and press SEL button
twice.
Range: 1000Ω Resolution: 0.1Ω Response time: 100ms
Audible beep: <30Ω Overload Protection: 600VDC/VACrms
Non Contact Voltage (NCV)
Use the non contact voltage (NCV) feature to test if a wire is hot or
not. Always test on a known live source before using. A red LED blinks
and beeping sound is emitted at >24VAC.
AC Voltage Detection Range: 24VAC to 600VAC (50-60Hz)
Amps AC (AAC) TrueRMS (50-60Hz)
Test any isolated power line. Select AAC/Hz dial position.
Ranges: 0 to 400AAC Resolution: 0.1A Crest factor: ≤ 3
Accuracy: ±(2.0% + 5 dgts) 50-60Hz
Jaw Opening: 1.2in (30 mm) Overload Protection: 400AAC
Frequency (Hz) Through Clamp
Measure frequency without using test leads, just use the clamp.
Turn dial to AAC/Hz and press SEL. Clamp Hz will be displayed.
Range: 20Hz to 400Hz Accuracy: ±(0.5% + 5)
Minimum current range: > 5AAC at 20 to 100Hz, >10AAC at 100
to 400Hz Resolution: 0.1Hz Overload Protection: 400AAC
BatteryReplacement
The battery must be replaced when the
battery icon is empty. "bAtt" will display with a
beeping sound. Meter will shut o in 5 seconds.
Disconnect and unplug leads, turn meter o
and remove battery cover.
Safety Information
• Never ground yourself when taking electrical
measurements. Do not touch exposed metal
pipes, outlets, xtures, etc., which might be
at ground potential, while taking measurements. Keep your body isolated from ground
by using dry clothing, rubber shoes, rubber
mats, or any approved insulating material.
• Disconnect the test leads before opening the
case. Inspect the test leads for damage to the
insulation or exposed wire. Replace if suspect.
Keep your ngers behind the nger guards
on the probes while taking measurements.
• When disconnecting from a circuit, disconnect
the “RED” lead rst, then the common “BLACK”
lead. Use one handed testing when possible.
Work with others.
• Turn o power to the circuit under test before
cutting, unsoldering, or breaking the circuit.
• Do not measure resistance (ohms) when circuit is powered. Isolate load by disconnecting
from circuit.
• Disconnect the meter from the circuit before
turning any inductor o, including motors,
transformers, and solenoids. High voltage
transients can damage the meter beyond
repair. Do not use during electrical storms.
• Do not apply more than rated voltages between input and ground.
• Isolate capacitors from system and discharge
them safely before testing.
All voltage tests: All voltage ranges will
withstand up to 600V. Do not apply more than
600VDC or AC rms.
Symbols used:
Caution, risk of electric shock
Caution, refer to manual.
Ground
Double insulation
Maintenance
Clean the exterior with a dry cloth. Do not use
liquid.
Temp. Calibration
For accuracies of ±1°F,
calibrate to a known
temperature. A glass
of stabilized ice water
is very close to 32°F
(0°C) and is usually very
convenient but any
known temperature can
be used.
1. Select the °F/ °C range.
2. Plug thermocouple to be
calibrated into the K-type jack.
3. Unscrew A and B and remove
the battery cover.
4. Stabilize a large cup of ice
water. Stir the ice with the
water until temperature stays
at 32°F (0°C).
5. Immerse the thermocouple
probe and let it stabilize. Keep
stirring to prevent micro-
environments.
6. Use a small screwdriver to
adjust calibration pot C to the
right of the battery as close to
32°F as you would like.
FCC Compliance and Advisory
Statement
This device complies with Part 15 of the FCC rules. Operation is
subject to the following two conditions: (1) this device may not cause
harmful interference, and (2) this device must accept any interference
received, including interference that may cause undesired operation.
This equipment has been tested and found to comply with the
limits for a Class B digital device, according to Part 15 of the FCC
rules. These limits are designed to provide reasonable protection
against harmful interference in a residential installation. This
equipment generates, uses and can radiate radio frequency energy
and if not installed and used in accordance with the instructions,
may cause harmful interference to radio communications. However,
there is no guarantee that interference will not occur in a particular
installation. If this equipment does cause harmful interference to
radio or television reception, which can be determined by turning
the equipment o and on, the user is encouraged to try correct the
interference by one or more of the following measures:
1.Reorient the receiving antenna.
2.Increase the separation between the equipment and receiver.
3.Connect the equipment into an outlet on a circuit dierent from
that to which the receiver is connected.
4.Consult the dealer or an experienced radio/TV technician for help.
Shielded interface cables must be used in order to comply with
emission limits.
FCC Caution: To assure continued compliance, any changes or
modications not expressly approved by the party responsible
for compliance could void the user's authority to operate this
equipment.
Limited Warranty
This meter is warranted against defects in material or
workmanship for one year from date of purchase from an authorized
Fieldpiece dealer. Fieldpiece will replace or repair the defective unit,
at its option, subject to verication of the defect.
This warranty does not apply to defects resulting from abuse,
neglect, accident, unauthorized repair, alteration, or unreasonable
use of the instrument.
Any implied warranties arising from the sale of a Fieldpiece
product, including but not limited to implied warranties of
merchantability and tness for a particular purpose, are limited
to the above. Fieldpiece shall not be liable for loss of use of the
instrument or other incidental or consequential damages, expenses,
or economic loss, or for any claim of such damage, expenses, or
economic loss.
State laws vary. The above limitations or exclusions may not
apply to you.
Obtaining Service
Email Fieldpiece warranty department at fpwarranty@eldpiece.
com for current xed price repair service. Send check or money
order made out to Fieldpiece Instruments for the amount quoted.
If your meter is under warranty there will be no cost for the repair/
replacement. Send your meter, freight prepaid, to Fieldpiece
Instruments. Send proof of date and location of purchase for
in-warranty service. The meter will be repaired or replaced, at the
option of Fieldpiece, and returned via least cost transportation.
For international customers, warranty for products purchased
outside of the U.S. should be handled through local distributors. Visit
our website to nd your local distributor.
www.eldpiece.com
© Fieldpiece Instruments, Inc 2014; v12