomed Waveform
The VIEW ALL feature is useful when you quickly need to see the complete waveform and then return to the zoomed part.
4 Display the complete waveform.
Press repeatedly to toggle between the zoomed part of the waveform and the complete waveform.
Turning Off the Zoom Function
5 Turn off the ZOOM function.
41
Fluke 196C/199C
Users Manual
Making Cursor Measurements
Cursors allow you to make precise digital measurements on waveforms. This can be done on live waveforms, recorded waveforms, and on saved waveforms.
Using Horizontal Cursors on a Waveform
To use the cursors for a voltage measurement, do the following:
1 From scope mode, display the
cursor key labels.
2 Press to highlight . Observe
that two horizontal cursors are displayed.
3 Highlight the upper cursor.
4 Move the upper cursor to the
desired position on the screen.
5 Highlight the lower cursor.
6 Move the lower cursor to the
desired position on the screen.
Note
Even when the key labels are not displayed at the bottom of the screen, you still can use the arrow keys. This allows full control of both cursors while having full screen view.
Figure 23. Voltage Measurement with Cursors
The screen shows the voltage difference between the two cursors and the voltage at the cursors. (See Figure 23.)
Use horizontal cursors to measure the amplitude, high or low value, or overshoot of a waveform.
42
Using Vertical Cursors on a Waveform
To use the cursors for a time measurement, do the following:
1 From scope mode, display the
cursor key labels.
2 Press to highlight . Observe
that two vertical cursors are displayed. Markers () identify the point where the cursors cross the waveform.
Using Replay, Zoom and Cursors
Making Cursor Measurements
Figure 24. Time Measurement with Cursors
4
3 If necessary, choose the trace:
TRACE A ,B, or M (Mathematics).
4 Highlight the left cursor.
5 Move the left cursor to the
desired position on the waveform.
6 Highlight the right cursor.
7 Move the right cursor to the
desired position on the waveform.
The screen shows the time difference between the cursors and the voltage difference between the two markers. (See Figure 24.)
8 Turn off the cursors.
43
Fluke 196C/199C
Users Manual
Using Cursors on a A+B, A-B or A*B Waveform
Cursor measurements on a A*B waveform give a reading in Watts if input A measures (milli)Volts and input B measures (milli)Amperes.
For other cursor measurements on a A+B, A-B or A*B waveform no reading will be available if the input A and input B measurement unit are different.
Making Rise Time Measurements
To measure rise time, do the following:
1 From scope mode, display the
cursor key labels.
2 Press to highlight (rise time).
Observe that two horizontal cursors are displayed.
3 If only one trace is displayed,
select MANUAL or AUTO (this automatically does steps 4 to 6). For multiple traces select the required trace A, B, or M (if a math function is active).
4 Move the upper cursor to 100%
of the trace height. A marker is shown at 90%.
5 Highlight the other cursor.
6 Move the lower cursor to 0% of
the trace height. A marker is shown at 10%.
The reading shows the risetime from 10%-90% of the trace amplitude.
Figure 25. Risetime Measurement
44
Chapter 5

Triggering on Waveforms

About this Chapter
This chapter provides an introduction to the trigger functions of the test tool. Triggering tells the test tool when to begin displaying the waveform. You can use fully automatic triggering, take control of one or more main trigger functions (semi-automatic triggering), or you can use dedicated trigger functions to capture special waveforms.
Following are some typical trigger applications:
Use the Connect-and-View™ function to have full automatic triggering and instant display of virtually any waveform.
If the signal is unstable or has a very low frequency, you can control the trigger level, slope, and trigger delay for a better view of the signal. (See next section.)
For dedicated applications, use one of the four manual trigger functions:
Edge triggering
External triggering
Video triggering
Pulse Width triggering
45
Fluke 196C/199C
Users Manual
Setting Trigger Level and Slope
The Connect-and-View™ function enables hands-off triggering to display complex unknown signals.
When your test tool is in manual range, do the following:
Perform an auto set. AUTO appears at the top right of the screen.
Automatic triggering assures a stable display of virtually any signal.
From this point, you can take over the basic trigger controls such as level, slope and delay. To optimize trigger level and slope manually, do the following:
1 Display the TRIGGER key labels.
2 Trigger on either positive slope or
negative slope of the chosen waveform.
3 Enable the arrow keys for manual
trigger level adjustment.
Figure 26. Screen with all Trigger Information
4 Adjust the trigger level.
Observe the trigger icon that indicates the trigger position, trigger level, and slope.
At the bottom of the screen the trigger parameters are displayed. (See Figure 26.) For example, means that input A is used as the trigger source with a positive slope.
When no trigger is found, the trigger parameters appear in gray.
46
Using Trigger Delay or Pre-trigger
You can begin to display the waveform some time before or after the trigger point has been detected. Initially, you have 2 divisions of pre-trigger view (negative delay).
To set the trigger delay, do the following:
5 Hold down to adjust the trigger
delay.
Triggering on Waveforms
Using Trigger Delay or Pre-trigger
5
Observe that the trigger icon on the screen moves to show the new trigger position. When the trigger position moves left off of the screen, the trigger icon changes into
to indicate that you have selected a trigger delay. Moving the trigger icon to the right on the display gives you a pre-trigger view.
In case of a trigger delay, the status at the bottom of the screen will change. For example:
This means that input A is used as the trigger source with a positive slope. The 500.0 ms indicates the (positive) delay between trigger point and waveform display.
