Flowline US06 Operating Manual

Echotouch®
Ultrasonic Level Transmitter
US01, US03, US06 & US12 Series Manual
  
Flowline, Inc. | 10500 Humbolt Street, Los Alamitos, CA 90720 p 562.598.3015 f 562.431.8507 w flowline.com MN300820 Rev B
Introduction / Table of Contents Step One
An ultrasonic sound wave is pulsed from the base of the transducer. The sound wave reflects against the process medium below it. The sound wave energy is returned to the transducer. The microprocessor based electronics measures the time of flight between the sound pulse generation and its receipt. This translates into the distance or range between the transducer and process media below.
NEW FEATURES
Reflective Ultrasonic Technology Simple configuration with WebCal
®
software or push button (some models) configuration
Increased temperature range Increased output filtering
TABLE OF CONTENTS
Introduction: ...................................................................................................................................................... 2
Specifications: ........................................................................................................................................... 4
Dimensions: .............................................................................................................................................. 5
Intrinsically Safe Control Drawings: .......................................................................................................... 7
Safety Precautions: ................................................................................................................................... 9
Getting Started: ................................................................................................................................................... 11
Setup Overview: ...................................................................................................................................... 11
Components: ........................................................................................................................................... 12
WebCal vs. Display Configuration: ......................................................................................................... 14
Understanding Sensor Height (Height): .................................................................................................. 15
Understanding Fill-Height (Fill-H): ........................................................................................................... 16
Sensor Output to Local Display:: ............................................................................................................ 18
Configuration Using WebCal
®
: ............................................................................................................................ 19
Step #1 – Install WebCal Software: ........................................................................................................ 20
USB Fob Interface: ...................................................................................................................... 21
Step #2 – Measure the Tank:: ................................................................................................................. 22
Step #3 – Sensor Configuration: ............................................................................................................. 23
Step #4 – Dimensional Entry: ................................................................................................................. 24
Step #5 – Tank Level Confirmation: ........................................................................................................ 25
Step #6 – Write to Unit: ........................................................................................................................... 25
Configuration Using Display: .............................................................................................................................. 26
Step #1 – Measure Tank: ........................................................................................................................ 27
Step #2 – Setting the Units of Measurement: ......................................................................................... 28
Step #3 – Setting the Height: .................................................................................................................. 29
Step #4 – Setting the Fill-H: .................................................................................................................... 30
Step #5 – Setting the Fail-Safe Current Output (LOST): ........................................................................ 31
Étape #1 - Mesurer le Tank: ................................................................................................................... 32
Étape #2 – Réglade des Unités de Mesure (UNITS): ............................................................................. 33
Étape #3 – Réglage de la Height: ........................................................................................................... 34
Étape #4 – Réglage de la Fill-H: ............................................................................................................. 35
Étape #5 – Comment Sélectionner une Sortie Actuelle Faible (LOST): ................................................. 36
| 2 MN300820 Rev B
Introduction / Table of Contents Step One
Installation: .......................................................................................................................................................... 37
Mounting Guide: ...................................................................................................................................... 37
Fitting Selection: ..................................................................................................................................... 39
Wiring: ................................................................................................................................................................ 42
Analog Output (4-20 mA): ....................................................................................................................... 42
Wire Connections: ................................................................................................................................... 43
Accessing the Terminals (US06 & US12 Series Only): .......................................................................... 43
Common Wiring to Displays, Controllers & PLCs: .................................................................................. 46
WebCal
Sensor Configuration: ............................................................................................................................. 49
Volumetric Configuration: ........................................................................................................................ 51
Tank Level Confirmation: ........................................................................................................................ 54
Write to Unit: ........................................................................................................................................... 55
Display Appendix: ............................................................................................................................................... 56
Air Gap vs Liquid Level: .......................................................................................................................... 56
How to reverse the current output (Rev mA): ......................................................................................... 57
How to Setup to Start-up Power: ............................................................................................................ 58
Values Menu: .......................................................................................................................................... 59
SETUP: ....................................................................................................................................... 59
Diagnostic (DIAG) parameters: ................................................................................................... 59
Reset: .......................................................................................................................................... 60
Factory Settings: ..................................................................................................................................... 61
User Settings: ......................................................................................................................................... 61
Troubleshooting: ..................................................................................................................................... 62
Warranty: ............................................................................................................................................................ 64
®
Appendix: ............................................................................................................................................ 49
MN300820 Rev B 3 |
Introduction (continued) Step One
SPECIFICATIONS
Electrical:
Supply Voltage: 14-28 Vdc Signal: 4 to 20 mA current loop; 22 mA
max. Signal invert: 4-20mA or 20-4mA Signal fail-safe: 4mA, 20mA, 21mA, 22mA or
hold last value
Mechanical (US06 & US12 series): Display Module: Removable via short cable and
plug
LCD: 6 characters, 10 mm (0.394”)
height
Buttons: Three, Up / DN / SELECT
Pin Connection: Display can be removed such
that the open connection allows
the sensor to be WebCal
capable.
Enclosure: Metal thermal head with skirt to
support transducer w/ conduit.
* Earth ground on package
external (min. 14 AWG)
Electrical Connections: Power & Return: 2-pin quick connect terminal
block
WebCal Connect.: 2-pin quick connect terminal
block
Display Connect.: 5-pin plug and socket Conduit Connection: Single, 1/2 inch NPT Transducer: Polyvinylidene Fluoride (PVDF) Terminal Cable Size: 16 to 24 AWG Process Mount: US06: 2" NPT or 2” G US12: 3" NPT or 3” G Gasket Viton A (G thread only) Classification: Intrinsically Safe Compliance: CE, RoHS
Mechanical (US01 & US03 series): Enclosure: Polypropylene (PP-FR GF30) Transducer: Polyvinylidene Fluoride (PVDF) Cable jacket mat’l: Polyvinyl Chloride (PVC) Cable type: 4-conductor with shield Cable length: 4’ (121.9 cm)
Cable Min. Temp.: 200°C Process Mount: 2" NPT or 2” G Gasket Viton A (G thread only) Classification: Intrinsically Safe
Compliance: CE, RoHS
Environmental: Temperature: F: -40° to 176° C: -40° to 80° Temp. compensation: Automatic Ambient temp.: F: -31° to 140° C: -35° to 60° Pressure: 30 psi, derated @ 1.667 psi per degree C above 25 °C Enclosure: US12: TYPE 4X, IP66 US06: TYPE 4X, IP66 US03: TYPE 6P, IP68 US01: TYPE 6P, IP68
Functional: Range: US12: 12 meters (39.37 feet) US06: 6 meters (19.69 feet) US03: 3 meters (9.84 feet) US01: 1.5 meters (4.92 feet) Dead band: US12: 45cm (18 inches) US06: 20cm (8 inches) US03: 10cm (4 inches) US01: 3.81cm (1.5 inches) Accuracy: US12, US06, US03: 0.2% of
Range in air at 20 °C US01: 0.125” (3mm) Resolution: 0.125" (3mm) Beam width: 2” (5cm) Menu: Self-scrolling (US06 & US12
only)
| 4 MN300820 Rev B
Introduction (continued) Step One
DIMENSIONS
Dimensions for US01 & US03 Series
Dimensions for US06 Series
MN300820 Rev B 5 |
Introduction (continued) Step One
DIMENSIONS
Dimensions for US12 Series
REFLECTIVE TECHNOLOGY
Condensation is the most common variable in liquid level applications. Condensation attenuates the acoustic signal of ultrasonic sensors with horizontal transducers, weakening their signal strength and signal to noise ratio by up to 50%, and substantially reducing their measurement reliability. At the core of Reflective Technology™ is a simple fact. Unlike flat horizontal surfaces, significant water droplets cannot adhere to smooth vertical surfaces. By orienting the transducer vertically, condensation runs off the transducer and does not affect sensor performance. The unimpeded transmit and receive signals are redirected to and from the liquid off a 45º reflector, delivering reliable level measurement.
FLAT TRANSDUCER REFLECTIVE TRANSDUCER
Signal Attenuation Reliable Measurement
| 6 MN300820 Rev B
Introduction (continued) Step One
INTRINSICALLY SAFE CONTROL DRAWING 600820 (US06 & US12 SERIES):
Intrinsically Safe Approvals:
Intrinsic Safe: 14-28 Vdc 22mA Output: 4-20mAdc Temp: -40°C to +80°C Type 4X/IP66: US06 & US12 series only Approval Agency:
CSAus
c
General: CE, RoHS
Intrinsically Safe Entity Parameters:
U I P L C
= 28 VDC
i/Vmax
= 100 mA
i/Imax
= 840 mW
i
= 2.46 mH
i
= 0 μF
i
CSA for U.S.: Class I, Div 1, Groups A, B, C and D T4; Class II, Div 1, Groups E, F, G T4; Class III, T135°C; Class I, Zone 0, IIC AEx ia T4 Ga, Class II, Zone 20, IIIC AEx ia, T135°C, Da CSA for Canada: Class I, Division 1, Groups A, B, C and D T4; Class II, Division 1, Groups E, F, G T4; Class III, T135°C; Ex ia IIC T4 Ga; Ex ia IIIC T135°C Da; IECEx: Ex ia IIC T4 Ga; Ex ia IIIC T135°C Da Tamb: -40°C tp 80°C
Install as per Control Drawing 600820
MN300820 Rev B 7 |
Introduction (continued) Step One
INTRINSICALLY SAFE CONTROL DRAWING 610125 (US01 & US03 SERIES):
Intrinsically Safe Approvals:
Intrinsic Safe: 14-28 Vdc 22mA Output: 4-20mAdc Temp: -40°C to +80°C Type 6P/IP68: US01 & US03 series only Approval Agency: General: CE, RoHS
Intrinsically Safe Entity Parameters:
U I P L C
= 28 VDC
i/Vmax
= 100 mA
i/Imax
= 840 mW
i
= 2.46 mH
i
= 0 μF
i
CSAus
c
CSA for U.S.: Class I, Div 1, Groups A, B, C and D T4; Class II, Div 1, Groups E, F, G T4; Class III, T135°C; Class I, Zone 0, IIC AEx ia T4 Ga, Class II, Zone 20, IIIC AEx ia, T135°C, Da CSA for Canada: Class I, Division 1, Groups A, B, C and D T4; Class II, Division 1, Groups E, F, G T4; Class III, T135°C; Ex ia IIC T4 Ga; Ex ia IIIC T135°C Da; IECEx: Ex ia IIC T4 Ga; Ex ia IIIC T135°C Da Tamb: -40°C tp 80°C
Install as per Control Drawing 610125
| 8 MN300820 Rev B
Introduction (continued) Step One
SAFETY PRECAUTIONS
About this Manual: PLEASE READ THE ENTIRE MANUAL PRIOR TO INSTALLING OR USING THIS
PRODUCT. This manual includes information on the EchoTouch
®
series Ultrasonic Level Transmitter from FLOWLINE. Please refer to the part number located on the switch label to verify the exact model configuration, which you have purchased.
