ScopeMeter 190C and 190B Series: Speed, performance and analysis power
For demanding applications, the ScopeMeter 190C and 190B Series high-performance oscilloscopes offer specifications usually found on top-end bench instruments. With up to 200 MHz bandwidth,
2.5 GS/s real time sampling and a deep memory of 27,500 points per input they’re ideal for engineers who need the full capabilities of a high-performance scope in a hand­held, battery powered instrument.
Dual input - 200, 100 or
60 MHz bandwidth
Now expanded with 3K memory and FFT Frequency Spectrum Analysis
Up to 2.5 GS/s real time
sampling per input High Waveform Resolution of
3000 datapoints per channel Frequency Spectrum using FFT
analysis (190C) Connect-and-View™automatic
triggering, a full range of man­ual trigger modes plus external triggering
Digital Persistence for analyzing
complex, dynamic signals like on an analog oscilloscope (190C)
Fast display update rate for
seeing dynamic behavior instantaneously
Automatic capture and replay of
100 screens 27,500 points and more per
input record length using ScopeRecord
1,000 V CAT II and 600 V CAT
III safety certified Up to 1,000 V independently
floating isolated inputs Four hours rechargeable
NiMH battery pack
mode
ScopeMeter 120 Series: Three-in-one simplicity
The compact ScopeMeter 120 Series is the rugged solution for industrial troubleshooting and installation applications. It’s a truly integrated test tool, with oscilloscope, multi­meter and “paperless” recorder in one affordable, easy-to-use instru­ment. Find fast answers to problems in machinery, instrumentation, control and power systems.
A dual input 40 MHz or 20 MHz
digital oscilloscope Two 5,000 counts true-rms
digital multimeters Cursor measurements (Fluke 124)
A dual input TrendPlot recorder
Connect-and-View trigger
simplicity for hands-off operation Shielded test leads for oscillo-
scope, resistance, continuity and capacitance measurements
Hz prob
idth, VPS40 1
e included standard
24
ull bandw
F
40 M
ith Fluke 1
w Up to seven hours batter
operation 600 V CAT III safety certified
Optically isolated RS-232
interface Rugged, compact case
0:1
y
Technical Specifications 190C and 190B Series
Oscilloscope Mode
Vertical Deflection
Fluke 199C Fluke 196C
99B Fluke 196B Fluke 192B
idth 200 MHz 100 MHz 60 MHz
Bandw Rise time
Bandwidth limiter: User selectable: 10 kHz, 20 MHz or off Number of inputs: 2 plus external trigger. All inputs isolated from each other and ground.
Input coupling: AC or DC, with ground level indicator Input sensitivity: 2 mV/div to 100 V/div (Fluke 190C
Series); 5 mV/div to 100 V/div (Fluke 190B Series) Normal/Invert: On both input channels; switched separately
Variable attenuator: Variable gain on input channel A Input voltage: 1000 V CAT II, 600 V CAT III rated -
See “general specifications” for further details.
Vertical resolution: 8 bit Accuracy: ± (1.5% of reading + 0.04 x range/div) Input impedance: 1 M± 1% // 15 pF ± 2 pF
Horizontal
Maximum 2.5 GS/s 1 GS/s 500 MS/s real-time sample rate
Number of 2 2 2 digitizers
Time base 5 ns/div 10 ns/div range to 5 s/div to 5 s/div
Maximum record length: 3000 points per input in Scope-mode; 27,500 points per input in ScopeRecordroll mode (5 ms/div to 2 min/div)
Accuracy: ± (0.01 % of reading + 1 pixel) Glitch capture: 50 nsec (at 5 µsec/div to 1 min/div);
250 nsec (at 2 min/div)
Display and Acquisition
Display 144 mm 144 mm
Display modes Input A, Input B, dual, average, Replay Persistence Digital Persistence: Persistence
modes
Visible screen width: 12 divisions Waveform mathematics: A + B, A - B, A * B, all
with user selectable scaling of resultant; A versus B (X - Y-mode); Frequency Spectrum using FFT analysis (190C only) Acquisition modes: Normal, auto, single shot, ScopeRecordroll, glitch capture, waveform compare, waveform compare with automatic “Pass / Fail testing” (in 199C and 196C only)
Trigger and Dela
Source: Input A, input B, exter All input references isolated from each other and from ground Modes: Automatic Connect-and­single shot, edge, dela pulsewidth, dual slope ( only)
Fluke 1
1.7 ns 3.5 ns 5.8 ns
Fluke 199C Fluke 196C Fluke 199B Fluke 196B Fluke 192B
Fluke 190C Fluke 190B
Full Color LCD Monochrome LCD
short/medium/long/infinite
on / off
y
.