When no trigger is found, the trigger parameters appear in gray.
Figure 27. Trigger Delay or Pre-trigger View
Figure 27 shows an example of a trigger delay of 500 ms (top) and an example of pre-trigger view of 8 divisions (bottom).
47
Fluke 196C/199C
Users Manual
Automatic Trigger Options
In the trigger menu, settings for automatic triggering can be changed as follows. (See also Chapter 1: “Displaying an Unknown Signal with Connect-and-View”)
1 Display the TRIGGER key labels.
Note
TRIGGER key labels can differ depending on
The the latest trigger function used.
2 Open the Trigger Options menu.
3 Open the Automatic Trigger
menu.
If the frequency range of the automatic triggering is set to > 15 Hz, the Connect-and-View™ function responds more quickly. The response is quicker because the test tool is instructed not to analyze low frequency signal components. However, when you measure frequencies lower than 15 Hz, the test tool must be instructed to analyze low frequency components for automatic triggering:
4 Select > 1 HZ and return to the
measurement screen.
48
Triggering on Waveforms
Triggering on Edges
5
Triggering on Edges
If the signal is instable or has a very low frequency, use edge triggering to obtain full manual trigger control.
To trigger on rising edges of the input A waveform, do the following:
1 Display the TRIGGER key labels.
2 Open the Trigger Options menu.
3 Open the Trigger on Edge menu.
When Free Run is selected, the test tool updates the screen even if there are no triggers. A trace always appears on the screen.
When On Trigger is selected, the test tool needs a trigger to display a waveform. Use this mode if you want to update the screen only when valid triggers occur.
When Single Shot is selected, the test tool waits for a trigger. After receiving a trigger, the waveform is displayed and the instruments is set to HOLD.
In most cases it is advised to use the Free Run mode:
4 Select Free Run, jump to Noise
reject Filter.
5 Set Noise reject Filter to Off.
Observe that the key labels at the bottom of the screen have adapted to allow further selection of specific edge trigger settings:
49
Fluke 196C/199C
Users Manual
Triggering on Noisy Waveforms
To reduce jitter on the screen when triggering on noisy waveforms, you can use a noise rejection filter. Continue from step 3 of the previous example as follows:
4 Select On Trigger, jump to Noise
reject Filter.
5 Set Noise reject Filter to On.
Observe that the trigger gap has increased. This is indicated by a taller trigger icon
.
Making a Single Acquisition
To catch single events, you can perform a single shot acquisition (one-time screen update). To set up the test tool for a single shot of the input A waveform, continue from step 3 again:
4 Select Single Shot.
5 Accept the settings.
The word WAITING appears at the top of the screen indicating that the test tool is waiting for a trigger. As soon as the test tool receives a trigger, the waveform is displayed and the instrument is set to hold. This is indicated by the word
HOLD at top of the screen.
50
The test tool will now have a screen like Figure 28.
6 Arm the test tool for a new single
shot.
Tip
The test tool stores all single shots in the replay memory. Use the Replay function to look at all the stored single shots.
Triggering on Waveforms
Triggering on Edges
Figure 28. Making a Single Shot Measurement
5
51
Fluke 196C/199C
Users Manual
Triggering on External Waveforms
Use external triggering when you want to display waveforms on inputs A and B while triggering on a third signal. You can choose external triggering with automatic triggering or with edge triggering.
1 Supply a signal to the red and black 4-mm
banana jack inputs. See Figure 29.
In this example you continue from the Trigger on Edges example. To choose the external signal as trigger source, continue as follows:
2 Display the TRIGGER (On Edges)
key labels.
3 Select Ext (external) edge trigger.
Observe that the key labels at the bottom of the screen have been adapted to allow selection of two different external trigger levels: 0.12 V and 1.2 V:
52
Figure 29. External Triggering
4 Select 1.2V under the Ext LEVEL
label.
From this point the trigger level is fixed and is compatible with logic signals.
Triggering on Video Signals
To trigger on a video signal, first select the standard of the video signal you are going to measure:
1 Apply a video signal to the red input A.
2 Display the TRIGGER key labels.
3 Open the Trigger Options menu.
Triggering on Waveforms
Triggering on Video Signals
Figure 30. Measuring Interlaced Video Signals
5
4 Select Video on A to open the
Trigger on Video menu.
5 Select positive signal polarity for
video signals with negative going sync pulses.
6
Trigger level and slope are now fixed.
Observe that the key labels at the bottom of the screen have been changed to allow further selection of specific video trigger settings:
Select the video standard and return.
53
Fluke 196C/199C
Users Manual
Triggering on Video Frames
Use FIELD 1 or FIELD 2 to trigger either on the first half of the frame (odd) or on the second half of the frame (even).
To trigger on the second half of the frame, do the following:
7 Choose FIELD 2.
The signal part of the even field is displayed on the screen.
Triggering on Video Lines
Use ALL LINES to trigger on all line synchronization pulses (horizontal synchronization).
7 Choose ALL LINES.
The signal of one line is displayed on the screen. The screen is updated with the signal of the next line immediately after the test tool triggers on the horizontal synchronization pulse.
To view a specific video line in more detail, you can select the line number. For example, to measure on video line 123, continue from step 6 as follows:
7 Enable video line selection.
8 Select number 123.
The signal of line 123 is displayed on the screen. Observe that the status line now also shows the selected line number. The screen is continuously updated with the signal of line 123.