User’s Responsibility for Safety: FLOWLINE manufactures a broad range of level sensing technologies. While each of these sensors is designed to operate in a wide variety of applications, it is the user’s responsibility to select a sensor model that is appropriate for the application, install it properly, perform tests of the installed system, and maintain all components. The failure to do so could result in property damage or serious injury.
Proper Installation and Handling: Only professional staff should install and/or repair this product. Never over tighten the sensor within the fitting. Always check for leaks prior to system start-up.
Wiring and Electrical: A supply voltage of 14 to 28 VDC is used to power the Echotouch®. Electrical wiring of the transmitter should be performed in accordance with all applicable national, state, and local codes.
Material Compatibility: The enclosure for the US06 & US12 series is made of metal and for the US01 and US03 series the enclosure is made of Polypropylene. The transducer is made of Polyvinylidene Fluoride (PVDF). Make sure that the model, which you have selected, is chemically compatible with the application media.
Enclosure: While the sensor housing is liquid-resistant the EchoTouch® is not designed to be operational when immersed. It should be mounted in such a way that the enclosure and transducer do not come into contact with the application media under normal operational conditions.
Warning - Potential Electrostatic Charging Hazard (Handling Static-Sensitive Circuits/Devices): When handling the transmitter, the technician should follow these guidelines to reduce any possible electrostatic charge build-up on the technician’s body and the electronic part.
1. Always touc h a known good ground source before handling the part. This should be repeated while handling the part and more frequently after sitting down from a standing position, sliding across the seat or walking a distance.
2. Avoid touching electrical terminals of the part unless making connections.
3. DO NOT open the US06 or US12 unit cover until it is time to configure.
Warning: Always use the Viton gasket when installing “G” threaded versions of EchoTouch® Warning: Make sure that all electrical wiring of the switch is in accordance with applicable codes. Warning: Use only specified replacement parts or Intrinsic Safety can be impaired. Warning: To prevent ignition of explosive atmospheres, disconnect power before servicing.
Make a Fail-Safe System: Design a fail-safe system that accommodates the possibility of switch and/or power failure. FLOWLINE recommends the use of redundant backup systems and alarms in addition to the primary system.
Flammable, Explosive or Hazardous Applications: Only the EchoTouch® is rated for use in hazardous locations. Refer to the certificate of compliance for all applicable intrinsically safe ratings and entity parameters for the EchoTouch
®
. Refer to the National Electric Code (NEC) for all applicable installation requirements in
hazardous locations.
MN300820 Rev B 9 |
Introduction (continued) Step One
PRÉCAUTIONS DE SÉCURITÉ
A Propos de ce Manuel: VEUILLEZ LIRE TOUT LE MANUEL AVANT D'INSTALLER OU D'UTILISER CE
PRODUIT. Ce manuel comprend des informations sur l'émetteur de niveau ultrasonique EchoTouch® de FLOWLINE. Reportez-vous au numéro de pièce situé sur l'étiquette du commutateur pour vérifier la configuration exacte du modèle que vous avez achetée.
Responsabilité de l'utilisateur pour la Sécurité: FLOWLINE fabrique une large gamme de technologies
de détection de niveau. Bien que chacun de ces capteurs soit conçu pour fonctionner dans une grande variété d'applications, il est de la responsabilité de l'utilisateur de sélectionner un modèle de capteur approprié à l'application, de l'installer correctement, d'effectuer des tests du système installé et de maintenir tous les composants. Le défaut de le faire pourrait entraîner des dommages matériels ou graves.
Installation et Manipulation Appropriées: seul le personnel professionnel doit installer et / ou réparer ce
produit. Ne jamais serrer le capteur sur le raccord. Vérifiez toujours les fuites avant le démarrage du système.
Câblage et électricité: une tension d'alimentation de 14 à 28 VDC est utilisée pour alimenter
l'Echotouch®. Le câblage électrique de l'émetteur doit être effectué conformément à tous les codes nationaux, d'état et locaux applicables.
Compatibilité des Matériaux: LeboîtierdessériesUS06etUS12estenmétaletceluidessériesUS01etUS03en
polypropylène. Le transducteur est en fluorure de polyvinylidène (PVDF). Assurez-vous que le modèle, que
vous avez sélectionné, est chimiquement compatible avec les supports d'application.
Boîtier: Bien que le boîtier du capteur soit résistant aux liquides, l'EchoTouch® n'est pas conçu pour être
opérationnel lorsqu'il est immergé. Il doit être monté de telle sorte que l'enceinte et le transducteur ne soient pas en contact avec le support d'application dans des conditions normales de fonctionnement.
Avertissement - Danger potentiel de charge électrostatique (manipulation de circuits / dispositifs
sensibles à l'électricité statique): lors de la manipulation de l'émetteur, le technicien doit suivre ces directives pour réduire toute accumulation de charge électrostatique possible sur le corps du technicien et la partie électronique.
1. Toujours toucher une bonne source de sol connue avant de manipuler la pièce. Cela devrait être répété
lors de la manipulation de la pièce et plus fréquemment après avoir reposé sur une position debout, glisser sur le siège ou marcher sur une distance.
2. Évitez de toucher les bornes électriques de la pièce à moins de faire des connexions.
3. N'ouvrez PAS le capot de l'appareil US06 ou US12 avant la configuration.
Avertissement: Utilisez toujours le joint Viton lors de l'installation des versions filetées "G" d'EchoTouch® Avertissement: Assurez-vous que tout le câblage électrique du commutateur est conforme aux codes
applicables.
Attention: utilisez uniquement des pièces de rechange spécifiées ou la sécurité intrinsèque peut être
altérée.
Avertissement: Pour éviter tout allumage d'atmosphères explosives, débranchez l'alimentation avant
l'entretien.
Effectuez un système anti-défaillance: Concevez un système de sécurité qui permet de commuter et /
ou de panne de courant. FLOWLINE recommande l'utilisation de systèmes de sauvegarde redondants et d'alarmes en plus du système primaire.
Applications inflammables, explosives ou dangereuses: seul EchoTouch® est destiné à être utilisé
dans des endroits dangereux. Reportez-vous au certificat de conformité pour toutes les cotes de sécurité intrinsèquement applicables et les paramètres de l'entité pour EchoTouch®. Reportez-vous au National Electric Code (NEC) pour toutes les exigences d'installation applicables dans les endroits dangereux.
| 10 MN300820 Rev B
Getting Started Section Two
SETUP OVERVIEW
Below highlights the initial steps in setting up your sensor for operation.
1. Check Part Number (Section Two) a. Confirm that the sensor’s part number matches the ordered part number and all components
are provided with the model delivered.
2. Measurements, Installation & Software (Section Two) a. Understand the location placement of the sensor relative to Measurement Range including
Sensor Height and Fill-Height settings.
b. Download the WebCal® software from www.flowline.com/webcal-software/.
3. Configure Sensor with WebCal™ (Section Three) a. Section 3 contains information on using the WebCal
4. Configure Sensor with Push Button (Section Four) a. Section 4 contains information on using the Push Button Display to configure the sensor. b. Only applies to the US06 & US12 Series.
5. Install the sensor (Section Five) a. Section 5 contains information about the sensor location placement and its mechanical
installation.
6. Wire the sensor (Section Six) a. Section 6 contains information about the sensor’s electrical wiring and power requirements.
Note: When configuring Echotouch
®
, choose either the WebCal® or Push Button method. Either method will accomplish the goal of sensor configuration. Changes to the configuration can be made using the alternative method. When beginning with one method, it is recommended to complete the configuration before using the other method to make any adjustments.
®
configuration software.
MN300820 Rev B 11 |
Getting Started (continued) Section Two
COMPONENTS
Echotouch® is offered in different models. Depending on the model purchased, you may or may not have been shipped all the components shown below. A “G” threaded process mounts require a Viton® gasket for installation and operation of Echotouch
®
.
P/N
Max.
Range
US12-0001-00 US12-0001-01 Included US12-0011-00
39.4’
(12m)
US12-0011-01 Included US06-0001-00 US06-0001-01 Included US06-0011-00
19.7’ (6m)
US06-0011-01 Included US03-0001-00 US03-0001-01 Included US03-0011-00
9.8’
(3m)
US03-0011-01 Included US01-0001-00 US01-0001-01 Included US01-0011-00
4.9’
(1.5m)
US01-0011-01 Included
Dead Band
18”
(45.7cm)
8”
(20cm)
4”
(10cm)
1.5”
(3.81cm)
Thread Fob
3” NPT
3” G
2” NPT
2” G
2” NPT
2” G
2” NPT
2” G
Not Included
Not Included
Not Included
Not Included
Not Included
Not Included
Not Included
Not Included
Enclosure/
Connection
TYPE 4X /
½” Conduit
TYPE 6P /
4’ Pigtail
Configuration
Push Button or
WebCal
WebCal
®
®
only
Viton Gasket (G threaded version only)
o Part # 200129 – (US06-0013-0_ only) o Part # 210157 – (US12-0013-0_ only)
Liquid Tight Fitting (US06 & US12 series only)
o Part # LM90-1001 o Note: If using the Liquid Tight Fitting on
the ½” conduit, the cable minimum is
0.170” (4.3mm) and the maximum is
0.450” (11.4mm).
Quick Start Guide
| 12 MN300820 Rev B
Getting Started (continued) Section Two
COMPOSANTS
Echotouch® est offert dans différents modèles. Selon le modèle acheté, vous pouvez ou non avoir envoyé tous les composants indiqués ci-dessous. Un montage fileté "G" nécessite un joint Viton® pour l'installation et l'utilisation d'Echotouch
®
.
P/N
Max.
Gamme
US12-0001-00 US12-0001-01 Inclus US12-0011-00
39.4’
(12m)
US12-0011-01 Inclus US06-0001-00 US06-0001-01 Inclus US06-0011-00
19.7’ (6m)
US06-0011-01 Inclus US03-0001-00 US03-0001-01 Inclus US03-0011-00
9.8’
(3m)
US03-0011-01 Inclus US01-0001-00 US01-0001-01 Inclus US01-0011-00
4.9’
(1.5m)
US01-0011-01 Inclus
Dead Band
18”
(45.7cm)
8”
(20cm)
8”
(20cm)
4”
(10cm)
Fil Fob
3” NPT
3” G
2” NPT
2” G
2” NPT
2” G
2” NPT
2” G
Not Inclus
Not Inclus
Not Inclus
Not Inclus
Not Inclus
Not Inclus
Not Inclus
Not Inclus
Enceinte/
Connexion
TYPE 4X /
½” Conduit
TYPE 6P /
4’ Natte
Configuration
Bouton
poussoir ou
WebCal®
WebCal
®
seulement
Viton Gasket (version filetée G uniquement)
o Part # 200129 – (US06-0013-0_ seulement) o Part # 210157 – (US12-0013-0_ seulement)
Liquid Tight Fitting (US06 & US12 série seulement)
o Part # LM90-1001 o Remarque: Si vous utilisez l'ajustement
liquide étanche sur le conduit ½ ", le minimum du câble est de 0,170" (4.3mm) et le maximum est de 0.450 "(11.4mm).
