Fieldpiece SC57 Operator's Manual

01
06
02
07
03
08
04
09
05
10
Quick Start
Receive a Wireless Reading
1. Select RECV on main dial of SC57.
2. Hold SYNC for one second.
3. Connect Fieldpiece accessory head to ET2W.
3. Turn ET2W to DC switch position.
4. Hold SYNC on ET2W for one second. When connection is made the wireless reading and
icons will display on the SC57 screen.
Certifications
UL 61010-1, Second Edition
IEC/EN61010-1, EMC EN61326-1 IEC/EN61010-2-032
FCC ID: VEARF915A
C-Tick (N22675)
WEEE
CATIII 600V, class II and pollution degree 2
indoor use comply with CE, RoHS compliant.
Description
As a proud owner of the first wireless Fieldpiece clamp meter, SC57, you are well on your way to eliminating tangled wires. SC57 will give you the power to receive measurements wirelessly from anywhere on the jobsite. For instance, you can transmit indoor wet bulb measurements for target superheat over the air while you work at the condenser.
Your SC57 can receive measurements wirelessly from any Fieldpiece accessory head connected to a Fieldpiece wireless transmitter like an ET2W or EH4W. Your SC57 comes with one ET2W wireless transmitter.
In addition, your SC57 can transmit wirelessly any measurement range on the dial to Fieldpiece wireless receivers; like EH4W, HG3, LT17AW, or another SC57.
SC57 is a swivel clamp meter designed for the HVAC/R technician. The swivel on the SC57 allows you to see the amperage reading, regardless of wire orientation. The jaw light and backlight
O F F
NCV
VAC
VDC
AAC
Hz
SC56
SEL
HOLD
INRUSH
make it easier to use in any lighting condition. True RMS helps you take more accurate voltage readings on variable speed drives.
Controls
Rotate dial to the function you want to use.
Hold for 1 second to connect with
other Fieldpiece wireless products
Cycle through parameters or ranges
within a dial position
Press to illuminate backlight for 3 min
Press to freeze current, maximum or
minimum reading
Fieldpiece
Wireless Swivel Clamp Meter
OPERATOR'S MANUAL
Model SC57
Display
SEL
HOLD
HOLD
MAX
MIN
SYNC
MAX
MIN
Wireless Icon
HOLD
HOLD
MAX
MIN
SYNC
MAX
MIN
Signal Strength Bars and Search Pattern indicator when sending a wireless measurement
SC57
SC57
O F F
WIRELESS ENABLED
SEL
HOLD
HOLD
MAX MIN
SYNC
MAX
MIN
Battery Life Indicator
O F F
M
FD
NCV
VAC
VDC
AAC
Hz
Hz
%
°C
SC56
SEL
HOLD
INRUSH
Auto Power O Enabled
O F F
M
FD
NCV
VAC
VDC
AAC
Hz
Hz
%
°C
SC56
SEL
HOLD
INRUSH
High Voltage Warning (+30V )
O
F F
SEL
HOLD
HOLD
MAX
MIN
SYNC
MAX
MIN
Data Hold Mode
SC57
SC57
O F F
WIRELESS ENABLED
SEL
HOLD
HOLD
MAX
MIN
SYNC
MAX
MIN
Maximum Reading
SC57
SC57
O
F F
SEL
HOLD
HOLD
MAX
MIN
SYNC
MAX
MIN
Minimum Reading
SEL
HOLD
INRUSH
Continuity Test
HOLD
INRUSH
Frequency Test (hertz)
NCV
SEL
HOLD
INRUSH
Resistance Test (ohms)
O
F F
NCV
VAC
VDC
AAC
Hz
SEL
HOLD
INRUSH
Capacitance Test (farads)
HOLD
INRUSH
Micro Unit (10−6, one millionth)
SEL
HOLD
HOLD
MAX
MIN
SYNC
MAX
MIN
M illi Unit (10−3, one thousandth)
Specifications
Wireless range: Up to 75 feet (23m) line of sight Minimum wireless distance: 1 foot (30cm) Wireless frequency: 910MHz to 920MHz (US), 868.1MHz to
868.5MHz (European)
Display: 10000 count display with backlight Overrange: (OL) or (-OL) is displayed Measurement rate: 2 times per second, nominal Zero: Automatic Operating environment: 32°F to 122°F (0°C to 50°C) at <70%
relative humidity
Storage temperature: -4°F to 140°F (-20°C to 60°C), 0 to 80% RH
(with battery removed)
Accuracy: Stated accuracy @ 73°F±9°F (23°C±5°C), <75%RH Temperature coecient: 0.1 x (specied accuracy) per °C [0°C to
18°C (32°F to 64°F), 28°C to 50° C(82°F to 122°F)]
Accuracy Specication: Calculate against a known value APO (Auto Power O): Approx. 30 minutes Power: Single standard 9-volt battery, NEDA 1604, JIS 006P, IEC 6F22 Battery life: 100 hours typical alkaline Low battery indication: Battery icon will be empty ( ). "batt"
displays along with a continuous beep when the battery voltage drops below the operating level. Meter shuts o in 5 seconds.
