Fieldpiece SC77 User Manual

Maintenance
Clean the exterior with clean dry
cloth. Do not use liquid.
Battery replacement: When the
multimeter displays " " the battery must be replaced. Disconnect and unplug leads, turn meter off, and remove the battery cover. Replace the battery with a NEDA type 1604 9V bat­tery.
Symbols used:
Caution, risk of electric shock
Caution, refer to manual.
Ground
Double insulation
For your safety...
General: Disconnect the test leads
All voltage tests: For model SC77
all voltage ranges will withstand up to 600VDC or 600VAC rms. Do not apply more than 600VDC or 600VAC rms.
AC tests: Disconnect the meter
from the circuit before turning any inductor off, including motors, trans­formers, and solenoids. High voltage transients can damage the meter beyond repair. Do not use during elec­trical storms.
Limited warranty
This meter is warranted against defects in material or workmanship for one years from date of purchase. Fieldpiece will replace or repair the defective unit, at its option, subject to verification of the defect.
This warranty does not apply to defects resulting from abuse, neglect, accident, unauthorized repair, alter­ation, or unreasonable use of the instrument.
Any implied warranties arising from the sale of a Fieldpiece product, including but not limited to implied war­ranties of merchantability and fitness for a particular purpose, are limited to the above. Fieldpiece shall not be liable for loss of use of the instrument or other incidental or consequential damages, expenses, or economic loss, or for any claim of such damage, expenses, or economic loss.
State laws vary. The above limita­tions or exclusions may not apply to you.
Obtaining service
Return any defective SC77 to Fieldpiece for service. For warranty service, please provide a proof of pur­chase. Contact Fieldpiece for out of warranty repair charges. Contact infor­mation can be found at our website www.fieldpiece.com.
OPERATOR’S MANUAL
400
DISCONNECT TESTLEADS
340
SC77
µ
Auto
4000m
400m
HOLD
NCV
m A K M Ω
m V V
APO HOLD n m m F AC DC
Works with Fieldpiece accessory heads!
AUTO RANGING T-RMS DIGITAL CLAMP METER
Model SC77
WARNINGS
DISCONNECT AND UNPLUG TEST
LEADS
before opening case.
TEST NCV FUNCTION ON KNOWN
LIVE WIRE before using.
DO NOT APPLY VOLTAGE greater than 30VAC or 60VDC to the thermocou­ple or the jacks when the rotary dial is on °F.
REMOVE
THE THERMOCOUPLE
when taking voltage measurements.
DISCONNECT
THE TEST LEADS when taking temperature measure­ments.
DO NOT APPLY VOLTAGE TO THE JACKS when the rotary dial is on microamps. Even low voltages can cause a current overload and potentially dam­age the circuit.
!
!
Works with Fieldpiece accessory heads
Connect your Fieldpiece accessory head to SC series meter through included deluxe silicone leads. Use 400mVDC or 4000mVDC range for most heads.
SC77 connected to
ARH4 through leads.
Available Fieldpiece accessory heads
There is a Fieldpiece accessory head available for just about any job. There are heads to measure tempera­ture, superheat, subcooling, RH%, wet bulb, dew point, vacuum (microns of mercury), manometer (inches of water column), amps AC&DC, high voltage,
CO, CO2, air velocity, and many more.
Non-contact voltage
With the NCV tab on the tip of the clamp close to an AC voltage, press the NCV button. The NCV LED will light and the beeper will beep. The NCV function is sensitive enough to detect 24VAC on thermostats.
Hi voltage indicator
In any VAC/VDC range, when you touch a voltage greater than 30V, the beeper will beep and the red Hi-V LED will blink. BE CAREFUL!
Current measurements
The fixed jaw is longer than the moveable jaw to make it easier to select just one wire from a bundle. With the jaws closed, separate one wire using the long fixed jaw. Slide it to the corner where the two jaws meet. Then open the jaws to let it in. You can select the wire without having to hold the jaws open.