Guide de démarrage rapide
MN300820 Rev B 13 |
Getting Started (continued) Step Two
WEBCAL® VS. DISPLAY CONFIGURATION
EchoTouch® can be configured using either the free WebCal® software and USB® Fob or with the internal display module. Either method will accomplish the goal of sensor configuration. Changes to the configuration can be made using the alternative method. Upon implementing one methodology, it is recommended to complete the configuration before using the other methodology in making any adjustments.
EchoTouch memory, so any settings configured before installation will not be lost when the sensor is powered down. To start, the following information is required:
Basic Tank Information
Power:
Note: The HEIGHT and Fill-Height (FILL-H) settings also determine the 4 to 20 mA current span. The HEIGHT setting determines the 4mA position and the FILL-H setting determines the 20 mA position.
®
can be configured before installation or after installation. The transmitter features non-volatile
o HEIGHT (Sensor Height) – Distance from
the bottom of the tank to the bottom of the sensor.
o FILL-H (Fill-Height) – Distance from the
bottom of the tank to the fill-height of the liquid.
o Provide input power to the Echotouch®
| 14 MN300820 Rev B
Getting Started (continued) Step Two
UNDERSTANDING SENSOR HEIGHT (HEIGHT)
This is a critical setting for EchoTouch®. Sensor Height (SH) defines the location of the sensor from the bottom of the tank. The value must take into account the shape of the tank and any risers, fittings, structures or extensions associated with the tank or the installation (see examples below). The reference point for definition of the Sensor Height is always the bottom of the mounting nut.
Sensor Off-center
Simple Vertical Tank Dome Top Raises HEIGHT
Changes HEIGHT
Simple Open Top Tank Sensor Extends into Sump
Simple Horizontal Tank Riser Elevates HEIGHT
Cone Bottom
Elevates HEIGHT
Mounting Fixture Elevates HEIGHT
MN300820 Rev B 15 |
Getting Started (continued) Step Two
UNDERSTANDING FILL-HEIGHT (FILL-H)
This is another critical setting for EchoTouch®. FILL-H (Fill-Height) defines the location of the highest point in the tank where the sensor will read level changes. When the level rises above FILL-H, the sensor wi ll read full (as long as the level does not enter the dead band). The reference point for definition of FILL-H is always from the bottom location of the Sensor Height. The value must take into account the sensors dead band, any risers, fittings, structures or extensions associated with the tank or the installation as well as the tanks geometry.
Example #1 – Flat Top Tanks Tank A is a vertical tank with a flat to p. The highest
value for FILL-H can be calculated by subtracting the Dead Band from the Sensor Height.
FILL-H = Sensor Height – Dead Band
Tank B is another vertical tank with a riser that
matches the dead band of the sensor. In this case, the FILL-H will be set to the top of the tank.
Note: The ratio of height to diameter of the riser must be no greater than 2:1. Any higher and the sensor will target the inside wall of the riser.
Example #2 – Dome Top Tanks Tank C has a tall enough dome whereas the Dead
Band is above the straight side of the tank. The FILL-H can be set to the top of the straight side (this is advantageous because the top of the straight side is typically a known volume of liquid).
The sensor is moved to a flat part of Tank D. As a result, the Dead Band is now below the straigh t si de of the tank. FILL-H cannot be set to match the straight side of the tank. Calculate FILL-H as follows:
FILL-H = Sensor Height – Dead Band
Example #3 – Horizontal Tanks Tank E is a horizontal tank with a rounded to p. The
highest value for FILL-H can be calculated by subtracting the Dead Band from the Sensor Height.
FILL-H = Sensor Height – Dead Band
Tank F is another horizontal tank with a riser that
matches the dead band of the sensor. In this case, the FILL-H will be set to the top of the tank.
Note: The
ratio of height to diameter of the riser must be no greater than 2:1. Any higher and the sensor will target the inside wall of the riser.
| 16 MN300820 Rev B
Getting Started (continued) Step Two
Example #4 – Cone Bottom Tank Tank G and Tank H are both vertical tanks that are
being supported above ground level with tank stands. The location of the Sensor Height (Tank G - bottom of tank or Tank H – ground level) will influence the FILL-H setting.
Note: FILL-H must always be referenced to the lowest location of the Sensor Height.
Since Tank G is off of the ground, the lowest location may be difficult to identify/measure. This location is required if the Volumetric Mode in WebCal
®
is used.
Tank H uses ground level as the main reference location. This method is easiest to use and ideal for using the Distance mode of WebCal
®
.
Cone and Rounded Bottom Tanks
The location of an EchoTouch
®
installed along the top of a cone or rounded bottom tank may have an effect on the installation of the sensor. Be sure to understand the geometry directly underneath the sensor. Cone or rounded bottom tanks will provide off angled surfaces that can reflect the ultrasonic sound energy away from the sensor as the liquid level is lowered.
#1 Center of Cone #2 Above an Angled Tank
Bottom (straight side only)
#3 Above an Angled Tank
Bottom (full range)
Does impede sensor’s performance. The angled bottom will reflect ultrasonic energy away from the
Tank
Geometry
Does not impede sensor’s performance. Sensor is able to track the entire range of the tank.
Does not impede sensor’s performance as long as the level remains within the straight side of the tank wall.
sensor. HEIGHT is set to the bottom of the tank, the sensor will function as long as the level is within the straight side of
Distance
Reading
HEIGHT is typically set to
the bottom of the tank. Sensor will track the full range of the tank.
HEIGHT is typically set to the bottom of the straight side. Sensor will only tack the straight side of the tank.
the tank.
MN300820 Rev B 17 |
Getting Started (continued) Step Two
SENSOR OUTPUT TO LOCAL DISPLAY/CONTROLLER
Echotouch® outputs a 4-20 mA signal to a local display/controller or to remote devices such as PLCs, SCADA, DCS or other displays/controllers. The 4-20 mA signal is set relative to the Sensor Height and Fill-Height settings. These settings create an operational range that can be translated into a level reading in defined units (i.e. inches, feet, gallons, meters, liters, etc.).
Tank A
In the Tank A example, the Sensor Height sets the 4mA to the bottom of the tank. Fill-Height sets the 20mA to the top of the straight side of the tank.
In the Tank B example, the 4-20mA sensor range is correlated to actual units of level measurement. The operational range now will have engineering values of 0 to 10 feet or 0 to 1000 gallons or 0 to 120 inches.
Tank B
Tank C
In the Tank C example, the 20mA setting was lowered by 1 foot, so the engineering values for the new shorter operational ranges are either 0 to 9 feet, 0 to 900 gallons or 0 to 108 inches.
| 18 MN300820 Rev B
Configuration Using WebCal
®
Step Three
EchoTouch
®
can be configured before installation. The transmitter features non-volatile memory, so any setting configured before installation will not be lost when the switch is powered down. To configure, follow the steps below:
1. Install WebCal a. Go to www.flowline.com/webCal-software/ b. Review how USB
®
software
®
and select language version.
Fob interfaces with EchoTouch® and your computer.
2. Measure the Tank a. Begin by measuring the key tank and fitting dimensions. b. Correct tank dimensions will result in accurate sensor measurement.
3. Sensor Configuration a. Configures Loop Fail-Safe, Output at Empty & Startup Condition for the sensor.
4. Dimensional Entry a. Distance Mode (default)
i. Basic information for operation (Sensor Height & Fill-Height).
5. Tank Level Confirmation a. Confirm the values are accurate for the application.
6. Write to Unit a. Uploads configuration into the sensor. b. Access to a customer wiring diagram specific to the relay configuration.
Note: When configuring EchoTouch
®
, choose either the WebCal® or Push Button method. Push Button is only available with the US06 and US12 series. Either method will accomplish the goal of sensor configuration. Changes to the configuration can be made using the alternative method. When beginning with one method, it is recommended to complete the configuration before using the other method to make any adjustments.
MN300820 Rev B 19 |
Configuration Using WebCal
STEP 1 – INSTALL WEBCAL® SOFTWARE
®
(continued) Step Three
EchoTouch® is configured through WebCal®*, a PC software program. WebCal® is a free download from
®
Flowline’s website. You must download and install WebCal
prior to plugging in the USB® Fob (LI99-
2001). Please go to .flowline.com/webcal-software/, and select your language version.
WEBCAL® SYSTEM REQUIREMENTS:
Windows® XP, Vista, 7, 8, 10 32 or 64-bit system 1 USB
®
2.0 port 10 mB hard drive space 256 mB RAM Internet connection
*
For complete information on the WebCal
®
software, please refer to the WebCal® manual located at
www.flowline.com/webcal-software/.
| 20 MN300820 Rev B
Configuration Using WebCal
USB® FOB INTERFACE:
®
(continued) Step Three
EchoTouch into your computer’s USB
®
communicates with WebCal®* through a USB® interface called a Fob. Before plugging your Fob
®
port, be sure that you have installed WebCal® (see www.flowline.com/webcal-
software/) on your computer.
Connect the red, green, white and black wires from Echotouch the screws on the terminals. Plug your Fob into the USB
®
®
into the correct terminals on the Fob. Tighten
port of your computer.
US01 & US03 Series only US06 & US12 Series only
LI99-2001 USB
®
Fob
Note: Only use the LI99­2001 USB® Fob (white).
The LI99-1001 USB® Fob (black) will not work with the EchoPod® series.
Wiring is identical for all EchoTouch
®
Series – Use only the Red, Black, Green and White wires.
The maximum cable distance
configuring the EchoTouch
Once EchoTouch
®
is configured and prior to installation, isolate the white and green wires from active
between the computer and EchoTouch® is 15’. This only applies for
®
.
power to prevent a short from the configuration circuit.
Note: When using the Fob, do not add VDC or VAC power. The Fob, when connected to the computer, will
provide the required power to the EchoTouch
Note: When connecting the sensor to the WebCal
®
.
®
Software, you must remove the display or the software
will not connect. Once completed, you can re-connect the display to the terminal.
* For complete information on the WebCal
www.flowline.com/webcal-software/.
®
software, please refer to the WebCal® manual located at
MN300820 Rev B 21 |
Configuration Using WebCal
STEP 2 - MEASURE THE TANK
®
(continued) Step Three
Measuring the tank is one of the most important aspects in configuring the sensor. When measuring the tank, take into account the location of the sensor with respect to fittings, risers, dome tops and bottoms, and identify where the measurements are taken from the sensor. The Sensor Height is influenced by the installation location. Sensors mounted along the sloped portion of the tank will result in a lower (HEIGHT) value. On the other hand, sensors installed in risers/nozzles will result in a higher (HEIGHT) value.
The basic measurements (Height and Fill-H) for configuration are described below:
1. HEIGHT (Sensor Height) - Distance from the bottom of the tank to the Bottom of Sensor.
a. The bottom of the tank will be the sensor’s zero
level / empty level.
b. This setting determines the 4mA location.