Dimensions: 258.3mm(H) x 71.2mm(W) x 42.7mm(D) Weight: Approx. 278g including battery Altitude: 6562 feet (2000m) Overload protection: 600VDC or 600VAC rms unless otherwise
stated in the individual test sections.
Test leads: Use UL listed test leads that comply to UL61010-031
rated CATIII 600V or above. Included test leads are gold-plated and have removeable safety caps.
Functions
Wireless Receiver Mode
Use your SC57 to wirelessly receive a live measurement you have set up at a diffeent location on the jobsite. Like receiving an indoor wet bulb temp. when your at the condensor.
Receiving Wireless Measurements
1. Select RECV switch position on SC57. Hold
SYNC button until meter beeps (>1 sec). Search
pattern initiates.
2. Connect and turn on any Fieldpiece accessory
head to the ET2W wireless transmitter.
3. Select DC switch position on ET2W for all acces-
sory heads except ACH4 (AC switch position).
4. Hold SYNC button on ET2W for one second.
5. The wireless measurement, signal strength,
and battery life of the ET2W will display in the
top-right corner of the SC57 screen. Note: If the ET2W is not connected within 2 min,
the SC57 will beep and stop searching.
Wireless Transmitter Mode
Transmit electrical measurements wirelessly from SC57 to Fieldpiece wireless receivers like EH4W, LT17AW, HG3 or another SC57. Leave the
SC57 behind a closed fan door and send the amp reading to the wireless receiver in your hand.
Sending Wireless Measurements
1. Select any switch position other than RECV.
2. Hold SYNC until meter beeps (>1 sec). Search pattern will initiate.
3. Select RECV switch position on the Fieldpiece wireless receiver. Hold SYNC button for one second.
4. The measurement from the SC57 will be displayed on the Fieldpiece wireless receiver.
Wireless Auto-Connection
When powered on, SC57 will search and connect to the last connected single-link wireless partner device.
Temperature (°F/°C)
Plug any K-type thermocouple directly into the meter to measure temperature. Cold junction is located inside the meter and allows for extremely accurate measurements even in rapidly changing ambient temperatures (going from rooftop to freezer). No adapter is required. See Temp Calibration section for calibration instructions.
Range: -40°F to 2200°F, (-40°C to 1200°C) Resolution: 0.1° Accuracy: ±(1°F) 32°F to 120°F, ±(1°C) 0°C to 49°C
±(1%+2°F) 120°F to 750°F, ±(1%+1°C) 49°C to 400°C ±(2%+6°F) -40°F to 32°F, ±(2%+3°C) -40°C to 0°C ±(2%+6°F) 750°F to 2200°F, ±(2%+3°C) 400°C to 1200°C
Sensor type: K-type thermocouple Overload protection: 30 VDC or 30VAC rms
Voltage AC (VAC) (50Hz-400Hz)
Test power lines (120, 220, 480), test 24V going to controls, and test for transformer failure.
Ranges: 0 to 600V Resolution: 0.1V Accuracy: ±(1.0% + 5 dgts) 50-100Hz
±(6%+5 dgts) 100-400Hz
True RMS: Yes Crest factor: ≤ 3 Input impedance: 1MΩ
Please operate the instrument following all instructions
of the operator's manual to avoid impairing the safety of the product.
SC57
AUTO OFFTRUE RMS
CAT III
600V
TEMP
K-TYPE
30V
MAX
SC57
O F F
WIRELESS ENABLED
SEL
HOLD
HOLD MAX MIN
SYNC
SYNC
MAX MIN
O F F
SC57
SC57
O F F
WIRELESS ENABLED
SEL
HOLD
HOLD
MAX
MIN
SYNC
SYNC
MAX
MIN
CAT.III
300V
400A
CLAMP
ACH4
AC Current Clamp
1AAC / 1mVAC 400AAC MAX
!
