Fieldpiece Fieldpiece ADMN2 Manual

Specifications
Pressure Ports: Two tube connectors for 4mm
(3/16 inch) I.D. flexible tubing
Units of Measure: inWC and mbar Accuracy:
±0.02inWC, ±(0.00inWC to 2.00inWC); ±1.5% FS, ±(2.00inWC to 60.0inWC); ±0.05mbar, ±(0.00mbar to 5.00mbar); ±1.5% FS, ±(5.00mbar to 150.0mbar);
Resolution: 0.1 or 0.01 depending on resolution
selected (If 0.01 resolution is selected you must divide displayed value by 10 for actual value).
Measurement Range:
inWC: -60.0 to 60.0 mBar: -150.0 to 150.0
Operating environment:
32ºF (0°C) to 122ºF (50°C), under 75%RH
Compatible media: Dry, non-corrosive gases Battery: Single standard 9-volt battery, NEDA
1604, JIS 006P, IEC 6F22.
Battery Life: 100 hours with low battery LED
indicator on accessory head.
Overrange: “OL” or “-OL” is displayed. Auto power off: 15 minutes Dimensions: 131mm(~5 3/16”)(H) x 67mm(~2
7/16”)(W) x 32mm(~1 1/4”)(D)
Weight: approximately 154g (~1/2 pound)
including battery
Resolution Modes (0.1 and 0.01)
There are two modes of resolution on the ADMN2. For most applications, 0.1 mode will be fine to use. When measuring low pressures, 0.01 will likely be the best mode for meaningful testing. The most common circumstance for 0.01 mode is measuring static pressure across a blower.
Only when using 0.01 mode, the display of your meter will show the true measurement X 10. So if the true measurement is 0.08, your meter will show
0.8. This is done to optimize resolution.
Warranty
The product is warranted to the original purchas­er against defects in material or workmanship for a period of one (1) year from the date of purchase. During the warranty period, Fieldpiece Instruments will, at its option, replace or repair the defective unit.
This warranty does not apply to defects resulting from abuse, neglect, accident, unauthorized repair, alteration, or unreasonable use of the instrument. Any implied warranty arising out of the sale of Fieldpiece's products including but not limited to implied warranties of merchantability and fitness for purpose are limited to the above. Fieldpiece shall not be liable for incidental or consequential dam­ages.
Service
Return any defective ADMN2 to Fieldpiece for warranty service along with proof of purchase. Contact Fieldpiece for out of warranty repair charges.
Dual-
P
ManomMae
Acces
ces
sory
Head
0.1
E
ME
0.01
(LCD/10)
Resolu
tion
tion
Dual Port Differential Manometer Accessory Head Model: ADMN2
OPERATOR’S MANUAL
Description
The ADMN2 is a portable dual port manometer accessory head. The ADMN2 is capable of taking gas pressure as well as static pressure. The ADMN2 also has a zero adjust. The auto-power-off function conserves battery life, but can be disabled for data logging if desired.
The ADMN2 will allow you to take gas pressure between -60inWC and 60inWC. The ADMN2 can measure P1 or P2 pressure readings or the differ­ence between P1 and P2. The ADMN2 also has two different measurement scales: inWC (inches of water column) and mBar. Static pressure is possible with the resolution of 0.01inWC or 0.01mBar.
Operation
1. Connect to COM and Volts jacks using any
Fieldpiece test leads with removable probe tips.
For Fieldpiece “stick-style” meters, EHDL1, or
DL3, slide the head directly over the meter.
2. Select the mVDC range.
3. Zero the ADMN2 by pressing the ZERO button
just before taking the pressure readings, while
at ambient pressure. This will zero both P1 and
P2.
4. Connect a single hose to get the gauge pres-
sure relative to the ambient or ZERO pressure.
5. Connect both hoses if you want to see relative
pressure, P1 minus P2.