2. FILL-H (Fill-Height) - Distance from the bottom of the tank to the level of liquid where the tank is full is FILL-H.
a. The FILL-H is typically set to the liquid full level. b. This setting determines the 20 mA location. c. 20mA cannot be set within the sensor’s dead
band [8” (20cm) from bottom of sensor]. Largest FILL-H value will be as follows:
Largest FILL-H = HEIGHT – 8” (20cm)
| 22 MN300820 Rev B
Configuration Using WebCal
®
(continued) Step Three
With EchoTouch
®
connected to your computer, open the WebCal®* software by clicking on the WebCal® icon. Follow steps 3-6 to configure the transmitter. Click “Help” in the lower right hand corner and open the help menu of WebCal
®
for instructions on WebCal®. If you need additional assistance using WebCal®, please
contact a Flowline applications engineer at (562) 598-3015. Note: For complete information on the WebCal
software, please refer to the WebCal
STEP 3 - SENSOR CONFIGURATION:
®
manual located at .flowline.com/webcal-software/.
Configures the Loop Fail-Safe, Output at Empty and Startup Condition for the sensor.
®
Note: When configuring EchoTouch®, choose either the WebCal® or Push Button method. Either method will accomplish the goal of sensor configuration. Changes to the configuration can be made using the alternative method. When beginning with one method, it is recommended to complete the configuration before using the other method to make any adjustments.
MN300820 Rev B 23 |
n
P
o
S
e
S
e
S
S
T
f
m
s
o
o
®
u
H
s
d
e
o
c
o
e
h
O
w
U
s
m
e
u
O
2
c
s
n
e
U
n
h
e
d
c
n
h
f
t
m
e
a
T
t
s
y
B
e
Co
STE
figurati
4 – DIMEN
n Using
IONAL EN
WebCal
RY:
(continu
d)
Step Thre
Distanc
change i category values ( will be s
Mode (de
n liquid leve . For Volu
ensor Heig
t in the unit
ensor Heig
osition (bott
p
ill-Height:
F
osition (bott
p
ault): Outp
l will reflect
etric output
ht and Fill-
shown un
ht: Sets th
m of tank)
Sets the lo
m of tank)
t of sensor linearly to t s, refer to V
eight) belo
er Height
location fo to the Mea
ation for 20 to the Full l
is based on e current o
LUMETRI
set the 4-
nits.
r 4mA. It is
urement lo
A. It is ba
vel positio
the distanc
tput. Note:
C CONFIG
0 mA curre
based on t
ation for th
ed on the
(see below
(height of
Most appli
RATION o t span for t
e distance sensor (bo
istance fro
).
liquid) in th
ations will f
page 40. e sensor.
rom the Em
tom of sen
the Empt
tank. Any
ll into this
he two
Both values
pty level
or).
level
| 24
MN300820 Rev
Configuration Using WebCal
STEP 5 – TANK LEVEL CONFIRMATION:
Verify the Height Units, Sensor Height & Fill-Height. All values were calculated in the previous Dimensional Entry window. To adjust these settings, click on Volumetric Modes.
®
(continued) Step Three

STEP 6 – WRITE TO UNIT:
This WebCal®* operation uploads configuration into the sensor. Other features in the section include providing a custom wiring diagram specific to the signal output and saving the configuration file to your hard drive.
VolumetricSensorOutput
TheVolumetric Mode buttonwillbehighlighted inBluewhen a volume outputisselected.
* For complete information on the WebCal
®
software, please refer to the WebCal® manual located at
flowline.com/webcal-software/.
MN300820 Rev B 25 |
Configuration Using Display Step Four
EchoTouch
®
can be configured before installation. The transmitter features non-volatile memory, so any setting configured before installation will not be lost when the switch is powered down. To configure, follow the steps below: (Pages 27 to 31)
1. Measure the Tank a. Begin by measuring the key tank and fitting dimensions. b. Correct tank dimensions will result in accurate sensor measurement.
2. Set the Units of Measurement (UNITS) a. Units can be configured in basic engineering units of length including Feet, Inches, Centimeters
or Meters.
3. Set the Sensor Height (HEIGHT) a. This is the empty setting (4mA) for the tank.
4. Set the Fill-Height (FILL-H) a. This is the full setting (20mA) for the tank.
5. Set the Fail-Safety (SAFE) a. This is the maximum measurement range for the sensor. The sensor will ignore all echo returns
beyond this setting.
NOTE: When configuring EchoTouch
accomplish the goal of sensor configuration. Changes to the configuration can be made using the alternative method. When beginning with one method, it is recommended to complete the configuration before using the other method to make any adjustments.
®
, choose either the WebCal® or Push Button method. Either method will
NOTE: Display configuration is only available with the US06 & US12 series.
EchoTouch® peut être configuré avant l'installation. L'émetteur dispose d'une mémoire non volatile, de sorte que tout réglage configuré avant l'installation ne sera pas perdu lorsque le commutateur est mis hors tension. Pour configurer, procédez comme suit: (Pages 32 à 36)
1. Mesurer le réservoir a. Commencez par mesurer le réservoir de clés et les dimensions de montage. b. Les dimensions correctes du réservoir entraîneront une mesure précise du capteur.
2. Définissez les unités de mesure (UNITS) a. Les unités peuvent être configurées dans des unités d'ingénierie de base comprenant des
pieds, des pouces, des centimètres ou des compteurs.
3. Réglez la hauteur du capteur (HAUTEUR) a. C'est le réglage vide (4mA) pour le réservoir.
4. Définissez la hauteur de remplissage (FILL-H) a. C'est le réglage complet (20mA) pour le réservoir.
5. Configurez la sécurité d'échec (SAFE) a. C'est la plage de mesure maximale pour le capteur. Le capteur ignore tous les retours d'écho
au-delà de ce paramètre.
REMARQUE: lors de la configuration d'EchoTouch®, choisissez la méthode WebCal® ou Push Button. L'une ou
l'autre méthode atteindra l'objectif de la configuration du capteur. Les modifications apportées à la configuration peuvent être effectuées en utilisant la méthode alternative. En commençant par une méthode, il est recommandé de compléter la configuration avant d'utiliser l'autre méthode pour procéder à des ajustements.
REMARQUE: la configuration de l'affichage n'est disponible qu'avec les sé ries US06 & US12.
| 26 MN300820 Rev B
Configuration Using Display (continued) Step Four
#1 - MEASURE THE TANK
Measuring the tank is one of the most important aspects in configuring the sensor. When measuring the tank, take into account the location of the sensor with respect to fittings, risers, dome tops and bottoms, and identify where the measurements are taken from the sensor. The Sensor Height is influenced by the installation location. Sensors mounted along the sloped portion of the tank will result in a lower (HEIGHT) value. On the other hand, sensors installed in risers/nozzles will result in a higher (HEIGHT) value.
The basic measurements (Height and Fill-H) for configuration are described below:
1. HEIGHT (Sensor Height) - Distance from the bottom of
the tank to the Bottom of Sensor.
a. The bottom of the tank will be the sensor’s zero
level / empty level.
a. This setting determines the 4mA location.
2. FILL-H (Fill-Height) - Distance from the bottom of the tank
to the level of liquid where the tank is full is FILL-H.
a. The FILL-H is typically set to the liquid full level. b. This setting determines the 20 mA location. c. 20mA cannot be set within the sensor’s dead band
[8” (20cm) from bottom of sensor]. Largest FILL-H value will be as follows:
Largest FILL-H = HEIGHT – 8” (20cm)
MN300820 Rev B 27 |
Configuration Using Display (continued) Step Four
TOP-LEVEL MENU
The TOP-LEVEL MENU will continue to scroll
through the items listed to the right until the SELECT button is pressed.
To return to the Normal / Operational mode of
Echotouch®, press SELECT while RUN appears in the display.
How to enter the TOP-LEVEL MENU
1. Press and hold OK (approximately 5 seconds) until
MENU is displayed.
2. The TOP-LEVEL MENU items will rotate through
display.
3. Use OK to choose a MENU item.
How to exit the TOP-LEVEL MENU
In TOP-LEVEL MENU, press SELECT when RUN appears.
STEP 2 - SETTING THE UNITS OF MEASUREMENT (UNITS)
1. In the TOP-LEVEL MENU mode, select UNITS.
2. Press OK to choose between INCHES, CM
(centimeter), FEET or METERS. a. To cancel UNITS, Select EXIT.
3. Select EXIT to return to the TOP-LEVEL MENU.
Note: EchoTouch® will adjust the display when measurement units are changed under UNITS. Example: If you are in inches and the display reads 60.0, the display will change to 5.00 when feet is selected as the unit of measurement.
| 28 MN300820 Rev B
Configuration (with Display…Continued) Step Four
STEP 3 - SETTING THE HEIGHT (SENSOR HEIGHT)
1. In TOP-LEVEL MENU mode, select TANK.
2. Select HEIGHT.
3. Using the UP and DOWN buttons, set the HEIGHT of the tank (transducer face to the bottom of the tank).
4. To enter the value, press and hold OK for 2 seconds and release. SAVED will display.
5. Select EXIT to return to TOP-LEVEL MENU.
Note: The HEIGHT setting also determines the 4 mA current location. To reverse the current output with 20mA at the bottom of the tank, use the REV MA function.
Note: To speed up the scrolling of the values on the display, hold down the OK button while holding down the UP or DOWN buttons.

MN300820 Rev B 29 |
Configuration (with Display…Continued) Step Four
STEP 4 - SETTING THE FILL-H (FILL-HEIGHT)
1. In TOP-LEVEL MENU mode, select TANK.
2. Select FILL-H.
3. Using the UP and DOWN buttons, set the Fill Height (FILL-H) of the tank (max. level of liquid to the bottom of the tank).
4. To enter the value, press and hold OK for 2 seconds and release. SAVED will display.
5. Select EXIT to return to TOP-LEVEL MENU.
Note: The Fill-Height (FILL-H) setting determines the 20 mA current span. To reverse the current output with 4mA being at the top of the tank, use the REV MA function.
Note: To speed up the scrolling of the values on the display, hold down the OK button while holding down the UP or DOWN buttons.

| 30 MN300820 Rev B
Configuration (with Display…Continued) Step Four
STEP 5 - HOW TO SELECT FAIL-SAFE CURRENT OUTPUT (LOST)
In the event the sensor does not receive an echo, the Fail-Safe Current Output or LOST setting can be set to output a current of 4mA, 20mA, 21mA, 22mA or Hold (last known value). During fail-safe, the display will read LOST.
1. In TOP-LEVEL MENU mode, select OUTPUT.
2. Select SAFE.
3. Select 4mA, 20mA, 21mA, 22mA or HOLD.
a. Press and hold OK on the
selected setting for 3 seconds then release.
b. To cancel SAFE, Select EXIT.