ET2W
LO BATT
SEND
RECEIVE
ON
AC DC
Wireless Transmitter
SYNC
ET2W
LO BATT
SEND
RECEIVE
ON
AC DC
Wireless Transmitter
SYNC
ET2W
LO BATT
SEND
RECEIVE
ON
AC DC
Wireless Transmitter
SYNC
AUTO OFF
ARH4
ET2W w/ ARH4
transmitting wirelessly
to SC57
SC57
AUTO OFFTRUE RMS
CAT III 600V
TEMP
K-TYPE
30V
MAX
SC57
O F F
WIRELESS ENABLED
SEL
HOLD
HOLD MAX
MIN
SYNC
SYNC
MAX MIN
AUTO OFFTRUE RMS
CAT III
600V
TEMP
K-TYPE
30V
MAX
Unplug Leads and Slide TEMP Switch to the Right
O F F
O F F
HOLD
HOLD MAX
MIN
SYNC
DC
SEL
HOLD
INRUSH
O F F
NCV
VAC
VDC
AAC
Hz
SEL
HOLD
INRUSH
HOLD
HOLD
MAX
MIN
11
16
12
17
13
18
14
19
15
20
MicroAmps DC (µADC)
Microamps for ame rectier diode test on a heater control. Connect leads between ame sensor probe and control module and turn heating unit on to read µA measurement. When the ame is on, there should be a measurable µADC signal, typically under 10µADC. Compare measurement to manufacture’s specication to determine if replacement is necessary.
Ranges: 1000µA Resolution: 0.1µA Accuracy: ±(1.0% + 5 dgts) Voltage burden: 1V Overload Protection: 600VDC or 600VAC rms
Frequency (Hz) Through Leads
Check variable frequency drives. Check incoming voltages to ensure they are cycling at 60Hz or desired frequency. Select VAC/Hz and press SEL button.
Ranges: 20 to 400Hz Resolution: 0.1Hz Accuracy: ±(0.5% + 5 dgts) Sensitivity: 5V rms on VAC range.
Capacitance (MFD)
Capacitors are one of the most failure prone components in a HVAC/R system. Set to MFD/Ω/
SEL
HOLD
INRUSH
to test motor start and run capacitors. Discharge the capacitor and disconnect from power and resistors between terminals before testing.
Ranges: 0 to 1000µF Resolution: 0.1µF Accuracy: ± (5% + 15 dgts) Overload Protection: 600VDC or 600VAC rms
Voltage DC (VDC)
Select VDC to measure DC voltage.
Ranges: 0 to 600V Accuracy: ±(1.0% + 5 dgts) on 600V range Resolution: 0.1V Input impedance: 1MΩ
Resistance (Ω)
Used for “ohming out” a motor. 0.1Ω resolution is necessary to test the resistance between the motor poles because the values are typically very low. Select MFD/Ω/
SEL
HOLD
INRUSH
and press SEL button once.
Ranges: 0 to 1000Ω, 1000Ω to 9999Ω Resolution: 0.1Ω Overload Protection: 600VDC/VAC rms Accuracy: ±(1.5% + 5 dgts)
Continuity (
SEL
HOLD
INRUSH
)
Use the continuity feature to test if a circuit is open or closed. Use this feature to check fuses as well. A steady “beep” and green LED indicate the circuit is good. Select MFD/Ω/
SEL
HOLD
INRUSH
and press SEL button
twice.
Range: 1000Ω Resolution: 0.1Ω Response time: 100ms Audible beep: <30Ω Overload Protection: 600VDC/VACrms
Non Contact Voltage (NCV)
Use the non contact voltage (NCV) feature to test if a wire is hot or not. Always test on a known live source before using. A red LED blinks and beeping sound is emitted at >24VAC.
AC Voltage Detection Range: 24VAC to 600VAC (50-60Hz)
Amps AC (AAC) TrueRMS (50-60Hz)
Test any isolated power line. Select AAC/Hz dial position.
Ranges: 0 to 400AAC Resolution: 0.1A Crest factor: ≤ 3 Accuracy: ±(2.0% + 5 dgts) 50-60Hz Jaw Opening: 1.2in (30 mm) Overload Protection: 400AAC
Frequency (Hz) Through Clamp
Measure frequency without using test leads, just use the clamp.
Turn dial to AAC/Hz and press SEL. Clamp Hz will be displayed.
Range: 20Hz to 400Hz Accuracy: ±(0.5% + 5) Minimum current range: > 5AAC at 20 to 100Hz, >10AAC at 100
to 400Hz Resolution: 0.1Hz Overload Protection: 400AAC
BatteryReplacement
The battery must be replaced when the battery icon is empty. "bAtt" will display with a beeping sound. Meter will shut o in 5 seconds. Disconnect and unplug leads, turn meter o and remove battery cover.
