FISCHER MS13 Operating Manual

Page 1
Operating manual
MS13
Digital pressure transmitter / switch
with colour change display
09005885 • BA_EN_MS13 • Rev. ST4-A • 10/16
*09005885*
Page 2
2 / 32 BA_EN_MS13
Masthead
Manufacturer:
FISCHER Mess- und Regeltechnik GmbH
Bielefelderstr. 37a D-32107 Bad Salzuflen
Telefon: +49 5222 974 0 Telefax: +49 5222 7170
eMail: info@fischermesstechnik.de web: www.fischermesstechnik.de
Technical editorial team:
Documentation representative: S. Richter Technical editor: R.Kleemann
All rights, also those to the translation, reserved. No part of these instructions may be reproduced or processed, duplicated or distributed using electronic sys­tems or any other form (print, photocopy, microfilm or another process) without the written consent of the company Fischer Mess- und Regeltechnik GmbH, Bad Salzuflen.
Reproduction for internal use is expressly allowed. Brand names and procedures are used for information purposes only and do
not take the respective patent situation into account. Great care was taken when compiling the texts and illustrations; nevertheless, errors cannot be ruled out. The company FISCHER Mess- und Regeltechnik GmbH will not accept any legal responsibility or liability for this.
Subject to technical amendments.
© FISCHER Mess- und Regeltechnik 2016
Version history
Rev. ST4-A 04/16 Version 1 (first edition)
Page 3
FISCHER Mess- und Regltechnik GmbH Table of Contents
BA_EN_MS13 3 / 32
Table of Contents
1 Product and functional description ..............................................................................................................4
1.1 Delivery scope ..........................................................................................................................................4
1.2 Performance features ...............................................................................................................................4
1.3 Intended use .............................................................................................................................................4
1.4 Product Overview .....................................................................................................................................5
1.5 Function diagram ......................................................................................................................................6
1.6 Design and mode of operation..................................................................................................................7
2 Installation and assembly..............................................................................................................................8
2.1 Generalities...............................................................................................................................................8
2.2 Process connection ..................................................................................................................................8
2.2.1 Measuring lines that need to be connected ..................................................................................9
2.2.2 Pressure surge absorption............................................................................................................9
2.3 Electrical connections .............................................................................................................................11
3 Commissioning.............................................................................................................................................12
3.1 General ...................................................................................................................................................12
3.2 Control elements.....................................................................................................................................12
3.2.1 LC display ...................................................................................................................................12
3.2.2 Keyboard..................................................................................................................................... 13
3.3 Menu levels.............................................................................................................................................14
3.3.1 Menu Level Switch points ...........................................................................................................14
3.3.2 Menu Level Input ........................................................................................................................15
3.3.3 Menu Level Measuring................................................................................................................ 16
3.3.4 Menu Level Output...................................................................................................................... 17
3.3.5 Menu Level Function................................................................................................................... 18
3.3.6 Menu Level Display..................................................................................................................... 20
3.3.7 Menu Level System .................................................................................................................... 22
4 Technical Data .............................................................................................................................................. 24
4.1 General Information ................................................................................................................................24
4.2 Input variables ........................................................................................................................................24
4.3 Output parameters..................................................................................................................................24
4.4 Measurement accuracy ..........................................................................................................................25
4.5 Auxiliary energy .....................................................................................................................................25
4.6 Application conditions .............................................................................................................................25
4.7 Display and operating interface ..............................................................................................................26
4.8 Construction design ................................................................................................................................26
4.8.1 Dimensional picture .................................................................................................................... 26
4.8.2 Process connection..................................................................................................................... 27
5 Order Codes..................................................................................................................................................28
5.1 Accessories ............................................................................................................................................29
6 Attachments..................................................................................................................................................30
Page 4
4 / 32 BA_EN_MS13
1 Product and functional description
1.1 Delivery scope
• MS13 Digital pressure transmitter / switch
• Operating instructions
The unit is supplied ex-works with standard parameters. For accessories, please see the list in section Order code [}29].
1.2 Performance features
Important features
• LCD colour change display
• Switchable pressure units
• 2 independent switching points with lots of configuration options
• Analogue signal output with possibility of characteristic curve spread, char­acteristic curve reversal and offset setting
• Characteristic curve implementation via table with max. 30 measuring points
• Complete adjustment of all parameters and measuring point protocol pos­sible through optional transmitter PC interface.
Typical applications
• Simple pump control systems
• Monitoring of pumps and compressors
• Filling level measuring
Application areas
• Technical facility equipment (TGA)
• Process engineering
• Process technology
• Environmental technology
1.3 Intended use
The MS13 is a pressure transmitter/ switch for measuring relative pressure. The unit is suitable for measuring pressure and under-pressure in non-agressive gas-like and fluid media.
The MS13 complies with the state-of-the-art and is safe; it also takes into ac­count the relevant regulations and EC directives. The manufacturer will not be liable for damage arising from incorrect or improper use.
The unit was designed bearing in mind all relevant factors that could impact on its safety. Also, the unit was produced, inspected and supplied with a user manual so that if it is used in unforeseeable reasonable conditions, its safety is guaranteed for its entire service life.
NOTICE
Soiled or aggressive media
Please contact the manufacturer before using this unit with dirty or aggressive media because the unit needs to be adapted for the specific customer in terms of the parts that come into contact with the media.