54
Triggering on Pulses
Use pulse width triggering to isolate and display specific pulses that you can qualify by time, such as glitches, missing pulses, bursts or signal dropouts.
Detecting Narrow Pulses
To set the test tool to trigger on narrow positive pulses shorter than 5 ms, do the following:
1 Apply a video signal to the red input A.
2 Display the TRIGGER key labels.
3 Open the Trigger Options menu.
Triggering on Waveforms
Triggering on Pulses
4 Select Pulse Width on A... to
open the Trigger on Pulse Width menu.
5 Select the positive pulse icon,
then jump to Condition.
6 Select <t, then jump to Update.
7 Select On Trigger.
The test tool is now prepared to trigger on narrow pulses only. Observe that the trigger key labels at the bottom of the screen have been adapted to set the pulse conditions:
5
55
Fluke 196C/199C
Users Manual
To set the pulse width to 5 ms, do the following:
7 Enable the arrow keys to adjust
the pulse width.
8 Select 5 ms.
All narrow positive pulses shorter than 5 ms are now displayed on the screen. (See Figure 31.)
Tip
The test tool stores all triggered screens in the replay memory. For example, if you setup your triggering for glitches, you can capture 100 glitches with time stamps. Use the look at all the stored glitches.
REPLAY key to
Figure 31. Triggering on Narrow Glitches
56
Triggering on Waveforms
Triggering on Pulses
5
Finding Missing Pulses
The next example covers finding missing pulses in a train of positive pulses. In this example it is assumed that the pulses have a 100 ms distance between the rising edges. If the time accidently increases to 200 ms, a pulse is missing. To set the test tool to trigger on such missing pulses, let it trigger on gaps bigger than about 150 ms. Do the following:
1 Display the TRIGGER key labels.
2 Open the Trigger Options menu.
3 Select Pulse Width on A... to
open the Trigger on Pulse Width menu.
4 Select the positive pulse icon to
trigger on the gap between the positive pulses, then jump to
Condition.
5 Select >t, then jump to Update.
6 Select On Trigger.
The test tool is now prepared to trigger on pulse gaps. Observe that the trigger menu at the bottom of the screen has been adapted to set the pulse condition:
57
Fluke 196C/199C
Users Manual
To set the pulse width to 150 ms, continue as follows:
7 Enable the arrow keys to adjust
the pulse width.
8 Select 150 ms.
Figure 32. Triggering on Missing Pulses
58
Chapter 6

Using Memory, PC and Printer

About this Chapter
This chapter provides a step-by-step introduction to the general functions of the test tool that can be used in the three main modes: Scope, Meter, or Recorder. You will find information on printer and computer communication at the end of this chapter.
Saving and Recalling
You can:
Save screens and setups to memory, and recall them again from memory. The test tool has 10 screen and setup memories and 2 record and setup memories.
Recall screens and recordings to analyze or print the screen image at a later date.
Recall a setup to continue a measurement with the recalled operating configuration.
59
Fluke 196C/199C
Users Manual
Saving Screens with Associated Setups
To save a screen in memory location 10, do the following:
1 Display the SAVE/PRINT key labels.
From this point the screen is frozen until you hide the
SAVE/PRINT key labels again.
2 Open the Save menu.
Observe that free memory locations are indicated by an open square (). Filled memory locations are indicated by a solid square ().
3 Highlight SCREEN+SETUP location 10.
4 Save the actual screen.
Note
The two record+setup memory locations store more than what is just visible on the screen. In TrendPlot or scope record mode the full recording is saved. In scope mode you can save all 100 replay screens in a single record+setup memory location.
60
Using Memory, PC and Printer
Saving and Recalling
6
Deleting Screens with Associated Setups
To delete all screens and associated setups, continue from step 2 of the previous example as follows:
3 Delete all saved screens and
setups.
To delete only one screen and setup, continue from step 2 of the previous example as follows:
3 Highlight SCREEN+SETUP location 5.
4 Delete the saved screen+setup of
memory location 5.
Recalling Screens with Associated Setups
To recall screen+setup 1, do the following:
1 Display the SAVE/PRINT key labels.
2 Open the Recall menu.
3 Highlight SCREEN+SETUP location 1.
4 Use RECALL to recall the saved
screen.
Observe that the recalled waveform is displayed and that
HOLD appears on the screen. From this point you can use
cursors and zoom for analysis or you can print the recalled screen.
To recall a sceen as a reference waveform to compare it with an actually measured waveform, continue from step 3 as follows:
61
Fluke 196C/199C
Users Manual
4 Use RECALL FOR REFERENCE to
recall the saved screen.
5 Resume the measurement. Both,
the reference screen and the measurement screen will be displayed.
Recalling a Setup Configuration
To recall the setup configuration from memory 1, do the following:
1 Display the SAVE/PRINT key labels.
2 Open the Recall menu.
3 Highlight SCREEN+SETUP location 1.
4 Use RECALL SETUP to recall the
saved setup.
Observe that RUN appears at the top right of the screen. From this point you continue in the new operating configuration.
62
Using Memory, PC and Printer
Documenting Screens
6
Viewing Stored Screens
To scroll through the memories while looking at the stored screens, do the following:
1 Display the SAVE/PRINT key labels.
2 Open the View... menu.
3 Highlight a screen location, and
view the screen properties.
4 View the screen, and open the
viewer.