4. Select EXIT to return to TOP-LEVEL MENU.
MN300820 Rev B 31 |
Configuration Using Display (continued) Step Four
ÉTAPE #1 - MESURER LE TANK
La mesure du réservoir est l'un des aspects les plus importants de la configuration du capteur. Lors de la mesure du réservoir, tenir compte de l'emplacement du capteur en ce qui concerne les raccords, les élévateurs, les dômes et les fonds, et identifier les mesures prises à partir du capteur. La hauteur du capteur est influencée par l'emplacement d'installation. Les capteurs montés le long de la partie inclinée du réservoir entraîneront une valeur inférieure (HAUTEUR). D'autre part, les capteurs installés dans les risers / buses entraîneront une valeur plus élevée (HAUTEUR).
Les mesures de base (Height et Fill-H) pour la configuration sont décrites ci-dessous :
1. HEIGHT (Hauteur du capteur) - Distance du fond du
réservoir au bas du capteur.
a. Le fond du réservoir correspondra au niveau zéro /
vide du capteur.
b. Ce paramètre détermine l'emplacement 4mA.
2. FILL-H (Fill-Height) - La distance du fond du réservoir au
niveau du liquide où le réservoir est plein est FILL-H.
a. Le FILL-H est généralement réglé sur le niveau
complet du liquide. b. Ce paramètre détermine l'emplacement de 20 mA. c. 20mA ne peut pas être réglé dans la bande morte
du capteur [8 "(20cm) du bas du capteur]. La plus
grande valeur FILL-H sera la suivante:
Le plus grand FILL-H = HEIGHT - 8 "(20cm)
| 32 MN300820 Rev B
n
N
Llele PdR
m
123
m
sRUN
P
123Ru
sls
o
O
s
K
r
®
a
z
s
m
g
O
e
H
m
z
t
P
n
e
a
s
c
s
c
®
z
N
e
e
L
o
N
U
L
C
®
m
e
i
(
à
s
O
E
e
U
N
U
p
M
N
s
E
e
a
f
u
h
p
r
d
r
e
l
e
R
é
N
n
C
X
O
E
e
p
r
Co
ME
U DE HAUT
e MENU T
figurati
s élément bouton O
n Using
NIVEAU
P-LEVEL
répertorié
soit enfon
Display
ontinuera
à droite ju
é.
continue
faire défile
qu'à ce qu
)
Step Fou
'Echotouch
Co
ment accé
. Appuye
. Les élé
. Utilisez
Co
ment quitt
Dan
apparaît.
ÉTA
. Dans le
our reveni
UN appar
seconde
l'afficha
le MENU
E #2 - RÉG
UNITS.
au mode
, appuye
ît à l'écran.
der au ME
et mainten
) jusqu'à c
ents TOP-
e.
K pour ch
r le MENU
AUT DE
LAGE DES
ode TOP-
Normal /
sur OK p
U DE NIV
z la touch
que MEN EVEL ME
isir un élém
DE NIVEA
IVEAU, ap
NITÉS DE
EVEL ME
pérationne
endant qu
AU SUPÉ
SELECT (
soit affich U tournent
ent MENU.
SUPÉRIE
uyez sur O
ESURE (U
U, sélectio
IEUR
environ 5
.
à travers
UR
K lorsque
ITS)
nez
. Appuye
. Sélectio
emarque:
nités de m
i vous êtes
'affichage p
électionné
CM (cen
a.
LEVEL.
sur SELE
imètre), FE
our annuler
nez EXIT p
EchoTouch
sure sont
en pouces
sse à 5,00
comme un
T pour choi ET ou MET
UNITS, sél
our revenir
réglera l'a
odifiées so
t que l'affic
lorsque les
té de mesu
ir entre IN
RS. ctionnez E u MENU T
fichage lors
s UNITS.
age indiqu
ieds sont
e.
HES,
IT.
P-
que les
xemple:
60.0,
MN30082
0 Rev B
33
|
Configuration (with Display…Continued) Step Four
ÉTAPE 3 - RÉGLAGE DE LA HAUTEUR (SENSOR HEIGHT)
1. En mode TOP-LEVEL MENU, sélectionnez TANK.
2. Sélectionnez HEIGHT.
3. À l'aide des boutons HAUT et BAS, réglez la HEIGHT du réservoir (face du transducteur au bas du réservoir).
4. Pour entrer la valeur, appuyez et maintenez la touche OK pendant 2 secondes et relâchez. SAVED s'affichera.
5. Sélectionnez EXIT pour revenir au MENU TOP-LEVEL.
Remarque: Le paramètre HEIGHT détermine également l'emplacement actuel de 4 mA. Pour inverser la sortie de courant avec 20 mA au bas du réservoir, utilisez la fonction REV MA.
Remarque: Pour accélérer le défilement des valeurs sur l'affichage, maintenez la touche OK enfoncée tout en maintenant enfoncés les boutons UP ou DOWN.

| 34 MN300820 Rev B
Configuration (with Display…Continued) Step Four
ÉTAPE 4 - RÉGLAGE DE LA FILL-H (FILL-HEIGHT)
1. En mode TOP-LEVEL MENU, sélectionnez TANK.
2. Sélectionnez FILL-H.
3. À l'aide des boutons HAUT et BAS, réglez la hauteur de remplissage (FILL-H) du réservoir (niveau maximum de liquide au fond du réservoir).
4. Pour entrer la valeur, appuyez et maintenez la touche OK pendant 2 secondes et relâchez. SAVED s'affichera.
5. Sélectionnez EXIT pour revenir au MENU TOP-LEVEL.
Remarque: Le paramètre de hauteur de remplissage (FILL-H) détermine la portée de courant de 20 mA. Pour inverser la sortie de courant avec 4mA en haut du réservoir, utilisez la fonction REV MA.
Remarque: Pour accélérer le défilement des valeurs sur l'affichage, maintenez la touche OK enfoncée tout en maintenant enfoncés les boutons UP ou DOWN.

MN300820 Rev B 35 |
Configuration (with Display…Continued) Step Four
ÉTAPE 5 - COMMENT SÉLECTIONNER UNE SORTIE ACTUELLE FAIBLE (LOST)
Dans le cas où le capteur ne reçoit pas d'écho, le réglage de sortie de courant sans échec ou LOST peut être configuré pour afficher un courant de 4mA, 20mA, 21mA, 22mA ou Hold (dernière valeur connue). En cas de panne de sécurité, l'affichage se lire LOST.
1. En mode TOP-LEVEL MENU, sélectionnez OUTPUT.
2. Sélectionnez SAFE.
3. Sélectionnez 4mA, 20mA, 21mA, 22mA ou HOLD.
a. Appuyez et maintenez OK sur
le réglage sélectionné pendant 3 secondes, puis relâchez.
b. Pour annuler SAFE,
sélectionnez EXIT.
4. Sélectionnez EXIT pour revenir au MENU TOP-LEVEL.
 
| 36 MN300820 Rev B
Installation Step Five
EchoTouch
®
should always be mounted perpendicular to the liquid surface (use the provided Viton® mounting gasket for G threaded versions only). Make sure that the fitting and transmitter threads are not damaged or worn. Always hand-tighten the transmitter within the fitting. Perform an installed leak test under normal process conditions prior to system start up.
Warning - Potential Electrostatic Charging Hazard (Handling Static-Sensitive Circuits/Devices): When handling the transmitter, the technician should follow these guidelines to reduce any possible electrostatic charge build-up on the technician’s body and the electronic part.
1. Always touc h a known good ground source before handling the part. This should be repeated while handling the part and more frequently after sitting down from a standing position, sliding across the seat or walking a distance.
2. Avoid touching electrical terminals of the part unless making connections.
3. DO NOT open the unit cover until it is time to calibrate.
MOUNTING GUIDE
1. Do not mount at an angle
2. Liquid should never enter the dead band
3. Side Wall: a. Mount at least 3” from the side wall
Do not install at an
angle relative to the
liquid.
4. Do not mount where obstacles will intrude on sensor’s
beam width
a. mount at least 3” from the side wall
5. Do not mount in a vacuum
6. Do not mount in the center of a dome top tank.
7. In cone bottom tank, position the sensor over the
Do not install within
3” of tank sidewall.
deepest part of the tank.
Installation in existing fittings: If the existing fitting is larger than the threads of the EchoTouch bushing such as the LM52-2400 (3” thread x 2” thread) or LM52-3400 (4” thread x 2” thread).
®
, select a reducer
Do not install with
objects in the
beam.
Do not install in
applications with
vacuum.
LM52-2400
Do not install in the
center of a dome
top tank.
MN300820 Rev B 37 |
Installation Step Five
EchoTouch
®
Viton
®
doit toujours être monté perpendiculairement à la surface du liquide (utilisez le joint de fixation
fourni uniquement pour les versions filetées G). Assurez-vous que les fils de montage et d'émetteur ne sont pas endommagés ou usés. Toujours serrer à la main l'émetteur dans le raccord. Effectuer un test d'étanchéité installé dans des conditions de processus normales avant le démarrage du système.
GUIDE DE MONTAGE
1. Ne pas monter à angle
2. Le liquide ne devrait jamais entrer dans la bande morte
3. Mur latéral:
a. Montez au moins 3 "de la paroi latérale
4. Ne pas monter où les obstacles s'immisceront sur la largeur du faisceau du capteur.
a. monter au moins 3 "de la paroi latérale
5. Ne pas monter sous vide
Ne pas installer à
un angle par
rapport au liquide.
Ne pas installer à
moins de 3 po de la
paroi latérale du
réservoir.
6. Ne pas monter au centre d'un réservoir en forme de dôme.
7. Dans le réservoir inférieur cône, positionnez le capteur sur la partie la plus profonde du réservoir.
Ne pas installer avec des objets
dans la poutre.
Installation dans les raccords existants: si le raccord existant est plus grand que les fils de l'EchoTouch®, sélectionnez une douille réductrice telle que le LM52-2400 (filetage 3 "x 2") ou LM52-3400 (filetage 4 "filetage x 2")
LM52-2400
Ne pas installer
dans des
applications avec
vide.
Ne pas installer au
centre d'un
réservoir en forme
de dôme.
| 38 MN300820 Rev B
t
c
a
1
2
o
T
n
d
d
8
Fth
A
A
d
T
8
n
C
a
D
e
e
e
f
h
e
u
a
d
n
:
e
w
o
c
C
n
m
5
d
e
c
d
a
a
d
u
n
R
9
r
Ap
a
)
r
s
)
i
t
g
d
s
2
a
w
p
m
e
s
N
a
o
f
p
k
o
a
e
s
9
e
s
u
o
d
8
c
m
e
p
s
n
e
s
a
e
T
e
n
h
e
d
A
x
s
e
e
1
p
l
l
i
a
d
t
e
y
g
e
0
s
u
e
e
e
k
y
w
f
n
t
h
Ins
allation
FITT
ING SELEC
Che
k the part
inst
lled in tank
the
G” threade
(continu
ION
umber to d
adapters,
EchoTouc
d)
termine th
langes, bra
®
.
required fi
kets or sta
ting mount
ndpipes.
ize and thr
ote: Alway
ad type. E
include th
choTouch®
gasket w
Step Fiv
is commonl
en installin
. Tank A
LM52-3
a.
b. c.