The device may only be used for the purpose stipulated by the manufacturer.
Page 5
FISCHER Mess- und Regltechnik GmbH Product and functional description | 1
BA_EN_MS13 5 / 32
1.4 Product Overview
All units of the series MS13 are supplied in an NG100 bayonet ring housing made of stainless steel.
1 2
3
4
5
6
7
Code U
Process connection
Illustration1: Product Overview
1 Colour change display 2 Bayonet ring housing 3 Keyboard 4 Process connection 5 Blowout plug 6 M12 plug 1 7 M12 plug 2
Process connection
The following options are available for the process connection.
Code L
Process connection
0 R
Illustration2: Process connection direction
Code 85 87 88 S1
G½ B ¼ -18 NPT EXTG¼ B Schrader ®
Illustration3: Process connections
Page 6
6 / 32 BA_EN_MS13
Code Process connection
85 Connection shanks with external thread G¼ B 87 Connection shanks with external thread G½ B 88 Connecting port with outer thread ¼ -18 NPT EXT S1 Schrader® screw connection inner thread 7/16 UNF
Electrical connections
The power is connected using two M12 connectors.
M12 flanged connector DIN EN 61076-2-101 coding A
Connector 1 5-pin Ms-nickel-plated Connector 2 4-pin Ms-nickel-plated
Nameplate
This type plate serves as an example of the information that is stated. The data shown is purely fictive, but does correspond to the actual conditions. For more information, please see the order code at the end of these instructions.
MESS- UND REGELTECHNIK GmbH
D-32107 Bad Salzuflen
Article no MS1303M87PKC0MV0 Measuring range 0-1.6 bar electrical output 4 - 20 mA / 3-wire
Oper. voltage 24V AC/DC P max. 3.2 Bar (45 PSI)
Prod. No. 1600624.01.001
Made in Germany
142
3
142
3
*
ge/gnswbr
wsblsw
br
ws
bl
+ Signal
+ supply
- Signal
- supply
( GND)
Plug guide
Switch contact 1
Switch contact 2
* bridged
internally
Order Codes
Serial number
Wiring diagrams
Technical data
Illustration4: Nameplate
1.5 Function diagram
SP2
1.275 bar
U/I
SP1
D
A
A
D
µC
IJK
3
4 5
6
7
2
1
Illustration5: Function diagram
Page 7
FISCHER Mess- und Regltechnik GmbH Product and functional description | 1
BA_EN_MS13 7 / 32
1 Measuring cell 2 A/D converter 3 Micro-controller 4 Keyboard 5 Switching outputs 6 LCD colour change display 7 D/A converter
1.6 Design and mode of operation
The device is based on a ceramic sensor element that is suitable for measuring over-pressure and under-pressure. The pressure deforms the measuring mem­brane causing a change in resistance on the attached measuring bridge. This change is evaluated by the device's electronics and transformed into the dis­play, switch contacts or an standardised analogue output signal.
The optional output signal can be dampened, spread, inverted and transformed via a table function even if it is non-linear. Overstepped limits can be visualised with the LCD colour change display.
The unit is configured with a keyboard or by means of remote configuration from a PC.
Page 8
8 / 32 BA_EN_MS13
2 Installation and assembly
2.1 Generalities
The instrument may only be installed and commissioned by specialized person­nel familiar with the installation, commissioning and operation of this product.
Specialized personnel are persons who can assess the work they have been assigned and recognize potential dangers by virtue of their specialized training, their skills and experience and their knowledge of the pertinent standards.
WARNING
Mounting pressure transmitters
During assembly, observe the respective national and international guidelines and safety regulations.
Only mount the pressure transmitter to systems that are depressurized. Only ever operate the unit within the permitted measuring range or below the max­imum overload.
Illustration6: Shutoff valve.
The device is set ex-works for vertical installation, however any installation posi­tion is possible.
To guarantee safe working conditions during installation and maintenance, suit­able stop valves must be fitted in the system (see accessories). By means of the manometer shutoff, the unit
• Can be depressurized or taken out of operation.
• Be disconnected from the power supply within the applicable system for re­pairs or inspections.
2.2 Process connection
• By authorized and qualified specialized personnel only.
• The pipes need to be depressurized when the instrument is being connec­ted.
• Appropriate steps must be taken to protect the device from pressure surges.
• Check that the device is suitable for the medium being measured.
• Maximum pressures must be observed (cf. Tech. data)
WARNING
Earth connection via the system earth
During assembly, ensure that the earth connection between the pressure trans­mitter and the system earth is ensured. The connection to the system earth is realised via the process connection. Therefore, never use an insulated Teflon tape or similar. Design the process connection acc. to EN 837 and use a suit­able flat seal.
Page 9
FISCHER Mess- und Regltechnik GmbH Installation and assembly | 2
BA_EN_MS13 9 / 32
2.2.1 Measuring lines that need to be connected
The following points need to be observed when connecting the pressure line:
• To ensure there is no influence on the measured values, severe bends and coils in the wire should be avoided.
• To prevent deposits, there should be a continuous incline or drop of at least 8%.
• When measuring steam pressure, a water bag-forming loop must be provided due to the temperature (see accessories).
Round shape U-shape
ca. 180
Ø20
56
155
ca. 200
ca. 275
ca. 130
Ø20
Ø56
SW27
Manometer connection
Connection spigot according to DIN EN 837-1
Illustration7: Siphon MZ1###
• The transmitter must be positioned below the measuring point for liquid measurements. Vent the pressure line before commissioning.
