Fieldpiece
Accessory Handle
Model EHDL1
OPERATOR'S MANUAL
Description
The model EHDL1 converts any
Fieldpiece accessory head into a stand alone
meter. Plug the EHDL1 into the bottom of
any Fieldpiece accessory head, select the
appropriate range, and read the display.
WARNING
!
Use only with Fieldpiece accessory
heads. Accessory heads deliver low voltage
(under 10V) to the EHDL1. Voltages higher
than 30VDC or 24VAC can damage the
EHDL1.
DO NOT CONNECT EHDL1 TO LINE
VOLTAGE!
How to use
1.Connect any of the Fieldpiece
accessory heads to the EDL1 by sliding it
onto the top of the handle.
2. Turn on and select either 200mV range
or 2000mV to best suit your measurement.
3. Select AC for current clamp or DC for
the rest of the accessory heads.
4. Follow any specific instructions to the
paticular head you are using and measure as
usual.
5. Press hold to hold current
measurement.
How to use MAX/MIN button
1. Press MAX/MIN to begin recording
MAX and MIN. MAX will be displayed.
2. Press MAX/MIN to alternate between
displaying the current MAX and MIN.
3. Press MAX/MIN and hold for 1 second
to erase MAX and MIN and display current
reading.
EHDL1
with ACM3
EHDL1
with ACH
EHDL1 with
AHDL1 and ACH
Specifications:
Display: 3 1/2 digit liquid crystal (LCD) with
a maximum read of 1999
Polarity: Automatic, positive implied,
negative polarity indication
Overrange: (OL) or (-OL) is displayed
Low battery indication: The "" is displayed
when the battery voltage drops below the
operating level
Measurement rate: 2.5 times per second,
nominal
Operating enviroment: 32oF to 122oF at
<70% relative humidity
Storage temperature: -4oF to 140oF, 0 to
80% R.H. with battery removed from
meter
Accuracy: Stated accuracy at 73oF ±9oF,
<75%relative humidity
Temperature Coefficient:
0.06 x (specified accuracy) per oF. (32oF
to 64oF, 82oF to 122oF)
Power: Single standard 9-volt battery
Battery Life: 200 hours typical with carbon-
zinc
Dimensions:
140mm (H) x 59.2mm (W) x 33mm (D)
Weight: Approx. 125g including battery
Saftey: Designed to meet CE-EMC
DC Volts:
Ranges: 200mV, 2000mV
Resolution: 0.1mV
Accuracy: ±(0.5%rdg + 1dgt)
Input impedance: 10MW
Overload protection: 30 VDC or 24VAC
AC Volts:
Ranges: 200mV, 2000mV
Resolution: O.1mV
Accuracy: ±(1.2%rdg + 3dgts) 50~500Hz
Input impedance: 10MW
Overload protection: 30 VDC or 24VAC
MAX/MIN Hold:
Operating on all ranges
Press 1 sec to exit
DATA Hold:
Operating on all ranges
Warranty
The EHDL1 is warranted to the original
purchaser against defects in material or
workmanship for a period of one year from
the date of purchase. During the warranty
period, Fieldpiece Instruments will, at its
option, replace or repair the defective unit.
This warranty does not apply to defects
resulting from abuse, neglect, accident,
unauthorized repair, alteration, or
unreasonable use of the instrument. Any
implied warranty arising out of the sale of
Fieldpiece's products including but not
limited to implied warranties of
merchantibility, and fitness for purpose, are
limited to the above.Fieldpiece shall not be
liable for incidental or consequential
damages.
State laws vary, so the above limitations
or exclusions may not apply to you. This
warranty gives you specific legal rights, and
you may also have other rights which vary
from state to state.
ServiceService
Service
ServiceService
There are no user serviceable parts
inside. Send any defective EHDL1 to
Fieldpiece Instruments for warranty service
along with proof of purchase.
Warranty service will typically be made
within 24 hours after receipt of defective unit.
Any non-working out-of-warranty product
may be sent to Fieldpiece Instruments for
service. Send the EHDL1 along with a check
or money order for $15 to Fieldpiece
Instruments. Your repaired or replaced
EHDL1 will be returned to you with a 6 month
warranty.
Fieldpiece Instruments, Inc.
580 West Central Ave.
Brea, CA 92821
www.fieldpiece.com
Phone: (714) 257-9060
Fax: (714) 257-9069
OPMANEHDL1V20NA