10
POSITIONING CALIBRATION
#5: Positioning calibration
Calibration is a process during which a device learns the position of
the limit switches and a motor characteristic.
Calibration is mandatory in order for the device to correctly recognize
a roller blind position.
The procedure consists of an automatic, full movement between the
limit switches (up, down, and up again).
Description:
– switch connected to the S1 terminal
– switch connected to the S2 terminal
Calibration using connected momentary switch
1. Make sure:
• the device is supplied,
• momentary switch is connected to S1 and S2 terminals,
• the device added to the Z-Wave network,
• parameter 151 is set to 1 or 2.
2. Press and hold or switch for at least 3 seconds
3. Press and hold the same switch again for at least 3 seconds.
4. Press and hold the same switch again for at least 3 seconds.
5. The device will perform the calibration process, completing
full cycle – up, down and up again.
6. Test whether the positioning works correctly.
Standard roller blind positioning
Calibration using the B-button
1. Make sure:
• the device is supplied,
• the device added to the Z-Wave network,
• parameter 151 is set to 1 or 2.
2. Press and hold the B-button.
3. Wait for the LED to glow white.
4. Quickly release and click the B-button again.
5. The device will perform the calibration process, completing
full cycle – up, down and up again
6. Test whether the positioning works correctly.
NOTE
If you notice the calibration process fail
(e.g. it does not start,
movement times are
really short or highly inaccurate), adjust
parameter 155 value
(e.g. reduce it by half).