• The transmitter must be positioned above the measuring point for gas measurements.
2.2.2 Pressure surge absorption
Pulsating pressure on the system side can lead to functional problems. We re­commend installing a damping element in the pressure connection lines as a protective measure.
a) Capillary throttle
Manometer connection
SW27 SW27
Connection spigot
according to DIN EN 837-1
305
110
Ø88
Ø4
Illustration8: Capillary throttle MZ400#
Page 10
10 / 32 BA_EN_MS13
b) Settable damping reactor
In operating mode, the damping throttle must be set so that the output signal follows the pressure changes with a delay.
20
66
19
SW27
Setting screw
Connection spigot according to DIN EN 837-1
Manometer connection
Illustration9: Damping reactor MZ410#
Page 11
FISCHER Mess- und Regltechnik GmbH Installation and assembly | 2
BA_EN_MS13 11 / 32
2.3 Electrical connections
• By authorized and qualified specialized personnel only.
• When connecting the unit, the national and international electro-technical regulations must be observed.
• Disconnect the system from the mains, before electrically connecting the device.
• Install the consumer-adapted fuses.
• Do not connect the connector if strained.
3-Leiterschaltung
Input
Output
AC
DC
Sensor
Signal
+U
b
- U
b
- Sig
+Sig
+U
b
- U
b
p
1
3
2
4
Connector 1: Supply and output signal
Illustration10: M12 connector 5­pin
Item Description Cable colour
A Coding A B Internal bridge 1 +U
b
Supply brown 2 -Sig Signal white 3 -U
b
Supply blue 4 +Sig Signal black 5 FE Functional earth Green/yellow
Connector 2: Switching outputs
Illustration11: M12 Plug 4-pin
Pos Description Cable colour
A Coding A 1 SP1 Switch point 1 (no) brown 2 SP2 Switch point 2 (no) white 3 SP2 Switch point 2 (com) blue 4 SP1 Switch point 1 (com) black
Page 12
12 / 32 BA_EN_MS13
3 Commissioning
3.1 General
All electrical supply, operating and measuring lines, and the pressure connec­tions must have been correctly installed before commissioning. All supply lines are arranged so that there are no mechanical forces acting on the device.
Check that the pressure connections do not leak before commissioning.
3.2 Control elements
3.2.1 LC display
1
3
4
2
Illustration12: LC display
1 Measurement display 4…6 digits 2 Bar chart display 3 Status display of the switch
points
4 Unit
Illustration13: LC display 2SP
The unit is shown to the right of the measured value. If the device is equipped with contacts, a closed contact is always symbolised by an inverted text "SP1" or "SP2". One exception is the 1-channel bar chart diagram. Here, the switching points are symbolised with simple numbers "12".
Page 13
FISCHER Mess- und Regltechnik GmbH Commissioning | 3
BA_EN_MS13 13 / 32
3.2.2 Keyboard
1
2
3
Illustration14: Keyboard
1 Page down menu Decrease value 2 Call up menu Save value OK 3 Page up menu Increase value
The individual menu items and parameters can be displayed using the buttons
and
. The respective menu item is selected or the parameters for making
changes are called up via the button
.
If a parameter can be changed, the display flashes. The change is made via the buttons
and
. The value is saved with the button
.
To leave a menu level or the entire menu, select the parameter "Quit" and press
.
Example: Switch-on point set switchpoint 1
In normal mode, press the button
to enter the menu. The menu level Switch
points appears. Press the enter key
again to call up the display parameter.
The first parameter Switch point 1 on is displayed. To change this parameter, press the button
again.
The device jumps to the input:
• The parameter is stated in the 1st line.
• The value that is to be changed is shown in the 2nd line, the display flashes.
• The input limits are displayed in the 3rd line (if there is one).
The required value is set with the buttons
and
and then confirmed with
.
Page 14
14 / 32 BA_EN_MS13
3.3 Menu levels
The menu levels are structured as follows:
End menu level
Menu Level Switch points
Menu Level Input
Menu Level Measuring
Menu Level Output
Menu Level Function
Menu Level Display
Menu Level System
End menu level
Reception
Output
Output
Illustration15: Menu levels
The following tables provide an overview of the parameters of the individual menu levels. In the Menu Level Systemyou can change to the respective na­tional language using the language parameter. You can see which languages are supported there.
3.3.1 Menu Level Switch points
Parameter name Description Value range
SP1 On Switch point 1 On MBA-50% … MBE+50% SP1 Off Switching point 1 off MBA-50% … MBE+50% SP1 delay Switching point 1 delay 0…1800 s SP1 Function Switching point 1 function NO, NC SP2 On Switch point 2 On MBA-50% … MBE+50% SP2 Off Switching point 2 off MBA-50% … MBE+50% SP2 delay Switching point 2 delay 0…1800 s SP2 Function Switching point 2 function NO, NC
The two switching outputs are configured by four parameters respectively. For the switch point 1 these are
Page 15
FISCHER Mess- und Regltechnik GmbH Commissioning | 3
BA_EN_MS13 15 / 32
SP1 On
SP1 Off
SP1 delay
SP1 Function
Accordingly for switch point 2:
SP2 On
SP2 Off
SP2 delay
SP2 Function
The function of the individual parameters is explained for both switch points us­ing Switch point 1 as an example.