6. Select P1, P2, or P1 - P2.
7. Select English (inches of water column) or Metric (mBar) units.
8. To disable the auto-power-off, remove the rub­ber covering and flip the switch below. Auto­power-off is helpful for conserving the battery; but if you are data logging with Fieldpiece model DL3, you will want to disable this feature.
9. If you are in an environment where the tempera­ture is noticeably changing while you are taking your reading, disconnect the meter from the hoses and ZERO it relative to ambient pressure before each reading.
Checking Gas Pressure on a Regulator
1. Screw the brass fitting into the pressure port on the regulator.
2. Put unit into operation (i.e. turn on the furnace and have furnace ignite, as if running it in nor­mal operation.)
3. This will give you the pressure coming out of the regulator.
4. If you suspect high or low inlet pressure into the regulator, the manometer can hook into the inlet port in the same manner it can connect into the outlet port. If you have a dual-port manometer, you can check both the inlet and the outlet simultaneously and see the pressure drop across the regulator.
5. See manufacturer's specification for the specifi-
cations on target inlet and outlet pressure for a given regulator or piece of combustion equip­ment.
Field Pressure Calibration
Remove the rubber covering the ZERO button. By pressing the ZERO button, both the P1 and P2 are zeroed to the ambient pressure. For this reason the calibration should be done when both P1 and P2 are disconnected from the hoses in order to properly calibrate to ambient pressure.
Connect Your Way
EHDL1 LT series DL3 HS series
Gas pressure measurement
(0.1 resolution mode)
Static pressure across a blower
(0.01 resolution mode)
0.8mVDC / 10 = 0.08inWC
OFF
2000
ADMN2
Dual-Port Manometer Accessory Head
P1
P2
P1-P2
ON
0.1 ENGLISH
METRIC
0.01
(LCD/10)
Lo Batt
Resolution
ZERO
AUTO
OFF
P1 P2
ADMN2
Dual-Port Manometer Accessory Head
P1
P2
P1-P2
ON
0.1 ENGLISH
METRIC
0.01
(LCD/10)
Lo Batt
Resolution
ZERO
AUTO
OFF
P1 P2
ADMN2
Dual-Port Manometer Accessory Head
P1
P2
P1-P2
ON
0.1 ENGLISH
METRIC
0.01
(LCD/10)
Lo Batt
Resolution
ZERO
AUTO
OFF
P1 P2
ADMN2
Dual-Port Manometer Accessory Head
P1
P2
P1-P2
ON
0.1 ENGLISH
METRIC
0.01
(LCD/10)
Lo Batt
Resolution
ZERO
AUTO
OFF
P1 P2
v13
P1 P2
Dual-Port Manometer Accessory Head
Resolution
(LCD/10)
P1-P2
P2 P1
ADMN2
ZERO
ON
Lo Batt
0.1
ENGLISH
0.01
METRIC
AUTO
OFF
0.1
0.01
P1 P2
(LCD/10)
Dual-Port
ON
Manometer Accessory
Lo Batt
Head
0.1 ENGLISH
Resolution
0.01 METRIC
(LCD/10)
P1-P2
P2 P1
AUTO
ZERO
OFF
3.0 inWC
ADMN2
To Meter
3.0 mVDC
0.1
P1 P2
Dual-Port
ON
0.01
(LCD/10)
Manometer Accessory Head
Resolution
P1-P2
To Me ter
(LCD/10)
P2 P1
ADMN2
Lo Batt
0.1
ENGLISH
0.01
METRIC
AUTO
ZERO
OFF
0.8 mVDC
0.05 inWC
-0.03 inWC
99 Washington Street Melrose, MA 02176
Phone 781-665-1400 Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
RMA316 Manometer Adapter
Used for 5/16" pressure outlet ports.
1. Shut off main gas supply to furnace.
2. Move the gas valve switch to "OFF" position.
3. Use a 3/32" hex wrench to loosen the outlet pressure tap screw. Rotate screw counter­clockwise one revolution to open.
