Fieldpiece SC76 User Manual

OPERATOR’S MANUAL
Works with Fieldpiece accessory heads!
AUTO RANGING DIGITAL CLAMP METER Model SC76
WARNINGS
DISCONNECT AND UNPLUG TEST
before opening case.
TEST NCV FUNCTION ON KNOWN
LIVE WIRE before using.
DO NOT APPLY VOLTAGE greater than 30VAC or 60VDC to the thermocou­ple or the jacks when the rotary dial is on °F.
REMOVE
THE THERMOCOUPLE
when taking voltage measurements.
DISCONNECT
THE TEST LEADS when taking temperature measure­ments.
DO NOT APPLY VOLTAGE TO THE JACKS when the rotary dial is on microamps. Even low voltages can cause a current overload and blow the fuse. Replace blown fuse to regain function.
!
Maintenance
Clean the exterior with clean dry
cloth. Do not use liquid.
Battery replacement: When the multimeter displays " " the battery must be replaced. Disconnect and unplug leads, turn meter off, and remove the battery cover. Replace the battery with a NEDA type 1604 9V bat­tery.
For your safety...
General: Disconnect the test leads before opening the case. Inspect the test leads for damage to the insulation or exposed metal. Replace if suspect. Never ground yourself when taking electrical measurements. Do not touch exposed metal pipes, outlets, fixtures, etc., which might be at ground poten­tial. Keep your body isolated from ground by using dry clothing, rubber shoes, rubber mats, or any approved insulating material. When disconnect­ing from a circuit, disconnect the "RED" lead first, then the common lead. Work with others. Use one hand for testing. Turn off power to the circuit under test before cutting, unsoldering, or breaking the circuit. Keep your fin­gers behind the finger guards on the probes. Do not measure resistance when circuit is powered. Do not apply more than rated voltage between input and ground.
All voltage tests: For model SC76 all voltage ranges will withstand up to 600VDC or 600VAC. Do not apply more than 600VDC or 600VAC.
AC tests: Disconnect the meter from the circuit before turning any inductor off, including motors, trans­formers, and solenoids. High voltage transients can damage the meter beyond repair. Do not use during elec­trical storms.
Symbols used:
Caution, risk of electric shock
Caution, refer to manual.
Ground
Double insulation
Limited warranty
This meter is warranted against defects in material or workmanship for one year from date of purchase. Fieldpiece will replace or repair the defective unit, at its option, subject to verification of the defect.
This warranty does not apply to defects resulting from abuse, neglect, accident, unauthorized repair, alter­ation, or unreasonable use of the instrument.
Any implied warranties arising from the sale of a Fieldpiece product, including but not limited to implied war­ranties of merchantability and fitness for a particular purpose, are limited to the above. Fieldpiece shall not be liable for loss of use of the instrument or other incidental or consequential damages, expenses, or economic loss, or for any claim of such damage, expenses, or economic loss.
State laws vary. The above limita­tions or exclusions may not apply to you.
!
Works with Fieldpiece accessory heads
Connect your Fieldpiece accessory head to SC series meter through included deluxe silicone leads. Use 340mVDC or 3400mVDC range for most heads.
SC76 connected to
ARH4 through leads.
Available Fieldpiece accessory heads
There is a Fieldpiece accessory head available for just about any job. There are heads to measure tempera­ture, RH%, wet bulb, dew point, vacu­um (microns of mercury), manometer (inches of water column), amps AC&DC, high voltage, CO, CO2, air
velocity, and many more.
Non-contact voltage
With the NCV tab on the tip of the clamp close to an AC voltage, press the NCV button. The NCV LED will light and the beeper will beep. The NCV function is sensitive enough to detect 24VAC on thermostats.
Hi voltage indicator
In any VAC/VDC range, when you touch a voltage greater than 30V, the beeper will beep and the red Hi-V LED will blink. BE CAREFUL!
Current measurements
The fixed jaw is longer than the moveable jaw to make it easier to select just one wire from a bundle. With the jaws closed, separate one wire using the long fixed jaw. Slide it to the corner where the two jaws meet. Then open the jaws to let it in. You can select the wire without having to hold the jaws open.
Microamps
For measuring the flame diode cur­rent in a heater control.
Capacitance
For motor-start and motor-run capacitors. Disconnect the capacitor from power first. Short the terminals to discharge the capacitors. Disconnect any resistors that might be between the terminals of the capacitor.
Temperature
Plug any K-type thermocouple directly into the meter to measure tem­perature. Temperature measurement will maintain good accuracy in fast changing environments. One thermo­couple is included. No adapter is required.
Display ºC or ºF
Remove back of meter, locate jumper on lower right corner of PCB (just below VR3) and close jumper to read temperature in °C.
Field Temp. calibration
For accuracies of ±1° calibrate the SC76 to a known temperature. A glass of stabilized ice water is very close to 32°F (0°C) and is usually very conven­ient.
1. Select the °F range on the SC76.
2. Remove back case. Hold the bat-
tery in place with a rubber band.
3. Stabilize a large cup of ice water.
4. Immerse the probe and let it stabi-
lize.
5. Adjust VR3 (lower right corner of
PCB) to get close to 32°F (0°C) then adjust VR4 (below VR3) to get within 0.1°F/C of 32°F (0°C).
6. To calibrate in °C, close the jumper that is to the left of VR3.
Using & storing test leads
Because the wire insulation is sili-
cone they will stay flexible in cold weather and will not melt if bumped by a soldering iron.