SP1 On defines the activation point, SP1 Off the deactivation point of switching
output 1. The values are shown in the valid unit and set accordingly. The values are shown in the valid unit and set accordingly. Both parameters can be set in­dependently over the entire value range.
The value range ranges from MBA – 50% to MBE + 50%. MBA stands for start of measuring range and MBE for the end of the measuring range.
Example: Measuring range = 0 … 160 kPa
The value range for this measuring range is -80 kPa … +240 kPa.
Together, the two parameters SP1 On and SP1 Off determine the switch func­tion of switching output 1:
• If SP1 On > SP1 Off, the output switches on, if the measured value ex­ceeds SP1 On. It is only switched off again if the measured value SP1 Off is undercut (hysteresis function).
• If SP1 On = SP1 Off, the output switches on if the measured value exceeds SP1 On and off if the measured value undercuts the same value (SP1 Off).
• If SP1 On < SP1 Off, the output switches on, if the measured value lies within these switch points: i.e:
SP1 On < Measured value < SP1 Off (window function).
SP1 Delay allows the reaction of the switch output to be delayed by between 0
and 1800 s. This parameter applies equally for switching on and off.
SP1 Function changes the function of the switching output 1. It is possible here
to define whether the contact should work as a open contact (NO) or a break contact (NC).
3.3.2 Menu Level Input
Parameter name Description Value range
Absorption Attenuation, damping 0…100 s Offset corr. Offset correction ⅓ basic measuring
range
Zero-pt. wind. Zero-point window ⅓ basic measuring
range
If there are unsteady pressure readings during operation, you can use the para­meters Absorption and Zero-pt. wind. to stabilise the reading and the output signal.
The parameter Absorption functions like a capillary throttle. However, it only acts on the display, output signal and switch points (if these exist) but not on the measuring cell itself.
You can set the response time to pressure jumps in the range 0.0 to 100 s.
Page 16
16 / 32 BA_EN_MS13
NOTICE
Response time
At maximum damping it can take over 2 minutes until the pressure jump from the nominal pressure 100% to 0% is also shown as zero in the display.
In many cases, unsteady readings are not a problem during normal operating mode, but this is not true for the idle state, i.e. if a measured value of zero is ex­pected. The parameter Zero-pt. wind. is designed to solve this. Its value defines a range around zero at which the measured value is set to zero (see fig.).
The display only stops showing zero when the pressure leaves the set window. When twice the window value is reached, the measuring pressure and the dis­play correspond again. This avoids jumps in the display.
Differential pressure
Display value
-x-2x x 2x
Zero-point window
Illustration16: Zero-point window
It makes sense to set the Offset (zero-point displacement) if, without differential pressure (remove measuring line), the display shows a value that is not zero. Before the offset correction, the zero-point window must be set to zero.
Select the Offset corr. parameter and correct the reading using the buttons or
until zero is shown in the display.
When setting the offset, the current measured value is displayed. The zero­point window is not active during the offset setting.
3.3.3 Menu Level Measuring
Parameter name Description Value range
MB start Measuring range start Basic measuring
range
MB end Measuring range end Basic measuring
range
Unit Measuring range unit bar, mbar, Pa, kPa,
MPa, psi, InWc, mmWs, mmHg
Limit Measuring range limit yes, no
The transmitter output signal primarily depends on the sensed pressure. How­ever, you have the option of adjusting the output signal to a large extent to suit your requirements.
Page 17
FISCHER Mess- und Regltechnik GmbH Commissioning | 3
BA_EN_MS13 17 / 32
NOTICE
Adjustment of the output signal
The basic measuring range (indicated on the type label) and the type of output signal (voltage / current) are not variable.
The parameters MB start and MB end initially define the two pressures between which the output signal will change at all. Both values are adjustable across the entire basic measuring range. The set values also refer to the pres­sure in the respective unit. However, the signal values (current / voltage) for ‘Start of measuring range’ and ‘End of measuring range’ are fixed.
If MB start is smaller than MB end, this is called an increasing characteristic curve; the output signal increases as the pressure increases.
If MB end is smaller than MB start, this is a decreasing characteristic curve and the output signal decreases as the pressure increases.
The difference between the values MB start and MB end must be at least 25 % of the basic measuring range.
You can select a unit other than the unit of the basic measuring range with the parameter Unit. The user should remember however that not every unit is suit­able. The conversion is automatic.
The parameter Limit allows the display, output and switching points to be lim­ited to the range between Start of measuring range and End of measuring range. This makes sense when content is measured to avoid "negative con­tents". If Limit is set to "no", those measured values that are greater or smaller than the end values are shown.
3.3.4 Menu Level Output
Parameter name Description Value range
min. output min. output
0.0 … 21.0 mA or
0.0 … 11.0 V
max. output max. output Error signal Measuring range unit
The parameters min. output, max. output and error signal define the limits of the output signal that may not be undercut or exceeded regardless of the pres­sure. The limit values take priority over the range defined by the MB start and
MB end parameters! These parameters primarily serve to prevent error mes-
sages in downstream systems caused by brief overstepping of measuring ranges.
The parameter min. output is usually only used for devices with an output sig­nal 4…20 mA because frequently values of below 3.8 mA are evaluated as er­ror signals.