5 Scroll through all stored screens.
Documenting Screens
With the FlukeView® software you can upload waveform data and screen bitmaps to your PC or notebook computer for further processing. Printing can also be done by connecting the test tool directly to a printer.
Connecting to a Computer
To connect the test tool to a PC or notebook computer and use the FlukeView software for Windows do the following:
Use the Optically Isolated RS-232 Adapter/Cable
(PM9080) to connect a computer to the OPTICAL PORT of the test tool. (See Figure 33.)
Figure 33. Connecting a Computer
®
(SW90W),
63
Fluke 196C/199C
Users Manual
Note
For information about installing and using the FlukeView ScopeMeter software, see the SW90W Users Manual.
A Software & Cable Carrying Case Kit is optionally available as model number SCC190.
Connecting to a Printer
To print a screen directly to a printer, use one of the following adapters:
The Optically Isolated RS-232 Adapter/Cable
(PM9080) to connect a serial printer to the OPTICAL PORT of the test tool. (See Figure 34.)
The Print Adapter Cable (PAC91, optional) to connect
a parallel printer to the OPTICAL PORT of the test tool. (See Figure 35.)
Before printing, you must setup the test tool for a specific printer.
64
Figure 34. Connecting a Serial Printer
Figure 35. Connecting a Parallel Printer
Using Memory, PC and Printer
Documenting Screens
6
Setting up the Printing Configuration
This example demonstrates how to set up the test tool for printing on a postscript printer with a 9600 baud rate:
1 Display the USER OPTIONS key
labels.
2 Open the User Options menu.
3 Open the Printer Setup submenu.
4 Select Postscript and jump to
Baud Rate.
5 Select a baud rate of 9600 and
return to normal mode.
Whenever possible, choose the option Postscript when printing screens. This option gives the best printing results. Consult the manual that came with your printer to find out whether it has Postscript printing possibilities.
Printing a Screen
To print the currently displayed screen, do the following:
1 Clear the menu if you do not want
to print it.
2 Display the SAVE/PRINT key labels.
3 Start printing.
A message appears at the bottom of the screen indicating that the test tool is busy printing.
Screens will be printed in black and white.
65
Fluke 196C/199C
Users Manual
66
Chapter 7

Tips

About this Chapter
This chapter gives you information and tips on how you can make the best use of the test tool.
Using the Standard Accessories
The following illustrations show the use of the standard accessories such as voltage probes, test leads, and the various clips.
67
Fluke 196C/199C
Users Manual
Figure 36. Probing Using 2-mm Heavy Duty
Test Probes
Warning
To avoid electrical shock or fire, do not connect the ground spring to voltages higher than 30 Vrms from earth ground.
Figure 38. Heavy Duty Fixed Connections for Scope
Measurements Using Industrial Alligator Clips
Figure 37. HF Voltage Probe Connection Using
Ground Spring
68
Figure 39. Electronic Connections for Scope
Measurements Using Hook Clips and Hook Clip
Grounding
Tips
Using the Standard Accessories
7
Figure 40. Electronic Connections for Scope
Measurements Using Hook Clips and Alligator Clip
Grounding
Figure 41. Fixed Electronic Connections for Scope
Measurements with Fixed External Triggering
Figure 42. Manual Probing for Meter Measurements
using 2-mm Test Probes
Figure 43. Heavy Duty Fixed Connections for Meter
Measurements Using Industrial Alligator Clips
69
Fluke 196C/199C
Users Manual
Using the Independently Floating Isolated Inputs
You can use the independently floating isolated inputs to measure signals that are independently floating from each other.
Independently floating isolated inputs offer additional safety and measurement capabilities compared to inputs with common references or grounds.
Measuring Using Independently Floating Isolated Inputs
The test tool has independently floating isolated inputs. Each input section (A, B, External Trigger / DMM) has its own signal input and its own reference input. The reference input of each input section is electrically isolated from the reference inputs of the other input sections. The isolated input architecture makes the test tool about as versatile as having three independent instruments. The advantages of having independently floating isolated inputs are:
It allows simultaneous measurement of independently floating signals.
Additional safety. Since the commons are not directly connected, the chance of causing short circuit when measuring multiple signals is greatly reduced.
Additional safety. When measuring in systems with multiple grounds, the ground currents induced are kept to a minimum.
Because the references are not connected together inside the test tool, each reference of the used inputs must be connected to a reference voltage.
Independently floating isolated inputs are still coupled by parasitic capacitance. This can occur between the input references and the environment, and between the input references mutually (see Figure 44). For this reason, you should connect the references to a system ground or another stable voltage. If the reference of an input is connected to a high speed and / or high voltage signal, you should be aware of parasitic capacitance. (See Figures 44, 45, 46 and 47.)
70
Tips
+
A
A
Using the Independently Floating Isolated Inputs
NALOG INPUT DIGITAL CONTROLLER
DIGITAL
ANALOG GROUND
GROUND
7
Figure 44. Parasitic capacitance between probes,
instrument and environment
MOTION/MOTOR CONTROLLER
DIGITAL GROUND
DC BUS
DC BUS
M
-
Figure 46. Correct connection of reference leads
NALOG INPUT DIGITAL CONTROLLER
DIGITAL
ANALOG GROUND
GROUND
Figure 47. Wrong connection of reference leads
Noise that is picked up by reference lead B can be
Figure 45. Parasitic capacitance between analog and
digital reference
transmitted by parasitic capacitance to the analog input amplifier.