free run,
ycle (190C
ideo, v
y, v
0C only), N
9
1
nal trigger input
,
View
ideo line, selectable
-c
Connect-and-View™: Advanced automatic triggering that recognizes signal patterns, automatically sets up and continuously adjusts triggering, time base and amplitude. Automatically displays stable waveforms of complex and dynamic signals like motor drive and control signals. Video triggering: NTSC, PAL, PAL+, SECAM. Includes field 1, field 2 and line select. Pulse width triggering: Pulse width qualified by time. Allows for triggering <t, >t, =t, t, where t is selec­table in minimal steps of 0.01 div. or 50 nsec. Time delay: One full screen of pre-trigger view or up to 100 screens (= 1200 divisions) of post-trigger delay Dual slope triggering: Both rising and falling transi­tions, when crossing the trigger level, initiate an acquisition (190C only) N-cycle triggering: Triggers on N-th occurrence of a trigger event; N to be set in the range 2 to 99 (190C only)
Automatic Capture of 100 Screens
The instrument ALWAYS memorizes the last 100 screens (no user setup required). When an anomaly occurs on screen, the REPLAY button can be pressed to review the full screen sequence over and over.
Instrument can be set up for triggering on glitches or intermittent anomalies and will operate in “baby-sit” mode and will capture 100 events.
Alternatively, the 199C and 196C can be set up in waveform compare mode to store only matching (“Pass”) or only non-matching (“Fail”) acquired wave­forms in the replay memory bank for further analysis. Replay: Manual or continuous replay. Displays the captured 100 screens as a “live” animation or under manual control. Each screen has date- and time-stamp. Replay storage: Up to 2 sets of 100 screens each can be saved for later recall and analysis
FFT - Frequency Spectrum Analysis (190C only)
Shows frequency content of oscilloscope waveform using Fast Fourier Transform
Window: Automatic, Hamming, Hanning or None Automatic window: Digitally re-samples acquired
waveform to get optimum frequency resolution in FFT resultant
Vertical scale: Linear / Logarithmic, in volts Frequency axis: Logarithmic; frequency range auto-
matically set as function of timebase range of oscillo­scope
Waveform Compare and Pass/Fail Testing
Waveform compare: Provides storage and display of a reference waveform for visual comparison with newly acquired waveforms. Reference is derived from an acquired waveform and can be modified in the ScopeMeter or externally using FlukeView®Software. Pass/Fail Testing (199C, 196C): In waveform compare mode, the Color Scopemeter can be set up to store only matching (“Pass”) or only non-matching (“Fail”) acquired waveforms in the replay memory bank for further analysis.
2 Fluke Corporation ScopeMeter 190C/190B Series and ScopeMeter 120 Series
Automatic Scope Measurements
ac rms, Vac+dc, Vpeak max, Vpeak min, Vpeak to
Vdc, V peak, Aac, Adc, Aac+dc, frequenc time, power factor, W
idth (pos/neg), dutycycle (pos/neg), temperature °C,
w temperature
Vpw
idth modulated motordrives and frequency inverters
w
°F, dBV, dBm into 50 and 600
m ac, Vpwm ac+dc for measurement on pulse
atts, VA, VA reactive, phase, pulse
y (Hz), risetime, fall-
Cursor Measurements
Source: Input A, input B or the Mathematical Result trace (excl. A vs B curve) Dual horizontal lines: Voltage at cursor 1 and 2, voltage b Dual vertical lines: Time b
etween cursors (in Hz), voltage between markers,
b risetime w
etween cursors (190C only), watts between cursors
b
190C only)
(
etween cursors
etween cursors, 1/T
ith markers, falltime with markers; Vrms
Single vertical line: Min-Max and Average voltage at cursor position; frequency and rms-value of individ­ual frequency component in FFT result (190C only) Zoom: Up to 16x horizontal zoom
Recorder Mode
ScopeRecord-Roll Mode
Dual input waveform storage mode
Source and display: Input A, Input B, Dual Memory depth: 27,500 points per input. Each point
consists of Min-Max pair Min-Max values: Min-Max values are measured at high sample rates ensuring capture and display of glitches