The max. output value can be used for the voltage and current to limit the max­imum value.
The value defined via the parameter Error signal is issued if the device detects an internal error and can no longer work correctly. It should be noted here that not all potential errors and faults can be detected by the device itself.
Page 18
18 / 32 BA_EN_MS13
3.3.5 Menu Level Function
The Function menu level is a variable menu whose appearance depends on the value of the Function parameter. There are linear, square rooted and table func­tions
Linear function
The input signal is linear before being sent to the display and the output. The range defined in the menu "Measuring" serves as the measuring range. If the function LINEAR is active, the other menu items are cancelled.
Parameter name Description Value range
Function Function Value = linear
Square rooted function
Here, the input signal is square rooted before being sent to the display and the output. This is necessary e.g. for flow measurements with differential pressure. A free unit can be defined for the display. To do this, the start and end of the display range and the number of decimal points are defined. It is also possible to define the unit with 4 characters.
Parameter name Description Value range
Function Function Value = square
rooted
MB decimal pl. Measuring range
decimal places
1234, 123.4, 12.34, 1,234, 12345, 123456
MB start Measuring range start -9999 … +9999 MB end Measuring range end -9999 … +9999 MB unit Measuring range unit 4 characters
The following section contains descriptions of the parameters MB decimal pl.,
MB start, MB end and MB unit to describe the table function.
Tables function
This function allows free adjustment of the input variable to the display and out­put via a table with up to 30 support points. A value pair comprising a measured value and display value is issued for every support point.
NOTICE
Change of parameter
When switching from TABLE to another function, the table is initialised again and the existing values are lost.
Parameter name Description Value range
Function Function Value = Table MB decimal pl. Measuring range
decimal places
1234, 123.4, 12.34, 1,234, 12345, 123456
MB start Measuring range start -9999 … +9999 MB end Measuring range end -9999 … +9999 MB unit Measuring range unit 4 characters No. of pairs Number of pairs n = 3…30 Value pair1 Value pair 1
MB-start … MB-end
Value pair2 Value pair 2 Value pair3 Value pair 3
Page 19
FISCHER Mess- und Regltechnik GmbH Commissioning | 3
BA_EN_MS13 19 / 32
Parameter name Description Value range
∙∙∙
Value pair30 Value pair 30
The display range is defined with the parameters MB decimal pl., MB start and
MB end. The user can select the configuration freely.
Using the parameter MB decimal pl., it is possible to select between a 5 or 6­digit presentation. The resolution is not increased. Only an extra zero or two zeros are added. This serves the correct display of larger values. The measur­ing range must be positive for the 6 digit presentation.
The MB unit gives the user the option of defining a completely independent unit. Letters, numbers or special characters can be used. The unit can be max. 4 characters long.
If the function TABLE is selected, then it is also necessary to state the No. of
pairs. It is defined here how many pairs of values (support points) are used in
the table. A table is made up of at least 3, max. 30 support points.
NOTICE
Number of value pairs
If the number of value pairs is changed, the table is initialised again and the ex­isting values are deleted.
Value pair
2
+14,6 mbar +8,6 %
+0,0 ... +100,0 mbar
1
2
1 input mark (value flashes) 2 allowed range of values
Illustration17: Value pair
Start End
Measuring range
Output / display [%]
100
0
0.0 100.0
Value pair 2
Illustration18: Table function (example)
The individual value pairs can be seen and changed with the Value pair1 to
Value pair30 parameters. A value pair comprises a measured value (left side)
and a display value (right side). The measured value must lie within the measur­ing range and the display value must lie within the defined "free unit". The re­spective limits are shown during input. The table must contain either increasing or decreasing values. the table must contain either continuously increasing or continuously failing values. A change from an increasing to a decreasing char­acteristic curve within a support point table is not allowed.
Page 20
20 / 32 BA_EN_MS13
3.3.6 Menu Level Display
The Display menu level is a variable menu whose appearance depends on the value of the colour parameter. In addition to the various colours for the back­ground lighting, there are also two auto-functions with colour switching avail­able.
Parameter name Description Value range
Colour Colour Off, red, green, yellow,
blue, pink, turquoise, white,
Auto1: Red-green Auto2: Red-yellow­green
Lighting Lighting time 0 s, 10 … 600 s Contrast Contrast 15 … 45 Bar chart Barchart display yes, no
The mots important parameter is Colour. A fixed colour can be defined for the background colour here. There are also two auto-functions with colour switching available. Alternatively, the background illumination can be permanently deac­tivated.
If permanent lighting is not required, the parameter Lighting time can be used to define when the lighting should be switched off after the last time a button is pressed. In addition to permanent lighting (0 s), automatic shut-down after 10… 600 s is also possible. The set time is only valid if the parameter Colour is not set to "off".
Amongst other things, the legibility of the display depends on the temperature and the reading angle. To ensure optimised legibility, the display can be adjus­ted using the parameter Contrast. When the contrast is changed, it is possible that the display appears empty or almost completely black. In this case, the contrast must be turned up or down.
Via the parameter Bar chart, the display can be switched between a display where the measured value is either shown in large digits or the display shows small digits and an additional barchart.
Auto1: Colour-change red to green
In the mode with the automatic colour switchover, it is possible to enter the re­quired switch thresholds "red-green switchover", "green-red switchover".
The switching thresholds can be moved within the measuring range. The series of switch points cannot be altered.