4. Connect the ADMN2 tubing to the smaller (3/16") end of the adapter and the other (5/16") end of the adapter into the of the adapter tube.
5. Slide the 5/16" adapter tube over the outlet pressure boss (port) to seal. Overlap the pres­sure boss by at least 3/8" to prevent leakage.
Outlet Pressure
Tap Screw
Outlet
Pressure
Boss
Inlet Pressure Tap Screw
Smaller end (3/16”) of the adapter connects to ADMN2 tubing. Remove the installed brass fitting from tube so the adapter can be inserted.
Inlet Pressure Boss
Adapter
(5/16” to 3/16”)
Adapter Tube
(5/16”)
More Products From Fieldpiece
Modular Expandability
Modular expandability is ability for accessory
heads and meters to change configurations to match the various needs of an HVAC/R technician.
Accessory heads (the sensors) send out a mV
signal, which represents the value of the measure­ment, to whatever meter is attached to it. Heads can attach directly to the top of a Stick meter, DL3 data logger, or EHDL1. They can also plug into any meter with mV ranges using ASLS2 leads.
Stick Meter
This is the heart of modular expandability. In addi­tion to being a full functioning multimeter, any acces­sory head can be used with it.
Model HS36
Non contact voltage
Magnetic hanger
Autoranging
Backlight
Temperature
Volts, amps, ohms
Frequency
Microfarads
Includes:
HS36 Meter
ACH4 Current Clamp ATB1 K-type TCouple ADLS2 Deluxe Leads
ANC1 Case
Accessory Heads
Accessory heads are the sensors of multiple parameters measured by technicians every day. They plug into a mV range (depending on the head) of a multimeter. The multimeter will display whatev­er the head is measuring. Instead of having to pur­chase and carry a separate instrument for each parameter, a technician can use multiple heads and a single multimeter to do the job.
Here are four of the many heads available:
AAV3 Air Velocity and Temperature
ARH4 Digital Psychrometer
AVG2 Digital Vacuum Gauge
AOX2 Combustion Check
Air Velocity & Temperature Head
English
READLOBATT
ON
LCDX 100
Average
(16sec)
Metric
Real time
AAV3
AUTO­OFF
English Metric
Ft/min M/s
KM/hrMPH ºF ºC
OFF
aaaaa
aaaaa
aaaaa
aaaaa
aaaaa
aaaaa
ASP2 Static Pressure Probe
For accurate static pressure measurements.
1. Connect ASP2 to ADMN2 via hose.
2. Insert ASP2 into drilled or pre-existing 1/4” hole.
3. Use allignment arrow on the ASP2 to direct probe into the air stream.
4. Read static pressure measurements from “stick­meter”, EHDL1 or record with DL3.
5. Remove probe and patch hole.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
CLAMP
300V
CAT.III
400A
AC Current Clamp
!
1AAC / 1mVAC 400AAC MAX
ON
H
STABLE
LOBATT
ENGLIS
SH
t
ng
r ation
METRIC
R134A
SC
Refrige
Superhea
and Subcooli
R404A
Temp
SH orSC
Pressu re
T/C
l
Ca
Set
M
AT
ON
Lo Batt
BLE
STA
Combustion
Check
ASX24
GAS
OIL
Type
Fuel
AUTO
OFF
T/C
2
CAL
2
O
CO
°F
%
TEMP
%
°C
AOX2
ACH4
O- F
AUT
OF
AirVelocity
&Temperature
Head
Eng
lish
Eng
Ft/min M/s
lish
Metri
c
Met
ºF ºC
ric
KM/hrMPH
OF
Real
F
time
AUT OFF
O-
AAV3
ON
LCD 100
X
READLOBATT
Aver
(16sec)
age
AUTO OFF
ARH4
Combustion Check
ON
Lo Batt
STABLE
GAS
Fuel Type
OIL
TEMP
%CO
2
%O
2
AUTO
T/C
°C °F
CAL
OFF
AOX2
Loading...