Time base range 5 ms/div to 1 min/div 2 min/div
Recorded timespan 6 sec to 24 hr 48 hr Glitch capture 50 ns 250 ns Sample rate 20 MS/s 4 MS/s Resolution 200 µsec to 2 sec 4.8 sec
ecording modes:
R
Single sweep, c Start-on-Trigger, Stop-on-Trigger Stop-on-Trigger (through external): ScopeRecord mode can be stopped by an individual trigger event, or by an interruption of a repetitive trigger signal
Horizontal scale: Time from start, time of day Zoom: Up to 100x Memory: Up to 2 dual input ScopeRecord waveforms
can be saved for later recall and analysis
TrendPlot™recording
Single or dual input electronic paperless chart recorder. Plots, displays and stores meter and scope measure­ments. Source and display: Input A, Input B or DMM input
ontinuous roll,
Memory depth: 18,000 points recording. Per record point a minimum, a maximum and an average value, plus a date and timestamp are recorded. Ranges: 5 s/div to 30 min/div in normal view mode; 5 min/div to 48 hr/div in view all mode, giving overview of total record Recorded timespan: Up to 22 days with a resolution of 1 minute Recording mode: Continuous roll for the duration of the full recordable timespan
Measurement speed: 5 measurements per second Horizontal scale: Time from start, time of day Zoom: Up to 64x zoom Memory: Up to 2 TrendPlot recordings can be saved
for later recall and analysis
Cursor measurements - all recorder modes
Source: Input A, B or DMM input Dual vertical lines: Min-Max or Average
voltage. Time between cursors Single vertical line: Min-Max or Average voltage. Absolute date and time or time from start
Meter Mode
Via 4 mm banana inputs. Fully isolated from scope inputs and scope ground. The specified accuracy is valid over the temperature range 18 °C to 28 °C (65 °F to 82 °F). Add 10 % of specified accuracy for each degree C below 18 °C or above 28 °C.
Maximum Resolution: 5,000 counts Voltmeter Ranges: 500 mV, 5 V, 50 V, 500 V, 1,000 V Accuracy:
Vdc ± (0.5 % + 5 counts) Vac true rms
15 Hz...60 Hz: ± (1 % + 10 counts) 60 Hz...1 kHz: ± (2.5 % + 15 counts)
Vac+dc true rms
dc...60 Hz: ± (1 % + 10 counts) 60 Hz...1 kHz: ± (2.5 % + 15 counts)
Ohms: Ranges: 500 , 5 k, 50 k, 500 k, 5 M, 30 M Accuracy: ± (0.6 % + 5 counts)
Other Meter Functions
Continuity: Beeper on < 50 (± 30 Ω) Diode test: Up to 2.8 V Amps: Adc, Aac, Aac+dc using an optional current
clamp or shunt. Scaling factors: 0.1 mV/A ... 100 V/A Temperature (°C, °F): With optional accessories. Scale factors 1 mV/°C or 1 mV/°F
Input impedance: 1 M± 1% // 10 pF ± 2 pF Advanced meter functions: Auto/manual ranging,
relative measurements (Zero reference), TrendPlot recording
opeMeter 1
Sc
/190B Series and ScopeMeter 120 Series Fluke Corporation 3
0C
9
General Specifications
Memory Save and Recall
Scope memories: 10 memory locations that each can contain two waveforms plus corresponding setup Recorder memories: 2 memory locations that each can contain 100 captured dual input scope screens, or a dual input ScopeRecord (27,500 Min-Max pairs per input), or a dual input Trendplot (18,000 min-max pairs + average values)
Real-Time Clock
Time and date stamp for ScopeRecord, 100 captured screens and TrendPlots
Mechanical Data
Size: 256 x 169 x 64 mm (10.1 x 6.6 x 2.5 in) Weight: 2 kg (4.4 lbs)
Case
Design: Rugged, shock proof with integrated protective holster
Drip and dust proof: IP51 according to IEC529 Shock and Vibration: Shock 30 g, Vibration
(sinusoidal) 3 g according to MIL-PRF-28800F Class 2
Display Bright full-color LCD Bright LCD
Brightness 80 Cd/m
Display size: 115.2 x 86.4 mm (4.54 x 3.4 in.) Resolution: 320 x 240 pixels Contrast and brightness: User adjustable,
temperature compensated
Power
Line power: Country specific line voltage adapter/battery charger included
Battery power: Rechargeable NiMH (installed) Battery operating time: 4 hours Battery charging time: 4 hours Battery power saving functions: Auto power down
with adjustable power down time. On screen battery status indicator
Fluke 190C Fluke 190B
with backlight, with backlight
2
using power adapter using power adapter
typ. 125 Cd/m2typ.