Parameter name Description Value range
Red-Gr. switch. Red-green switching MRS - 50% …
MRE + 50%
Gr-Red switch. Green-red switching Hysteresis Hysteresis 0.1 … 10.0 % Delay Delay 0 … 1800 s Colour Colour Off, red, green, yel-
low, blue, pink, tur­quoise, white,
Auto1: Red-green Auto2: Red-yellow­green
Lighting Lighting time 0 s, 10 … 600 s Contrast Contrast 15 … 45 Bar chart Barchart display yes, no
Page 21
FISCHER Mess- und Regltechnik GmbH Commissioning | 3
BA_EN_MS13 21 / 32
Basic measuring range
ME
F1 F2
MA
Illustration19: Function Auto1
MA MB-start Measuring range start
F1 Red-Gr. switch. Red-green switching F2 Gr-Red switch. Green-red switching
ME MB-end Measuring range end
The parameter Hysteresis can be used to prevent fast and unwanted colour changes. The hysteresis is set in the range 0.1... 10%.
NOTICE
Overlapping colour areas
Note: In the case of large hysteresis values, steps must be taken to ensure that the ranges of the individual colours do not overlap. Otherwise it is possible that the colour change may not function in the desired way.
The parameter Delay offers a further option to prevent unwanted colour changes. The colour change here can be delayed between 0…1800 s.
The parameter Lighting can be used to define when the lighting should be switched off after the last time a button is pressed. In addition to permanent lighting, automatic shut-down after 10…600 s is also possible. The set time is only valid if the parameter Colour is not set to "off". The lighting can be switched on permanently with the value 0s.
The legibility of the display can be adjusted using the parameter Contrast. When the contrast is changed, it is possible that the display appears empty or almost completely black. In this case, the contrast must be turned up or down again.
Via the parameter Bar chart, the display can be switched between a display where the measured value is either shown in large digits or the display shows small digits and an additional barchart.
Auto2: Colour-change red-yellow-green
In the Auto 2 mode with the automatic colour switchover, it is possible to enter the required switch thresholds "red-yellow switchover", "yellow-green switchover", green-yellow switchover, "yellow-red switchover".
The switching thresholds can be moved within the measuring range. The series of switch points cannot be altered.
Parameter name Description Value range
Red-Yell.switch. Red-yellow switchover
MRS - 50% … MRE + 50%
Yell.-Gr.switch. Yellow-green switchover Gr.-Yell. switch Green-yellow switchover Yell.-Red switch Yellow-red switchover Hysteresis Hysteresis 0.1 … 10.0 % Delay Delay 0 … 1800 s Colour Colour Off, red, green, yel-
low, blue, pink, tur­quoise, white,
Page 22
22 / 32 BA_EN_MS13
Parameter name Description Value range
Auto1: Red-green Auto2: Red-yellow­green
Lighting Lighting time 0 s, 10 … 600 s Contrast Contrast 15 … 45 Bar chart Barchart display yes, no
Basic measuring range
MA ME
F1 F2 F3 F4
Illustration20: Function Auto2
MA MB-start Measuring range start
F1 Red-Yell.switch. Colour-change red to yellow F2 Yell.-Gr.switch. Colour-change yellow to green F3 Gr.-Yell. switch Colour-change green to yellow F4 Yell.-Red switch Colour-change yellow to red
ME MB-end Measuring range end
In this menu the same parameters are used as those described in the previous sections.
NOTICE
Unused range
If a range is not to be used, the associated switch thresholds (F1…F4) can be set to the same value.
Example
The parameter Colour is set to Auto2. Only the green, yellow and red ranges are required here. To fade out the lower ranges red and yellow, the switch thresholds "red-yellow switching" and "yellow-green switching" are set to the start of the measuring range.
Basic measuring range
MA ME
F1
F2
F3 F4
Illustration21: Example Auto2
3.3.7 Menu Level System
Parameter name Description Value range
Language Language change DE, EN, FR, ES,
IT,PT,HU
Page 23
FISCHER Mess- und Regltechnik GmbH Commissioning | 3
BA_EN_MS13 23 / 32
Parameter name Description Value range
Software Info Information about the software Device type, serial
number, firmware version
Config Info Information about the configur-
ation
Basic measuring range, output signal, contacts
Key numbers Key numbers Operating time,
switch cycles of the contacts
Password Password 0/1…999 Load config. Load configuration Save config Save configuration
The user menu can be switched to German, English, French, Spanish, Italian, Portuguese or Hungarian using the parameter Language.
The menu items Software - Info and Config - Info provide information about the device. This information helps to answer questions about the device quickly.
• The device type, controller ID and the firmware version is shown in the soft-
ware info.
• The basic measuring range, the defined output signal and existing contacts are stated in the Config Info.
The Statistics provide information about the operating time and the relay switching cycles from the time of delivery. The operating time is shown in days (d) and hours (h)
A Password can be used to protect the menu against unauthorised access. The password is a figure from 1 to 999. The input 0 means that no password is active.
The password needs to be set if the user presses the button in normal mode to enter the menu. If a wrong password is entered, the system automatically jumps back to normal mode again. If no password is active, the display immediately jumps to the menu.
NOTICE
Forgotten password
The user cannot restore a forgotten password. Please contact the manufacturer in this case.