2” Tank A Socket x
(LM52-2
apter: Sel
90 for 3” se
or best res
read x thre
void tank a lways mou
i. Note
insid
apter
hread
90)
ct a tank a
nsors.
lts, select
d, reducer
apter (thre
t the tank a
Never mo
the tank.
w/ 3”x2”
(LM52-38
apter fittin
3” tank a
bushing.
d x thread)
apter so th
nt the tank
k Adapter
Ta
educer Bu
0 w/LM52-
, such as th
apter and
styles and/
e majority o
adapter u
hing
800)
e LM52-28
dd a reduc
r pipe stop fitting is o
side down
3” Tank A Socket x T
(LM52-3
0 tank ada
r bushing
forward of
tside the ta
r where th
apter
hread
90)
ter for 2” s
uch as the
the installe
k.
bulk of th
Tank
Thread
Do not u
nsors or th
LM52-240
transducer
e material i
dapter
Thread
e thread x
thr
ad
,
.
N
. Riser: I
a.
a. 2
te: Do not
stallations
ore Out C
c
n also be
iameter to
” (5 cm) dia
n
o taller than
Dia
4” (1 6” (1
xceed the
ith tall, na ncrete: onsidered
ore Height meter rise 12” (30 cm
R
iser Specif
In
er eter
2” (
cm)
0cm) 5cm)
imensions
row risers c
plications
s a riser ty for the dia should be
.
cations
Maximum
Height
4” (10cm) 8” (20cm)
12” (30cm)
listed abov
n impede t
here a tan
e applicati
eter of the
no taller th
.
he acoustic with a con n. In these core.
n 4” (10 c
signal.
rete ceiling
application
). Larger d
Note: If sensor to allow avoid th risers. exceeds the Ris Use a la takes i
Specific
that has be
follow a 2:
iameter rise
attempting
bove the to
for a higher
use of tal
he examp
the dimens
r Specific
rger tank a
to accoun
ations.
n cored o
ratio (Inn
rs should b
to raise th
of the tan
fill capacit and narro e to the le
ons listed i tions char
apter whic the Rise
t
r
,
t
.
r
MN30082
0 Rev B
39
|
t
3
4
t
U
n
-
r
h
t
p
e
8
t
u
a
,
r
l
l
(
T
d
e
p
0
s
o
h
r
h
n
w
s
d
s
s
s
p
p
M
0
d
s
s
h
B
e
Ins
allation
(continu
d)
Step Fiv
. Flange:
. Side Mo
such as
a.
2” Fla (LM52
mount b
If installing he LM52-2
se a flange
ge w/
2850)
unt Bracke
acket.
on a flange
50.
with a 3” th
3” F
: For insta
select a fla
ead and ad
ange w/ R
(LM52
lations in o
nge with a t
a 3” to 2”
ducer Bus
-3800)
en tanks a
read that i
educer bus
ing
d sumps, u
above the
hing to com
3”
(L
e the LM5
lane of the
lete the in
Flange
52-3850)
-1001 serie
flange,
tallation.
s side
LM5
-1001 Sho
Note: T method stand pi
e Side Mo
o secure st
e.
nt Bracket
nd pipes.
LM50 serie
here are to
) is not de
few threa
n
igned for u
s to properl
e with stan y hold the s
pipes or a
ensor and t
a
e
| 40
MN300820 Rev
Installation (continued) Step Five
5. Stand Pipe: A standpipe maybe used to dampen turbulence or when foam is present in the
application.
a) Pipe can be constructed from any material. b) Select a minimum 3” ID pipe or greater for the stand pipe.
i) Select a minimum 4” IS for the US12 series.
c) Use a coupling and reducer bushing to attach the
EchoTouch
®
to the pipe.
i) Use a reducer bushing such as LM52-2400 (3” Thread
x 2” Thread) fitting or the LM52-2410 (3” Slip x 2” Thread) fitting.
d) The pipe length should run the measurement span. The
bottom of the pipe should remain submerged at all times to prevent foam from entering the pipe.
e) Cut a 45° notch at the bottom of the pipe. Drill a
1/4”pressure equalization hole in the dead band.
f) The pumps should not drive liquid past the open end of the
stand pipe which causes the liquid in the pipe to oscillate.
Note: Never allow the bottom of the stand pipe to become exposed to air. This will break the liquid seal which will prevent echoes from returning back to the sensor.
EchoTouch®
UG06
3” x 2”
Reducer Bushing
(TxT)
3” Coupling
(S x T)
Vent Hole (1/4”)
3” PVC Pipe
EchoTouch
®
attached to a LM52-
2400 (3” x 2” reducer bushing) to
a Slip x Thread 3” Coupling.
Avoid the use of a tee within the
stand pipe. A tee can create false
signals impeding the sensor’s
performance.
MN300820 Rev B 41 |
Wiring Step Six
ANALOG OUTPUT (4-20 MA)
The analog output of the EchoTouch® is a loop powered 4-20 mA control circuit. The typical way to use this feature is to connect a positive supply to the (+) input and to sense the current flow out of the (-) output.
The cabling should be a shielded twisted pair to minimize EMI interference. Typically 20 to 24 gauge wire is used in this application.
| 42 MN300820 Rev B
Wiring Step Six
SORTIE ANALOGIQUE (4-20 MA)
La sortie analogique de l'EchoTouch® est un circuit de contrôle 4-20 mA alimenté en boucle. Le moyen typique d'utiliser cette fonction est de connecter une alimentation positive à l'entrée (+) et de détecter le flux de courant sortant de la sortie (-).
Le câblage doit être une paire torsadée blindée pour minimiser les interférences EMI. Typiquement, le fil de calibre 20 à 24 est utilisé dans cette application.
MN300820 Rev B 43 |
Wiring (continued) Step Seven
WIRE CONNECTIONS
US01 & US03 Series US06 & US12 Series (+) Power & (-) Return: (+) Power and (-) Return leads are for connection to a 24 VDC power supply or to a 4-
20 mA loop power source. The (+) and (-) wires can be extended up to 1,000 feet using a 22-gauge or larger wire.
Tx (Out)/W & Rx (In)/G: Tx (Out)/White and Rx (In)/Green leads are reserved for use with WebCal
®
. Note: The Green wire is not the Ground for the transmitter. In addition, these wires should not be connected to WebCal distance between the computer and Echotouch
®
while power is supplied from any source other than the LI99 series Fob. The maximum cable
®
is 15’ (4.6m). Note: Never allow the white or green wires to
touch any power supply.
GENERAL NOTES FOR ELECTRICAL CONNECTIONS, USAGE AND SAFETY
Where personal safety or significant property damage can occur due to a spill, the installation
must have a redundant backup safety system.
Wiring should always be completed by a licensed electrician. Supply voltage should never exceed 28 VDC. Use 16 to 24 AWG for connecting to terminals. Protect the sensor from excessive electrical spikes by isolating the power, whenever possible. The sensor materials must be chemically compatible with the liquids to be measured. Design a fail-safe system for possible sensor and/or power failure.
During power failure, relays will remain in their current state and will not change until power is
restored and the signal is reacquired.
ACCESSING THE TERMINALS (US06 & US12 SERIES ONLY)
To access the terminals, gently remove the display from the housing.
When wiring is completed, restore the display to the housing and replace the cover.
The US06 & US12 series can operate without the display. When wired and powered correctly, the sensor will continue to provide a proportional 4-20 mA signal.
| 44 MN300820 Rev B
Wiring (continued) Step Seven
CONNEXIONS DE FIL
US01 & US03 Series US06 & US12 Series
(+) Alimentation et (-) Retour: (+) Alimentation et (-) Les fils de retour sont destinés à être connectés à une
alimentation 24 V cc ou à une source d'alimentation en boucle 4-20 mA. Les fils (+) et (-) peuvent être prolongés jusqu'à 1000 pieds à l'aide d'un fil de calibre 22 ou supérieur.
Les connexions Tx (Sortie) / W & Rx (In) / G: Tx (Out) / White et Rx (In) / Green sont réservées pour une utilisation avec WebCal pas être connectés à WebCal distance maximale du câble entre l'ordinateur et EchoTouch
®
. Remarque: le fil vert n'est pas la terre de l'émetteur. En outre, ces fils ne doivent
®
lorsque l'alimentation est fournie par une source autre que la série LI99 Fob. La
®
est de 15' (4.6m). Remarque: N'autorisez jamais
les fils blanc ou vert à toucher n'importe quelle alimentation électrique.
NOTES GÉNÉRALES POUR CONNEXIONS ÉLECTRIQUES, UTILISATION ET SÉCURITÉ
Lorsque la sécurité personnelle ou des dégats matériels importants peuvent survenir en raison
d'un déversement, l'installation doit comporter un système de sécurité de secours redondant.
Le câblage doit toujours être complété par un électricien autorisé. La tension d'alimentation ne doit jamais dépasser 28 VDC. Protégez le capteur contre les pointes électriques excessives en isolant l'alimentation, chaque fois que
cela est possible.
Les matériaux du capteur doivent être chimiquement compatibles avec les liquides à mesurer. Conception d'un système de sécurité pour détecter un éventuel capteur et / ou une panne de courant.
Pendant la panne de courant, les relais resteront dans leur état actuel et ne changeront
pas jusqu'à ce que l'alimentation soit rétablie et que le signal soit réapprovisionné.
ACCÈS AUX TERMINAUX (US06 & US12 SERIES ONLY)
Pour accéder aux bornes, retirez doucement l'affichage du boîtier.
Lorsque le câblage est terminé, rétablissez l'affichage sur le boîtier et remettez le couvercle.
Les séries US06 & US12 peuvent fonctionner sans l'affichage. Lorsqu'il est câblé et alimenté correctement, le capteur continuera à fournir un signal proportionnel à 4-20 mA.
MN300820 Rev B 45 |
Wiring (continued) Step Six
COMMON WIRING TO DISPLAY, CONTROLLERS & PLC’S
Below is a quick review of wiring the EchoTouch® to common display, controllers and PLC’s.
DataView™ LI55 Series Level Controller
DataLoop™ LI23 Series Level Indicator without the backlight
DataLoop™ LI23 Series Level Indicator with the backlight
Note: Always refer to control drawing 600820 for the UG06 & US12 series and control drawing 610125 for
UG01 and UG03 series.
| 46 MN300820 Rev B
Wiring (continued) Step Six
COMMON WIRING TO DISPLAY, CONTROLLERS & PLC’S
Below is a quick review of wiring the EchoTouch® to common display, controllers and PLC’s.
Commander™ LI90 Series Multi-Tank Level Controller
DataPoint™ LC52 Series Level Controller JWA mode (Factory Setting)
GenericLoopPoweredDisplay
Note: Always refer to control drawing 600820 for the UG06 & US12 series and control drawing 610125 for UG01 and UG03 series.
MN300820 Rev B 47 |
Wiring (continued) Step Six
COMMON WIRING TO DISPLAY, CONTROLLERS & PLC’S
Below is a quick review of wiring the EchoTouch® to common display, controllers and PLC’s.