Use the single test probe holder on
clamp to make voltage testing easy.
For convenient lead storage, wrap
the leads as shown.
!
v16
SPECIFICATIONS
Display: LCD with a maximum read-
ing of 3400.
Analog bar graph: 34 segments with
measurements 20 times per sec­ond.
Data Hold: Operating on all ranges Range(RNG): Operating on all autor-
anges
Overrange: "OL" mark indication. Auto power off: Approx. 10 minutes. Measurement rate: 2 times per sec-
ond, nominal.
Operating environment: 32 to 122°F
(0 to 50°C) at <70% R.H.
Storage temperature: -4 to 140°F
(-20 to 60°C), 0 to 80% R.H. with battery removed.
Accuracy: Specifications good in
ambient conditions of 73°F ±9°F (23°C ±5°) , <75% relative humidity.
Temperature Coefficient: 0.1×(speci-
fied accuracy) per °F/°C; 32 to 64°F (0 to 19°C); 82 to 122°F (28 to 50°C).
Altitude: 6561.7 Feet (2000m). Jaw opening capability: 38mm con-
ductor.
Power: Single standard 9-volt battery,
NEDA 1604, JIS 006P, IEC 6F22.
Battery life: 300 hours typical with
alkaline.
Accessories: One pair test leads,
one pair of small alligator clips, k­type thermocouple, 9V battery (installed), and operating instruc­tions.
Safety: UL, CE, Cat III 600V, IEC
61010-1 (EN61010-1) and IEC61010-2-032 (EN61010-2-032).
DC volts
Ranges: 340mV, 3400mV, 34V, 340V,
600V
Resolution: 0.1mV Accuracy: ±(0.5% rdg + 1 dgt) Input impedance: 3MΩ on V inputs,
10MΩ on mV input
Overload protection: 600VDC or AC
rms
Normal mode rejection ratio: >50dB
@ >49Hz
Common mode rejection ratio:
>120dB up to 1000VDC
Transient protection: 6kV for 10μ
sec
AC volts(50Hz - 500Hz)
Ranges: 3.4V, 34V, 340V, 600V Resolution: 1mV Accuracy: ±(2.0% rdg + 4 dgts) 50 ~
500Hz
Input impedance: 3MΩ Overload protection: 600V DC or AC
rms
Transient protection: 6kV for 10μ
sec
AC current(50Hz - 60Hz)
Ranges: 34A, 300A Resolution: 0.01A DC Accuracy:
±(3.0% rdg + 5 dgts) 0 ~ 300A / 50 ~ 60Hz
Overload protection: 300AAC
DC current
Ranges: 340μA, 3400μA Resolution: 0.1μA Accuracy: ±(1.0% rdg + 1 dgt) Voltage burden: 1V, (8V on 3400μA
range)
Overload protection: 500VDC or AC
rms
Resistance
Ranges: 340Ω, 3.4kΩ, 34kΩ, 340kΩ,
3.4MΩ, 34MΩ
Resolution: 100mΩ Accuracy:
±(1.2% rdg + 4 dgts) on 340Ω to 340kΩ ranges ±(1.5% rdg + 4 dgts) on 3.4MΩ range ±(3.0% rdg + 5 dgts) on 34MΩ range
Open circuit volts: -0.45VDC typical,
(-1.2VDC on 340Ω range)
Overload protection: 500VDC or AC
rms
Temperature
Range: -30 to 1000°F (-35 to 600°C) Resolution: 0.1° Accuracy: ±1°F (32 to 120°F); ±0.5°C
(0 to 49°C),
±1% + 1.5°F (-4 to 750°F); ±1% +
1°C (-20 to 400°C),
±2% + 4°F (-30 to -4°F, 750 to 1000°F); ±2% +2°C (-35 to -20, 400 to 540°C)
Sensor type: K-type thermocouple Overload protection: 60 VDC or 30
VAC rms
Diode test
Test current: Approx. 1.0mA Resolution: 1mV Accuracy: ±(2% rdg + 3 dgts) Open circuit volts: 3.0Vdc typical Audible indication: Less than 0.2V Overload protection: 500VDC or AC
rms
Continuity
Audible indication: Less than 300Ω Response time: 500ms Green LED will be on continuously.
Capacitance
Range: 340μF Resolution: 0.1μF Accuracy: ±(3% rdg + 5 dgts) Test frequency: 16Hz Test voltage: <3.0V Overload protection: 500VDC or AC
rms
Hi-V(voltage) warning
>30VAC/DC meter beeps, red LED
blinks.
Non-contact voltage
Senses voltage 24V to 600V AC beeper chirps and big bright red LED comes on, works when meter dial is on any range.
Voltage
Resistance
DC Current (μADC)
Temperature
AC Current
Selecting ranges
For DC voltage, set the meter to the VDC parameter instead of VAC as shown above.
For all ranges and functions choose a range just above the value you expect. If display reads "OL" (over­load), select a higher range. If display shows less than three numbers, select a lower range for better resolution.
Capacitance
CE
This instrument complies with the requirements of the following European Community Directives: 89/336/EEC (Electromagnetic Compatibility) and 73/23/EEC (Low Voltage) as amended by 93/68/EEC (CE Marking).
Service
Return any defective SC76 to Fieldpiece for warranty service along with proof of purchase. Contact Fieldpiece for out of warranty repair charges.
www.fieldpiece.com
Insure the temperature being measured is stable. Maintain good contact between the thermocouple and what's being measured.
Disconnect test leads from
voltage before plugging
thermocouple in.
Loading...