The user can load a saved configuration via the menu item Load config. This means that a functional set of parameters can be loaded after trying out various settings.
The menu item Save config. serves to save the existing parameters in a pro­tected memory area. This is helpful if the settings of a functional device needs to be optimised. Save config. and Load config. can be used to quickly restore the initial status again.
NOTICE
Delivery condition
If the user has not yet saved a configuration, the default values (status on deliv­ery) are loaded. In this case, any measuring range spreads or switch points are reset and the device needs to be newly configured.
Page 24
24 / 32 BA_EN_MS13
4 Technical Data
4.1 General Information
Reference conditions (acc. to IEC 61298-1)
Temperature error +15 … +25 °C Relative humidity 45 … 75 % Air pressure 86 … 106 kPa 860 … 1060 mbar Installation position User-defined
4.2 Input variables
Relative pressure
Measuring range
Pressure safety Characteristic curve devi-
ation
Overpressure Bursting pres-
sure
Option Default
0…+1.6 bar 4 bar (56 PSI) 7 bar 0.5%FS 1.0 % FS 0…+2.5 bar 10 bar 15 bar 0.5%FS 1.0 % FS 0…+4 bar 10 bar 15 bar 0.5%FS 1.0 % FS 0…+6 bar 20 bar 35 bar 0.5%FS 1.0 % FS 0…+10 bar 40 bar 70 bar 0.5%FS 1.0 % FS 0…+16 bar 40 bar 70 bar 0.5%FS 1.0 % FS 0…+25 bar 100 bar 150 bar --- 1.0 % FS 0…+40 bar 100 bar 150 bar --- 1.0 % FS 0…+60 bar 200 bar 250 bar --- 1.0 % FS
Vacuum and ± measuring ranges
Measuring range
Pressure safety Characteristic curve devi-
ation
Overpressure Bursting pres-
sure
Option Default
0…-1 bar 4 bar (56 PSI) 7 bar --- 1.0 % FS
-1…0 bar 4 bar (56 PSI) 7 bar --- 1.0 % FS
-1…+0.6 bar 4 bar (56 PSI) 7 bar --- 1.0 % FS
-1…+1.5 bar 4 bar (56 PSI) 7 bar --- 1.0 % FS
-1…+3 bar 10 bar 15 bar --- 1.0 % FS
-1…+5 bar 20 bar 35 bar --- 1.0 % FS
-1…+9 bar 40 bar 70 bar --- 1.0 % FS
-1…+15 bar 40 bar 70 bar --- 1.0 % FS
-1…+24 bar 100 bar 150 bar --- 1.0 % FS
4.3 Output parameters
Analogue output
Output signal Signal range Apparent ohmic resistance
0…20 mA 0.0…21.0 mA RL ≤ 600 Ω 4…20 mA 0…10 V 0.0…11.0 V RL ≥ 2 kΩ
Page 25
FISCHER Mess- und Regltechnik GmbH Technical Data | 4
BA_EN_MS13 25 / 32
Switching outputs
2 potential-free relay contacts or 2 potential-free semiconductor switches (MOSFET)
Relay MOSFET
Progr. switching function Open contact (NO)
Break contact (NC)
One-pin activator (NO)
One-pin deactivator (NC) Max. switching voltage 32 V AC/DC 3…32 V AC/DC Max. switching current 2 A 0.25 A Max. switching output 60 W (VA) 8 W / 8 VA
RON ≤ 1 Ω
4.4 Measurement accuracy
Non-linearity Default < 1.0 % FS
Option
1)
< 0.5 % FS Hysteresis < 0.5 % FS Characteristic curve deviation
2)
Default 1.0 % Option
1)
0.5 %
Temperature drift Zero point 0.07 % FS/K
Measuring range 0.05 % FS/K
1)
only possible for certain measuring ranges
2)
incl. non-linearity and hysteresis
4.5 Auxiliary energy
Rated Voltage 24 V AC/DC Admissible operating voltage Ub = 12…32 V DC
Ub = 16.5…32 V AC
Electrical connection 5-pin M12 circular plug connector
4.6 Application conditions
Ambient conditions
Increase ambient temper­ature
-10 … +70 °C
Media temperature -10 … +70 °C Storage temperature -20 … +70 °C Enclosure protection class IP 65 acc. to EN 60529 EMC EN 61326-1:2013
EN 61326-2-3:2013
RoHS EN 50581:2012
Materials of the parts that come into contact with the surroundings
Housing Stainless steel 1.4301 Bayonet ring Stainless steel 1.4301 Front plate Aluminium Front film PET Polyethylene terephthalate Bayonet ring seal NBR Nitrile rubber Blowout plug FKM Fluorocarbon rubber
Materials of the parts that come into contact with the measuring medium
Process connection Stainless steel 1.4404 Measuring diaphragm ceramic Seal FKM Fluorocarbon rubber
Page 26
26 / 32 BA_EN_MS13
4.7 Display and operating interface
Advertisement
4...6-digit LCD, full graphic, colour backlighting
Programming
Attenuation 0.0…100.0s (jump response 10/90%) Switch output Switch-off point, switch-on point, delay (0...1800s),
function (NC / NO contact)
Measuring range unit bar, PSI, kPa, "free unit"È, starting value, end
value and decimal point for "free unit"
Output signal User-definable within the basic measuring range
(1)
Zero-point stabilising 0…⅓ of the basic measuring range
(2)
Zero point correction (off­set)
±⅓ of the basic measuring range
(3)
Implementation of charac­teristic curve
linear, square rooted, table with 3...30 support points
Password 001 ... 999 (000 = no password protection)
(1) Max. effective spread 4:1 (2) measured values around zero are set to zero. (3) To compensate different installation positions.