71
Fluke 196C/199C
Users Manual
Using the Tilt Stand
The test tool is equipped with a tilt stand, allowing viewing from an angle while placed on a table. From this position you can access the OPTICAL PORT at the side of the test tool. The typical position is shown in Figure 48.
Figure 48. Using the Tilt Stand
Resetting the Test Tool
If you want to reset the test tool to the factory settings, do the following:
1 Turn the test tool off.
2 Press and hold.
3 Press and release.
The test tool turns on, and you should hear a double beep, indicating the reset was successful.
4 Release.
Suppressing Key Labels and Menu’s
You can hide a menu or key label at any time:
Hide any key label or menu.
To display menus or key labels, press one of the yellow menu keys, e.g. the
SCOPE key.
72
Tips
Changing the Information Language
7
Changing the Information Language
During operation of the test tool, messages may appear at the bottom of the screen. You can select the language in which these messages are displayed. The combination of selectable languages (one or more) depends on the version ordered.
In this example you can select English or French . To change the language from English to French, do the following:
1 Display the USER key labels.
2
3
4
Open the Language Select menu.
Highlight FRENCH.
Accept French as the language.
Adjusting the Contrast and Brightness
To adjust the contrast and backlight brightness, do the following:
1 Display the USER key labels.
2 Enable the arrow keys for manual
contrast and backlight adjustment.
3 Adjust the contrast of the screen.
4 Change the backlight.
Note
The new contrast and brightness are stored until a new adjustment is made.
To save battery power, the test tool is in economic brightness mode when operated on the battery. The high brightness intensity increases when you connect the power adapter.
Note
Using dimmed light lengthens maximum battery power operation by about one hour.
73
Fluke 196C/199C
Users Manual
Changing the Display Color
To set the display to color or Black and White, do the following:
1 Display the USER key labels.
2 Open the User Options menu.
3 Open Display Options menu.
4 Choose display mode Color or
Black and White, and accept it.
Changing Date and Time
The test tool has a date and time clock. For example, to change the date to 19 April, 2002, do the following:
1 Display the USER key labels.
2 Open the User Options menu.
3 Open Date Adjust menu.
4 Choose 2002, jump to Month.
5 Choose 04, jump to Day.
74
Tips
Saving Battery Life
7
6 Choose 19, jump to Format.
7 Choose DD/MM/YY, accept the new
date.
You can change the time in a similar way by opening the Time Adjust menu (steps 2 and 3.)
Saving Battery Life
When operated on the battery (no battery charger connected), the test tool conserves power by shutting itself down. If you have not pressed a key for at least 30 minutes, the test tool turns itself off automatically.
Note
If the power adapter is connected, there is no automatic power shutdown.
Automatic power shutdown will not occur if TrendPlot or Scope Record is on, but the backlight will dim. Recording will continue even if the battery is low, and retention of memories is not jeopardized.
Setting the Power Down Timer
Initially the power shutdown time is 30 minutes. You can set the power shutdown time to 5 minutes as following:
1 Display the USER key labels.
2 Open the User Options menu.
3 Open the Battery Save Options
menu.
4 Select 5 Minutes.
75
Fluke 196C/199C
Users Manual
Changing the Auto Set Options
With the next procedure you can choose how auto set behaves when you press the
1 Display the USER key labels.
2 Open the User Options menu.
3 Open the Auto Set Adjust menu.
If the frequency range is set to > 15 Hz, the Connect-and­View function responds more quickly. The response is quicker because the test tool is instructed not to analyze low frequency signal components. However, when you measure frequencies lower than 15 Hz, the test tool must be instructed to analyze low frequency components for automatic triggering:
AUTO (auto set) key.
4 Select Signal > 1 Hz, then jump
to Coupling.
With the coupling option you can choose how auto set behaves. When you press the
AUTO (auto set) key, the
coupling can either be set to dc or left unchanged:
5 Select Unchanged.
Note
The auto set option for the signal frequency is similar to the automatic trigger option for the signal frequency. (See Chapter 5: “Automatic Trigger Options”). However, the auto set option determines the behavior of the auto set function and shows only effect when you press the auto set key.
76
Chapter 8

Maintaining the Test Tool

About this Chapter
This chapter covers basic maintenance procedures that can be performed by the user. For complete service, disassembly, repair, and calibration information, see the Service Manual. You will find the part number of the Service Manual in the section “Parts and Accessories” in this chapter.
Cleaning the Test Tool
Clean the test tool with a damp cloth and a mild soap. Do not use abrasives, solvents, or alcohol. These may damage the text on the test tool.
Storing the Test Tool
If you are storing the test tool for an extended period of time, charge the NiMH (Nickel-Metal Hydride) batteries before storing.
77
Fluke 196C/199C
Users Manual
Charging the Batteries
At delivery, the NiMH batteries may be empty and must be charged for 4 hours (with the test tool turned off) to reach full charge. When fully charged, the batteries provide 4 hours of use.
When battery power is used, the battery indicator at the top of the screen informs you about the condition of the batteries. The battery symbols are: symbol of operating time left.
To charge the batteries and power the instrument, connect the battery charger as shown in Figure 49. To charge the batteries more quickly, turn off the test tool.
indicates that there are typically five minutes
Caution
To avoid overheating of the batteries during charging, do not exceed the allowable ambient temperature given in the specifications.
Note
No damage will occur if the charger is connected for long periods, e.g., during the weekend. The instrument then automatically switches to trickle charging.