Microamps
For measuring the flame diode cur­rent in a heater control.
Capacitance
For motor-start and motor-run capacitors. Disconnect the capacitor from power first. Short the terminals to discharge the capacitors. Disconnect any resistors that might be between the terminals of the capacitor. The SC77 displays a “dsc” symbol when capacitor needs discharging.
Temperature
Plug any K-type thermocouple directly into the meter to measure tem­perature. Temperature measurement will maintain good accuracy in fast changing environments. One thermo­couple is included. No adapter is required.
Field Temp. calibration
For accuracies of ±1° calibrate the SC77 to a known temperature. A glass of stabilized ice water is very close to 32°F (0°C) and is usually very conven­ient.
1. Select the °F range on the SC77.
2. Remove back case. Hold the bat-
tery in place with a rubber band.
3. Stabilize a large cup of ice water by
stirring.
4. Immerse the probe and let it stabi-
lize. Continue stirring.
5. Adjust VR3 (lower right corner of
PCB) to get close to 32°F (0°C) then adjust VR4 (below VR3) to get within 0.1°F/C of 32°F (0°C).
6. To calibrate in °C, close the jumper
that is to the left of VR3.
Display ºC or ºF
Remove back of meter, locate jumper on lower right corner of PCB (just below VR3) and close jumper to read temperature in °C.
Using &storing test leads
Because the wire insulation is sili­cone they will stay flexible in cold weather and will not melt if bumped by a soldering iron.
Use the single test probe holder on clamp to make voltage testing easy.
For convenient lead storage, wrap the leads as shown.
!
SC77
APO HOLD n m m F AC DC
m A K M Ω
m V V
NCV
HOLD
Auto 4000m
µ
400m
400
F
C
C
400
F
1000
TEMP
DISCONNECT TESTLEADS
SPECIFICATIONS
Display: LCD with a maximum read-
ing of 4000.
Data Hold: Operating on all ranges Range(RNG): Operating on all autor-
anges
Overrange: "OL" mark indication. Auto power off: Approx. 10 minutes. Measurement rate: 2 times per sec-
ond, nominal.
Operating environment: 32 to 122°F
(0 to 50°C) at <70% R.H.
Storage temperature: -4 to 140°F
(-20 to 60°C), 0 to 80% R.H. with battery removed.
Accuracy: Specifications good in
ambient conditions of 73°F ±9°F (23°C ±5°) , <75% relative humidity.
Temperature Coefficient: 0.1 ×
(specified accuracy) per °F/°C; 32 to 66°F (0 to 19°C); 82 to 122°F (28 to 50°C).
Altitude: 6561.7 Feet (2000 meters). Jaw opening capability: 38mm con-
ductor.
Power: Single standard 9-volt battery,
NEDA 1604, JIS 006P, IEC 6F22.
Battery life: 100 hours carbon zinc
200 hours typical with alkaline.
Accessories: One pair test leads,
one pair of small alligator clips, k­type thermocouple, 9V battery (installed), and operating instruc­tions.
Safety: Designed to meet
IEC/EN61010-1 and IEC/EN61010­2-032, CATIII 600V, class II and pollution degree 2, indoor use
comply with CE. C-Tick.