Safety
Compliance:
N61010-1 (2nd Edition) Pollution degree 2
E
L 3111-1
U
AN/CSA C22.2 No.1010.1
C
NSI/ISA S82.01
A
Input Voltage Ratings
Maximum probe voltage: 1,000 V CAT II/600 V CAT III
(Maximum voltage between 10:1 probe tip (VPS200) and reference lead)
Floating voltage: 1,000 V CAT II/600 V CAT III
(Maximum voltage between earth ground and any terminal (signal input or shielding))
Independently isolated inputs:
1,000 V CAT II/600 V CAT III
(Maximum voltage between any terminal of one input or probe (VPS200) and any other terminal of another input or probe (VPS200))
Maximum voltage on BNC input directly (input A or B):
300 V CAT III
Maximum voltage on meter input:
1,000 V CAT II/600 V CAT III
Environmental
Operating temperature: 0 °C to +50 °C Storage temperature: -20 °C to +60 °C Humidity:
10 °C to 30 °C: 95% RH non condensing 30 °C to 40 °C: 75% RH non condensing 40 °C to 50 °C: 45% RH non condensing
Maximum operating altitude: 3,000 m (10,000 feet) Maximum storage altitude: 12 km (40,000 feet) Electro-Magnetic-Compatibility (EMC): EN 61326-1
for emission and immunity
Optically Isolated PC / Printer Interface
PC communication: Transfer instrument settings, screen images, waveform data and waveform
ompatible w
referenc for Window
T
FX optional P
es, c
®
s
ia optional P
o printer:
v
Supports H
/LQ, Seiko DPU-414 and Postscript printers via
AC 91
ith FlukeView
0
M9
aserjet
P L
80.
®
, D
®
eskJ
software
et, Epson
Warranty
3 years, parts and labor on mainframe instrument 1 year on accessories
Accessories
Standard Accessories Fluke 199C, Fluke 196C, Fluke 199B, Fluke 196B, Fluke 192B
Rechargeable battery pack (installed) BP190 Line voltage adapter / Battery charger BC190 Voltage probes and accessories 10:1 voltage probe (VPS200, 1 red + 1 grey) including hook clip, ground lead with mini alligator clip, ground lead with
Multimeter test leads TL75 Hard Point test lead set (1red, 1 black) User manual 10 language versions on CD-ROM, “Getting Started” booklet included with instrument
4 Fluke Corporation ScopeMeter 190C/190B Series and ScopeMeter 120 Series
hook clip, 4 mm add-on probe tip, ground lead to 4 mm banana plug
Technical Specifications
ScopeMeter 120 Series
Oscilloscope Mode
Vertical deflection
idth and Risetime Fluke 124 Fluke 123
Bandw
idth (risetime)
Bandw
with VPS40 probes 40 MHz 20 MHz
input A and B directly 40 MHz 20 MHz
ith STL120 Shielded Test Leads 12.5 MHz 12.5 MHz
w
Instrument risetime (input directly)
Number of inputs: 2 Input coupling: AC, DC with ground level indicator Input sensitivity: 5 mV to 500 V/div (with the
included VPS40 (Fluke 124) and STL120 shielded test leads measure up to 600 Vrms, CAT III) Input voltage: 600 V CAT III. See General Specifica­tions for more detailed information
Vertical resolution: 8 bit Accuracy: ± (2 % of reading + 0.05 x range/div) Input impedance: 1 M
shielded test leads; 1 M BB120; 5 M±1 % // 15.5 pF with VPS40, 10:1 Volt­age probe
±1 % // 225 pF with STL120
Horizontal
Max. sample rate (both channels simultaneously):
Fluke 124: 2.5 GS/s for repetitive signals; 25 MS/s for single shot Fluke 123: 1.25 GS/s for repetitive signals; 25 MS/s for single shot
Number of digitizers: 2 Time base range: 10 ns/div to 1 min/div (Fluke 124);
20 ns/div to 1 min/div (Fluke 123) Maximum record length: 512 Min-Max points per input
