FLIR A500-EST, A700-EST Specifications

FIXED MOUNT FLIR ESTTM THERMAL SCREENING SOLUTIONS
FLIR A500/A700-EST
FLIR A500-EST and A700-EST thermal cameras are non-contact screening tools that serve as a rst line of defense against potential health risks. These FLIR ESTTM thermal screening solutions detect and visualize heat to quickly identify individuals with an elevated skin temperature. With on-camera FLIR Screen­ESTTM Mode, the A500/A700-EST cameras can be deployed as a single, standalone screening station or in a network. Compatibility with industry standards such as Modbus TCP, MQTT, RESTful API for data transfer and RTSP for video makes third-party integration easy. The cameras can be connected to most Video Management Systems and are compatible with FLIR Screen-ESTTM Desktop software.
Contagions such as COVID-19, SARS, and other diseases can produce symptoms like elevated skin temperature—a possible sign of infection. While FLIR cameras are not capable of detecting or diagnosing viruses, these US FDA registered cameras represent a simple, preliminary measure for mitigating further contagion and possible rebound, providing the condence to return to normalcy.
www.ir.com/A500-A700-EST
EASY SET-UP & OPERATION
Begin screening quickly with limited ramp­up time and simple connections
• Integrated web browser with intuitive interface for simple camera set-up
• Standard Ethernet and Wi-Fi connectivity, and Power over Ether net for single- cable installation
• Integra tes easily into web services with R ESTful API over XML or JSON, while ONVIF compliance accommodates standard security VMS and NVR solutions
• Multiple case mounting points to support tripod or permanent xture installations
FAST, ACCURATE SCREENING
Performance hardware, smart analytics, and reliable calibration optimize the screening process
• On-camera FLIR Screen-EST Mode offers visual pass/fail graphic indicators and audible* alarms, enabling rapid decision-making
• Compatible with FLIR Screen-EST Desktop software, with automatic detection and measurement on faces for faster screening throughput
• Calibration with ambient drift compensation allows accurate screening without a reference
• Support for external blackbody compensation
*Via web interface
MAINTAINS SAFETY & PRIVACY
FLIR screening solutions are non-contact, safe, and effective
• On-camera FLIR Screen-EST Mode does not automatically store or record images or personal information
• Thermal imagery displays heat, not identifying facial features
• Thermal temperature measurement does not require personal contact and allows for social distancing in screening setups
SPECIFICATIONS
Imaging and optical data A500-EST A700-EST
Infrared resolution 464 × 3 48 pixels 640 × 4 80 pixels
Visual camera resolution 1280 × 960
Thermal resolution/NETD <40 mK@ 30°C/86°F (24° lens)
<30 mK @ 30°C/86°F (42° lens)
Lens 24° or 42°
Focal length 17 mm (24° lens) or 10 mm (42° lens)
Field of view 24° × 18° or 42° × 32°
Spatial resolution (IFOV) 0.90 mrad/pixel (24° lens)
1.66 mrad/pixel (42° lens)
0.66 mrad/pixel (24° lens)
1.20 mrad/pixel (42° lens)
Focus One-shot contrast, motorized, manual
Framerate 30 Hz
Detector data
Focal plane array/spectral range Uncooled microbolometer/7.5–14 µm
Detector pitch 17 µm 12 µm
Screening mode
Temperature measurement range 15°C to 45°C (59°F to 113°F)
Screening accuracy (drift) ±0.3°C (±0.5°F)
Image presentation
Digital data streaming Simultaneous thermal and visible
Command and control Ether net and Wi-Fi
Ethernet
On-C amera FLIR EST
TM
Mode
FLIR Screen -EST Mode is an on-camera method for simplied measure­ment of elevated skin temperature. This mode can display an alarm when a temperature greater than a user dened threshold is dete cted against a sample d average value. If the screening mode detects an individual with el­evat ed skin temperature , they can the n be eval uated using a me dical device such as a t hermome ter. In this way, FL IR Screen -EST Mo de provide s a faster, safer, and more reliable method for conducting elevated skin temperat ure screening.
FLIR ESTTM Desktop Software
Ether net connector type and standard M12 8-pin X-coded, Female; 1000 Mbps, IEEE 802.3
Ethernet power Power o ver Ethernet, PoE IEEE 802.3af class 3
Modbus TCP Master, Modbus TCP Slave, EthernetIP, IEEE 158 8, MQT T,
Ethernet protocols
SNMP, TCP, UDP, SNTP, RTSP, RTP, HTTP, HTTPS, ICMP, IGMP, sftp
(server), FTP (client) SMTP, DHCP, MDNS (Bonjour), uPnP
Wi-Fi
Conne ctor t ype and s tandard Female RP-SMA; IEEE802.11a/b/g/n
Connections Peer to peer (ad hoc)or infrastruct ure (net work)
General
Power PoE
External voltage Allowed range = 18-56 VDC, 8 W max
Size (L × W × H) 123 × 77 × 77 mm (4.84 × 3.03 × 3.0 3 in)
Weight 0.82 kg (1.8 lb)
Mounting Base mounting: 4 × M4 on 4 side s
Box contents Infrared camera with lens, Ethernet cable M12 to RJ45F (0.3 m/0.98 ft),
CORPORATE HEADQUARTERS
FLIR Systems, Inc. 27700 SW Parkway Ave. Wilsonville, OR 97070 USA
NASHUA
FLIR Systems, Inc. 9 Townsend West Nashua, NH 030 63 USA PH: +1 866. 477.3687
PH: +1 866. 477.3687
LAT IN A MERIC A
FLIR Systems Brasil Av. Antonio Bardella, 320 Sorocaba, SP 180 85 -852 Brasil PH: +55 15 32 38 8070
CANADA
FLIR Systems, Ltd. 3430 South Service Road, Suite 103 Burlington, ON L7N 3J5 Canada PH: +1 800.613.0507
Tripod mounting: UNC ¼”-20 on 2 sides
printed documentation including login for web interface
FLI R Screen- EST™ D esktop i s a compute r screening s oftw are for FL IR T-Series, Exx-Series, and Axxx-Series thermal imaging cameras. The sof tware deploys automatic measurement tools such as face detection and automatic average sampling to shor ten screening times for individuals to two seconds. F ast scre ening perfo rmance mak es FLIR Sc reen-ES T Desktop t he prefer red solutio n for screening application at entries, checkpoint s, and other high-t rafc areas while main taining recommende d social dista ncing guideline s.
DISCL AIMER: FLIR devices are intended for use as an adjunct to clinical pro ­cedures in the screening of skin sur face temperature. Various environmental and methodological factors can impact thermal imaging; t herefore, it should not be relied upon as t he sole determinant of a person’s body t emperature. Use of a me dical devic e will be neede d to identif y elevate d body temp erature.
www.ir.com NASDAQ: FLIR
Equipme nt describ ed herein is su bject to US ex port regulat ions and may re quire a licen se prior to exp ort. Diversi on contrar y to US law is prohibited . Imager y for illustration purp oses only. Sp ecica tions are sub ject to change w ithout not ice. ©20 20 FLIR Sy stems, Inc . All rights reserved. C reated: 07/10 /20
20- 0861-INS-A700 -EST- Datasheet- LTR