DC volts
Ranges: 400mV, 4000mV, 40V, 400V,
600V
Resolution: 0.1mV Accuracy: ±(0.5% rdg + 2 dgt) Input impedance: 3MΩ on V inputs,
10MΩ on mV input
Overload protection: 600VDC or AC
rms
Transient protection: 6kV for 10μ
sec
AC volts TrueRMS (50-500Hz)
Ranges: 4V, 40V, 400V, 600V Resolution: 1mV Accuracy: ±(2.0% rdg + 8 dgts) 50 ~
500Hz
Input impedance: 3MΩ Overload protection: 600V DC or AC
rms
Transient protection: 6kV for 10μ
sec
AC current TrueRMS (50-60Hz)
Ranges: 40A, 300A Resolution: 0.01A DC Accuracy:
±(3.0% rdg + 10 dgts) 0 ~ 300A / 50 ~ 60Hz
Overload protection: 300AAC
DC current
Ranges: 400μA Resolution: 0.1μA Accuracy: ±(1.0% rdg + 2 dgt) Voltage burden: 1V Overload protection: 500V DC or AC
rms
Resistance
Ranges: 400Ω, 4kΩ, 40kΩ, 400kΩ,
4MΩ, 40MΩ
Resolution: 0.1Ω Accuracy:
±(1.2% rdg + 4 dgts) on 400Ω to 400kΩ ranges ±(1.5% rdg + 4 dgts) on 4MΩ range ±(3.0% rdg + 5 dgts) on 40MΩ range
Open circuit volts: -0.45VDC typical,
(-1.2VDC on 400Ω range)
Overload protection: 500VDC or AC
rms
Temperature
Range: -30 to 1000°F (-34 to 538°C) Resolution: 0.1° Accuracy:
±1°F, 32 to 120°F (0 to 49°C) ±1%+1.5°F, -4 to 750°F (-20 to 399°C) ±2%+4°F, -30 to -4°F (-34 to -20°C) ±2%+4°F, 750 to 1000°F(399 to
538°C)
Sensor type: K-type thermocouple Overload protection: 60 VDC or 30
VAC rms
Continuity
Range: 400Ω Resolution: 1Ω Audible indication: Less than 25Ω Response time: 500ms Overload Protection: 500VDC or AC
rms
Green LED will be on continuously.
Capacitance
Range: 4μF, 4 0 μF, 400μF, 4mF 1mF = 1000μF i.e: 1.425mF = 1425μF
Resolution: 1nF Accuracy: ±(3% rdg + 10 dgts) on
4μF range
±(3% rdg + 5 dgts) on 40μF to
400μF ranges
±(5% rdg + 20 dgts) on 4mF range Input protection: 500VDC or AC rms
Hi-V(voltage) warning
>30VAC/DC meter beeps, red LED blinks.
Non-contact voltage
Senses voltage 24V to 600V AC (50Hz to 60Hz) beeper chirps and big bright red LED comes on, works when meter dial is on any range.
Voltage
Resistance
DC Current<4000μADC
Temperature
AC Current
Selecting ranges
For DC voltage, set the meter to the VDC parameter instead of VAC as shown above.
For all ranges and functions choose a range just above the value you expect. If display reads "OL" (over­load), select a higher range. If display shows less than three numbers, select a lower range for better resolution.
CE
This instrument complies with the requirements of the following European Community Directives: 89/336/EEC (Electromagnetic Compatibility) and 73/23/EEC (Low Voltage) as amended by 93/68/EEC (CE Marking).
Capacitance
True RMS
Digital multimeters use two different types of AC sensing. The most com­mon is average sensing, normalized to a true RMS value of a sine wave. The other is true RMS sensing. The actual true RMS value is sensed for a wave form within the limits of the crest factor. Either sensing method will give the same results on a clean sine wave but they may differ on a non-sinusoidal waveform.
400
1000
µ
Auto
4000m
400m
F
C
TEMP
400
F
C
Insure the temperature being meas­ured is stable. Maintain good contact between the thermocouple and what's being meas­ured.
400
1000
µ
Auto
4000m
400m
F
C
TEMP
400
F
C
Disconnect test leads from
voltage before plugging
thermocouple in.
Will display “dsc” if capacitor needs discharge.
v16
µ
400
F
C
400
1000
TEMP
Auto
400
F
C
TEMP
1000
4000m
400m
C
400
F
µ
F
400
C
TEMP
Auto
400m
C
400
F
1000
µ
Auto
4000m
400m
C
F
4000m
Auto
400
F
C
400
1000
TEMP
4000m
400m
C
F
µ
Loading...