Accuracy: ± (0.1 % of reading + 1 pixel) Glitch detect: 40 ns
y and acquisition
Displa
Display modes: Input A, input A and B, envelope, smooth Acquisition modes: Normal (including glitch capture), single shot, roll
Trigger and delay
Source: Input A, input B, external via optional ITP120 Modes: Automatic Connect-and-View, Free Run, Edge,
Single Shot, Video, Video Line Connect-and-View: Advanced automatic triggering that recognizes signal patterns and automatically sets up and continuously adjusts triggering, time base and amplitude. Automatically displays stable pictures of complex and dynamic signals like motor drive and control signals Video triggering: NTSC, PAL, PAL+, SECAM. Includes line select Time delay: Up to 10 divisions pre-trigger view
Measurements
Vdc, Vac, Vac+dc, Vpeak max, Vpeak min, Vpeak to peak, frequency (Hz), positive pulse width, negative pulse width, positive duty cycle, negative duty cycle, Amp ac, Amp dc, Amp ac+dc, Phase, Temperature °C, Temperature °F, dBV, dBm into 50 and 600 . (Amps, °C or °F with optional probes)
8.75 ns 17.5 ns
±1 % // 20 pF ± 3 pF with
Cursor Measurements (124 only)
ces:
Sour Modes: Single or dual vertical cursor, dual horizontal
cursor, rise- or falltime
Measur
Single vertical line: A time from start of rec Dual vertical lines: cursors, 1 Dual horizontal lines: High, low or V - readout, rise- and falltime: transition time, 0 %-level, 100 %­level, with markers at 10 % and 90 % Accuracy: As oscilloscope
Input A, Input B
ements:
verage, min value, max value,
ording in roll mode
V at markers, time between
/T between cursors (in Hz)
Dual Input Meter
The specified accuracy is valid over the temperature range 18 °C to 28 °C (64 °F to 82 °F). Add 10 % of specified accuracy for each °C below 18 °C or above 28 °C (64 °F to 82 °F) Max. meter bandwidth: 40 MHz (Fluke 124), 20 MHz (Fluke 123)
Voltage measurements
Measurement selection: Vdc, Vac rms, Vac+dc rms, Vpeak max, Vpeak min, V
Ranges: 500 mV, 5 V, 50 V, 500 V, 1250 V Full scale reading: 5,000 counts
Accuracy
± (0.5 % + 5 counts)
Vdc: Vac rms:
1 Hz to 60 Hz: ± (1 % + 10 counts) 60 Hz to 1 kHz: ± (2.5 % + 15 counts) 20 kHz to 1 MHz ± (5 % + 20 counts)
Vac+dc true-rms:
DC to 60 Hz: ± (1 % + 10 counts) 60 Hz to 1 kHz: ± (2.5 % + 15 counts) 20 kHz to 1 MHz (5 % + 20 counts)
Vpeak:
Max peak or Min peak: 5% of full scale Peak-to-peak: 10% of full scale
Ohms
Ranges: 500 , 5 k, 50 k, 500 k, 5 M, 30 M Max. resolution: 5,000 counts Accuracy: ± (0.6 % of reading + 5 counts)
Capacitance
anges:
R Max
ccuracy:±(2 % of reading + 10 counts)
A
50 nF to 500
. resolution:
5,000 c
Other meter functions
Frequency: Up to 70 MHz (Fluke 124) or up to 40 MHz (Fluke 123)
Continuity: Beeper on < 30 Diode test: Up to 2.8 V Amps: Amp dc, Amp ac, Amp ac+dc using an optional
current clamp or shunt. Scaling factors: 0.1 mV/Amp to 100 V/Amp Temperature (°C, °F): With optional accessories. Scale factors 1 mV/°C or 1 mV/°F
Number of inputs: 2 Input impedance: 1 M± 1 % // 10 pF ± 2 pF Advanced meter functions: Auto/manual ranging,
TouchHold®, Relative measurements (zero reference), TrendPlot recording
opeMeter 1
Sc
pk-pk
µF
ounts
/190B Series and ScopeMeter 120 Series Fluke Corporation 5
0C
9
Recorder Mode
Trendplot recording
Dual input electronic paperless chart recorder. Plots and displays the actual, minimum, maximum and average of any measurement.