. The
Figure 49. Charging the Batteries
78
Maintaining the Test Tool
Extending Battery Operation Time
8
Extending Battery Operation Time
Typically, NiMH batteries always meet the specified operating time. However, if the batteries have been extremely discharged (for example, when empty batteries were stored for a long period) it is possible that the battery condition has deteriorated.
To keep the batteries in optimal condition, observe the following guidelines:
Operate the test tool on batteries until the appears at the bottom of the screen. This indicates that the battery level is low and that the NiMH batteries need to be recharged.
To obtain optimal battery condition again, you can refresh the batteries. During a battery refresh, the batteries will be fully discharged and charged again. A complete refresh cycle takes about 12 hours and should be done about four times a year. You can check the latest battery refresh date. See section “Displaying Calibration Information”.
symbol
To refresh the battery, make sure that the test tool is line powered and proceed as follows:
1 Display the USER key labels.
2 Open the User Options menu.
A message appears asking whether you want to start the refresh cycle now.
3 Start the refresh cycle.
Do not disconnect the battery charger during the refresh cycle. This will interrupt the refresh cycle.
Note
After starting the refresh cycle, the screen will be black.
79
Fluke 196C/199C
Users Manual
Replacing the NiMH Battery Pack BP190
Usually it should not be necessary to replace the battery pack. However, if replacement is needed, this should be done by qualified personnel only. Contact your nearest Fluke center for more information.
Calibrating the Voltage Probes
To meet full user specifications, you need to adjust the red and gray voltage probes for optimal response. The calibration consists of a high frequency adjustment and a dc calibration for 10:1 probes. The dc calibration is not possible for 100:1 probes.
This example shows how to calibrate the 10:1 voltage probes:
1 Display the input A key labels.
2 Open the Probe on A menu.
3 Select Voltage, then jump to
Attenuation.
80
Figure 50. Adjusting Voltage Probes
If the 10:1 option is already selected, proceed with step 5.
4 Select 10:1, then return.
Maintaining the Test Tool
Calibrating the Voltage Probes
8
Repeat steps 2 and 3 and proceed as follows:
5 Select Probe Cal with the arrow
keys, then accept.
A message appears asking you whether to start the 10:1 probe calibration.
6 Start the probe calibration.
A message appears telling you how to connect the probe. Connect the red 10:1 voltage probe from the red input A jack to the red banana jack. Connect the reference lead to the black banana jack. (See Figure 50.)
7 Adjust the trimmer screw in the
probe housing until a pure square wave is displayed.
8 Continue with DC calibration.
Automatic DC calibration is only possible for 10:1 voltage probes.
The test tool automatically calibrates itself to the probe. During calibration you should not touch the probe. A message indicates when the DC calibration has completed successfully.
9 Return.
Repeat the procedure for the gray 10:1 voltage probe. Connect the grey 10:1 voltage probe from the grey input B jack to the red banana jack. Connect the reference lead to the black banana jack.
Note
When using 100:1 voltage probes, choose 100:1 attenuation to perform a HF adjustment. Automatic dc calibration is not possible with this probe type.
81
Fluke 196C/199C
Users Manual
Displaying Calibration Information
You can display version number and calibration date:
1 Display the USER key labels.
2 Open the Version & Calibration
menu.
The screen gives you information about the model number with software version, the calibration number with latest calibration date, and the latest battery refresh date.
3 Return.
Recalibration must be carried out by qualified personnel. Contact your local Fluke representative for recalibration.
Parts and Accessories
Standard Accessories
The following tables list the user-replaceable parts for the various test tool models. For additional optional accessories, see the ScopeMeter Accessories booklet.
To order replacement parts or additional accessories, contact your nearest service center.
82
Maintaining the Test Tool
Parts and Accessories
Table 1. Standard Accessories
Item Ordering Code
Battery Charger, available models:
Universal Europe 230 V, 50 and 60 Hz North America 120 V, 50 and 60 Hz United Kingdom 240 V, 50 and 60 Hz Japan 100 V, 50 and 60 Hz Australia 240 V, 50 and 60 Hz Universal 115 V/230 V, 50 and 60 Hz
*
The 230V rating of the BC190/808 is not for use in North America. A line plug adapter complying with the applicable National Requirements may be provided to alter the blade configurations for a specific country. The universal adapter is standard equipped with a North American line cord.
Voltage Probe Set (Red), designed for use with the Fluke ScopeMeter 190 series test tool.
The set includes the following items (not available separately):
10:1 Voltage Probe (red)
4-mm Test Probe for Probe Tip (red)
Hook Clip for Probe Tip (red)
Ground Lead with Hook Clip (red)
Ground Lead with Mini Alligator Clip (black)
Ground Spring for Probe Tip (black)
*
BC190/801 BC190/803 BC190/804 BC190/806 BC190/807 BC190/808
VPS200-R
8
83
Fluke 196C/199C
Users Manual
Item Ordering Code
Voltage Probe Set (Gray), designed for use with the Fluke ScopeMeter 190 series test tool.