4.8 Construction design
4.8.1 Dimensional picture
M12 plug 1 M12 plug 2
Blowout plug
4
68
203
Process connection
Ø100
51.6
40
Ø6
Illustration22: Dimensional drawing
Page 27
FISCHER Mess- und Regltechnik GmbH Technical Data | 4
BA_EN_MS13 27 / 32
4.8.2 Process connection
G½B G¼B ¼-18 NPT EXT Schrader
®
20
Code 87 85 88 S1
13
20
7.5
3
2
Ø6
Ø5
Ø2
Ø17.5
Ø9.5 ¼-18 NPT
7/16 UNF
Illustration23: Process connection
Port Material
G½ B Connection shanks with external thread 1.4404
G¼ B Connection shanks with external thread 1.4404
¼-18 NPT EXT Connection shanks with external thread 1.4404
7/16 UNF Connection with inner thread for the
Schrader®- screw connection >
1.4404
Page 28
28 / 32 BA_EN_MS13
5 Order Codes
Process connection
Output signal
operating voltage
Electrical connection
Measuring System / Design
M S 1 3
Type
Measuring range
0
1 2 5 6 7 8 9 10 11 123 4Code no.
Measurement accuracy
L V
Process connection direction
[1.2] Measuring range converted ranges
[bar] [kPa] [PSI]
03 0…1.6 bar 0 … 160 kPa 0 … 23.21 PSI 04 0…2.5 bar 0 … 250 kPa 0 … 36.26 PSI 05 0…4 bar 0 … 400 kPa 0 … 58.01 PSI 06 0…6 bar 0 … 600 kPa 0 … 87.02 PSI 07 0…10 bar 0 … 1000 kPa 0 … 145.0 PSI 08 0…16 bar 0 … 1600 kPa 0 … 232.1 PSI 09 0…25 bar 0 … 2500 kPa 0 … 362.6 PSI 10 0…40 bar 0 … 4000 kPa 0 … 580.1 PSI 11 0…60 bar 0 … 6000 kPa 0 … 870.2 PSI 31 -1…0 bar -100 … 0 kPa -14.50 … 0 PSI 32 -1…0.6 bar -100 … 60 kPa -14.50 … 8.702 PSI 33 -1…1.5 bar -100 … 150 kPa -14.50 … 21.75 PSI 34 -1…3 bar -100 … 300 kPa -14.50 … 43.51 PSI 35 -1…5 bar -100 … 500 kPa -14.50 … 72.52 PSI 36 -1…9 bar -100 … 900 kPa -14.50 … 130.5 PSI 37 -1…15 bar -100 … 15000 kPa -14.50 … 217.5 PSI 38 -1…24 bar -100 … 2400 kPa -14.50 … 348.1 PSI 39 0…-1 bar 0 … -100 kPa 0 … -14.50 PSI
[3] Measurement accuracy
M 1.0 % characteristic curve deviation
0 0.5 % characteristic curve deviation
[4.5] Process connection Material
85 Connection shanks with external thread G¼ B
1.4404
87 Connection shanks with external thread G½ B 88 Connecting port with outer thread ¼ -18 NPT EXT S1 Schrader® screw connection inner thread 7/16 UNF
Page 29
FISCHER Mess- und Regltechnik GmbH Order Codes | 5
BA_EN_MS13 29 / 32
[6.7] Output signal Operating voltage
EL 0 … 20 mA 24 V AC/DC 3-wire version PL 4 … 20 mA 24 V AC/DC 3-wire version CL 0 … 10 V 24 V AC/DC 3-wire version
[8] Measuring value display/switching elements
C 4...6-digit colour change LCD 2 relay contacts D 4...6-digit colour change LCD 2 semiconductor switches
[9] Process connection direction
L Port left R Port right 0 Port on top U Port below
[10] Electrical connection
L M12 connector socket Ms-nickel-plated
[11] Measuring System / Design
V FKM Fluororubber, Viton®
5.1 Accessories
Order no. Designation No. of
Poles
Length
06401993 Connection cable for switching outputs
with M12 connector
4-pin 2 m
06401994 Connection cable for switching outputs
with M12 connector
4-pin 5m
06401995 Connection cable for supply/signal
with M12 connector
5-pin 2 m
06401996 Connection cable for supply/signal
with M12 connector
5-pin 5 m
Accessories acc. to data sheet MZ
*)
MZ310# Wall mount acc. to DIN 16281 MZ1### Siphons MZ400# Capillary throttle coil MZ5### Manometer shutoff valve acc. to DIN 16270/16271 MZ6### Manometer shutoff valve acc. to DIN 16272
Parameter setting adapter
*)
EU05 Transmitter PC Interface incl. PC software
*)
A data sheet is available on our website or on request..
Page 30
30 / 32 BA_EN_MS13
6 Attachments
Illustration24: CE_DE_MS13
Page 31
FISCHER Mess- und Regltechnik GmbH Attachments | 6
BA_EN_MS13 31 / 32
Page 32
32 / 32 BA_EN_MS13
Loading...