Source and display: Input A, Input A and B Range: 15 s/div to 2 days per division (automatic) Recorded timespan: Up to 16 days with a resolution
of 1.5 hours Recording mode: Continuous with automatic vertical scaling and horizontal time compression Measurement speed: 2.5 measurements per second maximum Horizontal scale: Time from start
General Specifications
Case
Design: Rugged, shock proof with integrated protective holster
Drip and dust proof: IP51 according to IEC529 Shock and vibration: Shock 30 g Vibration 3 g
(sinusoidal) according to MIL-PRF-28800F Class 2
Display
Bright LCD with CCFL backlight, 60 (35) cd/m with (without) power adapter
Size: 72 x 72 mm (2.8 x 2.8 inch) Resolution: 240 x 240 pixels Contrast and brightness: User adjustable,
temperature compensated
Memory Save and Recall
20 (10 in Fluke 123) instrument screens with user set-ups and user text
Real-time clock
Time and date stamp TrendPlot recording
Power
Line power: Country specific line voltage adapter/batter
Battery p
in Fluke 1 Fluke 1
Battery oper
up to 5 hours using B
Battery char
(Fluke 1
Battery p
ith adjustable power down time. On-screen battery
w power indicator
y charger included
ower:
Rechargeable Ni-
MH BP130 (installed
24) or rechargeable NiCd BP120 (installed in
23)
ating time:
Up to 7 hours using B
P120
ging time:
5 hours (Fluke 1
24)
ower saving functions:
Auto power dow
2
P130,
23), 7 hours
n
Mechanical data
Size: 50 x 1
eight:1.2 kg (2.64 lbs)
W
15 x 232 mm (2 x 4.5 x 9.1 in)
Safety
Compliance: EN61010.1 (2nd Edition) Pollution degree 2, UL3111-1, CAN/CSA-C22.2 No. 1010.1, ANSI/ISA S82.01
Input voltage ratings
Maximum input voltage: 600 V CAT III
(Maximum voltage between input and reference lead)
Maximum input voltage using VPS40 Probe: 600 V CAT III, 1000 V CAT II
(Maximum voltage between
probe tip input and reference lead)
Floating voltage: 600 V CAT III
(Maximum voltage between earth ground and any terminal (signal input or reference lead))
Maximum voltage between reference leads:
Instrument has common grounds connected via self recovering fault protection. For different ground potential measurements between inputs use DP120 differential voltage probe
Environmental
Operating temperature: 0 °C to +50 °C Storage temperature: -20 °C to +60 °C Humidity:
10 °C to 30 °C, 95% RH non condensing; 30 °C to 40 °C, 75% RH non condensing; 40 °C to 50 °C, 45% RH non condensing Maximum operating altitude: 2,000 m (6,500 feet); 3,000 m (10,000 feet) voltages 400 V
Maximum storage altitude: 12 km (40,000 feet) Electro-Magnetic Compatibility:
Emission EN50081-1 (EN55022 and EN60555-2) Immunity EN50082-2 (IEC1000–4-2, -3, -4, -5)
Optically isolated PC/Printer interface
To printer: Supports HP Laserjet,®Deskjet,®Epson FX/LQ and postscript printers via optional PAC91 To PC: Transfer instrument settings, screen images and data, compatible with FlukeView®software for Windows®via optional PM9080
arranty
W
3 years, parts and lab
essories
1 year on ac
c
or on mainframe instrument
Accessories
Standard Accessories Fluke 123, Fluke 124
Rechargeable battery pack (installed) BP120 (Fluke 123)
Line voltage adapter / Battery charger PM8907 Voltage probes and accessories STL120 Shielded Test lead set; VPS40 high impedance 10:1 probe, 40 MHz (1 black, included with Fluke 124 only);
Multimeter test leads User manual 15 language versions on CD-ROM, “Getting Started” booklet included with instrument
6 Fluke Corporation ScopeMeter 190C/190B Series and ScopeMeter 120 Series
20 hook clips; g
C1
H
round leads w
ith mini alligator clips; AC120 alligator clips; BB120 BNC-to-Shielded banana adapter
TL75 Hard Point test lead (1 black)
30 (Fluke 1
P1
B
24)
International Safety Standards
Measurement Voltage
y Summary description
Categor
CAT IV Three phase at utility connection, any outdoors
CAT III Three-phase distribution (under 1,000 V),
AT II Single-phase receptable connected loads
C
AT I Electronic
C
onductors (under 1,000 V)
c
Outside and service entrance
• Service drop from pole to building Run between meter and panel
• Overhead line to detached building
• Underground line to well pump
y III source
ommercial lighting
0 ft)
including single phase c and distribution panels
• Feeders and short branch circuits Distribution panel devices
• Heavy appliance outlets with “short” connections to service entrance
Outlets and long branch circuits
• All outlets at more than 10 m (30 ft) from Categor All outlets at more than 20 m (6
• from Category IV source
• Electronic equipment
• Low energy equipment with transient limiting protection
To protect your instrument and – more importantly – yourself, choose a test tool that can withstand the electrical hazards present in the environment in which you plan to use it.