The set includes the following items (not available separately):
10:1 Voltage Probe (gray)
4-mm Test Probe for Probe Tip (gray)
Hook Clip for Probe Tip (gray)
Ground Lead with Hook Clip (gray)
Ground Lead with Mini Alligator Clip (black)
Flexible Test Leads (red and black)
Accessory Set (Red)
The set includes the following items (not available separately):
Industrial Alligator for Probe Tip (red)
2-mm Test Probe for Probe Tip (red)
Industrial Alligator for Banana Jack (red)
2-mm Test Probe for Banana Jack (red)
Ground Lead with 4-mm Banana Jack (black)
Accessory Set (Gray)
The set includes the following items (not available separately):
Industrial Alligator for Probe Tip (gray)
2-mm Test Probe for Probe Tip (gray)
Industrial Alligator for Banana Jack (gray)
2-mm Test Probe for Banana Jack (gray)
Ground Lead with 4-mm Banana Jack (black)
VPS200-G
TL24 (General Purpose Leads)
AS200-R
AS200-G
84
Maintaining the Test Tool
Parts and Accessories
Item Ordering Code
Replacement Set for Voltage Probe
The set includes the following items (not available separately):
2x 4-mm Test Probe for Probe Tip (red and gray)
3x Hook Clip for Probe Tip (2 red, 1 gray)
2x Ground Lead with Hook Clip (red and gray)
2x Ground Lead with Mini Alligator Clip (black)
5x Ground Spring for Probe Tip (black)
Table 2. Users Manuals
Item Ordering Code
Getting Started Manual (English)
Getting Started Manual (German)
Getting Started Manual (French)
Getting Started Manual (Spanish)
Getting Started Manual (Portuguese)
Getting Started Manual (Italian)
Getting Started Manual (Chinese)
Getting Started Manual (Japanese)
Getting Started Manual (Korean)
CD ROM with User Manual (all languages)
RS200
4822 872 30471
4822 872 30472
4822 872 30473
4822 872 30474
4822 872 30475
4822 872 30476
4822 872 30477
4822 872 30478
4822 872 30479
4022 240 12371
8
85
Fluke 196C/199C
Users Manual
Optional Accessories
Item Ordering Code
Software & Cable Carrying Case Kit
SCC190
Set contains the following parts:
Optically Isolated RS-232 Adapter/Cable
Hard Carrying Case
®
FlukeView 2000
ScopeMeter® Software for Windows 95® , 98® , Me® ,
®
and NT4
®
PM9080
C190
SW90W
Optically Isolated RS-232 Adapter/Cable PM9080
Hard Case C190
Soft Case C195
Current Shunt 4-20 mA CS20MA
Print Adapter Cable for Parallel Printers PAC91
Optional Service Manual
Item Ordering Code
Service Manual (English) 4822 872 05384
86
Troubleshooting
Maintaining the Test Tool
Troubleshooting
8
The Test Tool Does Not Start Up
The batteries may be completely empty. In this case the test tool will not start up, even if it is powered by the battery charger. Charge the batteries first: power the test tool with the battery charger without turning it on. Wait about 15 minutes and try turning on the test tool again.
The Test Tool Shuts Down After A Few Seconds
The batteries may be empty. Check the battery symbol at the top right of the screen. A indicates that the batteries are empty and must be charged.
symbol
The Screen Remains Black
Make sure that the test tool is on.
You might have a problem with the screen contrast.
Press
the arrow keys to adjust the contrast.
, then press
. Now you can use
The Operation Time Of Fully Charged Batteries Is Too Short
The batteries may be in poor condition. Refresh the batteries to optimize the condition of the batteries again. It is advised to refresh the batteries about four times a year.
The Printer Does Not Print
Make sure that the interface cable is properly connected between the test tool and the printer.
Make sure that you have selected the correct printer type. (See Chapter 6.)
Make sure that the baud rate matches with the printer. If not, select another baud rate. (See Chapter 6.)
If you are using the PAC91 (Print Adapter Cable), make sure that it is turned on.
87
Fluke 196C/199C
Users Manual
FlukeView Does Not Recognize The Test Tool
Make sure that the test tool is turned on.
Make sure that the interface cable is properly
connected between the test tool and the PC.
Make sure that the correct COM port has been selected in FlukeView. If not, change the COM port setting or connect the interface cable to another COM port.
Battery Operated Fluke Accessories Do Not Function
When using battery operated Fluke accessories, always first check the battery condition of the accessory with a Fluke multimeter.
88
Chapter 9

Specifications

Introduction
Performance Characteristics
FLUKE guarantees the properties expressed in numerical values with the stated tolerance. Specified non-tolerance numerical values indicate those that could be nominally expected from the mean of a range of identical ScopeMeter test tools.
Environmental Data
The environmental data mentioned in this manual are based on the results of the manufacturer’s verification procedures.
Safety Characteristics
The test tool has been designed and tested in accordance with Standards ANSI/ISA S82.01-1994, EN 61010.1 (1993) (IEC 1010-1), CAN/CSA-C22.2 No.1010.1-92 (including approval), UL3111-1 (including approval) Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use.
This manual contains information and warnings that must be followed by the user to ensure safe operation and to keep the instrument in a safe condition. Use of this equipment in a manner not specified by the manufacturer may impair protection provided by the equipment.