GenericPLC
Note: Always refer to control drawing 600820 for the UG06 & US12 series and control drawing 610125 for UG01 and UG03 series.

| 48 MN300820 Rev B
WebCal® Appendix Section Seven
This section of WebCal
®
is where you select the sensor configuration settings. Start from the top and work to the bottom, choosing the selections that are applicable to your application. All configuration settings must be selected before you can continue to the next step.
SENSOR CONFIGURATION
LOOP FAIL-SAFE
This feature allows you to select the fail-safe current output if the sensor looses echo confidence (LOST). When the sensor regains echo confidence, the output current will revert back to the current level condition.
Hold Last Value - The output will remain in the same state as
the last validated echo detected. Example: If the output was
6.7mA just prior to the lost signal, the sensor will continue to output 6.7mA until echo confidence is regained.
Empty - The output will revert to the current value for an
empty condition. When 4 mA at Bottom is selected, the sensor will output 4 mA during a fail-safe condition. If 20 mA at Bottom is selected, the sensor will output 20 mA during a fail-safe condition.
Full - The output will revert to the current value for a full
condition. When 4 mA at Bottom is selected, the sensor will output 20 mA during a fail-safe condition. If 20 mA at Bottom is selected, the sensor will output 4 mA during a fail­safe condition.
Overfill (21mA) - The sensor will output 21mA during a fail-
safe condition.
Overfill (22mA) - The sensor will output 22mA during a fail-
safe condition.
OUTPUT AT EMPTY
This feature allows you to select the orientation of the 4 to 20mA output (4 to 20 mA or 20 to 4 mA). Choose which output setting best fits the application. Factory default is 4mA at bottom and 20mA at top, as this configuration scenario is an industry standard. When connecting your sensor to a display, you must account for
your output orientation setting.
4mA at Bottom - The output current will be 4mA when the sensor measures an empty tank and 20mA
when the sensor measures a full tank.
20mA at Bottom - The output current will be 20mA when the sensor measures an empty tank and 4mA
when the sensor measures a full tank.Note: Right click on any item to open the help menu.
Note: Right click on any item to open the help menu. Note: To reset the configuration table, press the Clear Screen button
MN300820 Rev B 49 |
WebCal® Appendix (continued) Section Seven
SENSOR CONFIGURATION (CONTINUED)
STARTUP CONDITION
This feature allows you to select the startup current when power is first applied to the sensor. The sensor will consume the selected power while it is acquiring the liquid level. When the correct level has been identified, the output will adjust to the level output. Use this feature to avoid false alarms with the controller when power is first applied to the sensor.
Empty - The current output will revert to the selected current
value for an empty condition.
o 4 mA at Bottom - The sensor will output 4 mA while
the sensor powers up.
o 20 mA at Bottom - The sensor will output 20 mA while
the sensor powers up.
Mid Tank (12 mA) – The sensor will output 12 mA while the
sensor powers up.
Full - The output will revert to the selected current value for a
full condition.
o 4 mA at Bottom - The sensor w ill outp ut 20 mA while
powering up.
o 20 mA at Bottom - The sensor will ou tput 4 mA w hile the
sensor powers up.
Overfill (22mA) - The sensor will output 22mA while the
sensor powers up.
Note: Right click on any item to open the help menu. Note: To reset the configuration table, press the Clear Screen button
| 50 MN300820 Rev B
Appendix (WebCal) Section Seven
VOLUMETRIC CONFIGURATION
The sensor may be configured in volumetric units (Gallons or Liters) or Distance (Height of Liquid) units (inches, cm, feet or meters). WebCal default to Distance (Height of Liquid) with units of Inches. To change units or change from Distance to Volume, press the Volumetric Mode button as located near the center of the window.
Distance (Height of Liquid): When this is selected, the sensor will always output a linear output proportional to the height of liquid (regardless of the tank shape). In the two drawings below, the Sensor High is set to 64” and Fill-Height is set to 60”. If the tank is empty, the sensor will output 4mA. If the level is at 30”, then the sensor will output 12mA. Either tank can be configured with a display from 0” to 60” and the sensor will always output the height of liquid.
®
will
Volume: When volume is selected, the 4-20 mA output from the sensor will be proportional to the volume of the tank, not the height of the tank. This means that the current output will track the volume of the tank (in gallons or liters). Depending on the shape of the tank, the values may be a linear or non-linear scale. In the same examples below, the sensors are also configured with SH = 64” and FH = 60”. The current output in a linear tank will act as expected with volume increase matching equal changes to level. However, the current output in the Non-Linear tank will reflect the actual changes in volume. So even though the liquid height changes incrementally the same amount, the volume change will adjust throughout the entire span of the tank. The advantage of this is that sensor does all the math so you can read tank volume using a simple two-point display/controller such as the DataLoop
Linear Tank Example
LI25 series or DataView™ LI55 series.
Non-Linear Tank Example
Note: In the above illustration, 10” of liquid will
always be equal to 100 gallons of liquid (1” = 10 gallons).
Note: In the above illustration, 1” of liquid does not equal 10 gallons. The 10” at the bottom represents a rise of 62.8 gallons. As a change between 10” and 20” represents an increase of 109.6 gallons (i.e.
172.4 gallons – 62.8 gallons).
MN300820 Rev B 51 |
WebCal® Appendix (continued) Section Seven
Shape Selection Window: This window will shows the different tank shape options available in WebCal
Vertical Cylinder Vertical Cylinder with Cone Bottom Horizontal Cylinder with End caps Horizontal Cylinder with Spherical Ends Spherical Rectangular Strapping Table – Use this feature for manual entry
of measured tank distances and volumes.
Select any of the above tank shapes and press OK to confirm.
A. Dimensional Entry – Vertical Cylinder Example: Choose the Sensor Output Units as Distance or
Volume. After choosing the Sensor Output Units, select the units of measurement in the pull down to the left.
®
.
Units of Measurement
Distance Volume
Inches
Cm
Feet
Meters
Distance – Sensor Output Units (Vertical Cylinder Example):
Enter the dimensions of the tank. You must enter data in all fields shown.
Sensor Height: Distance from the bottom of the tank to the bottom of the sensor.
Fill Height: Distance from the bottom of the tank to the operational full level of liquid (20mA). This setting defines the location of full current output and is the top of the sensor’s measurement range.
Riser Height: Distance the sensor is above the top of the tank (including all fittings). Measured from the bottom of the sensor.
Gallons
Liters

| 52 MN300820 Rev B
WebCal® Appendix (continued) Section Seven
Volume – Sensor Output Units (Vertical Cylinder Example):
Enter the dimensions of the tank. You must enter data in all fields shown.
Sensor Height: Distance from the bottom of the tank to the bottom of the sensor.
Fill Height: Distance from the bottom of the tank to the operational full level of liquid (20mA). This setting defines the location of full current output and is the top of the sensor’s measurement range.
Riser Height: Distance the sensor is above the top of the tank (including all fittings). Measured from the bottom of the sensor.
Tank Height: Distance from the bottom of the tank to the top of the straight side wall. Diameter: Distance of the inside tank diameter.
Volume – Tank Capacity (Vertical Cylinder Example): After entering the dimensions, press the Capacity
button to show the Calculated Capacity of the tank. If the Calculated Capacity is slightly different than the expected capacity, click on the Adjust Capacity box and enter the expected capacity of the tank. If the Adjusted Capacity is more than 10% of the Calculated Capacity, recheck the dimensions information entered above.
When all dimensions are entered, press the Apply button to return to the previous Configuration window.
Apply – Transfers the dimensions to the original Configuration window.  Tanks – Returns to the previous Shape Selection window.  Cancel – Returns to the Configuration window without saving any information.  Help – Jumps to the Help menu.
MN300820 Rev B 53 |
WebCal® Appendix (continued) Section Seven
TANK LEVEL CONFIRMATION:
This section of WebCal® is where you confirm the values set in the previous step. The values were entered under the Dimensional entry window. To edit these settings, you must go back to the Dimensional entry window via the Volumetric Mode button.
Height Units: Units selected for configuration. When used as a device to measure the distance (height
of liquid), the options are inches, cm, feet or m. When used as a device to measure the volume of liquid, the options are gallons or liters.
Sensor Height: Distance from the bottom of the tank to the bottom of the sensor.  Fill-Height: Distance from the bottom of the tank to the operational full level of liquid (20mA). This
setting defines the location of full current output and is the top of the sensor’s measurement range.
Capacity: The total volume of the tank. Only shown when gallons or liters are selected.
Note: By extending the empty (4mA) to the bottom of the tank, the 4-20 mA output will track the volume of the tank. This allows any local display to read the actual volume of liquid without the need for any unique configuration. This feature is very useful with any non-linear tanks such as horizontal, spherical or tanks with cone bottoms.
HeightUnits
Capacity
SensorHeight
FillHeight
Volumetric Sensor Output: The volumetric mode button will be highlighted in Blue when a volume output is selected.
| 54 MN300820 Rev B
WebCal® Appendix (continued) Section Seven
WRITE TO UNIT
Write to Unit
After you have entered configurations, selected and configured the Tank Shape and entered the Tank Values, click “Write to Unit” and load the configuration into the memory of the sensor. When completed, this configuration will remain inside the sensor memory and will not change unless the sensor is connected to WebCal
®
and a new configuration is written to the sensor. Loss of power will not change or lose the configuration within sensor memory.
Next, use the file management features to save your configuration by clicking “Save Config File” and print your wiring diagram by clicking “Wiring Diagram.”
Save Config File” will save this configuration as a text file which can be loaded back into WebCal® by pressing the “Open Config File” button. It is good practice to save the configuration file for each different configuration with a unique name for easy identification. If using multiple sensors in identical applications, then use of a single configuration file is recommended.
Wiring diagram
Save Config File
Wiring Diagram” will display a PDF file showing the unique wiring for the specific configuration created in WebCal™. The PDF can be printed or emailed. It is good practice to save the wiring diagram as a backup
Advanced” is a feature setting designed to help solve performance or operational issues for specific applications. Changing these setting will alter the factory default performance or operation, of your sensor.
Increase Output Filtering: Placing a check mark in the box will increase the filtering (averaging) of the
analog output. Use this filter if the 4 to 20 mA output requires a smoother output for the application such as open channel flow measurement.
Decrease Output Filtering: Placing a check mark in the box
will eliminate all filtering (averaging) of the analog output which enables a pulse by pulse level reading. Use this filter to see changes in level after every echo pulse.
Note: Never check increase output filtering and decrease output filtering at the same time.
Stabilize Output in Dead Band: Placing a check mark in the box will activate a filter to hold the output
at Full if the level enters the dead band of the EchoPod
®
. This filter requires the level to leave the dead
band at a smooth and steady rate.
Reduce Fast Level Jump Response Time: Changes the sensor’s response time from 60 seconds to
10 seconds. This filter prevents the sensor from making a quick jump in level if a false signal suddenly appears. Change the response time if application involves expected quick level changes.
.