EN61010 establishes international safety require­ments for electrical measurement equipment. It separates the various electrical environments into
voltage-energy transients. To choose the right tool, the voltage rating alone does not determine the safety. It is the combination of voltage rating and installation category that determines maximum tran­sient withstand capability of the tool. CAT III rated instruments are recommended for measurement on industrial power distribution systems.
installation categories based on the danger from high
FlukeView®ScopeMeter®Software for Windows
FlukeView software adds PC power to your Fluke ScopeMeter Test Tools.
FlukeView Sc out of your Sc
Store instrument
ith Fluke 1
(w (Fluke 1
Copy color screen images into your reports and
documentation (color screen images with Fluke 190C Series only)
Capture and store waveform data from your
ScopeMeter on your PC Create and archive waveform references for auto-
matic (Fluke 190C Series) or visual (Fluke 190B and 190C Series) comparison
Use cursors for parameter measurement
Includes wa
analysis Copy waveform data into your spreadsheet for
detailed analysis Extended recording of up to four user-selected
measurements help you monitor and analyze slow moving signals and related events
Logging of other readings directly into other appli-
cation programs, eg., spreadsheet
opeMeter software helps you get more
opeMeter:
’s screen c
0C Series) or in black and white
9
0B and 1
9
20 Series)
veform analysis, e.g
opies on the P
., FFT spectrum
C, in c
olor
Add user text to instrument setups and send
these to the instrument for operator reference and instructions
Capture complete Replay sequence into the PC for
further analysis and documentation English, French and German versions included on a
single CD-ROM
Note: Some functionality may be available with spe models only
System re
Pentium 90 or better
CD-ROM drive
Window
T 4.0 / 2000 / X
N One free RS 232 port
M9080 Optically isolated
P
S 232 adapter/cable,
R
vailable separately or
a
cific ScopeMeter
quirements
®
s
95 / 98 / Me /
P
included in
CC190/SCC120 kit and in
S
opeMeter ‘S’ versions
Sc
Supported Instruments
Full support for Fluke 199C, 199B, 199, 196C, 196B, 196, 192B, 192, 124, 123
®
opeMeter 1
Sc
/190B Series and ScopeMeter 120 Series Fluke Corporation 7
0C
9
Selection Guide
90C Color ScopeMeter Series 190B ScopeMeter Series 120 ScopeMeter Series
1
99C Fluke 196C Fluke 199B Fluke 196B Fluke 192B Fluke 124 Fluke 123
Bandwidth 200 MHz 100 MHz 200 MHz 100 MHz 60 MHz 40 MHz 20 MHz Max real time sample rate 2.5 GS/s 1 GS/s 2.5 GS/s 1 GS/s 500 MS/s 25 MS/s 25 MS /s Max equivalent time
sample rate
y 144 mm full color LCD 144 mm monochrome LCD 102 mm monochrome LCD
Displa
ital persistence Yes, gives analog oscilloscope like
Dig
Envelope mode
aveform compare Visual reference and automatic Visual reference
W
FT analysis Windowing: Auto, Hamming,
F
Max record length
ope Mode: 3000 points per input channel
in Sc
opeRecord mode: 27,000 points per input or more (5 ms/div to 2 min/div)
in Sc
er of inputs 2 plus external / DMM input, all isolated from each other and from ground 2
Numb Number of digitizers 2 2 Independently floating
isolated inputs Input sensitiv Glitch capture Timebase range in
Scope mode Trigger types Connect-and-View
Scope measurements
Waveform mathematics A + B, A - B, A x B, A versus B (X-Y-mode, giving Lissajous diagrams)
ScopeRecord trigger modes Start on trigger, stop on trigger, stop in absence of trigger signal — Capture last 100 screens Automatic, with replay capability — Dual input TrendPlot Yes, with cursors and zoom Yes Memory for screens and 10 screens with set-up;
set-ups 5 more memories are made available upon registration of the ScopeMeter Memory for recordings Two, each can store 100 scope screens, a ScopeRecord or a TrendPlot True-rms multimeter 5000 counts, Volts, Amps, Ohms, Continuity, Diode, Temp Safety certified 1000 V CAT II / 600 V CAT III (instrument and included accessories) 600 V CAT III (instrument and
(EN61010-1) included accessories) Battery (installed) 4 hr Ni-MH BP190 7 hr NiMH 5 hr NiCd Line Power Adapter / battery-charger included Size 25 x 16.9 x 6.4 mm (10.1 x 6.7 x 2.5 in) 232 x 115 x 50 mm (9.2 x 4.5 x 2 in) Weight 2 kg (4.4 lb) 1.2 kg (2.64 lbs) PC and Printer Interface Using optional optically isolated RS-232 adapter/cable Warranty 3 years on main instrument, 1 year on the standard accessories
ity 2 mV/div to 100 V/div 5 mV/div. to 100 V/div 5 mV/div to 500 V/div
Fluke 1
waveform decay (user selectable)
Up to 3 ns using pulse w
+ Automatic Vrms and watts cursors + 26 26 automatic
measurement on cursor limited
+ Frequency Spectrum (FFT)
Yes Yes Yes
ass / Fail’ testing
‘P
Hanning or none
N-Cycle, Dual Slope
part of waveform measurements
2.5 GS/s 1.25 GS/s
——
——
512 min/max points per input
000 V between inputs, references and ground
Up to 1
idth triggering; 50 ns peak detect at 5 ms/div to 1 min/div
5 ns/div to 2 min/div
, Free Run, Single Shot, Edge, Delay, Video Frame, Video Line Connect-and-View
Selectable pulse width and external Single Shot, Edge, Video
7 cursor measurements, 30 automatic measurements
Detailed technical specifications, optional accessories and a virtual product demo can be found on the Fluke web site. Check it out at:
10 ns/div to 10 ns/div to 20 ns/div to
2 min/div 1 min/div 1 min/div
automatic measurements
ww.fluke.com/scopemeter
w
40 ns
20 10
, Free Run,
.