Safety Information
• Never ground yourself when taking electrical
measurements. Do not touch exposed metal pipes, outlets, xtures, etc., which might be at ground potential, while taking measure­ments. Keep your body isolated from ground by using dry clothing, rubber shoes, rubber mats, or any approved insulating material.
• Disconnect the test leads before opening the
case. Inspect the test leads for damage to the insulation or exposed wire. Replace if suspect. Keep your ngers behind the nger guards on the probes while taking measurements.
• When disconnecting from a circuit, disconnect
the “RED” lead rst, then the common “BLACK” lead. Use one handed testing when possible. Work with others.
• Turn o power to the circuit under test before
cutting, unsoldering, or breaking the circuit.
• Do not measure resistance (ohms) when cir­cuit is powered. Isolate load by disconnecting from circuit.
• Disconnect the meter from the circuit before
turning any inductor o, including motors, transformers, and solenoids. High voltage transients can damage the meter beyond
repair. Do not use during electrical storms.
• Do not apply more than rated voltages be­tween input and ground.
• Isolate capacitors from system and discharge
them safely before testing. All voltage tests: All voltage ranges will
withstand up to 600V. Do not apply more than 600VDC or AC rms.
Symbols used: Caution, risk of electric shock Caution, refer to manual. Ground Double insulation
Maintenance
Clean the exterior with a dry cloth. Do not use
liquid.
Temp. Calibration
For accuracies of ±1°F, calibrate to a known temperature. A glass of stabilized ice water is very close to 32°F (0°C) and is usually very convenient but any known temperature can be used.
1. Select the °F/ °C range.
2. Plug thermocouple to be
calibrated into the K-type jack.
3. Unscrew A and B and remove
the battery cover.
4. Stabilize a large cup of ice
water. Stir the ice with the
water until temperature stays
at 32°F (0°C).
5. Immerse the thermocouple
probe and let it stabilize. Keep
stirring to prevent micro-
environments.
6. Use a small screwdriver to
adjust calibration pot C to the
right of the battery as close to
32°F as you would like.
FCC Compliance and Advisory Statement
This device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
This equipment has been tested and found to comply with the limits for a Class B digital device, according to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment o and on, the user is encouraged to try correct the interference by one or more of the following measures:
1.Reorient the receiving antenna.
2.Increase the separation between the equipment and receiver.
3.Connect the equipment into an outlet on a circuit dierent from
that to which the receiver is connected.
4.Consult the dealer or an experienced radio/TV technician for help. Shielded interface cables must be used in order to comply with
emission limits. FCC Caution: To assure continued compliance, any changes or
modications not expressly approved by the party responsible
for compliance could void the user's authority to operate this
equipment.
Limited Warranty
This meter is warranted against defects in material or workmanship for one year from date of purchase from an authorized Fieldpiece dealer. Fieldpiece will replace or repair the defective unit, at its option, subject to verication of the defect.
This warranty does not apply to defects resulting from abuse, neglect, accident, unauthorized repair, alteration, or unreasonable use of the instrument.
Any implied warranties arising from the sale of a Fieldpiece product, including but not limited to implied warranties of merchantability and tness for a particular purpose, are limited to the above. Fieldpiece shall not be liable for loss of use of the instrument or other incidental or consequential damages, expenses, or economic loss, or for any claim of such damage, expenses, or economic loss.
State laws vary. The above limitations or exclusions may not apply to you.
Obtaining Service
Email Fieldpiece warranty department at fpwarranty@eldpiece. com for current xed price repair service. Send check or money order made out to Fieldpiece Instruments for the amount quoted. If your meter is under warranty there will be no cost for the repair/ replacement. Send your meter, freight prepaid, to Fieldpiece Instruments. Send proof of date and location of purchase for in-warranty service. The meter will be repaired or replaced, at the option of Fieldpiece, and returned via least cost transportation.
For international customers, warranty for products purchased outside of the U.S. should be handled through local distributors. Visit
our website to nd your local distributor.
www.eldpiece.com
© Fieldpiece Instruments, Inc 2014; v12
A
B
C
!
WARNINGS
DISCONNECT AND UNPLUG TEST LEADS before opening case. TEST NCV FUNCTION ON KNOWN LIVE WIRE before using. DO NOT APPLY VOLTAGE greater than 30VAC or 60VDC to the
thermocouple or the jacks when the rotary dial is on °F. (Use only k-type thermocouples)
DO NOT APPLY VOLTAGE TO THE JACKS when the rotary dial is on
microamps. Even low voltages can cause a current overload and potentially harm the meter.
!
O F F
O F F
O
F F
Discharge Cap First!
O F F
O F F
O F F
SELRNG
O F F
SELRNG
Loading...