Increase LOST Response Time: Changes the sensor’s response time from 60 seconds to 180
seconds. This filter sets the time the sensor waits before entering a LOST state. Change the response time if you want to delay the sensor from stating LOST.
MN300820 Rev B 55 |
Display Appendix Section Eight
AIR GAP VS. LIQUID LEVEL (HOW TO SELECT THE DISPLAY SETTING)
The display can be made to display either the height of liquid in the tank (LIQUID mode) or the amount of air in the tank (AIR mode). This function does not influence the current output, just the display on the sensor.
1. In TOP-LEVEL MENU mode, select DISPLAY.
2. Select LIQUID or AIR.
a. Press and hold OK on the selected setting for
3 seconds then release.
b. To cancel DISPLAY, Select EXIT.
3. Select EXIT to return to TOP-LEVEL MENU.
Liquid Mode Air Mode
| 56 MN300820 Rev B
Display Appendix (continued) Section Eight
HOW TO REVERSE THE CURRENT OUTPUT
The factory setting for Echotouch® has 4mA set at the bottom of the tank and 20mA at the top. If the application requires this output to be reversed (20mA at bottom and 4mA at top), then use the REV MA setting to reverse the current output.
With REV MA set to NORM (factory default), 4mA will be at the bottom and 20 mA at the top o f the t ank. With REV MA set to REV, 20mA will be at the bottom and 4mA at the top of the tank.
1. In TOP-LEVEL MENU mode, select
OUTPUT.
2. Select REV MA.
3. Select 4 BOTT or 4 TOP.
a. Press and hold OK on the selected
setting for 3 seconds then release.
a. To cancel REV MA, Select EXIT.
4. Select EXIT to return to TOP-LEVEL MENU.

MN300820 Rev B 57 |
Display Appendix (continued) Section Eight
HOW TO SETUP THE START-UP POWER (STRTUP):
During the window of time between when power is first applied to the sensor and the sensor acquires the level reading, the output current can be preset in order to avoid any false alarms during this window. The options for start-up are 4mA, 12mA, 20mA or 22mA.
1. In TOP-LEVEL MENU mode, select
OUTPUT.
2. Select STRTUP.
3. Select 4mA, 12mA, 20mA or 22mA.
a. Press and hold OK on the selected
setting for 3 seconds then release.
a. To cancel STRTUP, Select EXIT.
4. Select EXIT to return to TOP-LEVEL MENU.
| 58 MN300820 Rev B
Display Appendix (continued) Section Eight
VALUES MENU
Functions include providing setup information (SETUP), a diagnostic to be used in conjunction with a Flowline Representative (DIAG) and the ability to reset the Echotouch® (RESET).
SETUP
Will display the configuration settings for EchoTouch
®
.
1. In TOP-LEVEL MENU mode, select VALUES.
2. Select SETUP.
3. Setup will list the following information:
a. UNITS, b. DISPLY, c. REV MA, d. SAFE, e. HEIGHT, f. FILL-H g. START UP
4. Select EXIT to return to TOP-LEVEL MENU.
DIAGNOSTIC (DIAG) PARAMETERS
This mode runs diagnostic tests that confirm operation of EchoTouch
®
. This is a production test feature used by the factory to confirm operation.
1. In TOP-LEVEL MENU, select VALUES.
2. Select DIAG.
3. DIAG will list the following settings:
a. Display Test b. RANGE c. TEMP d. ECHO e. POWER f. VER
4. Select EXIT to return to TOP-LEVEL
MENU.
Note: This mode should only be used when supervised by a Flowline representative.
MN300820 Rev B 59 |
Display Appendix (continued) Section Eight
RESET
EchoTouch® enables the end user to reset the entire configuration back to the original factory settings.
2. In TOP-LEVEL MENU mode, select VALUES.
3. Select RESET.
4. Select YES. a. To cancel the reset, Select NO.
5. Select EXIT to return to TOP-LEVEL MENU.
 
| 60 MN300820 Rev B
®
Display Appendix (continued) Section Eight
FACTORY SETTINGS
Echotouch® Sensor Height Fill-Height
Loop Fail-
US12 Series US06 Series
472.4” (1200 cm) 454.4” (1154.3 cm)
236.2” (600 cm) 228.2” (579.7 cm)
US03 Series 118.1” (300 cm) 114.1” (290 cm) US01 Series 59.1” (150 cm) 57.6” (146.2 cm)
USER SETTINGS
Fill out the chart below and keep as a record of your configuration.
Tank
Height = Fill-H =
Units
Inches Feet cm Meter
Display
Air Liquid
Safe
Safe
Hold last
Output @
Empty
4mA at Bottom
Startup
Condition
Empty
22mA 21 mA 20mA Hold Last 4mA
Rev mA
4mA @ Bottom 4mA at Top
Start-up
4mA 12mA 20mA 22mA
TROUBLESHOOTING
PROBLEM SOLUTION
No Unit Detected in WebCal
®
.
If WebCal
Check that the LI99-2001 Fob is connected to the USB
cannot detect EchoTouch® when connected to the computer:
®
port.
Check that all four wires (Red, Black, White and Green) are securely
attached to the Fob.
Check Device Manager and confirm that both drivers (WebCal
®
Configuration & EchoFob) are present.
®
If the above do not work, reinstall the WebCal
software while the computer
is in Safe Mode.
Display shows FULL: Level of liquid is above the FILL-H setting: Check the FILL-H setting, making
sure the FILL-H setting is high enough so the level of liquid is below the FILL- H setting. The Fill-H setting is the distance from the bottom of the tank to the Full level of liquid.
MN300820 Rev B 61 |
Display Appendix (continued) Section Eight
TROUBLESHOOTING
PROBLEM SOLUTION
Display shows EMPTY: Level of liquid is beyond the HEIGHT setting: Check the HEIGHT setting,
making sure the HEIGHT setting is low enough so the level of liquid is above the HEIGHT setting.
Display shows LOST: Sensor is in a Fail-Safe state. The return sound pulses are not reaching the
transducer. First, cycle power off and on, waiting 10 seconds between the off and on states. If problem persists, check the installation fitting against the Installation instructions in the manual.
Display shows WARMUP: WARMUP appears when power is being applied to transmitter and the sensor
cannot acquire a valid return signal. This is similar to LOST state. Make sure the liquid level is within the operational range configured within the transmitter (i.e. Level is not further than HEIGHT setting and not above the FILL-H setting). If problem persists, check the installation fitting against the Installation instructions in the manual as well as check for a weak power supply or bad
wire connections. Display is opposite of the measured value:
Check the DISPLAY setting. AIR mode indicates the distance from the liquid
to the sensor. LIQUID mode indicates the height of liquid in the tank. Change
the DISPLAY mode from AIR to LIQUID or vice versa to correct. Transmitter indicates a current of 0 mA: Transmitter jumps to a current reading between 19 and 20 mA:
Check the wiring for an open circuit. An open circuit is the most common issue
with a 0 mA signal.
Check the installation of the transmitter. Bad installation fittings will cause
false signals near the top of the tank, which typically translates to a signal
between 19 and 20 mA. Also look for interference just below the transmitter. If
the transmitter is installed in a metal fitting, switch to a plastic fitting. Transmitter indicates a current over 23 mA:
Immediately check the wiring for a short circuit. The Echotouch® is current
limited to 22 mA. Anything above 23 mA indicates a short circuit. No unit detected error 1 Both errors can occur when the Display is still plugged in when connecting to Data page read error 2 Data page read error X (X is any number greater than
WebCal. Remove the display from its plug and try again.
Make sure display is removed. If issue persists, contact a Flowline
representative.
2)
-1 No device is attached to the USB Fob, or it is attached incorrectly. Please correct and try again.
1. Check the wires connected to the USB Fob as well as to the EchoPod terminals.
2. Make sure the cable length between the sensor and the computer is less than 15’ of total cable.
3. Make sure the display was removed from its plug.
4. If this persists, reboot computer in SAFE mode and re-install the WebCal installer program.
-2 Program to loader error open wires
Check the wires connected to the USB Fob as well as to the EchoPod terminals.

®
| 62 MN300820 Rev B
Display Appendix (continued) Section Eight
MN300820 Rev B 63 |
Warranty, Returns and Limitations Step Nine
WARRANTY
Flowline warrants to the original purchaser of its products that such products will be free from defects in material and workmanship under normal use and service in accordance with instructions furnished by Flowline for a period of two years from the date of manufacture of such products. Flowline's obligation under this warranty is solely and exclusively limited to the repair or replacement, at Flowline's option, of the products or components, which Flowline's examination determines to its satisfaction to be defective in material or workmanship within the warranty period. Flowline must be notified pursuant to the instructions below of any claim under this warranty within thirty (30) days of any claimed lack of conformity of the product. Any product repaired under this warranty will be warranted only for the remainder of the original warranty period. Any product provided as a replacement under this warranty will be warranted for the full two years from the date of manufacture.
RETURNS
Products cannot be returned to Flowline without Flowline's prior authorization. To return a product that is thought to be defective, go to flowline.com, and submit a customer return (MRA) request form and follow the instructions therein. All warranty and non-warranty product returns to Flowline must be shipped prepaid and insured. Flowline will not be responsible for any products lost or damaged in shipment.
LIMITATIONS
This warranty does not apply to products which: 1) are beyond the warranty period or are products for which the original purchaser does not follow the warranty procedures outlined above; 2) have been subjected to electrical, mechanical or chemical damage due to improper, accidental or negligent use; 3) have been modified or altered; 4) anyone other than service personnel authorized by Flowline have attempted to repair; 5) have been involved in accidents or natural disasters; or 6) are damaged during return shipment to Flowline. Flowline reserves the right to unilaterally waive this warranty and dispose of any product returned to Flowline where: 1) there is evidence of a potentially hazardous material present with the product; or 2) the product has remained unclaimed at Flowline for more than 30 days after Flowline has dutifully requested disposition. This warranty contains the sole express warranty made by Flowline in connection with its products. ALL IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION, THE WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSLY DISCLAIMED. The remedies of repair or replacement as stated above are the exclusive remedies for the breach of this warranty. IN NO EVENT SHALL FLOWLINE BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES OF ANY KIND INCLUDING PERSONAL OR REAL PROPERTY OR FOR INJURY TO ANY PERSON. THIS WARRANTY CONSTITUTES THE FINAL, COMPLETE AND EXCLUSIVE STATEMENT OF WARRANTY TERMS AND NO PERSON IS AUTHORIZED TO MAKE ANY OTHER WARRANTIES OR REPRESENTATIONS ON BEHALF OF FLOWLINE. This warranty will be interpreted pursuant to the laws of the State of California. If any portion of this warranty is held to be invalid or unenforceable for any reason, such finding will not invalidate any other provision of this warranty.
For complete product documentation, video training, and technical support, go to flowline.com. For phone support, call 562-598-3015 from 8am to 5pm PST, Mon - Fri. (Please make sure you have the Part and Serial number available.)
 
| 64 MN300820 Rev B
Loading...