Ordering Information
opeMeter (200 MHz / 2.5 GS/s)
Color Sc
S
/
99C
Fluke-1
Fluke-199C Color ScopeMeter (200 MHz / 2.5 GS /s) Fluke-196C/S Color ScopeMeter (100 MHz / 1 GS/s)
Fluke-196C Color ScopeMeter (100 MHz / 1 GS/s) Fluke-1
Fluke-1 Fluke-1
Fluke-196B ScopeMeter (100 MHz / 1 GS/s) Fluke-192B/S ScopeMeter (60 MHz / 500 MS/s) with
Fluke-192B ScopeMeter (6 Fluke-124/S Industrial Sc
Fluke-124 Industrial ScopeMeter, 40 MHz Fluke-123/S Industrial ScopeMeter, 20 MHz, with
Fluke-123 Industrial ScopeMeter, 20 MHz BP120 Rechargeable NiCd Battery for use with
with SCC190 kit
with SCC190 kit
opeMeter (200 M
Sc
S
/
99B
SCC190 kit
99B ScopeMeter (200 MHz / 2.5 GS/s)
ScopeMeter (100 MHz / 1 GS/s) with
S
/
96B
SCC190 kit
SCC190 kit
SCC120 kit
SCC120 kit
Fluke 120 Series
Hz / 2.5 G
Hz / 500 MS/s)
0 M
opeMeter, 40 M
S
Hz, w
/s) with
ith
y for use with
0 Series Sc
H Batter
S232 adapter/cable
opeMeters
ying case for
opeMeters
B
BP190 Rechargeable NiMH Battery for use with
SW90W FlukeView ScopeMeter Software for
M9
P SCC190 Software - Cable - Case kit for Fluke
C
S
C195 Durable, universal soft carrying case for
C190 Hard shell protective carrying case for
C1
C1
DP120 Differential Voltage Probe for use with
VPS40 40 MHz, 10:1 Voltage probe set for use
Fluke 120 Series
Fluke 190 Series ScopeMeters
Windows Optically isolated R
0
8
0
90 Series
1 Software - Cable - Case kit for Fluke
20
C1
120 Series
ScopeMeters and accessories
9
Fluke 1 Durable, protective soft carr
25
Fluke 120 Series ScopeMeters
20 Hard shell protective carrying case for
20 Series Sc
Fluke 1
Fluke 120 Series
with Fluke 120 Series
Rechargeable NiM
30
P1
Fluke. Keeping your world
up and running.
Fluke Corporation
O Box 9090, Everett, WA USA 98206
P
.
.V
Fluke Eur
PO Box 1186, 5602 BD Eindhoven, The N
F
In the U.S.A. (800) 443-5853 or Fax (425) 446-5116 In Europe/ Fax +31 (0) 40 2675 222 In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other countries (425) 446-5500 or F W
©2004 Fluke Corporation. All rights reserved. Trademark Printed in U.S.A. 3/2004 1629083 D-ENG-N Rev D
ope B
etherlands
e information call:
or mor
East/Africa +31 (0) 40 2675 200 or
-
M
6
1
ax (425) 446-5
eb ac
1
ess: http://www
c
s are the property of their respective ow
.fluke.c
om/
ners
.
Loading...