89
Fluke 196C/199C
Users Manual
Dual Input Oscilloscope
Isolated Inputs A and B (Vertical)
Bandwidth, DC Coupled
FLUKE 199C ....................................... 200 MHz (-3 dB)
FLUKE 196C ....................................... 100 MHz (-3 dB)
Lower Frequency Limit, AC Coupled
with 10:1 probe......................................... <2 Hz (-3 dB)
direct (1:1)................................................ <5 Hz (-3 dB)
Rise Time
FLUKE 199C ........................................................ 1.7 ns
FLUKE 196C ........................................................ 3.5 ns
Analog Bandwidth Limiters ............... 20 MHz and 10 kHz
Input Coupling.......................................................AC, DC
Polarity................................................... Normal, Inverted
Sensitivity Ranges
with 10:1 probe.............................. 50 mV to 1000 V/div
direct (1:1)......................................... 5 mV to 100 V/div
Trace Positioning Range................................±4 divisions
Input Impedance on BNC
DC Coupled........................1 MΩ (±1 %)//15 pF (±2 pF)
Max. Input Voltage
with 10:1 probe.........................................600 V CAT III
1000 V CAT II
direct (1:1) ................................................300 V CAT III
(For detailed specifications, see “Safety”)
Vertical Accuracy .....................±(1.5 % + 0.04 range/div)
Digitizer Resolution .................... 8 bits, separate digitizer
for each input
Horizontal
Maximum Time Base Speed:
FLUKE 199C .....................................................5 ns/div
FLUKE 196C .....................................................5 ns/div
Minimum Time Base Speed (Scope Record)..... 2 min/div
Real Time Sampling Rate (for both inputs simultaneously)
FLUKE199C:
5 ns to 2 µs /div ................................... up to 2.5 GS/s
5 µs to 120 s/div ............................................ 20 MS/s
FLUKE 196C:
5 ns to 2 µs /div ...................................... up to 1 GS/s
5 µs to 120 s/div ............................................ 20 MS/s
90
Specifications
Dual Input Oscilloscope
9
Record Length
Scope Record Mode........... 27500 points on each input
Scope Normal Mode............. 1000 points on each input
Scope Glitch Capture Mode ... 500 points on each input
Glitch Detection
5 µs to 120 s/div ........ displays glitches as fast as 50 ns
Waveform Display................ A, B, A+B, A-B, A*B, A vs B
Normal, Average (2,4,8,64 x), Persistence
Time Base Accuracy .........................................±100 ppm
Trigger and Delay
Trigger Modes....................................... Automatic, Edge,
External, Video, Pulse Width
Trigger Delay.................................. up to +1000 divisions
Pre Trigger View ............................ one full screen length
Max. Delay ..................................................... 10 seconds
Automatic Connect-and-View Trigger
Source............................................................... A, B, EXT
Slope....................................................Positive, Negative
Edge Trigger
Screen Update ........... Free Run, On Trigger, Single Shot
Source............................................................... A, B, EXT
Slope....................................................Positive, Negative
Trigger Level Control Range ......................... ±4 divisions
Trigger Sensitivity A and B
DC to 5 MHz at >5 mV/div ......................... 0.5 divisions
DC to 5 MHz at 5 mV/div ................................ 1 division
200 MHz (FLUKE 199C)................................. 1 division
250 MHz (FLUKE 199C)............................... 2 divisions
100 MHz (FLUKE 196C)................................. 1 division
150 MHz (FLUKE 196C)............................... 2 divisions
Isolated External Trigger
Bandwidth.............................................................. 10 kHz
Modes.....................................................Automatic, Edge
Trigger Levels (DC to 10 kHz)................... 120 mV, 1.2 V
Video Trigger
Standards ............................. PAL, PAL+, NTSC, SECAM
Modes...................... Lines, Line Select, Field 1 or Field 2
Source ............................................................................ A
Polarity ................................................ Positive, Negative
Sensitivity ...................................... 0.7 division sync level
91
Fluke 196C/199C
Users Manual
Pulse Width Trigger
Screen Update............................ On Trigger, Single Shot
Trigger Conditions............<T, >T, =T (±10 %), T(±10 %)
Source............................................................................ A
Polarity.....................................Positive or negative pulse
Pulse Time Adjustment Range ......... 0.01 div. to 250 div.
with a minimum of 300 ns (<T, >T) or 500 ns (=T, ≠T),
a maximum of 10 s,
and a resolution of 0.01 div. with a minimum of 50 ns
Continuous Auto Set
Autoranging attenuators and time base, automatic Connect-and-View™ triggering with automatic source selection.
Modes
Normal.................................... 15 Hz to max. bandwidth
Low Frequency......................... 1 Hz to max. bandwidth
Minimum Amplitude A and B
DC to 1 MHz........................................................ 10 mV
1 MHz to max. bandwidth.................................... 20 mV
Automatic Capturing Scope Screens
Capacity............................ 100 dual input scope Screens
For viewing screens, see Replay function.
Automatic Scope Measurements
The accuracy of all readings is within ± (% of reading + number of counts) from 18 °C to 28 °C. Add 0.1x (specific accuracy) for each °C below 18 °C or above 28 °C. For voltage measurements with 10:1 probe, add probe accuracy unless the probe has been calibrated on the test tool. At least 1.5 waveform period must be visible on the screen.
General
Inputs .....................................................................A and B
DC Common Mode Rejection (CMRR).................. >100 dB
AC Common Mode Rejection at 50, 60, or 400 Hz....... >60 dB
DC Voltage (VDC)
Maximum Voltage
with 10:1 probe..................................................... 1000 V
direct (1:1) .............................................................. 300 V
Maximum Resolution
with 10:1 probe........................................................ 1 mV
direct (1:1) ............................................................ 100 µV
Full Scale Reading .......................................... 1100 counts
Accuracy at 5 s to 5 µs/div ................... ±(1.5 % +5 counts)
Normal Mode AC Rejection at 50 or 60 Hz ............ >60 dB
92
Loading...